Skip to Content
BioengineeringBioengineering
  • Review
  • Open Access

24 November 2026

30 Pages

Implant Characteristics in Posterior Pedicle-Screw Fusion for Adolescent Idiopathic Scoliosis: Biomechanical Principles and Clinical Outcomes

,
,
and
1
2nd Clinical Department of General and Gastroenterological Surgery, Medical University of Bialystok Clinical Hospital, ul. M. Skłodowskiej-Curie 24a, 15-276 Bialystok, Poland
2
Department of Neurosurgery, Regional Specialized Hospital, ul. Dekerta 1, 66-400 Gorzow, Poland
3
Department of Neurosurgery, Polish Mother’s Memorial Hospital Research Institute, Rzgowska 281/289, 93-338 Lodz, Poland
*
Author to whom correspondence should be addressed.
Bioengineering2026, 13(12), 1431;https://doi.org/10.3390/bioengineering13121431 
(registering DOI)
This article belongs to the Special Issue Biomedical Engineering Strategies for Scoliosis Treatment

Abstract

This critical narrative review examines implant characteristics in posterior fusion for adolescent idiopathic scoliosis. Reported PubMed/MEDLINE and Google Scholar searches ended on 15 September 2026. Direct applicability required posterior-only, all-pedicle-screw fusion to be before the eighteenth birthday. Three nonrandomized implant comparisons qualified, including a combined-strategy cohort after author clarification of operative ages. Equal-diameter titanium and cobalt-chromium rods had identical observed mean thoracic correction without establishing equivalence. Monoaxial screws showed greater coronal correction than polyaxial screws without demonstrated patient-reported superiority; endpoint timing was inconsistent. Larger screws with thicker cobalt-chromium rods showed greater correction, but simultaneous component changes precluded attribution to one property. A separate eligible radiographic patient-reported outcome association study did not compare implants. The contextual Minimize Implants Maximize Outcomes trial established equivalence only for two-year coronal correction in primary thoracic curves of 45–65°; strict operative-age eligibility remained unresolved, and safety or patient-reported equivalence was not established. The age cutoff defines pediatric scope, not biological differences. These findings establish neither universal implant superiority nor a validated selection algorithm.

1. Introduction

Posterior fusion for idiopathic scoliosis aims to maintain deformity correction while providing safe fixation, appropriate spinal balance, and preservation of unfused mobile segments [1,2,3,4,5]. Relevant outcomes include coronal and sagittal alignment, rotation, patient-reported outcome measures (PROMs), and subsequent surgery [6,7]. Corrective maneuvers, releases, and anchor configuration influence the treatment delivered and can confound comparisons of implant specifications [8,9,10,11]. Historical or mixed-population placement studies and hybrid-construct comparisons provide background rather than direct estimates for the population defined in this review [1,2,3,4,9,12,13]. Screw-head design and correction technique should therefore be considered together [14,15]. Implant-related injury can involve screw trajectory, adjacent structures, or mechanical movement of the fixation [12,16,17]. Computational studies distinguish correction from loads at the bone–screw interface, while clinical rod-contour measurements show that implanted rods do not necessarily retain their intended geometry [18,19]. Studies of pedicle expansion address a separate anatomical question [20]. Similarly, simultaneously changing screw dimensions, rod diameter, and rod alloy evaluates a combined treatment strategy and cannot identify the effect of one component [21,22]. Deep infection and reoperation remain important outcomes, although broad surgical reviews and mixed-age cohorts cannot provide implant-specific risk estimates for the strict pediatric population [23,24].
Recent randomized density trials and registry studies extend the earlier rod literature, although their applicability depends on operative age, anchor composition, and the comparison being made [25,26,27]. This critical narrative review examines pedicle-screw dimensions and head design, rod diameter and alloy, and regional anchor distribution in posterior fusion for adolescent idiopathic scoliosis (AIS) performed before age 18. The sagittal analysis concerns thoracic kyphosis within the overall AIS sagittal profile, rather than structural kyphosis of another etiology. Table 1 summarizes the relationships between implant properties, construct mechanics, and clinical outcomes, together with the patient and operative factors that affect their interpretation. Previous reviews have synthesized rod-material, rod-diameter, or implant-density comparisons [28,29,30,31]. This review adds an explicit assessment of applicability to posterior-only, all-pedicle-screw fusion performed before age 18 and separates component-focused comparisons from combined implant strategies. It evaluates radiographic correction and maintenance alongside PROMs, implant-related events, and revision, with attention to outcome-specific follow-up and the definitions used in each study.
Table 1. Relationships between implant characteristics, construct mechanics, and clinical outcomes in posterior all-pedicle-screw fusion for AIS.

2. Materials and Methods

This critical narrative review appraises a nonexhaustive cited corpus within an intended publication window of 1 January 2006, through 17 October 2026; earlier technical reports provide historical background. According to the retained search account, the original searches used PubMed/MEDLINE and Google Scholar and were completed on 15 September 2026, with supplementary topic and title lookups and reference checking. Individual search dates, complete queries, exports, and screening counts are unavailable. The stated publication cutoff is 17 October 2026, not the original search completion date. The available records do not document a supplementary search covering 16 September through 17 October, so complete coverage through that cutoff cannot be established. An earlier supplementary version records targeted revision checks on 9 November 2026, and selected additional source checks were described for 10 November. These historical entries are distinguished from the present verification on 12 November 2026, in Supplementary Table S2. The revision verification described in Table S2 used DOI and title matching, PubMed-derived records through Europe PMC, accessible full texts and publisher pages, and targeted discovery through the research services identified in that table. It retained the publication cutoff and did not reconstruct a systematic search. Figure 1 describes the cited bibliography and its evidentiary roles, not records retrieved or screened.
Figure 1. The inventory contains 71 reference entries: 70 principal publications and one trial webpage. The linked Falick-Michaeli erratum is retained within its parent entry. Three publications provide confirmed implant comparisons; one is an eligible radiographic–PROM association study, and the remaining 67 entries provide contextual, mechanistic, secondary, methodological, or trial-documentation evidence. The reported original searches ended on 15 September 2026; individual search dates, complete queries, exports, and screening counts remain unavailable. Coverage through the stated publication cutoff of 17 October 2026, cannot be verified. These are bibliography counts, not search or study-selection counts. Supplementary Tables S1, S2 and S4 provide the classification and audit trail.
Direct clinical applicability required AIS and surgery before the patient’s eighteenth birthday. This threshold defines a focused pediatric operative population; it is not a threshold for skeletal maturity, bone quality, or implant behavior. Age at diagnosis, enrollment, pregnancy, or follow-up cannot substitute for operative age, and a mean age below 18 does not demonstrate that every operation met the criterion. Reported operative ages for the analyzed cohort or an extractable subgroup, or an explicit source-author clarification, were used to establish the age criterion. Recruitment criteria permitting age 18 or older do not by themselves establish that such patients underwent surgery. Where actual operative ages could not be verified, eligibility remained unresolved; where reported operative ages included 18 or older without an extractable eligible comparison, the study was outside the direct age scope. Earlier-onset idiopathic scoliosis was outside the direct scope unless an AIS subgroup was separable. These restrictions reduce direct coverage and can preferentially exclude otherwise informative studies because of reporting detail. The review therefore presents broader AIS evidence separately and discusses how interpretation changes when the age restriction is relaxed, without claiming a formal sensitivity meta-analysis.
The operative criterion required definitive posterior-only fusion with exclusively pedicle-screw anchors but did not require bilateral screws at every fused vertebra. Hybrid constructs, anterior or combined approaches, growing constructs, tethering, and nonidiopathic deformities were outside the direct synthesis unless an eligible subgroup was separable. They could remain as explicitly labeled contextual evidence. For example, Skalak et al. included both all-screw and screw-and-hook constructs despite a qualifying age range [32]. Population and construct eligibility were assessed separately from study design and risk of bias. Eligibility was classified as confirmed, unresolved, or outside the criteria; evidence roles distinguished implant comparisons, combined-strategy comparisons, association studies, descriptive safety reports, and mechanistic or secondary evidence. Appraisal considered allocation, baseline flexibility and alignment, co-interventions, outcome definitions and timing, missing data, and surveillance. Supplementary Table S1 gives source locations and reasons for the principal eligibility decisions. Contextual sources were used to address implant mechanics, broader AIS comparisons, outcome measurement, safety, or long-term clinical considerations; their inclusion does not imply exhaustive coverage of those topics. Laboratory, computational, and bone-regeneration studies provide indirect context and do not contribute to clinical implant comparisons.
Outcomes included coronal correction and maintenance, thoracic kyphosis with specified vertebral landmarks, lumbar lordosis, rotation and balance, questionnaire-specific PROMs, implant events, infection, pseudarthrosis, junctional complications, and unplanned reoperation. Extraction prioritized full-text methods for eligibility, results and tables for endpoint values and timing, and abstracts when full texts were unavailable. Study characteristics, implants, correction maneuvers, density denominators, response denominators, and adverse-event surveillance were recorded when recoverable. Endpoint-specific assessment times were distinguished from the minimum or mean follow-up of the overall cohort. Unreported information was distinguished from information not verifiable from the accessed source; neither was interpreted as a zero event. Conflicting passages remained unresolved unless clarified by the source authors; the reported corresponding-author clarification of operative ages in the study cohort [21] is documented in Supplementary Table S1 and Section S4. Bibliographic matching did not override discrepancies within a publication.
The main findings were organized around three confirmed implant comparisons: two focusing on rod alloy or screw-head design and one evaluating a combined implant strategy. The eligible radiographic–PROM association study was interpreted separately. Other comparisons were retained as contextual evidence with their applicability limits. Reviews assisted discovery and interpretation but added no independent participants [28,29,30,31]. No independent duplicate screening, formal risk-of-bias instrument, GRADE assessment, or clinical meta-analysis was used. Equivalence was accepted only for the specified endpoint and margin supported by its confidence interval. Calculations were limited to explicitly identified mechanical relationships and arithmetic illustrations. SANRA informed narrative reporting without a numerical quality score [33]. The historical pregnancy search accounts were referenced in PubMed/MEDLINE and Google Scholar. Adult age at pregnancy was permitted, while the index operation still had to satisfy the review’s age and construct criteria. Supplementary Section S3 and Table S3 separate radiographic, mechanical, symptom, and obstetric outcomes. Section S4 records source discrepancies and verification limits; Table S4 lists every cited source, its identifier, access level, and evidentiary role. AI assistance supported the literature discovery, source auditing, language editing, and figure preparation; its role and verification limits are disclosed in the acknowledgments and Supplementary Table S2.

