Abstract
Background/Objectives: Conventional diagnosis of inborn errors of metabolism (IEMs) requires multiple specimen types—urine organic acids, plasma amino acids, and serum acylcarnitines—analyzed on distinct analytical platforms. This multi-assay approach is labor-intensive and limits timely clinical decision making. We aimed to develop a fully automated serum-based LC–MS/MS platform for integrated quantitative metabolite profiling and to establish pediatric reference intervals (RIs) to support diagnostic interpretation. Methods: A fully automated LC–MS/MS system integrated with the CLAM-2030 automated pretreatment module was developed to enable simultaneous quantification of 25 organic acids, 8 amino acids, and 21 acylcarnitines. Analytical performance was assessed for linearity, limits of detection and quantification, precision and accuracy. Serum samples from 296 non-IEM children aged 0–6 years were analyzed to establish pediatric RIs using Box–Cox transformation and Gaussian modeling. Clinical utility was evaluated in sera from 89 patients diagnosed with IEM using z-score-based logistic regression models. Results: The method demonstrated excellent performance, with linearity (r2 > 0.99) across calibration ranges, limits of detection and quantification defined by signal-to-noise ratios > 3 and >10, and intra- and inter-assay precision < 15% CV for all 54 analytes. Twenty-one analytes met the acceptance criterion of ±20% accuracy at all quality-control levels. Pediatric RIs provided a quantitative framework for interpreting the metabolic abnormalities. In IEM patients, disease-specific metabolites were consistently outside the established ranges, and z-score-based logistic regression models successfully distinguished major IEM categories, including organic acidemias and long-chain fatty acid oxidation disorders. Conclusions: This fully automated, serum-based LC–MS/MS platform provides a clinically practical and quantitative framework for integrated metabolic profiling using pediatric RIs, supporting diagnosis and monitoring of IEMs in pediatric settings.
1. Introduction
Inborn errors of metabolism (IEMs) are genetic disorders that disrupt intermediary metabolic pathways, leading to the accumulation or deficiency of specific metabolites [1]. Collectively, IEMs affect approximately 1 in 2000 newborns worldwide, and their early and accurate evaluation is critical for clinical management, prognosis, and therapeutic monitoring [2,3].
Conventionally, differential diagnosis of IEMs requires the combination of analyses with multiple specimen types—urinary organic acids [4,5,6,7], plasma amino acids [8,9,10], and serum acylcarnitines [11]—using distinct analytical platforms. While this multi-specimen approach remains the diagnostic standard, it is labor-intensive, time-consuming, and unsuitable for rapid or repeated monitoring in clinical settings. The need for multiple sample collections can be particularly challenging in neonates and infants, where specimen volume and timing are often limiting factors [1,12,13].
Recent advances in liquid chromatography-tandem mass spectrometry (LC-MS/MS) have enabled simultaneous quantification of structurally diverse metabolites with high sensitivity and specificity [14]. Building on this, our previous study demonstrated that a fully automated LC–MS/MS system integrating the CLAM-2030 pretreatment module could quantify organic acids in serum samples within one hour, providing a clinically feasible alternative to gas chromatography-mass spectrometry [15]. Extending this approach to include amino acids and acylcarnitines within a single serum-based analytical framework would enable a comprehensive assessment of mitochondrial, amino acid and fatty acid oxidation pathways—key domains affected in many organic acidemias and fatty acid oxidation disorders.
Establishing pediatric reference intervals (RIs) for these serum metabolites is essential to translate such multiplex measurements into quantitative clinical interpretation. Pediatric metabolic profiles differ markedly from those of adults because of developmental changes in enzyme activity, nutritional status, and organ maturation [16]. Defining age-specific RIs, particularly in early children (0–6 years), allows objective quantification of metabolic deviations and facilitates longitudinal evaluation of disease activity or treatment response in patients with IEMs [17,18].
Moreover, the accumulation of large-scale, standardized metabolic data generated by such a fully automated platform provides an opportunity for data-driven diagnostic support. Quantitative deviations from pediatric RIs, when expressed as standardized z-scores, can be analyzed using machine-learning algorithms to identify disease-specific metabolic patterns. This integration of automation and computational modeling has the potential to transform metabolite profiling from qualitative pattern recognition to quantitative, reproducible diagnosis and classification of IEMs.
This study was designed to support diagnostic differentiation of inborn errors of metabolism included in current newborn screening programs in Japan (Supplemental Table S7).
In this study, we report the development and validation of a fully automated LC–MS/MS platform capable of simultaneously quantifying serum organic acids, amino acids, and acylcarnitines. We further establish pediatric (0–6 years) serum RIs for 54 metabolites and evaluate their clinical utility in patients with confirmed IEMs. Finally, we propose that quantitative, data-driven approaches may further advance precision diagnostics for inherited metabolic diseases.
2. Materials and Methods
2.1. Reagents
Ultrapure water, formic acid, methanol, and acetonitrile (FUJIFILM Wako Pure Chemical Corporation, Osaka, Japan) were used. 3-Nitrophenylhydrazine (3-NPH) and N-(3-dimethylaminopropyl)-N′-ethylcarbodiimide (EDC) hydrochloride were obtained from Sigma-Aldrich (St. Louis, MO, USA). Analytical and internal standards are listed in Supplemental Table S1.
2.2. Stock Solutions and QC Samples
Internal standards followed Table S1 (amino acids/acylcarnitines from NeoSMAAT kit (Sekisui Medical Co., Tokyo, Japan). Other analytes were dissolved in water and stored at −80 °C. Quality-control (QC) sera at three concentration levels—low, medium, and high quality control (LQC/MQC/HQC)—were prepared by spiking pooled serum and stored at −80 °C.
2.3. Sample Collection
All procedures were approved by the Kyushu University Ethics Committee (No. 22138, approved on 28 November 2024).
Children without inborn errors of metabolism.
We collected 296 residual serum samples from children aged 0–6 years who visited the outpatient clinic of the Department of Pediatrics, Kyushu University Hospital.
Inclusion criteria: creatinine < 0.48 mg/dL; AST < 68 U/L; no visible hemolysis.
