Abstract

Interest in local warming for chronic wound care has re-emerged as flexible heaters, embedded sensors, and smart dressings make controlled heat delivery increasingly feasible. This narrative translational review treats thermotherapy as a dose-dependent wound-interface intervention rather than a single warming technology. It distinguishes normothermic radiant warming, water-filtered infrared A, resistive Joule heating, photothermal systems, and temperature-monitoring technologies by mechanism, thermal context, therapeutic purpose, and evidence level. The review integrates wound pathophysiology, biological responses to temperature, thermal-dose concepts, emerging heater platforms, sensing and closed-loop control, clinical evidence, safety, and regulatory translation. Direct human treatment evidence remains concentrated in small historical radiant-warming and water-filtered infrared A studies, with limited additional local-heat evidence, and cannot be transferred directly to contemporary flexible, wireless, or photothermal platforms. For lower-temperature warming, heater setpoint alone is an inadequate dose descriptor, and the biological meaning of cumulative equivalent minutes at 43 °C (CEM43) has not been validated; measured time–temperature profiles, peak temperature, spatial variation, sensor location, and uncertainty are more informative. Clinical translation therefore depends on reproducible wound-interface dosing under realistic moisture, pressure, perfusion, dressing, motion, and workflow conditions, together with wound-specific dose models, transparent reporting, realistic interface validation, and architecture-specific regulatory evidence.

1. Introduction

Chronic wounds remain a substantial clinical and economic problem. Impaired perfusion, infection, and dysregulated matrix remodeling can slow repair or stop it altogether. Their prevalence is increasing with population aging and with the growing burden of diabetes, vascular disease, and immobility [1]. For patients and health systems, the consequences are considerable. Chronic wounds often require repeated clinic visits, hospital admissions, prolonged treatment, and substantial expenditure [2]. Current care is built around debridement, infection control, moisture balance, offloading, and correction of vascular disease where possible [3].
Nevertheless, many ulcers fail to heal even when these measures are optimized. This persistent gap points to an unmet need for adjunctive strategies that address microvascular and cellular dysfunction [4]. The biological rationale for controlled local warming is grounded in several processes involved in repair. Higher cutaneous temperature can increase microvascular blood flow [5] and tissue oxygen tension [6], while thermal stress can increase heat shock protein 70 (HSP70) and inflammatory markers in burn-blister fluid [7]. These effects make thermotherapy a plausible adjunct for selected chronic wounds. The clinical evidence, however, is still uneven. In high-risk settings, particularly ischemic, neuropathic, or infected tissue, safe and effective thermal dosing remains poorly defined. Recent advances in flexible materials and embedded electronics have changed what is technically possible for local heat delivery. Flexible liquid-metal heaters [8], graphene-based conductors [9], and hydrogel-integrated warming systems may provide more conformal and spatially uniform heat transfer than conventional rigid devices [10]. Temperature sensors combined with feedback electronics can adjust power output when contact pressure or ambient conditions change [11]. These features are relevant to chronic wounds because dressings must fit irregular surfaces, tolerate moisture and exudate, and sometimes function under intermittent compression. Early clinical and preclinical studies suggest that local warming may improve perfusion and some healing-related outcomes in chronic wounds, particularly venous leg ulcers, pressure injuries, and related nonhealing wounds [12,13,14]. These observations should be interpreted cautiously. Most studies are small, use diverse patient groups and treatment schedules, and report limited information on achieved interface temperature, temperature stability, spatial heat distribution, or cumulative thermal dose. This makes comparison across studies difficult and prevents confident definition of indications, contraindications, or safe dose ranges. Translation also requires regulatory and standards pathways that can accommodate multifunctional, digitally connected dressings. Existing regulations for powered heating pads [15] and standards for medical heating devices, including blankets, pads, and mattresses [16], mainly address short-term warming of intact skin. They do not fully capture the behavior of flexible, sensor-guided systems used for prolonged periods in moist and potentially contaminated wound environments. Developers therefore need evidence of long-term stability, moisture resistance, and reproducible thermal performance under realistic mechanical loading and microclimatic conditions [17]. Many prototypes are still tested on simplified bench models. Such models do not reproduce the dynamic wound interface, leaving an important gap between engineering performance claims and expected clinical behavior.

1.1. Search Strategy, Eligibility, and Evidence Appraisal

This narrative translational review used iterative, topic-specific literature searches from December 2025 through August 2026 across PubMed, Embase, Web of Science Core Collection, and IEEE Xplore. Searches were updated as the review themes were refined, and the evidence base was finalized with a literature cutoff of 31 August 2026. Supplementary Table S1 provides consolidated database-specific search expressions that operationalize the documented search concepts and support reproducible rerunning of the search framework.
The search concepts covered chronic wounds, diabetic foot ulcers, venous leg ulcers, pressure injuries, local warming or thermotherapy, noncontact normothermic wound therapy, thermal dose or CEM43, flexible heaters, smart dressings, photothermal systems, sensing and feedback control, and medical-device regulation. Searches were supplemented by backward citation tracking and by standards, guidance documents, and the device-oriented translational literature when directly relevant to safety, validation, or regulatory interpretation. Reproducibility is supported through the consolidated database-specific expressions in Supplementary Table S1, while the focused clinical selection audit is reported at candidate-report level in Table 1 and Supplementary Figure S1.
Table 1. Focused clinical-evidence classification and report-specific disposition.
For the focused clinical-evidence synthesis, reports were eligible for the core direct-treatment category when they involved human participants with an established chronic wound, evaluated a defined local thermal intervention intended as part of treatment, and reported wound-healing or directly wound-related clinical or physiological outcomes. Studies of intrinsically multimodal thermal technologies, such as water-filtered infrared A (wIRA), were eligible when the thermal intervention was a defined component of treatment, with outcomes interpreted according to the combined modality when thermal and nonthermal effects were inseparable. Studies combining local heating with an independent active co-intervention were classified as supportive/contextual evidence when the contribution of heating could not be evaluated separately. Prevention-only perioperative warming, monitoring-only studies, procedural warming, acute/postoperative wounds outside the chronic-wound treatment scope, and preclinical or bench studies were classified outside the core direct-treatment category. Clinically relevant reports were retained as supportive or contextual evidence where appropriate.
The screening workflow used first-author primary literature searching, initial screening, and eligibility assessment, followed by verification of the final clinical-study selection and classification decisions by two senior coauthors. Points requiring clarification were resolved through discussion. This workflow provided a documented primary assessment with senior coauthor verification appropriate to the narrative translational review design.
Twenty-five unique human reports entered the focused report-level clinical verification. Sixteen met the criteria for the core direct-treatment synthesis, seven were retained as supportive/contextual clinical evidence, one prevention-only report was retained as contextual evidence, and one report was classified outside thermotherapy evidence because it did not evaluate a therapeutic thermal intervention. These four mutually exclusive dispositions account for all 25 candidate reports. The counts represent the focused report-level clinical evidence set. Table 1 gives the disposition and study-specific rationale for each report, and Supplementary Figure S1 summarizes the focused selection flow.
The retained clinical reports were appraised at the study level for assignment method, unit of analysis, baseline comparability, attrition, co-interventions, temperature or dose measurement, adverse-event reporting, and outcome completeness. The evidence tables report only information directly supported by the accessible publication record, with unavailable items identified explicitly. Given the substantial heterogeneity in wound type, intervention modality, treatment schedule, outcome definition, and thermal-dose reporting, the synthesis was qualitative rather than meta-analytic. Generative AI assistance during manuscript preparation was used for language refinement, structural organization, reference-formatting assistance, internal consistency checks, and development of conceptual figure drafts under author guidance; all AI-assisted outputs were reviewed and edited by the authors before incorporation into the manuscript.

1.2. Review Scope and Translational Framework

The previous literature has addressed different components of thermotherapy translation largely in separate streams. Reviews of local warming have focused mainly on clinical efficacy and the limited chronic-wound treatment evidence [14,41]. The hyperthermia literature provides established temperature–time concepts such as CEM43, but these frameworks were developed principally for oncologic hyperthermia and their biological meaning for lower-temperature chronic-wound warming remains unvalidated [42,43]. In parallel, the recent smart-bandage and multifunctional-dressing literature has emphasized flexible materials, sensing, feedback control, wireless operation, and multimodal therapy [44,45,46,47], but these engineering capabilities do not by themselves establish a clinically justified wound-interface thermal dose. The present review brings these strands together through five linked questions: whether a thermal exposure can be defined, delivered, measured, controlled, and clinically justified across wound biology, material platform, interface metrology, clinical evidence quality, and regulatory verification. It also separates thermal modalities that differ in energy delivery and biological co-effects, distinguishes direct human treatment evidence from preclinical and monitoring analogies, and proposes wound-interface validation and reporting requirements. The purpose is translational synthesis rather than a pooled estimate of treatment effectiveness.

1.3. Chronic Wound Types and Pathophysiologic Barriers to Healing

Chronic wounds are clinically heterogeneous. The principal groups relevant to this review are venous leg ulcers, pressure injuries, diabetes-related foot ulcers, and arterial or mixed-etiology ulcers [1,4,14]. Their failure to heal arises from different initiating causes but commonly involves impaired perfusion or oxygen delivery, persistent inflammation, repeated mechanical loading or venous hypertension, cellular dysfunction, protease-dominant matrix degradation, and susceptibility to infection [4,48,49]. Diabetes-related foot ulcers additionally combine neuropathy, repetitive pressure, and variable ischemia; pressure injuries are driven by sustained or repetitive load and tissue deformation; venous ulcers occur in a milieu of venous hypertension and edema; and arterial or mixed ulcers may have limited capacity to increase perfusion when local metabolic demand rises. These distinctions are central to thermotherapy because the same external heat input can produce different tissue temperatures, heat-dissipation rates, and safety margins across wound types.
Refractory wounds should therefore not be treated as a uniform “cold wound” phenotype. Hypoxia can coexist with inflammation, edema, infection, neuropathy, or impaired vascular reserve, while surface temperature is also influenced by perfusion, ambient exposure, dressings, and inflammatory activity. Any thermotherapeutic rationale must remain wound-specific and adjunctive to etiologic care such as compression, offloading, vascular assessment, debridement, infection management, and moisture balance [4,50].

2. Biological Rationale and Biophysical Constraints of Thermotherapy in Chronic Wounds

2.1. Terminology, Temperature Ranges, and Temporal Context

Terminology in this field is inconsistent, so this review uses operational definitions rather than implying universally validated chronic-wound thresholds. “Normothermic wound warming” refers to restoring or maintaining the wound or periwound region near physiological tissue temperature; for synthesis in this review, approximately 36–38 °C is used as the normothermic/supportive-warming context, consistent with historical NNWT systems that commonly used a dressing target near 38 °C [24,26]. “Mild thermotherapy” is used here as the broader umbrella term for controlled local warming approximately 36–40 °C when the intended purpose is physiological support. For consistency within this review, “mild hyperthermia” refers operationally to the higher local-warming context of approximately 39–42 °C; broader hyperthermia literature may use wider ranges, including approximately 39–45 °C [51]. These temperature ranges overlap by design and are descriptive contexts rather than validated chronic-wound therapeutic windows. “Photothermal therapy” is defined by its mechanism, namely conversion of optical energy into heat, rather than by a fixed temperature range; the achieved temperature depends on the material, irradiance, exposure duration, and treatment purpose, and photothermal systems may also include nonthermal mechanisms such as photodynamic effects [52,53].
These temperature contexts are not biologically equivalent. Near-normothermic warming (~36–38 °C) primarily supports restoration of physiological tissue conditions and the rationale for perfusion and oxygen delivery, whereas higher local warming (~39–42 °C) can produce more pronounced vasodilation and engage temperature- and time-dependent heat-shock and immunomodulatory responses. With further temperature elevation or prolonged exposure, thermal stress and tissue-injury mechanisms become increasingly important. These ranges should therefore be interpreted as overlapping exposure contexts rather than fixed thresholds for angiogenesis, re-epithelialization, or matrix remodeling.
Temperature also changes during healing. In uncomplicated acute or surgical wounds, inflammatory activity is generally associated with an early temperature elevation, followed by progressive normalization as inflammation resolves and repair proceeds; a secondary rise may accompany infection [54]. Chronic wounds may not follow this trajectory because persistent inflammation, infection, ischemia, edema, neuropathy, and impaired perfusion can produce sustained, reduced, or spatially heterogeneous temperatures. Accordingly, healing phase should not be inferred from an absolute temperature alone; longitudinal wound and periwound measurements are more informative [54,55].

2.2. Vascular and Metabolic Responses to Warming

Perfusion is essential for oxygen and nutrient delivery, leukocyte trafficking, and matrix deposition during wound repair. Local warming causes cutaneous vasodilation through neurogenic and nitric oxide-mediated mechanisms. Increased blood flow may raise tissue oxygen tension, which is particularly relevant in chronic wounds where limited oxygen availability impairs collagen synthesis and fibroblast activity. Studies of localized warming in chronic wounds have reported improved periwound perfusion and oxygenation, with greater granulation in some settings [56]. Most of these studies, however, assess only superficial periwound tissue, and whether warming meaningfully affects deeper hypoxic wound regions remains uncertain. Vascular effects are closely linked to temperature-dependent changes in cellular metabolism and oxygen use. Mild warming can increase mitochondrial activity and accelerate biochemical processes involved in cell proliferation and matrix production, while cutaneous warming may further support tissue metabolism through nitric oxide release and endothelial activity [57]. These responses may be relevant in diabetic wounds, where fibroblast and immune-cell dysfunction can impair repair [58]. Their clinical significance remains uncertain because the available wound studies are limited and heterogeneous [41,56]. A 2025 protocol proposed a systematic review of vasomotor and sudomotor responses to increased core temperature in type 2 diabetes; because it is a protocol, it does not itself provide outcome evidence [59].

2.3. Cellular Stress, Immune Modulation, and Extracellular Matrix Dynamics

Heat also engages cellular stress-response pathways and can influence innate immune activity. Thermal stimulation can induce heat shock proteins, including HSP70, which stabilize proteins, limit aggregation, and support cell survival under stress. Increased HSP70 expression has been observed in thermally stressed skin, including burn-blister fluid and heat-shocked endothelial cells [7,60]. Post-translational modifications also participate in cellular heat-stress responses, although this evidence is not specific to chronic wounds [61]. These chaperone responses may help protect cells exposed to oxidative stress, inflammation, or mechanical strain in chronic wounds. Even so, the timing of heat-shock induction in chronic wound tissue is not well established, and the temperature–time combinations that provide cytoprotection without protein denaturation remain unclear [56]. Immune dysregulation is a defining feature of chronic wounds. Persistent pro-inflammatory signaling can keep macrophages in inflammatory phenotypes and delay the transition to reparative programs needed for resolution and tissue remodeling [62]. In experimental models, controlled mild warming has shifted macrophage polarization toward reparative, M2-like activity and improved wound repair. Some thermally delivered therapies have increased M2 markers and accelerated closure in diabetic wounds [63]. Thermal exposure can also alter innate immune signaling in a temperature-dependent manner, including changes in Toll-like receptor 4 (TLR4) expression and signaling under fever-range hyperthermia [64]. These changes could help move wounds away from persistent inflammation and toward healing. Still, immune responses to warming are likely to differ by tissue type, infection status, and comorbid disease. Findings from isolated cells or acute models may not reflect the inflammatory complexity of chronic wounds. Matrix remodeling is required for wound closure and depends on coordinated extracellular-matrix turnover by proteases and their endogenous inhibitors. Chronic wounds often have protease-dominant microenvironments, in which elevated matrix metalloproteinase (MMP) activity sustains matrix breakdown and delays healing [48]. Temperature can influence matrix-degrading pathways. Experimental skin studies show that repeated heat exposure can increase several MMPs and promote collagen and extracellular-matrix damage [65]. Because excessive proteolysis is already common in chronic wounds, poorly controlled thermotherapy could worsen destructive matrix degradation. The effect will likely depend on the baseline balance between proteases and their inhibitors. Future studies should therefore report a simple biomarker context, such as the ratio of MMPs to tissue inhibitors of metalloproteinases (TIMPs), together with clinical outcomes [48].

