Microextraction Technologies as Exposomic Sensors for On-Site Environmental Air Monitoring of Volatile Organic Compounds: A Review of Commercially Available Technologies

,
,
,
,
,
,
,
,
…
1
Department of Life Sciences, Health and Health Professions, Link Campus University, via del Casale di San Pio V 44, 00165 Rome, Italy
2
Department of Experimental and Clinical Medicine, University of Florence, Largo Giovanni Alessandro Brambilla 3, 50134 Florence, Italy
3
Division of Occupational Medicine, Careggi University Hospital, 50134 Florence, Italy
4
Institute of Atmospheric Pollution Research (IIA), National Research Council (CNR), Strada Provinciale 35d n. 9, 00010 Montelibretti, Italy

Abstract

Microextraction Technologies (METs) have emerged as pivotal exposomic sensors for the on-site monitoring of Volatile Organic Compounds (VOCs) in ambient air. By integrating sampling and sample preparation into a single step, METs provide solvent-free, miniaturized, and field-deployable solutions that align with the principles of green analytical chemistry. This review critically examines fourteen commercially available METs, selected for their demonstrated analytical performance, commercial accessibility, and validation in real-world environments. These devices represent the current state of practice in exposomics, enabling both short-term hotspot detection and long-term exposure assessment. Particular attention is given to their compatibility with transportable and portable detection platforms, including vehicle-mounted and hand-held gas chromatography/mass spectrometry systems, where METs function as front-end concentrators that enhance sensitivity and spatial resolution. This review further discusses emerging applications in wearable formats and unmanned aerial vehicles, underscoring the role of METs in bridging laboratory-grade precision with field-based exposome research. By situating METs within the broader exposomic workflow of sampling, detection, and interpretation, this work identifies current technological gaps and outlines priorities for advancing robust, scalable, and environmentally sustainable exposure assessment strategies.

1. Introduction

Environmental factors, particularly environmental pollutants and lifestyle behaviors, are among the main drivers of chronic non-communicable diseases, which now dominate global mortality. The chemical industry transforms raw materials into more than 350,000 distinct substances and products [1]. Consequently, unconsented exposure to these harmful substances constitutes a human rights violation. Du et al. [2] report an average lifetime cancer risk of 2.27 × 10−4, equivalent to an additional 2.27 cases per 10,000 exposed individuals, attributable to Hazardous Air Pollutants (HAPs). The Global Burden of Disease study reported that outdoor air pollution will cause approximately 6 to 9 million premature deaths annually by 2060. The associated global welfare losses, derived from individual willingness-to-pay estimates for reducing mortality risk, are expected to rise from USD 3 trillion in 2015 to USD 18–25 trillion by 2060. In addition, the welfare costs due to pain and suffering from pollution-related illnesses are likely to increase from around USD 300 billion in 2015 to about USD 2.2 trillion by 2060 [3]. The concept of the exposome was first proposed by Christopher Wild in 2005 [4], and the term encompasses the totality of human environmental exposures (defined as all non-genetic factors) from conception onwards. This framework complements the genome and emphasizes the urgent need for more comprehensive and precise environmental exposure data.
The general external exposome encompasses air quality determinants such as Volatile Organic Compounds (VOCs) [5]. VOCs are commonly defined as carbon-containing organic compounds with a boiling point below 373.15 K at 101 kPa and a vapor pressure above 13.3 Pa at 25 °C [6]. The EU Solvents Directive (1999/13/EC) provided a complementary criterion, defining VOCs as organic compounds with a vapor pressure of at least 10 Pa at 20 °C [7]. Emitted from diverse anthropogenic and natural sources, VOCs have attracted growing attention due to their widespread release and significant environmental and health impacts; they can also generate odor nuisance, adversely affecting quality of life [8]. Among VOCs, 187 have been classified as HAPs because of their demonstrated potential to pose serious risks to human health and ecosystems [9,10]. VOCs’ effective environmental quality assessment and pollution control depend fundamentally on analytical chemistry, which supplies the methods and techniques necessary to detect and quantify both inorganic and organic pollutants.
The concept of the exposome has seen a rapid growth in the literature, with PubMed results increasing sharplyfrom 202 between 2005 and 2015 to 3404 in 2015–2025. This expansion reflects two primary motivations: first, the simultaneous assessment of multiple environmental exposures promises a more comprehensive evaluation of their collective impact on human health; second, a high-precision characterization of environmental exposures enables integrated studies of gene–environment interactions. Several large-scale exposome projects have emerged, particularly in Europe, such as the HELIX initiative, which integrates early-life environmental hazards, ranging from chemical to physical stressors, and correlates them with child health, growth, and respiratory outcomes [11]. Another important project is EXPOsOMICS, which develops novel exposure assessment methodologies for air and water pollutants, linking them to the molecular biomarker effect [12]. The Heals consortium aims to construct individualized internal exposomes by combining omics data with biochemical biomonitoring [13]. In the United States, the HERCULES program provides infrastructure and expertise to advance exposome tool development, validation, and application in epidemiological research [14].
Since the 1960s, the introduction of comprehensive environmental regulations has driven an escalating demand for the quantitative analysis of environmental samples [15]. Such data underpin monitoring of air, water, and soil quality; the evaluation of contaminants distribution over space and time; the assessment of contamination severity; and tracking of remediation effectiveness. Currently, the exposome paradigm offers an integrative framework to consolidate fragmented environmental exposure data and to account for multifactorial stressors in disease etiology [16,17]. To operationalize the exposome in population studies, investigators have adopted geographic information systems, remote sensing platforms, smartphone-based personal exposure sensors, and high-throughput miniaturized air samplers [18]. On-site, continuous analysis of chemical stressors, conducted in situ rather than in centralized laboratories, shows its potential in exposomics, thanks to the accurate and timely assessment of environmental hazards [19]. Moreover, portable green analytical tools eliminate errors, delays, and costs associated with sample transport and storage, and lead to a reduction in the use of solvents by combining the sampling phase with the sample preparation phase [20]. Recent technological innovations have encompassed field-deployable separation and detection systems, such as compact Gas Chromatographs (GCs), Liquid Chromatographs (LCs), and Mass Spectrometers (MSs), alongside continuous monitoring devices and point-of-care platforms. Over the past decade, the extensive miniaturization of these instruments has driven significant reductions in operational time and cost [21,22,23,24,25,26,27,28]. Portable MS units are gaining traction and are set to transition from research settings to control laboratories, enabling on-site and at-line analyses. Today, both vehicle-portable and man-portable GC, LC, and MS systems are advancing rapidly: vehicle-portable platforms are evolving toward temporally and spatially resolved mapping once limited to stand-off methods, while man-portable formats are shrinking from bulky “luggable” to wearable backpacks and even handheld units. Portable, miniaturized sampling tools, such as Microextraction Technologies (METs) and GC, LC, and MS, offer the real-time, high-resolution chemical measurements essential for comprehensive multi-stressor evaluation and integrated gene–environment analyses in large-scale population studies, as required by exposomics.
The literature selection presented in this analysis was conducted by consulting the PubMed, Scopus, and Web of Science databases, examining contributions published from 2000 to 2025. The selection criteria favored studies reporting validations in real-world scenarios, the commercial accessibility of devices, and on-site analytical performance. In order to ensure a comprehensive overview of the principal technologies currently available on the market, the research was extended to gray literature, including application notes and technical data sheets provided directly by manufacturers (e.g., Markers International, Gerstel, Restek, and Supelco).This review situates METs as exposomic sensors applicable across multiple matrices, including air, water, and soil, where Volatile and Semi-Volatile Organic Compounds represent critical exposure pathways. By critically examining fourteen commercially available devices, selected for their analytical performance and field validation, we clarify how METs interact with portable and transportable detection systems to enable both short-term hotspot identification and long-term exposure assessment. In doing so, this review highlights their position within the exposomic workflow—sampling, detection, and interpretation—and identifies current technological gaps together with priorities for advancing robust, scalable, and environmentally sustainable exposure assessment.

2. Results and Discussion

Urban air quality continues to be a pivotal determinant of public health, with even short-term exposures to fine Particulate Matter (PM2.5) and particle-phase VOCs showing measurable impacts [29,30]. Primary pollutant concentrations can vary substantially over short spatial scales with a variability which is often poorly represented by routine monitoring networks but is critical for any accurate exposure assessment, air quality management, and environmental equity [31,32]. Technological advancements over recent decades have facilitated continuous monitoring and the implementation of Field Analytical Methods (FAMs), combining portable instrumentation, mobile laboratories, and regulatory monitoring networks such as the EPA Air Quality System (AQS) [33]. Frameworks such as the Triad Approach, integrating systematic planning, dynamic workplans, and field-based measurement technology, seek to optimize site assessment while reducing uncertainty in environmental decision-making [34]. Despite these advances, regulatory networks remain limited in their capacity to capture fine-scale spatiotemporal variability, particularly for primary pollutants, with implications for epidemiological studies and population exposure estimation [35,36,37,38,39]. To address these gaps, complementary approaches have been developed, including satellite remote sensing, chemical transport models, land use regression, and direct personal exposure measurements [40]. Each method presents distinct advantages and limitations, emphasizing the importance of integrating multiple approaches to resolve local pollution gradients and micro-scale variations [41]. Mobile monitoring using professionally driven fleet vehicles or public transport, alongside dense networks of low-cost sensors, has proven to be effective in mapping neighborhood-level differences and persistent hotspots within urban areas [42,43,44,45]. These strategies enhance the spatial resolution of air quality data, improve the characterization of human exposure, and provide a robust foundation for interpreting long-term environmental and health outcomes.

2.1. Classification METs

Across environmental matrices, including air, water, and soil, sampling and sample preparation typically account for 70–80% of the total analytical time. This finding has driven the development of solutions aiming to minimizing both the time and labor required during the preparation phase, while simultaneously addressing the need to enhance the sensitivity of various analytical methods [46]. By merging sampling and preparation into a single operation, METs streamline exposomic workflow sampling, detection, and interpretation, while reducing solvent use and enabling near real-time measurements. Although METs have historically been associated with grab sampling (instantaneous sampling), technological advances now allow for proper exposomic assessments through TWA (Time-Weighted Average) monitoring. Devices with planar geometries such as TF_SPME or silicone wristbands act as passive samplers, allowing exposure to be integrated over hours, days or weeks. This approach allows daily variations to be leveled out and the impact of transient interferents to be reduced, providing a measure of chronic contamination that is essential for epidemiological studies. The typical procedure for monitoring complex samples encompasses multiple stages: sampling, sample preparation, separation, quantification, statistical evaluation, and decision-making analysis. Analytical efficiency and accuracy depend critically on selecting a sampling technique tailored to the matrix and target compounds, since this choice determines sensitivity, selectivity, and regulatory compliance. Thus, defining an appropriate monitoring approach requires a comprehensive understanding of the available sampling and analytical methods. Samples collected for compliance purposes are often repurposed beyond their original intent, enabling the identification of pollution sources, the assessment of control measures’ effectiveness, and the evaluation of the relationships between pollution levels and public health outcomes. Since their introduction in 1989 (Figure 1), METs have evolved to meet these requirements, offering robust formats applicable across diverse environmental matrices and reducing both analysis time and uncertainty.
Figure 1. Timeline of METs development with respect to the concept of the exposome.
METs, by integrating the sampling and sample preparation steps into a single step, support the entire exposomic analytical workflow: sampling, detection, and interpretation. Furthermore, miniaturization and the reduced use of reagents (consistent with the principles of green analytical chemistry) are fundamental aspects for achieving near real-time measurements. The analytical performance is influenced by factors such as the volume and type of sorbent phase, device geometry, and extraction efficiency for target analytes, affecting both sensitivity and capacity.
Recent classifications of METs as exhaustive or non-exhaustive have practical implications for calibration strategies, sampling duration, and suitability for field deployment [47]. Conversely, earlier studies, such as those by Nerin et al. [48], classified METs according to conventional Solid-Phase Microextraction (SPME) fiber and in-tube developments, while Mehdinia et al. [49] distinguished between static and dynamic techniques. Nonetheless, the classification of METs methods remains neither fully clear nor comprehensive. This study proposes a coherent terminology for describing METs. To this end, terms recommended by the International Union of Pure and Applied Chemistry (IUPAC) for analytical Solid-Phase Microextraction (SPME) were adopted [50]. A classification system that categorizes METs into four groups based on geometry was developed, namely planar, in-tube, coated bar–disk, and SPME (Table 1). Similarly, METs’ characteristics are being evaluated, including the use of porous or liquid phases, active or passive samplers, short- and/or long-term sampling, personal and/or area sampling, and exhaustive or non-exhaustive approaches. In addition, considering the exposomic framework, a brief evaluation was carried out on their compatibility with drones, wearable devices, and automated collection boxes, which represent the most widespread and ready-to-use solutions for greater spatial and temporal coverage of contaminant dispersion. Drones enable targeted spatial and vertical sampling of airborne contaminants across heterogeneous landscapes and hard-to-reach locations, while wearables are optimized for personalized exposome assessment, capturing individual time-resolved exposures that reflect microenvironments and activities, and the latter, stationary, programmable samplers, provide continuous or scheduled sampling at fixed points, supporting long-term temporal trends and networked spatial coverage [51,52]. For environmental monitoring, geometry is critical, since sampler placement and positioning influence uptake kinetics and representativeness across air, water, and soil matrices. As noted, METs can be developed in various formats, but thin-film, high-surface area geometries are proposed as the optimal choice for environmental analysis in both laboratory and field settings.
Table 1. Overview of commercially available METs with their respective specifications, materials, and manufacturers. List of acronyms used in the table: Sorbent types: Polydimethylsiloxane (PDMS); Polyacrylate (PA); Polyethylene Glycol (PEG); Divinylbenzene (DVB); Carboxen® (CAR); Hydrophilic–Lipophilic Balanced (HLB); Polyacrylonitrile (PAN); Strong Cation Exchange (SCX); Strong Anion Exchange (SAX); Carbon Molecular Sieve (CMS); Porous Polymer Sorbent (PPS); Activated Carbon (AC); Carbon Wide Range (CWR). GC Injection Port: Thermal Desorption Units (TDUs); On-Port Solvent-Assisted Injection (OPSI); Sorbent Pen™ Desorption Unit (SPDU). All cells containing the label Not Available indicate that the corresponding information is not publicly available online. * Exact numbers depend on coating chemistry, sorbent, matrix, extraction/desorption solvents, thermal cycles, and mechanical handling.
This list above provides the framework for evaluating the commercially available METs discussed in the following sections, highlighting how geometry, sorbent chemistry, and sampling mode could determine their role as exposomic sensors (Table 1).