Exclusion criteria (prespecified):
- •
- Confirmed or suspected inborn errors of metabolism.
- •
- Acute-phase conditions or immediately post-infectious state.
- •
- Malignant disease.
- •
- Trisomy 21.
- •
- Ethnicity control: both parents self-identified as Japanese (to minimize inter-ethnic variability when establishing pediatric serum reference intervals).
- •
- Medications or therapies affecting metabolites: ongoing valproate treatment; carnitine supplementation; post-organ transplantation receiving immunosuppressive therapy; treatment with intravenous immunoglobulin preparations.
Age and sex distributions are summarized in Table S2.
Patients with IEMs.
Stored serum samples from 89 biochemically diagnosed, independent cases were included (Kyushu University, n = 5; Shimane University, n = 84), covering propionic acidemia (OMIM#606054), methylmalonic acidemia (OMIM#251000), isovaleric acidemia (OMIM#243500), 3-methylcrotonyl-CoA carboxylase deficiency (OMIM#210200), glutaric acidemia type II (OMIM#231680), medium-chain acyl-CoA dehydrogenase deficiency (OMIM#201450), very-long-chain acyl-CoA dehydrogenase deficiency (OMIM#201475), carnitine palmitoyltransferase II deficiency (OMIM#255110), mitochondrial trifunctional protein deficiency (OMIM#609015, #620300), phenylketonuria (OMIM#262600), and ornithine transcarbamylase deficiency (OMIM#311250). Case distribution is shown in Table S3.
2.4. Automated Sample Preparation
Protein precipitation and 3-NPH derivatization were performed on a CLAM™-2030 automated pretreatment system (Shimadzu, Kyoto, Japan) as previously described [15]. Pretreated extracts were analyzed on an LCMS™-8050 (Shimadzu, Kyoto, Japan) with a 5 µL injection volume.
2.5. LC–MS/MS Conditions
Chromatography was performed using a Shim-pack Scepter HD-C18-80 (150 × 2.1 mm, 3 µm) column (Shimadzu GLC, Kyoto, Japan). Mobile phases were (A) water with 0.1% formic acid and (B) acetonitrile with 0.1% formic acid, at 0.35 mL/min, with a 24-min gradient (16–70% B over 0–18 min, 95% B wash, and re-equilibration to 16% B). The column oven was maintained at 40 °C. Detection was performed by ESI-MRM with the following settings: nebulizing gas 3.0 L/min; drying gas 15.0 L/min; desolvation line 250 °C; heat block 400 °C; collision gas 230 kPa. Dwell times (22–163 ms) ensured ≥20 points per peak. MRM transitions for all 3-NPH-derivatized analytes are listed in Table S4. Four acylcarnitines (C16:1, C16OH, C18:1, C18:1OH) were empirically assigned using a mitochondrial trifunctional protein deficiency (OMIM #609015, #620300) serum, monitoring a characteristic product ion at m/z 220.1.
2.6. Method Validation
Linearity: Linearity was evaluated for 50 analytes, excluding four compounds without available reference standards (AC_C16:1, AC_C16OH, AC_C18:1, and AC_C18:1OH). Calibration samples were prepared by serial two-fold dilutions of the mixed standard solution to generate ten concentration levels. For analytes with internal standards, calibration curves were constructed from the relationship between the peak-area ratio (analyte/internal standard) and concentration; for those without internal standards, absolute peak areas were used. Linearity was considered acceptable when the coefficient of determination (r2) exceeded 0.99.
Sensitivity: Limit of detection (LOD) and quantification (LOQ) were defined at signal-to-noise ratios greater than 3 and 10, respectively, using five additional dilutions below the lowest calibrator.
Accuracy and Imprecision: For 34 analytes with internal standards, single-point calibration was applied with an accuracy acceptance was ±20%. Imprecision was evaluated at LQC/MQC/HQC (intra-/inter-day, n = 5 each), with CV < 15% as the criterion. For analytes without internal standards, variability was assessed from peak areas.
2.7. Data Analysis and Statistical Processing
All LC–MS/MS data were processed using LabSolutions software (version 5.97 SP1, Shimadzu Corporation, Kyoto, Japan). Peak areas were used as indicators of signal intensity. For pyruvic acid and 3-methylglutaconic acid, two peaks were observed and their summed values were used. Quantitative analytes below LOD were imputed with LOD values; others were set to a fixed area of 2000.
Reference intervals (RIs).
Pediatric RIs were established according to the harmonized protocol of the IFCC Committee on Reference Intervals and Decision Limits (C-RIDL) [19], which allows parametric or Box–Cox-transformed estimation after normalization and outlier exclusion. A parametric approach with Box–Cox transformation was used to approximate normality. After transformation, observations exceeding ±2.81 SD were excluded as outliers, and RI limits were derived by back-transforming ±1.96 SD to the original scale. When more than 2.5% of the control population fell below LOD, only the upper limit was reported. This C-RIDL-based procedure ensures international comparability and harmonization of RI estimates across institutions.
Z-score calculation and multivariate analysis.
Z-scores were computed after log10 transformation of quantitative values or peak areas, referenced against the mean and SD of the non-IEM children group (N = 296). For analytes with fewer than 100 valid data points, SD was adjusted so that a measurement corresponding to five times the LOD yielded a Z-score of 2.0. Multinomial logistic regression (MLR) models were constructed to evaluate whether diseases with related metabolic pathways could be discriminated from one another. The explanatory variables consisted of Z-scores representing standardized metabolite concentrations. For each disease category, a model was built using all metabolites known to be diagnostically relevant to that specific disorder, thereby including all established biomarker rather than restricting the model to only statistically significant variables. All modeling and performance evaluations were performed in R (version 4.4.0).