3. Direct Clinical Implant Comparisons

Three comparisons in the cited corpus met the population and construct criteria. Two focused on rod alloy or screw-head design, while the third compared combined implant strategies. Their findings are presented separately because a multicomponent comparison cannot identify the effect of one implant property. Eligibility does not imply low risk of bias. The eligible radiographic–PROM association study is discussed separately in Section 10; other clinical comparisons remain contextual.

3.1. Equal-Diameter Titanium and Cobalt-Chromium Rods

Yang et al. studied 45 patients aged 11–17 with Lenke 6 curves: 29 received 6.0 mm titanium rods, and 16 received 6.0 mm cobalt-chromium (CoCr) rods. Posterior segmental pedicle screws and double rod derotation were used, with at least five years of follow-up. Mean thoracic correction was 71.2% in both groups; no significant thoracic kyphosis or sagittal vertical axis difference was reported, while postoperative lumbar lordosis was greater with CoCr [6]. The study was described as prospective and nonrandomized. Groups were small, and follow-up was unequal; no validated PROM was collected. In addition, some subgroup counts in the source table do not reconcile with the reported group sizes, and lumbar correction percentages are assigned inconsistently between passages. The consistent thoracic result is retained, but those discrepancies are documented in Supplementary Section S4. The study does not establish material equivalence or a material-specific difference in uncommon complications.

3.2. Monoaxial and Polyaxial Pedicle Screws

Yang et al. compared 23 monoaxial and 23 polyaxial constructs in patients aged 11–17. Both used posterior segmental pedicle screws and rod derotation, but the cohorts underwent surgery in different years, and the polyaxial maneuver differed. Reported main-curve correction was 70.2% versus 65.3% (p = 0.040), without a significant apical derotation or Scoliosis Research Society-22 (SRS-22) questionnaire difference [14]. This was a retrospective comparison, not randomized treatment allocation. The source required at least two years of follow-up but labeled follow-up radiographs and the SRS-22 assessment as three-month outcomes, while other passages referred to final follow-up. Thus, the reported coronal difference cannot be presented as verified durable two-year superiority. Baseline sagittal differences, treatment era, co-interventions, and unclear endpoint timing limit inference.

3.3. Combined Implant-Strategy Comparison

Grabala et al. compared smaller screws with 5.5 mm titanium rods against larger screws with 6.0 mm cobalt-chromium rods in 260 patients [21]. The corresponding author clarification reported in Supplementary Table S1 confirmed that all patients underwent scoliosis surgery before age 18. The study therefore contributes a third eligible clinical comparison. Reported mean major-curve correction was 69.8% versus 74.5%, without a significant final SRS-22r total-score difference. Simultaneous changes in screw dimensions, rod diameter, and alloy preclude component-specific attribution. Source-data inconsistencies are addressed in Supplementary Section S4 and remain separate from the resolved age-eligibility question.

4. Biomechanical Framework for Pedicle-Screw–Rod Constructs

Loads are transferred through the rod, locked screw head, bone–screw interface, and instrumented spine. Increasing rod stiffness does not necessarily improve correction when anchor displacement, spinal resistance, or the chosen maneuver becomes limiting [18,34,35,36]. For an ideal solid circular beam, flexural rigidity is EI, where E is Young’s modulus and I is the second moment of area, I = πd4/64. At constant E, increasing diameter from 5.5 to 6.0 mm increases ideal flexural rigidity by approximately 42%. At constant diameter, a higher material modulus also increases ideal rigidity. These calculations explain a mechanical relationship but do not predict clinical correction. Actual rods are curved, plastically contoured, loaded nonlinearly and cyclically, and exposed to set-screw contact and local notching. Anchor spacing and loading direction also affect construct behavior. Relative to a 5.5 mm rod of the same material modulus, ideal rigidity ratios are 0.683, 1.000, 1.416, and 1.777 for diameters of 5.0, 5.5, 6.0, and 6.35 mm, respectively (Table 2 and Figure 2).
Table 2. Calculated ideal flexural rigidity at the same material modulus.
Figure 2. Calculated effect of rod diameter on ideal flexural rigidity at a constant material modulus. Relative rigidity was calculated as (d/5.5)4 for ideal solid circular rods using EI = Eπd4/64. Diameters of 5.0, 5.5, 6.0, and 6.35 mm give relative values of 0.683, 1.000, 1.416, and 1.777, respectively. The reference is a 5.5 mm rod of the same material modulus. These are elementary mechanical calculations, not patient or laboratory measurements. They do not compare alloys or predict clinical correction, fatigue endurance, or implant-related event rates.
A stiffer elastic component deforms less under a given load but requires more force to achieve the same deformation. Preserving rod contour and seating the rod are therefore different mechanical challenges. Elastic stiffness, yielding, plastic deformation, and fatigue resistance must also be distinguished. Lindsey et al. found reduced fatigue life of titanium implants after French-bender contouring [35]. Slivka et al. tested 4.5 mm rods, tube-bender contouring, and a single rebending step under different conditions [36]. These studies used different materials, diameters, contouring methods, and loading protocols, so their findings are not directly contradictory. Neither establishes pediatric rod-fracture risk or justifies repeated contouring of 5.5 or 6.0 mm rods. Failure location matters as well: a test ending in screw failure does not determine the rod’s ultimate endurance. Loading method, fixation geometry, and the failed component must be considered when relating bench findings to corrective maintenance, anchor integrity, and symptomatic failure [18,36]. Bone engagement depends on outer and core screw diameters, thread geometry, insertion depth, and pedicle anatomy. Nominal screw diameter is therefore an incomplete description of the interface [37]. In simulations based on five patient models, Clin et al. found that screw-head motion affected predicted correction and bone–screw forces, while greater implant density reduced predicted postoperative loads on individual screws [18]. These findings support considering correction and interface loading together. Commercial 5.5 mm rods also show supplier-related differences in measured stiffness and yield behavior within the same nominal alloy category, as demonstrated in four-point bending tests by Upasani et al. These are bench findings, not clinical superiority estimates [38].
The bone–screw interface also depends on living tissue and healing, not geometry alone. Li et al. describe interactions among innervation, vascular supply, and osteogenesis during bone regeneration [39]. This provides a biological framework for considering interface adaptation, but it does not establish that a particular pedicle-screw thread, diameter, or coating improves fixation in AIS. Meng et al. describe a trabeculae-like bone-filling material intended to support osteogenic cells and local delivery of therapeutic agents [40]. That biomaterial concept is distinct from load-bearing pedicle-screw fixation and supplies no clinical comparison of AIS implant designs.

5. Rod Diameter: Comparative Evidence and Interpretation

Isolating the effect of rod diameter requires comparable alloy, contour, anchor configuration, curve flexibility, and correction technique. Clinical reports often compare combined strategies, such as larger cobalt-chromium rods against smaller titanium rods. Prince et al. evaluated 352 patients treated with 5.5 or 6.35 mm rods, with rod choice related to patient and construct characteristics. Huang et al. compared 93 patients treated with those diameters in different periods and with different implant systems. Neither report was verified as an exclusively under-18, all-pedicle-screw comparison in the material assessed. Their findings therefore describe delivered treatment strategies rather than an isolated dose–response relationship for diameter. Table 3 distinguishes these studies from equal-diameter alloy comparisons [22,41].
Table 3. Rod diameter, alloy, and combined implant-strategy comparisons.
Diameter comparisons should distinguish initial correction from maintenance and report baseline standing and bending angles, early correction, and final alignment. Bowden et al. identified 75 AIS studies using pedicle-screw fixation systems, including 46 that reported rod materials or diameters. Their cross-study complication proportions do not isolate a diameter effect because populations, procedures, and surveillance differed [28]. Grabala et al. compared two combined strategies in 260 patients: smaller screws with 5.5 mm titanium rods (n = 132) and larger screws with 6.0 mm cobalt-chromium rods (n = 128). Reported mean major-curve correction was 69.8% and 74.5%, respectively [21]. The corresponding author clarified that the patient described as 18 years old in the source article’s Figure 1 underwent surgery several days before her eighteenth birthday and that the stated age referred to the first postoperative standing radiograph. The author also confirmed that every patient in the cohort underwent scoliosis surgery before age 18. The study therefore meets the age criterion and is included as an eligible combined-strategy comparison, as documented in Supplementary Table S1 and Section S4. These values are not used to infer a maintenance advantage. Simultaneous changes in screw size, rod diameter, and alloy preclude component-specific attribution.
Ruffilli et al. compared 5.5 mm CoCr rods in 30 patients with 6.0 mm titanium-alloy rods in 34 Lenke 1 patients, using high-density constructs and translation/direct vertebral rotation. CoCr was associated with greater coronal correction, whereas hypokyphotic patients had a greater increase in kyphosis with titanium. Both diameter and alloy differed, and the reported age range included 18, making this contextual evidence [43]. No verified 5.0 mm versus 5.5 or 6.0 mm comparison meeting the strict criteria was identified in the cited corpus. This limited coverage does not demonstrate inferiority of smaller rods or establish an optimal diameter.