2.1.1. Planar METs

The Thin-Film MET (TF-MET) has been commercially introduced by GERSTEL (Mülheim an der Ruhr, Germany) as TF-SPME as a planar microextraction device. Introduced in 2001 [53], it has since been employed for various gas and liquid chromatography applications [54]. Its large extractive-phase volume and thin-film geometry maximize the surface area, enhancing the analyte uptake and accelerating extraction kinetics [55]. The enhanced sensitivity of the TF-MET technique can be understood by applying the fundamental mass balance rule to SPME at equilibrium conditions [56]. Equation (1), derived from the mass balance under equilibrium conditions, defines the intrinsic relationship between the amount of analyte (the number of moles) extracted into the extractive phase at equilibrium (neeq) and its original concentration ( C s 0 ) in the sample. In this equation, Kes denotes the distribution constant of the analyte for the extractive phase and sample matrix, with Vs and Ve referring to the volume of the sample and extractive phase, respectively. Ultimately, this equation indicates mathematically that an increase in the volume of the extractive phase results in improved analytical sensitivity.
n e e q = K e s V s V e K e s V s + V s C s 0
In the abovementioned scenario, the equilibrium extraction time (te), defined as the time required to extract 95% of the equilibrium amount, can be calculated by Equation (2); where Ds, b, and δ refer to the diffusion coefficient of analyte in the sample, the thickness of the extractive phase, and the thickness of the boundary layer, respectively. This equation illustrates that for a given amount and type of extractive phase, a faster extraction equilibrium can be achieved when a thin layer of extractive phase is used.
t e = 3 δ K e s ( b ) / D s
Moreover, Equation (3) implies that the initial extraction rate, denoted as extracted amount (dn) per given time (dt), is directly proportional to the surface area of the extractive phase (A). Therefore, if a relatively large volume of extractive phase is spread over an ultra-thin layer, there will be a larger interface between the sample and the extractive phase, resulting in faster extraction kinetics.
d n d t = D s A δ C 0
Advances in TF-MET include specialized holders and samplers designed for diverse environments, such as rigid membranes for stability, Time-Weighted Average (TWA) devices for long-term monitoring, and robust blades for extreme conditions [57,58]. Hydrophilic–Lipophilic Balance (HLB) particles have been adopted to improve affinity for polar analytes [59]. TF-SPME devices are compatible with both gas and liquid chromatography, highlighting the critical role of sampler design in optimizing sensitivity and reproducibility [58]. Quintanilla et al. [60] applied a cellulose triacetate film doped with multiwalled carbon nanotubes for the monitoring of chlorpyrifos, tonalide, and triclosan in river water, and afforded limits of detection between 0.2 and 0.5 µg L−1, recoveries of 90–104%, and intra-film precision (RSD) < 7%. Notably, no significant matrix effects were observed for chlorpyrifos and tonalide, underscoring TF-MET’s robustness in complex environmental matrices. In the context of drinking water safety, thin-film SPME membranes coated with Hydrophilic–Lipophilic Balanced/polydimethylsiloxane phases have been paired with GC×GC–TOF MS to screen halogenated Disinfection By-Products (DBPs) at sub-ppb levels. The planar sorbent provided a 2.5-fold increase in extraction efficiency over fiber SPME [61]. Paper-based TF-MET patches offer an even more user-friendly format. Poojary et al. [62] proposed the use of disposable patches with divinylbenzene/PDMS/CNT coatings on cellulose paper for the quantification of 4-chlorophenol in water. Direct exposure for 10 min yielded a calibration range of 100–10,000 ng mL−1 and a limit of detection of 10 ng mL−1, validating the rapid field screening of this priority pollutant at regulatory threshold concentrations. Planar METs have also enabled faster and more sensitive air monitoring of explosives and other analytes, particularly when coupled with ion mobility spectrometry or advanced ionization techniques, including DART, DESI, SELDI, and MALDI [63,64,65,66].
Coated Blade Spray (CBS), developed by Gómez-Ríos and Pawliszyn [67], allowed for a direct and real-time analysis with minimal sample preparation, achieving low-picogram detection limits within three minutes and compatibility with portable and benchtop Mass Spectrometers [68]. Beyond its application to relatively simple matrices, CBS has been successfully tested on complex environmental and biological samples, where the selective polymer coating enables the efficient enrichment of target analytes. This capability demonstrates its potential as an exposomic sensor across diverse matrices, allowing direct coupling to mass spectrometry. Hu et al. [69], using CBS, developed a high-throughput and automated sample preparation method with the use of a 96-well plate system, enabling analysis in water of 21 drugs of abuse within 1 min per sample, while using only 8 µL of organic solvent for desorption and atmospheric pressure MS detection; coupling CBS and MS led to achieve low limits of detection from 1.5 to 9.0 ng L−1.
According to Yu et al. [70], CBS, functionalized with biocompatible polymers or hybrid nanosorbents, achieves comparable or lower limits of detection (0.01–0.5 µg L−1) for phenolic and polyaromatic pollutants than TF-MET, while offering superior mechanical stability and rapid thermal desorption directly within GC inlets. The planar, rigid format simplifies handling, facilitates automation, and withstands field conditions (temperature swings, biofouling, mechanical shock).
Taken together, TF-MET and CBS epitomize the ongoing miniaturization and greening of environmental analysis. Their high extraction capacity, minimal solvent use, and compatibility with advanced chromatographic–mass spectrometric techniques position them as front-line tools for on-site monitoring of legacy and emerging contaminants.

2.1.2. In-Tube METs

In parallel with planar TF-MET, in-tube or in-needle METs have been developed, offering exhaustive or non-exhaustive sampling, depending on the device design. Their compact geometry and compatibility with automated workflows make them particularly relevant for exposomic monitoring [71]. The first non-exhaustive in-tube device, the Solid Phase Dynamic Extraction SPDETM Magic Needle (Chromtech, 2000, Bad Camberg, Germany), consists of stainless-steel needles coated with sorbent and activated carbon, providing a larger stationary-phase volume than conventional SPME fibers and enabling automated, high-efficiency extraction using gas-tight syringes [72,73]. SPDE operates as a dynamic, non-equilibrium technique in which analyte enrichment is achieved through repeated suction cycles. However, an initial incubation period (typically 10–30 min at 60–70 °C) is required to promote analyte release into the headspace before dynamic extraction begins [74,75]. This dynamic approach confers a high enrichment capacity, with sensitivities reported up to ten-fold greater than static headspace methods for fire accelerant determination. Moreover, SPDE exhibits an exceptionally wide linear dynamic range, spanning nearly five orders of magnitude with PDMS coatings and exceeding six orders for VOCs in groundwater, underscoring its suitability for complex exposomic matrices.
Needle Trap Microextraction (NTME), introduced in 2001 by Pawliszyn’s group, serves as an exhaustive particle sampler, using commercial sorbents such as PDMS/DVB, Carboxen, Carbopack X, Tenax TA, and polymer beads. Combining various multiple sorbents or using nanoporous silica further enhances its performance [76,77,78]. Recent developments in Needle Trap Devices (NTDs) focus on advanced sorbent materials, including hybrid polymers and Metal–Organic Framework (MOF) coatings. For example, MIL-100(Fe) has been investigated as a MOF sorbent, offering enhanced selectivity and stability under field conditions. In addition, the robustness of NTDs has been confirmed: Anticó demonstrated that 22-gauge NTDs equipped with side-hole configurations and a thermally desorbed in-injector at 300 °C yielded reproducible sampling (RSD < 8%) and LODs of 0.01–0.1 µg m−3 for VOCs in indoor and outdoor air, with no carryover after 48 h of storage at 4 °C [79]. Direct thermal desorption within the injector preserves compound integrity and enables reproducible on-site airborne monitoring, eliminating the need for auxiliary gas lines or valves. This robustness is critical for exposomic studies that require reliable sampling under variable environmental conditions. In remote atmospheric campaigns, portable GC-MS platforms integrated with NTME sampling have quantified monoterpenes, aldehydes, amines, and anthropogenic VOCs at forest sites: Barreira et al. [80] reported that needle trap sampling followed by field-deployable GC-MS detected monoterpenes at 0.1–1 ppb, correlating diurnal emission patterns with particle number concentrations and illustrating the system’s utility for real-time ecosystem monitoring. Passive sampling variants of needle traps also exist [79]. Similarly, the NeedlEX device (Shinwa Chemical Industries, Kyoto, Japan) allows for repeated, automated air sampling of alcohols, organic solvents, amines, and fatty acids directly via GC injection without additional equipment [81]. In-Tube Extraction (ITEX) is a dynamic sampling method, often considered a cost-effective alternative to the purge and trap technique for enriching VOCs from aqueous samples. As a high-capacity enrichment technique (similar to SBSE), it provides high sensitivity; for example, the LOD for VOCs in honey has been reported in the range of 0.8 to 47 ng/g. ITEX demonstrates excellent precision and repeatability, with a relative standard deviation typically around 3% [82].
ITEX-InTube Extraction (CTC Analytics, Zwingen, Switzerland, 2008) uses a syringe-based microtrap packed with adsorbents (Tenax, Carbopack, Carbosieve, molecular sieve) to concentrate target compounds from air, which are then thermally desorbed directly into the GC injector. The ITEX-DHS approach employs a gas-tight syringe fitted with a micro-trap of adsorbent packing to repeatedly withdraw headspace from aqueous samples, enabling the dynamic enrichment of odorants such as geosmin, 2-methylisoborneol, 2,4,6-trichloroanisole, and isoborneol at sub-ppt concentrations. Under optimized conditions (60–100 repeated strokes, trap desorption at 250 °C), LODs down to 1–5 ppt were achieved, with linear ranges spanning three orders of magnitude and recoveries of 92–105% in river and drinking waters [83]. Full automation is achieved using modified software and autosamplers for preconditioning, internal standard addition, exhaustive extraction, and desorption [84,85,86].
Exhaustive METs include Sorbent Pens™ (Entech Instruments, Simi Valley, CA, USA, 2016), available in diffusive or active modes with a variety of sorbents for short- or long-term air sampling. PowerSorb® (Action Europe, 2017) is a polymer-based cylinder designed for thermal desorption or solvent extraction, applicable as a passive sampler or SPME alternative, with applications ranging from human scent detection to VOC monitoring for health and environmental studies [87,88,89,90,91,92]. Supplied in pre-packed vials under inert gas, the 20 mm × 2 mm polymer rod can be directly immersed in aqueous samples or deployed for dynamic and passive headspace sampling of VOCs without any prior conditioning steps.
The Getxent® microtube (Getxent, Biodesiv LLC, Switzerland) is a cost-effective polymer stick (3.8 × 30 mm) suitable for exposome analysis, compatible with thermal desorption below 100 °C [93,94]. Its simplicity and affordability make it attractive for large scale exposome studies, particularly where high sample throughput and minimal infrastructure are required. Taken together, in-tube METs provide versatile formats ranging from dynamic non-equilibrium enrichment (SPDE, ITEX) to exhaustive sorbent traps (NTME, Sorbent Pens). Their adaptability to complex matrices, compatibility with automation, and robustness under field conditions position them as essential tools for exposomic monitoring.

