  • Systematic Review
  • Open Access

26 November 2026

24 Pages

Three-Dimensional Accuracy of Static Guides and Dynamic Navigation for Palatal Miniscrew Insertion: A Systematic Review of Template Fabrication and Clinical Workflows

,
,
,
and
1
Department of Surgical Sciences, University of Cagliari, 09124 Cagliari, Italy
2
Private Practice, 1200 Geneva, Switzerland
3
Department of Surgical Sciences, School of Dental Medicine, University of Cagliari, 09124 Cagliari, Italy
4
Department of Surgical Sciences, Postgraduate School of Orthodontics, University of Cagliari, 09124 Cagliari, Italy

Abstract

Computer-guided insertion of palatal miniscrews can be performed with static surgical templates (s-CAS) or dynamic optical navigation (d-CAS), but their clinical transfer accuracy has not been systematically compared. This systematic review evaluated the three-dimensional placement accuracy of palatal miniscrews inserted with s-CAS and d-CAS. Following PRISMA 2020 (PROSPERO CRD420261380693), PubMed, Cochrane Library, Scopus, Web of Science, LILACS and Embase were searched in April–May 2026 for prospective clinical studies reporting angular, coronal and apical deviations between planned and achieved positions. Risk of bias was assessed with RoB 2 and ROBINS-I, certainty of evidence with GRADE, and findings were synthesized narratively. Nine studies (three randomized trials; up to 289 patients and 588 miniscrews) were included. Mean deviations of the s-CAS arms ranged from 3.61° to 7.15° (angular), 0.73 to 1.10 mm (coronal) and 1.25 to 2.00 mm (apical); those of the d-CAS arms, all from a single research group, ranged from 4.33° to 13.30°, 1.87 to 3.02 mm and 1.87 to 3.76 mm, respectively. In single randomized trials, full-arch guides reduced linear deviations and pressure-moulded guides reduced angular deviation. Static guides showed lower reported linear deviations than dynamic navigation, but all comparisons were indirect and measurement methods differed; certainty of evidence was moderate for s-CAS and very low for d-CAS.

1. Introduction

Temporary anchorage devices (TADs), particularly orthodontic miniscrews, have changed contemporary orthodontic and orthopedic biomechanics by providing absolute skeletal anchorage without relying on patient compliance [1,2,3]. Historically, conventional extraoral or tooth-borne anchorage systems (such as headgear, transpalatal arches, or Nance buttons) were associated with undesirable reciprocal side effects, including anchorage loss, tipping of anchor teeth, dental extrusion, and periodontal bone dehiscence [1,2,3,4,5,6,7]. Furthermore, while buccal interradicular sites were initially used for skeletal anchorage, their clinical success remains constrained by limited interradicular root clearance, the risk of iatrogenic periodontal ligament damage, and the physical obstruction of planned tooth movement pathways [1,2,3,8,9,10].
To overcome these anatomical and biomechanical limitations, the anterior palatal vault—specifically the median and paramedian region known as the “T-zone” located dorsal to the third palatal rugae—has emerged as the preferred site for skeletal anchorage [1,2,3,8,11,12,13,14,15]. Quantitative three-dimensional (3D) cone-beam computed tomography (CBCT) evaluations have consistently established that this area provides the thickest cortical and trabecular bone in the maxilla (averaging 6 to 10 mm in height), covered by a thin layer of attached keratinized mucosa [2,3,12,13,14,16,17]. Importantly, the paramedian T-zone is devoid of major neurovascular bundles and dental roots, demonstrating significantly higher clinical survival rates (95% to 99%) and stable peri-implant soft tissue responses compared to alveolar sites [1,2,3,8,12,13,14,17,18,19]. This anatomical foundation facilitates diverse orthopedic and orthodontic therapies, including Miniscrew-Assisted Rapid Palatal Expansion (MARPE/MSE) in late adolescents and adults [1,2,4,5,6,16,20,21,22,23], maxillary protraction protocols in skeletal Class III malocclusions [2,4,6,16,24], maxillary molar distalization and space closure [1,2,3,8,15,25,26], and the traction of impacted teeth [1,2,3,18].
Despite the favorable anatomy of the anterior palate, conventional direct freehand insertion presents substantial clinical challenges [1,3,10,25]. Freehand placement relies entirely on direct visual inspection and mucosal landmarks, which often leads to significant spatial discrepancies in insertion depth, angulation, and inter-implant parallelism [1,3,10,25]. Although clinical landmarks provide an anatomical safety margin relative to the central incisor roots [1], uncontrolled angulation or excessive miniscrew length increases the risk of encroaching upon the incisive canal, nasal cavity floor, or impacted tooth follicles, particularly in hyperdivergent or steep palatal vault morphologies [1,2,3,14,16]. Moreover, bicortical anchorage—the simultaneous engagement of the oral palatal cortex and the nasal cortical floor—is less predictable with manual insertion, although it is important for withstanding heavy orthopedic forces and preventing miniscrew tipping during maxillary expansion [1,2,5,16,22,23]. Furthermore, freehand placement precludes the feasibility of a digital “one-visit protocol”, wherein miniscrews and prefabricated CAD/CAM skeletal appliances are delivered simultaneously in a single clinical appointment [1,2,3,8,15,25,27].
Currently, computer-guided skeletal anchorage is categorized into two main operative modalities [1,3,10]. On the one hand, Static Computer-Assisted Surgery (s-CAS) utilizes patient-specific 3D-printed, milled, or thermoformed templates with rigid sleeves designed on superimposed surface scans and CBCT or cephalometric datasets to mechanically constrain drill trajectory and depth [1,8,27]. On the other hand, Dynamic Computer-Assisted Surgery (d-CAS) employs real-time optical tracking navigation systems that project the handpiece trajectory directly onto pre-operative CBCT scans, eliminating the need for physical templates while permitting intraoperative modifications [10].
Although digital workflows and virtual planning are rapidly expanding [2,3,5,16,17], clinical investigations indicate that placement accuracy is not uniform across computer-guided workflows, but is potentially modulated by template architecture, fabrication methodologies, digital planning platforms, anatomical palatal variations, and operator-related learning curves [1,2,3,8,10,16,27]. Previous systematic reviews have addressed guided miniscrew insertion, but they combined clinical, cadaveric and in vitro data on static guide designs in the anterior palate [28], synthesized clinical studies of 3D templates without a specific focus on palatal sites or dynamic navigation [29], or evaluated interradicular rather than palatal sites [30]. A synthesis restricted to prospective clinical studies of palatal miniscrews that considers static and dynamic guidance together and accounts for participant overlap between reports is therefore lacking. Therefore, the aim of this systematic review is to evaluate and compare the three-dimensional placement accuracy—specifically analyzing angular deviation, coronal head linear discrepancy, and apical tip displacement—of palatal orthodontic miniscrews inserted via static surgical guides and dynamic optical navigation across different template fabrication protocols and clinical workflows.

2. Materials and Methods

2.1. Protocol Registration

The current systematic review was conducted in adherence to the PRISMA guidelines, which outline the Preferred Reporting Items for Systematic Reviews and Meta-Analyses [31]. The review protocol was formally registered within PROSPERO, the International Prospective Register of Systematic Reviews, under the registration number CRD420261380693. The primary research question was formulated as follows: “In patients requiring palatal miniscrew insertion, how does static computer-assisted surgery (s-CAS)—across different template fabrication protocols and designs—compare to real-time dynamic navigation (d-CAS) in terms of 3D spatial placement accuracy, operative efficiency, and clinical outcomes?”. The criteria were defined according to the PICO framework:
  • (P) Population: Patients requiring palatal orthodontic temporary anchorage devices (TADs)/miniscrews.
  • Intervention: Static computer-assisted surgery (s-CAS) utilizing patient-specific surgical templates (3D-printed, pressure-moulded thermoformed, or CAD/CAM milled).
  • (C) Comparison: Dynamic computer-assisted surgery (d-CAS) with real-time optical navigation, or alternative surgical template fabrication designs/protocols (e.g., full-arch vs. skeletonized; 3D-printed vs. thermoformed/milled).
  • (O) Outcomes: Primary: Placement accuracy measures (angular deviation in degrees, 3D coronal/head linear deviation in mm, 3D apical/tip linear deviation in mm). Secondary: vertical depth deviation, operative time, learning-curve trends, miniscrew success after orthodontic loading, complications (e.g., root contact, nasal floor perforation), and patient-reported pain.