3. Results
3.1. Fully Automated Serum LC-MS/MS Platform
Building on our previous report, we refined a fully automated workflow integrating on-instrument sample preparation (CLAM-2030) with LC–MS/MS [15] to enable simultaneous quantification of 25 organic acids, 8 amino acids, and 21 acylcarnitines from a single serum specimen. The total analysis time was within 60 min per sample, supporting routine implementation in clinical laboratories. All analytes were derivatized with 3-nitrophenylhydrazine (3-NPH) and monitored in multiple reaction monitoring (MRM) mode. For four acylcarnitines (C16:1, C16OH, C18:1, and C18:1OH), MRM transitions were determined empirically using serum from a patient with mitochondrial trifunctional protein (TFP) deficiency, with precursor ions yielding the characteristic fragment at m/z 220.1. A complete list of transitions is provided in Table S4, and representative ion chromatograms are shown in Figure 1. Theses refinements extend the analytical scope from organic acids alone to a multidimensional metabolic panel encompassing mitochondrial, amino acid, and fatty acid oxidation pathways, enabling comprehensive metabolic profiling from serum.
Figure 1.
MRM chromatograms obtained using the developed LC-MS/MS method. (A) Chromatogram of organic acids analyzed using the standard solution. (B) Chromatogram of amino acids and acylcarnitines analyzed using the standard solution. For AC_C16:1, C16OH, C18:1, C18:1OH, standards were not available; therefore, retention times and MRM transitions were confirmed using a serum sample from a patient with mitochondrial trifunctional protein deficiency. Representative chromatograms shown here were obtained from normal serum.
3.2. Linearity, Sensitivity, Accuracy and Imprecision
Calibration curves were prepared for 50 analytes with reference standards. Acceptable linearity (r2 > 0.99) was confirmed across eight or more concentration levels for all analytes (Table 1). Limits of detection (LOD) and quantification (LOQ) were determined by serial dilutions (signal-to-noise ratios of >3 and >10, respectively). Accuracy was assessed for 34 analytes with internal standards using single-point calibration: 21 analytes showed deviations within ±20% at all quality-control levels. Imprecision was <15% coefficient of variation (CV) for all 54 analytes in both intra- and inter-day assessments, consistent with ICH M10 and FDA bioanalytical guidelines [20,21] (Table 2). These results confirm the analytical robustness and quantitative reliability of the automated LC-MS/MS platform for clinical application.
Table 1.
Linearity and sensitivity.
Table 2.
Accuracy and imprecision.
3.3. Pediatric Serum RIs (0–6 Years)
To enable quantitative interpretation of serum metabolite profiles, pediatric RIs were established for 54 analytes using serum samples from 296 non-IEM children aged 0–6 years. Descriptive statistics of this non-IEM cohort are provided in Table S5.
Reference intervals were calculated according to the statistical approach described in Section 2.7, following the harmonized IFCC Committee on Reference Intervals and Decision Limits (C-RIDL) protocol [19]. The resulting pediatric serum RIs are summarized in Table 3. For analytes in which more than 2.5% of the control population had concentrations below the limit of detection, only the upper reference limit is reported. This comprehensive dataset provides the first serum-based pediatric RI framework encompassing organic acids, amino acids, and acylcarnitines, thereby supporting quantitative evaluation and longitudinal monitoring of metabolic disorders. The established RIs also can form the foundation for data-driven disease classification models using machine-learning approaches.
Table 3.
Reference intervals for 54 serum analytes in children aged 0–6 years.
3.4. Clinical Validation Across Inborn Errors of Metabolism
Clinical performance was evaluated in 89 biochemically diagnosed cases encompassing multiple IEM categories. Compared with non-IEM children, each disease group exhibited at least one metabolite with a Z-score > 3.0, indicating marked deviation from the reference distribution (Figure 2). The altered profiles corresponded well to known disease pathophysiology. For example, glutaric acidemia type II (multiple acyl-CoA dehydrogenase deficiency, OMIM #231680) exhibited elevations of short- to long-chain acylcarnitines together with organic acids such as glutaric and 2-hydroxyglutaric acid [22,23,24]. While individual quantification of these organic acids in serum is available in select clinical laboratories, the present platform enables their simultaneous and quantitative assessment together with acylcarnitines within a single serum-based analytical workflow. These findings demonstrated that disease-specific metabolic signatures can be captured quantitatively from serum in a single analytical run, including metabolites that have traditionally required urine-based testing. To explore whether these quantitative serum data could further support diagnostic discrimination, we next conducted a pilot statistical modeling analysis using machine-learning-oriented approaches.
Figure 2.
Z-scores of diagnostic markers in metabolic disorders. For each disorder, the corresponding diagnostic markers are plotted. Black circles indicate non-IEM children (n = 296), while red circles represent affected patients.
3.5. Pilot Multinominal Logistic Regression Analysis for Machine-Learning-Assisted Diagnosis
As a proof-of-concept step toward data-driven diagnostics, we constructed pilot multinomial logistic regression (MLR) models using z-score-normalized values of representative diagnostic metabolites to distinguish among key IEM categories. The analyses targeted three disease group: (i) propionic acidemia (PA) and methylmalonic acidemia (MMA); (ii) isovaleric acidemia (IVA) and 3-methylcrotonyl-CoA carboxylase deficiency (MCC); (iii) long-chain fatty acid oxidation disorders, including very-long-chain acyl-CoA dehydrogenase (VLCAD), carnitine palmitoyltransferase II (CPT2), and mitochondrial trifunctional protein (TFP) deficiencies.
Figure 3A shows the MLR results for differentiating PA and MMA based on serum metabolite profiles. Among the 54 quantified metabolites, five diagnostic markers—3-hydroxypropionic acid, methylcitric acid, methylmalonic acid, propionylcarnitine (AC_C3), and methylmalonylcarnitine (AC_C4DC)—were incorporated into the model. Linear predictors were calculated as weighted sums of Z-scores, reflecting each metabolite’s regression coefficient. Distinct separation of PA and MMA was observed along the predictor axis, with odds ratios confirming strong contributions of propionate-pathway metabolites. The confusion matrix demonstrated perfect classification performance, correctly identifying all PA (n = 18) and MMA (n = 25) cases without misclassification among 341 non-IEM children.
Figure 3.