2.4. Heat Transfer at the Wound Interface and Implications for Local Temperature Control

The therapeutic effect of thermotherapy depends on heat transfer from the device, across the wound surface, and into deeper tissue. Tissue water content, local blood flow, and the thermal properties of dressings together determine how fast and how deeply tissue warms [66]. Moist granulation tissue under occlusive or hydrogel dressings conducts heat more efficiently than dry or scaly skin. Exudative areas may therefore reach higher temperatures than adjacent regions. Hydrogel dressings used at pressure-prone interfaces also alter contact and local moisture conditions [67]. Many bioheat models used for cutaneous heating originate from intact or homogeneous perfused-tissue assumptions [68]. Chronic wounds rarely satisfy these assumptions. Hydration is uneven, perfusion varies, deeper structures may be exposed, and irregular contours can produce complex temperature gradients over space and time. Clinical thermography of pressure ulcers further demonstrates spatial heterogeneity in wound and periwound temperatures [69]. Current models therefore provide only approximate guidance for chronic-wound thermotherapy and require wound-specific adaptation and validation. These mechanistic and interface uncertainties explain the need for wound-specific dose definitions and measurement requirements, as discussed in Section 3.

3. Thermal Dose Concepts and Biophysical Models for Chronic Wounds

3.1. Thermal Dose Frameworks: Foundations and Limitations

Thermal dose describes the combined effect of temperature and exposure time on tissue response. CEM43 was developed in oncology hyperthermia to normalize variable time–temperature histories to equivalent minutes at 43 °C [42,43]. In the conventional discrete formulation:
C E M 43 = ∑ i Δ t i   R ( 43 − T i )
where Δti is the duration of interval i in minutes, Ti is the measured temperature (°C) at the specified measurement site during that interval, and R is the empirical time–temperature factor, conventionally 0.5 for Ti > 43 °C and 0.25 for Ti < 43 °C [42,43,51]. At Ti = 43 °C, the exponent is zero, so each minute contributes one equivalent minute. The discrete calculation therefore assumes that Ti adequately represents the measured temperature over each interval and that interval contributions are additive after conversion to the 43 °C reference. It requires a measured temperature–time history from a stated measurement location; a nominal heater setpoint should not be substituted for Ti. The metric is rooted in heat-induced cytotoxicity and does not explicitly represent perfusion-mediated oxygenation or other lower-temperature physiological effects [51]. This limitation matters for chronic-wound interventions that may target near-normothermic temperatures rather than cytotoxic hyperthermia; historical NNWT studies, for example, used a dressing target of 38 °C [24,26]. A CEM43 value can be calculated mathematically at such temperatures, but its biological interpretation as a wound-healing dose has not been validated. Accordingly, CEM43 should not be used as the sole dose descriptor for normothermic wound warming.
Any dose estimate must specify measurement location because heater setpoint, heater-surface temperature, dressing-interface temperature, wound-bed surface temperature, periwound skin temperature, and deeper tissue temperature describe different thermal quantities. Where CEM43 is reported, the underlying temperature–time record and sensor location should accompany the value. For lower-temperature wound warming, the primary dose description should be based on the measured thermal exposure itself: complete time–temperature profile, peak temperature, exposure duration, spatial range or gradient, measurement location, sensor accuracy or uncertainty, and relevant dressing and perfusion conditions. CEM43 can be reported secondarily as a hyperthermia-derived normalization descriptor while wound-specific biological validation is established [43]. Figure 1 summarizes these wound-interface heat-transfer and measurement considerations.
Figure 1. Wound-interface heat transfer: (A) surface heating; (B) idealized versus variable interface temperatures; (C) perfusion- and moisture-mediated heat removal.
Table 2 groups thermal-dose concepts and practical dosing approaches relevant to local heating. Models developed for oncology hyperthermia, intact skin, and conventional surface warming provide useful terminology and safety concepts, with wound-specific validation required for application to chronic-wound protocols [66,70]. In this context, CEM43 is used as a hyperthermia-derived normalization metric, while lower-temperature wound-healing dose is characterized directly from measured thermal exposure.
Table 2. Thermal dose frameworks and practical dosing approaches relevant to chronic wound thermotherapy.
In wound tissue, temperature influences linked vascular, metabolic, immune, and matrix pathways, but the evidence does not support a single universal therapeutic temperature. In healthy skin, local warming above approximately 37 °C produces progressively stronger cutaneous vasodilation, with a substantial nitric-oxide-dependent component near the higher end of commonly studied local-heating ranges [5,57]. At higher temperatures or longer exposures, physiological stimulation increasingly overlaps with cellular stress and thermal-injury mechanisms [73,74]. Table 3 therefore uses operational temperature contexts rather than fixed chronic-wound thresholds.
Table 3. Operational temperature contexts and reported physiological or biological responses relevant to wound healing.

3.2. Determinants of Delivered Thermal Dose at the Wound Interface

At the wound interface, delivered dose is governed by surface geometry, hydration, dressing layers, contact pressure, and local perfusion. Moist or exudative tissue may conduct heat more readily than dry or callused margins. Movement and dressing compression can also change contact during treatment [76,77]. Flat, dry bench substrates therefore offer limited assurance of clinical performance [78]. Because hydrogel, occlusive, absorbent, and foam dressings differ in heat retention and moisture handling, each device-dressing combination should be tested under standardized wet, compressed, and clinically realistic conditions [79,80,81]. Heat penetration reflects a balance between conduction through the wound bed and convective heat removal by circulating blood. Conductive transfer depends mainly on tissue composition and water content, whereas convective transfer depends on local perfusion [82]. When perfusion rises during warming, convective cooling may limit further temperature rise in deeper layers and help stabilize oxygen tension. Studies of localized warming in vascular-compromised tissue support this feedback behavior and show that changes in blood flow affect local oxygen tension and temperature stability. In ischemic tissue, reduced perfusion weakens this protective effect. A surface temperature that appears acceptable may therefore create greater local overheating risk. Because most biophysical models do not incorporate these perfusion-dependent differences, their predictions may be unreliable in patients with significant vascular disease [70,72,82]. Reliable real-time temperature sensing is indispensable for safe dosing. Thin-film thermistors embedded in flexible heaters can give local measurements with good sensitivity and rapid response, but repeated bending, moisture ingress, and aging of encapsulation materials may cause drift [83,84]. In a closed-loop system, even a small sensor error can lead to unrecognized overheating or under-treatment because the controller responds to the reported temperature rather than the true tissue value. Few thermotherapy prototypes report calibration data, error margins, or response latency under realistic humidity, pressure, and movement. As a result, it is difficult to judge the reliability of temperature control or safety limits [85].

3.3. Implications for Wound-Specific Dosing and Device Design

Overall, the biological evidence provides a strong rationale for controlled, wound-specific investigation. Mild warming may improve perfusion, oxygenation, inflammatory signaling, heat-shock responses, and matrix remodeling, with supporting evidence spanning intact skin, acute models, and heterogeneous wound studies [41,86]. For device design and clinical protocols, the central requirement is to define the therapeutic window for each wound context. Future dosing studies should report achieved tissue-facing temperature, exposure duration, dressing configuration, moisture state, perfusion status, and clinically meaningful outcomes, including granulation, re-epithelialization, closure, pain, and tissue injury [41,43,51]. Figure 2 marks the intended transition from mild warming that may support repair to higher exposures associated with thermal stress and highlights the importance of integrating device setpoint with tissue risk, moisture, perfusion, and control reliability when defining safety margins [73,74].
Figure 2. Mechanistic pathways linking controlled local warming to chronic-wound healing and thermal risk. The 36–40 °C label denotes the operational mild-thermotherapy range used in this review, not a validated universal chronic-wound therapeutic window. Responses depend on exposure duration, modality, tissue perfusion, and measurement location. Arrows indicate conceptual relationships among thermal exposure, biological responses, and wound outcomes. The green-to-orange gradient indicates a qualitative transition from therapeutic to excessive thermal exposure and does not represent a quantitative scale. Abbreviation: HSP, heat shock protein.
For chronic wounds, thermotherapy therefore requires dose limits measured at the wound interface rather than inferred from nominal device settings. The following sections consider whether current device architectures, sensors, and control strategies can deliver such doses under realistic clinical conditions.

4. Device Architectures and Smart Thermotherapeutic Dressings

4.1. Heat-Delivery Principles and Device Design Space

Temperature-related wound technologies should be classified by mechanism and intended role before evidence is compared. Table 4 distinguishes direct therapeutic warming from multimodal photothermal systems and monitoring-only technologies, and separates direct human chronic-wound treatment evidence from engineering, preclinical, monitoring, and multimodal evidence. These distinctions reduce the risk of findings from one modality being generalized to technologies that differ in energy delivery, temperature context, therapeutic purpose, or biological co-interventions.
Table 4. Classification of temperature-related wound technologies by mechanism, intended role, control, and evidence context.
Figure 3 presents a qualitative engineering design map rather than an evidence ranking. The horizontal axis represents increasing conformability or substrate design flexibility, while the vertical axis represents increasing control complexity from open-loop operation toward sensor-integrated, adaptive, and connected control. Device positions are conceptual and indicate typical design tendencies rather than measured performance values [52,96,110].
Figure 3. Qualitative engineering design space for thermotherapeutic devices. The horizontal axis represents increasing conformability or substrate flexibility, and the vertical axis increasing control complexity from open-loop to adaptive or connected control. Symbol positions are conceptual and do not indicate quantitative performance, clinical maturity, or comparative efficacy.

4.2. Heater Platforms and Emerging Materials Systems

Heater platforms should be judged by the reliability of interface dosing rather than by heat generation alone. Conventional powered heating pads are familiar, simple, and already covered by established safety frameworks [15]. Their limited conformability can lead to uneven contact over irregular anatomy or layered dressings. Burns from body-heating devices have been documented, reinforcing the need for careful placement and monitoring [111]. Most conventional pads use thermostatic control or cut-offs rather than distributed interface sensing; advanced wearable systems can integrate real-time sensing and control [112]. FDA labeling guidance specifically warns against use over insensitive skin or in the presence of poor circulation [113]. Because existing standards mainly address dry blankets, pads, and mattresses, direct evidence for conventional heating systems over moist wound-dressing interfaces remains limited [16]. Liquid-metal microheaters use gallium-based conductive alloys, such as eutectic gallium-indium (EGaIn), embedded in soft elastomers through microchannels or patterned traces to produce deformable Joule heating [87]. These designs can maintain electrical continuity during bending and stretching while improving conformal contact. Better contact reduces interfacial air gaps that would otherwise impair heat transfer and may support more uniform surface heating than rigid traces on soft substrates [88]. Bench and in vivo demonstrations have shown rapid thermal ramping and relatively stable temperature fields under moderate strain, which is consistent with possible use on irregular skin and tissue surfaces. Leakage, channel fracture, and interface degradation remain important failure modes. Long-term stability during repeated handling and wet, variable-contact use is still insufficiently characterized [8,89,114]. Graphene and metallic thin-film heaters can provide thin, rapidly responsive, conformal heating layers for dressings or textiles, and they can be paired with sensing or wireless modules [90,91,115,116]. Their main translational risks include fatigue cracking, delamination, defect-related hot spots, and performance drift in moist, protein-rich wound environments. Across flexible-heater platforms, evidence remains limited for electrical safety, durability, and reproducible dosing under wound-like moisture, shear, and dressing-change conditions [92,93,94]. Hydrogel-integrated heaters provide close contact, high flexibility, and hydrated heat transfer across uneven wound surfaces. Hydrogels can also support drug-delivery functions, creating a route to combined thermal and pharmacologic treatment [97]. Swelling, drying, leachables, sterilization, and repeated handling can nevertheless change electrical output and mechanical integrity. These material changes should be treated as dose-control risks, not simply as fabrication problems [92,97]. Transient and bioresorbable platforms offer a route to short-duration devices that function during early repair and then resorb, avoiding device retrieval. An integrated flexible heater has been demonstrated in a biodegradable nanofibrous platform for on-demand drug delivery, although not as a clinically validated wound-warming system [98]. This approach is appealing for postoperative wounds and deep or difficult-to-access cavities, where removal may add risk. Most prototypes, however, still depend mainly on passive thermal behavior, including device geometry and intrinsic resistive properties, rather than closed-loop temperature sensing and feedback. During hydrolysis or corrosion, materials and interfaces change. Stiffness, electrical continuity, and contact conditions may shift, moving the spatial temperature profile away from the intended dose. The local and systemic tolerance of degradation by-products in inflamed, ischemic, or infection-prone tissue also remains incompletely characterized for many material classes. Regulatory translation must therefore justify both device performance over time and biological safety throughout degradation and resorption [99,100,101]. Photothermal wound dressings typically incorporate light-absorbing nanomaterials, such as Ti3C2Tx MXene, into hydrogel matrices so that external irradiation is converted into localized heat at the wound interface. MXene-enabled designs can heat rapidly and with spatial confinement. They are often combined with antimicrobial mechanisms, including photothermal killing, heat-triggered drug release, and MXene-associated antibacterial activity. In infected-wound animal models, the combination of bacterial suppression and thermal support has improved wound closure [102,103]. Recent preclinical platforms include a thermal-feedback hydrogel with self-limiting photothermal and thermoelectric functions, a long-acting mild-hyperthermia microneedle patch, and an alginate dressing with controlled heat release [117,118,119]. These systems broaden the engineering design space but do not constitute clinical efficacy evidence. Despite these preclinical results, nanomaterial safety remains a major translational concern. Depending on composition, surface terminations, oxidation state, and dose, MXenes and related two-dimensional particles may induce reactive-species-mediated stress responses, interact with blood and immune cells, or transform in ways that make in vivo behavior difficult to predict. Long-term retention and clearance are still incompletely defined for many MXene formulations. In chronic wounds, impaired vascular or lymphatic drainage may prolong residence time and alter local immune reactions. These uncertainties limit clinical viability even when laboratory and small-animal results are favorable [120,121].
Photothermal dressings are best interpreted as multimodal systems rather than as direct equivalents of normothermic warming. Combined photothermal-photodynamic therapy (PTT/PDT) couples light-to-heat conversion with photosensitizer-mediated reactive oxygen species generation, providing complementary antibacterial mechanisms relevant to infected wounds. Direct preclinical wound evidence includes a multifunctional bilayer hydrogel in an MRSA-infected rat wound model, in which PTT/PDT provided antibacterial control together with staged bioactive release supporting subsequent repair [122]. In such platforms, healing reflects the combined contribution of thermal, photodynamic, drug, material, and wound-microenvironment effects. Multifunctional antibacterial nanoplatforms can further combine PTT/PDT with chemodynamic effects [123], while photocatalytic materials such as graphitic carbon nitride provide ROS-mediated antibacterial activity and have been reviewed for biological applications including wound healing [124]. Other wound-oriented photothermal systems combine heating with drug release or material-mediated antibacterial activity [52,53,102,103]. Irradiance, exposure time, measured temperature, reactive-oxygen mechanisms, and co-interventions should therefore be reported as distinct components of the treatment description.