2.1.3. Coated Bar–Disk METs

The oldest METs in terms of coated bar conception, the Stir Bar Sorptive Extraction (SBSE) technique, patented in the US in 2002 [95], was proposed by Baltussen in 1999 [96]. The device is a magnetic stir bar enclosed in a glass shell covered with a sorbent layer with a PDMS sorbent layer (0.5–1 mm), commercially available as Twister (Gerstel GmbH) [97]. These bars come in two sizes for different sample volumes, and in 2012, an Ethylene Glycol (EG) silicone sorbent phase was introduced to extend the range of extractable compounds [98,99,100]. SBSE is characterized by a very high extraction capacity thanks to the thick layer of PDMS used, with almost quantitative recoveries for nonpolar compounds (log Kow > 5). PDMS coatings exhibit high durability, remaining stable for at least 100 extraction cycles, allowing a large number of samplings with costs (around 100$ each SBSE). Concerning airborne pollutants, for Polycyclic Aromatic Hydrocarbons (PAHs), extending SBSE sampling durations to 60 min and coupling to thermal desorption GC–MS achieved sub-ng m−3 sensitivity. SBSEs are particularly well suited to the study of long-term environmental exposure and the monitoring of chronic contamination. Thanks to their high capacity and robustness, passive sampling provided by SBSEs offers an integrated measure of exposure that is not altered or limited by specific conditions on a given day, leveling out variables and reducing the weight of interferents in analytical evaluation. The MonoTrapTM—Monolithic Material Sorptive Extraction (MMSE) uses a porous silica monolithic hybrid surface functionalized with activated carbon, graphite, PDMS, and C18 groups, providing a high adsorption capacity. Extraction occurs primarily by adsorption with minimal solvent consumption and no need for preconditioning. Two adsorption chemistries are available: silica-like for nonpolar analytes and silica–carbon for enhanced polar compound retention. The innovative RGC18-TD adsorbent (graphitic carbon/ODS) offers high sensitivity; moreover, the implementation of vacuum-assisted extraction (Vac-HS-MMSE) reduces kinetic equilibrium times. It enhances sensitivity, with signal intensities increasing by roughly one order of magnitude for higher boiling analytes. This makes the technique effective for the analysis of odorous VOCs (aldehydes, ketones, alcohols) with very low odor thresholds (OTs), such as the compounds emitted in the environment by hot-mix asphalt during road surfacing [101].
MMSE devices are offered in rod and disk formats and can be reused after flushing [102,103,104]. In dynamic flow trials, 45 min sampling at 150 mL min−1 allowed for the determination of phenolic and nitro-aromatic compounds with LODs of 0.2–1.0 µg m−3, recoveries ranging from 65 to 92%, and RSDs under 12%. The rigid bar format facilitates direct thermal desorption in GC inlets (250–300 °C) and resists mechanical stresses encountered during portable or remote deployments [105]. The Magic Chemisorber (Frontiers Laboratories, 2004) is a solid-phase diffusive extraction device originally coated with a 500 µm PDMS layer, providing a high extraction capacity. Sampling is performed by direct immersion in aqueous samples or in the headspace phase. The device is effective for non-polar analytes such as PAHs and Polychlorinated Biphenyls (PCBs). Studies have reported excellent recoveries for PAHs and PCBs in water (96% to 110%) and a very low LOD of less than 2.7 ng L−1 for PAHs. Desorption occurs thermally in a pyrolysis–GC-MS system, with excellent reproducibility. In 2019, a PEG phase was introduced on the outer wall of a deactivated stainless-steel tube, enabling selective sampling of polar compounds [106,107]. Together, SBSE, MMSE, and the Magic Chemisorber exemplify the coated-bar MET family, which, thanks to their high sorbent capacity, mechanical robustness, and compatibility with thermal desorption, make them particularly suitable for long-term exposomic monitoring across different matrices, especially when integrated measures of chronic exposure are required.

2.1.4. SPME METs

The SPME, first commercialized by Supelco in 1993 and improved in 2001 with a customized holder, has significantly influenced sample preparation in analytical chemistry by offering miniaturized, solvent-free workflows, short extraction times, low cost, and easy coupling to GC and LC [47]. These advantages have led to widespread applications in environmental studies and on-site analysis [108,109], particularly for VOC sampling, allowing for short- or long-term exposure, on-fiber derivatization, on-line sampling with autosamplers, and compatibility with field GC or GC-MS systems [110,111,112,113,114,115,116]. Advancements include Restek’s SPME Arrows (2015), which feature a larger-diameter, robust probe with increased sorbent volume for higher sensitivity and throughput, and Markes International’s HiSorb probes (2016), consisting of a thin inert rod wrapped with PDMS sorbent for non-exhaustive gaseous analyte extraction [117,118,119,120,121]. A direct comparison of the technologies reveals that extraction capacity is closely related to the volume of the extraction phase. While a conventional SPME fiber offers a limited volume (approximately 0.9 µL), Arrow SPME (3.8–11.8 µL) and SBSE/Twister (63–126 µL) provide a significant increase in analytical sensitivity, which is essential for detecting traces in complex matrices. In terms of operational robustness, Twister devices maintain stability for over 100 extraction cycles, while NeedlEx fibers are reusable for approximately 25–30 analyses. The use of Nitinol (NiTi) cores or StableFlex fibers has further improved mechanical durability compared with traditional fused silica. For environmental sampling, especially in polluted air, the use of porous adsorbent phases (e.g., Carboxen) can be subject to the displacement effect (competitive adsorption), in which high-concentration interferents compete for active sites, displacing target analytes. Recent advances in coating chemistry, encompassing inorganic materials (metal–organic frameworks, carbon nanotubes, graphene), organic polymers (polymeric ionic liquids, molecularly imprinted polymers), and hybrid nanocomposites, have substantially enhanced extraction capacity, thermal and solvent stability, and analyte specificity [122]. Although early SPME applications mostly involved laboratory analysis following field sampling, some recent studies have demonstrated its effectiveness for immediate on-site analysis. Gorecki and Pawliszyn showed SPME as a viable method for high-speed field GC separations [123]. Koziel et al. applied SPME for formaldehyde detection and other organic analytes in indoor air and field settings [124,125,126]. Recently, Strano et al. integrated conventional SPME fibers alongside photoionization (PID) and Metal-OXide (MOX) sensors on a small unmanned aerial vehicle to tackle airborne threats and fugitive industrial emissions. Using benchtop validation, multiple SPME coatings (e.g., PDMS, polyacrylate) were exposed to calibration atmospheres of VOCs and nerve agent simulants, yielding limits of detection in the sub-µg m−3 range (e.g., ≈0.1 µg m−3 for toluene and 0.05 µg m−3 for dimethyl methylphosphonate). The combined real-time response of the PID/MOX array guided flight paths to hotspots, while post-flight GC-MS analysis of the fibers provided compound-specific identification and quantitation, with recoveries between 85 and 110% and RSDs below 10% [127].
Other field GC-MS applications include the rapid detection of thermal degradation products from riot control agents [128]. A critical example of MET application in the exposome field is monitoring in the coffee processing industry, an environment characterized by the coexistence of carbon monoxide (CO), alpha-diketones, and bio-reactive dusts. In these scenarios, the grinding process represents the epicenter of risk, with diacetyl peaks that can exceed 9000 ppb. While photoionization detectors (PIDs) provide real-time data, METs coupled with GC-MS are indispensable for chemical speciation, allowing differentiation between diacetyl and 2,3-pentanedione, which is essential for the prevention of bronchiolitis obliterans. The adoption of analytically designed localized ventilation (LEV) has been shown to reduce exposure to al-alpha-diketones by up to 16 times.

2.2. Air Sampling and Monitoring

2.2.1. Stationary Sampling

Recent advances in on-line SPME monitoring and automated sampling systems have significantly improved analytical workflows for environmental and field studies. While conventional stationary samplers provide accurate measurements at fixed locations, mobile laboratories positioned close to sampling sites enable on-line sampling, often via Teflon lines. Commercially available, fully automated miniaturized techniques rely primarily on three-axis (xyz) autosamplers, which have proliferated over the last decade, particularly when coupled with GC. These systems, featuring modularity and flexibility, support microextraction techniques for air sampling and can be equipped with accessories such as barcode readers, robotic syringe changers, decappers, heating and stirring modules, and LAN connectivity to enhance usability and productivity [129,130]. CTC Analytics AG pioneered commercial xyz autosamplers, releasing the first GC liquid autosampler (A200S) in 1986, followed by the PAL system in 1998 and subsequent expansions including the HTX PAL, PAL RTC, RSI, and LSI systems (2003–2014). Today, various companies, Leap Technology, Chromtech, Markes International, GERSTEL, KONIK Group, EST Analytical, Moduvision Technologies, and PAS Technology, produce customized autosamplers to meet diverse analytical needs and improve workflow efficiency. In parallel, portable sequential sampling devices have been developed for environmental pollution assessment, capable of remotely collecting gases, vapors, and Particulate Matter (PM) via tubes, filters, denuders, canisters, or bags, often in compliance with EPA speciation requirements. These systems are increasingly applied in industrial hygiene for area sampling. Examples include CEH DELT®, GasCheck™ (AMS Analitica, Pesaro, Italy), SASS™ (Met One Instruments, Grants Pass, OR, USA), URG-2000, Partisol® Model 2300, GAS08/16, Echo TUBE, and MTS-32™ (Markes International, Bridgend, Wales (UK)). Innovative solutions such as ODORPREP® (Lab Service Analytica, Anzola dell'Emilia, Italy) offer real-time, automated, remote-activated sampling of industrial odor emissions, compliant with European Standard EN 13725 [131], and supported by Horizon 2020 funding (Project OdorPrep, G.A. 756865). Complementing this, OdorBOT enables citizen reporting of odor nuisances via a Telegram™ interface, collecting and mapping real-time data for subsequent analysis [132].

2.2.2. Mobile Sampling

Mobile and wearable sampling approaches have advanced individual exposure and exposome monitoring by enabling the high-resolution characterization of chemical interactions in diverse microenvironments. Mobile systems, including automobiles, drones, helicopters, or boats, allow the rapid deployment of SPME samplers or mobile GC systems with adapted power and carrier gas supplies. These roving systems provide high-speed, repeated measurements that capture transient events and map sharp spatial concentration gradients, including point sources [133,134,135]. On the road, vehicle-mounted SPME–GC systems have likewise been employed for high-resolution mapping of urban VOC plumes. One example includes Kore PTR 3c, equipped with Live-Drive Technology and Instavac Systems for near-instant VOC measurements, and drone-mounted TF-SPME devices for the on-site screening of pollutants in hard-to-reach water bodies [59]. METs represent a powerful tool for environmental and exposome assessment, as they are easily integrable with unmanned aerial vehicles (UAVs) or drones; miniaturization and generally no need for reagent consumption are crucial, as they help to drastically reduce overall mass and energy requirements.
Field studies and reviews have demonstrated the effectiveness of drone-compatible METs. Notable examples include a TF-SPME drone water sampler with a payload of about 12 g, coupled with a portable GC-MS for on-site screening, and an aerial SPME Arrow campaign that collected 135 samples at heights between 50 and 400 m and identified 48 VOCs [136].
Nowadays, the main challenges concern the need to further improve energy efficiency, reduce mass, and ensure the stability of portable analytical configurations (such as miniaturized GC-MS), that accompany METs analyses. Wearable samplers are deployed on participants or in microenvironments to assess inhalation exposure, using either active devices (pump-assisted quantitative sampling) or passive devices (time-integrated sampling via diffusion). Early developments include wearable SPME necklaces and pins for ketamine monitoring using ionic liquid coatings [137]. More recent innovations, such as the Fresh Air Wristband Clip, integrate PDMS sorbent bars and triethanolamine-coated pads to monitor VOCs, PAHs, and NO2 [138,139,140,141]. Sorbent membranes have also been incorporated into badges, brooches, and lapel diffusion tubes, enabling personal exposure assessment. Silicone wristbands have proven effective for the multi-hour to multi-week monitoring of environmental and occupational pollutants, including polybrominated diphenyl ethers, organophosphate flame retardants, and pesticide residues, and have been successfully applied across diverse populations and geographic locations. By integrating exposures over extended periods, these devices provide a more comprehensive picture of chronic contamination that is not distorted by day-to-day variability. In this way, transient fluctuations and analytical interferents are leveled out, reducing their impact on data interpretation and supporting long-term exposomic studies of environmental exposure. In one field deployment, participants wore silicone wristbands continuously for ten days; the analysis revealed a median of 43 unique chemicals per device, with individual analyte masses spanning 0.1–500 ng per band [52]. Recently, miniaturized electrochemical and optical sensors have been integrated into wearable badges, vests, or backpacks to provide real-time measurements of gases and particulates. Urban cohort studies using wearable PM2.5 monitors recorded personal exposures averaging 28 ± 12 µg m−3, approximately 23% higher than concurrent fixed-site measurements, highlighting substantial exposure variability during commuting and indoor–outdoor transitions [142].

