A Linear and High-Sensitivity Microwave Biosensor on a FR-4 Substrate for Aqueous Glucose Monitoring Using a Concentric Square-Shaped Split-Ring Resonator

,
,
,
and
1
Electronics and Micro-Electronic Laboratory (LEµE), Bd de L’environnement, Monastir 5000, Tunisia
2
RF2S Spectrum Solutions, 18 Street of the Faïencerie, 33300 Bordeaux, France
3
University Grenoble-Alpes, Grenoble-INP/LCIS, 26000 Valence, France
4
Higher Institute of Applied Sciences and Technology of Sousse, University of Sousse, Street Taher Ben Achour, Sousse 4003, Tunisia
This article belongs to the Section Biosensors

Abstract

Non-invasive glucose monitoring remains a significant challenge in diabetes management, with existing approaches often limited by poor accuracy, high cost, or patient discomfort. Microwave-based biosensors offer a promising label-free alternative by exploiting the dielectric contrast between glucose and water. This paper presents a compact, dual-band concentric square-shaped split-ring resonator (SRR-type) biosensor fabricated on a low-cost FR-4 substrate for aqueous glucose detection. The sensor leverages electric field confinement in inter-ring gaps to transduce glucose-induced permittivity changes into measurable shifts in resonance frequency and reflection coefficient. Experimental results demonstrate a linear, monotonic response across the clinical range up to 250 mg/dL, with a frequency-domain sensitivity of 1.964 MHz/(mg/dL) and amplitude-domain sensitivity of 0.0332 dB/(mg/dL), achieving high coefficients of determination (R2 = 0.9956 and 0.9927, respectively). The design achieves a normalized size of 0.137 λg2, combining high sensitivity and compact size within a scalable platform. Operating in the UWB-adjacent band (2.76–3.25 GHz), the proposed biosensor provides a practical, reproducible, and PCB-compatible solution for next-generation label-free glucose monitoring.

1. Introduction

Diabetes mellitus (DM) is a chronic metabolic disorder characterized by persistent hyperglycemia resulting from defects in insulin secretion, insulin action, or both. It represents a major and growing global health challenge: the number of adults living with diabetes rose from 422 million in 2014 to 537 million in 2021 and is projected to reach 643 million by 2030 and 783 million by 2045 [1]. DM encompasses several subtypes with distinct etiologies and clinical trajectories. Type 1 diabetes (T1D) arises from autoimmune destruction of pancreatic β-cells, leading to absolute insulin deficiency, typically diagnosed in childhood or adolescence. Type 2 diabetes (T2D), the most prevalent form, stems from a combination of insulin resistance and progressive β-cell dysfunction and is increasingly observed in younger populations due to rising obesity and sedentary lifestyles. Additional forms include gestational diabetes mellitus (GDM), monogenic diabetes syndromes, cystic fibrosis–related diabetes (CFRD), and drug- or chemical-induced diabetes [1,2]. This heterogeneity underscores the need for personalized, accurate, and frequent glucose monitoring to guide therapeutic decisions.
Sustained hyperglycemia significantly elevates the risk of microvascular (e.g., retinopathy, nephropathy, neuropathy) and macrovascular (e.g., cardiovascular disease) complications [3,4]. Consequently, tight glycemic control, enabled by reliable blood glucose level (BGL) monitoring, is critical to mitigating long-term morbidity [5]. Conventional self-monitoring of blood glucose (SMBG) relies on invasive finger-prick sampling coupled with electrochemical test strips, a method widely considered as the clinical gold standard [6,7,8,9]. However, its invasiveness, patient discomfort, recurring consumable costs, and risk of infection often result in poor adherence, particularly in pediatric and elderly populations. These limitations have spurred intense research into non-invasive, painless, and continuous glucose monitoring alternatives [10,11,12,13,14,15,16,17,18,19].
Non-invasive approaches have explored glucose detection in alternative biofluids such as saliva [20,21], urine [22], sweat [23,24,25], and tears [26,27]. Yet, these media often exhibit weak, delayed, or inconsistent correlation with blood glucose concentrations, limiting their clinical reliability. Other strategies, including reverse iontophoresis [28,29], bio-impedance spectroscopy [30], and optical techniques such as near-infrared (NIR) [31,32], Raman [33,34], contact lenses [35,36,37] and photoacoustic spectroscopy [38,39,40,41], have demonstrated potential but face persistent challenges related to low signal-to-noise ratio, environmental interference (e.g., temperature, hydration), high system cost, and limited portability. Transdermal methods that extract interstitial fluid via reverse iontophoresis followed by electrochemical detection further suffer from skin irritation, sensor drift, and the need for external current injection [19,42].
In this context, microwave- and radiofrequency (RF)-based biosensing has emerged as a promising label-free alternative for non-invasive glucose monitoring [43,44,45,46,47]. As shown in Figure 1, the approach exploits the fact that glucose molecules in interstitial fluid or aqueous media form hydrogen bonds with water, reducing the solution’s polarity and dipole mobility. This leads to a measurable, monotonic decrease in the complex permittivity, specifically the real part (εr) and loss tangent (tanδ), in the 0.5–10 GHz range [48]. Even small changes in glucose concentration thus perturb the electromagnetic (EM) response of a nearby resonant structure, enabling detection without chemical reagents. Critically, the clinical reliability of RF- and microwave-based glucose estimates has been validated using Clarke Error Grid analysis, with recent studies reporting over 95% of predicted values falling within Zone A (clinically accurate) and Zone B (benign errors), confirming their suitability as a label-free sensing platform for future non-invasive diabetes management [44].
Figure 1. Conceptual overview of the non-invasive glucose monitoring process (label-free sensor).
RF and microwave techniques offer deeper tissue penetration, minimal scattering, and immunity to optical interference, enabling robust examination of dielectric properties in biological domain. These methods detect shifts in resonant frequency, amplitude, or phase of planar structures, such as antennas [49], split-ring resonators [50,51], or transmission lines [52,53], in response to dielectric changes induced by glucose. Key advantages include low fabrication cost, compatibility with standard printed circuit board (PCB) processes, potential for miniaturization, and seamless integration with wireless readout systems [54,55].
A wide variety of microwave resonator topologies have been explored for glucose sensing, including patch resonators [56], microstrip bandpass filters [57], complementary split-ring resonators (CSRRs) [58,59,60,61,62,63,64], coplanar waveguide (CPW)-loaded resonators [65,66], interdigital capacitors (IDCs) [67], cavity resonators [68], and meandered-line structures [69]. While these designs demonstrate feasibility, many suffer from limited sensitivity, large dimension, or complex fabrication. Notably, planar resonators with concentric geometries [70], which leverage strong field confinement in inter-ring gaps, remain underexplored despite their potential for enhanced dielectric sensitivity and compact integration.
The Federal Communications Commission’s (FCC) 2002 allocation of the 3.1–10.6 GHz band for unlicensed ultra-wideband (UWB) communications has further accelerated interest in compact, low-power biosensors operating in this spectrum. UWB-compatible architectures offer broad operational bandwidth, high temporal resolution, and immunity to multipath interference, attributes well-suited for real-time, continuous monitoring in biological environments [71,72]. Within this framework, planar resonators featuring concentric square metallic rings present a compelling design: their sharp corners and inter-ring gaps concentrate electric fields, enhancing interaction with the analyte and improving sensitivity to localized dielectric changes induced by glucose in adjacent aqueous samples.
The purpose of this study is to investigate the feasibility of a planar concentric square-ring microwave resonator for label-free glucose detection in aqueous solutions. The work focuses on characterizing the electromagnetic response of the proposed sensor to glucose-induced variations in complex permittivity across the physiologically relevant concentration range (up to 250 mg/dL). By integrating full-wave electromagnetic simulation with experimental validation, this study aims to elucidate the underlying sensing mechanism and evaluate the performance of this simple, PCB-compatible topology as a viable platform for microwave-based biosensing.
The remainder of this paper is organized as follows: Section 2 presents the biosensor design and electromagnetic analysis, including the geometry, equivalent circuit model, and simulation results. Section 3 details the preparation of glucose aqueous solutions and their dielectric characterization across the 1–10 GHz band. Section 4 describes the sensor’s response to glucose concentrations, including simulated and measured S11 trends. Section 5 validates the design through fabrication and experimental measurements in air and with glucose solutions. Section 6 analyzes the regression models, quantifies sensitivity, and compares performance against state-of-the-art sensors. Finally, Section 7 concludes the paper with a summary and future perspectives.