6. Contextual Evidence on Rod Alloy

The directly applicable equal-diameter comparison is presented in Section 3.1. The broader AIS literature adds an important randomized comparison: Sakai et al. permitted recruitment of girls aged 10–19 years and randomized 69 patients to 6.0 mm titanium or CoCr rods. No significant one-year radiographic or questionnaire differences were reported. Without a verified under-18 subgroup, this remains contextual randomized evidence and cannot establish strict-age equivalence or equal late revision risk [42]. Other comparisons address different contrasts. Lamerain’s 2014 study compared CoCr with stainless steel, Sabah compared equal-diameter titanium with CoCr, and Etemadifar used a mixed CoCr–titanium pair against bilateral titanium [45]. Strict operative-age eligibility was not established from the accessed material, and these comparisons should not be merged into a single uniform alloy effect [45,46,47]. Kumar et al.’s 2026 review synthesized six comparisons and reported no significant differences in the analyzed outcomes [29]. It supplies broader AIS context, not independent patients or an exclusively under-18 estimate [29]. Differences from kyphosis-focused syntheses require attention to included studies, comparators, and measurement endpoints [28].
Todderud et al.’s registry analysis included 1348 Lenke 1–4 patients, with markedly unequal groups: 42 received 5.5 mm titanium, 651 5.5 mm CoCr, 586 5.5 mm steel, 46 6.0 mm titanium, and 23 6.0 mm CoCr rods. The largest observed kyphosis gain was with 6.0 mm titanium. Strict operative-age and exclusive-anchor eligibility remained unresolved. This contextual association does not establish causal superiority, particularly given potential center, surgeon, era, and treatment-selection effects [26]. Table 3 preserves individual group sizes rather than implying equal precision across the registry comparisons.

7. Pedicle Screw Diameter, Purchase, and Anatomic Safety

Screw diameter affects both the potential mechanical purchase and the requirement imposed on a pedicle. A thoracic pedicle that is substantially narrower than the proposed implant presents a different risk-benefit calculation from a broad lumbar pedicle. The relevant dimensions include the narrowest transverse diameter, cortical thickness, pedicle orientation, and vertebral rotation. AIS alters these structures asymmetrically; concave pedicles are often anatomically constrained, and the location of the spinal cord and major vessels makes a seemingly small deviation clinically consequential [12,16,37]. Sarwahi et al. compared 6.5 mm screws at all levels with 5.0–5.5 mm screws at all levels in 412 AIS patients. Analyses included flexibility strata and matching by surgeon and treatment year. The larger-screw group had greater coronal correction and postoperative kyphosis; 87.6% improved by at least one apical rotation grade. This percentage describes patients crossing a grading threshold, not 87.6% rotational correction. CT-based accuracy assessment was limited to patients with available scans, and the complete operative-age range was not established. The study therefore provides relevant observational context, without defining a universally safe size or a verified exclusively under-18 effect [37].
Screw-size studies should record planned and inserted dimensions at each level, because grouping by maximum, mean, or prescribed all-level diameter defines different exposures. Length changes the engaged trajectory and should be analyzed separately from diameter; no qualifying comparison isolated its effect on correction or revision. A screw-to-pedicle ratio must specify the imaging plane and whether the denominator is the outer cortical or inner cancellous diameter. The eligible evidence does not establish a universal ratio. Placement accuracy, corrective performance, and clinical injury require separate assessment: successful insertion does not demonstrate superior correction, and absence of neurologic injury does not establish cortical integrity [17,37].

Contextual Screw-Head Comparisons

The confirmed monoaxial/polyaxial comparison is presented in Section 3.2. Contextual studies evaluate different head designs and maneuvers. Dalal et al. compared 95 uniplanar with 115 polyaxial bilateral screw constructs using 5.5 mm steel rods. Coronal correction did not differ significantly, while residual apical rotation grades favored uniplanar screws at six weeks and one year [48]. Hu et al. compared uniplanar screws with polyaxial screws plus vertebral derotation in 50 Lenke 5C patients. Coronal correction was 74.4% and 73.9%; apical derotation and two-year SRS-22 self-image favored the uniplanar technique [49]. The complete operative-age range was not verified for either study, so neither contributes a confirmed strict-age implant effect. Future comparisons should specify screw-head design by level and side, rod characteristics, regional density, derotation technique, baseline sagittal alignment, and thoracoplasty.

8. Anchor Density and Strategic Distribution

Global screw density measures anchor quantity across the fusion span; regional distribution identifies placement at the apex, concavity, and terminal levels. Because low-, moderate-, and high-density labels are not standardized, Table 4 retains each study’s thresholds and denominator [25,27,50].
Table 4. Screw size, head design, density, and regional distribution.
Studies with unresolved eligibility remain contextual. Numerical comparisons are descriptive unless a formal equivalence design is identified. Computational findings from Wang et al. [34] distinguish apical derotation from overall anchor loading and are not clinical outcomes. Here, screw density is the number of pedicle screws divided by the number of vertebrae in the fusion span, including vertebrae without screws. When expressed as site occupancy, the denominator is the number of potential bilateral pedicle sites. Thus, 1.4 screws per vertebra corresponds to 70% bilateral site occupancy. This definition is not imposed retrospectively on studies using different denominators: vertebrae, motion segments, and anchors restricted to the main curve are not interchangeable. Conversion requires a clearly defined fusion span and anchoring pattern. Figure 3 illustrates how identical global density can conceal different regional distributions.
Figure 3. Global screw density and regional anchor distribution are distinct construct characteristics. Three illustrative configurations span ten generic vertebral levels with twenty potential bilateral pedicle sites. Configuration (A) contains twenty screws, corresponding to 2.0 screws per vertebra and 100% site occupancy. Configurations (B,C) each contain fourteen screws, corresponding to 1.4 screws per vertebra and 70% site occupancy, but differ in regional distribution. Within the illustrative V4–V6 region, (B) occupies six sites and (C) occupies four. All vertebrae within the fusion span are included in the denominator, including levels without screws. Filled circles indicate implanted screws, and open circles indicate unoccupied sites. These configurations are examples for explaining measurement, not recommended operative patterns.
The Minimize Implants Maximize Outcomes (MIMO) trial randomized 211 adolescents with primary thoracic curves of 45–65° to high-density constructs (n = 108; target ≥ 1.8 implants per fused level) or low-density constructs (n = 103; target ≤ 1.4). At two years, intention-to-treat coronal correction was 67.6% and 65.7%, respectively. The difference, low minus high density, was −1.9 percentage points (95% CI, −6.1 to 2.2), within the prespecified equivalence margin of ±10 percentage points. Five patients in the high-density group and six in the low-density group underwent reoperation. The trial establishes equivalence for its specified correction endpoint, although differing upper-age descriptions in recruitment documentation leave strict operative-age eligibility unresolved [25,51]. This finding does not establish equivalence for other radiographic outcomes, safety, or PROMs. For illustration, a 10-percentage-point correction margin corresponds to 6° for a 60° baseline curve; this is an arithmetic example, not a validated clinical-importance threshold. The small number of reoperations cannot establish equivalence of uncommon events. The trial also does not define density selection for highly rigid curves, more complex deformities, or other correction maneuvers [25].
Jackson et al. compared 1225 patients with high-density constructs and 640 with moderate-density constructs. Reported categories were >1.8 and approximately 1.4–1.8 screws per level. Both groups had a mean final main-curve angle of 19°, without a significant difference in SRS-22 scores. In the Lenke 2 subgroup, higher density was associated with a small coronal advantage without a corresponding PROM benefit. The complete operative-age range remained unresolved. These contextual registry findings do not replace a prespecified equivalence test [27]. Shen et al. compared 28 low-density and 34 high-density Lenke 1 cases using the cohort-based threshold of 1.61 screws per level. Two-week correction was 67.9% and 74.3%, respectively, and final correction was 65.0% and 69.1%. Neither correction-percentage comparison was statistically significant. The low-density group had a shorter operative time (278.4 vs. 331.0 min) and lower blood loss (823.6 vs. 1010.9 mL). SRS-22 domains did not differ at follow-up. Only mean operative ages were identified, and the allocation was retrospective. The early absolute difference in Cobb angles and the later correction percentages should be kept as separate outcomes rather than merged into an identity of correction [50]. The small, randomized trial by Mostafa et al. [52] and the separate meta-analyses by Aoun et al. and Zheng et al. [30,31] provide additional evidence on implant density. These publications differ in study design, eligibility criteria, and density definitions; their findings are therefore used as contextual evidence. Wang et al. used 30 patient-specific geometries and 600 simulations to study translation followed by apical vertebral derotation. They identified a correlation between apical rotation correction after the additional derotation maneuver and local apical density (r = 0.825). Greater overall density reduced predicted bone–screw forces. These computational results suggest a mechanism by which similar coronal correction can coexist with different rotational correction and anchor loading. Though they do not define the clinical threshold, they provide support for the practice of recording regional distribution in addition to the total screw count [34]. An anchor map should include concave and convex apical screws, bilateral terminal fixation, skipped vertebrae, and participation in the reduction maneuver, including deviation from the initial pattern. Regional purchase and distribution distinguish a planned low-density strategy from arbitrary screw omission and prevent total screw count from substituting for a complete construct description [18,34].