2.2.3. Transportable and Portable Instruments

Two complementary deployment strategies have emerged to bridge the gap between field sampling and laboratory analysis. Vehicle-based mobile laboratories, which require the installation of compact benchtop GC or LC systems in trailers or vans, maintain stable power supplies and controlled environmental conditions that closely replicate those of fixed laboratories. This approach dramatically reduces the time between sample collection and data generation, allowing complex GC-MS and LC-MS workflows to be carried out directly at the point of interest. At the same time, advances in microfabrication and power management have led to the development of truly portable chromatographs, which integrate pumps, capillary columns, detectors, and electronics within a single enclosure. According to Sharma et al. [143], such systems must operate autonomously for at least eight hours, which is ideal for monitoring occupational exposure, which normally refers to intervals or limits over eight hours, using battery or solar power; achieve rapid thermal equilibration; and resist extremes of temperature and humidity, dust, and mechanical shock. They must also deliver separations and sensitivities comparable to those of benchtop instruments, all while minimizing the use of toxic solvents and extra-column volume in order to uphold the principles of green chemistry. Together, these developments enable high-performance, on-site chromatographic analysis in diverse and challenging field environments.
Transportable MS, LC, and GC
Recent advances in analytical instrumentation have enabled rapid, on-site chemical analysis by integrating MS and chromatography with miniaturized and portable sampling techniques (Table 2). MS technologies are classified into beam-based (sector, time-of-flight, quadrupole) and trapping-based (ion trap) analyzers [122]. Among these, Ambient Mass Spectrometry (AMS) has revolutionized direct analysis, allowing the rapid detection of analytes in complex matrices without extensive sample preparation or chromatographic separation [144,145]. Representative ionization techniques include Low-Temperature Plasma (LTP), miniFAPA, Desorption Electrospray Ionization (DESI), and Direct Analysis in Real Time (DART), with DESI relying on electrospray mechanisms and DART on Atmospheric Pressure Chemical Ionization (APCI). A wide array of other ESI- and APCI-related methods, such as paper spray, touch spray, CBS, and Probe ElectroSpray Ionization (PESI), further extend the analytical capabilities of AMS [146,147,148,149,150,151,152,153,154,155,156]. DESI has been already used for airborne particulate monitoring: dynamic DESI sampling of filter-mounted urban aerosols enabled the quantification of nitro-PAHs down to 0.1 ng m−3, correlating concentration hotspots with traffic patterns and seasonal biomass burning events [145]. The integration of AMS with microextraction techniques has been achieved through devices such as the Transmission Module™ (TM) and SPME-TM, which use polymer-coated stainless-steel meshes to concurrently isolate and enrich analytes while minimizing interferences that could suppress or enhance ionization. Coupled with high-resolution instruments such as AccuTOF-DART™, this approach produces results comparable to traditional GC-MS within seconds [157,158,159,160]. On-site chemical analysis is further facilitated by miniature MSs, including SPIMS 2000 and systems from BaySpec Inc. (San Jose, CA, USA), enabling the rapid profiling of VOCs and other target compounds in field conditions [160,161]. Parallel developments in chromatography complement these advances. Portable and transportable SPME and Twister applications allow direct coupling to conventional or fast GC systems, improving sample throughput and analytical efficiency. Commercial high-speed GC systems, such as Trace 1300 (Thermo Fisher, Waltham, MA, USA), Nexis GC 2030 (Shimadzu, Kyoto, Japan), Clarus SQ 8 (PerkinElmer, Waltham, MA, USA), and Agilent (Santa Clara, CA, USA) Intuvo 9000, enable ultra-fast analysis and facilitate on-site environmental and industrial hygiene applications. Similarly, compact high-pressure LC systems provide versatile, portable solutions for in-field analysis. These instruments allow precise separation with minimal laboratory footprint and energy consumption, particularly when coupled with MS for real-time detection [162]. Miniaturized systems such as the MiD platform (Microsaic Systems, Sand Hutton, England) integrate chip-based MS detection with LC, enabling efficient direct analysis of complex samples in the field, chemical reaction monitoring, or environmental pollutant assessments. These technologies reduce the need for laborious sample preparation and long chromatographic separations, streamline workflows, and expand the capacity for environmental, industrial, and occupational monitoring under real-world conditions.
Table 2. Overview of commercially available transportable GC, LC, and MS with their respective specifications and applications.
Man-Portable MS, GC, and LC Systems
Nowadays, modern portable instruments allow rapid, high-resolution measurements of gases, vapors, and liquids directly in the field (Table 3). Membrane Introduction Mass Spectrometry (MIMS) utilizes membrane sampling probes to allow the selective permeation of gaseous or aqueous molecules into the mass spectrometer for qualitative and quantitative analysis. As MIMS does not require prior sample preparation, it is ideally suited for portable MS systems and on-site applications [163,164,165,166]. MIMS does not require energy support to power mechanical pumps (rotary or turbomolecular) to achieve a vacuum; at most it can compensate for the use of these with chemical pumps or ion pumps. Extensions of MIMS, including Trap-and-Release (T&R) MIMS and Single-Sided MIMS (SS-MIMS), enable the sensitive detection of polar Semi-Volatile Organic Compounds (SVOCs) through preconcentration within ultra-thin membranes followed by rapid thermal desorption [167,168,169,170,171,172]. In 2002, Fiber Introduction Mass Spectrometry (FIMS) was introduced, coupling SPME directly to MS. FIMS uses PDMS-coated fibers to achieve selective adsorption, efficient preconcentration, and single-sided desorption of VOCs and SVOCs, enabling high-throughput analysis at trace levels [173]. Commercial portable systems implementing these technologies include VapourSense Q-Technologies’ portable quadrupole MS and 1st Detect’s MMS-1000™, a miniature ion trap MS with a rapid response and an integrated pre-concentrator for enhanced sensitivity. Parallel developments in portable GC have expanded field capabilities. Portable GC and GC-MS systems, including backpack-style instruments such as HAPSITE, Teledyne FLIR Griffin G510, Torion T-9, and EXPEC-3500, allow rapid on-site measurements of gases, vapors, and airborne pollutants, with response times as low as one second. These systems often combine MEMS-based pre-concentrators, SPME fibers, and miniaturized columns to reduce sample preparation and enhance sensitivity; for example, Torion T-9 was applied on field evaluations of roadside air emissions, reporting LODs of 0.02–0.05 µg m−3 for BTEX, during continuous 5 min sampling cycles at 30 km h−1 vehicle speed [174]. FLIR’s Griffin G510 pairs a high-speed, low-thermal mass GC column with a linear quadrupole MS, operating in two modes, Survey (real-time vapor screening) and Confirmation (split/splitless injection of vapors, liquids, or solids via SPME/Touch-And-Go probes), enabling the comprehensive detection of chemical agents, industrial toxicants, and environmental pollutants. In ambient air trials, the G510 achieved LODs of 0.05–0.10 µg m−3 for BTEX in <4 min Survey scans (100 mL min−1 flow) and confirmed semi-volatile PAHs at sub-ng m−3 levels after 15 min thermal desorption runs [175]. Miniaturized and portable high-pressure LC systems complement these gas-phase analyses. Advances in column miniaturization, including capillary and micro-chip formats, combined with low-flow pumps and sensitive detection, allow portable LC systems such as Focus LC Compact (Axcend®) to perform on-site separation and analysis with minimal solvent consumption. While early portable LC relied on conventional columns and detectors, recent developments integrate UV absorption or MS for detection, enabling field-based chemical monitoring and reaction analysis [176,177,178,179,180]. These innovative transportable instruments reduce laborious sample preparation, enable near real-time measurements, and facilitate the monitoring of dynamic chemical processes across diverse matrices, including gases, vapors, liquids, and complex environmental or industrial samples, combining laboratory-grade performance with field portability. Despite the advantages of on-site monitoring, portable instruments have inherent limitations compared with benchtop systems, including shorter columns (which reduce chromatographic resolution) and mass analyzers with reduced performance. METs therefore serve as front-end concentrators necessary to compensate for the lower intrinsic sensitivity of portable instrumentation.
Table 3. Overview of commercially available portable GC, LC and MS with their respective specifications and applications.

3. Conclusions

The METs coupled with portable chromatographic and MS platforms potentially enable high-fidelity, on-field measurements of airborne pollutants within the exposome framework. By integrating sampling, preconcentration, and analysis into compact, automated workflows, MET-GC, MET-LC, and portable MS systems deliver sub-ppb detection limits, rapid turnaround, and minimal solvent use, thereby meeting the dual demands of green analytical chemistry and exposome-scale exposure assessment. Although progress has been made, bringing these tools into everyday sampling environments will depend on their ability to perform reliably under tough conditions, such as extreme temperatures, high humidity, mechanical shocks, and regular vibrations. To move from prototypes to networks that can truly support epidemiological work, it is equally important to focus on affordable manufacturing, the use of open-source hardware and software, and clear, standardized protocols for calibration, data management, and consistency across devices. Looking forward, the convergence of MET-based preconcentration, ambient ionization MS, microfluidic LC modules, and wireless telemetry promises fully instrumented exposome observatories. This review provides an evaluation of the fourteen commercial METs available for in situ environmental sampling of VOCs, highlighting their capabilities to meet short- and long-term sampling requirements in exposomics. Despite significant progress, METs face technical limitations that shape future research directions. There are physical and chemical limits to miniaturization in order to maintain a high analytical performance. In addition, long-term stability and calibration drift management pose a significant challenge, particularly in extreme environmental conditions. Although METs are mature technologies ready for field use, challenges remain regarding protocol standardization and calibration in extreme environmental conditions (humidity and temperature), which can cause drift in results. The transition to a comprehensive monitoring network will require the greater production of independent validation data and the acceptance of microextraction methods by regulatory bodies in order to overcome the current regulatory gap. Future priorities will therefore include developing new energy solutions to support the integration of unmanned systems, as well as enhancing automation and data integration with detection and interpretation systems.

Author Contributions

Conceptualization, N.M., S.D. and R.G.; methodology, I.R. and G.C.; investigation, E.G., F.C. and D.C.; writing—original draft preparation, S.D., N.M., I.R. and G.C.; writing—review and editing, all authors; visualization, A.B. and V.T.; supervision, I.S. and R.G.; project administration, R.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Acknowledgments

This manuscript has been developed based on the concept of the project FIN.E.ODOR. (FINger-print ed Emissioni ODORigene di conglomerati bituminosi) by SITEBassociazione Strade ITaliane E Bitumi (Bologna, Italy) and the PIN-Polo Universitario Città di Prato (Prato, Italy) of the University of Florence. This work was also carried out thanks to the technical scientific collaboration between ARPAT and the University of Florence “CARETTERIZZAZIONE DELLE EMISSIONI ODORIGENE ATTRAVERSO LA CREAZIONE DI PROFILI ANALITICI CARATTERISTICI”. We would like to thank and remember the contribution made by the Mieczysław Sajewicz (23 March 1951–6 June 2025)—a colleague and a sincere friend, who suddenly passed. Mietek was not only a successful researcher in his professional field—the chemical analysis of organic compounds, mainly mass spectrometry, as well as spontaneous chiral inversion processes and vibrational processes—he was also an extremely talented person in other areas that require both high intellect and quickness of the mind.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
METsMicroextraction Technologies
HAPsHazardous Air Pollutants
VOCsVolatile Organic Compounds
GCsGas Chromatographs
LCsLiquid Chromatographs
MSsMass Spectrometers
PMParticulate Matter
FAMsField Analytical Methods
AQSAir Quality System
IUPACInternational Union of Pure and Applied Chemistry
SPMESolid-Phase Microextraction
TWATime-Weighted Average
HLBHydrophilic–Lipophilic Balance
CBSCoated Blade Spray
NTMENeedle Trap Microextraction
LODLimit of Detection
MMSEMonoTrapTM—Monolithic Material Sorptive Extraction
AMSAmbient Mass Spectrometry
LTPLow-Temperature Plasma
DESIDesorption Electrospray Ionization
DARTDirect Analysis in Real Time
APCIAtmospheric Pressure Chemical Ionization
PESIProbe Electrospray Ionization
TMTransmission Module
MIMSMembrane Introduction Mass Spectrometry
SVOCsSemi-Volatile Organic Compounds
FIMSFiber Introduction Mass Spectrometry
TF-METThin-Film Microextraction Technologies
TF-SPMEThin-Film Solid-Phase Microextraction
PDMSPolydimethylsiloxane
PAPolyacrylate
PEGPolyethylene Glycol
DVBDivinylbenzene
CARCarboxen®
HLBHydrophilic–Lipophilic Balanced
PANPolyacrylonitrile
SCXStrong Cation Exchange
SAXStrong Anion Exchange
CMSCarbon Molecular Sieve
PPSPorous Polymer Sorbent
ACActivated Carbon
CWRCarbon Wide Range
TDUsThermal Desorption Units
OPSIOn-Port Solvent-Assisted Injection
SPDUSorbent Pen™ Desorption Unit