2.2. Eligibility Criteria

To ensure methodological rigor and homogeneity of the extracted statistical data, literature selection was performed by applying strict inclusion and exclusion criteria structured according to the PICO framework:
Inclusion Criteria:
  • Population (P): Human patients (children, adolescents, or adults) requiring orthodontic or orthopedic skeletal anchorage via temporary anchorage devices (TADs)/miniscrews inserted in the median or paramedian regions of the anterior palatal vault.
  • Intervention (I): Palatal miniscrew insertion planned and performed using computer-assisted static surgical guides (s-CAS), either 3D-printed (SLA/DLP), CAD/CAM-milled or pressure-moulded/thermoformed on digitally planned models, or dynamic computer-assisted surgery (d-CAS) using real-time optical tracking navigation systems.
  • Comparison (C): Dynamic computer-assisted surgery (d-CAS) using real-time optical tracking navigation systems, or alternative surgical template fabrication designs/protocols (e.g., full-arch vs. skeletonized; 3D-printed vs. thermoformed/milled).
  • Outcomes (O): Primary Outcomes: Quantitative, three-dimensional placement accuracy metrics determined by superimposing pre-operative virtual planning and post-operative models/CBCT scans, including: (1) angular deviation/axis tilt in degrees (°), (2) coronal/entry linear deviation at the miniscrew head in millimeters (mm), and (3) apical/tip linear deviation at the miniscrew apex in millimeters (mm). Secondary Outcomes: Vertical depth linear deviation (mm), operative surgical duration (minutes:seconds), learning curve trends, TAD clinical success/stability rate after orthodontic loading, the incidence of iatrogenic injuries (e.g., dental root damage, nasal floor perforation), and patient-reported pain.
  • Study Design (S): Prospective randomized controlled trials (RCTs), prospective parallel-group clinical trials, and prospective single-arm or comparative in vivo clinical studies published in peer-reviewed journals. Single-arm studies and studies whose internal comparison did not concern the guidance modality (e.g., insertion side, palatal morphology or operator) were eligible because they report the accuracy of a guided protocol against the virtual plan.
Exclusion Criteria:
  • Preclinical in vitro simulations, artificial typodont models, animal studies, or human cadaveric studies lacking an in vivo clinical patient validation phase.
  • Non-guided, purely freehand manual miniscrew insertions without static template guidance or dynamic navigation assistance.
  • Miniscrews placed in anatomical regions other than the palate (e.g., buccal interradicular, infrazygomatic crest, or mandibular buccal shelf).
  • Studies failing to quantitatively report 3D linear or angular deviations between planned and achieved positions via tridimensional superimposition.
  • Case reports, case series, narrative reviews, systematic reviews, conference abstracts, and non-peer-reviewed literature.

2.3. Information Sources

A comprehensive literature search was performed between 10 April and 11 May 2026 by two researchers independently (C.S., C.F.) to identify articles relevant to the study question. The following electronic databases were investigated: PubMed, Cochrane Library, Scopus, Web of Science, LILACS, and Embase. A manual search of the reference lists of all included studies and relevant reviews was also performed to identify further eligible articles. Trial registries and grey literature were not searched. No restrictions regarding publication date or language were applied.

2.4. Search Strategy

The search strategy was developed according to the PICO framework, combining free-text keywords and, in Embase, Emtree terms. It was first developed for PubMed and subsequently adapted using broader keyword combinations in the remaining databases to account for differences in indexing engines and syntax sensitivity (Table 1). In Scopus, the query was executed without parentheses around the initial OR operator, which broadened the search yield (87 records) rather than restricting it.
Table 1. Search strategy for the six databases (PubMed, Cochrane Library, Scopus, Web of Science, LILACS and Embase).

2.5. Data Collection

The review process initially involved screening titles and abstracts, followed by full-text assessment of eligible papers. Two reviewers (C.S., C.F.) independently conducted the screening process; discordant evaluations were resolved through discussion with two additional authors (C.D., A.V.). Duplicates were removed using reference management software and verified manually. To assess the level of inter-examiner agreement, Cohen’s kappa coefficient was calculated [32] (Cohen’s κ = 0.86), indicating almost perfect agreement.
Following full-text review, studies meeting the predefined eligibility criteria were included. Data extracted included first author, publication year, country of origin, study design, cohort characteristics (sample size, age, sex), technology used, virtual planning and analysis software, and primary and secondary accuracy outcomes. For multi-arm studies, data were extracted separately for each arm. Three-dimensional deviations were preferred; single-plane (horizontal, vertical or planar angular) values were recorded separately and were not combined with three-dimensional values. Reports from the same research group were scrutinized for potential participant overlap (recruitment period, study center, ethical approval code, and outcome ranges).

2.6. Risk of Bias Assessment

The risk of bias within the included studies was appraised independently by two reviewers, with any discrepancies resolved through discussion and consensus. In accordance with PRISMA guidelines, specific assessment tools tailored to the experimental design of the included investigations were employed. For Randomized Controlled Trials (RCTs), methodological quality was evaluated using the revised Cochrane Risk of Bias tool (RoB 2) [33], which appraises five standard bias domains: the randomization process, deviations from intended interventions, missing outcome data, measurement of the outcome, and selection of the reported result, leading to an overall classification of “Low risk”, “Some concerns”, or “High risk”. Conversely, non-randomized prospective clinical studies were evaluated using the Cochrane ROBINS-I tool (Risk Of Bias In Non-randomized Studies of Interventions) [34], which examines seven distinct potential sources of bias, specifically covering confounding, participant selection, classification of interventions, deviations from intended interventions, missing data, outcome measurement, and selection of the reported result. For unified visual presentation alongside RoB 2, the seven ROBINS-I domains were mapped onto five harmonized columns (D1–D5) by grouping participant selection and intervention classification into D2, while keeping all domain-level judgments aligned with the full seven-domain assessment.

2.7. Certainty Assessment

The overall certainty of the evidence for primary outcomes was evaluated using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) approach [35]. The certainty of evidence was categorized as High, Moderate, Low, or Very Low. The assessment was initiated at “High” for randomized trials and also for non-randomized studies assessed with ROBINS-I, in which confounding and selection bias are addressed within the risk-of-bias domain [36]. Potential downgrading was evaluated across five domains: risk of bias, inconsistency, indirectness, imprecision, and publication bias, differences in reference points, imaging modalities and outcome definitions, as well as the lack of head-to-head trials, were considered under indirectness; participant overlap between reports from the same research group was considered as an additional reason for downgrading.

2.8. Data Synthesis

Given the clinical and methodological heterogeneity of the included studies (differing guide designs, manufacturing protocols, surgical platforms, 3D vs. planar angular metrics, and absence of dispersion measures in certain reports) and the documented participant overlap among reports originating from the same research group, quantitative data pooling via meta-analysis was not performed, and findings were synthesized narratively following the Synthesis Without Meta-analysis (SWiM) reporting guideline [37]. For each study arm, the mean ± standard deviation of angular, coronal (head/entry), and apical (tip) deviations were tabulated, and within-study comparisons were presented with corresponding p-values. The ranges of reported means across study arms were used to descriptively summarize static (s-CAS) and dynamic (d-CAS) guidance; these ranges represent descriptive summaries rather than pooled estimates. The mean of each arm was displayed with its 95% confidence interval in a forest plot without a summary estimate; confidence intervals were calculated from the reported means, standard deviations and numbers of miniscrews. Because most patients received two miniscrews, the miniscrew-level analyses of the primary studies do not account for clustering within patients, and their precision may be overestimated. Because no study compared s-CAS and d-CAS directly within the same trial, all comparisons between modalities are indirect. Non-independent reports from the same research team were identified, flagged in Table 2, and their cohorts were not counted cumulatively as independent participants.
Table 2. Main characteristics of the nine studies included in the review.

3. Results

3.1. Selection of Sources of Evidence

A comprehensive electronic database search yielded a total of 206 records: PubMed (n = 39), Cochrane Library (n = 51), Scopus (n = 87), Web of Science (n = 4), LILACS (n = 10), and Embase (n = 15). Prior to screening, 123 duplicate records were identified and systematically removed. No records were flagged as ineligible by automated tools or removed for other reasons.
Consequently, 83 unique records underwent preliminary screening based on title and abstract evaluation. Of these, 72 records were excluded as they failed to meet the eligibility criteria, specifically case reports or case series (n = 23), preclinical in vitro or cadaveric designs (n = 12), inappropriate study focus/title (n = 12), wrong intervention (n = 10), wrong outcome (n = 9), inappropriate study population (n = 5), and general narrative reviews (n = 1).
For the remaining 11 records, full-text retrieval was sought. Two reports could not be successfully retrieved, leaving 9 reports to be assessed for eligibility via full-text evaluation. Following full-text analysis, all 9 clinical studies met all predefined eligibility criteria and were definitively included in the qualitative synthesis of this systematic review. These 9 included reports derive from at least seven independent patient cohorts, as three reports originated from the same research institution and may involve overlapping participants. The complete multi-stage selection process, mapping the progression from initial database identification to final study inclusion, is visually summarized in the PRISMA 2020 flow diagram (Figure 1) [31].
Figure 1. PRISMA 2020 flow diagram of the literature search and study selection process.