9. Three-Dimensional Correction and Interpretation of Radiographic Quality

Coronal Cobb correction is the most commonly reported postoperative endpoint, but it provides an incomplete assessment of deformity correction. A patient may have a high coronal correction percentage alongside residual apical rotation, inadequate thoracic kyphosis, shoulder imbalance, coronal decompensation, or an inappropriate distal fusion level [6,7]. Correction percentage is calculated as [(preoperative Cobb angle−postoperative Cobb angle)/preoperative Cobb angle] × 100. Comparisons may be misleading when baseline severity or flexibility differs: correction from 55° to 15° is 72.7%, whereas correction from 90° to 35° is 61.1%. These percentages alone do not capture the mechanical demands of flexible and rigid curves. Comparative analyses should account for side-bending or traction flexibility, curve pattern, rotation, and baseline sagittal profile. Primary curve location and severity may also interact with rod contour and implant rigidity.
Sagittal alignment should retain the original vertebral landmarks: T2–T5, T2–T12, T4–T12, and T5–T12 describe different segments. The desirable change depends on the baseline profile, and increased kyphosis or lordosis is not automatically an improvement [6,43]. Lamerain’s single-arm CoCr series included patients aged 18 and lacked a material-control group, so it cannot establish material superiority in the strict population [53].
Rod contour and spinal alignment are also distinct. In a contextual 27-patient series, Cidambi et al. measured 5.5 mm ultrahigh-strength steel rods before insertion and at a mean of five weeks after surgery. Concave rods flattened by approximately 21°, whereas convex contour increased by approximately 2°; spinal T5–T12 kyphosis was maintained [19]. Kokabu et al. examined contour change in relation to facetectomy and screw distribution, but the cohort included ages 12–20 and constructs with hooks, placing it outside the direct criteria [54]. CT rotation, projection-based rotation grades, rib prominence, and clinical trunk rotation cannot be pooled as one measure of three-dimensional correction [14,48,49]. Alignment on the initial standing radiograph must be distinguished from maintenance and adaptation during follow-up. An early endpoint cannot be relabeled as a long-term result solely because the cohort had a longer minimum follow-up, as illustrated by the inconsistent timing in the monoaxial/polyaxial comparison [14]. Coronal correction, sagittal alignment, balance, rotation, maintenance, and PROMs should be reported together. A composite success score would require validation of its components and weights from both clinical and patient perspectives.

10. Patient-Reported Outcomes and Clinical Relevance

Patient-reported outcomes were interpreted according to the questionnaire version, domain, baseline score, and assessment interval. SRS-22, SRS-22r, and SRS-30 are related but distinct instruments. Total score, self-image, pain, function, mental health, and satisfaction address different aspects of health. Improvement in one domain does not establish improvement across all domains. Table 5 presents extracted values and identifies missing baseline information and uncertain follow-up [7,14,27,42].
Table 5. Patient-reported outcomes by instrument, domain, and assessment interval.
Berlin et al. evaluated 51 patients aged 11–17 years who underwent posterior pedicle-screw dual-rod fusion, with a mean follow-up of 16.6 months. SRS-30 completion was 88.2%. The mean total score increased from 3.6 to 4.2, but function/activity did not improve significantly, and curve correction did not correlate with satisfaction. This eligible association study supports separate assessment of radiographic and patient-reported outcomes; it does not estimate an effect of rod alloy, screw size, or density. Nor does the absence of a correlation establish that residual deformity is clinically unimportant [7]. The eligible screw-head comparison reported SRS-22 totals of 4.2 and 4.3 without a significant difference, but the assessment interval was inconsistent; the eligible alloy comparison collected no validated PROM [6,14]. These findings support only the conclusion that patient-reported superiority was not demonstrated in the two component-focused comparisons. Contextual studies add domain-specific observations, including better self-image with the uniplanar technique in Hu et al., while the broader-age alloy trial and density registry reported no significant questionnaire differences [27,42,49].
In the eligible combined-strategy cohort, SRS-22r totals reportedly increased from 3.86 to 4.36 with smaller screws/titanium rods and from 3.82 to 4.26 with larger screws/CoCr rods, without a significant final between-group total difference [21]. Response denominators and internal source inconsistencies limit interpretation. Group means cannot reconstruct MCID responder proportions. Comparative PROM analyses require baseline adjustment, specified domains and time points, and transparent response denominators; adjustment for postoperative alignment also requires care because alignment may mediate part of an implant effect.

Meaningful Change, Measurement Error, and Between-Group Interpretation

Carreon et al. calculated MCIDs for the SRS-22 for AIS surgery as 0.20 for pain, 0.98 for appearance, and 0.08 for activity; the MCID for activity was lower than the measurement error. Analyzing 1281 surgical AIS patients, Kelly et al. reported minimum detectable measurement differences (MDMDs) for the SRS-22r as approximately 0.30 for pain, 0.24 for activity, and 0.30 for self-image. These measurement studies contribute to the interpretation but are not comparative implant-effect cohorts eligible under current criteria [55,56]. A within-person change threshold should not automatically be used as the required difference between group means. Patient-perceived importance and change exceeding measurement error are related but distinct concepts. A proper implant comparison should include a prespecified questionnaire and domain, baseline adjustment, and between-group differences with confidence intervals, and, when a defensible threshold exists, the proportion of patients experiencing meaningful change. Failure to reach statistical significance between the groups does not prove the equivalence of patient benefit, and a significant between-group difference does not prove clinically meaningful benefits [55,56]. The available baseline-to-follow-up domain data and responder counts were insufficient to reconstruct an MCID responder analysis for the above head-design and rod comparison studies. No uniform threshold was used across all questionnaire versions [7,14,42,49,55,56]. Serial PROM reporting should include response denominators, characteristics of responders and nonresponders (if available), and sensitivity analysis, since missing data can be associated with pain, satisfaction, or subsequent surgery.

11. Complications, Implant Failure, and Revision Surgery

Complication reporting should separate perioperative injury, delayed implant events, and alignment failure [11,23]. Potential tradeoffs include screw purchase versus pedicle-wall violation, rod contour retention versus anchor demand, and fewer insertion sites versus greater load per remaining anchor [1,17,18,20]. Radiographic breach must be distinguished from neurologic, vascular, or visceral injury. Observed accuracy depends on placement technique, imaging modality, breach threshold, and routine versus symptom-driven CT surveillance [1,2,16]. Report per-screw accuracy separately from per-patient risk and account for multiple implants or events within one patient. Potentially overlapping event categories cannot be added to estimate the number of patients with any complication; no such totals were reconstructed here.
Klyce et al. evaluated 1057 patients and 19,569 screws, reporting plowing in 4.5% of patients and approximately 0.4% of screws. Operative-age eligibility remains unresolved. Plowing was reported in 48 patients and 72 screws. It was associated with lower implant density; 13 affected patients had loss of correction, and revision was more frequent among patients with plowing. These observational associations do not establish causation. Plowing remains distinct from medial cortical breach and should not be assumed to be clinically inconsequential [17]. Implant fracture should be characterized by timing, location, symptoms, alignment, fusion assessment, and treatment. Early loss of fixation, a later asymptomatic radiographic fracture, and painful failure associated with nonunion are clinically different events. Revision is a separate endpoint because it additionally reflects symptoms, structural consequences, and treatment decisions. The outcome table therefore separates the reported device event from the operation it prompted and links each revision to its documented indication. A recent multicenter description by La Poche et al. identified 22 AIS patients revised for confirmed pseudarthrosis among 3532 registry patients. Median time to revision was 2.5 years; 17 failures involved the distal construct. Reported patterns included screw loosening, rod fracture, and screw breakage, and all revisions were posterior-only. Because exact operative ages and original anchor composition were not verified, the study is used as contextual evidence about failure presentation, not as a target-population incidence estimate or alloy comparison. Its contribution is to link mechanical findings to a defined clinical diagnosis and the actual revision performed [57]. Bowden’s lower pooled reoperation proportion with 5.5 mm than 6.0 mm rods may reflect case mix, follow-up, or operative strategy rather than diameter [28]. MIMO remains contextual because strict operative ages are unresolved; its equivalence finding concerns two-year coronal correction in 45–65° primary thoracic curves and cannot establish safety or PROM equivalence [25,51].
Perioperative events also require consideration of co-interventions. Thoracoplasty was frequent in the monoaxial/polyaxial study, so chest-tube insertion and hemothorax cannot be attributed automatically to screw-head design. Small infection counts and differences in treatment era also prevent a component-specific causal interpretation. Likewise, shorter operative time or lower blood loss in density studies does not establish a causal effect of density on infection. Blood-loss comparisons require information about tranexamic acid and other hemostatic measures [58]. Table 6 retains observed events without treating unreported endpoints as zero [14,50]. Broader spinal-surgery evidence associates malnutrition with infection risk, so nutritional status should be treated as a potential confounder rather than an implant effect [59].
Table 6. Implant-related events, other complications, and subsequent operations.
Pseudarthrosis is a biological and mechanical outcome [24]. Solid fusion reduces dependence on implants for long-term stability. Implant failure may accompany nonunion, while excessive motion or inadequate load sharing may contribute to nonunion. Distinguishing these pathways requires explicit fusion criteria, serial imaging, and revision findings [24,60]. Rod fracture alone does not prove pseudarthrosis, and stable plain radiographs do not definitively establish fusion. Junctional outcomes should specify the measured angle, diagnostic threshold, symptoms, structural failure, and revision. Rod properties, alignment, baseline anatomy, and fusion endpoints must be considered together. Available comparisons do not separate their independent effects sufficiently to rank alloys by junctional safety [26]. A difference in mean proximal junctional angle is not equivalent to a difference in structural failure incidence.
Revision-free survival requires a common time origin, accurate revision dates, adequate surveillance, and appropriate censoring. The 2026 meta-analysis found no statistically significant difference in complication rates, but sparse events and variable age criteria limit interpretation [29]. Comparisons across treatment eras must account for follow-up duration because longer observation creates more opportunity to detect late events. Time to first reoperation, recurrent procedures, and loss to follow-up should be reported separately. Patients cannot be assumed revision-free beyond their last observed contact. Comparisons of uncommon harms should include affected-patient counts, denominators, observation periods, indications, and confidence intervals.