2. Biosensor Design and Electromagnetic Analysis

2.1. Sensor Geometry and Fabrication

Building upon the electromagnetic sensing principles outlined in the Introduction, this work presents the design and analysis of a compact planar microwave biosensor based on a multi-resonant concentric square-ring architecture, engineered for high sensitivity to dielectric perturbations induced by glucose concentration variations in aqueous solutions.
The proposed structure consists of four concentric square metallic rings patterned on the top layer of an FR-4 epoxy-glass substrate (εr = 4.3, tanδ = 0.025, thickness = 1.6 mm). An asymmetric slot is incorporated into each ring to break geometric symmetry and enable resonant excitation under linear polarization. Although conceptually inspired by resonant ring structures, the implemented design functions as a coupled SRR array, as the resonant elements are conductive traces on the signal layer rather than apertures in the ground plane.
A single 50 Ω microstrip feedline is positioned adjacent to the outermost ring, providing capacitive coupling to the resonator array. The bottom layer of the substrate is fully metallized to form a continuous ground plane, ensuring stable electromagnetic confinement and minimizing radiation losses. While full two-port S-parameter simulations (S11 and S21) were performed for comprehensive characterization, the sensor operates primarily in reflection mode (S11) in practical use, with the feedline exciting the resonators via near-field coupling.
To achieve multi-band operation within the 2–7 GHz range, the rings are designed with progressively decreasing side lengths (9 mm, 7 mm, 5 mm, and 3 mm) and non-uniform inter-ring spacing. The asymmetric slots, 0.94 mm wide for the outer three rings and 0.4 mm for the innermost ring, further lift modal degeneracy, induce mode splitting, and broaden the effective bandwidth. This deliberate geometric asymmetry promotes mode hybridization and spectral separation of resonances, yielding multiple electromagnetic signatures that respond differentially to changes in the complex permittivity of the overlying glucose solution.
The sensor layout and layer configuration are illustrated in Figure 2. Figure 2a shows the top-layer geometric configuration, including the concentric rings, asymmetric slots, and microstrip feedline. Figure 2b displays the cross-sectional side view, depicting the substrate stack: top and bottom copper layers (each 35 μm thick), FR-4 dielectric substrate, and solid ground plane.
Figure 2. Geometry and layer stack-up of the proposed concentric square-ring microwave biosensor: (a) top view and (b) cross-sectional side view.
All critical geometric parameters, including ring side lengths, slot widths, inter-ring gaps, and feedline dimensions, are summarized in Table 1.
Table 1. Geometric parameters of the concentric square-ring biosensor.

2.2. Equivalent Circuit Model

The concentric square-ring structure introduces an inductive effect through the metallic arms, while the asymmetric gaps between adjacent ring segments exhibit a shunt capacitive effect with the surrounding dielectric medium. The capacitive coupling between neighboring rings creates a mutual impedance network that lowers the fundamental resonant frequency and enhances design compactness. The resulting equivalent LC resonance governs the sensor’s electromagnetic response. In this configuration, each square metallic ring functions as an inductor, while the gaps between ring segments and the fringing fields between adjacent rings act as distributed capacitors. The conductor width is W = 0.4 mm, and the slot gaps serve as the primary capacitive regions.
The effective inductance Leff of each ring is determined by its side length a, conductor width W, and geometry. For a square loop, the inductance is expressed as [73]:
L e f f = μ 0 a π l n 2 a W + 0.384 + W 4 a
where a takes values of 9 mm, 7 mm, 5 mm, and 3 mm for rings 1 through 4, respectively, and μ0 is the permeability of free space.
The effective capacitance Ceff associated with each slot is modeled by accounting for both parallel-plate and fringing-field contributions [73,74]:
C e f f = ε r ε 0 A g 1 + g π W l n 4 W g
where A is the effective slot area, g is the gap width, εr = 4.3, and ε0 is the permittivity of free space.
As illustrated in Figure 3, the equivalent circuit model comprises the following elements: L1 and C1 represent the feedline coupling and edge fringing with the host medium; the mutual coupling between rings is captured by two parallel LC branches (L2, L4, C2) and (L3, L5, C3); and the grounding effect through the substrate is modeled by L6 and C4. This lumped-element representation accurately reflects the distributed electromagnetic behavior of the four-ring concentric square-split resonator and explains the observed frequency shift in response to glucose-induced changes in the dielectric environment.
Figure 3. Lumped-element equivalent circuit model of the proposed biosensor.
To explicitly map the equivalent circuit elements to the physical structure, the following correspondence is established:
  • L1 and C1 represent the feedline coupling and fringing capacitance to the outermost ring (Ring 1, 9 mm).
  • L2 and C2 model the inductance and gap capacitance of the second ring (Ring 2, 7 mm), with L4 capturing its mutual coupling to Ring 1.
  • L3 and C3 correspond to the third ring (Ring 3, 5 mm), with L5 representing its mutual coupling to the innermost ring (Ring 4, 3 mm).
  • L6 and C4 describe the grounding effect of Ring 4 through the FR-4 substrate, capturing its interaction with the bottom ground plane.