3.2. Characteristics of the Included Studies

The general and methodological characteristics of the 9 included studies are systematically summarized in Table 2. The final selection comprises a diverse yet complementary body of clinical evidence, published between 2022 and 2026. Among the included literature, three studies are structured as Randomized Controlled Trials (RCTs) [41,42,43], while the remaining six reports are prospective non-randomized or single-arm clinical cohort investigations [38,39,40,44,45,46]. In the study by Brilli et al. (2025) [41], participants were randomly allocated to operators of different generations and all of them were treated with d-CAS; no included study compared s-CAS and d-CAS directly.
The cumulative sample size across all included papers encompasses 289 patients requiring palatal temporary anchorage devices (TADs) for orthodontic purposes, with a total of 588 miniscrews analyzed in vivo. In all investigations, miniscrews were inserted into the anterior palatal vault, predominantly targeting the paramedian or median areas to achieve stable cortical bone anchorage. These totals are the sum of the samples reported in each paper and probably overestimate the number of independent participants. The three reports by Brilli et al. (2024, 2025, and 2026; 119 patients and 238 miniscrews as reported) [39,40,41] come from the same research group and center, used the same dynamic navigation system in overlapping recruitment periods (from November–December 2022), and share identical ranges of the deviation values (e.g., maximum angular deviation of 28.4° and maximum apical deviation of 5.85 mm in all three reports); some patients may therefore appear in more than one report. Excluding these three reports, the remaining six studies included 170 patients and 350 miniscrews. The number of independent participants is therefore estimated between 242 and 289 patients (494–588 miniscrews) and cannot be determined more precisely from the published data.
Regarding the surgical approach, the literature exhibits a technological transition: the majority of the studies evaluate static computer-aided surgery (s-CAS) utilizing 3D-printed stereolithographic (SLA) or thermoformed surgical templates [38,42,43,44,45,46], whereas recent investigations introduce and validate dynamic navigation systems (d-CAS) [39,40,41]. Digital planning was based on cone-beam computed tomography (CBCT) matched with intraoral scans in seven studies, whereas two trials combined intraoral scans with lateral cephalograms [42,43]. Post-operative positions were recorded with intraoral scans of the miniscrew heads or scan bodies in the static studies and with CBCT in the dynamic studies, and deviations were computed with different software and reference points (Supplementary Table S2). The outcome data extracted from each study arm are summarized in Table 3, and the comparisons made within each study are reported in Table 4. For Al-Gazzawi et al. [38], only the in vivo data were considered.
Table 3. Accuracy outcomes by study arm (mean ± SD).
Table 4. Comparisons made within each study and corresponding p-values.

3.3. Results of Individual Sources of Evidence

Table 3 reports the accuracy outcomes (mean ± SD) for each study arm and Table 4 the comparisons made within each study, with the corresponding p-values. The following sections describe only the results of the nine included studies.
The narrative synthesis of the included clinical investigations shows that the deviations between the planned and the achieved position of palatal miniscrews varied with the navigation modality (static vs. dynamic), guide design, manufacturing process, and clinical execution.
Across the included studies, the mean deviations of the static (s-CAS) arms [38,42,43,44,45,46] ranged from 3.61° to 7.15° for angulation, from 0.73 to 1.10 mm at the coronal head, and from 1.25 to 2.00 mm at the apex, whereas those of the dynamic (d-CAS) arms [39,40,41] ranged from 4.33° to 13.30°, from 1.87 to 3.02 mm, and from 1.87 to 3.76 mm, respectively (Table 3; Figure 2). Linear deviations were therefore lower in the s-CAS arms, whereas the angular ranges partly overlapped. These are ranges of the reported means, not pooled estimates, and the two modalities were not compared directly within any study. The planar angles reported by Palone et al. (2026) [44] and the laboratory-step deviation reported by Pozzan et al. (2022) [46], which measures the guide before clinical placement, were not included in the angular range of the static arms; likewise, the single-plane (horizontal) head discrepancies reported by Al-Gazzawi et al. (2024) [38] were not included in the coronal range. In a sensitivity evaluation restricting the d-CAS data to the single largest and randomized cohort from this research group to eliminate any potential participant duplication (Brilli et al., 2025 [41]; 72 patients, 144 miniscrews), mean deviations ranged from 4.33° to 8.89° for angulation, 1.87 to 2.82 mm at the coronal entry point, and 1.87 to 3.31 mm at the apex, confirming that linear discrepancies remained higher than in the s-CAS arms while angular ranges continued to overlap.
Figure 2. Mean deviation and 95% confidence interval of each study arm for angular, coronal and apical deviation. No summary estimate is shown. Confidence intervals were calculated from the reported means and standard deviations; n = number of miniscrews per arm, except Möhlhenrich et al. (2026) [43], for whom n = patients because values were averaged per patient; because patient-level intraclass correlation coefficients were unavailable, miniscrew-level confidence intervals do not account for within-patient clustering and should be interpreted as potentially overprecise. All dynamic navigation arms derive from three reports of the same research group that may share participants. Single-plane values (Al-Gazzawi et al. [38], horizontal discrepancies; Palone et al. [44], planar angles) are not shown [39,40,41,42,43,44,45,46].

3.3.1. Static Computer-Assisted Surgery (s-CAS) Outcomes

Across the static guided protocols, three-dimensional printing and thermoforming workflows achieved low linear deviations at the miniscrew head (0.73–1.10 mm; below 1 mm in all arms except the skeletonized guide, 1.01 mm, and the multicortical guide, 1.10 mm) [42,43,44].
Al-Gazzawi et al. (2024) [38] evaluated a rigid, full-arch 3D-printed resin template with an integrated vertical depth stop, reporting mean horizontal discrepancies at the miniscrew head of 0.32 ± 0.15 mm on the left side and 0.35 ± 0.16 mm on the right side, with mean vertical discrepancies of 0.34 ± 0.17 mm and 0.31 ± 0.17 mm, respectively; three-dimensional and apical deviations were not reported. No statistically significant differences were observed based on sidedness (p = 0.43 for horizontal and p = 0.39 for vertical planes).
Hoffmann et al. (2025) [42] compared two CAD/CAM guide architectures, demonstrating that a full-arch design provided significantly superior linear accuracy at both the head (0.73 ± 0.47 mm vs. 1.01 ± 0.55 mm; p = 0.021) and the apex (1.47 ± 1.30 mm vs. 2.00 ± 1.01 mm; p = 0.011) compared to a skeletonized design. Angular deviation was also lower in the full-arch group (5.55 ± 4.88° vs. 6.44 ± 3.36°), although this difference did not reach statistical significance (p = 0.082). A 100% clinical success rate was maintained after 3 months of orthodontic loading.
Möhlhenrich et al. (2026) [43] compared conventional pressure-moulded (PM) thermoformed guides with 3D-printed (3DP) guides, revealing that PM templates achieved significantly lower overall 3D angular deviations (3.61 ± 1.29° vs. 5.77 ± 2.45°; p = 0.006), an advantage particularly evident in paramedian sites (3.41 ± 0.79° vs. 6.13 ± 2.48°; p = 0.018). Linear deviations at the shoulder (0.81 ± 0.27 mm vs. 0.84 ± 0.34 mm; p = 0.778) and tip (1.25 ± 0.40 mm vs. 1.49 ± 0.58 mm; p = 0.199) showed no significant inter-group differences.
Palone et al. (2026) [44] evaluated multicortical (bi- and tricortical) insertion using rigid DLP-printed guides with 7 mm metal sleeves, observing a mean total head deviation of 1.10 ± 1.01 mm, a total tip deviation of 1.71 ± 1.10 mm, and an angular discrepancy of 5.90 ± 6.06° in the transverse plane. Multiple linear regression confirmed that clinical and structural variables—including insertion side (p = 0.407–0.583), frontal insertion angle (p = 0.132–0.452), and miniscrew length (p = 0.207–0.540)—did not significantly impact placement accuracy.
Pozzan et al. (2022) [46] measured the angular deviation separately for the laboratory step (2.12 ± 1.62°), the clinical step (6.23 ± 3.75°), and the whole procedure (5.70 ± 3.42°). The deviation of the laboratory step was significantly lower than those of the clinical step and of the whole procedure (both p < 0.001), whereas the clinical step and the whole procedure did not differ significantly (p = 0.231). The three measurements are not additive.
Ronsivalle et al. (2023) [45] compared open-system digital planning platforms, reporting that generic restorative implant software (Blue Sky Plan: 7.15 ± 1.09° right, 6.19 ± 0.80° left) achieved angular placement accuracy comparable to dedicated orthodontic software (Dolphin: 6.74 ± 1.23° right, 5.79 ± 0.95° left) without statistically significant inter-software differences (p = 0.440 and p = 0.320, respectively). However, an operator handedness/sidedness effect was detected, with right-sided insertions displaying significantly higher angular discrepancies (p ≤ 0.019).