References

  1. Wang, Z.; Walker, G.W.; Muir, D.C.G.; Nagatani-Yoshida, K. Toward a Global Understanding of Chemical Pollution: A First Comprehensive Analysis of National and Regional Chemical Inventories. Environ. Sci. Technol. 2020, 54, 2575–2584. [Google Scholar] [CrossRef] [Scilit]
  2. Du, Z.; Mo, J.; Zhang, Y. Risk Assessment of Population Inhalation Exposure to Volatile Organic Compounds and Carbonyls in Urban China. Environ. Int. 2014, 73, 33–45. [Google Scholar] [CrossRef] [Scilit]
  3. OECD (Ed.) The Economic Consequences of Outdoor Air Pollution; OECD Publishing: Paris, France, 2016. [Google Scholar]
  4. Wild, C.P. Complementing the Genome with an “Exposome”: The Outstanding Challenge of Environmental Exposure Measurement in Molecular Epidemiology. Cancer Epidemiol. Biomark. Prev. 2005, 14, 1847–1850. [Google Scholar] [CrossRef] [Scilit]
  5. Vrijheid, M. The Exposome: A New Paradigm to Study the Impact of Environment on Health. Thorax 2014, 69, 876–878. [Google Scholar] [CrossRef] [Scilit]
  6. Wang, L.K.; Pereira, N.C.; Hung, Y.-T. Air Pollution Control Engineering; Handbook of Environmental Engineering; Humana Press Inc.: Totowa, NJ, USA, 2004; Volume 1. [Google Scholar]
  7. Demeestere, K.; Dewulf, J.; De Witte, B.; Van Langenhove, H. Sample Preparation for the Analysis of Volatile Organic Compounds in Air and Water Matrices. J. Chromatogr. A 2007, 1153, 130–144. [Google Scholar] [CrossRef] [Scilit]
  8. Wojnarowska, M.; Plichta, G.; Sagan, A.; Plichta, J.; Stobiecka, J.; Sołtysik, M. Odour Nuisance and Urban Residents’ Quality of Life: A Case Study in Kraków’s in Plaszow District. Urban Clim. 2020, 34, 100704. [Google Scholar] [CrossRef] [Scilit]
  9. U.S. Environmental Protection Agency-EPA. What Are Hazardous Air Pollutants? U.S. Environmental Protection Agency-EPA: Washington, DC, USA, 2024. [Google Scholar]
  10. U.S. Environmental Protection Agency-EPA. Health Effects Notebook for Hazardous Air Pollutants; U.S. Environmental Protection Agency-EPA: Washington, DC, USA, 2024. [Google Scholar]
  11. Athlete Project 2020 From Womb to Tomb: Tackling Diseases by Knowing the Surrounding Influences 2023. Available online: https://projects.research-and-innovation.ec.europa.eu/en/horizon-magazine/womb-tomb-tackling-diseases-knowing-surrounding-influences (accessed on 2 September 2025).
  12. Vineis, P.; Chadeau-Hyam, M.; Gmuender, H.; Gulliver, J.; Herceg, Z.; Kleinjans, J.; Kogevinas, M.; Kyrtopoulos, S.; Nieuwenhuijsen, M.; Phillips, D.H.; et al. The Exposome in Practice: Design of the EXPOsOMICS Project. Int. J. Hyg. Environ. Health 2017, 220, 142–151. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. HEALS. Health and Environment-Wide Associations Based on Large Population Surveys. The Largest Research Project in Europe on Environment and Health. Available online: https://www.heals-eu.eu/ (accessed on 2 September 2025).
  14. Emory-Rollings school of public health Hercules-Learning How Environmental Exposures Affect Health and Community Well-Being. Available online: https://emoryhercules.com/ (accessed on 2 September 2025).
  15. Anastas, P.T. Green Chemistry and the Role of Analytical Methodology Development. Crit. Rev. Anal. Chem. 1999, 29, 167–175. [Google Scholar] [CrossRef] [Scilit]
  16. Vitale, C.M.; Price, E.J.; Miller, G.W.; David, A.; Antignac, J.-P.; Barouki, R.; Klánová, J. Analytical Strategies for Chemical Exposomics: Exploring Limits and Feasibility. Exposome 2021, 1, osab003. [Google Scholar] [CrossRef] [Scilit]
  17. Shukla, R.; Luo, J.; LeMasters, G.K.; Grinshpun, S.A.; Martuzevicius, D. Sampling over Time: Developing a Cost Effective and Precise Exposure Assessment Program. J. Environ. Monit. 2005, 7, 603. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Price, E.J.; Vitale, C.M.; Miller, G.W.; David, A.; Barouki, R.; Audouze, K.; Walker, D.I.; Antignac, J.-P.; Coumoul, X.; Bessonneau, V.; et al. Merging the Exposome in an Integrated Framework for “Omic” Sciences. iScience 2021, 25, 3. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Gao, P. The Exposome in the Era of One Health. Environ. Sci. Technol. 2021, 55, 2790–2799. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Pawliszyn, J. On-Site Determinations, a “Green” and More Efficient Approach to Analytical Measurement. TrAC Trends Anal. Chem. 2016, 85, 1. [Google Scholar] [CrossRef] [Scilit]
  21. Snyder, D.T.; Pulliam, C.J.; Ouyang, Z.; Cooks, R.G. Miniature and Fieldable Mass Spectrometers: Recent Advances. Anal. Chem. 2016, 88, 2–29. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Lawton, Z.E.; Traub, A.; Fatigante, W.L.; Mancias, J.; O’Leary, A.E.; Hall, S.E.; Wieland, J.R.; Oberacher, H.; Gizzi, M.C.; Mulligan, C.C. Analytical Validation of a Portable Mass Spectrometer Featuring Interchangeable, Ambient Ionization Sources for High Throughput Forensic Evidence Screening. J. Am. Soc. Mass Spectrom. 2017, 28, 1048–1059. [Google Scholar] [CrossRef] [Scilit]
  23. Zhang, X.; Zhang, H.; Yu, K.; Li, N.; Liu, Y.; Liu, X.; Zhang, H.; Yang, B.; Wu, W.; Gao, J.; et al. Rapid Monitoring Approach for Microplastics Using Portable Pyrolysis-Mass Spectrometry. Anal. Chem. 2020, 92, 4656–4662. [Google Scholar] [CrossRef] [Scilit]
  24. Fiorentin, T.R.; Logan, B.K.; Martin, D.M.; Browne, T.; Rieders, E.F. Assessment of a Portable Quadrupole-Based Gas Chromatography Mass Spectrometry for Seized Drug Analysis. Forensic Sci. Int. 2020, 313, 110342. [Google Scholar] [CrossRef] [Scilit]
  25. Wang, W.; Wang, S.; Xu, C.; Li, H.; Xing, Y.; Hou, K.; Li, H. Rapid Screening of Trace Volatile and Nonvolatile Illegal Drugs by Miniature Ion Trap Mass Spectrometry: Synchronized Flash-Thermal-Desorption Purging and Ion Injection. Anal. Chem. 2019, 91, 10212–10220. [Google Scholar] [CrossRef] [Scilit]
  26. Zhou, C.; Wu, H.; Zhang, X.; Zhang, Y.; Xie, W.; Xu, W. High-Throughput and Direct Sample Screening Using a Laser Spray Ionization Miniature Mass Spectrometer. Anal. Chem. 2019, 91, 8808–8813. [Google Scholar] [CrossRef] [Scilit]
  27. Fedick, P.W.; Pu, F.; Morato, N.M.; Cooks, R.G. Identification and Confirmation of Fentanyls on Paper Using Portable Surface Enhanced Raman Spectroscopy and Paper Spray Ionization Mass Spectrometry. J. Am. Soc. Mass Spectrom. 2020, 31, 735–741. [Google Scholar] [CrossRef] [Scilit]
  28. Pulliam, C.J.; Bain, R.M.; Osswald, H.L.; Snyder, D.T.; Fedick, P.W.; Ayrton, S.T.; Flick, T.G.; Cooks, R.G. Simultaneous Online Monitoring of Multiple Reactions Using a Miniature Mass Spectrometer. Anal. Chem. 2017, 89, 6969–6975. [Google Scholar] [CrossRef] [Scilit]
  29. Marshall, J.D.; Apte, J.S.; Coggins, J.S.; Goodkind, A.L. Blue Skies Bluer? Environ. Sci. Technol. 2015, 49, 13929–13936. [Google Scholar] [CrossRef] [Scilit]
  30. Di, Q.; Dai, L.; Wang, Y.; Zanobetti, A.; Choirat, C.; Schwartz, J.D.; Dominici, F. Association of Short-Term Exposure to Air Pollution with Mortality in Older Adults. JAMA 2017, 318, 2446. [Google Scholar] [CrossRef] [Scilit]
  31. Marshall, J.D.; Nethery, E.; Brauer, M. Within-Urban Variability in Ambient Air Pollution: Comparison of Estimation Methods. Atmos. Environ. 2008, 42, 1359–1369. [Google Scholar] [CrossRef] [Scilit]
  32. Morello-Frosch, R.; Pastor, M.; Sadd, J. Environmental Justice and Southern California’s “Riskscape”: The Distribution of Air Toxics Exposures and Health Risks among Diverse Communities. Urban Aff. Rev. 2001, 36, 551–578. [Google Scholar] [CrossRef] [Scilit]
  33. The Agilent Mobile Laboratory for Chemical and Biological Defense 2003. Available online: https://www.agilent.com/Library/technicaloverviews/Public/5988-9152EN.pdf?srsltid=AfmBOoqmERiHXL-15N5AtN6EER-VbC7FjbJj__1pN8A_aDISa_9BXRfi (accessed on 2 September 2025).
  34. EPA Region 6 Mobile Laboratory 2025. Available online: https://www.epa.gov/aboutepa/epa-region-6-mobile-laboratory (accessed on 2 September 2025).
  35. Di, Q.; Wang, Y.; Zanobetti, A.; Wang, Y.; Koutrakis, P.; Choirat, C.; Dominici, F.; Schwartz, J.D. Air Pollution and Mortality in the Medicare Population. N. Engl. J. Med. 2017, 376, 2513–2522. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. Carrothers, T.J.; Evans, J.S. Assessing the Impact of Differential Measurement Error on Estimates of Fine Particle Mortality. J. Air Waste Manag. Assoc. 2000, 50, 65–74. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Zeger, S.L.; Thomas, D.; Dominici, F.; Samet, J.M.; Schwartz, J.; Dockery, D.; Cohen, A. Exposure Measurement Error in Time-Series Studies of Air Pollution: Concepts and Consequences. Environ. Health Perspect. 2000, 108, 419–426. [Google Scholar] [CrossRef]
  38. Goldman, G.T.; Mulholland, J.A.; Russell, A.G.; Srivastava, A.; Strickland, M.J.; Klein, M.; Waller, L.A.; Tolbert, P.E.; Edgerton, E.S. Ambient Air Pollutant Measurement Error: Characterization and Impacts in a Time-Series Epidemiologic Study in Atlanta. Environ. Sci. Technol. 2010, 44, 7692–7698. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Goldman, G.T.; Mulholland, J.A.; Russell, A.G.; Gass, K.; Strickland, M.J.; Tolbert, P.E. Characterization of Ambient Air Pollution Measurement Error in a Time-Series Health Study Using a Geostatistical Simulation Approach. Atmos. Environ. 2012, 57, 101–108. [Google Scholar] [CrossRef] [Scilit]
  40. Apte, J.S.; Messier, K.P.; Gani, S.; Brauer, M.; Kirchstetter, T.W.; Lunden, M.M.; Marshall, J.D.; Portier, C.J.; Vermeulen, R.C.H.; Hamburg, S.P. High-Resolution Air Pollution Mapping with Google Street View Cars: Exploiting Big Data. Environ. Sci. Technol. 2017, 51, 6999–7008. [Google Scholar] [CrossRef] [Scilit]
  41. Snyder, E.G.; Watkins, T.H.; Solomon, P.A.; Thoma, E.D.; Williams, R.W.; Hagler, G.S.W.; Shelow, D.; Hindin, D.A.; Kilaru, V.J.; Preuss, P.W. The Changing Paradigm of Air Pollution Monitoring. Environ. Sci. Technol. 2013, 47, 11369–11377. [Google Scholar] [CrossRef] [Scilit]
  42. Hasenfratz, D.; Saukh, O.; Walser, C.; Hueglin, C.; Fierz, M.; Arn, T.; Beutel, J.; Thiele, L. Deriving High-Resolution Urban Air Pollution Maps Using Mobile Sensor Nodes. Pervasive Mob. Comput. 2015, 16, 268–285. [Google Scholar] [CrossRef] [Scilit]
  43. VoPham, T.; Bertrand, K.A.; Jones, R.R.; Deziel, N.C.; DuPré, N.C.; James, P.; Liu, Y.; Vieira, V.M.; Tamimi, R.M.; Hart, J.E.; et al. Dioxin Exposure and Breast Cancer Risk in a Prospective Cohort Study. Environ. Res. 2020, 186, 109516. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Kumar, A.; Flewitt, A.J. An Approach to Simultaneously Test Multiple Devices for High-Throughput Production of Thin-Film Electronics. J. Disp. Technol. 2016, 12, 240–246. [Google Scholar] [CrossRef] [Scilit]
  45. Steffens, N.K.; Haslam, S.A.; Schuh, S.C.; Jetten, J.; Van Dick, R. A Meta-Analytic Review of Social Identification and Health in Organizational Contexts. Personal. Soc. Psychol. Rev. 2017, 21, 303–335. [Google Scholar] [CrossRef] [Scilit]
  46. Lan, H. Development of Materials and Methodologies for Microextraction Techniques; Department of Chemistry, Faculty of Science, University of Helsinki: Helsinki, Finland, 2019. [Google Scholar]
  47. Dugheri, S.; Mucci, N.; Cappelli, G.; Trevisani, L.; Bonari, A.; Bucaletti, E.; Squillaci, D.; Arcangeli, G. Advanced Solid-Phase Microextraction Techniques and Related Automation: A Review of Commercially Available Technologies. J. Anal. Methods Chem. 2022, 2022, 8690569. [Google Scholar] [CrossRef] [Scilit]