2.3. Electromagnetic Simulation and Resonance Analysis

Full-wave electromagnetic simulations were performed using CST Studio Suite® (Microwave Studio®) with open (radiation) boundary conditions and adaptive mesh refinement to ensure numerical convergence. A systematic parametric study was conducted to trace the evolution of resonant behavior as rings were incrementally added. Figure 4a illustrates the sequential design stages, from a single ring to the complete four-ring configuration, while Figure 4b shows the corresponding simulated reflection coefficients (S11).
Figure 4. (a) Sequential design stages of the concentric square-shaped SRR and (b) corresponding simulated reflection coefficient (S11) responses.
The resonance progression is summarized as follows. A single ring exhibits a fundamental resonance at 6.74 GHz. Adding a second ring introduces a lower-frequency mode at 3.67 GHz, attributed to increased effective inductance and inter-ring capacitance. Incorporating a third ring generates a hybridized resonance at 3.44 GHz, while the lower mode shifts to 3.25 GHz due to enhanced mutual coupling. The complete four-ring configuration stabilizes into two dominant resonances at 3.27 GHz and 5.16 GHz, with a deep return loss of −19.65 dB at the lower band, indicating strong field confinement and a high quality factor (Q).
This dual-band response enables redundant and complementary sensing channels within a compact slot area of 9 × 9 mm2, enhancing detection reliability and enabling discrimination of subtle dielectric changes induced by varying glucose concentrations. The electric field is highly concentrated in the inter-ring gaps and asymmetric slot regions, precisely where the aqueous sample is deposited, maximizing interaction with the analyte and forming the foundation for high-sensitivity, label-free glucose detection.
To further elucidate the physical mechanism behind the sensor’s high sensitivity, we present in Figure 5 the simulated electric field (E-field) magnitude distribution at the primary resonance frequency (3.27 GHz). The E-field is strongly confined within the inter-ring gaps and asymmetric slots, with peak magnitudes exceeding 65,000 V/m. This intense field localization, particularly in the innermost gap region, maximizes energy coupling to the overlying aqueous sample, thereby amplifying the dielectric perturbation effect. This field concentration directly explains the high sensitivity and identifies the optimal sensing area as the region defined by the innermost ring and its slot.
Figure 5. Simulated electric field (E-field) magnitude distribution at 3.27 GHz.

3. Preparation and Dielectric Characterization of Glucose Solutions

Concentrations of aqueous glucose solutions at 20, 60, 100, 140, 200, and 250 mg/dL were prepared independently using analytical-grade D-glucose Dextrose (≥99.5%, Bulk®, Mirków, Poland) and deionized water (resistivity ≥ 18.2 MΩ cm) to avoid cumulative dilution errors. Solutions were characterized using a Keysight N1501A dielectric probe connected to an Agilent N5222A vector network analyzer (VNA). Measurements were performed over 1–10 GHz, aligned with the sensor’s operating band, and preceded by standard open/short/water calibration. All dielectric measurements were conducted in a temperature-controlled ambient test room at 25 ± 1 °C. Each solution was allowed to equilibrate for sufficient time after preparation to ensure thermal homogeneity before measurement. Four trials per concentration ensured statistical reliability. The complex permittivity εr∗ = εr′ − jεr′′ was extracted, with εr′′ = σ/(ωε0).
As shown in Figure 6, εr′ decreases monotonically with both frequency and glucose concentration, while tanδ = εr′′/εr′ increases with frequency but varies minimally with glucose.
Figure 6. Mean dielectric properties of glucose solutions versus frequency for various concentrations: (a) relative permittivity (ε′) and (b) loss tangent (tanδ).
This confirms that glucose reduces solution polarizability (due to its lower dipole mobility compared to water) [75], while losses remain dominated by water relaxation. Standard deviations are low (≈10−3 for εr′, ≈10−5 for tanδ; see Figure 7), confirming high reproducibility.
Figure 7. Standard deviation of dielectric properties across repeated measurements for various glucose concentrations: (a) relative permittivity (ε′) and (b) loss tangent (tanδ).

4. Biosensor Response and Glucose Detection

Full-wave simulations evaluated the sensor’s response to glucose solutions using a polyethylene container (εr = 2.2, inner diameter = 19 mm) placed over the resonator array. The complex permittivity of each solution at 3.25 GHz (from Section 2.3) was assigned to the liquid domain.
Figure 8 shows the simulated reflection coefficient (S11) across 2–7 GHz. Two resonances appear near 3 GHz and 5 GHz. As glucose concentration increases, the primary resonance at 2.754 GHz shifts upward by 482 MHz and deepens from −15.2 dB to <−22.2 dB, indicating enhanced impedance matching and stronger field-analyte interaction. The monotonic, linear trend confirms that both frequency shift and S11 amplitude serve as robust indicators for glucose quantification. These dual-parameter responses validate the sensor’s potential for label-free, contactless monitoring.
Figure 8. Simulated S11 response versus frequency for different glucose concentrations.

5. Experimental Validation

5.1. Sensor Fabrication and Characterization in Air

The measurement setup for the dual-band concentric square-shaped SRR sensor is shown in Figure 9a. To validate the electromagnetic design, the reflection coefficient (S11) of the unloaded sensor was measured in air using an Agilent N5222A VNA. Prior to data acquisition, the VNA’s Ports were calibrated using the standard short-open-load (SOL) calibration procedure to ensure measurement accuracy.
Figure 9. (a) Measurement test-bench setup, (b) comparison of simulated and measured reflection coefficient (S11) of the biosensor in air.
As shown in Figure 9b, the measured response exhibits two distinct resonances at 3.32 GHz and 5.59 GHz, in close agreement with simulated results (Figure 4b). Minor discrepancies in resonance frequency (<2.5%) and depth are attributed to manufacturing tolerances (e.g., trace width variations), substrate inhomogeneity, calibration uncertainties, and connector repeatability, factors commonly observed in planar microwave biosensor validation. This strong correlation confirms the accuracy of the electromagnetic model and establishes a reliable baseline for liquid-loading experiments.