Multinomial logistic regression models for representative inborn errors of metabolism. (A) Propionic acidemia (PA) and methylmalonic acidemia (MMA). (B) Isovaleric acidemia (IVA) and 3-methylcrotonyl-CoA carboxylase deficiency (MCC). (C) Long-chain fatty acid oxidation disorders including very-long-chain acyl-CoA dehydrogenase (VLCAD), carnitine palmitoyltransferase II (CPT2), and mitochondrial trifunctional protein (TFP) deficiencies. For each model, the left panel shows the distribution of linear predictors for each disease versus other groups. The upper-right table summarize odds ratio derived from logistic regression for representative diagnostic metabolites. For simplicity, 95% confidence intervals were omitted and are denoted as “*”, as nearly all extended beyond the analytical range (approximately <0.01–>999) due to small sample sizes and sparse events in these rare disease cohorts. the lower-right tables show the confusion matrix comparing predicted and actual diagnostic categories.
Figure 3B,C presents similar MLR frameworks applied to the other disease groups. The model for IVA and MCC accurately distinguished between these two branched-chain amino acid disorders (Figure 3B), while the model for long-chain fatty acid oxidation disorders successfully separated VLCAD, CPT2, and TFP deficiencies with clear boundaries of linear predictors (Figure 3C). Across all models, predicted classifications showed strong concordance with biochemical diagnoses, supporting the feasibility of quantitative, serum-based metabolite profiling for multivariate disease discrimination. These proof-of-concept results highlight the potential of integrating standardized metabolomic quantification with machine-learning-assisted diagnostic systems, underscoring the long-term applicability of the fully automated serum LC–MS/MS platform in precision metabolic medicine.
4. Discussion
We developed and validated a fully automated serum LC–MS/MS workflow that consolidates second-line testing for IEMs—including organic acids, amino acids, and acylcarnitines—into a single analytical run from one serum specimen. The total analysis time of within 60 min per sample. By integrating on-instrument sample preparation with multiplexed LC–MS/MS detection, the platform directly addresses two entrenched limitations of the current multi-assay paradigm: fragmentation of testing across distinct specimen types and platforms, and pre-analytical complexity and delays, particularly when urinary GC–MS is outsourced [12,13]. Automation minimizes operator-dependent variability, enables predictable batching, and reduces staffing burden. These advantages facilitate implementation in routine clinical laboratories, where rapid turnaround time, reproducibility, and timely availability of results are critical operational priorities, particularly when centralized reference testing would otherwise delay clinical decision-making. Together, these features position the system as a practical, high-performance framework for standardized metabolic assessment, enabling timely clinical decision-making in both acute presentations and chronic care settings.
4.1. Analytical and Operational Performance
From an analytical perspective, the validated linear ranges encompassed physiological concentrations for most target metabolites observed in non-IEM children (n = 296), allowing quantitative reporting against the newly established serum-based pediatric RIs. Despite employing single-point calibration for 34 internal-standardized analytes, imprecision was uniformly <15% CV across all 54 targets, in line with ICH M10 and FDA bioanalytical guidelines [20,21]. This trade-off between analytical complexity and operational feasibility is justified in the context of acute metabolic crises, where turnaround time often outweighs incremental precision gains from multipoint calibration. Two metabolites—3-hydroxybutyric acid and glutamine—exceeded the upper limit of linearity in some specimens, suggesting the need for future optimization of MRM parameters, extended calibration ranges, or adaptive dilution strategies.
Although the majority of analytes demonstrated acceptable analytical performance, several metabolites, including methylcitric acid, isovalerylglycine, 3-methylcrotonylglycine, AC_C6, and C18, exhibited notable systematic deviations from nominal concentrations across concentration levels. For AC_C6, the absence of a structurally matched stable isotope-labeled internal standard may have contributed to reduced quantitative accuracy. For the other metabolites, isotopically matched internal standards were used; however, residual systematic bias may arise from compound-specific analytical characteristics under multiplex conditions. In particular, for 3-methylcrotonylglycine, incomplete chromatographic resolution from the isobaric compound tiglylglycine may have contributed to the observed positive deviation. This reflects the inherent challenge of resolving structurally related isomers in a high-throughput LC–MS/MS platform. A similar consideration applies to AC_C5, where isovalerylcarnitine and 2-methylbutyrylcarnitine were evaluated using authentic reference standards. Although complete baseline separation was not achieved, a consistent difference in retention time allowed practical chromatographic differentiation under the present conditions (Supplemental Figure S1). Despite these limitations in absolute quantification for selected analytes, clinical interpretation in the present study was based on integrated disease-specific metabolic profiles rather than isolated single-analyte values. Therefore, while caution is warranted when interpreting absolute concentrations for these compounds, the overall diagnostic performance and pattern-recognition capability of the method remain robust.
Operationally, full automation standardizes pre-analytical steps such as protein precipitation and 3-NPH derivatization, reducing variability due to manual timing, temperature, or reagent handling. Consolidating testing into a single serum matrix simplifies sample routing, reduces matrix heterogeneity across assays, and facilitates integration with Laboratory Information Systems (LIS). In scenarios where urine collection is delayed or infeasible—such as in neonates [25] or critically ill patients—the single-serum workflow prevents diagnostic interruptions and ensures continuous clinical support. Although current guidelines recommend heparinized plasma for acylcarnitine analysis [11], serum was used in this study, reflecting both the availability of archived patient samples and the routine use of serum in clinical biochemical testing at our institution.
4.2. Clinical Significance of a Unified Serum Platform
Clinically, this enables integrated metabolic assessment using the same sample typically collected for standard biochemistry, allowing early diagnosis, follow-up, and therapy monitoring. The unified workflow supports real-time decision-making and could shorten the time between suspicion and intervention. The quantitative data obtained also permit longitudinal follow-up of disease activity, therapeutic response, and residual metabolic load, transforming static diagnostic testing into dynamic clinical monitoring.
