Anatomically-Based Biomagnetic Source Imaging of Gastric Slow Waves

,
,
,
,
,
,
and
1
Auckland Bioengineering Institute, The University of Auckland, Auckland 1010, New Zealand
2
Department of Surgery, Vanderbilt University Medical Center, Nashville, TN 37232, USA
*
Author to whom correspondence should be addressed.

Abstract

Gastric slow waves (SWs) coordinate stomach contractions, and disruptions in these contractions are linked to motility disorders. Non-invasive characterization of SWs can aid in diagnosing motility disorders, but existing analysis techniques lack spatial specificity and direct mapping to stomach anatomy. A novel anatomy-based biomagnetic source imaging method was developed and validated to characterize gastric SW activity from non-invasive magnetogastrograms (MGG) using subject-specific anatomical models of the stomach and torso. A distributed source space was formed from spatial basis functions placed over the stomach surfaces, and the optimal weighting of these basis functions that best reproduced the MGG signals was estimated using a regularized least-squares framework. The method was validated against in vivo SW recordings using both simulated and experimental MGG. The source imaging method successfully reconstructed simultaneous SW events and their propagation patterns. SWs estimated from both simulated and experimental MGG showed spatial correlation scores of up to 0.59 and position errors as low as 3.5% of the proximal-distal axis of the stomach (i.e., 7.6 mm) compared to the underlying SW activity. The method accurately identified SW propagation patterns, even in the presence of multiple simultaneously active SW events. The addition of noise up to a 5 dB signal-to-noise ratio resulted in less than 0.4 mm reduction in accuracy, demonstrating the robustness of the method. The developed method effectively identified spatiotemporal dysrhythmias by separating and characterizing multiple simultaneous SW events. This represents significant potential for developing personalized biomarkers for the non-invasive diagnosis of gastric motility disorders.

1. Introduction

The stomach plays a significant role in digestion as gastric contractions are crucial for breaking down ingested food. These contractions are regulated in part by rhythmic bioelectrical activities called slow waves (SWs) [1]. In a healthy intact stomach, SWs arise from a pacemaker region located on the greater curvature and propagate as regular and organized events in the same direction as contractions [2]. Disruptions in this electrical conduction system are linked with gastric motility disorders [2], which severely impair the quality of life and affect up to 10% of the population [3,4]. However, the diagnosis of many gastric motility disorders remains largely symptom-based, with no routinely available objective biomarkers that reliably identify the underlying gastric dysfunction [5,6]. Therefore, characterization of SW dysrhythmias is a promising option for the assessment of gastric dysfunction phenotypes and could enable treatment stratification.
The most accurate recordings of SW propagation have been made from the gastric serosa using dense electrode arrays, which is a technique known as high-resolution mapping [2]. However, it has limited utility as a routine diagnostic tool because it is highly invasive. Gastric SWs can also be measured non-invasively as electrogastrograms (EGG) using cutaneous electrodes [7,8]. Additionally, the magnetic fields generated by gastric SWs can be recorded non-invasively as magnetogastrograms (MGG) using superconducting quantum interference devices (SQUIDs) positioned above the abdomen [9,10]. MGG signals have the advantage of being less attenuated by the abdominal wall than EGG signals, and hence MGG is more sensitive to sources located deeper within the torso [11].
The non-invasive detection of SW dysrhythmias is challenging because they can occur within the normal SW frequency range and can be spatially complex [2]. The estimation of SW propagation speed and direction from both EGG and MGG has been previously demonstrated [9,12,13], but these methods cannot fully characterize the underlying SW activity. In particular, the separation of multiple sources is crucial for dysrhythmia detection because multiple SW events are typically simultaneously present in both the healthy and diseased stomach [2].
The characterization of gastric SWs from non-invasive modalities is confounded by the high variability of stomach shape, position, and orientation both within the population and over time within individuals [14,15]. Subject-specific gastric anatomy can be used to provide context for the interpretation of conventional EGG and MGG analysis methods, but can also be directly integrated into analysis methods for more realistic solutions. Anatomy-based source imaging has been widely utilized to assess the function of the human brain [16,17] and heart [18,19]. Similarly, constraining estimates of gastric SW activity to the surface of subject-specific stomach models can be effective in the localization of simultaneously active events in different stomach regions. Anatomically-based gastric source imaging methods have been investigated computationally [20,21], but validation against invasive SW measurements has not been achieved.
We recently developed a gastric electroanatomical mapping system that employed an electromagnetic coil localization technique to acquire the anatomy of the stomach and to register serosal recording electrodes to the stomach geometry [22]. In this study, a novel anatomy-based biomagnetic source imaging method was developed to reconstruct SW activity from non-invasive MGG recordings and validated for the first time using experimental serosal recordings.

2. Materials and Methods

The developed biomagnetic source imaging method reconstructs gastric SW activity from MGG recordings using subject-specific anatomical models of the stomach and torso. Reconstruction is performed using a forward model that computes the magnetic fields for a given source distribution over the stomach anatomy, and an inverse solution that recovers the contributing source distribution from the measured fields.

2.1. Magnetic Field Computation—Forward Problem

When a field source is embedded within a conducting medium, generated magnetic fields (MFs) result from both the primary source itself and a secondary source produced by a fictitious current at the interface between regions of different electrical conductivity [23,24]. Using the torso as the volume conductor, the MF at a sensor location was then computed as follows [25,26]:
B = B d + B v
where B d is the MF generated by the primary source and B v is the MF due to the secondary source. The MF at a field point (i.e., a sensor location) r s due to a primary dipole source ( ρ ) was computed from [26]:
B d ( r s ) = μ 0 ρ × r 4 π r 3
where μ 0 is the permeability of free space, r is the displacement vector from ρ to r s , r is the absolute distance of r , and × represents the cross product. The MF due to the secondary source was computed from [25,26]:
B v ( r s ) = − μ 0 4 π ∑ j = 1 N σ j ∫ G ∇ ϕ ( r d ) × r r 3 d Ω j
where N is the number of surfaces enclosing regions of different conductivity, σ j is the constant conductivity of the j-th region defined by the surface Ω j , and ϕ ( r d ) is the electrical potential on Ω j due to the dipole source.

2.2. Source Imaging—Inverse Problem

2.2.1. Source Space Construction

A distributed source space was constructed for each subject-specific stomach geometry by placing spatial basis functions (i.e., electric dipoles) with a fixed position and orientation over the stomach surface. SW events typically form ring-shaped wavefronts with depolarization occurring at the leading edge and repolarization occurring at the trailing edge. Therefore, rings of electric dipoles oriented normal to the surface in the mucosal direction were used as basis functions because their resultant MFs well-approximated the MFs produced by a ring-shaped SW event.
Basis function dipoles were placed along 24 splines that extended from the lower esophageal sphincter to the pyloric sphincter (Figure 1a). The dipole magnitudes were scaled using the distances between neighboring splines and basis functions. Then, the dipoles were grouped into circumferential rings (Figure 1a), and the mean position of each dipole group was used to define the proximal-distal axis of the stomach. Therefore, each dipole group was indexed to a corresponding position along this axis (Figure 1b).
Figure 1. Overview of the basis-function source model and transmembrane potential approximation. (a) A representative circumferential ring of basis function dipoles (green) that were placed along splines (black) extending from the lower esophageal sphincter to the pyloric sphincter. (b) The proximal-distal organ axis with the corresponding location (green x) of the basis function ring in (a). (c) Approximation of transmembrane potentials from extracellular SW recordings to simulate MGG signals. The transmembrane potential for each grid point on the stomach geometry was estimated by aligning SW events recorded from the gastric serosa (top) with a SW template (bottom) using activation times.
The forward magnetic field solution for each basis function group was computed as described in Section 2.1 and a lead-field matrix ( L ) was then constructed from these solutions. Each column in L contained the magnetic field gradients at every sensor generated by a basis function group. L is related to the MGG measurements B by:
B = L W + ϵ
where W is an array of weights that defines the contribution of each basis function group, and ϵ is noise.