4.3. Smart Dressings, Sensing, and Closed-Loop Control

Smart dressings add sensing, feedback control, and workflow integration in an effort to narrow the gap between setpoint and delivered dose. Wireless hydrogel-based bioelectronic systems can combine microheaters, temperature sensors, and therapeutic reservoirs within conformal dressings, allowing programmable dosing and remote adjustment without disturbing the dressing [44,108,125]. The same integration introduces risks related to power management, packaging, cybersecurity, sensor-heater coupling, cost, and fault handling, especially during long wear in moist and mechanically variable wounds. Large-animal bioelectronic wound platforms provide a useful reference point for thermotherapy translation, particularly for packaging durability, wireless operation, and multi-day function in wet and mechanically dynamic wound environments. Baniya et al., for example, reported a wireless wound-healing actuator tested in a porcine model under conditions closer to clinical use. Although that system delivered electrical and drug-based therapy rather than heat, it illustrates the level of sealing, robustness, and workflow compatibility that future thermotherapeutic dressings will need to demonstrate [109].
Temperature sensing is the safety-critical component in closed-loop thermotherapy. Thin-film thermistors or resistance temperature detectors (RTDs) placed near the interface can provide fast local measurements, but bending, pressure variation, moisture ingress, and aging of the encapsulation can introduce calibration drift or signal artifacts [83,84,85]. Noncontact infrared thermography avoids direct contact, but exudate and dressing materials can alter emissivity and measurement conditions [104]. Wound-bed thermal imaging may nevertheless contribute to infection assessment [105]. Few prototypes quantify sensor accuracy, latency, and drift under the combined humidity, pressure, and movement conditions expected in chronic wound care. Integrated dressings should report sensor location, heater-sensor separation, encapsulation thickness, calibration method, measurement uncertainty, control logic, overshoot limits, and fault detection. Without this information, the displayed temperature should not be assumed to represent the true wound-interface temperature. Figure 4 shows how heater design, sensing position, electronics, exudate, contact pressure, and dressing materials together determine delivered dose and control reliability [92,110,126].
Figure 4. Conceptual architecture of a sensor-guided thermotherapeutic dressing with feedback control. Temperature and auxiliary sensors inform a controller that adjusts heater output relative to the prescribed setpoint; sensor position is important because the measured temperature may not represent the actual wound-interface temperature.
Overall, the field has moved away from rigid heaters and toward conformal, sensor-integrated, and multifunctional platforms. The clinical value of this added complexity will depend on whether it produces stable, measured, and safe interface temperatures during realistic use, not only during dry bench testing [127,128,129].

5. Regulatory, Safety, and Validation Considerations

5.1. Regulatory Pathways, Technical Standards, and Performance Validation

Regulatory classification depends on the product’s intended purpose, primary mode of action, energy source, software functions, duration and site of contact, and any combination with drugs or biologics; no single pathway covers all wound-thermotherapy platforms. In the United States, conventional powered heating pads are regulated under 21 CFR 890.5740 as Class II devices [15]. In the European Union, classification under the Medical Device Regulation depends on intended purpose and risk; active therapeutic devices that exchange energy are addressed under Rule 9, with class depending on whether energy delivery is potentially hazardous [130,131]. Accordingly, the technology archetypes in Table 5 are not assigned universal regulatory classes or submission pathways. The applicable pathway must be determined from the specific product configuration, intended purpose, mode of action, constituent functions, and jurisdiction. Flexible sensor-controlled heaters, externally activated photothermal materials, wireless closed-loop dressings, and heating-plus-drug systems can therefore require different regulatory analyses [130,131,132]. For eligible novel low- to moderate-risk devices without a legally marketed predicate, the U.S. FDA De Novo pathway can provide a classification route; whether it applies depends on the specific intended use and risk profile [133].
Table 5. Device-specific regulatory and validation considerations for representative thermotherapeutic wound technologies.
Horizontal standards remain relevant but must be matched to the device architecture. IEC 60601-1 addresses basic safety and essential performance for medical electrical equipment; IEC 60601-2-35 addresses heating devices using blankets, pads, or mattresses; IEC 60601-1-10 addresses physiologic closed-loop controllers; and IEC 60601-1-2 addresses electromagnetic compatibility [16,134,135,136]. ISO 14971 provides the risk-management framework [17]. Table 5 therefore separates established regulatory and standards frameworks that may apply to a given product from product-specific classification considerations and from the wound-interface validation criteria proposed by the authors for chronic-wound translation. To make these distinctions concrete, Table 5 is organized around representative device examples that span conventional warming, sensor-integrated heating, externally activated photothermal systems, connected closed-loop dressings, and heating combined with drug delivery.
Device safety depends on predictable thermal output, reliable insulation, and robust control during exposure to exudate, high humidity, compression, offloading, repeated handling, and variable contact pressure [17,83,84,85,140,141]. Under International Organization for Standardization (ISO) 14971, these conditions should be represented explicitly in hazard analysis, risk controls, verification testing, and residual-risk justification. Relevant failure modes include power instability, insulation breach, sensor drift, and control faults [17].
Many prototypes still demonstrate safety only under ideal laboratory conditions. Long-duration electrical and thermal performance in wet, mechanically stressed, dressing-compatible environments is rarely reported in a form that would support risk-based regulatory review [142]. Performance claims should be anchored to delivered thermal exposure. Premarket evidence should quantify heating rate, steady-state stability, peak temperature, spatial gradients, exposure duration, and the test conditions used to obtain those values. Because many wound-heating studies report only nominal setpoints or treatment schedules, standardized interface-dose reporting under moisture, pressure, and dressing variation is important for regulatory credibility as well as clinical interpretation [143,144,145].

5.2. Human Factors, Usability, and Workflow-Related Risks

Human factors engineering focuses on how clinicians and patients interact with devices and is an expected part of US and EU medical-device development when safe use must be demonstrated under intended conditions. Evaluations typically address interface clarity, ease of placement, alarm visibility, and the adequacy of labeling and instructions because these features influence whether a device can be used safely and consistently. For chronic-wound thermotherapy, human factors assessment also points to the need for simple temperature controls, clear treatment-progress indicators, and explicit warnings for high-risk conditions such as ischemia or impaired sensation [113,139]. Published usability testing remains limited for many thermotherapeutic systems, especially in settings that resemble clinical workflows or home-based care. This matters because foreseeable use errors can materially change delivered thermal exposure. IEC 62366-1 requires a usability-engineering process focused on safety-related use errors [146]. Heating-pad recalls involving overheating, electrical faults, and burn injuries illustrate the consequences of control or manufacturing failures [147].

5.3. Cybersecurity and Data Governance

As thermotherapeutic devices incorporate wireless sensing or remote monitoring, cybersecurity becomes part of the regulatory assessment of connected medical devices. Remote temperature monitoring in high-risk diabetic foot care illustrates why secure communication and protected data transmission matter when streamed measurements may influence clinical decisions [106]. Current regulatory expectations emphasize documented controls, including authentication, encryption, secure update mechanisms, and lifecycle processes that reduce the likelihood of unauthorized changes to device parameters. Many wireless thermotherapy prototypes still do not report cybersecurity assessment methods or validation evidence. Connected systems therefore require cybersecurity risk management across the product life cycle and evidence appropriate to premarket and postmarket oversight [137,148,149].

5.4. Quality Management and Implications for Developers

Manufacturers must meet quality-system requirements that formalize design controls, risk documentation, verification and validation, and post-market quality activities across the device life cycle. In the United States, the FDA Quality Management System Regulation aligns quality expectations with ISO 13485:2016 by incorporation by reference [150,151,152]. This reinforces the need for documented processes that support consistent design, testing, and change control. For thermotherapeutic devices, performance testing should reflect foreseeable use conditions, including moisture exposure, repeated handling, and interactions with dressing materials that can alter heat transfer and electrical integrity. Clear documentation of design assumptions, test criteria, and performance limits is necessary for regulatory acceptance and for judging whether bench evidence is transferable to clinical use. Many academic prototypes are not developed within a regulatory-grade quality system, and missing verification evidence can become a primary barrier to translation even when the basic device function appears promising. These expectations directly affect whether a promising heater can become a reproducible wound-care device. A wound-relevant verification framework is still needed to connect electrical safety, heat uniformity, sensing reliability, software behavior, human factors, and post-market surveillance for flexible and connected systems [130,134,151]. The clinical evidence should therefore be interpreted within this regulatory and validation context.

6. Clinical Evidence and Translational Implications

6.1. Overview of Clinical Studies and Wound Types

Direct human treatment evidence for established chronic wounds is concentrated in historical radiant or noncontact normothermic wound therapy (NNWT) studies, together with a limited number of wIRA and other local-heat reports. These studies provide the clinical foundation for the present synthesis, while contemporary flexible, wireless, and photothermal platforms are considered according to their own engineering and preclinical evidence. Schumann et al. [12] is included as modality-specific wIRA + VIS evidence and interpreted in relation to the combined intervention. Scott et al. [13] is retained as prevention-related contextual evidence because it evaluated intraoperative forced-air warming for prevention of postoperative pressure ulcers. Galdino-Junior et al. [39] is retained as supportive/contextual clinical evidence because it evaluated warmed saline during dressing changes. The systematic review by Yue et al. [14] screened 139 records and found no RCT meeting its prespecified eligibility criteria. Of the three full-text reports assessed, one was considered nonrandomized, one evaluated warming together with co-interventions, and one reported use of a random-number table with insufficient methodological detail to establish randomization. The present synthesis therefore distinguishes the design terminology reported in individual primary publications from eligibility under Yue et al.’s later systematic-review criteria. Table 6 and Supplementary Tables S2 and S3 use design terminology supported by the primary reports.
Table 6. Quantitative outcome signals, thermal-dose limitations, and interpretation of direct clinical thermotherapy studies.
The clinical relevance of local warming depends on wound context and should not be inferred from the presence of treatment evidence alone. The available direct human evidence is concentrated primarily in venous, pressure, and diabetes-related wounds, whereas evidence for arterial/ischemic and mixed-etiology wounds is more limited. Differences in evidence availability should be considered separately from wound-specific factors that may influence the safety and reproducibility of thermal intervention, including impaired perfusion, neuropathy or sensory loss, infection, mechanical loading, and interaction with standard etiologic care. Postoperative warming evidence considered in this review is contextual and prevention-related rather than evidence of thermotherapy for established chronic wounds. Accordingly, the present synthesis does not assign numerical evidence or safety rankings to individual wound types.
Supplementary Table S2 and Table 6 summarize the core direct-treatment reports using information directly supported by the accessible publication record. Supplementary Table S2 reports total sample size and treatment/control group sizes where available, distinguishes enrolled or randomized populations from analyzed/completer populations where relevant, and identifies whether the reported unit is the patient or the wound. For studies in which a participant contributed more than one wound, patient and wound counts are reported separately and the analytical level used in the primary study is identified. The table also summarizes assignment/comparator, baseline context, thermal intervention, target or device setpoint, achieved or measured temperature, measurement site, follow-up, attrition, and adverse-event reporting. Target/setpoint and measured wound or periwound temperature are reported as separate quantities. Follow-up is presented as the documented observation period or stopping rule, and attrition is reported by distinguishing enrolled or randomized populations from analyzed or completer populations where these differ. Adverse-event reporting is recorded according to the wording of the primary publication. Table 6 presents study-specific quantitative outcome signals together with the corresponding thermal-dose and methodological context, including healing rate, proportion healed, time to prespecified wound-area thresholds, wound-area or volume reduction, recurrence, pain scores, and associated significance values where reported. Supplementary Table S3 provides a study-level qualitative risk-of-bias appraisal based on assignment, attrition and unit of analysis, co-interventions or blinding, and thermal-dose measurement. Across these tables, only values and study characteristics supported by the accessible source record are reported, and unavailable items are identified explicitly.

6.2. Thermal Dosing, Temperature Control, and Microcirculatory Outcomes

The main gap in clinical reporting is characterization of the delivered thermal exposure. Most studies describe an intended setpoint, nominal device behavior, or treatment schedule instead of continuous interface temperature, spatial uniformity, peak temperature, recovery after dressing changes, and cumulative exposure. Without these data, outcomes cannot be linked confidently to biological effect, device performance, or application variability [41,57]. Microcirculatory measures should be interpreted as mechanistic intermediates, not as substitutes for healing endpoints. Local warming can increase superficial flow and oxygenation, but future trials should test whether deeper hypoxic wound regions receive meaningful oxygen delivery. This will require dose logs paired with prespecified perfusion or oxygenation measures and closure-related outcomes [56,153]. Near-infrared spectroscopy is another noninvasive approach for characterizing wound-tissue oxygenation and perfusion and may complement, rather than replace, healing endpoints [154]. Patient-reported outcomes and safety events should be collected prospectively, including heat-related comfort, pain during dressing changes, adherence, maceration, burns, worsening infection, and device-related pressure injury [39,41]. Because heat-pain profiles vary even among pain-free volunteers, thermal discomfort should be measured directly rather than inferred from device temperature alone [155].

6.3. Methodological Quality and Outcome Measures

Supplementary Table S3 shows that the principal limitations are small samples, incomplete allocation or attrition reporting, wound-versus-patient unit issues, lack of blinding in several reports, co-interventions, and incomplete thermal dosimetry. Yue et al. screened 139 records and assessed three full-text reports, none of which met the review’s prespecified inclusion criteria [14]. Because that review does not provide a study-by-study crosswalk for the historical NNWT reports summarized here, the present review does not retrospectively reclassify those studies. Randomized-design terminology therefore reports the description supported by the primary publication and is kept distinct from eligibility under the systematic-review criteria of Yue et al. [14]. Future protocols should follow SPIRIT guidance, completed randomized trials should follow CONSORT, and device-based thermotherapy should be described with TIDieR-level detail, including temperature-control logic, sensor placement, dressing stack, co-interventions, adherence, and adverse-event capture [156,157,158].

6.4. Wound Settings Most Likely to Benefit and Major Risk Groups

Current evidence positions thermotherapy as an investigational adjunct rather than routine care. The most plausible candidate wounds are those with impaired but recoverable perfusion, delayed granulation, persistent inflammation, and preserved capacity for microvascular response. Venous leg ulcers and pressure injuries provide the clearest human treatment signals to date, although the evidence base remains limited and methodologically heterogeneous [41].
Severe ischemia, neuro-ischemic disease, diabetic neuropathy, active infection, and mixed-etiology ulcers remain high-risk settings. Poor perfusion reduces heat removal, sensory loss may mask overheating, peripheral arterial disease can blunt vascular redistribution, and infection may increase metabolic demand in oxygen-limited tissue. Until temperature limits, exposure times, stopping rules, and monitoring requirements are defined, unsupervised use in these groups is not justified [159,160,161,162,163,164].

6.5. Thermotherapy in Combination with Standard and Adjunctive Wound Care

Established chronic-wound care is etiology-specific and can include compression for venous disease, offloading for diabetes-related foot disease, pressure-injury management, debridement, infection management, moisture balance, and NPWT in selected wounds [159,165,166,167,168,169,170]. NPWT has randomized clinical evidence in diabetic foot ulcers and has also been studied with wound thermometry; these data concern NPWT itself and do not establish benefit from adding heat [171,172,173]. Thermotherapy should therefore be studied as an adjunct within established care pathways rather than as an isolated intervention. Potential interactions with NPWT, compression, offloading, and antimicrobial strategies require direct testing because each can alter moisture, contact, perfusion, or local thermal distribution. Table 7 summarizes possible roles, major cautions, evidence position, and knowledge gaps by wound context.
Table 7. Potential indications, cautions, and knowledge gaps for thermotherapy in common wound scenarios.
Compression therapy is central to venous leg ulcer care, and superficial warming has shown potential in this setting. If heating devices are placed beneath or near compression systems, pressure may improve contact and heat transfer. At the same time, trapped moisture and an altered microclimate may increase local thermal exposure. The interaction could strengthen treatment effects, but it could also raise risk, particularly in patients with mixed arterial disease [164]. Existing studies address wound warming [14], compression-related pressure and moisture effects [174], or compression sensing [175] separately; they do not establish temperature distribution, vascular response, or safety limits for combined warming and compression. Debridement removes devitalized tissue and bacterial burden and may create a short interval in which improved perfusion could support early granulation. Applying heat after debridement is therefore plausible, but it has not been tested in controlled clinical studies. Dressing changes also alter moisture, temperature, and interface contact, all of which influence heat transfer. Future protocols should specify the timing of thermotherapy relative to debridement, cleansing, dressing changes, compression, and offloading so that variability in delivered dose is reduced [50,176,177]. Chronic wounds often carry a high bacterial burden. Warming could theoretically support host clearance by improving perfusion, or it could be paired with temperature-responsive antimicrobial dressings and controlled drug release. These combined strategies remain largely preclinical. Clinical studies are needed to determine whether warming improves antimicrobial efficacy, changes bacterial growth, alters exudate dynamics, or creates safety concerns in infected or heavily colonized wounds [10,178,179].