4.3. Establishment and Interpretation of Pediatric RIs
A major strength of this study is the establishment of serum-based pediatric RIs for 54 metabolites across 0–6 years of age. Given the strong age dependence of metabolic fluxes—driven by growth, nutrition, and enzyme maturation—age-adjusted RIs are essential for accurate interpretation. The parametric Box–Cox framework with outlier exclusion (±2.81 SD) and back-transformation (±1.96 SD) provided clinically interpretable limits even for skewed distributions. For analytes with >2.5% of measurements below the detection limit, only upper limits were defined, reflecting empirical detectability in pediatric serum.
Our serum RIs complement existing plasma-based catalogs [26,27,28] and address known matrix-dependent differences. For instance, several amino acids exhibit higher concentrations in serum than plasma [29,30]. Thus, serum-specific RIs provide practical interpretive value for laboratories operating primarily on serum workflows. We also compiled previously reported plasma-based reference data alongside our serum results for comparison (Supplemental Table S6), facilitating inter-laboratory harmonization. The availability of these RIs enables quantitative interpretation of metabolite deviations, transforming conventional qualitative pattern recognition into standardized, quantitative disease assessment.
While the Clinical and Laboratory Standards Institute (CLSI) EP28-A3c guideline [31] is widely recognized as the primary reference for establishing RIs, we adopted the IFCC C-RIDL protocol to ensure alignment with international harmonization initiatives specifically designed for multicenter, population-based studies [19,32]. The C-RIDL approach allows the use of parametric normalization (e.g., Box–Cox or logarithmic transformation) and supports direct comparison of datasets across institutions and ethnicities, which is particularly important for pediatric cohorts where large sample sizes are difficult to obtain at a single site. Moreover, the C-RIDL framework has been successfully applied in Asian pediatric populations and provides a validated pathway for indirect transfer and harmonization of RIs [32]. Therefore, although not the standard CLSI pathway, the IFCC C-RIDL methodology was intentionally selected to ensure global comparability and reproducibility of pediatric reference data.
4.4. Quantitative Disease Evaluation and Machine-Learning-Assisted Modeling
Clinical validation in 89 biochemically confirmed IEM cases demonstrated that disease-specific metabolic patterns can be captured directly from serum, even for metabolites currently measured in urine. Representative examples include glutaric acidemia type II, characterized by elevated organic acids (glutaric, 2-hydroxyglutaric) and broad acylcarnitine accumulation [22,23,24], and propionic and methylmalonic acidemias, which exhibit disease-specific combinations of methylcitric acid, propionylcarnitine (AC_C3), and methylmalonic acid reflecting the site of enzymatic blockage [33,34]. In the present study, the MMA cases were included as biochemically confirmed methylmalonic aciduria based on elevated methylmalonic acid levels; however, detailed subclassification (e.g., mut, cblA/B versus cblC-related combined MMA) was not available for all archived samples. In patients with propionic acidemia (PA) and methylmalonic acidemia (MMA), plasma-based analyses have shown that related organic acids can be directly quantified and that their concentrations are associated with disease status and metabolic control [35,36]. These observations were consistent with known biochemical pathways and confirm the diagnostic fidelity of serum-based analysis.
To explore the potential of data-driven interpretation, we applied pilot multinomial logistic regression models using z-score–normalized metabolite data derived from the pediatric RIs. The models successfully discriminated key IEM categories: propionic/methylmalonic acidemias, isovaleric/3-methylcrotonyl-CoA carboxylase deficiencies (IVA/MCC), and long-chain fatty acid oxidation disorders (VLCAD, CPT2, TFP deficiencies). Among the predefined markers, methylmalonic acid and acylcarnitine C4DC were more strongly associated with MMA, whereas 3-hydroxypropionic acid, methylcitric acid, and C3-carnitine showed higher odds ratios in PA—findings concordant with established biochemical distinctions described in consensus guidelines [33]. Although exploratory, these results demonstrate that quantitative serum data preserve diagnostic information suitable for multivariate, machine-learning-assisted classification.
While further work is needed to expand model size and diversity, this proof-of-concept underscores the feasibility of combining standardized quantitative metabolite data with statistical modeling to improve reproducibility and objectivity in IEM diagnostics. In the long term, integration of automated LC–MS/MS workflows with machine-learning-assisted interpretation could facilitate semi-automated diagnostic support in routine practice, analogous to algorithmic decision systems in clinical chemistry or hematology.
4.5. Limitation and Future Directions
This study has several limitations. The number of available cases for each IEM category was relatively small, limiting statistical power and exploration of rarer phenotypes. Commercial internal standards were available for 34 of the 54 target analytes; for the remaining compounds, their unavailability necessitated semi-quantitative assessment and may have affected cross-lot comparability. Another analytical limitation relates to the chromatographic resolution of certain isobaric amino acids. Under the multiplex LC–MS/MS conditions used in this study, complete chromatographic separation between leucine and isoleucine was not achieved, and hydroxyproline, which is also isobaric with leucine, was not specifically evaluated as an independent analyte. However, in human serum leucine is typically present at substantially higher concentrations than hydroxyproline, and leucine and isoleucine are generally present in comparable ranges and biologically correlated as branched-chain amino acids. Because the present study focuses on relative metabolite quantification within a uniformly processed sample set rather than diagnostic discrimination of individual branched-chain amino acids, this limitation is unlikely to substantially affect the comparative analyses performed in this work. Nevertheless, applications requiring strict discrimination among leucine, isoleucine, and hydroxyproline would require dedicated chromatographic separation.
Another limitation is that total homocysteine (tHcy), an essential marker for the differential diagnosis between isolated methylmalonic acidemia and combined MMA with homocystinuria, was not included in the present analyte panel. In clinical practice, diagnosis of MMA typically follows a stepwise biochemical algorithm: detection of elevated methylmalonic acid during initial metabolic screening, followed by targeted measurement of tHcy to distinguish isolated MMA from combined cobalamin-related disorders, and subsequent confirmation by molecular genetic testing. Therefore, the present multiplex metabolite platform is intended to function primarily as a biochemical screening and metabolic monitoring tool within this established diagnostic workflow rather than as a stand-alone assay for subtype discrimination.