3.3.2. Dynamic Computer-Assisted Surgery (d-CAS) Outcomes

The three included studies on dynamic optical navigation (Navident) were all carried out by the same research group (see Section 3.2) [39,40,41].
Brilli et al. (2024) [39] evaluated the influence of palatal vault shape in 24 patients (48 miniscrews) and found higher angular and apical deviations in steep/high palates (13.30 ± 7.77° and 3.76 ± 1.59 mm) than in medium (7.11 ± 5.70° and 2.71 ± 1.44 mm) and low/flat palates (4.92 ± 3.15° and 2.43 ± 1.16 mm; p = 0.008 for angle and p = 0.036 for Apex-3D, as reported in the abstract; the two values are inverted in the Results section of the original article). Post hoc tests confirmed the differences for angular deviation (steep versus medium and versus flat palates) but not for Apex-3D. Entry point deviations were not significantly influenced by vault shape (p = 0.130).
Brilli et al. (2025) [41] investigated the influence of the operator generation in 72 patients (144 miniscrews) who were randomly allocated to operators of generation X, Y, or Z; all patients were treated with d-CAS. Generation Z operators achieved significantly lower angular deviations (4.33 ± 1.92°) and lower deviations at the entry point (1.87 ± 0.64 mm) and at the apex (1.87 ± 0.68 mm) than generation X and Y operators (p = 0.003 for angle; p < 0.001 for entry and apex).
Brilli et al. (2026) [40] evaluated a prospective single-arm cohort of 23 patients (46 miniscrews) and reported mean deviations of 8.50 ± 6.69° (range 0.00–28.40°) for angulation, 2.55 ± 1.12 mm (range 0.83–5.20 mm) at the entry point, 3.11 ± 1.49 mm (range 0.44–5.85 mm) at the apex, and 0.87 ± 0.80 mm (range 0.01–3.05 mm) for vertical depth at the apex. The duration of the intervention decreased significantly with the number of surgeries performed (r = −0.69, p < 0.001), whereas the accuracy variables were not significantly correlated with the number of surgeries (p > 0.05). The d-CAS results of Brilli et al. (2024, 2025a, and 2025b) [39,40,41] are reported separately and were not pooled, because the reports may share participants.

3.4. Primary Outcomes Analysis

3.4.1. Accuracy of Computer-Assisted Palatal Miniscrew Insertion

The primary outcomes of the included studies were the deviations between the planned and the achieved miniscrew position: angular deviation, linear coronal (head/entry) deviation, and linear apical (tip) deviation [38,39,40,41,42,43,44,45,46]. The magnitude of the deviations varied with the guidance modality, fabrication method, template architecture, and palatal morphology [39,42,43,45].

3.4.2. Angular Deviation

In static systems (s-CAS), mean angular deviations ranged from 3.61° to 7.15° across the static arms [42,43,45,46]. Pressure-moulded thermoformed guides demonstrated the lowest angular deviation (3.61 ± 1.29°) and differed significantly from 3D-printed guides (5.77 ± 2.45°, p = 0.006) [43]. Full-arch guides showed a numerically lower deviation (5.55 ± 4.88°) than skeletonized guides (6.44 ± 3.36°), without reaching statistical significance (p = 0.082) [42]. In the investigation by Pozzan et al. (2022) [46], the deviation of the clinical step was 6.23 ± 3.75° and that of the complete procedure was 5.70 ± 3.42° (laboratory step: 2.12 ± 1.62°). Palone et al. (2026) [44] reported planar angles (XY 5.90 ± 6.06°, YZ 2.65 ± 2.47°, XZ 3.02 ± 2.65°), which are not directly comparable with three-dimensional angular deviations and are not included in the range above.
In dynamic navigation (d-CAS), mean angular deviations ranged from 4.33° to 13.30° across the dynamic arms, with the single-arm study reporting an overall mean of 8.50 ± 6.69° (range 0.00–28.40°) [40]. Steep and high palatal vaults displayed significantly greater angular discrepancies (13.30 ± 7.77°) compared to flat vaults (4.92 ± 3.15°; p = 0.008) [39]. Furthermore, generation Z operators achieved lower deviations (4.33 ± 1.92°) than generation X and Y operators (8.42 ± 6.80° and 8.89 ± 6.50°; p = 0.003) [41]. These dynamic arms derive from reports originating from a single institution that may share participants.

3.4.3. Coronal Linear Discrepancy (Entry Point/Miniscrew Head)

In static systems (s-CAS), mean three-dimensional coronal deviations ranged from 0.73 ± 0.47 mm (full-arch 3D-printed guides) to 1.10 ± 1.01 mm (multicortical DLP guides) [42,44], remaining below 1 mm in all arms except the skeletonized guides (1.01 ± 0.55 mm) and multicortical guides. Al-Gazzawi et al. [38] reported horizontal head discrepancies below 0.5 mm with a rigid 3D-printed guide, which are not directly comparable with three-dimensional values. Full-arch guides provided significantly lower coronal deviations (0.73 ± 0.47 mm) than skeletonized designs (1.01 ± 0.55 mm; p = 0.021) [42], whereas pressure-moulded and 3D-printed guides showed no significant difference (0.81 ± 0.27 vs. 0.84 ± 0.34 mm; p = 0.778) [43].
In dynamic navigation (d-CAS), mean coronal deviations ranged from 1.87 to 3.02 mm across the dynamic arms (individual values 0.83–5.20 mm in the single-arm cohort) [39,40,41] and were higher than in the static arms, although they were measured on post-operative CBCT rather than on intraoral scans and the two modalities were not compared directly within any clinical trial. Entry point deviations were not significantly influenced by palatal vault shape (p = 0.130) [39], whereas generation Z operators achieved significantly lower deviations (1.87 ± 0.64 mm) than generation X and Y operators (2.48 ± 1.18 and 2.82 ± 0.95 mm; p < 0.001) [41].

3.4.4. Apical Linear Discrepancy (Apical Tip)

In static systems (s-CAS), mean apical deviations ranged from 1.25 ± 0.40 mm (pressure-moulded guides) to 2.00 ± 1.01 mm (skeletonized guides) and reached 1.71 ± 1.10 mm with multicortical DLP guides [42,43,44]. Full-arch guides showed significantly lower apical deviations (1.47 ± 1.30 mm) than skeletonized guides (2.00 ± 1.01 mm; p = 0.011) [42], whereas pressure-moulded and 3D-printed guides did not differ significantly (1.25 ± 0.40 vs. 1.49 ± 0.58 mm; p = 0.199) [43]. In all static studies that reported both parameters, apical deviations were consistently larger than coronal deviations.
In dynamic navigation (d-CAS), mean apical deviations ranged from 1.87 to 3.76 mm across the dynamic arms, with the single-arm study reporting a mean of 3.11 ± 1.49 mm (range 0.44–5.85 mm) [40]. Steep and high palates exhibited higher apical deviations (3.76 ± 1.59 mm) than low and flat palates (2.43 ± 1.16 mm; omnibus p = 0.036, although pairwise post hoc comparisons were not statistically significant) [39] and generation Z operators demonstrated lower apical discrepancies (1.87 ± 0.68 mm) compared to generation X (3.31 ± 1.42 mm) and generation Y operators (3.07 ± 1.39 mm; p < 0.001) [41].

3.5. Secondary Outcomes Analysis

Across the nine included studies, vertical depth deviation was reported in three studies [38,40,44], operative duration in one [40], learning-curve or operator-experience effects in two [40,41], post-loading miniscrew stability in one [42], and radiological root assessment in one [39], whereas nasal floor perforation and patient-reported pain were not reported in any included study. In addition to the primary spatial deviations, several secondary clinical outcomes were documented across the included literature, including vertical depth deviation [38,40,44], operative surgical duration [40], and learning curve trends [40,41].
Regarding vertical depth linear deviation, static surgical guides featuring mechanical depth stops achieved high precision, with reported discrepancies of 0.34 ± 0.17 mm on the left side and 0.31 ± 0.17 mm on the right side in rigid full-arch designs [38], and 0.81 ± 0.92 mm at the miniscrew head and 0.89 ± 0.89 mm at the apex in multicortical guides [44]. In contrast, dynamic optical navigation yielded a mean apical vertical discrepancy of 0.87 ± 0.80 mm, with individual clinical values ranging between 0.01 and 3.05 mm [40].
Operative surgical duration and operator experience were appraised in dynamic workflows. In the prospective single-arm dynamic cohort, the duration of the surgical intervention decreased significantly with the cumulative number of surgeries performed (r = −0.69, p < 0.001), reflecting improved procedural efficiency [40]. However, the spatial accuracy parameters did not show significant linear improvements over time (p > 0.05) [40]. In addition, the operator’s generational cohort was associated with dynamic navigation accuracy, as Generation Z operators showed significantly lower angular and linear deviations than Generation X and Y operators (post hoc p ≤ 0.006) [41].
Complications and patient-reported pain were not systematically reported. Brilli et al. (2024) [39] observed no radiological signs of root damage on post-operative CBCT, and Hoffmann et al. (2025) [42] reported no miniscrew failures, inflammation or increased mobility after three months of orthodontic loading; none of the included studies reported data on nasal floor perforation or patient-reported pain.