6.6. Care Models, Home-Based Use, and Remote Monitoring

Remote temperature monitoring can detect early tissue changes in high-risk diabetic feet and has been associated with reduced ulcer formation and amputation in real-world programs [107]. These findings support temperature-informed care, but they should not be interpreted as evidence for therapeutic heating. In home-based thermotherapy, remote monitoring may eventually support supervised dosing, adherence tracking, and alerts. Home wearable and digital foot-care systems have been demonstrated [180,181], but adherence remains a documented limitation in remote temperature monitoring [182,183]. Data quality, cybersecurity, sensor stability, usability, and safe long-wear performance still require direct evaluation [148,184]. Implementation will depend on workflow as much as device design. Qualitative research on virtual chronic-wound care similarly highlights the importance of patient education, multidimensional care needs, and implementation support [185]. Clinicians, caregivers, and patients need simple controls, clear alarms, placement guidance, instructions on dressing compatibility, and explicit responses to fault conditions. Correct use should be verified in representative-user studies because placement or operating errors can produce variable dosing and inconsistent outcomes [139,186]. Overall, the clinical evidence supports controlled investigation rather than routine adoption. In research settings, thermotherapy should remain an adjunct to standard care in carefully selected wounds, with particular caution in poor perfusion, reduced sensation, active infection, or mixed etiology [159,168].

7. Future Directions and Roadmap

7.1. Mechanistic Gaps and Wound-Specific Dose Frameworks

Future mechanistic studies should move beyond intact-skin physiology and simplified cell or bench models. Experimental systems need to reproduce exudate, dressing layers, heterogeneous tissue, impaired perfusion, bacterial burden, and mechanical loading. These variables determine both heat distribution and biological response. Priority assessments include perfusion, oxygenation, inflammatory mediators, heat-shock responses, matrix remodeling, and, where feasible, direct measurements in human chronic-wound tissue [55,75,187]. CEM43 and related metrics remain useful for comparison, but they should not be used as standalone dose definitions for chronic wounds. Wound-specific models should include geometry, moisture, dressing configuration, tissue heterogeneity, and perfusion-dependent heat removal. General hyperthermia modeling confirms that perfusion can materially change heat retention [188]. Wound-specific models should therefore be validated against temperature maps, perfusion measures, and clinical outcomes [42]. Dedicated safety studies are especially needed in patients with poor perfusion, neuropathy, and mixed etiologies. These patients may need perfusion support but may also have reduced ability to dissipate or perceive harmful heat. Conservative ceilings, exposure durations, stopping rules, and monitoring requirements should be defined before unsupervised or home-based use is considered [189,190].

7.2. Translational Readiness and Engineering Constraints

Because thermotherapeutic platforms differ in mechanism and maturity, translational assessment is most informative when it identifies the failure mode that most directly threatens wound-interface dose validity. Table 8 therefore summarizes one priority failure mode for each emerging heater class and the evidence needed to address it. These priorities are deliberately narrower than the broader platform classification in Table 4 and focus on whether the intended thermal exposure remains stable under moisture, deformation, pressure, optical attenuation, and material change [45,46,47,53,114,116,191,192,193,194,195,196,197,198,199].
Table 8. Single critical wound-interface failure mode and verification evidence for emerging thermotherapeutic platforms.

7.3. Standardization of Endpoints, Trial Design, and Combined-Modality Protocols

Future trials should make thermal interventions reproducible. Existing core-outcome and outcome-classification work in complex wounds, venous leg ulcers, and pressure-ulcer research underscores the need to harmonize wound outcomes, while chronic wounds also impose important quality-of-life burdens [200,201,202,203]. Protocols should follow SPIRIT, completed randomized trials should follow CONSORT, and device-based interventions should be described using TIDieR. Reports should include sensor placement, control logic, dressing stack, treatment schedule, co-interventions, adherence, and adverse-event capture [156,157,158,201]. A minimum reporting set should include wound diagnosis and etiology; perfusion and sensory status; target and measured interface temperature; sensor type, location, calibration, response time, and uncertainty; duration, frequency, adherence, dressing configuration, moisture condition, co-interventions, spatial uniformity, peak temperature, adverse thermal events, wound-area method, and patient-relevant outcomes [156,157,158]. Wound-area measurement methods should be validated and described explicitly [204,205]. Protocols that combine thermotherapy with compression, NPWT, debridement, antimicrobial dressings, or bacterial-burden management should test synergy rather than assume it. They should specify how timing and co-interventions affect perfusion, drug delivery, moisture balance, contact pressure, exudate handling, sensor accuracy, and thermal distribution [156,157,201].

7.4. Regulatory Evolution, Digital Health, and Intelligent Control

Physical wound-interface durability and platform-specific failure modes are summarized in Table 8. For connected and adaptive thermotherapy systems, additional translational requirements arise from sensing, software, communications, cybersecurity, closed-loop behavior, and long-term data quality. Remote monitoring and automatic control may eventually support supervised home thermotherapy, adherence tracking, and alerts. Cybersecurity and closed-loop control requirements are addressed by FDA guidance [148,206], while foot-monitoring studies identify adherence and implementation limitations [107,180,183]. Sensor accuracy, data quality, privacy, and long-term performance in the home still require direct evaluation. Adaptive wound bioelectronic platforms provide useful design analogies for thermotherapy because they combine distributed sensing, wireless communication, and closed-loop control. Systems such as a-Heal show how diagnostics and therapy may eventually be integrated in advanced wound devices. They should be interpreted as translational analogies rather than direct evidence for thermal wound treatment [207]. Adaptive algorithms could personalize heating according to perfusion, moisture, temperature history, or healing trajectory. They also introduce verification and regulatory challenges. Recent FDA guidance on artificial intelligence (AI)-enabled device software and predetermined change control plans is relevant to this direction [208,209,210]. Before intelligent heat-control systems can be used routinely, developers will need hardware-in-the-loop testing, fault handling, overshoot limits, drift detection, cybersecurity controls, and monitored feasibility studies under realistic wound conditions [206,207,208,209,210].

7.5. Integrated Roadmap Priorities

As summarized in Figure 5, progress toward routine wound-care use will require five linked advances: wound-specific dose definition, validation under realistic interface conditions, standardized reporting, clinically informative trials, and translation pathways that address usability, manufacturing, cybersecurity, safety, and regulation [137,139,207].
Figure 5. Translational roadmap for thermotherapeutic wound devices. The upper sequence summarizes five development priorities from mechanistic dose–response studies to regulatory translation; the lower band highlights cross-cutting priorities for implementation, durability, usability, and adaptive monitoring. Arrows indicate a conceptual progression rather than a validated linear pathway.
Table 9 summarizes priority research questions across chronic-wound mechanisms, dose modeling, device durability, high-risk patient groups, combined treatments, hybrid-device standards, remote monitoring, and adaptive control. For each area, it identifies suggested methods, core measures, and translational relevance so that future studies are easier to compare and more useful for device development and clinical decision-making.
Table 9. Priority research questions and proposed approaches for advancing thermotherapeutic wound devices.
In practical terms, priority should be given to wound-relevant mechanistic studies, moist and pressurized interface validation, early safety studies in high-risk limbs, controlled combination protocols, and regulatory-grade evidence for connected or adaptive devices.

8. Conclusions

Thermotherapy is biologically plausible for selected chronic wounds, but it is not yet a practice-defining adjunct. This review treats local warming as a wound-interface, dose-controlled device intervention and separates modalities that differ substantially in energy delivery and biological co-effects. Clinical credibility depends on whether a defined exposure can be specified, delivered, measured, controlled, and justified under the biological, mechanical, and workflow conditions of chronic-wound care. Evidence from historical NNWT or wIRA should not be transferred directly to Joule-heated, photothermal, wireless, or multimodal platforms without direct supporting data.
Flexible heaters, hydrogel systems, photothermal materials, bioresorbable devices, and sensor-guided dressings broaden the technical options for local warming, with interface-level evidence providing the basis for translation. For lower-temperature wound warming, complete measured time–temperature profiles, peak temperature, spatial variation, measurement location, and sensor uncertainty provide the most informative dose description, while CEM43 can serve as a secondary normalization metric during wound-specific biological validation. Future work should emphasize wound-specific dosing frameworks, realistic interface validation, standardized clinical reporting and risk-of-bias assessment, and controlled evaluation alongside established wound-care practices. Together, these steps provide a clear evidence pathway toward safe, reproducible, and clinically credible thermotherapy modalities.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bioengineering13101126/s1, Table S1: Consolidated database-specific search expressions corresponding to the iterative search framework; Figure S1: Focused clinical-evidence selection flow for the direct-treatment synthesis; Table S2: Direct human studies of local warming for established chronic wounds: design, sample, thermal intervention, target/setpoint and achieved/measured temperature, follow-up, attrition, and adverse-event reporting; Table S3: Study-level qualitative risk-of-bias appraisal for direct clinical thermotherapy reports.

Author Contributions

Conceptualization, S.A.H.; methodology, S.A.H. and B.A.-N.; literature investigation, S.A.H. and M.A.-A.; visualization, S.A.H.; writing—original draft preparation, S.A.H.; writing—review and editing, S.A.H., B.A.-N., M.A.-A., A.-R.A.-H. and O.K.; supervision, O.K.; project administration, S.A.H. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable. This narrative review did not involve human participants or animal subjects.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Acknowledgments