In addition, most patient samples analyzed in this study were archived specimens obtained from external institutions, and detailed clinical information regarding treatment status or sampling time points was not consistently available. Therefore, the present dataset does not allow direct evaluation of longitudinal metabolite changes during therapy. Prospective studies with well-defined pre- and post-treatment sampling will be required to formally validate the clinical utility of this platform for therapeutic monitoring.
The current panel does not include certain newborn-screening-relevant metabolites such as argininosuccinic acid and succinylacetone [37,38]. In several specimens, concentrations of 3-hydroxybutyric acid and glutamine exceeded the validated upper limit of linearity. These metabolites are known to increase markedly during acute metabolic decompensations such as ketosis or hyperammonemic crises in patients with inborn errors of metabolism. In such cases, dilution and re-analysis can be readily implemented within the current LC–MS/MS workflow, and future methodological refinements will aim to extend the calibration range to improve quantitative performance at very high concentrations. Although our pediatric cohort was relatively large, RIs were not partitioned by narrow age bands or sex, and neonatal time points (e.g., postnatal day 3–5)—which correspond to the standard sampling window for newborn screening [39]—were not represented. Finally, inter-laboratory comparisons and participation in external quality assessment (EQA) schemes were not performed but will be essential for broader validation and proficiency alignment. External validation through international collaborative studies will be essential to generalize these RIs beyond the Japanese population.
Future studies should aim to expand the analyte library, optimize analytical ranges, and establish age- and sex-partitioned RIs, including neonatal subsets. Collaborative, multi-institutional efforts combining clinical, biochemical, and genetic data will be crucial to build robust machine-learning-assisted models and validate them across diverse populations. Further integration into laboratory information systems may support automated flagging, trend analysis, and harmonized reporting across metabolic centers.
In summary, this study provides a foundation for serum-based, quantitative, and fully automated assessment of inborn errors of metabolism. By defining pediatric RIs and demonstrating proof-of-concept machine-learning-assisted classification, we show that serum metabolomics can unify second-line testing, shorten diagnostic turnaround, and enable data-driven interpretation. This platform represents a practical step toward precision metabolic diagnostics that are standardized, reproducible, and operationally feasible for modern clinical laboratories.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics16060911/s1, Table S1. Standards and ISTD. Table S2. Demographic characteristics (age and sex distribution) of children without inborn errors of metabolism. Table S3. Distribution of IEM Patient cohort. Table S4. Multiple reaction monitoring transitions of 3-Nitrophenylhydrazine(3-NPH) derivatized compounds. Table S5. Summary statistics of metabolite concentrations in children without inborn errors of metabolism. Table S6. Serum pediatric RIs (0–6 years) derived in this study versus published plasma reference limits across targeted metabolites. Table S7. Inborn error s of metabolism included in current newborn metabolic screening programs in Japan. Figure S1 Representative MRM chromatograms of isovalerylcarnitine and 2-methylbutyrylcarnitine. References [40,41,42,43,44,45,46] are cited in the supplementary materials.
Author Contributions
D.S., Y.U. and T.N. were responsible for the conceptualization and design of the research. Y.U., S.M. and A.F. performed all experiments. Y.U. conducted the data analysis and drafted the manuscript. D.S. and Y.K. critically revised and finalized the manuscript. Y.M., V.T., H.K. and M.M. provided clinical samples, patient information, and clinical consultation. T.H. and A.S. supervised the overall research process and provided administrative and technical oversight. All authors have read and agreed to the published version of the manuscript.
Funding
This study was supported by research grant from Shimadzu corporation and JSPS KAKENHI Grant Number JP24K18860.
Institutional Review Board Statement
This study was approved by the Ethics Committee of Kyushu University (approval number: 22138, approved on 28 November 2024). All procedures were conducted in accordance with the ethical standards of the institutional and national committee on human experimentation and with the Declaration of Helsinki (1975, as revised in 2000).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.
Acknowledgments
We are grateful to Ayako Hidaka and Chihiro Yanagiuchi for their valuable technical supports.
Conflicts of Interest
Author Tsuyoshi Nakanishi is an employee of Shimadzu Corporation. The company collaborated with Kyushu University on this research project and provided partial research support. The remaining authors declare that they have no conflicts of interest. The authors declare that this study received funding from Shimadzu corporation. The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication.
References
- Saudubray, J.-M.; Garcia-Cazorla, À. Inborn Errors of Metabolism Overview: Pathophysiology, Manifestations, Evaluation, and Management. Pediatr. Clin. N. Am. 2018, 65, 179–208. [Google Scholar] [CrossRef] [Scilit]
- Guerrero, R.B.; Salazar, D.; Tanpaiboon, P. Laboratory Diagnostic Approaches in Metabolic Disorders. Ann. Transl. Med. 2018, 6, 470. [Google Scholar] [CrossRef] [Scilit]
- Driesen, K.; Witters, P. Understanding Inborn Errors of Metabolism through Metabolomics. Metabolites 2022, 12, 398. [Google Scholar] [CrossRef] [Scilit]