3.6. Risk of Bias Assessment Within Studies

The methodological quality and risk of bias of the nine included studies were critically appraised according to their specific study design. Randomized controlled trials (RCTs) were evaluated using the revised Cochrane Risk of Bias tool (RoB 2) [33], whereas non-randomized prospective clinical studies were evaluated using the Risk Of Bias In Non-randomized Studies of Interventions (ROBINS-I) tool [34].
Overall, the included evidence showed a low-to-moderate risk of bias (Figure 3). In randomized trials [41,42,43], the primary methodological constraint related to the impossibility of blinding the operating surgeons due to the overt nature of the interventions (different guide designs or manufacturing methods, or operators of different generations). However, objective three-dimensional computerized outcome assessments (automated voxel/ICP superimpositions and coordinate analysis algorithms) substantially mitigated measurement and detection bias. In non-randomized prospective cohorts [38,39,40,44,45,46], potential confounding was primarily linked to baseline anatomical variations (e.g., palatal vault depth and slope) and operator experience levels. The possible overlap of participants among the three reports of the same research group is not captured by standard risk-of-bias tools and was formally taken into account in the assessment of the certainty of the evidence.
Figure 3. Risk of bias assessment for the 9 included studies [38,39,40,41,42,43,44,45,46], according to RoB 2 [33] and ROBINS-I [34], detailing judgments across evaluated domains. Note: For RoB 2 (RCTs), D1 = Bias arising from the randomization process; D2 = Bias in selection of participants/classification of interventions; D3 = Bias due to missing outcome data; D4 = Bias in measurement of the outcome; D5 = Bias in selection of the reported result. For ROBINS-I (Non-randomized studies), D1 = Bias due to confounding; D2 = Bias in selection of participants/classification of interventions; D3 = Bias due to missing data; D4 = Bias in measurement of outcomes; D5 = Bias in selection of the reported result.

Summary of Methodological Quality

In randomized trials appraised via RoB 2 [33], all 3 RCTs [41,42,43] demonstrated low risk regarding the randomization process, missing outcome data, and selective outcome reporting, with “some concerns” arising solely from the lack of clinician blinding, an unavoidable feature of surgical device trials. In prospective non-randomized cohorts evaluated with ROBINS-I [34], 3 studies [38,44,45] exhibited a low overall risk of bias due to well-controlled protocols, whereas 3 investigations [39,40,46] were judged as having moderate overall risk attributable to confounding from anatomical palatal slopes, single-arm cohort designs, or operator learning curves.

3.7. Certainty of the Evidence (GRADE)

The overall certainty of the evidence for primary outcomes was evaluated using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) approach [35] (Table 5). For static computer-assisted surgery (s-CAS), the evidence was rated as Moderate certainty, downgraded by one level for indirectness due to variations in reference points, imaging modalities, and guide designs across studies as well as the absence of direct comparisons with d-CAS. Risk of bias was judged as not serious because primary accuracy outcomes were measured using automated, objective 3D digital superimposition algorithms that minimized detection bias in both randomized [42,43] and non-randomized cohorts [38,44,45,46]; inconsistency was rated as not serious given the narrow and consistent ranges of mean deviations across all static arms (coronal: 0.73–1.10 mm; apical: 1.25–2.00 mm); and imprecision was judged as not serious because cumulative sample sizes (192 to 244 miniscrews) yielded narrow confidence intervals around the arm means (Figure 2), while methodological heterogeneity was already penalized under indirectness to avoid double-downgrading. For dynamic computer-assisted surgery (d-CAS), the evidence was rated as Very Low certainty, downgraded due to risk of bias, inconsistency (for angular and apical deviations), indirectness, imprecision, and the risk of cohort overlap across reports originating from a single research center.
Table 5. GRADE Summary of Findings for Primary Accuracy Outcomes.