2.2.2. Regularized Inverse Solution

Penalized linear regression was used to estimate the spatial contribution of each basis function group (i.e., distribution of SW activity) from MGG using,
W ^ ( t ) = arg min W ≥ 0 ∥ B ( t ) − L W ( t ) ∥ 2 2 + λ ∥ W ( t ) ∥ 2 2
where t is time, W ^ ( t ) are the estimated weight values for the basis function groups, and ∥ W ( t ) ∥ 2 2 is an L 2 -norm regularization term that is scaled by the regularization parameter λ .
The first term in Equation (5) compared the measured MF gradients ( B ( t ) ) for each sensor with the values defined by the lead-field matrix L and scaled by the weights W ( t ) . The second term is a penalty that acted to avoid over-fitting by encouraging weight values to be small and to have a Gaussian distribution.
Weight values were restricted to be positive or zero because an inverted basis function orientation is not physiological. Also, the fundus was excluded in source imaging because this region of the stomach does not exhibit SWs in vivo [27]. This was achieved by discarding the columns of L that corresponded to the most proximal third of the stomach.
The quasi-Newton optimization method in MATLAB (Version R2021b, MathWorks Inc., Natick, MA, USA) was used to solve Equation (5) for each time-step. Therefore, the inverse solution W ^ ( t ) for each MGG time-series resulted in a 2D matrix, where the rows correspond to weight values for different locations along the proximal-distal axis, and the columns are solutions for different time-points.

2.2.3. Regularization Parameter Sensitivity

A parameter sweep was performed for the regularization parameter, λ , for each SW activity pattern to evaluate the sensitivity of source imaging to the strength of regularization. The parameter sweep was done by computing the 2D correlation between the source imaging solution and the ground-truth activity. The λ value resulting in the largest correlation score was determined as the optimal regularization parameter.

2.3. In Silico Validation

The source imaging method was first applied to simulated MGG derived from experimental SW data previously acquired from two pigs [22], where the recorded activation sequences had been extrapolated onto the corresponding stomach geometry. These activation sequences were used to define the primary dipole sources by approximating the underlying transmembrane potential distribution.

2.3.1. Approximation of Transmembrane Potentials

Transmembrane potentials over the stomach geometry were approximated from the extracellular SW recordings using detected activation times and a gastric cell model [28]. First, a transmembrane SW template was computed by solving the cell model for a single cell over one SW cycle. The duration of the template, defined by the activation-recovery interval, was scaled to 4.3 s, corresponding to the mean interval obtained from monophasic recordings in pigs [29]. The template was then assigned to each of the 280,000 grid points over the stomach geometry, and the depolarization phase of the template was temporally aligned with the corresponding activation time at that grid point, as illustrated in Figure 1c. This process was applied to all activation times at each grid point, resulting in an approximation of transmembrane potential activity over multiple SW cycles. Finally, electric dipole sources were defined from spatial gradients of transmembrane potentials as follows:
ρ = − σ ∇ V m
where ρ is the dipole at a grid point, σ corresponds to the tissue conductivity, and V m is the transmembrane potential. Dipoles were then vectorially summed to create equivalent dipole vectors within each element of the stomach mesh, and each equivalent dipole vector was positioned at the mean location of the contributing solution points weighted by dipole magnitudes [30].

2.3.2. Simulated Slow Wave Patterns

Four SW datasets were used for validation, representing two distinct propagation patterns: (i) a 1.9-cpm bradygastric activity (i.e., less than 3.2 cpm) produced by an ectopic pacemaker recorded in one pig and (ii) a 1.2-cpm bradygastric activity with an anterograde (proximal to distal) propagation recorded in the second pig. Additionally, a 3.2-cpm normogastric activity with the same activation site and propagation pattern was constructed from each of these two cases by repeating the SW activity with reduced periods between consecutive events. The frequency in these additional datasets corresponded to the mean SW frequency reported for pigs [27].

2.3.3. The Impact of Noise

The impact of uncorrelated measurement noise was assessed by adding Gaussian white noise to simulated MGG signals. Seven noise levels with signal-to-noise ratios that ranged between 20 dB and − 10 dB at 5 dB intervals were investigated. White noise was added to the MGG signals to generate 10 independent noisy datasets at each noise level using the awgn function in MATLAB.

2.4. In Vivo Validation

The animal study protocol was approved by the Institutional Animal Care and Use Committee of Vanderbilt University. The experimental procedures for simultaneous MGG and high-resolution serosal SW recordings, along with the associated signal processing methods, have been previously described in detail [22]. Briefly, MGG signals were recorded using a SQUID gradiometer array, downsampled from 3 kHz to 30 Hz, and bandpass filtered between 1 and 50 cpm. Gastric activity was isolated using second-order blind identification (SOBI), followed by a 5-s moving average filter. Simultaneously, gastric SW potentials were recorded using high-resolution electrode arrays placed on the anterior and posterior serosal surfaces of the stomach. Activation times were identified from these recordings and interpolated over the reconstructed subject-specific stomach geometries to generate three-dimensional SW activation maps [22]. For validation, biomagnetic source imaging was applied to the processed MGG signals corresponding to the two recorded SW propagation patterns (see Section 2.3.2), and the contributions of the spatial basis functions along the proximal-distal stomach axis were estimated. The resulting spatiotemporal SW activity distributions were then compared with the corresponding ground-truth activity.

