Abstract
Neonatal continuous kidney replacement therapy (CKRT) is complicated by technical constraints, circuit clotting, and thrombocytopenia. We report a term female neonate with severe birth asphyxia, disseminated intravascular coagulation, and acute kidney injury who required prolonged CKRT with recurrent circuit clotting and repeated platelet transfusions. CKRT was initiated on day of life (DOL) 5, and persistent anuria prompted transfer to a specialized pediatric center for consideration of peritoneal dialysis on DOL 38. Three intentionally selected circuits were examined qualitatively by scanning electron microscopy (SEM): a cellulose triacetate membrane with post-dilution on ACH-Σ, and polyarylethersulfone membranes with combined pre- and post-dilution on Prismaflex at pre-blood-pump (PBP) flow rates of 100 and 300 mL/h. Circuit durations were 5 h 30 min, 6 h 40 min, and 7 h 53 min, respectively. Retrieved membranes and pump-tubing surfaces showed blood cell-like structures, platelet-like aggregates, and fibrin-like networks. Adherent structures appeared relatively sparse in selected images from the PBP 300-mL/h circuit despite its longer exposure. These morphology-based, nonquantitative observations cannot establish the effects of individual treatment factors or a causal relationship between CKRT and thrombocytopenia. Further standardized studies are needed to investigate blood–material interactions during neonatal CKRT.