Pregnancy as a Special Long-Term Outcome

Pregnancy is relevant to long-term counseling after AIS fusion, but the available studies do not compare rod alloy, rod diameter, screw design, or regional anchor density as determinants of pregnancy-related implant failure. In the pedicle-screw cohort of Grabala et al., 62 women with subsequent pregnancy and 66 without pregnancy had reported mean correction losses of 3.5° ± 3° and 4.5° ± 3°, respectively (p = 0.379), with a final mean major-curve angle of 20° in both groups. The full operative-age range was not verified. These findings provide context about average alignment maintenance, without establishing equivalence or excluding uncommon individual failures [61].
Pain, mode of delivery, and obstetric complications are distinct from screw loosening, rod fracture, pseudarthrosis, or revision. The available reports do not establish pregnancy-specific superiority of any implant configuration or justify construct selection on that basis [62,63,64,65,66,67]. Pregnancy is treated here as a special long-term outcome rather than direct evidence for implant selection. The source-specific appraisal is provided in Supplementary Section S3 and Table S3.

12. Interaction Between Implant Choice and Correction Technique

Surgeon and technique are integral to the treatment comparison. Implant changes may alter the correction maneuver, rod contouring, anchor selection, and use of releases. In a regression analysis of hypokyphotic AIS, Monazzam et al. found that surgeon identity predicted kyphosis correction. Strict operative-age eligibility remained unresolved. This association supports accounting for surgeon and operative practice; it does not establish that surgeon identity alone determines correction or that rod properties have no effect [68].
Osteotomy use must also be specified. Posterior column osteotomy increases spinal mobility and may reduce the force required for correction. If larger rods are selected for more rigid curves that also undergo more extensive release, implant choice and co-intervention become confounded. Adjustment for preoperative Cobb angle alone does not resolve this problem without information on bending flexibility, rotation, sagittal profile, and release technique [44,54]. In the 58-patient comparison performed by Jabbouri et al., thoracic kyphosis was within its 20–40° range in 95.5% of patients who received pre-contoured rods compared to 61.1% of surgeon-contoured rods at mean follow-ups of 9.7 and 11.3 months, respectively (22 and 36 patients). These short and unequal intervals do not establish durable correction. The pre-contoured group predominantly received CoCr rods, whereas the surgeon-contoured group predominantly received titanium rods. Both groups used 5.5 or 6.0 mm rods, but the distribution of rod diameters by group was not reported. A CoCr-only subgroup analysis included 20 patients with pre-contoured rods and six with surgeon-contoured rods; the TK target of 20–40° was achieved at final follow-up in 100% and 66.7%, respectively. This small nonrandomized subgroup comparison reduces alloy confounding but does not isolate the effect of contouring accuracy. Strict operative-age eligibility remained unresolved. The overall comparison therefore evaluates a planning and implant package [44]. Another templating series extended to age 21 is outside the strict target population [44,69]. A device comparison can help to understand the meaning of the term “patient-specific”. This term can imply either the target chosen from the planning model, the intraoperative template, the digitally guided bending, or the factory-contoured rod, each representing a different intervention. The plan can also affect fusion levels, releases, and correction sequence before the insertion of the rod. To isolate manufacturing or contouring accuracy, groups should have comparable alignment targets and operative protocols; all effects of a planning package cannot be attributed to the rod [44,69]. More extensive releases or osteotomies can increase operative burden [11,58]. TXA exposure should be recorded in blood-loss comparisons because antifibrinolytic use can affect bleeding independently of implant selection [70,71]. Fusion-level selection should be specified as an operative exposure rather than inferred from implant type. Comparisons should standardize or stratify uppermost and lowermost instrumented vertebrae (UIV and LIV), describe the remaining mobile lumbar segments, and define adding-on and balance outcomes. Berlin et al. examined LIV-related measures and PROM associations but did not establish that a particular rod prevents adding-on or permits a shorter fusion. Fusion extent should therefore be a prespecified variable in implant-outcome analyses [7].

13. Discussion: Comparative Effectiveness and the Limits of Component-Level Claims

The principal clinical evidence comprises three nonrandomized implant comparisons with confirmed population and construct eligibility: Yang et al.’s equal-diameter alloy comparison, Yang et al.’s screw-head comparison, and Grabala et al.’s combined-strategy comparison [6,14,21]. None establishes a universally preferable implant. The alloy study found identical observed thoracic correction without a validated PROM; the screw-head study found a coronal difference without a demonstrated SRS-22 advantage and with uncertain outcome timing. The combined-strategy study reported greater correction with larger screws and thicker CoCr rods, but did not isolate the effect of an individual component or demonstrate a final SRS-22r total-score advantage. Table 7 summarizes these limitations. Berlin et al.’s eligible association study concerns the relationship between radiographs and patient experience and cannot estimate an implant effect [7]. The count refers to the appraised cited corpus, not an exhaustive set of all potentially eligible studies.
Table 7. Core clinical evidence and the limits of implant-specific inference.
Eligibility, internal validity, and clinical importance should be considered separately. The strict age rule narrows the pediatric question but excludes or leaves unresolved studies that may be informative for older adolescents. In a broader AIS synthesis, Sakai et al.’s randomized trial, which permitted recruitment at ages 10–19, would add evidence against an assumed universal one-year alloy advantage, and MIMO would provide endpoint-specific equivalence of two-year coronal correction for primary thoracic curves of 45–65°. Neither supplies a separately verified under-18 estimate here. MIMO does not establish equivalence of PROMs, uncommon complications, or density choices in other curve populations [25,42,51]. This qualitative comparison indicates how the evidence base broadens; it is not a pooled sensitivity analysis and cannot establish a different strict-age treatment effect.
Combined-system comparisons are legitimate clinical questions, but simultaneous changes in screws, rods, planning, or maneuvers prevent attribution to a single component [21,41,44]. Small event counts, unequal surveillance, and possible cohort overlap further constrain comparisons. Absence of statistical significance does not demonstrate equivalence, significance in one subgroup alone does not establish interaction, and within-person MCID thresholds do not automatically define between-group superiority [55,56]. Possible cohort overlap is another concern in registry studies and computational analyses. Publications from the same multicenter network may share participants, although common investigators alone do not prove overlap. Wang et al. used geometries from the MIMO research program for computational modeling, so those simulations do not constitute an independent clinical cohort. Such work can explain mechanisms but cannot be added to trial enrollment to enlarge the clinical evidence base. Study identifiers, enrollment periods, centers, and eligibility criteria should be compared before any aggregation [25,26,27,34].

14. Clinical Implications and Limits of Inference

The three eligible comparisons support a limited conclusion: neither the tested alloy choice, screw-head design, nor combined implant strategy has demonstrated overall clinical superiority in the defined pediatric population [6,14,21]. The combined-strategy results cannot isolate the contributions of screw dimensions, rod diameter, and alloy. These comparisons do not establish preferred rod or screw diameters, a density target, or equivalence of rare complications. These are boundaries of the available comparative evidence, rather than prescriptive implant recommendations.
Anatomy, bone purchase, curve flexibility, regional alignment goals, and the correction maneuver remain considerations supported mainly by biomechanics and broader clinical context [18,37,68]. Greater rod rigidity may improve contour retention while increasing the force needed for seating. This mechanical trade-off explains a planning consideration; it does not validate a rule for choosing an alloy or diameter. The review presents no selection algorithm or universal threshold.
MIMO remains contextual because strict operative-age eligibility is unresolved. Its clinical implication is confined to the tested two-year coronal correction equivalence endpoint in primary thoracic curves of 45–65°; it does not establish equivalent safety, PROMs, or density choices for other curves [25,51]. Proposed regional anchor maps and component-specific comparisons are research priorities, not validated treatment rules.

15. Priorities for Future Research

Future prospective comparisons should define the implant change, verify operative ages, and document baseline flexibility, curve pattern, sagittal profile, and co-interventions. Diameter comparisons should keep alloy and operative strategy comparable; alloy comparisons should use a common diameter and record intended and retained rod geometry. Radiographic assessment, PROM domains, response rates, and clinically important adverse events should be prespecified. Necessary intraoperative modifications should be documented rather than concealed.
A factorial design could separate alloy and diameter effects and test their interaction if equipoise and sufficient enrollment are achievable. Multicenter follow-up and registry linkage would improve assessment of uncommon or late revisions. Detailed reporting should include level-specific screw dimensions and trajectories, regional anchor distribution, fusion endpoints, releases, nutritional status, and infection-related covariates. Serial rod-contour and alignment measurements could test whether mechanical differences translate into sustained patient benefit [19,44]. Table 8 presents a proposed minimum reporting set rather than an established reporting standard. Biomimetic architecture and local biological support provide a separate translational research question [39,40]. The cited regeneration and bone-filling studies do not establish improved screw purchase, fusion, or AIS outcomes. Any subsequent fixation study would need to distinguish biological effects from implant stiffness and correction technique.
Table 8. Minimum reporting set for a clinically interpretable implant comparison.
The minimum dataset in Table 8 connects implant and operative details with serial imaging, PROMs, and adjudicated events. Long-term comparisons should also evaluate cost-effectiveness and patient preferences, because a modest radiographic gain may not justify additional implants or operative burden. Prospective registry linkage should capture late and repeated reoperations, including those performed outside the original institution, with adjudicated indications. Patient-level analyses should use the patient as the unit of analysis; implant-level analyses should account for clustering of screws within patients and patients within centers.