2.5. Accuracy Metrics

  • Spatial Error: The recorded SW activation times on the stomach geometry were divided into time segments with a step of 0.33 s. For each segment, the mean position of the grid points activating within that segment was computed and mapped to the proximal-distal stomach axis by finding the nearest location on the axis, defining the ground-truth SW position along the axis at that time. These ground-truth positions were delayed by half the duration of the assumed SW morphology (2.15 s) so that they corresponded to the midpoint of the SW event rather than the leading edge of the wavefront. The estimated SW position was computed from the source imaging solution as the mean position of the basis function groups weighted by the estimated weight values ( W ^ ). The spatial error was then quantified as the separation between the ground-truth and estimated positions along the axis over the course of a SW cycle.
  • Spatial Correlation: As propagating SW events are distributed over stomach regions, rather than existing at discrete locations, it is valuable to compare the spatial distribution of SW estimates against that of the ground-truth SW activity. Therefore, the 3D maps of transmembrane potentials were collapsed down to the proximal-distal stomach axis using the percentage of depolarized solution points within each circumferential ring segment. Depolarization values were binned using a threshold of 25% to create a binary representation of the depolarization state as follows: greater than or equal to 25% depolarization was classified as substantial SW activity and assigned a value of 1; otherwise, the activity was classified as a lack of substantial SW activity and assigned a value of 0. A 2D matrix was constructed from the binary representation of SW activity, with rows that represented locations along the proximal-distal axis and columns that represented time-points. This binary 2D matrix was correlated against the source imaging solutions, which were analogous 2D matrices, using the corr2 function in MATLAB. For the 2D matrices A and B , the correlation was computed using the following expression:
    C O R 2 D ( A , B ) = ∑ m ∑ n A m n − A ¯ B m n − B ¯ ∑ m ∑ n A m n − A ¯ 2 ∑ m ∑ n B m n − B ¯ 2
    where m and n correspond to rows and columns in the 2D matrices, A ¯ is the mean of A and B ¯ is the mean of B .

3. Results

3.1. Forward Problem Solutions

The transmembrane potentials approximated from the SW activation sequences captured the spatial and temporal features of SW propagation. For the activation maps in Figure 2a, Figure 2b shows heat maps of the approximated transmembrane potentials over the 3D stomach models at three time-points within a SW cycle. The transmembrane potentials formed organized wavefronts that followed the recorded activation patterns (as shown by the potential traces at three stomach locations in Figure 2c), and wavefront width varied with the propagation speed.
Figure 2. Overview of the forward model from slow wave activation to simulated MGG signals. (a) Extrapolated slow wave activation sequences used for the validation, with electrode positions (black dots) used in high-resolution extracellular recordings and four representative magnetic field sensor positions (pink diamonds). (b) Estimated transmembrane potential and equivalent current dipoles (purple arrows) for three time-points during the propagation cycle of an event. (c) Approximated transmembrane potential traces for three stomach locations. (d) Experimental and simulated MGG for four sensors along the proximal-distal axis with correlation coefficients.
The equivalent current dipoles computed from the transmembrane potentials are shown in Figure 2b. At the first time-point, SW events were initially represented by dipoles located near the leading edges of the wavefronts and oriented in the direction of propagation. After wavefronts were established, the depolarization and repolarization phases were either represented separately by opposing dipoles positioned at the leading and trailing edges or by net dipoles positioned at the approximate center of the wavefronts. This dipole configuration varied with the location of events relative to the element boundaries of the stomach mesh. Finally, when SW events exited the stomach, there was a period with only the repolarization phase, during which all dipoles were oriented opposite to the direction of propagation.
The simulated and experimental MGG signals showed deflections at similar times across several sensor locations for both SW activity patterns (Figure 2d). The simulated signals generally reproduced the timing and polarity of the main experimental deflections, although they were often sharper and shorter in duration. Agreement varied spatially across the sensor array, with correlation coefficients ranging from 0.10 to 0.74 for the ectopic activity pattern and from 0.41 to 0.67 for the anterograde activity pattern.

3.2. Source Imaging Solutions

3.2.1. Regularization Parameter Sensitivity

The regularization parameter ( λ ) sweep for all the simulated MGG datasets had a single broad peak at a λ value from 1.98 × 10 − 3 to 6.90 × 10 − 3 (Table 1). The parameter sweep for a representative case (bradygastric ectopic activity) is shown in Figure 3a. The mapping of the ground-truth activity along the proximal-distal axis for this dataset showed activity initially occurring within the distal corpus and then moving proximally and distally in agreement with the 3D activation maps (Figure 3b-i). The depolarization percentage for circumferential stomach segments varied throughout the propagation cycle as the configuration and orientation of the wavefronts changed.
Table 1. Source imaging accuracy for optimal regularization parameter ( λ ) values. Correlation is the 2D correlation between source imaging solutions and the ground-truth activity maps along the proximal-distal axis, and position error is reported as the percentage of the proximal-distal axis.
Figure 3. Regularization parameter sensitivity analysis for a representative activity pattern (bradygastric ectopic activity). (a) Parameter sweep of the regularization parameter ( λ ) against the 2D correlation between the source imaging solution and the ground-truth activity. Black arrows indicate the selected regularization parameter values. (b) Heat maps of SW activity along the proximal-distal stomach axis. (i) Ground-truth activity represented as the percentage of depolarized solution-points within circumferential stomach segments. Source imaging solutions are shown for (iii) the optimum level of regularization ( λ O P T ), as well as for an order of magnitude (ii) below and (iv) above the optimum. Non-zero weight values indicate SW activity. Red dashed arrows indicate the direction of SW propagation.
Source imaging using regularization parameter values lower than the optimum resulted in relatively unstructured solutions composed of multiple localized centers (Figure 3(bii)). Source imaging using the optimal level of regularization produced solutions that had fewer, more distributed centers with propagation that was in general agreement with the ground-truth activity (Figure 3(biii)). Parameter values above the optimum were similar to the optimum solutions, but had more distributed centers with less defined or absent connections. For example, the optimum solution for the bradygastric ectopic activity displayed relatively continuous propagation from the distal corpus in both a proximal and distal direction. In contrast, a parameter value an order of magnitude above the optimum provided a solution with activity initially in the distal corpus and then in both the proximal and distal stomach, but these centers were relatively isolated from one another (Figure 3(biv)).

3.2.2. Bradygastric Activity

Source imaging solutions with the optimal λ for the bradygastric activity patterns are shown in Figure 4, and the solutions are compared against the transmembrane potentials at three time-points during the propagation cycle in Figure 4c. The first time-point for the ectopic activity was shortly after the initiation of the event, when the wavefront was on the greater curvature within the distal corpus. Source imaging for this time-point correctly indicated activity within the distal corpus near the position of the wavefront. The proximal and distal aspects of the event were largely separated in the second time-point and were propagating in opposite directions. At this time, source imaging indicated activity near the ground-truth proximal and distal activity; however, estimates were relatively low in magnitude and distributed. At the third time-point, the proximal and distal wavefronts were fully separated and were within the mid-corpus and distal antrum, respectively. Source imaging indicated activity at these locations with larger and more localized weights than in the previous time-point. The correlation with the ground-truth activity was 0.44, and the mean position error was 3.5 ± 2.3% of the organ axis (7.6 ± 5.0 mm).
Figure 4. Source imaging for bradygastric MGG simulated from experimentally recorded activation sequences using the optimal level of regularization. (a) Ground-truth SW activity along the proximal-distal axis represented as the percentage of depolarized solution points within circumferential rings. (b) Source imaging results. Weight values for the basis function groups along the proximal-distal axis, where non-zero weight values indicate SW activity. (c) Comparison of SW potentials (red/white) and source imaging estimations (rainbow, red = 1 and blue = 0 as scaled by the colormap in (b)) for three time-points during the propagation cycle of an event. Red dashed and black arrows indicate the direction of wavefront propagation.
The first time-point for the anterograde activity corresponded to shortly after the initiation of the event, when the wavefront was on the greater curvature within the proximal corpus. Source imaging for this time-point correctly indicated activity within the proximal corpus near the position of the wavefront. The estimated activity was wider than the underlying SW event, but the largest weight values were near the position of the wavefront. The wavefront formed a complete and relatively narrow circumferential ring within the distal corpus at the second time-point. Source imaging failed to detect this activity and incorrectly estimated activity within the antrum, albeit weakly. At the third time-point, the wavefront was within the antrum and was substantially wider compared to the previous time-point. Source imaging correctly indicated activity at the location of the wavefront. The correlation with the ground-truth activity was 0.59, and the mean position error was 6.4 ± 5.1% of the organ axis (12.9 ± 10.4 mm).