- Tanaka, K.; West-Dull, A.; Hine, D.G.; Lynn, T.B.; Lowe, T. Gas-Chromatographic Method of Analysis for Urinary Organic Acids. II. Description of the Procedure, and Its Application to Diagnosis of Patients with Organic Acidurias. Clin. Chem. 1980, 26, 1847–1853. [Google Scholar] [CrossRef] [Scilit]
- Matsumoto, I.; Kuhara, T. A New Chemical Diagnostic Method for Inborn Errors of Metabolism by Mass Spectrometry—Rapid, Practical, and Simultaneous Urinary Metabolites Analysis. Mass. Spectrom. Rev. 1996, 15, 43–57. [Google Scholar] [CrossRef]
- Ohie, T.; Fu, X.; Iga, M.; Kimura, M.; Yamaguchi, S. Gas Chromatography-Mass Spectrometry with Tert.-Butyldimethylsilyl Derivation: Use of the Simplified Sample Preparations and the Automated Data System to Screen for Organic Acidemias. J. Chromatogr. B Biomed. Sci. Appl. 2000, 746, 63–73. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Carling, R.S.; Witek, K.; Emmett, E.C.; Gallagher, C.; Moat, S.J. Urine Organic Acid Metabolomic Profiling by Gas Chromatography Mass Spectrometry: Assessment of Solvent Extract Evaporation Parameters on the Recovery of Key Diagnostic Metabolites. Clin. Chim. Acta 2025, 565, 120015. [Google Scholar] [CrossRef] [Scilit]
- Prinsen, H.C.M.T.; Schiebergen-Bronkhorst, B.G.M.; Roeleveld, M.W.; Jans, J.J.M.; de Sain-van der Velden, M.G.M.; Visser, G.; van Hasselt, P.M.; Verhoeven-Duif, N.M. Rapid Quantification of Underivatized Amino Acids in Plasma by Hydrophilic Interaction Liquid Chromatography (HILIC) Coupled with Tandem Mass-Spectrometry. J. Inherit. Metab. Dis. 2016, 39, 651–660. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sharer, J.D.; De Biase, I.; Matern, D.; Young, S.; Bennett, M.J.; Tolun, A.A. Laboratory Analysis of Amino Acids, 2018 Revision: A Technical Standard of the American College of Medical Genetics and Genomics (ACMG). Genet. Med. 2018, 20, 1499–1507. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Phipps, W.S.; Crossley, E.; Boriack, R.; Jones, P.M.; Patel, K. Quantitative Amino Acid Analysis by Liquid Chromatography-Tandem Mass Spectrometry Using Low Cost Derivatization and an Automated Liquid Handler. JIMD Rep. 2020, 51, 62–69. [Google Scholar] [CrossRef] [Scilit]
- Miller, M.J.; Cusmano-Ozog, K.; Oglesbee, D.; Young, S. Laboratory Analysis of Acylcarnitines, 2020 Update: A Technical Standard of the American College of Medical Genetics and Genomics (ACMG). Genet. Med. 2021, 23, 249–258. [Google Scholar] [CrossRef] [Scilit]
- Phiri, M.M.; Davoren, E.; Vorster, B.C. Miniaturization and Automation Protocol of a Urinary Organic Acid Liquid-Liquid Extraction Method on GC-MS. Molecules 2023, 28, 5927. [Google Scholar] [CrossRef] [Scilit]
- Xiao, Y.; Sun, C.; Leung, E.K.Y. Urine Organic Acid Testing in the Clinical Laboratory: The Past, Current, and Future. Encyclopedia 2025, 5, 153. [Google Scholar] [CrossRef] [Scilit]
- Pitt, J.J. Principles and Applications of Liquid Chromatography-Mass Spectrometry in Clinical Biochemistry. Clin. Biochem. Rev. 2009, 30, 19–34. [Google Scholar]
- Ueyanagi, Y.; Setoyama, D.; Kawakami, D.; Mushimoto, Y.; Matsumoto, S.; Hotta, T.; Kang, D. Fully Automated Quantitative Measurement of Serum Organic Acids via LC-MS/MS for the Diagnosis of Organic Acidemias: Establishment of an Automation System and a Proof-of-Concept Validation. Diagnostics 2021, 11, 2195. [Google Scholar] [CrossRef] [Scilit]
- Ellul, S.; Wake, M.; Clifford, S.A.; Lange, K.; Würtz, P.; Juonala, M.; Dwyer, T.; Carlin, J.B.; Burgner, D.P.; Saffery, R. Metabolomics: Population Epidemiology and Concordance in Australian Children Aged 11–12 Years and Their Parents. BMJ Open 2019, 9, 106–117. [Google Scholar] [CrossRef] [Scilit]
- Peng, G.; Tang, Y.; Cowan, T.M.; Enns, G.M.; Zhao, H.; Scharfe, C. Reducing False-Positive Results in Newborn Screening Using Machine Learning. Int. J. Neonatal Screen. 2020, 6, 16. [Google Scholar] [CrossRef] [Scilit]
- Zhou, M.; Deng, L.; Huang, Y.; Xiao, Y.; Wen, J.; Liu, N.; Zeng, Y.; Zhang, H. Application of the Artificial Intelligence Algorithm Model for Screening of Inborn Errors of Metabolism. Front. Pediatr. 2022, 10, 855943. [Google Scholar] [CrossRef] [Scilit]
- Ichihara, K.; Ozarda, Y.; Barth, J.H.; Klee, G.; Qiu, L.; Erasmus, R.; Borai, A.; Evgina, S.; Ashavaid, T.; Khan, D.; et al. A Global Multicenter Study on Reference Values: 1. Assessment of Methods for Derivation and Comparison of Reference Intervals. Clin. Chim. Acta 2017, 467, 70–82. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- European Medicines Agency. ICH Guideline M10 on Bioanalytical Method Validation and Study Sample Analysis_Step 5; European Medicines Agency (EMA): Amsterdam, The Netherlands, 2022; pp. 1–45. [Google Scholar]
- U.S. Food and Drug Administration (FDA). Bioanalytical Method Validation Guidance for Industry; U.S. Department of Health and Human Services: Silver Spring, MD, USA, 2018; pp. 1–40.
- Yotsumoto, Y.; Hasegawa, Y.; Fukuda, S.; Kobayashi, H.; Endo, M.; Fukao, T.; Yamaguchi, S. Clinical and Molecular Investigations of Japanese Cases of Glutaric Acidemia Type 2. Mol. Genet. Metab. 2008, 94, 61–67. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Grünert, S.C. Clinical and Genetical Heterogeneity of Late-Onset Multiple Acyl-Coenzyme A Dehydrogenase Deficiency. Orphanet J. Rare Dis. 2014, 9, 117. [Google Scholar] [CrossRef] [Scilit]
- Siano, M.A.; Mandato, C.; Nazzaro, L.; Iannicelli, G.; Ciccarelli, G.P.; Barretta, F.; Mazzaccara, C.; Ruoppolo, M.; Frisso, G.; Baldi, C.; et al. Hepatic Presentation of Late-Onset Multiple Acyl-CoA Dehydrogenase Deficiency (MADD): Case Report and Systematic Review. Front. Pediatr. 2021, 9, 672004. [Google Scholar] [CrossRef] [Scilit]
- Kaufman, J.; Fitzpatrick, P.; Tosif, S.; Hopper, S.M.; Donath, S.M.; Bryant, P.A.; Babl, F.E. Faster Clean Catch Urine Collection (Quick-Wee Method) from Infants: Randomised Controlled Trial. BMJ 2017, 357, j1341. [Google Scholar] [CrossRef] [Scilit]
- Seattle Children’s Hospital, Department of Laboratory Medicine STAT Beta-Hydroxybutyrate (STAT BOHB)—Seattle Children’s Hospital. Available online: https://seattlechildrenslab.testcatalog.org/show/STAT-BOHB-1 (accessed on 31 October 2025).