16. Limitations of This Review

The reported original searches ended on 15 September 2026, but the complete search record and screening counts are unavailable. The stated publication cutoff is 17 October 2026; no supplementary search covering the intervening period is documented, so complete coverage through that date cannot be confirmed. Requiring documented surgery before age 18 and exclusively pedicle-screw fixation deliberately restricts the review’s applicability and makes eligibility dependent on the detail reported in each source. Unresolved eligibility therefore reflects insufficient or conflicting information, rather than evidence that a cohort necessarily included ineligible patients. The broader AIS literature includes randomized comparisons whose eligibility for the direct synthesis could not be confirmed; however, the available data did not permit estimation of effects specifically in patients operated on before age 18 or a quantitative sensitivity analysis. Accordingly, the conclusions concern the evidence appraised for this strictly defined population and should not be interpreted as a comprehensive assessment of implant strategies across all AIS surgery.
Methodological appraisal was qualitative, without formal risk-of-bias grading. SANRA informed reporting but was not used to assess the certainty of treatment effects [33]. The three eligible comparisons have study-specific limitations, including nonrandom allocation, small samples in the two component-focused studies, inconsistencies in source tables, uncertain outcome timing, and incomplete reporting of patient-reported outcomes or safety. Interpretation of contextual studies is further constrained by heterogeneous outcomes, possible cohort overlap, and differences in follow-up duration and surveillance. Laboratory and computational studies provide mechanistic explanations but do not establish clinical superiority. Table S4 distinguishes full-text assessment, abstract-level checks, and verification against publisher records. Confirmation of bibliographic identity does not establish the accuracy of every reported statistic or guarantee that all retractions have been identified. The review tables summarize published observations and explicitly identified calculations; they do not constitute a pooled patient-level dataset or a validated model for implant selection.

17. Conclusions

In the appraised cited corpus, three nonrandomized implant comparisons had confirmed eligibility for posterior all-pedicle-screw fusion before age 18. The equal-diameter alloy study reported identical mean thoracic correction but did not collect a validated PROM. The screw-head study reported greater coronal correction with monoaxial screws without a demonstrated SRS-22 advantage, and its assessment timing was inconsistent. The combined-strategy study, included after corresponding-author confirmation of cohort-wide operative ages, reported greater correction with larger screws and thicker CoCr rods but could not isolate individual component effects or establish a final SRS-22r total-score advantage. None establishes overall implant superiority. The separately eligible radiographic–PROM study is associative and does not compare devices.
Broader AIS comparisons remain contextual when age or construct eligibility is unresolved or outside the criteria. MIMO supports only its prespecified two-year coronal correction equivalence endpoint for primary thoracic curves of 45–65°, with strict operative-age eligibility unresolved; safety, PROM equivalence, and universal density selection do not follow. The strict age rule limits direct applicability rather than identifying a biological threshold. Biomechanical reasoning informs research and planning considerations, but the reviewed evidence does not validate an implant-selection algorithm.

Supplementary Materials

The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/bioengineering13121431/s1. The accompanying supplementary document contains Sections S1–S4 and Table S1–S4. Table S1: Study eligibility, source locations and consequences for synthesis. Table S2: Historical search account and dated revision verification log. Table S3: Pregnancy after AIS fusion and the limits of implant-specific inference. Table S4: Complete cited-source inventory, access level and claim-specific appraisal. Section S3 provides the detailed pregnancy appraisal; Section S4 documents source discrepancies and verification boundaries. Reference numbers are synchronized with the main manuscript.

Author Contributions

Conceptualization, M.G., J.S. and P.G.; methodology, M.G. and P.K.; software, P.G., J.S. and P.K.; formal analysis, P.G., P.K., M.G. and J.S.; investigation, P.G., M.G. and P.K.; resources, P.G., P.K., J.S. and M.G.; data curation, M.G., P.K. and P.G.; writing—original draft preparation, P.G., J.S. and M.G.; writing—review and editing, P.G.; visualization, P.K. and P.G.; supervision, P.G.; project administration, M.G. and P.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable. This narrative review examines the published literature and contains no newly collected participant data or identifiable patient images.

Data Availability Statement

No new participant-level dataset was generated. Evidence is available in the cited publications and the accompanying supplementary appraisal. The original search logs are not recoverable. The operative-age clarification for [21] is summarized in Supplementary Table S1 and Section S4; independent patient-level records were not examined for this review.