  48. Nerín, C.; Salafranca, J.; Aznar, M.; Batlle, R. Critical Review on Recent Developments in Solventless Techniques for Extraction of Analytes. Anal. Bioanal. Chem. 2009, 393, 809–833. [Google Scholar] [CrossRef] [Scilit]
  49. Mehdinia, A.; Aziz-Zanjani, M.O. Advances for Sensitive, Rapid and Selective Extraction in Different Configurations of Solid-Phase Microextraction. TrAC Trends Anal. Chem. 2013, 51, 13–22. [Google Scholar] [CrossRef] [Scilit]
  50. Poole, C.; Mester, Z.; Miró, M.; Pedersen-Bjergaard, S.; Pawliszyn, J. Glossary of Terms Used in Extraction (IUPAC Recommendations 2016). Chem. Int. 2016, 38, 26. [Google Scholar] [CrossRef] [Scilit]
  51. Kokate, P.; Middey, A.; Sadistap, S.; Sarode, G.; Narayan, A. Review on Drone-Assisted Air-Quality Monitoring Systems. Drones Auton. Veh. 2024, 1, 10005. [Google Scholar] [CrossRef] [Scilit]
  52. Okeme, J.O.; Koelmel, J.P.; Johnson, E.; Lin, E.Z.; Dong, G.; Pollitt, K.J.G. Wearable Passive Samplers for Assessing Envi-ronmental Exposure to Organic Chemicals: Current Approaches and Future Directions. Curr. Environ. Health Rep. 2023, 10, 84–98. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  53. Wilcockson, J.B.; Gobas, F.A.P.C. Thin-Film Solid-Phase Extraction to Measure Fugacities of Organic Chemicals with Low Volatility in Biological Samples. Environ. Sci. Technol. 2001, 35, 1425–1431. [Google Scholar] [CrossRef] [Scilit]
  54. Bruheim, I.; Liu, X.; Pawliszyn, J. Thin-Film Microextraction. Anal. Chem. 2003, 75, 1002–1010. [Google Scholar] [CrossRef] [Scilit]
  55. Murtada, K.; Galpin, V.; Grandy, J.J.; Singh, V.; Sanchez, F.; Pawliszyn, J. Development of Porous Carbon/Polydimethylsiloxane Thin-Film Solid-Phase Microextraction Membranes to Facilitate on-Site Sampling of Volatile Organic Compounds. Sustain. Chem. Pharm. 2021, 21, 100435. [Google Scholar] [CrossRef] [Scilit]
  56. Olcer, Y.A.; Tascon, M.; Eroglu, A.E.; Boyacı, E. Thin Film Microextraction: Towards Faster and More Sensitive Microextraction. TrAC Trends Anal. Chem. 2019, 113, 93–101. [Google Scholar] [CrossRef] [Scilit]
  57. Yu, L.-D.; Tong, Y.-J.; Gong, X.; Lan, B.; Zhu, F.; Ouyang, G. Remote Sampling of Persistent Organic Pollutants by a Home-Made Thin Film Device. Anal. Chim. Acta 2025, 1334, 343422. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  58. Ahmadi, F.; Sparham, C.; Boyacı, E.; Pawliszyn, J. Time Weighted Average Concentration Monitoring Based on Thin Film Solid Phase Microextraction. Environ. Sci. Technol. 2017, 51, 3929–3937. [Google Scholar] [CrossRef] [Scilit]
  59. Grandy, J.J.; Singh, V.; Lashgari, M.; Gauthier, M.; Pawliszyn, J. Development of a Hydrophilic Lipophilic Balanced Thin Film Solid Phase Microextraction Device for Balanced Determination of Volatile Organic Compounds. Anal. Chem. 2018, 90, 14072–14080. [Google Scholar] [CrossRef] [Scilit]
  60. Quintanilla, I.; Perelló, C.; Merlo, F.; Profumo, A.; Fontàs, C.; Anticó, E. Multiwalled Carbon Nanotubes Embedded in a Polymeric Matrix as a New Material for Thin Film Microextraction (TFME) in Organic Pollutant Monitoring. Polymers 2023, 15, 314. [Google Scholar] [CrossRef] [Scilit]
  61. Wheatley, J.; Murtada, K.; McGregor, L.; Edwards, M.; Pawliszyn, J. Monitoring Disinfection By-Products in Water Using Thin-Film, SPME with GCxGC-TOF MS. In Proceedings of the SepSolve Analytical; SepSolve Analytical: Peterborough, UK, 2022. [Google Scholar]
  62. Poojary, H.; Das, P.P.; Koo, S.; Ghosh, C. Development of a Paper-Based Disposal Thin-Film Solid-Phase Microextraction Tool for the Quantification of Environmentally Hazardous 4-Chlorophenol in Water. ACS Omega 2025, 10, 1657–1666. [Google Scholar] [CrossRef] [Scilit]
  63. Guerra, P.; Lai, H.; Almirall, J.R. Analysis of the Volatile Chemical Markers of Explosives Using Novel Solid Phase Microextraction Coupled to Ion Mobility Spectrometry. J. Sep. Sci. 2008, 31, 2891–2898. [Google Scholar] [CrossRef] [Scilit]
  64. Deng, J.; Yang, Y.; Wang, X.; Luan, T. Strategies for Coupling Solid-Phase Microextraction with Mass Spectrometry. TrAC Trends Anal. Chem. 2014, 55, 55–67. [Google Scholar] [CrossRef] [Scilit]
  65. Mirnaghi, F.S.; Pawliszyn, J. Reusable Solid-Phase Microextraction Coating for Direct Immersion Whole-Blood Analysis and Extracted Blood Spot Sampling Coupled with Liquid Chromatography–Tandem Mass Spectrometry and Direct Analysis in Real Time–Tandem Mass Spectrometry. Anal. Chem. 2012, 84, 8301–8309. [Google Scholar] [CrossRef] [Scilit]
  66. Rodriguez-Lafuente, A.; Mirnaghi, F.S.; Pawliszyn, J. Determination of Cocaine and Methadone in Urine Samples by Thin-Film Solid-Phase Microextraction and Direct Analysis in Real Time (DART) Coupled with Tandem Mass Spectrometry. Anal. Bioanal. Chem. 2013, 405, 9723–9727. [Google Scholar] [CrossRef] [Scilit]
  67. Gómez-Ríos, G.A.; Pawliszyn, J. Development of Coated Blade Spray Ionization Mass Spectrometry for the Quantitation of Target Analytes Present in Complex Matrices. Angew. Chem. Int. Ed. 2014, 53, 14503–14507. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  68. Interchim. OPSI/CMS Analysis: Direct Screening of Cocaine on Paper Currency; Interchim: Montluçon, France, 2021. [Google Scholar]
  69. Hu, K.; Zhou, W.; Yang, C.; Wang, Y.; Jiang, R.W.; Zhang, Z.; Pawliszyn, J. Rapid screening of drugs in environmental water using metal organic framework/Ti3C2Tx composite coated blade spray-mass spectrometry. J. Hazard. Mater. 2024, 472, 134609. [Google Scholar] [CrossRef] [Scilit]
  70. Yu, M.; Roszkowska, A.; Pawliszyn, J. In Vivo Solid-Phase Microextraction and Applications in Environmental Sciences. ACS Environ. Au 2022, 2, 30–41. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  71. Ruiz-Jimenez, J.; Lan, H.; Leleev, Y.; Hartonen, K.; Riekkola, M.-L. Comparison of Multiple Calibration Approaches for the Determination of Volatile Organic Compounds in Air Samples by Solid Phase Microextraction Arrow and In-Tube Extraction. J. Chromatogr. A 2020, 1616, 460825. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  72. Musshoff, F.; Lachenmeier, D.W.; Kroener, L.; Madea, B. Automated Headspace Solid-Phase Dynamic Extraction for the Determination of Amphetamines and Synthetic Designer Drugs in Hair Samples. J. Chromatogr. A 2002, 958, 231–238. [Google Scholar] [CrossRef] [Scilit]
  73. Lan, H.; Hartonen, K.; Riekkola, M.-L. Miniaturised Air Sampling Techniques for Analysis of Volatile Organic Compounds in Air. TrAC Trends Anal. Chem. 2020, 126, 115873. [Google Scholar] [CrossRef] [Scilit]
  74. Kędziora-Koch, K.; Wasiak, W. Needle-based extraction techniques with protected sorbent as powerful sample preparation tools to gas chromatographic analysis: Trends in application. J. Chromatogr. A 2018, 1565, 1–18. [Google Scholar] [CrossRef] [Scilit]
  75. Jochmann, M.A.; Kmiecik, M.P.; Schmidt, T.C. Solid-phase dynamic extraction for the enrichment of polar volatile organic compounds from water. J. Chromatogr. A 2006, 1115, 208–216. [Google Scholar] [CrossRef] [Scilit]
  76. Poormohammadi, A.; Bahrami, A.; Farhadian, M.; Ghorbani Shahna, F.; Ghiasvand, A. Development of Carbotrap B-Packed Needle Trap Device for Determination of Volatile Organic Compounds in Air. J. Chromatogr. A 2017, 1527, 33–42. [Google Scholar] [CrossRef] [Scilit]
  77. Ghalichi Zave, Z.; Bahrami, A.; Ghorbani Shahna, F.; Farhadian, M. Application of a Needle Trap Device Packed with XAD-2 Polyaniline Composite for Sampling Naphthalene and Phenanthrene in Air. J. Chromatogr. A 2019, 1602, 74–82. [Google Scholar] [CrossRef] [Scilit]
  78. Maleki, S.; Hashemi, P.; Rasolzadeh, F.; Maleki, S.; Ghiasvand, A.R. A Needle Trap Device Packed with Nanoporous Silica Sorbents for Separation and Gas Chromatographic Determination of Polycyclic Aromatic Hydrocarbons in Contaminated Soils. J. Chromatogr. Sci. 2018, 56, 771–778. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  79. Asl-Hariri, S.; Gómez-Ríos, G.A.; Gionfriddo, E.; Dawes, P.; Pawliszyn, J. Development of Needle Trap Technology for On-Site Determinations: Active and Passive Sampling. Anal. Chem. 2014, 86, 5889–5897. [Google Scholar] [CrossRef] [Scilit]
  80. Barreira, L.M.F.; Xue, Y.; Duporté, G.; Parshintsev, J.; Hartonen, K.; Jussila, M.; Kulmala, M.; Riekkola, M.-L. Potential of Needle Trap Microextraction–Portable Gas Chromatography–Mass Spectrometry for Measurement of Atmospheric Volatile Compounds. Atmos. Meas. Tech. 2016, 9, 3661–3671. [Google Scholar] [CrossRef] [Scilit]
  81. Iwai, M.; Kondo, F.; Suzuki, T.; Ogawa, T.; Seno, H. Quantification of Ethanol in Whole Blood by Extraction Using NeedlEx® and Gas Chromatography/Mass Spectrometry. Med. Mass Spectrom. 2019, 3, 30–34. [Google Scholar]
  82. NeeedlEX: Specification and Prices. Available online: https://shinwa-cpc.co.jp/en/products/needlex/ (accessed on 2 September 2025).
  83. Bowerbank, S.; Romano, A. In-Tube Extraction Dynamic Headspace (ITEX-DHS) Sampling Technique Coupled to GC-MS for Sensitive Determination of Odorants in Water. Thermo Fish. Sci. 2020. [Google Scholar]
  84. Laaks, J.; Jochmann, M.A.; Schilling, B.; Schmidt, T.C. Optimization Strategies of In-Tube Extraction (ITEX) Methods. Anal. Bioanal. Chem. 2015, 407, 6827–6838. [Google Scholar] [CrossRef] [Scilit]
  85. Lan, H.; Holopainen, J.; Hartonen, K.; Jussila, M.; Ritala, M.; Riekkola, M.-L. Fully Automated Online Dynamic In-Tube Extraction for Continuous Sampling of Volatile Organic Compounds in Air. Anal. Chem. 2019, 91, 8507–8515. [Google Scholar] [CrossRef] [Scilit]
  86. Jochmann, M.A.; Yuan, X.; Schilling, B.; Schmidt, T.C. In-Tube Extraction for Enrichment of Volatile Organic Hydrocarbons from Aqueous Samples. J. Chromatogr. A 2008, 1179, 96–105. [Google Scholar] [CrossRef] [Scilit]
  87. Audette, E.; Dumont, P.-L.; Crispino, F. PowerSorb® for Forensic Investigation of VOC Traces: Application on Perfume Traces. Forensic Sci. Int. 2025, 376, 112605. [Google Scholar] [CrossRef] [Scilit]
  88. Cuzuel, V.; Portas, E.; Cognon, G.; Rivals, I.; Heulard, F.; Thiébaut, D.; Vial, J. Sampling Method Development and Optimization in View of Human Hand Odor Analysis by Thermal Desorption Coupled with Gas Chromatography and Mass Spectrometry. Anal. Bioanal. Chem. 2017, 409, 5113–5124. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  89. Cuzuel, V.; Leconte, R.; Cognon, G.; Thiébaut, D.; Vial, J.; Sauleau, C.; Rivals, I. Human Odor and Forensics: Towards Bayesian Suspect Identification Using GC × GC–MS Characterization of Hand Odor. J. Chromatogr. B 2018, 1092, 379–385. [Google Scholar] [CrossRef] [Scilit]
  90. Duemichen, E.; Eisentraut, P.; Celina, M.; Braun, U. Automated Thermal Extraction-Desorption Gas Chromatography Mass Spectrometry: A Multifunctional Tool for Comprehensive Characterization of Polymers and Their Degradation Products. J. Chromatogr. A 2019, 1592, 133–142. [Google Scholar] [CrossRef] [Scilit]
  91. Müller, A.; Kocher, B.; Altmann, K.; Braun, U. Determination of Tire Wear Markers in Soil Samples and Their Distribution in a Roadside Soil. Chemosphere 2022, 294, 133653. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  92. Leemans, M.; Cuzuel, V.; Bauër, P.; Baba Aissa, H.; Cournelle, G.; Baelde, A.; Thuleau, A.; Cognon, G.; Pouget, N.; Guillot, E.; et al. Screening of Breast Cancer from Sweat Samples Analyzed by 2-Dimensional Gas Chromatography-Mass Spectrometry: A Preliminary Study. Cancers 2023, 15, 2939. [Google Scholar] [CrossRef] [Scilit]