3.2.3. Normogastric Activity

Source imaging solutions for the 3.2 cpm activity patterns using the optimal λ are shown in Figure 5. Biomagnetic source imaging successfully identified multiple simultaneously propagating events (Figure 5a,b). This was more evident in Subject 2, where multiple events with an anterograde propagation were temporally overlapped.
Figure 5. Source imaging for normogastric MGG simulated from experimentally recorded activation sequences using the optimal level of regularization. (a) Ground-truth SWs. (b) Source imaging results. (c) Comparison of SW potentials (red/white) and source imaging estimations (rainbow as scaled by the colormap in (b)).
The solutions are compared against the transmembrane potentials at three time-points during the propagation cycle of a SW event in Figure 5c. Similar to the 1.9 cpm activity in Subject 1, the ectopic activation site was successfully identified at the first time-point. The source imaging also estimated a potential event in the proximal corpus, for which the weight values were noticeably smaller. This was likely due to the termination of the earlier event. The source imaging results in the second and third time-points in the 3.2 cpm activity were similar to those of the 1.9 cpm activity, as there was no significant difference in the SW characteristics at those time-points between the two activities. The correlation with the ground-truth activity was 0.30, and the mean position error was 3.9 ± 3.1% of the organ axis (8.5 ± 6.7 mm).
Similarly, at the first time-point in Subject 2, there was a SW event with an incomplete circumferential wavefront within the proximal corpus, while another event forming a complete circumferential wavefront was within the antrum. Source imaging for this time-point indicated SW activity at both of these locations. A wavefront with a complete and relatively narrow circumferential ring was recorded within the distal corpus in the second time-point. Similar to the 1.2 cpm case, source imaging failed to detect this activity and incorrectly indicated activity within the antrum with low weight values. The third time-point largely mirrored the first, as the wavefronts were within both the proximal corpus and antrum. Source imaging for this time-point correctly indicated SW activities at both of these locations. Discrete wavefront position estimates largely followed the anterograde propagation of the underlying wavefront. The mean position error was 6.0 ± 4.0% of the organ axis (12.2 ± 8.1 mm), while the correlation with the ground-truth activity was 0.44.

3.3. The Effect of Measurement Noise

The impact of noise on source imaging performance is shown in Figure 6. The propagation patterns estimated at 20 dB were effectively identical to those obtained from the noiseless MGG. This is supported by the <0.01 change in mean correlation values and <0.1% (<0.1 mm) change in mean position errors at 20 dB compared to the noiseless results for all four activity patterns. Increasing the noise level to 5 dB also did not substantially change the estimations, and the change in mean position errors was <0.2% (<0.4 mm), and the change in mean correlation values was <0.04 compared to the noiseless case for all four activity patterns. The noise levels between 0 dB and −10 dB increased the magnitude of errors compared to the lower noise levels. The change in mean position errors compared to the noiseless case was within the range of 1.4–3.1% at the − 10 dB noise level. Furthermore, mean correlation values for all activity patterns converged on approximately 0.2 at this noise level, and changes compared to noiseless MGG ranged between − 0.10 and − 0.34 .
Figure 6. The effect of uncorrelated measurement noise on source imaging accuracy. (a) Correlations between the source imaging solutions and ground-truth activity along the proximal-distal axis. (b) Position error reported as the percentage of the proximal-distal axis. Error bars are standard deviations.

3.4. In Vivo Validation

Source imaging results for the in vivo MGG recordings are shown in Figure 7. In Subject 1 (the 1.9 cpm ectopic activity), the retrograde component of propagation was clearly identifiable from source imaging, whereas the anterograde component was mainly missing in the estimation (Figure 7b). Source imaging results and transmembrane potential over the 3D stomach models are compared for four time-points in Figure 7c for a 120 s time-segment of MGG. The first time-point was shortly after the initiation of the event, when the ectopic wavefront was on the greater curvature within the distal corpus. Source imaging for this time-point indicated activity within the distal corpus, with higher weight values near the position of the wavefront. The proximal and distal components of the event were largely separated at the second time-point and were propagating in opposite directions; however, source imaging results were largely unchanged compared to the first time-point. At the third time-point, the proximal and distal wavefronts were fully separated within the proximal corpus and distal antrum, respectively. Source imaging indicated retrograde propagation from the pacemaking site, but the estimated position lagged behind the ground-truth SW position. Additionally, the distal wavefront was not detected. At the fourth time-point, the proximal wavefront was in the process of exiting the coverage of the recording array. Source imaging results indicated SW activity across a broad region within the proximal stomach that included the ground-truth position. Discrete SW positions estimated from the weight values largely tracked with the retrograde propagation of the underlying proximal event. Mean position error was 11.9 ± 7.0% of the organ axis (25.7 ± 15.1 mm).
Figure 7. Source imaging for experimental MGG using the optimal level of regularization. (a) Ground- truth SWs. (b) Source imaging results. (c) Comparison of SW potentials (red/white) and source imaging estimations (rainbow as scaled by the colormap in (b)).
In Subject 2, anterograde propagation was clearly identifiable from source imaging for the experimental 1.2 cpm anterograde SW activity (Figure 7b). Source imaging results and transmembrane potential over the 3D stomach model are compared in Figure 7c for four time-points during the propagation of the second event in the 120 s time-segment of MGG. The first time-point was approximately 8 s prior to the initiation of the event. Source imaging results indicated SW activity within the proximal corpus at this time. At the second time-point, the wavefront was on the greater curvature within the proximal corpus, and source imaging results indicated activity within the distal corpus. The wavefront formed a complete and relatively narrow circumferential ring within the distal corpus at the third time-point. Similar to the results for simulated MGG, source imaging failed to detect this activity and indicated activity within the antrum. At the fourth time-point, the wavefront was within the antrum and was substantially wider than at the previous time-point, and source imaging correctly indicated activity within the antrum. Discrete SW positions estimated from the weight values largely tracked with the anterograde propagation of the underlying wavefront with a mean position error of 16.0 ± 8.8% of the organ axis (32.6 ± 17.9 mm).