5.2. Sensor Response to Glucose Solutions

The sensor’s response to aqueous glucose solutions was evaluated by placing the polyethylene container filled with 5 mL of solution directly above the resonator array. Measurements were performed across the same concentration range used in simulations up to 250 mg/dL. All S11 measurements were performed using a single fabricated sensor, with five independent readings per glucose concentration. This experiment is established under an ambient temperature and controlled environment (25 ± 1 °C), with a sufficient equilibration period for each solution to ensure thermal stability and reproducible dielectric properties.
Figure 10 shows the measured S11 response across all glucose concentrations, while simulated results are shown in Figure 8. Both datasets exhibit a consistent, monotonic trend: as glucose concentration increases, the primary resonance (near 2.76 GHz) shifts to higher frequencies and the return-loss minimum deepens (i.e., S11 becomes more negative). These changes confirm that the dielectric properties of the overlying medium directly modulate the sensor’s electromagnetic behavior.
Figure 10. Measured S11 response versus frequency for different glucose concentrations.
The frequency upshift arises from the reduced effective permittivity of the glucose-water mixture. Glucose molecules possess a lower dipole moment and polarizability than water. Thus increasing their concentration lowers the solution’s complex permittivity, raising the resonant frequency.
Critically, the equivalent model of the proposed sensor (see Figure 3) explains the glucose-induced frequency upshift. Indeed, when an aqueous glucose solution is placed over the sensor, the reduced relative permittivity (εr′) of the medium decreases the effective capacitance (primarily C2 and C3, which are most exposed to the analyte). From the resonance condition f 0 = 1 / 2 π L e f f C e f f , a reduction in Ceff leads to an increase in f0, precisely the upward shift observed experimentally. This mechanistic link between the equivalent circuit and the sensing phenomenon is thus clearly articulated.
Concurrently, the deeper resonance (from −13.4 dB to −21.7 dB experimentally) indicates improved impedance matching and enhanced electromagnetic energy transfer into the lossy medium.
While qualitative agreement is excellent, minor quantitative differences exist. Measured resonances are slightly shifted (~10–15 MHz) relative to simulation, likely due to unmodeled fringing fields, finite sample volume, and substrate moisture absorption. Experimental curves also appear smoother and exhibit greater loss, consistent with additional dissipative mechanisms, such as conductor roughness, FR-4 dielectric losses, and connector mismatches, not fully captured in the idealized simulation.
Nevertheless, the monotonicity, reproducibility, and directional response across the full physiological range validate the sensor’s operating principle. These results demonstrate the feasibility of planar microwave resonators for label-free, glucose detection based on dielectric contrast, paving the way for integration into wearable or point-of-care monitoring platforms.
To further evaluate the sensor’s response, the transmission coefficient (S21) across the same glucose concentrations was measured. As shown in Figure 11, S21 exhibits a monotonic upward frequency shift from 2.85 GHz to 2.9 GHz, with a corresponding deepening of the transmission minimum from −12.5 dB to −15.5 dB. While S21 confirms the same trend as S11, its frequency shifts (~50 MHz) and resonance depth (−15.5 dB) are much smaller than those observed in S11 (491 MHz, −21.7 dB). This indicates that S11 provides a stronger, more robust signature of the dielectric change, likely due to stronger field confinement at the feedline and resonator interface. For this proposed design, S11 is thus the preferred metric for glucose detection.
Figure 11. Measured S21 response versus frequency for different glucose concentrations.

6. Regression Analysis, Sensitivity, and Performance Comparison

6.1. Regression Modeling and Sensitivity Analysis

Figure 12a presents the resonance frequency of the primary mode as a function of glucose concentration, while Figure 12b shows the corresponding reflection coefficient (S11). Both simulated (dotted line) and experimental (dashed line) datasets cover the physiological range up to 250 mg/dL.
Figure 12. Simulated and measured (a) resonance frequency and (b) reflection coefficient as functions of glucose concentration, with nonlinear regression fits.
All reported S11 responses are based on 5 independent measurements per concentration, with a standard deviation of less than 86 MHz for resonance frequency and less than 1 dB for |S11| magnitude, confirming high reproducibility as illustrated in Figure 13 and summarized in Table 2.
Figure 13. Measured error bars (±1 SD) represent variability across 5 independent measurements: (a) resonance frequency and (b) reflection coefficient.
Table 2. Mean and standard deviation for resonance frequency (Fr) and reflection coefficient (S11).
The resonance frequency increases from 2.761 GHz (DI water) to 3.252 GHz (250 mg/dL), while the S11 magnitude decreases from −13.4 dB to −21.7 dB, reflecting enhanced impedance matching with rising glucose concentration.
Although the response appears nearly linear, a more accurate empirical description is achieved using nonlinear regression models that capture subtle deviations from linearity [50,60,61,62,63,64,74,75]. For the resonance frequency, a two-phase exponential association model provides an excellent fit (R2 = 0.9956):
F r ( m e a s ) = y 0 + Y 1 1 − e − x / ρ 1 + Y 2 1 − e − x / ρ 2
while Fr is the predicted resonance frequency in GHz, y0 is the baseline frequency offset, Y1, Y2 are amplitude coefficients, ρ1, ρ2 are concentration constants in mg/dL governing the rate of response.
The fitted parameters are summarized in Table 3.
Table 3. Two-phase exponential association parameters for resonance frequency regression.
The dominance of the fast concentration constant (ρ1 ≪ ρ2) indicates an initial rapid frequency shift at low concentrations, followed by a slower saturation trend, behavior consistent with glucose-induced dielectric saturation in aqueous environment [60,61,62].
For the reflection coefficient, a third-order polynomial yields the best fit (R2 = 0.9927):
S 11 ( m e a s ) = a 3 x 3 + a 2 x 2 + a 1 x + a 0
with coefficients listed in Table 4.
Table 4. Polynomial coefficients for reflection coefficient (S11) regression.
The high coefficients of determination (R2 > 0.99) confirm that both models accurately describe the sensor’s response. The minor offset (~15.4 MHz) between simulated and measured baselines is attributed to real-world factors such as substrate losses, humidity, and alignment tolerances, effects absent in idealized simulations.
To contextualize our results against clinical standards, we compare our sensor’s output with reference values via Clarke Error Grid analysis. As shown in Figure 14, 85% of data points fall within Zone A (clinically accurate), while 15% lie in Zone B (benign errors), well within the safety margins for diabetes management [44]. This demonstrates that our label-free approach achieves accuracy comparable to invasive methods, without requiring consumables or blood extraction.
Figure 14. Clarke Error Grid analysis of predicted vs. reference glucose concentrations.
The sensitivity of a biosensor quantifies its ability to transduce minute changes in analyte concentration into measurable electromagnetic responses. For the proposed glucose biosensor, two complementary sensitivity metrics are defined: the frequency-domain sensitivity SFr = ΔFr/ΔGc and the amplitude-domain sensitivity SS11 = Δ|S11|/ΔGc, where ΔFr and Δ|S11| represent the shift in resonant frequency and reflection coefficient magnitude, respectively, in response to a glucose concentration variation ΔGc.
Over the physiological range up to 250 mg/dL, the sensor achieves a frequency sensitivity of 1.964 MHz/(mg/dL), enabling resolution of clinically relevant glucose fluctuations. The concurrent deepening of the reflection coefficient (from −13.4 dB to −21.7 dB) yields a robust amplitude-domain sensitivity, providing an independent validation channel that enhances resilience to environmental and instrumental drift.
For cross-sensor comparison, the normalized frequency sensitivity is also evaluated as [76,77]:
S n = 1 f 0 Δ F r Δ G c
where f0 = 2.76 GHz is the baseline resonance frequency (measured with the empty dielectric container in place). This yields Sn = 0.071 (mg/dL)−1, a figure of merit that accounts for operating frequency and facilitates fair comparison across resonant biosensors.
Together, these metrics confirm that the concentric square-ring topology effectively couples glucose-induced dielectric changes to measurable RF signatures, establishing a reliable foundation for label-free, glucose monitoring.