  93. Getxent Blank Getxent Tubes. Available online: https://getxent.com/products/getxent-tubes-x50?srsltid=AfmBOop0RDkQIroaIpOfqnbUMrCbZmUjnAiJAre8Yvty4GU_1oE--EUV (accessed on 2 September 2025).
  94. Boudard, E.; Fisson, L.; Moumane, N.; Dugay, J.; Vial, J.; Thiébaut, D. Study of Sampling Phases for Body Odor Sampling Prior to Analysis by TD-GC × GC/ToFMS. Anal. Bioanal. Chem. 2025, 417, 3177–3190. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  95. Sandra, P.; Baltussen, E.; David, F. Method for Solid-Phase Microextraction and Analysis, and a Collector for This Method. U.S. Patent No. 6,815,216, 9 November 2004. [Google Scholar]
  96. Baltussen, E.; Sandra, P.; David, F.; Cramers, C. Stir Bar Sorptive Extraction (SBSE), a Novel Extraction Technique for Aqueous Samples: Theory and Principles. J. Microcolumn Sep. 1999, 11, 737–747. [Google Scholar] [CrossRef] [Scilit]
  97. Hasan, C.K.; Ghiasvand, A.; Lewis, T.W.; Nesterenko, P.N.; Paull, B. Recent Advances in Stir-Bar Sorptive Extraction: Coatings, Technical Improvements, and Applications. Anal. Chim. Acta 2020, 1139, 222–240. [Google Scholar] [CrossRef] [Scilit]
  98. Soini, H.A.; Bruce, K.E.; Wiesler, D.; David, F.; Sandra, P.; Novotny, M.V. Stir Bar Sorptive Extraction: A New Quantitative and Comprehensive Sampling Technique for Determination of Chemical Signal Profiles from Biological Media. J. Chem. Ecol. 2005, 31, 377–392. [Google Scholar] [CrossRef] [Scilit]
  99. David, F.; Ochiai, N.; Sandra, P. Two Decades of Stir Bar Sorptive Extraction: A Retrospective and Future Outlook. TrAC Trends Anal. Chem. 2019, 112, 102–111. [Google Scholar] [CrossRef] [Scilit]
  100. Zhu, L.; Xu, G.; Wei, F.; Yang, J.; Hu, Q. Determination of Melamine in Powdered Milk by Molecularly Imprinted Stir Bar Sorptive Extraction Coupled with HPLC. J. Colloid Interface Sci. 2015, 454, 8–13. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  101. Dugheri, S.; Cappelli, G.; Fanfani, N.; Squillaci, D.; Rapi, I.; Venturini, L.; Vita, C.; Gori, R.; Sirini, P.; Cipriano, D.; et al. Vacuum-Assisted MonoTrapTM Extraction for Volatile Organic Compounds (VOCs) Profiling from Hot Mix Asphalt. Molecules 2024, 29, 4943. [Google Scholar] [CrossRef] [Scilit]
  102. Jang, H.-J.; Son, H.-H.; Lee, D.-S. Optimization of Disk Sorptive Extraction Based on Monolithic Material for the Determination of Aroma Compounds from Lantana camara L. by Gas Chromatography-Mass Spectrometry. Bull. Korean Chem. Soc. 2011, 32, 4275–4280. [Google Scholar] [CrossRef] [Scilit]
  103. Hirose, N.; Maeda, G.; Miyagi, K.; Wada, K.; Ohta, H. Simplified Analysis of Flavor in Shiikuwasha Products Using Monolithic Silica Adsorbents “MonoTrap” and GC-MS. Shimadzu Tech. Rep. 2017, 146, 1–4. [Google Scholar]
  104. Ding, L.; Wang, L.; Nian, L.; Tang, M.; Yuan, R.; Shi, A.; Shi, M.; Han, Y.; Liu, M.; Zhang, Y.; et al. Non-Targeted Screening of Volatile Organic Compounds in a Museum in China Using GC-Orbitrap Mass Spectrometry. Sci. Total Environ. 2022, 835, 155277. [Google Scholar] [CrossRef] [Scilit]
  105. Ahmad, S.M.; Calado, B.B.C.; Oliveira, M.N.; Neng, N.R.; Nogueira, J.M.F. Bar Adsorptive Microextraction Coated with Carbon-Based Phase Mixtures for Performance-Enhancement to Monitor Selected Benzotriazoles, Benzothiazoles, and Benzenesulfonamides in Environmental Water Matrices. Molecules 2020, 25, 2133. [Google Scholar] [CrossRef] [Scilit]
  106. Frontier Lab Magic Chemisorber PEG (MC-PEG). Available online: https://www.frontier-lab.com/products/multi-functional-pyrolysis-system/17867/ (accessed on 2 September 2025).
  107. Wang, L.; Hosaka, A.; Watanabe, C.; Ohtani, H.; Tsuge, S. Development of a Novel Solid-Phase Extraction Element for Thermal Desorption Gas Chromatography Analysis. J. Chromatogr. A 2004, 1035, 277–279. [Google Scholar] [CrossRef] [Scilit]
  108. Murtada, K. Trends in Nanomaterial-Based Solid-Phase Microextraction with a Focus on Environmental Applications—A Review. Trends Environ. Anal. Chem. 2020, 25, e00077. [Google Scholar] [CrossRef] [Scilit]
  109. Kenessov, B.; Koziel, J.A.; Bakaikina, N.V.; Orazbayeva, D. Perspectives and Challenges of On-Site Quantification of Organic Pollutants in Soils Using Solid-Phase Microextraction. TrAC Trends Anal. Chem. 2016, 85, 111–122. [Google Scholar] [CrossRef] [Scilit]
  110. Pacenti, M.; Dugheri, S.; Gagliano-Candela, R.; Strisciullo, G.; Franchi, E.; Degli Esposti, F.; Perchiazzi, N.; Boccalon, P.; Arcangeli, G.; Cupelli, V. Analysis of 2-Chloroacetophenone in Air by Multi-Fiber Solid-Phase Microextraction and Fast Gas Chromatography-Mass Spectrometry. Acta Chromatogr. 2009, 21, 379–397. [Google Scholar] [CrossRef] [Scilit]
  111. Pacenti, M.; Lofrumento, C.; Dugheri, S.; Zoppi, A.; Borsi, I.; Speranza, A.; Boccalon, P.; Arcangeli, G.; Antoniucci, A.; Castellucci, E.M.; et al. Physicochemical Characterization of Exhaust Particulates from Gasoline and Diesel Engines by Solid-Phase Micro Extraction Sampling and Combined Raman Microspectroscopic/Fast Gas-Chromotography Mass Spectrometry Analysis. Eur. J. Inflamm. 2009, 7, 25–37. [Google Scholar] [CrossRef] [Scilit]
  112. Duan, C.; Shen, Z.; Wu, D.; Guan, Y. Recent Developments in Solid-Phase Microextraction for on-Site Sampling and Sample Preparation. TrAC Trends Anal. Chem. 2011, 30, 1568–1574. [Google Scholar] [CrossRef] [Scilit]
  113. Dugheri, S.; Bonari, A.; Pompilio, I.; Colpo, M.; Montalti, M.; Mucci, N.; Arcangeli, G. Assessment of Occupational Exposure to Gaseous Peracetic Acid. Int. J. Occup. Med. Environ. Health 2018, 31, 527–535. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  114. Larroque, V.; Desauziers, V.; Mocho, P. Development of a Solid Phase Microextraction (SPME) Method for the Sampling of VOC Traces in Indoor Air. J. Environ. Monit. 2006, 8, 106–111. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  115. Chen, J.; Kuang, Y.; Feng, X.; Mao, C.; Zheng, J.; Ouyang, G. Solid Phase Microextraction Coupled with Portable Gas Chromatography-Mass Spectrometry for on-Site Analysis. Green Anal. Chem. 2024, 11, 100164. [Google Scholar] [CrossRef] [Scilit]
  116. Dugheri, S.; Cappelli, G.; Rapi, I.; Gori, R.; Venturini, L.; Fanfani, N.; Vita, C.; Guerriero, E.; Cipriano, D.; Bartolucci, G.L.; et al. Rapid On-Field Monitoring for Odor-Active Homologous Aliphatic Aldehydes and Ketones from Hot Mix Asphalt Emission by Dynamic-SPME Air Sampling with Online Gas Chromatographic Analysis. Molecules 2025, 30, 3545. [Google Scholar] [CrossRef] [Scilit]
  117. Herrington, J.S.; Gómez-Ríos, G.A.; Myers, C.; Stidsen, G.; Bell, D.S. Hunting Molecules in Complex Matrices with SPME Arrows: A Review. Separations 2020, 7, 12. [Google Scholar] [CrossRef] [Scilit]
  118. Brown, R.W.; Mayser, J.P.; Widdowson, C.; Chadwick, D.R.; Jones, D.L. Dependence of Thermal Desorption Method for Profiling Volatile Organic Compound (VOC) Emissions from Soil. Soil Biol. Biochem. 2021, 160, 108313. [Google Scholar] [CrossRef] [Scilit]
  119. Butler, K.; Johnston, N.; Hebdon, K.; Anderson, A.; Pace, A.; Skibiel, A. Measuring Air Toxic Gases in Dairy Cows Exposed to Wildfire Smoke. In Proceedings of the Idaho Conference on Undergraduate Research, Boise, ID, USA, 16–17 July 2025. [Google Scholar]
  120. Kaikiti, K.; Omirou, M.; Ioannides, I.M.; Agapiou, A. Mapping Soil VOCs Using Three Green Sample Extraction Techniques and GC–MS. Microchem. J. 2024, 207, 111851. [Google Scholar] [CrossRef] [Scilit]
  121. Elia, S.; Stylianou, M.; Agapiou, A. Advanced Micro-Extraction Techniques (SPME, HiSorb) for the Determination of Goat Cheese Whey Wastewater VOCs. J. Environ. Manag. 2024, 351, 119934. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  122. Hou, X.; Wang, L.; Guo, Y. Recent Developments in Solid-Phase Microextraction Coatings for Environmental and Biological Analysis. Chem. Lett. 2017, 46, 1444–1455. [Google Scholar] [CrossRef] [Scilit]
  123. Gorecki, T.; Pawliszyn, J. Field-Portable Solid-Phase Microextraction/Fast GC System for Trace Analysis. Field Anal. Chem. Technol. 1997, 1, 277–284. [Google Scholar] [CrossRef] [Scilit]
  124. Koziel, J.; Jia, M.; Khaled, A.; Noah, J.; Pawliszyn, J. Field Air Analysis with SPME Device. Anal. Chim. Acta 1999, 400, 153–162. [Google Scholar] [CrossRef] [Scilit]
  125. Koziel, J.A.; Noah, J.; Pawliszyn, J. Field Sampling and Determination of Formaldehyde in Indoor Air with Solid-Phase Microextraction and On-Fiber Derivatization. Environ. Sci. Technol. 2001, 35, 1481–1486. [Google Scholar] [CrossRef] [Scilit]
  126. Jia, M.; Koziel, J.; Pawliszyn, J. Fast Field Sampling/Sample Preparation and Quantification of Volatile Organic Compounds in Indoor Air by Solid-Phase Microextraction and Portable Gas Chromatography. Field Anal. Chem. Technol. 2000, 4, 73–84. [Google Scholar] [CrossRef] [Scilit]
  127. Strano, M.; Ricci, D.; Pigro, M.C.; Martellucci, L.; Gaudio, P.; Malizia, A.; Di Giovanni, D. Evaluation of an Innovative Proposal for the Integration of Chemical Sensors with SPME Fibers on UAVs, for Forensic, Military, and Civil Applications. Eur. Phys. J. Plus 2025, 140, 276. [Google Scholar] [CrossRef] [Scilit]
  128. Smith, P.A.; Kluchinsky, T.A.; Savage, P.B.; Erickson, R.P.; Lee, A.P.; Williams, K.; Stevens, M.; Thomas, R.J. Traditional Sampling with Laboratory Analysis and Solid Phase Microextraction Sampling with Field Gas Chromatography/Mass Spectrometry by Military Industrial Hygienists. AIHA J. 2002, 63, 284–292. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  129. Mordor Intelligence. Portable Chromatography Systems Market Size & Share Analysis—Growth Trends & Forecasts (2025–2030). Available online: https://www.mordorintelligence.com/industry-reports/portable-chromatography-systems-market (accessed on 2 September 2025).
  130. Zou, J.; Sai, T.; Duan, S.; Winniford, B.; Zhang, D. Automated Method for Short-Chain Aldehydes Emission Measurement by Dynamic Solid-Phase Microextraction on-Fiber Derivatization GC-MSD Coupled with a Flow-Cell. J. Chromatogr. A 2022, 1671, 462996. [Google Scholar] [CrossRef] [Scilit]
  131. UNI EN 13725; Air quality—Stationary Source Emissions—Determination of Odour Concentration by Dynamic Olfactometry. Versione 2022. Comite Europe en de Normalisation (CEN): Brussels, Belgium, 2022.
  132. Romain, A.C.; Fays, S. Odour Sampling System with Remote Triggering: Feedback from a Belgian Experience. In Proceedings of the 7th International Conference on Environmental Odour Monitoring & Control, Taormina, Italy, 19–21 April 2021. [Google Scholar]
  133. Negi, I.; Tsow, F.; Tanwar, K.; Zhang, L.; Iglesias, R.A.; Chen, C.; Rai, A.; Forzani, E.S.; Tao, N. Novel Monitor Paradigm for Real-Time Exposure Assessment. J. Expo. Sci. Environ. Epidemiol. 2011, 21, 419–426. [Google Scholar] [CrossRef] [Scilit]
  134. Air & Waste Management Association. Field Analytical Methods for Hazardous Wastes and Toxic Chemicals: Proceedings of a Specialty Conference; VIP (Series); Air & Waste Management Association: Pittsburgh, PA, USA, 1997. [Google Scholar]
  135. Toscano, P.; Gioli, B.; Dugheri, S.; Salvini, A.; Matese, A.; Bonacchi, A.; Zaldei, A.; Cupelli, V.; Miglietta, F. Locating Industrial VOC Sources with Aircraft Observations. Environ. Pollut. 2011, 159, 1174–1182. [Google Scholar] [CrossRef] [Scilit]