4. Discussion

This study demonstrates the feasibility of a novel anatomically based source imaging approach for resolving gastric SW activity from non-invasive MGG measurements. A distributed source space was generated by placing spatial basis functions over the surface of subject-specific stomach models. The optimal weighting of these basis functions, which best reproduced the MGG signals, was estimated using a regularized least-squares framework. Previously acquired simultaneous gastric SW and MGG recordings, together with the corresponding subject-specific anatomies, were used to validate the method [22]. The estimated weight values represented the distribution of SW activity across the stomach, and their accuracy was assessed against the underlying SW potentials. To our knowledge, this is the first study to validate the spatiotemporal reconstruction of gastric SWs from non-invasive measurements against experimental serosal recordings.
Our proposed source imaging method utilized the subject-specific stomach anatomy and enabled the characterization of SW activity along the proximal-to-distal stomach axis. Incorporating subject anatomy in the localization of bioelectrical activities significantly improved the spatiotemporal characterization of electrical signals in the brain and heart and provided new clinical biomarkers [16,17,18,19]. Similarly, our results have shown that normal and dysrhythmic propagation patterns were successfully identified from the simulated MGG data with high spatial correlation values and low position errors compared to the underlying SW activity, providing evidence that the method was spatially accurate (see Table 1). While the experimental MGG data showed limited quantitative spatial agreement with the underlying SW activity, the method was able to correctly identify the ectopic activation site and associated retrograde propagation in Subject 1, as well as the anterograde propagation pattern in Subject 2, despite a temporal lag. This represents a significant improvement over existing non-invasive (EGG or MGG) analysis methods, where SW activity is mostly assessed in the frequency domain and complex spatial characteristics of SWs cannot be reliably identified [8,31].
The provision of SW activity estimates mapped directly to subject-specific stomach anatomy enabled the successful localization of the dysrhythmic events in the case of ectopic activity, which can significantly aid in the identification of dysrhythmias and the diagnosis of motility disorders. Furthermore, the capacity to identify and separate SW events, as demonstrated for the simulated 3.2 cpm ectopic and anterograde cases, enables the characterization of activity patterns with multiple simultaneously active events. This also represents a significant advancement because existing non-invasive EGG/MGG analysis methods cannot spatially separate simultaneously active events, which are common in both the healthy and dysrhythmic stomach [2], and instead provide insights on the aggregate activity [9,12,13].
The developed source imaging technique was first validated using simulated MGG data, which was computed from the serosal measurements by approximating transmembrane potentials using SW activation times and a morphology template constructed from the experimentally validated mathematical SW model [28]. This provided a sufficient approximation of the SW activity for forward modeling. The estimated SW transmembrane potentials formed organized wavefronts with widths that varied with propagation speed, which is consistent with organ-level simulations [13,32]. To date, MGG has been primarily computed from 3D organ-level SW simulations tuned to replicate experimental observations, and the lack of corresponding experimental MGG data to compare against has prevented the direct validation of simulation methods [13,32]. In this study, equivalent current dipoles, which are commonly used in the forward problem of bioelectrical activities, were quantified from the approximated transmembrane potentials to compute MGG signals. This approach enabled the comparison of experimental and simulated MGG for the same SW activity for the first time, which is important for the development and validation of forward models.
The simulated MGG signals correlated with the experimental measurements at the SQUID sensors, with correlation coefficients reaching up to 0.74. The differences between simulated and experimental MGG could be explained by a combination of forward modeling errors and challenges relating to MGG signal processing. Previous computational investigations have demonstrated that biomagnetic sensor dimensions and source-to-sensor distance can influence measurement performance [33]. Although the forward model used in this study accounted for the SQUID gradiometer configuration, the finite sensing area of the individual pickup coils was not explicitly considered. This simplification may introduce discrepancies between modelled and measured magnetic fields, potentially influencing source reconstruction accuracy. In addition, the MGG simulations used homogeneous volume conductor models and neglected signal attenuation due to the tissue layers of the abdomen, but the impact of these tissue layers is expected to be relatively small [11,25].
Some of the propagation patterns estimated from source imaging were spatially discontinuous, as there were regions of high weight values that were not connected. For example, the locations of SW activity within the corpus and antrum for the anterograde activity patterns had an indistinct or absent connection despite representing a continuous period of propagation. In particular, relatively narrow circumferential wavefronts within the distal corpus were not accurately identified in either simulated or experimental MGG, whereas reconstruction was more accurate when these wavefronts widened towards the antrum. A possible explanation for this is the low magnitude of MGG signals during this stage of propagation, which may be influenced by the orientation and spatial configuration of the wavefronts relative to the magnetic field sensors. Low-magnitude MGG signals produce small residual errors, and regularization encourages weight values towards zero. Furthermore, the simplified source representation using circumferential basis function groups may limit the reconstruction of certain wavefront configurations such as circumferential propagation. The estimation of an activation sequence instead of single time-point activity, as commonly done for ECG source imaging, would address this issue and enable the estimation of continuous propagation patterns [34,35]. However, a significantly different approach would be required for the estimation of activation sequences in the stomach compared to the heart to account for the simultaneous propagation of multiple events.
The source imaging method was substantially robust against uncorrelated measurement noise up to a signal-to-noise ratio of 5 dB (see Figure 6). The signal-to-noise ratio for MGG would be expected to be greater than 5 dB given the magnitude of MGG, the sensitivity of SQUID gradiometers, and the effectiveness of shielding [36]. However, the use of additive Gaussian white noise does not fully capture the characteristics of experimental MGG recordings since sampling rate, measurement bandwidth, and filtering can also influence the effective noise variance and temporal correlations [37], potentially affecting reconstruction accuracy under realistic acquisition conditions. In addition, correlated noise is also present in MGG measurements. This interference is of varied origin, but biological sources are the most significant contributors because external noise sources are dampened by the magnetic shielding and because the gradiometer sensor arrangement is most sensitive to nearby sources. These noise sources can be effectively minimized using source separation methods such as SOBI because they generally occur at distinct frequencies [38,39]. Furthermore, results also suggested that the impact of uncorrelated measurement noise is largely additive.
The gastric electroanatomical mapping technique [22] enabled the in vivo validation by recording ground-truth serosal SW activity simultaneously with MGG measurements. Source imaging using experimental MGG recordings correctly identified anterograde propagation and the retrograde component of ectopic activity, although the estimated activity patterns were spatially and temporally offset from the ground-truth activity. The characterization of the ectopic activity pattern, in particular, is significant because the direction of propagation was not identifiable from the MGG analysis [22]. However, the quantitative accuracy metrics computed did not fully reflect these qualitative observations, and relatively lower agreement with the ground-truth spatiotemporal activity distributions was observed compared to simulated MGG. This reduced accuracy may be partly attributed to discrepancies between simulated and experimental MGG, which may arise from limitations in both MGG signal processing and forward modeling. In particular, the SOBI method used in this study primarily relies on temporal information and may not fully preserve the spatial coherence of the magnetic field distributions. These discrepancies were reflected in the multiple localized maxima observed in experimental MGG, compared with the single, more spatially extended maximum observed in simulated MGG. Furthermore, the low spatial correlations may partly reflect the evaluation methodology, which is sensitive to spatial and temporal offsets. Despite these quantitative limitations, the ability to identify abnormal SW sources and characterize their propagation patterns may still provide clinically relevant information.
Distributed basis functions were used in the source imaging, and clustering them into circumferential groups simplified the source space to the proximal-distal axis and made the source imaging problem tractable; this is an approach that has previously been used for EGG source imaging [20]. This simplification meant that the estimated activity patterns only reflected anterograde and retrograde propagation. Propagation along the proximal-distal stomach axis is generally dominant because of the formation of annular wavefronts after rapid circumferential propagation [2]. Therefore, the primary components of propagation should remain identifiable despite this simplification. Relaxing this constraint may enable circumferential propagation to be identified; however, alternative spatial and temporal constraints would be required to achieve realistic solutions. Potential constraints include restricting the time-course of estimates to have realistic activation and refractory periods and making activation dependent on the activation history of neighboring solution points.
The regularized least-squares framework with non-negative constraints used in this study provides a computationally tractable approach to the ill-posed inverse problem by limiting source magnitudes and preventing reversal of the predefined dipole orientations. Compared with less constrained distributed source reconstruction methods, the predefined source orientations and circumferential grouping reduce the number of unknown parameters and simplify the estimation of gastric SW activity. However, the use of L2 regularization may suppress weaker source activity, while the independent reconstruction of individual time-points does not explicitly enforce temporal continuity. Alternative approaches incorporating sparsity-promoting regularization or spatiotemporal constraints may improve the localization and continuity of reconstructed SW events, but they may also introduce greater computational complexity and additional assumptions about the spatial and temporal characteristics of gastric SW activity. Another important consideration within the proposed framework is the selection of the regularization parameter, which controls the balance between data fidelity and solution regularization. In this study, the regularization parameter was selected based on its correlation with ground-truth SW activity to assess reconstruction performance under optimal regularization conditions. In practice, clinical applications would require parameter selection independently of ground-truth measurements. Established methods that have been previously used in cardiac and brain source imaging, such as the L-curve method [40,41], and generalized cross-validation [42], could provide alternatives for selecting the regularization parameter directly from MGG measurements. The suitability of these approaches for gastric biomagnetic source imaging requires further investigation.
While this study demonstrated the potential of MGG source imaging for the non-invasive characterization of gastric SWs, validation was limited to data from two pigs, representing a restricted range of activity patterns and anatomical geometries. In addition, the SW activity recorded in both pigs exhibited bradygastric rhythms (<3.2 cpm), with ectopic activation observed in one pig. Such abnormal SW patterns, including ectopic activation, are commonly observed in pig studies, likely due to the effects of surgery, anesthesia, and organ handling [43]. Simulating normal-rhythm activity (3.2 cpm) using the experimentally recorded SW patterns enabled the proposed method to be evaluated under normal-rhythm conditions, thereby allowing the influence of SW frequency on source imaging accuracy to be assessed. However, a broader range of SW activity patterns and anatomical geometries, along with a larger number of subjects, is required in future investigations to test how generalizable the results are.
Additionally, an alternative non-invasive method for acquiring stomach geometry would be required for clinical translation. Although magnetic resonance imaging, computed tomography, or X-ray imaging can be used to reconstruct stomach geometry, these measurements are typically acquired in a separate imaging environment and may involve a substantial time delay [44]. Ideally, stomach geometry should be acquired close to the time of MGG recording to minimize source-imaging errors arising from changes in gastric geometry due to body position, gastric filling, digestion, or other physiological factors. Three-dimensional ultrasound represents the most viable bedside alternative for acquiring contemporaneous subject-specific stomach geometry [45,46], as it could be performed alongside MGG measurements and has previously achieved reconstruction errors of approximately 2–3 mm for cardiac and torso geometry [47,48]. The feasibility and accuracy of source imaging using stomach geometries acquired with 3D ultrasound therefore require further in vivo validation. In addition to anatomical imaging, translating the proposed method to human subjects will require reliable MGG acquisition under clinical conditions. Conventional SQUID-based systems require cryogenic cooling and magnetic shielding, resulting in substantial infrastructure and operational costs, which limit their widespread clinical use [10]. Optically pumped magnetometers (OPMs) offer a promising alternative, providing high magnetic field sensitivity without cryogenic cooling and enabling more flexible sensor configurations. Recent studies have also demonstrated their feasibility for recording gastric SW activity [49,50]. However, magnetic shielding and environmental interference remain important challenges for OPM-based measurements, and further validation would be required to establish their suitability for anatomy-based gastric source imaging.