During the preparation of this manuscript, the authors used ChatGPT (OpenAI; web application, accessed during manuscript preparation from December 2025 to September 2026; multiple model versions were used as the service was updated) for language refinement, structural organization, reference-formatting assistance, and development of conceptual figure drafts under author guidance. The authors reviewed and edited all outputs and took full responsibility for the content of this publication.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Sen, C.K. Human Wounds and Its Burden: An Updated Compendium of Estimates. Adv. Wound Care 2019, 8, 39–48. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Nussbaum, S.R.; Carter, M.J.; Fife, C.E.; DaVanzo, J.; Haught, R.; Nusgart, M.; Cartwright, D. An Economic Evaluation of the Impact, Cost, and Medicare Policy Implications of Chronic Nonhealing Wounds. Value Health 2018, 21, 27–32. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Sibbald, R.G.; Elliott, J.A.; Persaud-Jaimangal, R.; Goodman, L.; Armstrong, D.G.; Harley, C.; Coelho, S.; Xi, N.; Evans, R.; Mayer, D.O.; et al. Wound Bed Preparation 2021. Adv. Skin Wound Care 2021, 34, 183–195. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Frykberg, R.G.; Banks, J. Challenges in the Treatment of Chronic Wounds. Adv. Wound Care 2015, 4, 560–582. [Google Scholar] [CrossRef] [Scilit]
  5. Johnson, J.M.; Kellogg, D.L. Local thermal control of the human cutaneous circulation. J. Appl. Physiol. 2010, 109, 1229–1238. [Google Scholar] [CrossRef] [Scilit]
  6. Horwitz, O.; Peirce, G.; Montgomery, H. Oxygen Tension of Tissues by the Polarographic Method. Circulation 1951, 4, 111–115. [Google Scholar] [CrossRef] [Scilit]
  7. Yoo, K.; Suh, K.Y.; Choi, G.H.; Kwak, I.-S.; Seo, D.K.; Kym, D.; Yoon, H.; Cho, Y.S.; Kim, H.O. Serial Changes of Heat Shock Protein 70 and Interleukin-8 in Burn Blister Fluid. Ann. Dermatol. 2017, 29, 194. [Google Scholar] [CrossRef] [Scilit]
  8. Li, Y.; Zhang, H.; Ye, Z.; Liu, M.; Liu, W.; Li, Z.; Gui, L. A Novel Flexible Liquid Metal Microheater with a Textured Structure. Micromachines 2024, 15, 75. [Google Scholar] [CrossRef] [Scilit]
  9. Kang, M.; Jeong, H.; Park, S.-W.; Hong, J.; Lee, H.; Chae, Y.; Yang, S.; Ahn, J.-H. Wireless graphene-based thermal patch for obtaining temperature distribution and performing thermography. Sci. Adv. 2022, 8, eabm6693. [Google Scholar] [CrossRef] [Scilit]
  10. Ding, J.; Jiang, J.; Tian, Y.; Su, B.; Zeng, M.; Wu, C.; Wei, D.; Sun, J.; Luo, H.; Fan, H. Temperature-Responsive Hydrogel System Integrating Wound Temperature Monitoring and On-demand Drug Release for Sequentially Inflammatory Process Regulation of Wound Healing. ACS Appl. Mater. Interfaces 2024, 16, 67444–67457. [Google Scholar] [CrossRef] [Scilit]
  11. Ju, Y.S. Thermal management and control of wearable devices. iScience 2022, 25, 104587. [Google Scholar] [CrossRef] [Scilit]
  12. Schumann, H.; Calow, T.; Weckesser, S.; Müller, M.; Hoffmann, G. Water-filtered infrared A for the treatment of chronic venous stasis ulcers of the lower legs at home: A randomized controlled blinded study. Br. J. Dermatol. 2011, 165, 541–551. [Google Scholar] [CrossRef] [Scilit]
  13. Scott, E.M.; Leaper, D.J.; Clark, M.; Kelly, P.J. Effects of Warming Therapy on Pressure Ulcers-a Randomized Trial. AORN J. 2001, 73, 921–938. [Google Scholar] [CrossRef] [Scilit]
  14. Yue, J.-H.; Zhang, S.-J.; Sun, Q.; Sun, Z.-R.; Wang, X.-X.; Golianu, B.; Lu, Y.; Zhang, Q. Local warming therapy for treating chronic wounds. Medicine 2018, 97, e9931. [Google Scholar] [CrossRef] [Scilit]
  15. U.S. Government Publishing Office. 21 CFR 890.5740—Powered Heating Pad. Available online: https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-890/subpart-F/section-890.5740 (accessed on 31 August 2026).
  16. International Electrotechnical Commission. IEC 60601-2-35:2020+AMD1:2023 CSV, Medical Electrical Equipment—Part 2-35: Particular Requirements for the Basic Safety and Essential Performance of Heating Devices Using Blankets, Pads or Mattresses and Intended for Heating in Medical Use; IEC: Geneva, Switzerland, 2023; Available online: https://webstore.iec.ch/en/publication/68375 (accessed on 31 August 2026).
  17. ISO 14971:2019; Medical Devices—Application of Risk Management to Medical Devices. International Organization for Standardization (ISO): Geneva, Switzerland, 2019.
  18. Robinson, C.; Santilli, S.M. Warm-Up Active Wound Therapy: A novel approach to the management of chronic venous stasis ulcers. J. Vasc. Nurs. 1998, 16, 38–42. [Google Scholar] [CrossRef] [Scilit]
  19. Santilli, S.M.; Valusek, P.A.; Robinson, C. Use of a noncontact radiant heat bandage for the treatment of chronic venous stasis ulcers. Adv. Wound Care 1999, 12, 89–93. [Google Scholar]
  20. Cherry, G.W.; Wilson, J. The treatment of ambulatory venous ulcer patients with warming therapy. Ostomy Wound Manag. 1999, 45, 65–70. [Google Scholar]
  21. Price, P.; Bale, S.; Crook, H.; Harding, K. The effect of a radiant heat dressing on pressure ulcers. J. Wound Care 2000, 9, 201–205. [Google Scholar] [CrossRef] [Scilit]
  22. Kloth, L.C.; Berman, J.E.; Dumit-Minkel, S.; Sutton, C.H.; Papanek, P.E.; Wurzel, J. Effects of a normothermic dressing on pressure ulcer healing. Adv. Skin Wound Care 2000, 13, 69–74. [Google Scholar]
  23. Biland, L.; Barras, J. Die wassergefilterte Infrarot-A-Hyperthermie zur Behandlung venöser Ulcera. Hefte Wundbehand 2001, 5, 41. [Google Scholar]
  24. Whitney, J. Treatment of pressure ulcers with noncontact normothermic wound therapy: Healing and warming effects. J. Wound Ostomy Cont. Nurs. 2001, 28, 244–252. [Google Scholar] [CrossRef] [Scilit]
  25. McCulloch, J.; Knight, C.A. Noncontact normothermic wound therapy and offloading in the treatment of neuropathic foot ulcers in patients with diabetes. Ostomy Wound Manag. 2002, 48, 38–44. [Google Scholar]
  26. Kloth, L.C.; Berman, J.E.; Nett, M.; Papanek, P.E.; Dumit-Minkel, S. A Randomized Controlled Clinical Trial to Evaluate the Effects of Noncontact Normothermic Wound Therapy on Chronic Full-thickness Pressure Ulcers. Adv. Skin Wound Care 2002, 15, 270–276. [Google Scholar] [CrossRef] [Scilit]
  27. Alvarez, O.M.; Rogers, R.S.; Booker, J.G.; Patel, M. Effect of noncontact normothermic wound therapy on the healing of neuropathic (diabetic) foot ulcers: An interim analysis of 20 patients. J. Foot Ankle Surg. 2003, 42, 30–35. [Google Scholar] [CrossRef] [Scilit]
  28. Karr, J.C. External Thermoregulation of Wounds Associated with Lower-Extremity Osteomyelitis. J. Am. Podiatr. Med. Assoc. 2003, 93, 18–22. [Google Scholar] [CrossRef] [Scilit][Green Version]
  29. Thomas, D.R.; Diebold, M.R.; Eggemeyer, L.M. A controlled, randomized, comparative study of a radiant heat bandage on the healing of stage 3-4 pressure ulcers: A pilot study. J. Am. Med. Dir. Assoc. 2005, 6, 46–49. [Google Scholar] [CrossRef]
  30. Alvarez, O.; Patel, M.; Rogers, R.; Booker, J. Effect of non-contact normothermic wound therapy on the healing of diabetic neuropathic foot ulcers. J. Tissue Viability 2006, 16, 8–11. [Google Scholar] [CrossRef] [Scilit]
  31. Mercer, J.B.; Nielsen, S.P.; Hoffmann, G. Improvement of wound healing by water-filtered infrared-A (wIRA) in patients with chronic venous stasis ulcers of the lower legs including evaluation using infrared thermography. GMS Ger. Med. Sci. 2008, 6, Doc11. [Google Scholar]
  32. Petrofsky, J.S.; Lawson, D.; Berk, L.; Suh, H. Enhanced healing of diabetic foot ulcers using local heat and electrical stimulation for 30 min three times per week. J. Diabetes 2010, 2, 41–46. [Google Scholar] [CrossRef] [Scilit]
  33. Hoffmann, G. Improvement of wound healing in chronic ulcers by hyperbaric oxygenation and by waterfiltered ultrared A induced localized hyperthermia. Adv. Exp. Med. Biol. 1994, 345, 181–188. [Google Scholar] [CrossRef] [Scilit]
  34. Ellis, S.L.; Finn, P.; Noone, M.; Leaper, D.J. Eradication of methicillin-resistant Staphylococcus aureus from pressure sores using warming therapy. Surg. Infect. 2003, 4, 53–55. [Google Scholar] [CrossRef] [Scilit]
  35. Lawson, D.; Petrofsky, J.S. A randomized control study on the effect of biphasic electrical stimulation in a warm room on skin blood flow and healing rates in chronic wounds of patients with and without diabetes. Med. Sci. Monit. 2007, 13, CR258–CR263. [Google Scholar]
  36. Petrofsky, J.S.; Lawson, D.; Suh, H.J.; Rossi, C.; Zapata, K.; Broadwell, E.; Littleton, L. The influence of local versus global heat on the healing of chronic wounds in patients with diabetes. Diabetes Technol. Ther. 2007, 9, 535–544. [Google Scholar] [CrossRef] [Scilit]
  37. Petrofsky, J.S.; Schwab, E.; Lo, T.; Cuneo, M.; Lawson, D. The thermal effect on the blood flow response to electrical stimulation. Med. Sci. Monit. 2007, 13, CR498–CR504. [Google Scholar]
  38. Suh, H.; Petrofsky, J.S.; Lo, T.; Lawson, D.; Yu, T.; Pfeifer, T.M.; Morawski, T. The combined effect of a three-channel electrode delivery system with local heat on the healing of chronic wounds. Diabetes Technol. Ther. 2009, 11, 681–688. [Google Scholar] [CrossRef] [Scilit]
  39. Galdino-Júnior, H.; Martins de Oliveira Segundo, E.; Gomes Malaquias, S.; Vinaud, M.C.; Pereira, L.V.; Guimarães, R.A.; Bachion, M.M. Effect of Heated Saline Solution on Pain Intensity, Wound Bed Temperature, and Comfort during Chronic Wound Dressing Changes: Crossover Randomized Clinical Trial. Adv. Skin Wound Care 2024, 37, 1–8. [Google Scholar] [CrossRef] [Scilit]
  40. Suh, H.; Petrofsky, J.; Fish, A.; Hernandez, V.; Mendoza, E.; Collins, K.; Yang, T.; Abdul, A.; Batt, J.; Lawson, D. A new electrode design to improve outcomes in the treatment of chronic non-healing wounds in diabetes. Diabetes Technol. Ther. 2009, 11, 315–322. [Google Scholar] [CrossRef] [Scilit]
  41. Haesler, E. Local warming/heat therapy for chronic wounds: A WHAM evidence summary. Wound Pract. Res. 2023, 31, 90–92. [Google Scholar] [CrossRef] [Scilit]
  42. Sapareto, S.A.; Dewey, W.C. Thermal dose determination in cancer therapy. Int. J. Radiat. Oncol. Biol. Phys. 1984, 10, 787–800. [Google Scholar] [CrossRef] [Scilit]
  43. van Rhoon, G.C. Is CEM43 still a relevant thermal dose parameter for hyperthermia treatment monitoring? Int. J. Hyperth. 2016, 32, 50–62. [Google Scholar] [CrossRef] [Scilit]
  44. Hossain, M.I.; Zahid, M.S.; Chowdhury, M.A.; Hossain, M.M.M.; Hossain, N.; Islam, M.A.; Mobarak, M.H. Smart bandage: A device for wound monitoring and targeted treatment. Results Chem. 2024, 7, 101292. [Google Scholar] [CrossRef] [Scilit]
  45. Moradifar, F.; Sepahdoost, N.; Tavakoli, P.; Mirzapoor, A. Multi-functional dressings for recovery and screenable treatment of wounds: A review. Heliyon 2025, 11, e41465. [Google Scholar] [CrossRef] [Scilit]
  46. Wu, P.; Jin, L.; Jiang, W.; Zhou, Y.; Lin, L.; Lin, H.; Chen, H. Smart bandages for wound monitoring and treatment. Biosens. Bioelectron. 2025, 283, 117522. [Google Scholar] [CrossRef] [Scilit]
  47. Niu, Y.; Zhao, Z.; Yang, L.; Lv, D.; Sun, R.; Zhang, T.; Li, Y.; Bao, Q.; Zhang, M.; Wang, L.; et al. Towards Intelligent Wound Care: Hydrogel-Based Wearable Monitoring and Therapeutic Platforms. Polymers 2025, 17, 1881. [Google Scholar] [CrossRef] [Scilit]
  48. Kolahreez, D.; Ghasemi-Mobarakeh, L.; Liebner, F.; Alihosseini, F.; Quartinello, F.; Guebitz, G.M.; Ribitsch, D. Approaches to Control and Monitor Protease Levels in Chronic Wounds. Adv. Ther. 2024, 7, 2300396. [Google Scholar] [CrossRef] [Scilit]
  49. Falanga, V.; Isseroff, R.R.; Soulika, A.M.; Romanelli, M.; Margolis, D.; Kapp, S.; Granick, M.; Harding, K. Chronic wounds. Nat. Rev. Dis. Primers 2022, 8, 50. [Google Scholar] [CrossRef] [Scilit]
  50. Eriksson, E.; Liu, P.Y.; Schultz, G.S.; Martins-Green, M.M.; Tanaka, R.; Weir, D.; Gould, L.J.; Armstrong, D.G.; Gibbons, G.W.; Wolcott, R.; et al. Chronic wounds: Treatment consensus. Wound Repair. Regen. 2022, 30, 156–171. [Google Scholar] [CrossRef] [Scilit]
  51. Datta, N.R.; Marder, D.; Datta, S.; Meister, A.; Puric, E.; Stutz, E.; Rogers, S.; Eberle, B.; Timm, O.; Staruch, M.; et al. Quantification of thermal dose in moderate clinical hyperthermia with radiotherapy: A relook using temperature-time area under the curve (AUC). Int. J. Hyperth. 2021, 38, 296–307. [Google Scholar] [CrossRef] [Scilit]
  52. Huang, K.; Liu, W.; Wei, W.; Zhao, Y.; Zhuang, P.; Wang, X.; Wang, Y.; Hu, Y.; Dai, H. Photothermal Hydrogel Encapsulating Intelligently Bacteria-Capturing Bio-MOF for Infectious Wound Healing. ACS Nano 2022, 16, 19491–19508. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  53. Xiao, Z.; Liu, J.; Song, Z.; Du, T.; Du, X. Advances in hydrogels combined with photothermal/photodynamic therapy for bacterial infection. J. Mater. Chem. B 2025, 13, 15145–15166. [Google Scholar] [CrossRef] [Scilit]
  54. Fridberg, M.; Bafor, A.; Iobst, C.A.; Laugesen, B.; Jepsen, J.F.; Rahbek, O.; Kold, S. The role of thermography in assessment of wounds. A Scoping Rev. Inj. 2024, 55, 111833. [Google Scholar] [CrossRef] [Scilit]
  55. Derwin, R.; Patton, D.; Strapp, H.; Moore, Z. Wound pH and temperature as predictors of healing: An observational study. J. Wound Care 2023, 32, 302–310. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  56. Huang, J.; Fan, C.; Ma, Y.; Huang, G. Exploring Thermal Dynamics in Wound Healing: The Impact of Temperature and Microenvironment. Clin. Cosmet. Investig. Dermatol. 2024, 17, 1251–1258. [Google Scholar] [CrossRef] [Scilit]
  57. Wolf, S.T.; Dillon, G.A.; Alexander, L.M.; Kenney, W.L.; Stanhewicz, A.E. Quantification and interpretation of nitric oxide-dependent cutaneous vasodilation during local heating. J. Appl. Physiol. 2024, 137, 1418–1424. [Google Scholar] [CrossRef] [Scilit]
  58. Wang, R.; Gu, S.; Kim, Y.H.; Lee, A.; Lin, H.; Jiang, D. Diabetic Wound Repair: From Mechanism to Therapeutic Opportunities. MedComm 2025, 6, e70406. [Google Scholar] [CrossRef] [Scilit]