Acknowledgments

ChatGPT (OpenAI; https://chatgpt.com/; last used 12 November 2026) assisted with language revision, manuscript organization, reference numbering, and programmatic figure preparation. Supplementary Table S2 distinguishes the research services, historical checks, and verification limits. AI-generated summaries were not treated as primary evidence. The authors retain responsibility for evaluating the cited evidence and approving the final manuscript.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Grabala, P.; Helenius, I.J.; Kowalski, P.; Grabala, M.; Zacha, S.; Deszczynski, J.M.; Albrewczynski, T.; Galgano, M.A.; Buchowski, J.M.; Chamberlin, K.; et al. The Child’s Age and the Size of the Curvature Do Not Affect the Accuracy of Screw Placement with the Free-Hand Technique in Spinal Deformities in Children and Adolescents. J. Clin. Med. 2023, 12, 3954. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Suk, S.I.; Kim, W.J.; Lee, S.M.; Kim, J.H.; Chung, E.R. Thoracic pedicle screw fixation in spinal deformities: Are they really safe? Spine 2001, 26, 2049–2057. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Kim, Y.J.; Lenke, L.G.; Bridwell, K.H.; Cho, Y.S.; Riew, K.D. Free hand pedicle screw placement in the thoracic spine: Is it safe? Spine 2004, 29, 333–342. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Kuklo, T.R.; Lenke, L.G.; O’Brien, M.F.; Lehman, R.A., Jr.; Polly, D.W., Jr.; Schroeder, T.M. Accuracy and efficacy of thoracic pedicle screws in curves more than 90 degrees. Spine 2005, 30, 222–226. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Suk, S.I. Pedicle screw instrumentation for adolescent idiopathic scoliosis: The insertion technique, the fusion levels and direct vertebral rotation. Clin. Orthop. Surg. 2011, 3, 89–100. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Yang, J.H.; Suh, S.W.; Chang, D.G. Comparison of surgical correction rates between titanium and cobalt-chrome-alloy as rod materials in adolescent idiopathic scoliosis. Sci. Rep. 2020, 10, 10053. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Berlin, C.; Tielemann, S.; Quante, M.; Halm, H. Correlation of radiographic parameters and patient satisfaction in adolescent idiopathic scoliosis treated with posterior screw-dual-rod instrumentation. Eur. Spine J. 2023, 32, 3140–3148. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Huang, Y.; Zhu, C.; Wang, Y.; Feng, G.; Liu, L. A Novel Deformity Correction Manipulation System for Better Correction of Large Thoracic Scoliosis. Orthop. Surg. 2024, 16, 2436–2446. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Daher, M.; Aoun, M.; Kreichati, G.; Kharrat, K.; Sebaaly, A. Hybrid vs all pedicle screws constructs in adolescent idiopathic scoliosis: A metaanalysis of clinical and radiological outcomes. Spine Deform. 2024, 12, 1229–1239. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Pankowski, R.; Roclawski, M.; Ceynowa, M.; Mikulicz, M.; Mazurek, T.; Kloc, W. Direct Vertebral Rotation Versus Single Concave Rod Rotation: Low-dose Intraoperative Computed Tomography Evaluation of Spine Derotation in Adolescent Idiopathic Scoliosis Surgery. Spine 2016, 41, 864–871. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Grabala, P.; Gregorczyk, J.; Fani, N.; Galgano, M.A.; Grabala, M. Surgical Treatment Strategies for Severe and Neglected Spinal Deformities in Children and Adolescents without the Use of Radical Three-Column Osteotomies. J. Clin. Med. 2024, 13, 4824. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Chan, C.Y.W.; Kwan, M.K. Safety of Pedicle Screws in Adolescent Idiopathic Scoliosis Surgery. Asian Spine J. 2017, 11, 998–1007. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Noe, M.C.; Link, R.C.; Warren, J.R.; Etebari, C.V.; Whitmire, M.H.; Anderson, J.T.; Schwend, R.M. Three-dimensional deformity correction in adolescent idiopathic scoliosis patients: What are the benefits of hybrid apical sublaminar bands versus all-pedicle screws? J. Pediatr. Orthop. B 2025, 34, 367–374. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Yang, J.H.; Kim, H.J.; Chang, T.Y.; Suh, S.W.; Chang, D.G. Comparative Analysis of Monoaxial and Polyaxial Pedicle Screws in the Surgical Correction of Adolescent Idiopathic Scoliosis. J. Clin. Med. 2024, 13, 2689. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. De Salvatore, S.; Asunis, E.; Oggiano, L.; Brigato, P.; Palombi, D.; Ruzzini, L.; Costici, P.F. Correction of adolescent idiopathic scoliosis using the convex rod de-rotation maneuver: A systematic review. Eur. J. Orthop. Surg. Traumatol. 2025, 35, 319. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Zaghloul, K.M.; Tawfik, A.M.; Gadelhak, B.; Sobh, H.M.; Batouty, N.M.; Sobh, D.M. Understanding the Aorta-Spine Relation in Idiopathic Scoliosis: Value of Noncontrast CT-Derived Curved Coronal Reformats and 3D Volume Images. Int. J. Spine Surg. 2021, 15, 818–825. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Klyce, W.; Suresh, K.V.; Jain, A.; Parent, S.; Shah, S.A.; Cahill, P.J.; George, S.; Clements, D.; Upasani, S.; Yaszay, B.; et al. Pedicle Screw Plowing in Adolescent Idiopathic Scoliosis: How Common Is It and Is It a Problem? Spine 2022, 47, 873–878. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Clin, J.; Le Navéaux, F.; Driscoll, M.; Mac-Thiong, J.M.; Labelle, H.; Parent, S.; Shah, S.A.; Lonner, B.S.; Newton, P.O.; Serhan, H. Biomechanical Comparison of the Load-Sharing Capacity of High and Low Implant Density Constructs With Three Types of Pedicle Screws for the Instrumentation of Adolescent Idiopathic Scoliosis. Spine Deform. 2019, 7, 2–10. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Cidambi, K.R.; Glaser, D.A.; Bastrom, T.P.; Nunn, T.N.; Ono, T.; Newton, P.O. Postoperative changes in spinal rod contour in adolescent idiopathic scoliosis: An in vivo deformation study. Spine 2012, 37, 1566–1572. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Grabala, P.; Helenius, I.J.; Grabala, M.; Shah, S.A. Influences of Increasing Pedicle Screw Diameter on Widening Vertebral Pedicle Size during Surgery in Spinal Deformities in Children and Adolescents without Higher Risk of Pedicle and Vertebral Breaches. J. Clin. Med. 2023, 12, 5368. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. Grabala, P.; Kowalski, P.; Grabala, M. The Influence of Increased Pedicle Screw Diameter and Thicker Rods on Surgical Results in Adolescents Undergoing Posterior Spinal Fusion for Idiopathic Scoliosis. J. Clin. Med. 2024, 13, 2174. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Prince, D.E.; Matsumoto, H.; Chan, C.M.; Gomez, J.A.; Hyman, J.E.; Roye, D.P., Jr.; Vitale, M.G. The effect of rod diameter on correction of adolescent idiopathic scoliosis at two years follow-up. J. Pediatr. Orthop. 2014, 34, 22–28. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Suszczyńska, J.; Grabala, M.; Grabala, P. Modulation of Surgical Site Infection Risk in Spinal and Thoracic Surgeries Through Operative Parameters: A Narrative Review. J. Clin. Med. 2025, 14, 8124. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Jamnik, A.A.; Datcu, A.M.; Lachmann, E.; Patibandla, S.D.; Thornberg, D.; Jo, C.H.; Morris, W.Z.; Ramo, B.; Johnson, M. Repeat surgical interventions following “definitive” instrumentation and fusion for idiopathic scoliosis: A 30-year update. Spine Deform. 2024, 12, 99–107. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Larson, A.N.; Polly, D.W.; Sponseller, P.D.; Kelly, M.P.; Richards, B.S.; Garg, S.; Parent, S.; Shah, S.A.; Weinstein, S.L.; Crawford, C.H., 3rd; et al. The Minimize Implants Maximize Outcomes Study Group. The Effect of Implant Density on Adolescent Idiopathic Scoliosis Fusion: Results of the Minimize Implants Maximize Outcomes Randomized Clinical Trial. J. Bone Jt. Surg. Am. 2024, 106, 180–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Todderud, J.E.; Nugraha, H.K.; Kelly, M.P.; Pahys, J.; Shah, S.; Fletcher, N.; Newton, P.O.; Sucato, D.; Yaszay, B.; Larson, A.N.; et al. Rod options to outcomes: Comparing sagittal correction in pediatric posterior spinal fusion by rod size and material. Spine Deform. 2025, 13, 1949–1958. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Jackson, T.J.; Shah, S.A.; Shufflebarger, H.L.; Sucato, D.J.; Garg, S.; Sponseller, P.; Clements, D., 3rd; Fletcher, N.D.; Newton, P.O.; Samdani, A.F.; et al. High-density and moderate-density implant constructs for adolescent idiopathic scoliosis have equivalent clinical and radiographic outcomes at 2 years. Spine Deform. 2025, 13, 773–787. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Bowden, D.; Michielli, A.; Merrill, M.; Will, S. Systematic review and meta-analysis for the impact of rod materials and sizes in the surgical treatment of adolescent idiopathic scoliosis. Spine Deform. 2022, 10, 1245–1263. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Kumar, V.; Ifthekar, S.; Raj, V.; Vatkar, A.; Barik, S. Is There a Difference in the Outcomes of Adolescent Idiopathic Scoliosis Deformity Correction with the Type of Rod (Titanium vs. Cobalt-chrome) Used? A Systematic Review and Meta-analysis. J. Orthop. Case Rep. 2026, 16, 591–598. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Aoun, M.; Daher, M.; Bizdikian, A.J.; Kreichati, G.; Kharrat, K.; Sebaaly, A. Implant density in adolescent idiopathic scoliosis: A meta-analysis of clinical and radiological outcomes. Spine Deform. 2024, 12, 909–921. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Zheng, B.; Zhou, Q.; Liu, X.; Ma, K.; Qiang, Z. Low-density pedicle screw in adolescent idiopathic scoliosis: A systematic review and meta-analysis of 1,762 patients. Front Surg. 2025, 12, 1607323. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  32. Skalak, T.J.; Gagnier, J.; Caird, M.S.; Farley, F.A.; Li, Y. Higher pedicle screw density does not improve curve correction in Lenke 2 adolescent idiopathic scoliosis. J. Orthop. Surg. Res. 2021, 16, 276. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  33. Baethge, C.; Goldbeck-Wood, S.; Mertens, S. SANRA-a scale for the quality assessment of narrative review articles. Res. Integr. Peer Rev. 2019, 4, 5. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  34. Wang, X.; Larson, A.N.; Polly, D.W., Jr.; Aubin, C.E. Biomechanical Computational Study of Pedicle Screw Position and Density in Adolescent Idiopathic Scoliosis Instrumentation. Spine 2023, 48, 1436–1445. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  35. Lindsey, C.; Deviren, V.; Xu, Z.; Yeh, R.F.; Puttlitz, C.M. The effects of rod contouring on spinal construct fatigue strength. Spine 2006, 31, 1680–1687. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. Slivka, M.A.; Fan, Y.K.; Eck, J.C. The Effect of Contouring on Fatigue Strength of Spinal Rods: Is it Okay to Re-bend and Which Materials Are Best? Spine Deform. 2013, 1, 395–400. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Sarwahi, V.; Hasan, S.; Koutsogiannis, P.; Visahan, K.; Rao, H.; Patil, A.; Lo, Y.; Amaral, T.; DiMauro, J.P. Effect of Pedicle Screw Size on Surgical Outcomes Following Surgery for 412 Adolescent Idiopathic Scoliosis Patients. Spine 2023, 48, 1544–1551. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Upasani, V.V.; Kluck, D.G.; Farnsworth, C.L.; Shah, N.U.; Yaszay, B.; Caffrey, J.P.; Newton, P.O. Variation in Mechanical Properties of 5.5 mm Cobalt Chromium and Titanium Spinal Rods Across Suppliers. J. Orthop. Res. 2025, 43, 1493–1500. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Li, X.L.; Zhao, Y.Q.; Miao, L.; An, Y.X.; Wu, F.; Han, J.Y.; Han, J.Y.; Tay, F.R.; Mu, Z.; Jiao, Y.; et al. Strategies for promoting neurovascularization in bone regeneration. Mil. Med. Res. 2025, 12, 9. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Meng, F.; Yu, Y.; Yu, J.; Chen, H.; Ma, Y.; Chen, Y.; Wang, Y.; Jiang, J.; He, X.; Zhan, Z.; et al. A trabeculae-like biomimetic bone-filling material as a potential organoid for bone defect treatment. Organoid Res. 2025, 1, 025040003. [Google Scholar] [CrossRef] [Scilit]