5. Conclusions

This study presents a novel anatomy-based biomagnetic source imaging method for characterizing gastric SW activity using non-invasive magnetogastrography measurements, providing the first-ever biophysical validation of the inverse solution for gastric potentials. By integrating subject-specific stomach models, the method successfully identified key propagation characteristics, including anterograde propagation and the retrograde component of ectopic activity, in both simulated and experimental MGG data, although quantitative reconstruction accuracy was lower for the experimental data. The ability to resolve multiple simultaneous SW events and localize dysrhythmic origins offers a significant advancement in the non-invasive assessment of gastric motility disorders. These findings highlight the potential of the proposed method for identifying gastric SW abnormalities and supporting future clinical diagnostic applications. However, validation was restricted to two pig subjects. Further studies involving larger and more diverse cohorts are necessary to better evaluate reconstruction accuracy, especially in MGG measurements, and establish the generalizability and clinical applicability of the method.

Author Contributions

Conceptualization, R.A. and L.K.C.; methodology, C.E.D. and R.A.; software, C.E.D.; validation, C.E.D., N.D.M. and S.S.; formal analysis, C.E.D. and R.A.; data curation, C.E.D. and R.A.; writing—original draft preparation, C.E.D. and R.A.; writing—review and editing, L.K.C., N.D.M., S.S., N.P., P.D. and L.A.B.; visualization, C.E.D.; supervision, R.A. and L.K.C.; project administration, R.A.; funding acquisition, R.A. and L.A.B. All authors have read and agreed to the published version of the manuscript.

Funding

This research was supported in part by the Marsden Fund Council of New Zealand, the Auckland Bioengineering Institute, and the National Institutes of Health, USA. C.E. Drake was supported by a University of Auckland Doctoral Scholarship.

Institutional Review Board Statement

The animal protocols were approved by the Institutional Animal Care and Use Committee at Vanderbilt University and overseen by the Office of Animal Welfare and Assurance in the Division of Animal Care under protocol M1600245-01, Noninvasive markers of functional nausea in children.

Data Availability Statement

The experimental and simulated magnetogastrography data, gastric slow-wave potentials, and subject-specific anatomical data supporting this study are publicly available in Figshare at https://doi.org/10.6084/m9.figshare.34397580.

Conflicts of Interest

L.K. Cheng, P. Du, and N. Paskaranandavadivel hold intellectual property in the field of gastric electrophysiology and are shareholders in FlexiMap Ltd. P. Du is also a shareholder in Alimetry Ltd. The remaining authors have no competing interests to declare.