  59. Gkiata, P.; Vliora, M.; Bargiota, A.; Flouris, A.D. The effect of increased core body temperature on the vasomotor and sudomotor responses of patients with type 2 diabetes: Protocol of a systematic literature review and meta-analysis. Syst. Rev. 2025, 14, 247. [Google Scholar] [CrossRef] [Scilit]
  60. Coimbra, L.B.C.T.; Pinto-Martinez, A.; Pavan, I.C.B.; Melo, E.G.; Araujo, T.L.S. Dynamics of heat shock protein 70 kDa in heat-shocked and hypoxic human endothelial cells. Cell Stress Chaperones 2025, 30, 100085. [Google Scholar] [CrossRef] [Scilit]
  61. Cui, Q.; Jia, K.; Li, F.; Zheng, J.; Wang, F. Post-translational modifications in heat stress-related diseases. Front. Mol. Biosci. 2025, 12, 1666874. [Google Scholar] [CrossRef] [Scilit]
  62. Dai, Y.; Chen, Y. Targeting persistently activated inflammatory microenvironment to promote chronic wound healing. Front. Immunol. 2025, 16, 1708358. [Google Scholar] [CrossRef] [Scilit]
  63. He, M.; Huang, X.; Liu, T.; Yang, Q.; Zheng, Y.; Wu, Y.; Zhang, Y.; Liang, Y.; Liao, Z.; Peng, S.; et al. Microwave therapy promotes wound healing in diabetic mice by activating IL-33/ST2-mediated M2 macrophage polarization. Sci. Rep. 2025, 15, 37984. [Google Scholar] [CrossRef] [Scilit]
  64. Yan, X.; Xiu, F.; An, H.; Wang, X.; Wang, J.; Cao, X. Fever range temperature promotes TLR4 expression and signaling in dendritic cells. Life Sci. 2007, 80, 307–313. [Google Scholar] [CrossRef] [Scilit]
  65. Shin, M.H.; Seo, J.-E.; Kim, Y.K.; Kim, K.H.; Chung, J.H. Chronic heat treatment causes skin wrinkle formation and oxidative damage in hairless mice. Mech. Ageing Dev. 2012, 133, 92–98. [Google Scholar] [CrossRef] [Scilit]
  66. Gefen, A. The Role of the Thermal Conductivity of Dressings in Prevention and Treatment of Wounds. Wounds Int. 2021, 12, 18–24. Available online: https://woundsinternational.com/journal-articles/the-role-of-the-thermal-conductivity-of-dressings-in-prevention-and-treatment-of-wounds/ (accessed on 31 August 2026).
  67. Grigatti, A.; Gefen, A. What makes a hydrogel-based dressing advantageous for the prevention of medical device-related pressure ulcers. Int. Wound J. 2022, 19, 515–530. [Google Scholar] [CrossRef] [Scilit]
  68. Shrivastava, D.; Vaughan, J.T. A Generic Bioheat Transfer Thermal Model for a Perfused Tissue. J. Biomech. Eng. 2009, 131, 074506. [Google Scholar] [CrossRef] [Scilit]
  69. Cwajda-Bialasik, J.; Filipska-Blejder, K.; Moscicka, P. Medical thermography in the diagnosis of pressure ulcers: A narrative review. J. Wound Manag. 2025, 26, 100–110. [Google Scholar] [CrossRef] [Scilit]
  70. Nam, H.J.; Wee, S.Y.; Kim, S.Y.; Jeong, H.G.; Lee, D.W.; Byeon, J.-Y.; Park, S.-H.; Choi, H.J. The correlation between transcutaneous oxygen pressure (TcPO2) and forward-looking infrared (FLIR) thermography in the evaluation of lower extremity perfusion according to angiosome. Int. Wound J. 2024, 21, e14431. [Google Scholar] [CrossRef] [Scilit]
  71. Bernasconi, S.; Angelucci, A.; De Cesari, A.; Masotti, A.; Pandocchi, M.; Vacca, F.; Zhao, X.; Paganelli, C.; Aliverti, A. Recent Technologies for Transcutaneous Oxygen and Carbon Dioxide Monitoring. Diagnostics 2024, 14, 785. [Google Scholar] [CrossRef] [Scilit]
  72. Katsui, S.; Yamamoto, Y.; Kudo, T. Development of a Non-Invasive Local Heating Load Test to Detect Severe Limb Ischemia Within 200 seconds. Vasc. Endovasc. Surg. 2023, 57, 19–25. [Google Scholar] [CrossRef] [Scilit]
  73. Foster, K.R.; Ziskin, M.C.; Balzano, Q. Time-temperature Thresholds and Safety Factors for Thermal Hazards from Radiofrequency Energy above 6 GHz. Health Phys. 2021, 121, 234–247. [Google Scholar] [CrossRef] [Scilit]
  74. Yarmolenko, P.S.; Moon, E.J.; Landon, C.; Manzoor, A.; Hochman, D.W.; Viglianti, B.L.; Dewhirst, M.W. Thresholds for thermal damage to normal tissues: An update. Int. J. Hyperth. 2011, 27, 320–343. [Google Scholar] [CrossRef] [Scilit]
  75. Wiegand, C.; Hipler, U.-C.; Elsner, P.; Tittelbach, J. Keratinocyte and Fibroblast Wound Healing In Vitro Is Repressed by Non-Optimal Conditions but the Reparative Potential Can Be Improved by Water-Filtered Infrared A. Biomedicines 2021, 9, 1802. [Google Scholar] [CrossRef] [Scilit]
  76. Amrani, G.; Peko, L.; Hoffer, O.; Ovadia-Blechman, Z.; Gefen, A. The microclimate under dressings applied to intact weight-bearing skin: Infrared thermography studies. Clin. Biomech. 2020, 75, 104994. [Google Scholar] [CrossRef] [Scilit]
  77. Rauscher, M.; Rauscher, A.; Hu, L.Y.; Schlitt, H.J.; Krauß, S.; Illg, C.; Reis Wolfertstetter, P.; Hofmann, A.; Knorr, C.; Denzinger, M. Influence of Accumulation of Humidity under Wound Dressings and Effects on Transepidermal Water Loss (TEWL) and Skin Hydration. Appl. Sci. 2024, 14, 7739. [Google Scholar] [CrossRef] [Scilit]
  78. MacRae, B.A.; Spengler, C.M.; Psikuta, A.; Rossi, R.M.; Annaheim, S. A Thermal Skin Model for Comparing Contact Skin Temperature Sensors and Assessing Measurement Errors. Sensors 2021, 21, 4906. [Google Scholar] [CrossRef] [Scilit]
  79. Schwartz, D.; Gefen, A. An integrated experimental-computational study of the microclimate under dressings applied to intact weight-bearing skin. Int. Wound J. 2020, 17, 562–577. [Google Scholar] [CrossRef] [Scilit]
  80. Nuutila, K.; Eriksson, E. Moist Wound Healing with Commonly Available Dressings. Adv. Wound Care 2021, 10, 685–698. [Google Scholar] [CrossRef] [Scilit]
  81. Call, E.; Oberg, C.; Streit, I.; Rappl, L.M. Comparing fluid handling and microclimate conditions under superabsorbent polymer and superabsorbent foam dressings over an artificial wound. WCET J. 2019, 39, 11–23. [Google Scholar] [CrossRef] [Scilit]
  82. Yuan, X.; Fu, J.; Mao, Y.; Li, J.; Zhang, Y.; Zhao, Y.; Zhang, B.; Yin, R.; Zhang, H. Advances in bioheat transfer models for hyperthermia: A comprehensive review and future directions. Results Eng. 2025, 28, 107499. [Google Scholar] [CrossRef] [Scilit]
  83. Yoo, S.; Yang, T.; Park, M.; Jeong, H.; Lee, Y.J.; Cho, D.; Kim, J.; Kwak, S.S.; Shin, J.; Park, Y.; et al. Responsive materials and mechanisms as thermal safety systems for skin-interfaced electronic devices. Nat. Commun. 2023, 14, 1024. [Google Scholar] [CrossRef] [Scilit]
  84. Liu, L.; Dou, Y.; Wang, J.; Zhao, Y.; Kong, W.; Ma, C.; He, D.; Wang, H.; Zhang, H.; Chang, A.; et al. Recent Advances in Flexible Temperature Sensors: Materials, Mechanism, Fabrication, and Applications. Adv. Sci. 2024, 11, 2405003. [Google Scholar] [CrossRef] [Scilit]
  85. Luo, Y.; Abidian, M.R.; Ahn, J.-H.; Akinwande, D.; Andrews, A.M.; Antonietti, M.; Bao, Z.; Berggren, M.; Berkey, C.A.; Bettinger, C.J.; et al. Technology Roadmap for Flexible Sensors. ACS Nano 2023, 17, 5211–5295. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  86. Piazena, H.; Vaupel, P.; Thomsen, A.R. Clinical wIRA-hyperthermia: Heating properties and effectiveness in lower trunk regions and its accordance with ESHO quality criteria for superficial hyperthermia. Int. J. Hyperth. 2023, 40, 2244208. [Google Scholar] [CrossRef] [Scilit]
  87. Chen, S.; Cui, Z.; Wang, H.; Wang, X.; Liu, J. Liquid metal flexible electronics: Past, present, and future. Appl. Phys. Rev. 2023, 10, 021308. [Google Scholar] [CrossRef] [Scilit]
  88. Wang, Y.; Yu, Z.; Mao, G.; Liu, Y.; Liu, G.; Shang, J.; Qu, S.; Chen, Q.; Li, R.-W. Printable Liquid-Metal@PDMS Stretchable Heater with High Stretchability and Dynamic Stability for Wearable Thermotherapy. Adv. Mater. Technol. 2019, 4, 1800435. [Google Scholar] [CrossRef] [Scilit]
  89. Ma, X.; Wu, X.; Cao, S.; Zhao, Y.; Lin, Y.; Xu, Y.; Ning, X.; Kong, D. Stretchable and Skin-Attachable Electronic Device for Remotely Controlled Wearable Cancer Therapy. Adv. Sci. 2023, 10, 2205343. [Google Scholar] [CrossRef] [Scilit]
  90. Traiwattanapong, W.; Molahalli, V.; Pattanaporkratana, A.; Chattham, N. Recent Developments in Thermally Stable Transparent Thin Films for Heater Applications: A Systematic Review. Nanomaterials 2024, 14, 2011. [Google Scholar] [CrossRef] [Scilit]
  91. Pavlou, C.; Pastore Carbone, M.G.; Manikas, A.; Tsakonas, C.; Koutroumanis, N.; Galiotis, C. Highly efficient and responsive thin heaters based on CVD graphene/polyetherimide nanolaminates for next-gen thermal management applications. Chem. Eng. J. 2024, 497, 154744. [Google Scholar] [CrossRef] [Scilit]
  92. Pang, Q.; Yang, F.; Jiang, Z.; Wu, K.; Hou, R.; Zhu, Y. Smart wound dressing for advanced wound management: Real-time monitoring and on-demand treatment. Mater. Des. 2023, 229, 111917. [Google Scholar] [CrossRef] [Scilit]
  93. Mohammed, D.W. Mechanical Flexibility and Electrical Reliability of ZnO-Al Thin Films on Polymer Substrates Under Different External Deformation. Adv. Mater. Sci. 2024, 24, 32–45. [Google Scholar] [CrossRef] [Scilit]
  94. Patel, S.; Ershad, F.; Zhao, M.; Isseroff, R.R.; Duan, B.; Zhou, Y.; Wang, Y.; Yu, C. Wearable electronics for skin wound monitoring and healing. Soft Sci. 2022, 2, 9. [Google Scholar] [CrossRef] [Scilit]
  95. Yang, G.; Xie, B.; Tian, J.; Zhang, Y.; Liu, Y.; Gong, H.; Pang, B.; Qiu, Y.; Gong, C.; Bu, T.; et al. A Wireless Hydrogel Thermotherapy System with Adhesion-Customizable Interfaces for Accelerating Wound Healing. Adv. Mater. 2026, 38, e13302. [Google Scholar] [CrossRef] [Scilit]
  96. Du, N.; Fan, Y.; Zhang, Y.; Huang, H.; Lyu, Y.; Cai, R.; Zhang, Y.; Zhang, T.; Guan, Y.; Nan, K. Wireless, Programmable, and Refillable Hydrogel Bioelectronics for Enhanced Diabetic Wound Healing. Adv. Sci. 2024, 11, 2407820. [Google Scholar] [CrossRef] [Scilit]
  97. Damiri, F.; Simińska-Stanny, J.; Rasoulian, F.; Pinal, C.; Kaczmarek-Szczepanska, B.; Patel, C.D.; Patel, H.M.; Papadimitirou, S.A.; Jaradat, E.; Paiva-Santos, A.C.; et al. Emerging trends in polysaccharide-based smart PEGylated hydrogels for biomedical applications. Carbohydr. Polym. 2026, 375, 124707. [Google Scholar] [CrossRef] [Scilit]
  98. Tamayol, A.; Hassani Najafabadi, A.; Mostafalu, P.; Yetisen, A.K.; Commotto, M.; Aldhahri, M.; Abdel-Wahab, M.S.; Izadi Najafabadi, Z.; Latifi, S.; Akbari, M.; et al. Biodegradable elastic nanofibrous platforms with integrated flexible heaters for on-demand drug delivery. Sci. Rep. 2017, 7, 9220. [Google Scholar] [CrossRef] [Scilit]
  99. Zhang, Y.; Lee, G.; Li, S.; Hu, Z.; Zhao, K.; Rogers, J.A. Advances in Bioresorbable Materials and Electronics. Chem. Rev. 2023, 123, 11722–11773. [Google Scholar] [CrossRef] [Scilit]
  100. Hwangbo, J.; Seo, H.; Sim, G.; Avila, R.; Nair, M.; Kim, B.; Choi, Y. Bioresorbable polymers for electronic medicine. Cell Rep. Phys. Sci. 2024, 5, 102099. [Google Scholar] [CrossRef] [Scilit]
  101. Abdel-Mottaleb, M.M.A.; Lunter, D.J. Bioabsorbable medical devices. In Medical Devices in Modern Healthcare; Jana, S., Manna, S., Eds.; Academic Press: London, UK, 2026; Chapter 7; pp. 287–307. [Google Scholar] [CrossRef] [Scilit]
  102. Park, H.; Kim, J.-U.; Kim, S.; Hwang, N.S.; Kim, H.D. Sprayable Ti3C2 MXene hydrogel for wound healing and drug release system. Mater. Today Bio 2023, 23, 100881. [Google Scholar] [CrossRef] [Scilit]
  103. Liu, M.; Zheng, L.; Zha, K.; Yang, Y.; Hu, Y.; Chen, K.; Wang, F.; Zhang, K.; Liu, W.; Mi, B.; et al. Cu(II)@MXene based photothermal hydrogel with antioxidative and antibacterial properties for the infected wounds. Front. Bioeng. Biotechnol. 2023, 11, 1308184. [Google Scholar] [CrossRef] [Scilit]
  104. Ramirez-GarciaLuna, J.L.; Bartlett, R.; Arriaga-Caballero, J.E.; Fraser, R.D.J.; Saiko, G. Infrared Thermography in Wound Care, Surgery, and Sports Medicine: A Review. Front. Physiol. 2022, 13, 838528. [Google Scholar] [CrossRef] [Scilit]
  105. Collins, A.R.; O’Connor, G.M.; Ryan, D.A.; Parmeter, M.; Dinneen, S.; Gethin, G. Wound Bed Temperature has Potential to Indicate Infection Status: A Cross-Sectional Study. Wound Repair. Regen. 2025, 33, e70072. [Google Scholar] [CrossRef] [Scilit]
  106. Reyzelman, A.M.; Koelewyn, K.; Murphy, M.; Shen, X.; Yu, E.; Pillai, R.; Fu, J.; Scholten, H.J.; Ma, R. Continuous Temperature-Monitoring Socks for Home Use in Patients with Diabetes: Observational Study. J. Med. Internet Res. 2018, 20, e12460. [Google Scholar] [CrossRef] [Scilit]
  107. Shih, C.-D.; Scholten, H.J.; Ripp, G.; Srikanth, K.; Smith, C.; Ma, R.; Fu, J.; Reyzelman, A.M. Effectiveness of a Continuous Remote Temperature Monitoring Program to Reduce Foot Ulcers and Amputations: Multicenter Postmarket Registry Study. JMIR Diabetes 2024, 9, e46096. [Google Scholar] [CrossRef] [Scilit]
  108. Shirzaei Sani, E.; Xu, C.; Wang, C.; Song, Y.; Min, J.; Tu, J.; Solomon, S.A.; Li, J.; Banks, J.L.; Armstrong, D.G.; et al. A stretchable wireless wearable bioelectronic system for multiplexed monitoring and combination treatment of infected chronic wounds. Sci. Adv. 2023, 9, eadf7388. [Google Scholar] [CrossRef] [Scilit]
  109. Baniya, P.; Tebyani, M.; Hee, W.S.; Li, H.; Asefifeyzabadi, N.; Schorger, K.; Yang, H.-Y.; Gallegos, A.; Zhu, K.; Lu, F.; et al. Development of a wireless bioelectronic actuator for wound healing in a porcine model. Sci. Rep. 2025, 15, 39106. [Google Scholar] [CrossRef] [Scilit]
  110. Pang, Q.; Lou, D.; Li, S.; Wang, G.; Qiao, B.; Dong, S.; Ma, L.; Gao, C.; Wu, Z. Smart Flexible Electronics-Integrated Wound Dressing for Real-Time Monitoring and On-Demand Treatment of Infected Wounds. Adv. Sci. 2020, 7, 1902673. [Google Scholar] [CrossRef] [Scilit]
  111. Kornhaber, R.; Visentin, D.; West, S.; Haik, J.; Cleary, M. Burns Sustained from Body Heating Devices: An Integrative Review. Wounds 2020, 32, 123–133. Available online: https://www.hmpgloballearningnetwork.com/site/wounds/reviews/burns-sustained-body-heating-devices-integrative-review (accessed on 31 August 2026).
  112. Liu, Z.; Wang, X.; He, Y.; Hong, W.; Sun, P.; Liu, W.; Ye, D.; Yang, Z.; Wang, X.; Wu, M.; et al. Stretchable multifunctional wearable system for real-time and on-demand thermotherapy of arthritis. Microsyst. Nanoeng. 2025, 11, 84. [Google Scholar] [CrossRef] [Scilit]