  41. Huang, T.H.; Ma, H.L.; Wang, S.T.; Chou, P.H.; Ying, S.H.; Liu, C.L.; Yu, W.K.; Chang, M.C. Does the size of the rod affect the surgical results in adolescent idiopathic scoliosis? 5.5-mm versus 6.35-mm rod. Spine J. 2014, 14, 1545–1550. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Sakai, D.; Tanaka, M.; Takahashi, J.; Taniguchi, Y.; Schol, J.; Hiyama, A.; Misawa, H.; Kuraishi, S.; Oba, H.; Matsubayashi, Y.; et al. Cobalt-chromium versus titanium alloy rods for correction of adolescent idiopathic scoliosis based on 1-year follow-up: A multicenter randomized controlled clinical trial. J. Neurosurg. Spine 2021, 34, 897–906. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  43. Ruffilli, A.; Fiore, M.; Viroli, G.; Barile, F.; Manzetti, M.; Martikos, K.; Greggi, T.; Faldini, C. 5.5-mm Cobalt-Chrome vs 6-mm Titanium Alloy Rods in Surgical Treatment of Lenke 1 Adolescent Idiopathic Scoliosis With High-Density Pedicle Screws and Direct Vertebral Rotation on Differently Shaped Rods: A Retrospective Comparative Cohort Study. Int. J. Spine Surg. 2023, 17, 25–31. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Jabbouri, S.S.; Joo, P.; David, W.B.; Jeong, S.; Moran, J.; Jonnalagadda, A.; Tuason, D. Pre-contoured patient-specific rods result in superior immediate sagittal plane alignment than surgeon contoured rods in adolescent idiopathic scoliosis. J. Spine Surg. 2024, 10, 177–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. Lamerain, M.; Bachy, M.; Delpont, M.; Kabbaj, R.; Mary, P.; Vialle, R. CoCr rods provide better frontal correction of adolescent idiopathic scoliosis treated by all-pedicle screw fixation. Eur. Spine J. 2014, 23, 1190–1196. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  46. Sabah, Y.; Clément, J.L.; Solla, F.; Rosello, O.; Rampal, V. Cobalt-chrome and titanium alloy rods provide similar coronal and sagittal correction in adolescent idiopathic scoliosis. Orthop. Traumatol. Surg. Res. 2018, 104, 1073–1077. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. Etemadifar, M.R.; Andalib, A.; Rahimian, A.; Nodushan, S.M.H.T. Cobalt chromium-Titanium rods versus Titanium-Titanium rods for treatment of adolescent idiopathic scoliosis; which type of rod has better postoperative outcomes? Rev. Assoc. Med. Bras. 2018, 64, 1085–1090. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Dalal, A.; Upasani, V.V.; Bastrom, T.P.; Yaszay, B.; Shah, S.A.; Shufflebarger, H.L.; Newton, P.O. Apical vertebral rotation in adolescent idiopathic scoliosis: Comparison of uniplanar and polyaxial pedicle screws. J. Spinal Disord. Tech 2011, 24, 251–257. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Hu, Z.; Zhao, Z.; Li, J.; Tseng, C.C.; Qiu, Y.; Cheng, J.C.; Zhu, Z.; Liu, Z. Comparison of Clinical and Radiologic Outcome of Three-Dimensional Correction in Lenke 5C Curve: Uniplanar Versus Polyaxial Pedicle Screws. World Neurosurg. 2018, 114, e729–e734. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  50. Shen, M.; Jiang, H.; Luo, M.; Wang, W.; Li, N.; Wang, L.; Xia, L. Comparison of low density and high density pedicle screw instrumentation in Lenke 1 adolescent idiopathic scoliosis. BMC Musculoskelet. Disord. 2017, 18, 336. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  51. Mayo Clinic. MIMO Clinical Trial. Clinical Trial NCT01792609; Protocol MIMO1208M18202. Available online: https://www.mayo.edu/research/clinical-trials/cls-20149145 (accessed on 12 November 2026).
  52. Mostafa, H.E.; Moustafa, A.S.S.; El Zayat, H.H.; Koheil, A.; Ragab, O.A.A. Comparative study of low-density versus high-density pedicular screw implantations in management of adolescent idiopathic scoliosis: A randomized clinical trial. Egypt J. Neurosurg. 2024, 39, 52. [Google Scholar] [CrossRef] [Scilit]
  53. Lamerain, M.; Bachy, M.; Dubory, A.; Kabbaj, R.; Scemama, C.; Vialle, R. All-Pedicle Screw Fixation With 6-mm-Diameter Cobalt-Chromium Rods Provides Optimized Sagittal Correction of Adolescent Idiopathic Scoliosis. Clin. Spine Surg. 2017, 30, E857–E863. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  54. Kokabu, T.; Sudo, H.; Abe, Y.; Ito, M.; Ito, Y.M.; Iwasaki, N. Effects of Multilevel Facetectomy and Screw Density on Postoperative Changes in Spinal Rod Contour in Thoracic Adolescent Idiopathic Scoliosis Surgery. PLoS ONE 2016, 11, e0161906. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  55. Carreon, L.Y.; Sanders, J.O.; Diab, M.; Sucato, D.J.; Sturm, P.F.; Glassman, S.D. The minimum clinically important difference in Scoliosis Research Society-22 Appearance, Activity, And Pain domains after surgical correction of adolescent idiopathic scoliosis. Spine 2010, 35, 2079–2083. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  56. Kelly, M.P.; Lenke, L.G.; Sponseller, P.D.; Pahys, J.M.; Bastrom, T.P.; Lonner, B.S.; Abel, M.F. The minimum detectable measurement difference for the Scoliosis Research Society-22r in adolescent idiopathic scoliosis: A comparison with the minimum clinically important difference. Spine J. 2019, 19, 1319–1323. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  57. La Poche, A.; Benes, G.; Hariharan, A.; Jain, A.; Samdani, A.; Vorhies, J.; Gabos, P.; Newton, P.O.; Yaszay, B.; Louer, C.; et al. Pseudoarthrosis After Posterior Spinal Fusion in Adolescent Idiopathic Scoliosis: A Multicenter Analysis of Revision Strategies and Outcomes. Spine 2026. epub ahead of printing. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  58. Rybaczek, M.; Kowalski, P.; Mariak, Z.; Grabala, M.; Suszczyńska, J.; Łysoń, T.; Grabala, P. Safety in Spine Surgery: Risk Factors for Intraoperative Blood Loss and Management Strategies. Life 2025, 15, 1615. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  59. Tsantes, A.G.; Papadopoulos, D.V.; Lytras, T.; Tsantes, A.E.; Mavrogenis, A.F.; Koulouvaris, P.; Gelalis, I.D.; Ploumis, A.; Korompilias, A.V.; Benzakour, T.; et al. Association of malnutrition with surgical site infection following spinal surgery: Systematic review and meta-analysis. J. Hosp. Infect. 2020, 104, 111–119. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  60. Lenke, L.G.; Lee, V.; Hassan, F.M. Revision of Surgery for Adolescent Idiopathic Scoliosis: Reasons, Treatments, and Clinical Management with Case Examples. J. Clin. Med. 2024, 13, 2233. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  61. Grabala, P.; Helenius, I.; Shah, S.A.; Larson, A.N.; Buchowski, J.M.; Latalski, M.; Grabala, M.; Guszczyn, T. Impact of Pregnancy on Loss of Deformity Correction After Pedicle Screw Instrumentation for Adolescent Idiopathic Scoliosis. World Neurosurg. 2020, 139, e121–e126. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  62. Singh, M.; Yan, Z.; Daher, M.; Hanna, J.; Diebo, B.G.; Daniels, A.H. Curve Progression and Clinical Outcomes in Pregnant Females with Adolescent Idiopathic Scoliosis: A Systematic Review and Meta-Analysis. World Neurosurg. 2024, 190, e525–e536. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  63. Nandoliya, K.R.; Sadagopan, N.S.; Alwakeal, A.; Kemeny, H.; Cloney, M.; Dahdaleh, N.S.; Koski, T.; El Tecle, N. Adolescent Idiopathic Scoliosis and Pregnancy. Cureus 2023, 15, e46782. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  64. Falick-Michaeli, T.; Schroeder, J.E.; Barzilay, Y.; Luria, M.; Itshayek, E.; Kaplan, L. Adolescent Idiopathic Scoliosis and Pregnancy: An Unsolved Paradigm. Glob. Spine J. 2015, 5, 179–184, Erratum in Global Spine J. 2016, 6, 626. https://doi.org/10.1055/s-0036-1586242. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  65. Grabala, P.; Helenius, I.; Buchowski, J.M.; Larson, A.N.; Shah, S.A. Back Pain and Outcomes of Pregnancy After Instrumented Spinal Fusion for Adolescent Idiopathic Scoliosis. World Neurosurg. 2019, 124, e404–e410. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  66. Landrum, M.; Nocka, H.R.; Ashebo, L.; Hilmara, D.; MacAlpine, E.; Flynn, J.M.; Ho, M.; Newton, P.O.; Sponseller, P.D.; Lonner, B.S.; et al. Pregnancy and Childbirth After Spinal Fusion for Adolescent Idiopathic Scoliosis. J. Pediatr. Orthop. 2023, 43, 620–625. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  67. Pappajohn, A.F.; Seddio, A.E.; Jahagirdar, O.B.; Islam, W.; Jabbouri, S.S.; Tuason, D.A.; Bahtiyar, M.O.; Grauer, J.N. Pregnant adolescent idiopathic scoliosis patients with prior fusion are more often recommended cesarian section delivery, without clear differences in maternal or fetal complications: A retrospective matched-cohort analysis. Spine Deform. 2026, 14, 1715–1723. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  68. Monazzam, S.; Newton, P.O.; Bastrom, T.P.; Yaszay, B.; Harms Study Group. Multicenter Comparison of the Factors Important in Restoring Thoracic Kyphosis During Posterior Instrumentation for Adolescent Idiopathic Scoliosis. Spine Deform. 2013, 1, 359–364. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  69. Marya, S.; Elmalky, M.; Schroeder, A.; Tambe, A. Correction of Thoracic Hypokyphosis in Adolescent Scoliosis Using Patient-Specific Rod Templating. Healthcare 2023, 11, 980. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  70. Liu, X.; Ma, Z.; Wang, H.; Zhang, X.; Li, S.; Zhang, M.; An, J.; Luo, Z. Effectiveness and Safety of High-Dose Tranexamic Acid in Adolescent Idiopathic Scoliosis Surgery: A Meta-Analysis and Systematic Review. World Neurosurg. 2024, 191, 39–48. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  71. Huang, L.; Li, P.; Gu, X.; Zhao, R.; Ma, X.; Wei, X. The administration of tranexamic acid for corrective surgery involving eight or more spinal levels: A systematic review and meta-analysis. Asian J. Surg. 2022, 45, 681–688. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Article metric data becomes available approximately 24 hours after publication online.