References

  1. Huizinga, J.D.; Lammers, W.J.E.P. Gut peristalsis is governed by a multitude of cooperating mechanisms. Am. J. Physiol. Gastrointest. Liver Physiol. 2009, 296, G1–G8. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. O’Grady, G.; Gharibans, A.A.; Du, P.; Huizinga, J.D. The gastric conduction system in health and disease: A translational review. Am. J. Physiol. Gastrointest. Liver Physiol. 2021, 321, G527–G542. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Sperber, A.D.; Bangdiwala, S.I.; Drossman, D.A.; Ghoshal, U.C.; Simren, M.; Tack, J.; Whitehead, W.E.; Dumitrascu, D.L.; Fang, X.; Fukudo, S.; et al. Worldwide Prevalence and Burden of Functional Gastrointestinal Disorders, Results of Rome Foundation Global Study. Gastroenterology 2021, 160, 99–114.e3. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Huang, I.H.; Schol, J.; Khatun, R.; Carbone, F.; Van den Houte, K.; Colomier, E.; Balsiger, L.M.; Törnblom, H.; Vanuytsel, T.; Sundelin, E.; et al. Worldwide prevalence and burden of gastroparesis-like symptoms as defined by the United European Gastroenterology (UEG) and European Society for Neurogastroenterology and Motility (ESNM) consensus on gastroparesis. United Eur. Gastroenterol. J. 2022, 10, 888–897. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Fox, M.R.; Kahrilas, P.J.; Roman, S.; Gyawali, C.P.; Scott, S.M.; Rao, S.S.; Keller, J.; Camilleri, M. International Working Group for Disorders of Gastrointestinal Motility and Function. Clinical measurement of gastrointestinal motility and function: Who, when and which test? Nat. Rev. Gastroenterol. Hepatol. 2018, 15, 568–579. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Avci, R.; Miller, K.J.; Paskaranandavadivel, N.; Bradshaw, L.A.; Huckabee, M.L.; Cheng, L.K. Bioelectrical Signals for the Diagnosis and Therapy of Functional Gastrointestinal Disorders. Appl. Sci. 2020, 10, 8102. [Google Scholar] [CrossRef] [Scilit]
  7. Yin, J.; Chen, J.D.Z. Electrogastrography: Methodology, validation and applications. J. Neurogastroenterol. Motil. 2013, 19, 5–17. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Oczka, D.; Augustynek, M.; Penhaker, M.; Kubicek, J. Electrogastrography measurement systems and analysis methods used in clinical practice and research: Comprehensive review. Front. Med. 2024, 11, 1369753. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Bradshaw, L.A.; Kim, J.H.; Somarajan, S.; Richards, W.O.; Cheng, L.K. Characterization of electrophysiological propagation by multichannel sensors. IEEE Trans. Biomed. Eng. 2016, 63, 1751–1759. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Avci, R.; Nagahawatte, N.D.; Palmada, N.; Du, P.; Bradshaw, L.A.; Cheng, L.K. Advances in sensing stomach function with biomagnetism. In Advances in Bioelectromagnetism; Academic Press: Cambridge, MA, USA, 2026; pp. 259–271. [Google Scholar] [CrossRef] [Scilit]
  11. Kim, J.H.K.; Pullan, A.J.; Bradshaw, L.A.; Cheng, L.K. Influence of body parameters on gastric bioelectric and biomagnetic fields in a realistic volume conductor. Physiol. Meas. 2012, 33, 545–556. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Gharibans, A.A.; Kim, S.; Kunkel, D.; Coleman, T.P. High-Resolution Electrogastrogram: A Novel, Noninvasive Method for Determining Gastric Slow-Wave Direction and Speed. IEEE Trans. Biomed. Eng. 2017, 64, 807–815. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Avci, R.; Eichler, C.E.; Paskaranandavadivel, N.; Du, P.; Angeli-Gordon, T.R.; Bradshaw, L.A.; Cheng, L.K. Characterizing Spatial Signatures of Gastric Electrical Activity Using Biomagnetic Source Localization. IEEE Trans. Biomed. Eng. 2022, 69, 3551–3558. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Lee, S.O.; Arwani, R.; McNeilly, S.; Kunkel, S.; Dadparvar, S.; Maurer, A.H.; Parkman, H.P. Variation of Stomach Shapes in Gastric Emptying Scintigraphy: Correlation With Gastric Emptying Results, Body Weight, and Symptoms. Neurogastroenterol. Motil. 2025, 37, e14968. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Ruenruaysab, K.; Calder, S.; Hayes, T.; Andrews, C.; O’Grady, G.; Gharibans, A.; Du, P. Effects of Anatomical Variations of the Stomach on Body-Surface Gastric Mapping Investigated Using a Large Population-Based Multiscale Simulation Approach. IEEE Trans. Biomed. Eng. 2022, 69, 1369–1377. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Hamid, L.; Habboush, N.; Stern, P.; Japaridze, N.; Aydin, Ü.; Wolters, C.H.; Claussen, J.C.; Heute, U.; Stephani, U.; Galka, A.; et al. Source imaging of deep-brain activity using the regional spatiotemporal Kalman filter. Comput. Methods Programs Biomed. 2021, 200, 105830. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. An, N.; Gao, Z.; Li, W.; Cao, F.; Wang, W.; Xu, W.; Wang, C.; Xiang, M.; Gao, Y.; Wang, D.; et al. Source localization comparison and combination of OPM-MEG and fMRI to detect sensorimotor cortex responses. Comput. Methods Programs Biomed. 2024, 254, 108292. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Ramanathan, C.; Ghanem, R.N.; Jia, P.; Ryu, K.; Rudy, Y. Noninvasive electrocardiographic imaging for cardiac electrophysiology and arrhythmia. Nat. Med. 2004, 10, 422–428. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Wang, Y.; Cuculich, P.S.; Zhang, J.; Desouza, K.A.; Vijayakumar, R.; Chen, J.; Faddis, M.N.; Lindsay, B.D.; Smith, T.W.; Rudy, Y. Noninvasive electroanatomic mapping of human ventricular arrhythmias with electrocardiographic imaging. Sci. Transl. Med. 2011, 3, 98ra84. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Allegra, A.B.; Gharibans, A.A.; Schamberg, G.E.; Kunkel, D.C.; Coleman, T.P. Bayesian inverse methods for spatiotemporal characterization of gastric electrical activity from cutaneous multi-electrode recordings. PLoS ONE 2019, 14, e0220315. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. Eichler, C.E.; Cheng, L.K.; Paskaranandavadivel, N.; Angeli-Gordon, T.R.; Du, P.; Bradshaw, L.A.; Avci, R. Anatomically Constrained Gastric Slow Wave Localization using Biomagnetic Data. Proc. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2022, 2022, 3935–3938. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Drake, C.E.; Cheng, L.K.; Muszynski, N.D.; Somarajan, S.; Paskaranandavadivel, N.; Angeli-Gordon, T.R.; Du, P.; Bradshaw, L.A.; Avci, R. Electroanatomical mapping of the stomach with simultaneous biomagnetic measurements. Comput. Biol. Med. 2023, 165, 107384. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Mosher, J.; Leahy, R.; Lewis, P. EEG and MEG: Forward solutions for inverse methods. IEEE Trans. Biomed. Eng. 1999, 46, 245–259. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Plonsey, R. The nature of sources of bioelectric and biomagnetic fields. Biophys. J. 1982, 39, 309–312. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Geselowitz, D. On the magnetic field generated outside an inhomogeneous volume conductor by internal current sources. IEEE Trans. Magn. 1970, 6, 346–347. [Google Scholar] [CrossRef] [Scilit]