  136. Grandy, J.J.; Galpin, V.; Singh, V.; Pawliszyn, J. Development of a drone-based thin-film solid-phase microextraction water sampler to facilitate on-site screening of environmental pollutants. Anal. Chem. 2020, 92, 12917–12924. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  137. Bianchi, F.; Bisceglie, F.; Dugheri, S.; Arcangeli, G.; Cupelli, V.; Del Borrello, E.; Sidisky, L.; Careri, M. Ionic Liquid-Based Solid Phase Microextraction Necklaces for the Environmental Monitoring of Ketamine. J. Chromatogr. A 2014, 1331, 1–9. [Google Scholar] [CrossRef] [Scilit]
  138. Lin, E.Z.; Esenther, S.; Mascelloni, M.; Irfan, F.; Godri Pollitt, K.J. The Fresh Air Wristband: A Wearable Air Pollutant Sampler. Environ. Sci. Technol. Lett. 2020, 7, 308–314. [Google Scholar] [CrossRef] [Scilit]
  139. Guo, P.; Lin, E.Z.; Koelmel, J.P.; Ding, E.; Gao, Y.; Deng, F.; Dong, H.; Liu, Y.; Cha, Y.; Fang, J.; et al. Exploring Personal Chemical Exposures in China with Wearable Air Pollutant Monitors: A Repeated-Measure Study in Healthy Older Adults in Jinan, China. Environ. Int. 2021, 156, 106709. [Google Scholar] [CrossRef] [Scilit]
  140. Koelmel, J.P.; Lin, E.Z.; DeLay, K.; Williams, A.J.; Zhou, Y.; Bornman, R.; Obida, M.; Chevrier, J.; Pollitt, K. Assessing the External Exposome of South African Children Using Wearable Passive Samplers and High-Resolution Mass Spectrometry. Environ. Sci. Technol. 2021, 56, 2191–2203. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  141. Fuentes, Z.C.; Schwartz, Y.L.; Robuck, A.R.; Walker, D.I. Operationalizing the Exposome Using Passive Silicone Samplers. Curr. Pollut. Rep. 2022, 8, 1–29. [Google Scholar] [CrossRef] [Scilit]
  142. Helbig, C.; Ueberham, M.; Becker, A.M.; Marquart, H.; Schlink, U. Wearable Sensors for Human Environmental Exposure in Urban Settings. Curr. Pollut. Rep. 2021, 7, 417–433. [Google Scholar] [CrossRef] [Scilit]
  143. Sharma, S.; Tolley, L.T.; Tolley, H.D.; Plistil, A.; Stearns, S.D.; Lee, M.L. Hand-Portable Liquid Chromatographic Instrumentation. J. Chromatogr. A 2015, 1421, 38–47. [Google Scholar] [CrossRef] [Scilit]
  144. Peacock, P.M.; Zhang, W.-J.; Trimpin, S. Advances in Ionization for Mass Spectrometry. Anal. Chem. 2017, 89, 372–388. [Google Scholar] [CrossRef] [Scilit]
  145. Takáts, Z.; Wiseman, J.M.; Gologan, B.; Cooks, R.G. Mass Spectrometry Sampling Under Ambient Conditions with Desorption Electrospray Ionization. Science 2004, 306, 471–473. [Google Scholar] [CrossRef] [Scilit]
  146. Harper, J.D.; Charipar, N.A.; Mulligan, C.C.; Zhang, X.; Cooks, R.G.; Ouyang, Z. Low-Temperature Plasma Probe for Ambient Desorption Ionization. Anal. Chem. 2008, 80, 9097–9104. [Google Scholar] [CrossRef] [Scilit]
  147. Jecklin, M.C.; Schmid, S.; Urban, P.L.; Amantonico, A.; Zenobi, R. Miniature Flowing Atmospheric-pressure Afterglow Ion Source for Facile Interfacing of CE with MS. Electrophoresis 2010, 31, 3597–3605. [Google Scholar] [CrossRef] [Scilit]
  148. D’Agostino, P.A.; Chenier, C.L.; Hancock, J.R.; Lepage, C.R.J. Desorption Electrospray Ionisation Mass Spectrometric Analysis of Chemical Warfare Agents from Solid-phase Microextraction Fibers. Rapid Commun. Mass Spectrom. 2007, 21, 543–549. [Google Scholar] [CrossRef] [Scilit]
  149. Cajka, T.; Riddellova, K.; Tomaniova, M.; Hajslova, J. Recognition of Beer Brand Based on Multivariate Analysis of Volatile Fingerprint. J. Chromatogr. A 2010, 1217, 4195–4203. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  150. D’Agostino, P.A.; Hancock, J.R.; Chenier, C.L.; Lepage, C.R.J. Liquid Chromatography Electrospray Tandem Mass Spectrometric and Desorption Electrospray Ionization Tandem Mass Spectrometric Analysis of Chemical Warfare Agents in Office Media Typically Collected during a Forensic Investigation. J. Chromatogr. A 2006, 1110, 86–94. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  151. Haunschmidt, M.; Klampfl, C.W.; Buchberger, W.; Hertsens, R. Determination of Organic UV Filters in Water by Stir Bar Sorptive Extraction and Direct Analysis in Real-Time Mass Spectrometry. Anal. Bioanal. Chem. 2010, 397, 269–275. [Google Scholar] [CrossRef] [Scilit]
  152. Bridoux, M.C.; Malandain, H.; Leprince, F.; Progent, F.; Machuron-Mandard, X. Quantitative Analysis of Phosphoric Acid Esters in Aqueous Samples by Isotope Dilution Stir-Bar Sorptive Extraction Combined with Direct Analysis in Real Time (DART)-Orbitrap Mass Spectrometry. Anal. Chim. Acta 2015, 869, 1–10. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  153. Mielczarek, P.; Silberring, J.; Smoluch, M. Miniaturization in Mass Spectrometry. Mass Spectrom. Rev. 2020, 39, 453–470. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  154. Nemes, P.; Vertes, A. Laser Ablation Electrospray Ionization for Atmospheric Pressure, in Vivo, and Imaging Mass Spectrometry. Anal. Chem. 2007, 79, 8098–8106. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  155. Zhang, W.; Wang, X.; Xia, Y.; Ouyang, Z. Ambient Ionization and Miniature Mass Spectrometry Systems for Disease Diagnosis and Therapeutic Monitoring. Theranostics 2017, 7, 2968–2981. [Google Scholar] [CrossRef] [Scilit]
  156. Venter, A.; Nefliu, M.; Graham Cooks, R. Ambient Desorption Ionization Mass Spectrometry. TrAC Trends Anal. Chem. 2008, 27, 284–290. [Google Scholar] [CrossRef] [Scilit]
  157. Dane, A.J.; Cody, R.B. Using Solid Phase Microextraction with ACCUTOF-DART™ for Fragrance Analysis. LC GC N. Am. 2009, FEV, 46. [Google Scholar]
  158. Tzanetou, E.N.; Kasiotis, K.M. A mini review on solid phase micro-extraction applications in mass spectrometry detection of toxins. World J. Anal. Chem. 2013, 1, 14. [Google Scholar]
  159. Pawliszyn, J.B.; Rios, G.A.G. Probe for Extraction of Molecules of Interest from a Sample. U.S. Patent No. 9,733,234, 15 August 2017. [Google Scholar]
  160. Gómez-Ríos, G.A.; Pawliszyn, J. Solid Phase Microextraction (SPME)-Transmission Mode (TM) Pushes down Detection Limits in Direct Analysis in Real Time (DART). Chem. Commun. 2014, 50, 12937–12940. [Google Scholar] [CrossRef] [Scilit]
  161. Li, L.; Chen, T.-C.; Ren, Y.; Hendricks, P.I.; Cooks, R.G.; Ouyang, Z. Mini 12, Miniature Mass Spectrometer for Clinical and Other Applications—Introduction and Characterization. Anal. Chem. 2014, 86, 2909–2916. [Google Scholar] [CrossRef] [Scilit]
  162. Guo, X.; Bai, H.; Ma, X.; Li, J.; Ren, Y.; Ouyang, Z.; Ma, Q. Online Coupling of an Electrochemically Fabricated Solid-Phase Microextraction Probe and a Miniature Mass Spectrometer for Enrichment and Analysis of Chemical Contaminants in Infant Drinks. Anal. Chim. Acta 2020, 1098, 66–74. [Google Scholar] [CrossRef] [Scilit]
  163. Johnson, R.C.; Cooks, R.G.; Allen, T.M.; Cisper, M.E.; Hemberger, P.H. Membrane Introduction Mass Spectrometry: Trends and Applications. Mass Spectrom. Rev. 2000, 19, 1–37. [Google Scholar] [CrossRef] [Scilit]
  164. Thompson, A.J.; Creba, A.S.; Ferguson, R.M.; Krogh, E.T.; Gill, C.G. A Coaxially Heated Membrane Introduction Mass Spectrometry Interface for the Rapid and Sensitive On-line Measurement of Volatile and Semi-volatile Organic Contaminants in Air and Water at Parts-per-trillion Levels. Rapid Commun. Mass Spectrom. 2006, 20, 2000–2008. [Google Scholar] [CrossRef] [Scilit]
  165. Cotte-Rodríguez, I.; Handberg, E.; Noll, R.J.; Kilgour, D.P.A.; Cooks, R.G. Improved Detection of Low Vapor Pressure Compounds in Air by Serial Combination of Single-Sided Membrane Introduction with Fiber Introduction Mass Spectrometry (SS-MIMS-FIMS). Analyst 2005, 130, 679–686. [Google Scholar] [CrossRef] [Scilit]
  166. Boscaini, E.; Alexander, M.L.; Prazeller, P.; Märk, T.D. Investigation of Fundamental Physical Properties of a Polydimethylsiloxane (PDMS) Membrane Using a Proton Transfer Reaction Mass Spectrometer (PTRMS). Int. J. Mass Spectrom. 2004, 239, 179–186. [Google Scholar] [CrossRef] [Scilit]
  167. Lauritsen, F.R.; Ketola, R.A. Quantitative Determination of Semivolatile Organic Compounds in Solution Using Trap-and-Release Membrane Inlet Mass Spectrometry. Anal. Chem. 1997, 69, 4917–4922. [Google Scholar] [CrossRef] [Scilit]
  168. Matz, G.; Lennemann, F. On-Line Monitoring of Biotechnological Processes by Gas Chromatographic-Mass Spectrometric Analysis of Fermentation Suspensions. J. Chromatogr. A 1996, 750, 141–149. [Google Scholar] [CrossRef] [Scilit]
  169. Mendes, M.A.; Eberlin, M.N. Trace Level Analysis of VOCs and Semi-VOCs in Aqueous Solution Using a Direct Insertion Membrane Probe and Trap and Release Membrane Introduction Mass Spectrometry. Analyst 2000, 125, 21–24. [Google Scholar] [CrossRef] [Scilit]
  170. Haddad, G.; Schouwenburg, S.; Altesha, A.; Xu, W.; Liu, G. Using Breath Analysis as a Screening Tool to Detect Gastric Cancer: A Systematic Review. J. Breath Res. 2020, 15, 016013. [Google Scholar] [CrossRef] [Scilit]
  171. Vellasco, A.P.; Haddad, R.; Eberlin, M.N.; Höehr, N.F. Combined Cysteine and Homocysteine Quantitation in Plasma by Trap and Release Membrane Introduction Mass Spectrometry. Analyst 2002, 127, 1050–1053. [Google Scholar] [CrossRef] [Scilit]
  172. Riter, L.S.; Takáts, Z.; Cooks, R.G. Single-Sided Membrane Introduction Mass Spectrometry for on-Line Determination of Semi-Volatile Organic Compounds in Air. Analyst 2001, 126, 1980–1984. [Google Scholar] [CrossRef] [Scilit]
  173. Meurer, E.C.; Tomazela, D.M.; Silva, R.C.; Augusto, F.; Eberlin, M.N. Fiber Introduction Mass Spectrometry: Fully Direct Coupling of Solid-Phase Microextraction with Mass Spectrometry. Anal. Chem. 2002, 74, 5688–5692. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  174. Torion T-9 GCMS-Product Note. Available online: https://perkinelmer.widen.net/content/uaki2tuark/pdf/prd-torion-t-9-portable-gcms.pdf?u=3ezgsz (accessed on 2 September 2025).
  175. FLIR Griffin G510. Available online: https://defense.flir.com/defense-products/griffin-g510/ (accessed on 2 September 2025).
  176. Hicks, M.B.; Mattern, K.; Fine, J.; Grosser, S.; Patel, D.; Weisel, L.; Aggarwal, P. Portable Capillary LC for In-line UV Monitoring and MS Detection: Comparable Sensitivity and Much Lower Solvent Consumption. J. Sep. Sci. 2023, 46, 2300300. [Google Scholar] [CrossRef] [Scilit]
  177. Chatzimichail, S.; Casey, D.; Salehi-Reyhani, A. Zero Electrical Power Pump for Portable High-Performance Liquid Chromatography. Analyst 2019, 144, 6207–6213. [Google Scholar] [CrossRef] [Scilit]
  178. Foster, S.W.; Xie, X.; Pham, M.; Peaden, P.A.; Patil, L.M.; Tolley, L.T.; Farnsworth, P.B.; Tolley, H.D.; Lee, M.L.; Grinias, J.P. Portable Capillary Liquid Chromatography for Pharmaceutical and Illicit Drug Analysis. J. Sep. Sci. 2020, 43, 1623–1627. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  179. Chatzimichail, S.; Rahimi, F.; Saifuddin, A.; Surman, A.J.; Taylor-Robinson, S.D.; Salehi-Reyhani, A. Hand-Portable HPLC with Broadband Spectral Detection Enables Analysis of Complex Polycyclic Aromatic Hydrocarbon Mixtures. Commun. Chem. 2021, 4, 17. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  180. LCGC International-Supplements. The Potential for Portable Capillary Liquid Chromatography 2020. Available online: https://www.chromatographyonline.com/journals/lcgc-supplements/special-issues-06-02-2020 (accessed on 2 September 2025).
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.