4. Discussion

The present systematic review provides a comprehensive qualitative synthesis of contemporary clinical evidence regarding the three-dimensional placement accuracy of orthodontic temporary anchorage devices (TADs) placed in the anterior palatal vault using static computer-assisted surgery (s-CAS) and dynamic computer-assisted surgery (d-CAS). Our findings suggest that the magnitude of the spatial error between virtual planning and clinical execution varies with the type of guidance, template design, manufacturing method and recipient-site anatomy [38,39,40,41,42,43,44,45,46]. A difference was observed between the two approaches: static templates showed coronal discrepancies below 1 mm in most arms [42,43], whereas dynamic navigation showed mean entry deviations between 1.87 and 3.02 mm and mean angular deviations up to 13.30°, which varied with palatal vault slope and operator generation [39,40,41]. Because these data derive from different studies, centres and measurement methods, this difference is indirect.
These findings expand upon earlier systematic reviews on computer-guided orthodontic miniscrew insertion [28,29,30]. Stucki et al. [28] reported that static guide design influences placement accuracy in the anterior palate, but combined in vitro, cadaveric, and clinical investigations without evaluating dynamic navigation. Similarly, Jedliński et al. [29] confirmed that 3D-printed templates improve positional accuracy compared to freehand insertion across mixed anatomical regions, while Santmartí-Oliver et al. [30] demonstrated reduced linear and angular deviations in buccal interradicular sites, where root proximity imposes stricter spatial constraints than in the anterior palatal vault. By restricting inclusion exclusively to prospective in vivo studies in the palate and accounting for cohort overlap in d-CAS [39,40,41], the present synthesis provides a more targeted clinical benchmark.
This difference may be explained, at least in part, by the mechanical differences between guided and tracked freehand insertion [39,42]. Static guides operate as rigid mechanical conduits that completely constrain the bur along three spatial coordinates and establish an absolute insertion stop, thereby reducing the influence of patient movements or minor tremors [38,42,43]. Conversely, dynamic navigation does not mechanically stabilize the handpiece; rather, it provides an optical feedback loop that requires the surgeon to continuously execute active neuromuscular compensations, translating a virtual trajectory visualized on an external two-dimensional monitor into precise physical movements inside the oral cavity [39,40,41]. Dynamic guidance may therefore be more vulnerable to instantaneous perceptual lag, fulcrum instability, and minute visual-motor calibration errors that physical guide sleeves largely limit [39,41].
Within static workflows, template architecture influenced positional accuracy in one RCT, with full-arch tooth-supported templates showing lower coronal and apical deviations than skeletonized designs [42]. Biomechanically, it may be hypothesized that broader contralateral dental coverage distributes insertion counter-torques across multiple abutments, reducing micro-rocking or elastic flexion during placement, whereas narrow skeletonized connectors may be more prone to deformation under hand pressure [42]. Similarly, manufacturing technology affected static precision in a single RCT, where pressure-moulded thermoformed templates achieved lower angular deviations than 3D-printed resin guides [43]; however, because the 3D-printed guides were supported on only three teeth per side, differences in guide extension may also have contributed [43]. From a material standpoint, these differences could hypothetically reflect anisotropic polymerization shrinkage or post-curing warpage in vat-photopolymerized resins compared to the more uniform adaptation of thermoformed polymers over stone casts [43]. Nevertheless, static templates remain subject to clinical error accumulation, as step-by-step analyses showed that the clinical placement phase accounts for the largest share of angular divergence compared to the laboratory fabrication step [46]. This intraoperative discrepancy may hypothetically arise from the functional clearance required between the drill shaft and the metallic guide sleeve, as well as limited visual access, restricted irrigation, and uneven compression of resilient palatal mucosa during seating [38,42,44,46]. These physical constraints are avoided by dynamic computer-assisted surgery, which utilizes real-time optical tracking of passive infrared reflectors attached to the patient and handpiece [39,40,41]. By eliminating physical resin templates, d-CAS allows direct visualization of the insertion site and the flexibility to reorient the insertion trajectory intraoperatively if unforeseen bone deficiencies are encountered [39,40,41]. However, the precision of d-CAS decreased in steep, high-arched palates compared to flat palatal vaults [39]. Biomechanically, this phenomenon may be related to the unfavorable angle of incidence between the drill axis and the steep palatal cortical incline [39]. When a self-drilling miniscrew contacts an inclined, highly calcified cortical bone plate at an oblique trajectory without physical sleeve containment, the cutting flutes experience lateral deflection forces [39]. This may induce cortical skiving or tangential slippage along the palatal wall before positive cortical purchase can be established, driving the miniscrew apex away from its planned skeletal target [39].
Moreover, operator-related factors appear to influence dynamic navigation performance, as Generation Z operators achieved better angular and linear accuracy than Generation X and Y operators in one randomized study [41]. Dynamic navigation requires hand–eye decoupling, because the surgeon manipulates the handpiece within the oral cavity while looking at a monitor [40,41]. The authors hypothesized that younger operators, more familiar with screen-based tasks, adapt more easily to this setting [41]; however, digital literacy was not measured, all operators were novices with the navigation system, and the Generation Z operators were final-year students, so other operator characteristics may also explain the difference. Interestingly, while procedural surgical duration decreased markedly with ongoing clinical repetition, geometric spatial accuracy metrics did not improve significantly over the learning curve, suggesting that the residual error of d-CAS may depend on factors such as tracking resolution, handpiece ergonomics and bone surface resistance rather than on procedural speed [40].
From a clinical and biological standpoint, the magnitude of apical tip displacement is a key safety parameter [39,42,44]. In the anterior palate, although the median and paramedian T-zone provides an average anatomical clearance from the incisor roots, uncontrolled apical deviation may cause iatrogenic damage, including mechanical contact with adjacent permanent dental roots, iatrogenic invasion of the incisive canal, and unwanted breach of the nasal cavity floor [38,39,42]. Furthermore, achieving rigid bicortical anchorage—defined as simultaneous engagement of both the oral cortical plate and the cortical floor of the nasal cavity—is a biological prerequisite for heavy orthopedic therapies such as Miniscrew-Assisted Rapid Palatal Expansion (MARPE) [44]. As documented in critical appraisal of dentoalveolar and periodontal complications following MARPE, failure to establish bicortical anchorage or excessive miniscrew tipping under heavy lateral activation vectors creates stress concentrations at the palatal bone margins, resulting in microcracks, miniscrew loosening, alveolar dehiscence, and undesirable dental buccal tipping [47]. In the two RCTs, mean apical deviations of full-arch, pressure-moulded and 3D-printed guides ranged from 1.25 to 1.49 mm [42,43], whereas mean apical deviations exceeded 3 mm in four of the seven dynamic navigation arms, particularly in steep palates [39,40,41]; whether this difference translates into a clinically relevant difference in safety for orthopedic protocols requiring multicortical anchorage [44] remains to be established.
Methodologically, the literature exhibits substantial variance in how three-dimensional discrepancies are calculated, representing a major hurdle for direct scientific comparability [38,39,40,41,42,43,44,45,46]. Studies employed disparate spatial reference axes, varying surface-to-surface inspection algorithms (iterative closest point registration vs. custom mechanical cylinder matching), and distinct angular projections (true 3D Euclidean angles vs. planar transverse projections) [42,44,46]. In addition, static studies recorded the post-operative position on intraoral scans of the miniscrew head or scan bodies, from which the apex was reconstructed virtually, whereas dynamic studies measured both points on post-operative CBCT [38,39,40,41,42,43,44,45,46]. Specifically, intraoral optical scanning offers sub-50-micrometer surface resolution without metallic artifacts, whereas post-operative CBCT is constrained by voxel size (typically 0.15–0.30 mm) and titanium beam-hardening artifacts that can blur miniscrew boundaries and inflate apparent linear discrepancies by several tenths of a millimeter independently of actual surgical placement error. As demonstrated by comprehensive methodological umbrella assessments of transfer accuracy across digital orthodontic workflows, spatial error must not be viewed as an absolute physical entity, but rather as an operational construct heavily dictated by the mathematical parameters of the superimposition pipeline, region-of-interest segmentation boundaries, and alignment algorithms employed [48]. Furthermore, post-operative clinical validation poses an inherent ethical conflict between measurement precision and patient radiation exposure (ALADA principles) [38,42]. While intraoral surface optical scanning provides a radiation-free method to evaluate coronal head discrepancy, it completely fails to quantify apical tip displacement, root proximity, or nasal cortical engagement [38,42,43]. Conversely, post-operative low-dose CBCT captures true 3D skeletal positions, but introduces ionizing radiation and is affected by beam hardening and streak artifacts generated by titanium or stainless steel miniscrew threads, which complicate automated surface registration [39,44]. In this scenario, the clinical application of deep-learning algorithms and automated artificial intelligence (AI) CBCT segmentation tools may help to overcome these diagnostic limitations, offering automated, operator-independent maxillary bone mapping and artifact-suppressed 3D reconstruction that could improve future pre-operative planning and post-operative accuracy assessments [49].
Finally, several critical limitations inherent to the current evidence base must be acknowledged. Foremost among these is the complete absence of prospective, head-to-head randomized controlled trials comparing s-CAS directly against d-CAS within the same clinical sample; consequently, all comparative assertions between static and dynamic guidance in this review remain indirect and descriptive [38,39,40,41,42,43,44,45,46]. Secondly, the entire body of clinical evidence regarding dynamic navigation originates exclusively from a single academic research team and surgical center, characterized by overlapping recruitment periods and identical outcome dispersion ranges [39,40,41]. This introduces a risk of participant overlap and institutional center bias, which limits the external generalizability of dynamic navigation performance across different clinical settings [39,40,41]. Thirdly, the electronic search strings varied across databases, lacked explicit navigation terms outside PubMed, and did not include parentheses in the Scopus query, while trial registries and grey literature were not searched; although this may have affected retrieval specificity or omitted unpublished studies, the risk of missing eligible clinical trials was mitigated by manual screening of reference lists and cross-checking with prior systematic reviews [28,29,30].

5. Conclusions

In the included studies, both static computer-assisted surgery (s-CAS) and dynamic computer-assisted surgery (d-CAS) achieved mean deviations compatible with miniscrew placement in the anterior palatal T-zone; however, complications such as root contact or nasal floor perforation were rarely reported, and no threshold of clinically acceptable accuracy has been established.
Static surgical templates showed lower reported linear deviations than dynamic navigation, with mean coronal deviations below 1 mm in most arms; this comparison is indirect and may be partly influenced by differences in measurement methods. Within static workflows, single randomized trials showed lower linear deviations with full-arch than with skeletonized guides and lower angular deviation with pressure-moulded than with 3D-printed guides; the mechanisms underlying these differences remain hypothetical.
Dynamic navigation allows direct visual access and intraoperative flexibility without physical templates, but its accuracy was lower in steep palatal vaults and varied with the operator’s generational cohort; all dynamic navigation data derive from a single research group.
Given the current absence of direct head-to-head clinical trials and the potential participant overlap within the dynamic navigation literature, well-designed multicenter randomized controlled trials directly contrasting static and dynamic workflows under standardized superimposition pipelines, possibly supported by automated segmentation tools, are warranted to define evidence-based clinical protocols.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bioengineering13121450/s1, Table S1: PRISMA 2020 checklist; Table S2: Methodological characteristics of the included studies relevant to accuracy measurement.

Author Contributions

Conceptualization, A.V. and E.S.; methodology, C.S. and A.V.; investigation, C.S. and C.F.; data curation, C.S. and C.D.; validation, C.D. and A.V.; formal analysis, C.S. and A.V.; writing—original draft preparation, C.S.; writing—review and editing, C.F., C.D., A.V. and E.S.; visualization, C.D.; supervision, A.V. and E.S.; project administration, E.S. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created in this study. All data analyzed are contained in the cited articles and in the tables of this review.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
3DThree-dimensional
CAD/CAMComputer-aided design/computer-aided manufacturing
CBCTCone-beam computed tomography
d-CASDynamic computer-assisted surgery
DLPDigital light processing
GRADEGrading of Recommendations Assessment, Development and Evaluation
ICPIterative closest point
MARPEMiniscrew-assisted rapid palatal expansion
MSEMaxillary skeletal expander
PICOPopulation, Intervention, Comparison, Outcome
PMPressure-moulded
PRISMAPreferred Reporting Items for Systematic Reviews and Meta-Analyses
PROSPEROInternational Prospective Register of Systematic Reviews
RCTRandomized controlled trial
RoB 2Revised Cochrane risk-of-bias tool for randomized trials
ROBINS-IRisk Of Bias In Non-randomized Studies of Interventions
s-CASStatic computer-assisted surgery
SDStandard deviation
SLAStereolithography
STLStandard tessellation language
TADTemporary anchorage device