6.2. Comparative Performance Analysis

Table 5 presents a detailed performance comparison between the proposed concentric square SRR biosensor and state-of-the-art microwave glucose sensors.
Table 5. Performance comparison of the proposed biosensor with state-of-the art microwave glucose sensors (SRR-type).
The proposed sensor achieves a frequency-domain sensitivity of 1.964 MHz/(mg/dL) and an amplitude-domain sensitivity of 0.0332 dB/(mg/dL) over the clinically relevant range up to 250 mg/dL, with high linearity (R2 = 0.9956 for frequency, R2 = 0.9927 for S11). This represents the highest sensitivity among all FR-4-based in vitro microwave glucose sensors, surpassing the 1.73 MHz/(mg/dL) reported by Harnsoongnoen et al. [62] (on DiClad880, 1800 mm2) by 13% and the 0.438 MHz/(mg/dL) of Saleh et al. [50] (on FR-4) by 4.5×, while also outperforming other FR-4-based design [63]. The sensor’s compact size of 25 × 15 mm2 (375 mm2), 4.8× smaller than [62] and 24× smaller than [78], is further validated by its normalized size of 0.137 λg2, which is 3.2× smaller than [62] (0.438 λg2) and 6× smaller than [63] (0.820 λg2), demonstrating superior miniaturization relative to operating scale. Critically, this high performance is achieved using standard FR-4 substrate (εr = 4.3), avoiding the need for costly materials like RO4350 [60] or RT5880 [78,79], thus enabling low-cost, PCB-compatible mass production without compromising sensitivity.
Unlike most referenced works that rely on a single sensing parameter (e.g., frequency-only in [61] or amplitude-only in [78]), our design provides dual-parameter readout (frequency and S11), offering cross-validation and enhanced robustness against environmental drift. Furthermore, all measurements were conducted in a controlled in vitro aqueous environment (0–250 mg/dL), ensuring reproducibility and direct correlation with dielectric theory, unlike in vivo studies [78] that suffer from biological variability. The normalized sensitivity (Sn = 7.11 × 10−4 (mg/dL)−1) further confirms competitive performance when accounting for operating frequency.
In summary, the proposed biosensor delivers a good trade-off between sensitivity, miniaturization, cost, linearity, and robustness, a combination unmatched by any existing FR-4-based microwave glucose sensor.
Although the present study focuses on in vitro measurements in aqueous glucose solutions, the proposed sensing concept can be extended toward more complex biological environments. Future work will focus on testing with interstitial fluid simulants, integration with compact RF readout circuits, and validation under real-world conditions, including temperature variations and ionic interference. We also plan to incorporate on-chip temperature sensing and real-time compensation algorithms to ensure robustness in uncontrolled thermal environments. Additionally, long-term stability and reusability will be assessed through biofouling resistance tests, drift analysis over days to weeks, and encapsulation with biocompatible membranes (e.g., PDMS or Nafion) to enable reliable operation in complex biological settings.

7. Conclusions

This work demonstrates a high-performance, planar microwave biosensor based on a concentric square-shaped split-ring resonator (SRR) architecture for label-free glucose detection in aqueous solutions. Fabricated on a low-cost FR-4 substrate, the sensor exploits dielectric perturbations induced by glucose to produce measurable shifts in both resonance frequency and reflection coefficient magnitude. Experimental validation confirms a linear and highly reproducible response over the physiological range up to 250 mg/dL, with a frequency sensitivity of 1.964 MHz/(mg/dL) and amplitude sensitivity of 0.0332 dB/(mg/dL).
The dual-parameter readout enhances reliability, while the compact size (25 × 15 mm2), or 0.137 λg2) and PCB compatibility ensure low-cost manufacturability and potential for wearable integration. Unlike in vivo studies that are influenced by biological variability, this work establishes performance in a controlled aqueous environment, enabling direct correlation with dielectric theory and facilitating sensor calibration. The high linearity (R2 > 0.99) and contactless operation further support clinical translation.
The proposed sensor operates in the 2.76–3.25 GHz band, adjacent to the FCC-allocated UWB spectrum (3.1–10.6 GHz), which is designated for low-power medical and communication devices. Simulations under typical EMI conditions (e.g., 2.4 GHz Wi-Fi, 5 GHz LTE) show frequency shifts of <1 MHz, less than 0.2% of the total glucose-induced response (491 MHz), confirming high immunity to real-world RF noise. Furthermore, the compact size (25 × 15 mm2), PCB-compatible fabrication, and reflection mode S11 readout enable seamless integration with low-power wireless modules (e.g., BLE or UWB transceivers) for real-time data transmission to smartphones or cloud-based analytics platforms, paving the way for wearable, connected glucose monitoring systems.
In summary, the proposed SRR-type biosensor strikes an optimal balance between sensitivity, size, cost, and robustness, marking a significant step toward practical, accessible, and a promising label-free biosensor for future pain-free, glucose monitoring for people with diabetes.

Author Contributions

Conceptualization, all authors; methodology, all authors; software, K.J.; validation, all authors; formal analysis, all authors; investigation, all authors; resources, D.K., P.L.-A. and H.G.; data curation, all authors; writing—original draft preparation, K.J. and S.S.; writing—review and editing, S.S., D.K., P.L.-A. and H.G.; visualization, all authors.; supervision, S.S., D.K., P.L.-A. and H.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.

Conflicts of Interest

Author Sehmi Saad was employed by the company RF2S Spectrum Solutions. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Abbreviations

The following abbreviations are used in this manuscript:
ABPSArtificial Blood Plasma Solution
ASRRAsymmetric Split Ring Resonator
BGLBlood Glucose Level
BLEBluetooth Low Energy
CFRDCystic Fibrosis–Related Diabetes
CPWCoplanar Waveguide
CSRRComplementary Split-Ring Resonators
CSTComputer Simulation Technology
DCCDouble Concentric Circular
DI WaterDeionized Water
DMDiabetes Mellitus
EMElectromagnetic
EMIElectromagnetic Interference
FCCFederal Communications Commission
FR-4Flame Retardant 4
GDMGestational Diabetes Mellitus
IDCInterdigital Capacitors
ISMIndustrial-Scientific-Medical
LTELong Term Evolution
MUTMaterial Under Test
N/ANot Available
NIRNear-Infrared
PDMSPolydimethylsiloxane
PCBPrinted Circuit Board
RFRadio Frequency
SDStandard Deviation
SMBGSelf-Monitoring of Blood Glucose
SOLShort-Open-Load
SRRSplit-Ring Resonator
T1DType 1 Diabetes
T2DType 2 Diabetes
UWBUltra-Wide Band
VNAVector Network Analyzer
Wi-FiWireless-Fidelity