  26. Sarvas, J. Basic mathematical and electromagnetic concepts of the biomagnetic inverse problem. Phys. Med. Biol. 1987, 32, 11–22. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Egbuji, J.U.; O’Grady, G.; Du, P.; Cheng, L.K.; Lammers, W.J.; Windsor, J.A.; Pullan, A.J. Origin, propagation and regional characteristics of porcine gastric slow wave activity determined by high-resolution mapping. Neurogastroenterol. Motil. 2010, 22, e292–300. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Corrias, A.; Buist, M.L. Quantitative cellular description of gastric slow wave activity. Am. J. Physiol. Gastrointest. Liver Physiol. 2008, 294, G989–G995. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Han, H.; Cheng, L.K.; Avci, R.; Paskaranandavadivel, N. Quantification of Gastric Slow Wave Velocity Using Bipolar High-Resolution Recordings. IEEE Trans. Biomed. Eng. 2022, 69, 1063–1071. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Pullan, A.; Cheng, L.; Yassi, R.; Buist, M. Modelling gastrointestinal bioelectric activity. Prog. Biophys. Mol. Biol. 2004, 85, 523–550. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Foong, D.; Calder, S.; Varghese, C.; Schamberg, G.; Xu, W.; Daker, C.; Ho, V.; Andrews, C.N.; Gharibans, A.A.; O’Grady, G. Gastric Alimetry® Test Interpretation in Gastroduodenal Disorders: Review and Recommendations. J. Clin. Med. 2023, 12, 6436. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  32. Eichler, C.E.; Cheng, L.K.; Paskaranandavadivel, N.; Du, P.; Bradshaw, L.A.; Avci, R. Effects of magnetogastrography sensor configurations in tracking slow wave propagation. Comput. Biol. Med. 2021, 129, 104169. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  33. Sun, C.; Liang, Y.; Yang, X.; Zhao, B.; Zhang, P.; Liu, S.; Yang, D.; Wu, T.; Zhang, J.; Guo, H. Optimizing biomagnetic sensor performance through in silico diagnostics: A novel approach with BEST (Biomagnetism Evaluation via Simulated Testing). Iscience 2024, 27, 110167. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  34. Graham, A.J.; Orini, M.; Zacur, E.; Dhillon, G.; Daw, H.; Srinivasan, N.T.; Lane, J.D.; Cambridge, A.; Garcia, J.; O’Reilly, N.J.; et al. Simultaneous Comparison of Electrocardiographic Imaging and Epicardial Contact Mapping in Structural Heart Disease. Circ. Arrhythm. Electrophysiol. 2019, 12, e007120. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  35. Roudijk, R.W.; Boonstra, M.J.; Brummel, R.; Kassenberg, W.; Blom, L.J.; Oostendorp, T.F.; Te Riele, A.S.J.M.; van der Heijden, J.F.; Asselbergs, F.W.; van Dam, P.M.; et al. Comparing Non-invasive Inverse Electrocardiography With Invasive Endocardial and Epicardial Electroanatomical Mapping During Sinus Rhythm. Front. Physiol. 2021, 12, 730736. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. Wang, J.; Yang, K.; Yang, R.; Kong, X.; Chen, W. SQUID Gradiometer Module for Fetal Magnetocardiography Measurements Inside a Thin Magnetically Shielded Room. IEEE Trans. Appl. Supercond. 2019, 29, 1600204. [Google Scholar] [CrossRef] [Scilit]
  37. Yu, X.; Jia, R.; Feng, X.; Hao, X.; Sun, C. Metric-anchored noise translation for acquisition-aware biomagnetism simulations. Measurement 2026, 286, 122359. [Google Scholar] [CrossRef] [Scilit]
  38. Belouchrani, A.; Abed-Meraim, K.; Cardoso, J.F.; Moulines, E. A blind source separation technique using second-order statistics. IEEE Trans. Signal Process. 1997, 45, 434–444. [Google Scholar] [CrossRef] [Scilit]
  39. Erickson, J.C.; Obioha, C.; Goodale, A.; Bradshaw, L.A.; Richards, W.O. Detection of small bowel slow-wave frequencies from noninvasive biomagnetic measurements. IEEE Trans. Biomed. Eng. 2009, 56, 2181–2189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Hansen, P.C. Analysis of discrete ill-posed problems by means of the L-curve. SIAM Rev. 1992, 34, 561–580. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Molero, R.; Martínez-Pérez, M.; Herrero-Martín, C.; Reventós-Presmanes, J.; Roca-Luque, I.; Mont, L.; Climent, A.M.; Guillem, M.S. Improving electrocardiographic imaging solutions: A comprehensive study on regularization parameter selection in L-curve optimization in the atria. Comput. Biol. Med. 2024, 182, 109141. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Golub, G.H.; Heath, M.; Wahba, G. Generalized cross-validation as a method for choosing a good ridge parameter. Technometrics 1979, 21, 215–223. [Google Scholar] [CrossRef]
  43. O’Grady, G.; Egbuji, J.U.; Du, P.; Lammers, W.J.E.P.; Cheng, L.K.; Windsor, J.A.; Pullan, A.J. High-resolution spatial analysis of slow wave initiation and conduction in porcine gastric dysrhythmia. Neurogastroenterol. Motil. 2011, 23, e345–e355. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Gopirajah, R.; Anandharamakrishnan, C. Advancement of Imaging and Modeling Techniques for Understanding Gastric Physical Forces on Food. Food Eng. Rev. 2016, 8, 323–335. [Google Scholar] [CrossRef] [Scilit]
  45. Liao, D.; Gregersen, H.; Hausken, T.; Gilja, O.H.; Mundt, M.; Kassab, G. Analysis of surface geometry of the human stomach using real-time 3-D ultrasonography in vivo. Neurogastroenterol. Motil. 2004, 16, 315–324. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  46. Steinsvik, E.K.; Hatlebakk, J.G.; Hausken, T.; Nylund, K.; Gilja, O.H. Ultrasound imaging for assessing functions of the GI tract. Physiol. Meas. 2021, 42, 024002. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. Queirós, S.; Vilaça, J.L.; Morais, P.; Fonseca, J.C.; D’hooge, J.; Barbosa, D. Fast left ventricle tracking using localized anatomical affine optical flow. Int. J. Numer. Method Biomed. Eng. 2017, 33, e2871. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Cheng, L.K.; Sands, G.B.; French, R.; Withy, S.; Wong, S.; Legget, M.; Smith, W.; Pullan, A.J. Rapid construction of a patient-specific torso model from 3D ultrasound for non-invasive imaging of cardiac electrophysiology. Med. Biol. Eng. Comput. 2005, 43, 325–330. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Liang, S.; Gao, K.; He, J.; Jia, Y.; Jiao, H.; Feng, L. System and approach to detecting of gastric slow wave and environmental noise suppression based on optically pumped magnetometer. Biocybern. Biomed. Eng. 2024, 44, 1–8. [Google Scholar] [CrossRef] [Scilit]
  50. Zhang, Y.; He, X.; Li, H.; Zhang, L.; Yan, B.; Peng, Y.; Ge, Q.; Ding, M.; Leng, Y. A potential approach to detecting of gastrointestinal slow waves based on optically pumped magnetometers array. Biocybern. Biomed. Eng. 2025, 45, 549–559. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.