References

  1. Iodice, G.; Ludwig, B. Insertion guide for palatal TADs: Advantages and disadvantages from the clinical point of view. Semin. Orthod. 2025, 31, 91–97. [Google Scholar] [CrossRef] [Scilit]
  2. Wilmes, B.; Drescher, D. CAD-CAM Workflows for Palatal TAD Anchored Appliances. Semin. Orthod. 2023, 29, 51–59. [Google Scholar] [CrossRef] [Scilit]
  3. Wilmes, B.; Drescher, D. Conventional versus digital workflows for palatal TADs? Semin. Orthod. 2025, 31, 81–90. [Google Scholar] [CrossRef] [Scilit]
  4. Garib, D.; Miranda, F.; Palomo, J.M.; Pugliese, F.; Da Cunha Bastos, J.C.; Dos Santos, A.M.; Janson, G. Orthopedic outcomes of hybrid and conventional Hyrax expanders: Secondary data analysis from a randomized clinical trial. Angle Orthod. 2021, 91, 178–186. [Google Scholar]
  5. Lo Giudice, A.; Quinzi, V.; Ronsivalle, V.; Martina, S.; Bennici, O.; Isola, G. Description of a Digital Work-Flow for CBCT-Guided Construction of Micro-Implant Supported Maxillary Skeletal Expander. Materials 2020, 13, 1815. [Google Scholar] [CrossRef] [Scilit]
  6. Maino, G.; Turci, Y.; Arreghini, A.; Paoletto, E.; Siciliani, G.; Lombardo, L. Skeletal and Dentoalveolar Effects of Hybrid Rapid Palatal Expansion and Facemask Treatment in Growing Skeletal Class III Patients. Am. J. Orthod. Dentofac. Orthop. 2018, 153, 262–268. [Google Scholar] [CrossRef] [Scilit]
  7. Verma, S.; Kumar, V.; Verma, R.K.; Singh, S.P.; Sharma, S. Effectiveness of Mini-screw and Nance palatal arch for anchorage control following segmental maxillary canine retraction in adolescents with angle Class I crowding using 3D digital models: A randomized clinical trial. J. World Fed. Orthod. 2025, 14, 259–265. [Google Scholar] [CrossRef] [Scilit]
  8. Wilmes, B.; Vasudavan, S.; Drescher, D. CAD-CAM-Fabricated Mini-Implant Insertion Guides for the Delivery of a Distalization Appliance in a Single Appointment. Am. J. Orthod. Dentofac. Orthop. 2019, 156, 148–156. [Google Scholar] [CrossRef] [Scilit]
  9. Iodice, G.; Nanda, R.; Drago, S.; Repetto, L.; Tonoli, G.; Silvestrini-Biavati, A.; Migliorati, M. Accuracy of Direct Insertion of TADs in the Anterior Palate with Respect to a 3D-Assisted Digital Insertion Virtual Planning. Orthod. Craniofac. Res. 2022, 25, 192–198. [Google Scholar] [CrossRef] [Scilit]
  10. Suzuki, S.; Teixeira, T.M.; Pelegrine, A.; Teixeira, M.; Garcez, A.; Suzuki, H. Navigated vs. Freehand Miniscrew Insertion: Enhancing Precision for Predictable Orthodontic Anchorage. J. World Fed. Orthod. 2025, 14, 451–452. [Google Scholar] [CrossRef] [Scilit]
  11. Chen, G.; Chen, S.; Zhang, X.Y.; Jiang, R.P.; Liu, Y.; Shi, F.H.; Xu, T.-M. Stable region for maxillary dental cast superimposition in adults, studied with the aid of stable miniscrews. Orthod. Craniofac. Res. 2011, 14, 70–79. [Google Scholar] [CrossRef] [Scilit]
  12. Gracco, A.; Lombardo, L.; Cozzani, M.; Siciliani, G. Quantitative cone-beam computed tomography evaluation of palatal bone thickness for orthodontic miniscrew placement. Am. J. Orthod. Dentofacial Orthop. 2008, 134, 361–369. [Google Scholar] [CrossRef] [Scilit]
  13. Gracco, A.; Lombardo, L.; Cozzani, M.; Siciliani, G. Assessment of palatal bone thickness in adults with cone beam computerised tomography. Australas. Orthod. J. 2007, 23, 109–113. [Google Scholar] [CrossRef] [Scilit]
  14. Poon, Y.-C.; Chang, H.-P.; Tseng, Y.-C.; Chou, S.-T.; Cheng, J.-H.; Liu, P.-H.; Pan, C.-Y. Palatal bone thickness and associated factors in adult miniscrew placements: A cone-beam computed tomography study. Kaohsiung J. Med. Sci. 2015, 31, 265–270. [Google Scholar] [CrossRef] [Scilit]
  15. Wilhelmy, L.; Willmann, J.H.; Tarraf, N.E.; Wilmes, B.; Drescher, D. Maxillary Space Closure Using a Digital Manufactured Mesialslider in a Single Appointment Workflow. Korean J. Orthod. 2022, 52, 236–245. [Google Scholar] [CrossRef] [Scilit]
  16. Cantarella, D.; Savio, G.; Grigolato, L.; Zanata, P.; Berveglieri, C.; Lo Giudice, A.; Isola, G.; Del Fabbro, M.; Moon, W. A new methodology for the digital planning of micro-implant-supported maxillary skeletal expansion. Med. Devices Evid. Res. 2020, 13, 93–106. [Google Scholar] [CrossRef] [Scilit]
  17. Tao, T.; Zou, K.; Jiang, R.; He, K.; He, X.; Zhang, M.; Wu, Z.; Shen, X.; Yuan, X.; Lai, W.; et al. Artificial Intelligence-Assisted Determination of Available Sites for Palatal Orthodontic Mini Implants Based on Palatal Thickness through CBCT. Orthod. Craniofac. Res. 2023, 26, 491–499. [Google Scholar] [CrossRef] [Scilit]
  18. Altieri, F.; Luzzi, V.; Mezio, M.; Polimeni, A.; Cassetta, M. Computer-Guided Miniscrew Insertion in the Paramedian and Parapalatal Areas of the Palatal Vault: Failure Rate and Learning Curve Required to Obtain Predictable Results. Int. J. Comput. Dent. 2024, 27, 345–353. [Google Scholar]
  19. De Stefano, A.; Guarnieri, R.; Fiorelli, B.; Barbato, E.; Galluccio, G. Clinical Experience and Digital Knowledge in Virtual Planning of Palatal Orthodontic Miniscrew Insertion. Appl. Sci. 2023, 13, 7474. [Google Scholar] [CrossRef] [Scilit]
  20. Cantarella, D.; Karanxha, L.; Zanata, P.; Moschik, C.; Torres, A.; Savio, G.; Del Fabbro, M.; Moon, W. Digital Planning and Manufacturing of Maxillary Skeletal Expander for Patients with Thin Palatal Bone. Med. Devices Evid. Res. 2021, 14, 299–311. [Google Scholar] [CrossRef] [Scilit]
  21. Fouda, A.S.; Afify, A.K.; Aboulfotouh, M.H.; Attia, K.H.; Abouelezz, A.M.; Elkordy, S.A. Dental arch changes after anterior open bite treatment in the mixed dentition produced by miniscrew-supported palatal crib vs conventional fixed palatal crib: A randomized clinical trial. Angle Orthod. 2022, 92, 487–496. [Google Scholar] [CrossRef] [Scilit]
  22. Benetti, M.; Montresor, L.; Cantarella, D.; Zerman, N.; Spinas, E. Does Miniscrew-Assisted Rapid Palatal Expansion Influence Upper Airway in Adult Patients? A Scoping Review. Dent. J. 2024, 12, 60. [Google Scholar] [CrossRef] [Scilit]
  23. Sicca, N.; Benedetti, G.; Nieri, A.; Vitale, S.; Lopponi, G.; Mura, S.; Verdecchia, A.; Spinas, E. Comparison of Side Effects Between Miniscrew-Assisted Rapid Palatal Expansion (MARPE) and Surgically Assisted Rapid Palatal Expansion (SARPE) in Adult Patients: A Scoping Review. Dent. J. 2025, 13, 47. [Google Scholar] [CrossRef] [Scilit]
  24. Irlandese, G.; Perrotta, G.; Marsili, V.; Carboni, L.; Verdecchia, A.; Spinas, E. Comparative Evaluation of the Effects of Miniscrew and Miniplate Skeletal Anchorage in the Orthopedic Treatment of Growing Class III Malocclusion: A Systematic Review and Meta-Analysis. Bioengineering 2025, 12, 1065. [Google Scholar] [CrossRef] [Scilit]
  25. Vachiramon, A.; Urata, M.; Kyung, H.M.; Yamashita, D.-D.; Yen, S.L.-K. Clinical Applications of Orthodontic Microimplant Anchorage in Craniofacial Patients. Cleft Palate Craniofac. J. 2009, 46, 136–146. [Google Scholar] [CrossRef] [Scilit]