References

  1. Magliano, D.J.; Boyko, E.J. IDF Diabetes Atlas, 10th ed.; International Diabetes Federation: Brussels, Belgium, 2021. [Google Scholar] [PubMed]
  2. Scully, K.J.; Sherwood, J.S.; Martin, K.; Ruazol, M.; Marchetti, P.; Larkin, M.; Zheng, H.; Sawicki, G.S.; Uluer, A.; Neuringer, I.; et al. Continuous glucose monitoring and HbA1c in cystic fibrosis: Clinical correlations and implications for CFRD diagnosis. J. Clin. Endocrinol. Metab. 2022, 107, e1444–e1454. [Google Scholar] [CrossRef] [Scilit]
  3. Fowler, M.J. Microvascular and macrovascular complications of diabetes. Clin. Diabetes 2008, 26, 77–82. [Google Scholar] [CrossRef] [Scilit]
  4. Zakir, M.; Ahuja, N.; Surksha, M.A.; Sachdev, R.; Kalariya, Y.; Nasir, M.; Kashif, M.; Shahzeen, F.; Tayyab, A.; Khan, M.S.M.; et al. Cardiovascular complications of diabetes: From microvascular to macrovascular pathways. Cureus 2023, 15, e45835. [Google Scholar] [CrossRef] [Scilit]
  5. Šimat, M.; Janković Makek, M.; Mičetić, M. Treatment of Diabetes Mellitus by Acupuncture: Dynamics of Blood Glucose Level and Its Mathematical Modelling. Sci 2023, 5, 38. [Google Scholar] [CrossRef] [Scilit]
  6. Clarke, W.L.; Cox, D.; Gonder-Frederick, L.A.; Carter, W.; Pohl, S.L. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987, 10, 622–628. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Montagnana, M.; Caputo, M.; Giavarina, D.; Lippi, G. Overview on self-monitoring of blood glucose. Clin. Chim. Acta 2009, 402, 7–13. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Clar, C.; Barnard, K.D.; Cummins, E.; Royle, P.; Waugh, N. Self-monitoring of blood glucose in type 2 diabetes: Systematic review. Health Technol. Assess. 2010, 14, 1–140. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Nichols, S.P.; Koh, A.; Storm, W.L.; Shin, J.H.; Schoenfisch, M.H. Biocompatible materials for continuous glucose monitoring devices. Chem. Rev. 2013, 113, 2528–2549. [Google Scholar] [CrossRef] [Scilit]
  10. Villena Gonzales, W.; Mobashsher, A.T.; Abbosh, A. The progress of glucose monitoring—A review of invasive to minimally and non-invasive techniques, devices and sensors. Sensors 2019, 19, 800. [Google Scholar] [CrossRef] [Scilit]
  11. Shokrekhodaei, M.; Quinones, S. Review of non-invasive glucose sensing techniques: Optical, electrical and breath acetone. Sensors 2020, 20, 1251. [Google Scholar] [CrossRef] [Scilit]
  12. Tang, L.; Chang, S.J.; Chen, C.J.; Liu, J.T. Non-invasive blood glucose monitoring technology: A review. Sensors 2020, 20, 6925. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Jing, Y.; Chang, S.J.; Chen, C.J.; Liu, J.T. Glucose monitoring sensors: History, principle, and challenges. J. Electrochem. Soc. 2022, 169, 057514. [Google Scholar] [CrossRef] [Scilit]
  14. Xue, Y.; Thalmayer, A.S.; Zeising, S.; Fischer, G.; Lübke, M. Commercial and scientific solutions for blood glucose monitoring—A review. Sensors 2022, 22, 425. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Reddy, V.S.; Agarwal, B.; Ye, Z.; Zhang, C.; Roy, K.; Chinnappan, A.; Narayan, R.J.; Ramakrishna, S.; Ghosh, R. Recent advancement in biofluid-based glucose sensors using invasive, minimally invasive, and non-invasive technologies: A review. Nanomaterials 2022, 12, 1082. [Google Scholar] [CrossRef] [Scilit]
  16. Hina, A.; Saadeh, W. Noninvasive blood glucose monitoring systems using near-infrared technology—A review. Sensors 2022, 22, 4855. [Google Scholar] [CrossRef] [Scilit]
  17. Laha, S.; Rajput, A.; Laha, S.S.; Jadhav, R. A concise and systematic review on non-invasive glucose monitoring for potential diabetes management. Biosensors 2022, 12, 965. [Google Scholar] [CrossRef] [Scilit]
  18. Di Filippo, D.; Sunstrum, F.N.; Khan, J.U.; Welsh, A.W. Non-invasive glucose sensing technologies and products: A comprehensive review for researchers and clinicians. Sensors 2023, 23, 9130. [Google Scholar] [CrossRef] [Scilit]
  19. Wang, M.; Zheng, J.; Zhang, G.; Lu, S.; Zhou, J. Wearable Electrochemical Glucose Sensors for Fluid Monitoring: Advances and Challenges in Non-Invasive and Minimally Invasive Technologies. Biosensors 2025, 15, 309. [Google Scholar] [CrossRef] [Scilit]
  20. Zhang, W.; Du, Y.; Wang, M.L. Noninvasive glucose monitoring using saliva nano-biosensor. Sens. Bio-Sens. Res. 2015, 4, 23–29. [Google Scholar] [CrossRef] [Scilit]
  21. Soni, A.; Jha, S.K. Smartphone based non-invasive salivary glucose biosensor. Anal. Chim. Acta 2017, 996, 54–63. [Google Scholar] [CrossRef] [Scilit]
  22. Prasad, S.N.; Weerathunge, P.; Karim, N.; Anderson, S.; Hashmi, S.; Mariathomas, P.D.; Bansal, V.; Ramanathan, R. Non-invasive detection of glucose in human urine using a color-generating copper NanoZyme. Anal. Bioanal. Chem. 2021, 413, 1279–1291. [Google Scholar] [CrossRef] [Scilit]
  23. Kim, J.; Campbell, A.S.; Wang, J. Wearable non-invasive epidermal glucose sensors: A review. Talanta 2018, 177, 163–170. [Google Scholar] [CrossRef] [Scilit]
  24. Xu, J.; Fang, Y.; Chen, J. Wearable biosensors for non-invasive sweat diagnostics. Biosensors 2021, 11, 245. [Google Scholar] [CrossRef] [Scilit]
  25. Zafar, H.; Channa, A.; Jeoti, V.; Stojanović, G.M. Comprehensive review on wearable sweat-glucose sensors for continuous glucose monitoring. Sensors 2022, 22, 638. [Google Scholar] [CrossRef] [Scilit]
  26. Chu, M.X.; Miyajima, K.; Takahashi, D.; Arakawa, T.; Sano, K.; Sawada, S.-I.; Kudo, H.; Iwasaki, Y.; Akiyoshi, K.; Mochizuki, M.; et al. Soft contact lens biosensor for in situ monitoring of tear glucose as non-invasive blood sugar assessment. Talanta 2011, 83, 960–965. [Google Scholar] [CrossRef] [Scilit]
  27. Sempionatto, J.R.; Brazaca, L.C.; García-Carmona, L.; Bolat, G.; Campbell, A.S.; Martin, A.; Tang, G.; Shah, R.; Mishra, R.K.; Kim, J.; et al. Eyeglasses-based tear biosensing system: Non-invasive detection of alcohol, vitamins and glucose. Biosens. Bioelectron. 2019, 137, 161–170. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Cheng, Y.; Gong, X.; Yang, J.; Zheng, G.; Zheng, Y.; Li, Y.; Xu, Y.; Nie, G.; Xie, X.; Chen, M.; et al. A touch-actuated glucose sensor fully integrated with microneedle array and reverse iontophoresis for diabetes monitoring. Biosens. Bioelectron. 2022, 203, 114026. [Google Scholar] [CrossRef] [Scilit]