  113. U.S. Food and Drug Administration. Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Heating and Cooling Devices; Center for Devices and Radiological Health: Rockville, MD, USA, 1995. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/guidance-document-preparation-premarket-notification-510k-applications-heating-and-cooling-devices (accessed on 31 August 2026).
  114. Kim, S.; Saito, M.; Wei, Y.; Bhuyan, P.; Choe, M.; Fujie, T.; Mondal, K.; Park, S. Stretchable and wearable polymeric heaters and strain sensors fabricated using liquid metals. Sens. Actuators A Phys. 2023, 355, 114317. [Google Scholar] [CrossRef] [Scilit]
  115. Choi, M.J.; Yoon, D.H.; Park, Y.S.; Nam, H.; Kim, G.H. Development of a Flexible and Conductive Heating Membrane via BSA-Assisted Electroless Plating on Electrospun PVDF-HFP Nanofibers. Appl. Sci. 2025, 15, 8023. [Google Scholar] [CrossRef] [Scilit]
  116. Chen, J.; Shi, Y.; Ying, B.; Hu, Y.; Gao, Y.; Luo, S.; Liu, X. Kirigami-enabled stretchable laser-induced graphene heaters for wearable thermotherapy. Mater. Horiz. 2024, 11, 2010–2020. [Google Scholar] [CrossRef] [Scilit]
  117. Yang, Y.; Fang, Q.; Shu, T.; Liu, W.; Sun, Y.; Liu, X.; Li, M.; Guo, B. Thermal-feedback modality switchable hydrogel with thermoelectric self-modulation for stage-adaptive repair of infected diabetic wound. Nat. Commun. 2026, 17, 9849. [Google Scholar] [CrossRef] [Scilit]
  118. Zhang, X.; Zhang, M.; Li, F.; Yang, W.; Wang, H.; Wang, Y.; Weng, J.; Li, J.; Chen, X. Engineered Long-Acting Mild Hyperthermia Microneedle Patches for Personalized Therapy of Infected Diabetic Wounds. ACS Appl. Mater. Interfaces 2026, 18, 46096–46111. [Google Scholar] [CrossRef] [Scilit]
  119. Zhao, X.; Liu, Z.; Zhang, X.; Li, X.; Xue, W.; Liu, Y.; Liu, K.; Wu, Z.; Lv, F. Non-phase change arylazopyrazole-containing alginate dressings with controlled heat release providing “hot spring” effect for wound healing. Int. J. Biol. Macromol. 2026, 376, 153539. [Google Scholar] [CrossRef] [Scilit]
  120. Vasyukova, I.A.; Zakharova, O.V.; Kuznetsov, D.V.; Gusev, A.A. Synthesis, Toxicity Assessment, Environmental and Biomedical Applications of MXenes: A Review. Nanomaterials 2022, 12, 1797. [Google Scholar] [CrossRef] [Scilit]
  121. Ramezani Farani, M.; Mirzaee, D.; Hatami, A.; Kumar, K.; Ghoreishian, S.M.; Huh, Y.S. Biocompatibility and immunomodulation of MXenes for targeted delivery of bioactive agents and drugs. Bioact. Mater. 2026, 55, 546–567. [Google Scholar] [CrossRef] [Scilit]
  122. Zha, K.; Zhang, W.; Hu, W.; Tan, M.; Zhang, S.; Yu, Y.; Gou, S.; Bu, P.; Zhou, B.; Zou, Y.; et al. Three-Step Regenerative Strategy: Multifunctional Bilayer Hydrogel for Combined Photothermal/Photodynamic Therapy to Promote Drug-Resistant Bacteria-Infected Wound Healing. Adv. Funct. Mater. 2024, 34, 2308145. [Google Scholar] [CrossRef] [Scilit]
  123. Ge, L.; Yuan, S.; Wang, X.; Li, Y.; Chen, D.; Wang, Y.; Wang, M.; Cao, Y.; Wu, Q. Multifunctional Ti3C2@PDA/CuO2 Nanoplatform for Synergistic Photothermal, Photodynamic and Photothermal-Enhanced Chemodynamic Antibacterial Therapy. ACS Appl. Nano Mater. 2024, 7, 27387–27399. [Google Scholar] [CrossRef] [Scilit]
  124. An, J.; Zhao, H.; Jia, Z.; Zhao, C.; Cui, C.; Meng, F.; Sheng, L.; Wen, M.; Zheng, Y.; Xi, T. Design and Engineering of Photocatalytic Graphitic Carbon Nitride for Environmental and Biological Disinfection. ACS ES&T Eng. 2025, 5, 2138–2169. [Google Scholar] [CrossRef] [Scilit]
  125. Ge, Z.; Guo, W.; Tao, Y.; Sun, H.; Meng, X.; Cao, L.; Zhang, S.; Liu, W.; Akhtar, M.L.; Li, Y.; et al. Wireless and Closed-Loop Smart Dressing for Exudate Management and On-Demand Treatment of Chronic Wounds. Adv. Mater. 2023, 35, 2304005. [Google Scholar] [CrossRef] [Scilit]
  126. Park, M.; Yoo, J.-Y.; Yang, T.; Jung, Y.H.; Vázquez-Guardado, A.; Li, S.; Kim, J.-H.; Shin, J.; Maeng, W.-Y.; Lee, G.; et al. Skin-integrated systems for power efficient, programmable thermal sensations across large body areas. Proc. Natl. Acad. Sci. USA 2023, 120, e2217828120. [Google Scholar] [CrossRef] [Scilit]
  127. Alberts, A.; Tudorache, D.-I.; Niculescu, A.-G.; Grumezescu, A.M. Advancements in Wound Dressing Materials: Highlighting Recent Progress in Hydrogels, Foams, and Antimicrobial Dressings. Gels 2025, 11, 123. [Google Scholar] [CrossRef] [Scilit]
  128. Khadeeja Thanha, K.; Ayisha Sana, P.; Ajesh, J.; Naseef, P.; Tharayil, H.; Lubaib, P.; Pramod, K. Advanced smart bioelectronics for wound healing: Biosensing, drug delivery, and artificial intelligence. Int. J. Pharm. 2025, 684, 126098. [Google Scholar] [CrossRef] [Scilit]
  129. Zhao, X.; Xu, H.; Sun, Y.; Yang, Y.; Guo, B. Self-Adaptive Wound Dressings for Wound Healing and Repair. Adv. Mater. 2026, 38, e15854. [Google Scholar] [CrossRef] [Scilit]
  130. European Parliament; Council of the European Union. Regulation (EU) 2017/745 of 5 April 2017 on Medical Devices. Off. J. Eur. Union 2017, L117, 1–175. Available online: https://eur-lex.europa.eu/eli/reg/2017/745/oj/eng (accessed on 31 August 2026).
  131. Medical Device Coordination Group. MDCG 2021-24: Guidance on Classification of Medical Devices; European Commission: Brussels, Belgium, 2021; Available online: https://health.ec.europa.eu/latest-updates/mdcg-2021-24-guidance-classification-medical-devices-2021-10-04_en (accessed on 31 August 2026).
  132. Singh, M.V.; Apshingekar, P.; Gandhi, S.; Singh, O.V. US regulatory compliance for medical combination products: An overview. Front. Med. Technol. 2024, 6, 1486318. [Google Scholar] [CrossRef] [Scilit]
  133. U.S. Food and Drug Administration. De Novo Classification Request. Available online: https://www.fda.gov/medical-devices/premarket-submissions-selecting-and-preparing-correct-submission/de-novo-classification-request (accessed on 31 August 2026).
  134. International Electrotechnical Commission. IEC 60601-1:2005+AMD1:2012+AMD2:2020 CSV, Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential Performance; IEC: Geneva, Switzerland, 2020. [Google Scholar]
  135. International Electrotechnical Commission. IEC 60601-1-10:2007+AMD1:2013+AMD2:2020 CSV, Medical Electrical Equipment—Part 1-10: General Requirements for Basic Safety and Essential Performance—Collateral Standard: Requirements for the Development of Physiologic Closed-Loop Controllers; IEC: Geneva, Switzerland, 2020; Available online: https://webstore.iec.ch/en/publication/67383 (accessed on 31 August 2026).
  136. International Electrotechnical Commission. IEC 60601-1-2:2014+AMD1:2020 CSV, Medical Electrical Equipment—Part 1-2: General Requirements for Basic Safety and Essential Performance—Collateral Standard: Electromagnetic Disturbances—Requirements and Tests; IEC: Geneva, Switzerland, 2020; Available online: https://webstore.iec.ch/en/publication/67554 (accessed on 31 August 2026).
  137. U.S. Food and Drug Administration. Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions; Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2026. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/cybersecurity-medical-devices-quality-management-system-considerations-and-content-premarket (accessed on 31 August 2026).
  138. International Electrotechnical Commission. IEC 81001-5-1:2021, Health Software and Health IT Systems Safety, Effectiveness and Security—Part 5-1: Security—Activities in the Product Life Cycle; IEC: Geneva, Switzerland, 2021. [Google Scholar]
  139. U.S. Food and Drug Administration. Applying Human Factors and Usability Engineering to Medical Devices; Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2016. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/applying-human-factors-and-usability-engineering-medical-devices (accessed on 31 August 2026).
  140. Beach, C.; Cooper, G.; Weightman, A.; Hodson-Tole, E.F.; Reeves, N.D.; Casson, A.J. Monitoring of Dynamic Plantar Foot Temperatures in Diabetes with Personalised 3D-Printed Wearables. Sensors 2021, 21, 1717. [Google Scholar] [CrossRef] [Scilit]
  141. Sanchez-Jimenez, J.L.; Priego-Quesada, J.I.; Gisbert-Ruiz, M.J.; Cibrian-Ortiz de Anda, R.M.; Perez-Soriano, P.; Aparicio, I. Effect of Compression Tights on Skin Temperature in Women with Lipedema. Appl. Sci. 2023, 13, 1133. [Google Scholar] [CrossRef] [Scilit]
  142. Lee, B.W.; Kang, J.; Yoo, J.-Y.; Won, S.M. Textile electronic systems for therapeutic applications. npj Flex. Electron. 2025, 9, 118. [Google Scholar] [CrossRef] [Scilit]
  143. Nemomssa, H.D.; Bossuyt, F.; Vandecasteele, B.; De Pauw, H.; Gidi, N.W.; Bauwens, P. Revolutionizing Patient Care: A Comprehensive Review of Recent Advances in Flexible Printed Heaters for Wearable Medical Applications. Actuators 2025, 14, 1. [Google Scholar] [CrossRef] [Scilit]
  144. Pollini, M.; Striani, R.; Paladini, F.; Kiani, A.; Acocella, M.R.; Esposito Corcione, C. Nanotechnological Antibacterial and Conductive Wound Dressings for Pressure Ulcer Prevention. Nanomaterials 2024, 14, 1309. [Google Scholar] [CrossRef] [Scilit]
  145. U.S. Food and Drug Administration. Evaluation of Thermal Effects of Medical Devices That Produce Tissue Heating and/or Cooling; Draft Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2024. Available online: https://www.fda.gov/media/177004/download (accessed on 31 August 2026).
  146. International Electrotechnical Commission. IEC 62366-1:2015+AMD1:2020 CSV, Medical Devices—Part 1: Application of Usability Engineering to Medical Devices; IEC: Geneva, Switzerland, 2020; Available online: https://webstore.iec.ch/en/publication/67220 (accessed on 31 August 2026).
  147. U.S. Food and Drug Administration. Whele LLC Announces National Voluntary Recall of Mighty Bliss Electric Heating Pad Due to Product Safety Concerns. 24 October 2022. Available online: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/whele-llc-announces-national-voluntary-recall-mighty-bliss-electric-heating-pad-due-product-safety (accessed on 31 August 2026).
  148. U.S. Food and Drug Administration. Postmarket Management of Cybersecurity in Medical Devices; Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2016. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/postmarket-management-cybersecurity-medical-devices (accessed on 31 August 2026).
  149. International Medical Device Regulators Forum. Principles and Practices for Medical Device Cybersecurity; IMDRF/CYBER WG/N60FINAL:2020; IMDRF. 2020. Available online: https://www.imdrf.org/sites/default/files/docs/imdrf/final/technical/imdrf-tech-200318-pp-mdc-n60.pdf (accessed on 31 August 2026).
  150. ISO 13485:2016; Medical Devices—Quality Management Systems—Requirements for Regulatory Purposes. International Organization for Standardization (ISO): Geneva, Switzerland, 2016.
  151. U.S. Food and Drug Administration. Quality Management System Regulation (QMSR). Available online: https://www.fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr (accessed on 31 August 2026).
  152. U.S. Government Publishing Office. 21 CFR Part 820—Quality Management System Regulation. Available online: https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (accessed on 31 August 2026).
  153. Leenstra, B.; Wijnand, J.; Verhoeven, B.; Koning, O.; Teraa, M.; Verhaar, M.C.; de Borst, G.J. Applicability of Transcutaneous Oxygen Tension Measurement in the Assessment of Chronic Limb-Threatening Ischemia. Angiology 2020, 71, 208–216. [Google Scholar] [CrossRef] [Scilit]
  154. Oropallo, A.; Ortega-Loayza, A.G.; Korzendorfer, H.; James, F.; Dotson, P.; Khimchenko, A.; Driver, V.R.; Sonenblum, S.E. Advancing Chronic Wound Care with Near-Infrared Spectroscopy Imaging: Clinical Applications, Measurement Parameters, and Insights into Healing Dynamics. Wounds 2025, 37, 384–392. Available online: https://pubmed.ncbi.nlm.nih.gov/41270199/ (accessed on 31 August 2026).
  155. Potvin, S.; Vincenot, M.; Haroune, L.; Giguère, C.-É.; Gendron, L.; Léonard, G.; Marchand, S. Data-driven dynamic profiles of tonic heat pain perception in pain-free volunteers are associated with differences in anandamide levels. Sci. Rep. 2024, 14, 17238. [Google Scholar] [CrossRef] [Scilit]
  156. Hoffmann, T.C.; Glasziou, P.P.; Boutron, I.; Milne, R.; Perera, R.; Moher, D.; Altman, D.G.; Barbour, V.; Macdonald, H.; Johnston, M.; et al. Better reporting of interventions: Template for intervention description and replication (TIDieR) checklist and guide. BMJ 2014, 348, g1687. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  157. Schulz, K.F.; Altman, D.G.; Moher, D.; for the CONSORT Group. CONSORT 2010 Statement: Updated guidelines for reporting parallel group randomised trials. BMJ 2010, 340, c332. [Google Scholar] [CrossRef] [Scilit]
  158. Chan, A.-W.; Tetzlaff, J.M.; Altman, D.G.; Laupacis, A.; Gøtzsche, P.C.; Krleža-Jerić, K.; Hróbjartsson, A.; Mann, H.; Dickersin, K.; Berlin, J.A.; et al. SPIRIT 2013 Statement: Defining Standard Protocol Items for Clinical Trials. Ann. Intern. Med. 2013, 158, 200–207. [Google Scholar] [CrossRef] [Scilit]
  159. Schaper, N.C.; van Netten, J.J.; Apelqvist, J.; Bus, S.A.; Fitridge, R.; Game, F.; Monteiro-Soares, M.; Senneville, E.; the IWGDF Editorial Board. Practical guidelines on the prevention and management of diabetes-related foot disease (IWGDF 2023 update). Diabetes Metab. Res. Rev. 2024, 40, e3657. [Google Scholar] [CrossRef] [Scilit]
  160. The Chinese Burn Association, the Yangtze River Delta Integrated Diabetic Foot Alliance, and the Editorial Committee of the Chinese Journal of Burns and Wound Repair; Luo, G.; Liu, Y.; Wang, A. Practical guidelines for the prevention and management of diabetic foot disease in China. Burns Trauma. 2025, 13, tkaf064. [Google Scholar] [CrossRef] [Scilit]
  161. Chang, W.-Y.; Liu, H.-H.; Huang, D.-W.; Chou, Y.-Y.; Ou, K.-L.; Wang, C.-H.; Dai, N.-T.; Tzeng, Y.-S. Severe burn injury from the common Asian practice of heat application in patients with diabetic neuropathy. Int. Wound J. 2022, 19, 845–852. [Google Scholar] [CrossRef] [Scilit]