  26. Busson, F.; Charavet, C. Homemade CAD-CAM surgical guide for palatal mini-screw insertion for maxillary molar distalization providing anchorage. A two times original protocol. L’Orthod. Francaise 2024, 95, 189–203. [Google Scholar]
  27. Schwärzler, A.; Ludwig, B.; Chitan, P.; Lettner, S.; Sagl, B.; Jonke, E. Transfer Accuracy of 3D Printed versus CAD/CAM Milled Surgical Guides for Temporary Orthodontic Implants: A Preclinical Micro CT Study. J. Dent. 2024, 146, 105060. [Google Scholar] [CrossRef] [Scilit]
  28. Stucki, L.; Kauke-Navarro, M.; Schwab, J.; Hoenigl, A.; Safi, A.-F. Impact of guide design and accuracy of mini-implant placement in the anterior palate: A systematic review. Int. Orthod. 2026, 24, 101169. [Google Scholar] [CrossRef] [Scilit]
  29. Jedliński, M.; Janiszewska-Olszowska, J.; Mazur, M.; Ottolenghi, L.; Grocholewicz, K.; Galluccio, G. Guided Insertion of Temporary Anchorage Device in Form of Orthodontic Titanium Miniscrews with Customized 3D Templates—A Systematic Review with Meta-Analysis of Clinical Studies. Coatings 2021, 11, 1488. [Google Scholar] [CrossRef] [Scilit]
  30. Santmartí-Oliver, M.; Jorba-García, A.; Moya-Martínez, T.; de-la-Rosa-Gay, C.; Camps-Font, O. Safety and Accuracy of Guided Interradicular Miniscrew Insertion: A Systematic Review and Meta-Analysis. J. Clin. Med. 2024, 13, 7697. [Google Scholar] [CrossRef] [Scilit]
  31. Page, M.J.; McKenzie, J.E.; Bossuyt, P.M.; Boutron, I.; Hoffmann, T.C.; Mulrow, C.D.; Shamseer, L.; Tetzlaff, J.M.; Akl, E.A.; Brennan, S.E.; et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ 2021, 372, n71. [Google Scholar] [CrossRef] [Scilit]
  32. Cohen, J. A Coefficient of Agreement for Nominal Scales. Educ. Psychol. Meas. 1960, 20, 37–46. [Google Scholar] [CrossRef] [Scilit]
  33. Sterne, J.A.C.; Savović, J.; Page, M.J.; Elbers, R.G.; Blencowe, N.S.; Boutron, I.; Cates, C.J.; Cheng, H.-Y.; Corbett, M.S.; Eldridge, S.M.; et al. RoB 2: A Revised Tool for Assessing Risk of Bias in Randomised Trials. BMJ 2019, 366, l4898. [Google Scholar] [CrossRef] [Scilit]
  34. Sterne, J.A.; Hernán, M.A.; Reeves, B.C.; Savović, J.; Berkman, N.D.; Viswanathan, M.; Henry, D.; Altman, D.G.; Ansari, M.T.; Boutron, I.; et al. ROBINS-I: A Tool for Assessing Risk of Bias in Non-Randomised Studies of Interventions. BMJ 2016, 355, i4919. [Google Scholar] [CrossRef] [Scilit]
  35. Guyatt, G.H.; Oxman, A.D.; Vist, G.E.; Kunz, R.; Falck-Ytter, Y.; Alonso-Coello, P.; Schünemann, H.J.; GRADE Working Group. GRADE: An emerging consensus on rating quality of evidence and strength of recommendations. BMJ 2008, 336, 924–926. [Google Scholar] [CrossRef] [Scilit]
  36. Schünemann, H.J.; Cuello, C.; Akl, E.A.; Mustafa, R.A.; Meerpohl, J.J.; Thayer, K.; Morgan, R.L.; Gartlehner, G.; Kunz, R.; Katikireddi, S.V.; et al. GRADE guidelines: 18. How ROBINS-I and other tools to assess risk of bias in nonrandomized studies should be used to rate the certainty of a body of evidence. J. Clin. Epidemiol. 2019, 111, 105–114. [Google Scholar] [CrossRef] [Scilit]
  37. Campbell, M.; McKenzie, J.E.; Sowden, A.; Katikireddi, S.V.; Brennan, S.E.; Ellis, S.; Hartmann-Boyce, J.; Ryan, R.; Shepperd, S.; Thomas, J.; et al. Synthesis without meta-analysis (SWiM) in systematic reviews: Reporting guideline. BMJ 2020, 368, l6890. [Google Scholar] [CrossRef] [Scilit]
  38. Al-Gazzawi, A.M.Q.; Knode, V.; Ludwig, B.; Othman, A.; Salamini, A.; Pandis, N.; Fleming, P.S. Midpalatal Miniscrew Insertion: The Accuracy of Digital Planning and Surgical Placement. Am. J. Orthod. Dentofac. Orthop. 2024, 166, 69–75. [Google Scholar] [CrossRef] [Scilit]
  39. Brilli, D.; Cauli, I.; Cassetta, M. How Palatal Vault Morphology and Screw Length Influence the Accuracy of Dynamic Computer-Guided Orthodontic Miniscrew Insertion. A Prospective Clinical Study. J. Dent. 2024, 146, 105093. [Google Scholar] [CrossRef] [Scilit]
  40. Brilli, D.; Giansanti, M.; Bertoldo, S.; Cauli, I.; Cassetta, M. Dynamic Navigation System Accuracy in Orthodontic Miniscrew Insertion in the Palatal Vault: A Prospective Single-Arm Clinical Study. Int. J. Comput. Dent. 2026, 29, 23–31. [Google Scholar]
  41. Brilli, D.; Giansanti, M.; Cassetta, M. Does the Operator’s Generational Cohort Influence the Accuracy of Orthodontic Miniscrews Insertion in the Palatine Vault in Dynamic Computer-Guided Surgery? A Randomized Prospective Study. Oral Maxillofac. Surg. 2025, 29, 120. [Google Scholar] [CrossRef] [Scilit]
  42. Hoffmann, L.; Kakoschke, T.K.; Keller, A.; Hötzel, L.; Baumert, U.; Otto, S.; Wichelhaus, A.; Sabbagh, H. Accuracy of Guided Insertion of Orthodontic Temporary Anchorage Devices Comparing Two Different 3D Printed Surgical Guide Designs: A Randomized Controlled Trial. Sci. Rep. 2025, 15, 31749. [Google Scholar] [CrossRef] [Scilit]
  43. Möhlhenrich, S.C.; Ayad, I.; Linnerz, F.; Xi, T.; Vinayahalingam, S.; Bergé, S.; Danesh, G.; Chhatwani, S. Clinical Transfer Accuracy of Pressure-Moulded versus 3D-Printed Drilling Guides for Orthodontic Mini-Implants in the Anterior Palate: A Randomized Prospective Clinical Study. Sci. Rep. 2026, 16, 50346. [Google Scholar] [CrossRef] [Scilit]
  44. Palone, M.; Pepe, F.; Grotteria, C.; Cremonini, F.; Maino, G.B.; Brucculeri, L.; Lombardo, L. Factors Influencing the Accuracy of Guided Palatal Miniscrew Insertion in Young Adults: Prospective in Vivo Study. J. Orofac. Orthop. 2026. [Google Scholar] [CrossRef] [Scilit]
  45. Ronsivalle, V.; Venezia, P.; Bennici, O.; D’Antò, V.; Leonardi, R.; Lo Giudice, A. Accuracy of Digital Workflow for Placing Orthodontic Miniscrews Using Generic and Licensed Open Systems. A 3d Imaging Analysis of Non-Native.Stl Files for Guided Protocols. BMC Oral Health 2023, 23, 494. [Google Scholar] [CrossRef] [Scilit]
  46. Pozzan, L.; Migliorati, M.; Dinelli, L.; Riatti, R.; Torelli, L.; Di Lenarda, R.; Contardo, L. Accuracy of the digital workflow for guided insertion of orthodontic palatal TADs: A step-by-step 3D analysis. Prog. Orthod. 2022, 23, 27. [Google Scholar] [CrossRef] [Scilit]
  47. Frenna, B.; Grimaldi, R.; Fiandaca, S.; Basha, R.; Caprio, M.; Rizzo, G.E.M.; Verdecchia, A.; Spinas, E. Diagnostic Assessment of Periodontal and Dentoalveolar Complications Following Mini-Screw-Assisted Rapid Palatal Expansion in Adults and Late Adolescents: A Systematic Review. Diagnostics 2026, 16, 352. [Google Scholar] [CrossRef] [Scilit]
  48. Lalli, E.; Verdecchia, A.; Parrini, S.; Rossini, G.; Malagraba, F.E.; Beti, M.M.; Marchese, E.; Spinas, E. Transfer Accuracy in Digital Indirect Bonding: A Methodological Umbrella Review of Definitions, Measurement Frameworks, and Evidence Synthesis. Bioengineering 2026, 13, 607. [Google Scholar] [CrossRef] [Scilit]
  49. Iacob, A.M.; Verdecchia, A.; García-Mesa, Y.; Spinas, E.; Cobo, T. Effectiveness of automated segmentation of maxillofacial structures in cone-beam computed tomography images using artificial intelligence: A systematic review. Int. Orthod. 2026, 24, 101081. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.