  29. Li, X.; Li, T.; Liu, B.; Hu, N.; Hu, T.; Ni, Z. Boosting the performance of an iontophoretic biosensing system with a graphene aerogel and Prussian blue for highly sensitive and noninvasive glucose monitoring. Analyst 2023, 148, 3184–3192. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Gelao, G.; Marani, R.; Carriero, V.; Perri, A.G. Design of a dielectric spectroscopy sensor for continuous and non-invasive blood glucose monitoring. Int. J. Adv. Technol. Eng. Explor. 2012, 3, 55–64. [Google Scholar]
  31. Alsunaidi, B.; Althobaiti, M.; Tamal, M.; Albaker, W.; Al-Naib, I. A review of non-invasive optical systems for continuous blood glucose monitoring. Sensors 2021, 21, 6820. [Google Scholar] [CrossRef] [Scilit]
  32. Davison, N.B.; Gaffney, C.J.; Kerns, J.G.; Zhuang, Q.D. Recent progress and perspectives on non-invasive glucose sensors. Diabetology 2022, 3, 56–71. [Google Scholar] [CrossRef] [Scilit]
  33. Li, N.; Zang, H.; Sun, H.; Jiao, X.; Wang, K.; Liu, T.C.Y.; Meng, Y. A noninvasive accurate measurement of blood glucose levels with Raman spectroscopy of blood in microvessels. Molecules 2019, 24, 1500. [Google Scholar] [CrossRef] [Scilit]
  34. Todaro, B.; Begarani, F.; Sartori, F.; Luin, S. Is Raman the best strategy towards the development of non-invasive continuous glucose monitoring devices for diabetes management? Front. Chem. 2022, 10, 994272. [Google Scholar] [CrossRef] [Scilit]
  35. Lin, Y.-R.; Hung, C.-C.; Chiu, H.-Y.; Chang, P.-H.; Li, B.-R.; Cheng, S.-J.; Yang, J.-W.; Lin, S.-F.; Chen, G.-Y. Noninvasive glucose monitoring with a contact lens and smartphone. Sensors 2018, 18, 3208. [Google Scholar] [CrossRef] [Scilit]
  36. Bamgboje, D.; Christoulakis, I.; Smanis, I.; Chavan, G.; Shah, R.; Malekzadeh, M.; Violaris, I.; Giannakeas, N.; Tsipouras, M.; Kalafatakis, K.; et al. Continuous non-invasive glucose monitoring via contact lenses: Current approaches and future perspectives. Biosensors 2021, 11, 189. [Google Scholar] [CrossRef] [Scilit]
  37. Kazanskiy, N.L.; Khonina, S.N.; Butt, M.A. Smart contact lenses—A step towards non-invasive continuous eye health monitoring. Biosensors 2023, 13, 933. [Google Scholar] [CrossRef] [Scilit]
  38. Tanaka, Y.; Tajima, T.; Seyama, M.; Waki, K. Differential continuous wave photoacoustic spectroscopy for non-invasive glucose monitoring. IEEE Sens. J. 2019, 20, 4453–4458. [Google Scholar] [CrossRef] [Scilit]
  39. Kottmann, J.; Rey, J.M.; Sigrist, M.W. Mid-Infrared photoacoustic detection of glucose in human skin: Towards non-invasive diagnostics. Sensors 2016, 16, 1663. [Google Scholar] [CrossRef] [Scilit]
  40. Kaysir, M.R.; Song, J.; Rassel, S.; Aloraynan, A.; Ban, D. Progress and perspectives of mid-infrared photoacoustic spectroscopy for non-invasive glucose detection. Biosensors 2023, 13, 716. [Google Scholar] [CrossRef] [Scilit]
  41. Kaysir, M.R.; Zaman, T.M.; Rassel, S.; Wang, J.; Ban, D. Photoacoustic resonators for non-invasive blood glucose detection through photoacoustic spectroscopy: A systematic review. Sensors 2024, 24, 6963. [Google Scholar] [CrossRef] [Scilit]
  42. Bandodkar, A.J.; Wang, J. Non-invasive wearable electrochemical sensors: A review. Trends Biotechnol. 2014, 32, 363–371. [Google Scholar] [CrossRef] [Scilit]
  43. Yilmaz, T.; Foster, R.; Hao, Y. Radio-frequency and microwave techniques for non-invasive measurement of blood glucose levels. Diagnostics 2019, 9, 6. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Kandwal, A.; Liu, L.W.; Deen, M.J.; Jasrotia, R.; Kanaujia, B.K.; Nie, Z. Electromagnetic wave sensors for noninvasive blood glucose monitoring: Review and recent developments. IEEE Trans. Instrum. Meas. 2023, 72, 8007215. [Google Scholar] [CrossRef] [Scilit]
  45. Jain, P.; Joshi, A.M.; Mohanty, S.P.; Cenkeramaddi, L.R. Non-Invasive Glucose Measurement Technologies: Recent Advancements and Future Challenges. IEEE Access 2024, 12, 61907–61936. [Google Scholar] [CrossRef] [Scilit]
  46. Kandwal, A.; Sharma, Y.D.; Jasrotia, R.; Kit, C.C.; Lakshmaiya, N.; Sillanpää, M.; Liu, L.W.; Igbe, T.; Kumari, A.; Sharma, R.; et al. A comprehensive review on electromagnetic wave based non-invasive glucose monitoring in microwave frequencies. Heliyon 2024, 10, e37825. [Google Scholar] [CrossRef] [Scilit]
  47. Alsultani, A.B.; Kovács, K.; Chase, J.G.; Benyo, B. Advances in invasive and non-invasive glucose monitoring: A review of microwave-based sensors. Sens. Actuators Rep. 2025, 9, 100332. [Google Scholar] [CrossRef] [Scilit]
  48. Zakaria, A.A.; Allam, A.; AbdelRahman, A.B. Microwave Based Non-Invasive Blood Glucose Sensors: Key Design Parameters and Case-Informed Evaluation. IEEE Access 2025, 13, 154695–154711. [Google Scholar] [CrossRef] [Scilit]
  49. Shi, J.; Fernández-García, R.; Gil, I. Sensor technologies for non-invasive blood glucose monitoring. Sensors 2025, 25, 3591. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  50. Saleh, G.; Ateeq, I.S.; Al-Naib, I. Glucose level sensing using single asymmetric split ring resonator. Sensors 2021, 21, 2945. [Google Scholar] [CrossRef] [Scilit]
  51. Elsheakh, D.N.; Mohamed, E.H.; Eldamak, A.R. Blood glucose monitoring biosensor based on multiband split-ring resonator monopole antenna. Biosensors 2025, 15, 250. [Google Scholar] [CrossRef] [Scilit]
  52. Harnsoongnoen, S.; Wanthong, A. Coplanar waveguide transmission line loaded with electric-LC resonator for determination of glucose concentration sensing. IEEE Sens. J. 2017, 17, 1635–1640. [Google Scholar] [CrossRef] [Scilit]
  53. Wu, W.J.; Shang, L.; Zhao, W.S.; Wang, W. Fluidic Glucose Measurement Based on A Differential Microwave Sensing System with Combination of Multi-Stepped-Impedance Transmission Lines. IEEE Sens. J. 2024, 24, 28805–28817. [Google Scholar] [CrossRef] [Scilit]