  162. Nerone, V.S.; Springer, K.D.; Atway, S.A. Diabetic Foot Burns: A Case Series. J. Foot Ankle Surg. 2014, 53, 453–455. [Google Scholar] [CrossRef] [Scilit]
  163. Parshakov, A.; Zubareva, N.; Podtaev, S.; Frick, P. Detection of Endothelial Dysfunction Using Skin Temperature Oscillations Analysis During Local Heating in Patients with Peripheral Arterial Disease. Microcirculation 2016, 23, 406–415. [Google Scholar] [CrossRef] [Scilit]
  164. Mosti, G.; Iabichella, M.L.; Partsch, H. Compression therapy in mixed ulcers increases venous output and arterial perfusion. J. Vasc. Surg. 2012, 55, 122–128. [Google Scholar] [CrossRef] [Scilit]
  165. Gould, L.J.; Alderden, J.; Aslam, R.; Barbul, A.; Bogie, K.M.; El Masry, M.; Graves, L.Y.; White-Chu, E.F.; Ahmed, A.; Boanca, K.; et al. WHS guidelines for the treatment of pressure ulcers-2023 update. Wound Repair. Regen. 2024, 32, 6–33. [Google Scholar] [CrossRef] [Scilit]
  166. International Working Group on the Diabetic Foot. IWGDF Guidelines (2023 Update). Available online: https://iwgdfguidelines.org/guidelines-2023/ (accessed on 31 August 2026).
  167. Lohr, J.M.; Raffetto, J.D.; Dexter, D.J.; Regulski, M.J.; Edmonds, M.E.; Ozsvath, K.J.; Blakely, M.M. A synergistic multimodality treatment approach to address the key drivers of wound chronicity. J. Vasc. Surg. Venous Lymphat. Disord. 2026, 14, 102348. [Google Scholar] [CrossRef] [Scilit]
  168. Norton, L.; Parslow, N.; Ho, C.; O’Sullivan-Drombolis, D.; Rogers, A.; Parsons, A.; Moss, L. Best practice recommendations for the prevention and management of pressure injuries. In Best Practice Recommendations for Skin Health and Wound Management 2025; Kuhnke, J.L., Burrows, C.A., Evans, R.M., Orsted, H.L., Rosenthal, S., Eds.; Wounds Canada: Toronto, ON, Canada, 2025; Available online: https://www.woundscanada.ca/docman/public/health-care-professional/bpr-2025/3515-7-bpr-chapter-7-2025/file (accessed on 31 August 2026). [CrossRef] [Scilit]
  169. Venous Leg Ulcer Guideline Committee. Clinical Practice Guideline for Venous Leg Ulcers: Assessment, 2nd ed.; Haesler, E., Ed.; Hong Kong Enterostomal Therapists Association; New Zealand Wound Care Society; Wounds Australia; Wound Healing Society of Singapore: Singapore, 2025; ISBN 978-0-6489801-1-7. Available online: https://d122d2wjqead0l.cloudfront.net/AP/Int/woundsaus/uploads/Publications/venous%20leg%20ulcer%20assessment%20section-2025.pdf (accessed on 31 August 2026).
  170. Keast, D.; Angel, D.; Weir, D.; Haesler, E.; Ousey, K.; Carville, K.; Waters, N.; Idensohn, P.; Swanson, T.; Bjarnsholt, T. Wound Infection in Clinical Practice: Principles of Best Practice, 3rd ed.; Wounds International: London, UK, 2022; Available online: https://woundsinternational.com/consensus-documents/wound-infection-in-clinical-practice-principles-of-best-practice/ (accessed on 31 August 2026).
  171. Maranna, H.; Lal, P.; Mishra, A.; Bains, L.; Sawant, G.; Bhatia, R.; Kumar, P.; Beg, M.Y. Negative pressure wound therapy in grade 1 and 2 diabetic foot ulcers: A randomized controlled study. Diabetes Metab. Syndr. 2021, 15, 365–371. [Google Scholar] [CrossRef] [Scilit]
  172. Huang, Y.; Yu, Z.; Xu, M.; Zhao, X.; Tang, Y.; Luo, L.; Deng, D.; Chen, M. Negative pressure wound therapy promotes wound healing by down-regulating miR-155 expression in granulation tissue of diabetic foot ulcers. Sci. Rep. 2025, 15, 6733. [Google Scholar] [CrossRef] [Scilit]
  173. Srivastava, V.; Meena, R.N.; Pratap, A.; Verma, A.K.; Ansari, M.A.; Mishra, S.P. Effect of negative pressure wound therapy on wound thermometry in diabetic foot ulcers. J. Fam. Med. Prim. Care 2022, 11, 7001–7007. [Google Scholar] [CrossRef] [Scilit]
  174. Pillai, P.S.; Islam, M.A.; Rathour, R.; Kumar, B.; Das, A.; Alagirusamy, R. Optimizing the pressure change of compression bandage due to temperature and moisture using response surface methodology. Prog. Eng. Sci. 2025, 2, 100118. [Google Scholar] [CrossRef] [Scilit]
  175. Park, Y.; Kwon, K.; Kwak, S.S.; Yang, D.S.; Kwak, J.W.; Luan, H.; Chung, T.S.; Chun, K.S.; Kim, J.U.; Jang, H.; et al. Wireless, skin-interfaced sensors for compression therapy. Sci. Adv. 2020, 6, eabe1655. [Google Scholar] [CrossRef] [Scilit]
  176. McGuiness, W.; Vella, E.; Harrison, D. Influence of dressing changes on wound temperature. J. Wound Care 2004, 13, 383–385. [Google Scholar] [CrossRef] [Scilit]
  177. Gefen, A.; Fisk, J.; Bunker, J.; Bagshaw, L.E. Moisture-Responsive Thermal Conductivity Properties of Hydrofiber Versus Polyurethane Foam: Implications for Pressure Injury Prevention. Int. Wound J. 2026, 23, e70849. [Google Scholar] [CrossRef] [Scilit]
  178. Huang, Y.; Song, M.; Li, X.; Du, Y.; Gao, Z.; Zhao, Y.-Q.; Li, C.; Yan, H.; Mo, X.; Wang, C.; et al. Temperature-responsive self-contraction nanofiber/hydrogel composite dressing facilitates the healing of diabetic-infected wounds. Mater. Today Bio 2024, 28, 101214. [Google Scholar] [CrossRef] [Scilit]
  179. Diao, J.; Meng, Y.; Wang, L.; Zhao, G.; Sun, B.; Du, S.; Niu, Y.; He, G.; Yu, M.; Jiang, X. Temperature/pH Co-responsive CS/NIPAM-Based hydrogel with controlled release of Ga3+ for improved infected wound healing. Chem. Eng. Sci. 2024, 294, 120104. [Google Scholar] [CrossRef] [Scilit]
  180. Matijevich, E.; Minty, E.; Bray, E.; Bachus, C.; Hajizadeh, M.; Liden, B. A Multi-Faceted Digital Health Solution for Monitoring and Managing Diabetic Foot Ulcer Risk: A Case Series. Sensors 2024, 24, 2675. [Google Scholar] [CrossRef] [Scilit]
  181. Li, M.; Wang, C.; Yu, Q.; Chen, H.; Ma, Y.; Wei, L.; Wu, M.X.; Yao, M.; Lu, M. A wearable and stretchable dual-wavelength LED device for home care of chronic infected wounds. Nat. Commun. 2024, 15, 9380. [Google Scholar] [CrossRef] [Scilit]
  182. Littman, A.J.; Timmons, A.K.; Korpak, A.; Chan, K.C.G.; Jones, K.T.; Shirley, S.; Nordrum, K.; Robbins, J.; Masadeh, S.; Moy, E. Remote Foot Temperature Monitoring Among Veterans: Large Observational Study of Noncompliance and Its Correlates. JMIR Diabetes 2024, 9, e53083. [Google Scholar] [CrossRef] [Scilit]
  183. Rovers, F.; Van Netten, J.; Busch-Westbroek, T.; Aan de Stegge, W.; Bus, S. Adherence to at-Home Monitoring of Foot Temperatures in People with Diabetes at High Risk of Ulceration. Int. J. Low. Extrem. Wounds 2025, 24, 691–699. [Google Scholar] [CrossRef] [Scilit]
  184. Ferreira, R.G.; Silva, A.P.; Nunes-Pereira, J. Current On-Skin Flexible Sensors, Materials, Manufacturing Approaches, and Study Trends for Health Monitoring: A Review. ACS Sens. 2024, 9, 1104–1133. [Google Scholar] [CrossRef] [Scilit]
  185. Babaei, N.; Zamanzadeh, V.; Valizadeh, L.; Lotfi, M.; Avazeh, M. Care needs of patients with chronic wounds for implementing a virtual care program: A qualitative study. PLoS ONE 2025, 20, e0339895. [Google Scholar] [CrossRef] [Scilit]
  186. U.S. Food and Drug Administration. Human Factors and Medical Devices. Available online: https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/human-factors-and-medical-devices (accessed on 31 August 2026).
  187. Liu, H.; Yan, X.-Y.; Li, G.-Q.; Wang, B.-N.; Wang, D.; Zhang, Y.-H.; Guo, J.-L. Evaluation of wound temperature monitoring at various anatomical sites in the management of patients with diabetic foot undergoing microcirculation reconstruction. J. Orthop. Surg. Res. 2024, 19, 776. [Google Scholar] [CrossRef] [Scilit]
  188. Maniotis, N.; Mitropoulos, S.; Vordos, N.; Tsiantos, V. The Role of Blood Perfusion in the Thermal Interaction Between Magnetic Nanoparticles and Cancerous Tumors: A Computational Study. Magnetism 2025, 5, 6. [Google Scholar] [CrossRef] [Scilit]
  189. Lapitan, D.; Xing, X.; Glazkov, A.; Krasulina, K.; Kovaleva, Y.; Glazkova, P.; Britvin, T.; Zagarov, S.; Larkov, R.; Rogatkin, D. Diabetes-Associated Changes in the Toe Photoplethysmogram During Local Heating Test. J. Biophotonics 2025, 18, e202500190. [Google Scholar] [CrossRef] [Scilit]
  190. Koïtka, A.; Abraham, P.; Bouhanick, B.; Sigaudo-Roussel, D.; Demiot, C.; Saumet, J.L. Impaired Pressure-Induced Vasodilation at the Foot in Young Adults with Type 1 Diabetes. Diabetes 2004, 53, 721–725. [Google Scholar] [CrossRef] [Scilit]
  191. Huang, X.; Zhang, Q.; Yang, Y.; Chow, L.; Ma, J.; Xu, G.; Guo, F.; He, X.; Li, Z.; Zhao, G.; et al. A skin-interfaced three-dimensional closed-loop sensing and therapeutic electronic wound bandage. Nat. Commun. 2025, 16, 5782. [Google Scholar] [CrossRef] [Scilit]
  192. Park, Y.; Jung, J.; Lee, Y.; Lee, D.; Vlassak, J.J.; Park, Y.-L. Liquid-metal micro-networks with strain-induced conductivity for soft electronics and robotic skin. npj Flex. Electron. 2022, 6, 81. [Google Scholar] [CrossRef] [Scilit]
  193. Zhu, Y.; Shi, Z.; Zhao, Y.; Bu, S.; Hu, Z.; Liao, J.; Lu, Q.; Zhou, C.; Guo, B.; Shang, M.; et al. Recent trends in the transfer of graphene films. Nanoscale 2024, 16, 7862–7873. [Google Scholar] [CrossRef] [Scilit]
  194. Cai, C.; Li, W.; Zhang, X.; Cheng, B.; Chen, S.; Zhang, Y. Natural Polymer-Based Hydrogel Dressings for Wound Healing. Adv. Wound Care 2025, 14, 295–322. [Google Scholar] [CrossRef] [Scilit]
  195. Feng, F.; Zhao, Z.; Li, J.; Huang, Y.; Chen, W. Multifunctional dressings for wound exudate management. Prog. Mater. Sci. 2024, 146, 101328. [Google Scholar] [CrossRef] [Scilit]
  196. Nygren, E.; Malone, M.; Gefen, A. Laboratory Evaluations of Wound Dressings: Key Advances to Reflect Clinical Reality. Int. Wound J. 2025, 22, e70333. [Google Scholar] [CrossRef] [Scilit]
  197. Su, R.; Wang, L.; Han, F.; Bian, S.; Meng, F.; Qi, W.; Zhai, X.; Li, H.; Wu, J.; Pan, X.; et al. A highly stretchable smart dressing for wound infection monitoring and treatment. Mater. Today Bio 2024, 26, 101107. [Google Scholar] [CrossRef] [Scilit]
  198. Guzman, S.; Call, K.; Russon, M.; Jellum, S.; Fisk, J.; Call, E. Evaluation of Shear Force Redistribution and Microclimate in Foam Dressings Indicated for Pressure Injury Prevention. Adv. Skin Wound Care 2025, 38, 183–188. [Google Scholar] [CrossRef] [Scilit]
  199. Goshisht, M.K.; Goshisht, A.; Bajpai, A.; Bajpai, A. Recent advances in biomedical applications of smart nanomaterials: A comprehensive review. RSC Pharm. 2025, 2, 1227–1267. [Google Scholar] [CrossRef] [Scilit]
  200. Raepsaet, C.; Alves, P.; Cullen, B.; Gefen, A.; Lázaro-Martínez, J.L.; Lev-Tov, H.; Najafi, B.; Santamaria, N.; Sharpe, A.; Swanson, T.; et al. The development of a core outcome set for clinical effectiveness studies of bordered foam dressings in the treatment of complex wounds. J. Tissue Viability 2023, 32, 430–436. [Google Scholar] [CrossRef] [Scilit]
  201. Hallas, S.; Nelson, E.A.; O’Meara, S.; Gethin, G.; Adderley, U.; Meskell, P.; Nixon, J.; O’Loughlin, A.; Probst, S.; Tawfick, W.; et al. Development of a core outcome set for use in research evaluations of interventions for venous leg ulceration: International eDelphi consensus. J. Tissue Viability 2024, 33, 324–331. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  202. Lechner, A.; Kottner, J.; Coleman, S.; Muir, D.; Beeckman, D.; Chaboyer, W.; Cuddigan, J.; Moore, Z.; Rutherford, C.; Schmitt, J.; et al. Outcomes for Pressure Ulcer Trials (OUTPUTs) project: Review and classification of outcomes reported in pressure ulcer prevention research. Br. J. Dermatol. 2021, 184, 617–626. [Google Scholar] [CrossRef] [Scilit]
  203. Redmond, M.C.; Gethin, G.; Finn, D.P. A Review of Chronic Wounds and Their Impact on Negative Affect, Cognition, and Quality of Life. Int. Wound J. 2025, 22, e70748. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  204. Khan, S.; Banu, S.A.; Mamachan, M.; Vinodhkumar, O.R.; Manjusha, K.M.; Kumar, R.; Pawde, A.M.; Dhama, K.; Pal, A. Comparative assessment of smartphone-based digital planimetry for wound area measurement. Narra J. 2025, 5, e1987. [Google Scholar] [CrossRef] [Scilit]
  205. Casanova-Lozano, L.; Reifs-Jiménez, D.; Martí-Ejarque, M.; Reig-Bolaño, R.; Grau-Carrión, S. Evaluation of Two Digital Wound Area Measurement Methods Using a Non-Randomized, Single-Center, Controlled Clinical Trial. Electronics 2024, 13, 2390. [Google Scholar] [CrossRef] [Scilit]
  206. U.S. Food and Drug Administration. Technical Considerations for Medical Devices with Physiologic Closed-Loop Control Technology; Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2023. Available online: https://www.fda.gov/media/154994/download (accessed on 31 August 2026).
  207. Li, H.; Yang, H.-Y.; Lu, F.; Hee, W.S.; Asefifeyzabadi, N.; Baniya, P.; Gallegos, A.; Schorger, K.; Zhu, K.; Recendez, C.; et al. Towards adaptive bioelectronic wound therapy with integrated real-time diagnostics and machine learning-driven closed-loop control. npj Biomed. Innov. 2025, 2, 31. [Google Scholar] [CrossRef] [Scilit]
  208. U.S. Food and Drug Administration. Predetermined Change Control Plans for Medical Devices; Draft Guidance for Industry and FDA Staff; FDA: Silver Spring, MD, USA, August 2024. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/predetermined-change-control-plans-medical-devices (accessed on 31 August 2026).
  209. U.S. Food and Drug Administration. Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations; Draft Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2025. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/artificial-intelligence-enabled-device-software-functions-lifecycle-management-and-marketing (accessed on 31 August 2026).
  210. U.S. Food and Drug Administration. Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions; Guidance for Industry and Food and Drug Administration Staff; FDA: Silver Spring, MD, USA, 2025. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence (accessed on 31 August 2026).
  211. Chan, A.-W.; Tetzlaff, J.M.; Gøtzsche, P.C.; Altman, D.G.; Mann, H.; Berlin, J.A.; Dickersin, K.; Hróbjartsson, A.; Schulz, K.F.; Parulekar, W.R.; et al. SPIRIT 2013 explanation and elaboration: Guidance for protocols of clinical trials. BMJ 2013, 346, e7586. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.