  54. Shamkhalichenar, H.; Bueche, C.J.; Choi, J.W. Printed circuit board (PCB) technology for electrochemical sensors and sensing platforms. Biosensors 2020, 10, 159. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  55. Zhao, W.; Tian, S.; Huang, L.; Liu, K.; Dong, L. The review of Lab-on-PCB for biomedical application. Electrophoresis 2020, 41, 1433–1445. [Google Scholar] [CrossRef] [Scilit]
  56. Yilmaz, T.; Foster, R.; Hao, Y. Broadband tissue mimicking phantoms and a patch resonator for evaluating noninvasive monitoring of blood glucose levels. IEEE Trans. Antennas Propag. 2014, 62, 3064–3075. [Google Scholar] [CrossRef] [Scilit]
  57. Baghbani, R.; Rad, M.A.; Pourziad, A. Microwave sensor for non-invasive glucose measurements design and implementation of a novel linear. IET Wirel. Sens. Syst. 2015, 5, 51–57. [Google Scholar] [CrossRef] [Scilit]
  58. Jang, C.; Park, J.K.; Lee, H.J.; Yun, G.H.; Yook, J.G. Non-invasive fluidic glucose detection based on dual microwave complementary split ring resonators with a switching circuit for environmental effect elimination. IEEE Sens. J. 2020, 20, 8520–8527. [Google Scholar] [CrossRef] [Scilit]
  59. Hosseinzadeh, S.; Pakvarjouy, M. Microwave-based single port glucose sensor arranging interdigital and complementary split ring resonator. Measurement 2025, 246, 116695. [Google Scholar] [CrossRef] [Scilit]
  60. Ebrahimi, A.; Scott, J.; Ghorbani, K. Microwave reflective biosensor for glucose level detection in aqueous solutions. Sens. Actuators A Phys. 2020, 301, 111662. [Google Scholar] [CrossRef] [Scilit]
  61. Gao, M.; Qiang, T.; Ma, Y.; Liang, J.; Jiang, Y. RFID-based microwave biosensor for non-contact detection of glucose solution. Biosensors 2021, 11, 480. [Google Scholar] [CrossRef] [Scilit]
  62. Harnsoongnoen, S.; Buranrat, B. Advances in a microwave sensor-type interdigital capacitor with a hexagonal complementary split-ring resonator for glucose level measurement. Chemosensors 2023, 11, 257. [Google Scholar] [CrossRef] [Scilit]
  63. Yang, S.; Gao, S.; Zhuang, Y.; Hu, W.; Zhao, J.; Yi, Z. Non-destructive sensor for glucose solution concentration detection using electromagnetic technology. Micromachines 2024, 15, 758. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  64. Rattan, P.; Houngkamhang, N.; Orankitanun, T.; Phasukkit, P. Metamaterial microwave sensor for glucose level measurement based on strip line with complementary split ring resonator. Phys. Status Solidi A Appl. Mater. Sci. 2025, 222, 2400180. [Google Scholar] [CrossRef] [Scilit]
  65. Yi, Z.; Wang, C. Noninvasive glucose sensors using defective-ground-structure coplanar waveguide. IEEE Sens. J. 2022, 23, 195–201. [Google Scholar] [CrossRef] [Scilit]
  66. Sameer, M.; Agarwal, P. Coplanar waveguide microwave sensor for label-free real-time glucose detection. Radioengineering 2019, 28, 491. [Google Scholar] [CrossRef] [Scilit]
  67. Khan, M.R.R.; Khalilian, A.; Kang, S.W. Fast, highly-sensitive, and wide-dynamic-range interdigitated capacitor glucose biosensor using solvatochromic dye-containing sensing membrane. Sensors 2016, 16, 265. [Google Scholar] [CrossRef] [Scilit]
  68. Wang, S.; Wan, C.; Zheng, Y. Microwave characterization of glucose level using composite PMS-HMSIW cavity resonator integrated with microfluidic channel. IEEE Trans. Instrum. Meas. 2024, 73, 6009116. [Google Scholar] [CrossRef] [Scilit]
  69. Ma, Y.; Qiang, T.; Gao, M.; Liang, J.; Jiang, Y. Quantitative, temperature-calibrated and real-time glucose biosensor based on symmetrical-meandering-type resistor and intertwined capacitor structure. Biosensors 2021, 11, 484. [Google Scholar] [CrossRef] [Scilit]
  70. Juan, C.G.; Potelon, B.; Quendo, C.; Bronchalo, E. Microwave planar resonant solutions for glucose concentration sensing: A systematic review. Appl. Sci. 2021, 11, 7018. [Google Scholar] [CrossRef] [Scilit]
  71. Mahmud, M.Z.; Islam, M.T.; Misran, N.; Almutairi, A.F.; Cho, M. Ultra-wideband (UWB) antenna sensor based microwave breast imaging: A review. Sensors 2018, 18, 2951. [Google Scholar] [CrossRef] [Scilit]
  72. Tiberi, G.; Ghavami, M. Ultra-wideband (UWB) systems in biomedical sensing. Sensors 2022, 22, 4403. [Google Scholar] [CrossRef] [Scilit]
  73. Bahl, I.J.; Bhartia, P. Microwave Solid State Circuit Design; Wiley-Interscience: Hoboken, NJ, USA, 2003. [Google Scholar]
  74. Almutairi, A.F.; Islam, M.S.; Samsuzzaman, M.; Islam, M.T.; Misran, N.; Islam, M.T. A complementary split ring resonator based metamaterial with effective medium ratio for C-band microwave applications. Results Phys. 2019, 15, 102675. [Google Scholar] [CrossRef] [Scilit]
  75. Turgul, V.; Kale, I. Characterization of the complex permittivity of glucose/water solutions for noninvasive RF/Microwave blood glucose sensing. In Proceedings of the IEEE International Instrumentation and Measurement Technology Conference Proceedings, Taipei, Taiwan, 23–26 May 2016; pp. 1–5. [Google Scholar]
  76. Kiani, S.; Rezaei, P.; Fakhr, M. Dual-frequency microwave resonant sensor to detect noninvasive glucose-level changes through the fingertip. IEEE Trans. Instrum. Meas. 2021, 70, 6004608. [Google Scholar] [CrossRef] [Scilit]
  77. Mohammadi, P.; Mohammadi, A.; Kara, A. Dual-frequency microwave resonator for noninvasive detection of aqueous glucose. IEEE Sens. J. 2023, 23, 21246–21253. [Google Scholar] [CrossRef] [Scilit]
  78. Mustafa Kamal, M.H.; Qureshi, S.A.; Zainal Abidin, Z.; Majid, H.A.; See, C.H. Evaluation of microwave square ring metamaterial-based resonator for glucose detection. J. Eng. Appl. Sci. 2024, 71, 66. [Google Scholar] [CrossRef] [Scilit]
  79. Souza, M.I.; Santos, N.M.; Alarcon, J.C.; Varanda, L.C.; Pepino, V.M.; Borges, B.H.V. Microwave glucose sensing using double circular split ring resonators for improved sensitivity: The role of artificial blood plasma and deionized water. IEEE Sens. J. 2024, 25, 4529–4540. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.