- Reuter, S.E.; Evans, A.M.; Chace, D.H.; Fornasini, G. Determination of the Reference Range of Endogenous Plasma Carnitines in Healthy Adults. Ann. Clin. Biochem. 2008, 45, 585–592. [Google Scholar] [CrossRef] [Scilit]
- Abildgaard, A.; Knudsen, C.S.; Hoejskov, C.S.; Greibe, E.; Parkner, T. Reference Intervals for Plasma Vitamin B12 and Plasma/Serum Methylmalonic Acid in Danish Children, Adults and Elderly. Clin. Chim. Acta 2022, 525, 62–68. [Google Scholar] [CrossRef] [Scilit]
- Lee, H.-O.; Uzzo, R.G.; Kister, D.; Kruger, W.D. Combination of Serum Histidine and Plasma Tryptophan as a Potential Biomarker to Detect Clear Cell Renal Cell Carcinoma. J. Transl. Med. 2017, 15, 72. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yu, Z.; Kastenmüller, G.; He, Y.; Belcredi, P.; Möller, G.; Prehn, C.; Mendes, J.; Wahl, S.; Roemisch-Margl, W.; Ceglarek, U.; et al. Differences between Human Plasma and Serum Metabolite Profiles. PLoS ONE 2011, 6, e21230. [Google Scholar] [CrossRef] [Scilit]
- Clinical and Laboratory Standards Institute. Clinical and Laboratory Standards Institute Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, 3rd ed.; Approved Guideline; Clinical and Laboratory Standards Institute: Wayne, PA, USA, 2010; p. 76. [Google Scholar]
- Klee, G.G.; Ichihara, K.; Ozarda, Y.; Baumann, N.A.; Straseski, J.; Bryant, S.C.; Wood-Wentz, C.M. Reference Intervals: Comparison of Calculation Methods and Evaluation of Procedures for Merging Reference Measurements From Two US Medical Centers. Am. J. Clin. Pathol. 2018, 150, 545–554. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Baumgartner, M.R.; Hörster, F.; Dionisi-Vici, C.; Haliloglu, G.; Karall, D.; Chapman, K.A.; Huemer, M.; Hochuli, M.; Assoun, M.; Ballhausen, D.; et al. Proposed Guidelines for the Diagnosis and Management of Methylmalonic and Propionic Acidemia. Orphanet J. Rare Dis. 2014, 9, 130. [Google Scholar] [CrossRef] [Scilit]
- Fraser, J.L.; Venditti, C.P. Methylmalonic and Propionic Acidemias: Clinical Management Update. Curr. Opin. Pediatr. 2016, 28, 682. [Google Scholar] [CrossRef] [Scilit]
- Haijes, H.A.; Jans, J.J.M.; van der Ham, M.; van Hasselt, P.M.; Verhoeven-Duif, N.M. Understanding Acute Metabolic Decompensation in Propionic and Methylmalonic Acidemias: A Deep Metabolic Phenotyping Approach. Orphanet J. Rare Dis. 2020, 15, 68. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Maines, E.; Catesini, G.; Boenzi, S.; Mosca, A.; Candusso, M.; Dello Strologo, L.; Martinelli, D.; Maiorana, A.; Liguori, A.; Olivieri, G.; et al. Plasma Methylcitric Acid and Its Correlations with Other Disease Biomarkers: The Impact in the Follow up of Patients with Propionic and Methylmalonic Acidemia. J. Inherit. Metab. Dis. 2020, 43, 1173–1185. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Häberle, J.; Boddaert, N.; Burlina, A.; Chakrapani, A.; Dixon, M.; Huemer, M.; Karall, D.; Martinelli, D.; Crespo, P.S.; Santer, R.; et al. Suggested Guidelines for the Diagnosis and Management of Urea Cycle Disorders. Orphanet J. Rare Dis. 2012, 7, 32. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stinton, C.; Geppert, J.; Freeman, K.; Clarke, A.; Johnson, S.; Fraser, H.; Sutcliffe, P.; Taylor-Phillips, S. Newborn Screening for Tyrosinemia Type 1 Using Succinylacetone—A Systematic Review of Test Accuracy. Orphanet J. Rare Dis. 2017, 12, 48. [Google Scholar] [CrossRef] [Scilit]
- Odenwald, B.; Brockow, I.; Hanauer, M.; Lüders, A.; Nennstiel, U. Is Our Newborn Screening Working Well? A Literature Review of Quality Requirements for Newborn Blood Spot Screening (NBS) Infrastructure and Procedures. Int. J. Neonatal Screen. 2023, 9, 35. [Google Scholar] [CrossRef] [Scilit]
- Mitochondrial Metabolites, Plasma-Mayo Clinic Laboratories|Pediatric Catalog. Available online: https://pediatric.testcatalog.org/show/MMPP (accessed on 18 September 2025).
- Lactate, Plasma-Mayo Clinic Laboratories Extended Catalog. Available online: https://www.testcatalog.org/show/LACS1 (accessed on 18 September 2025).
- Pyruvic Acid, Blood-Network Reference Lab. Available online: https://nrl.testcatalog.org/show/PYR (accessed on 18 September 2025).
- Hydroxyglutaric Acids, Glutaric Acid, Ethylmalonic Acid, and Methylsuccinic Acid, Plasma-Logan Health Medical Center Laboratory Test Catalog. Available online: https://logan.testcatalog.org/show/HGEMP (accessed on 18 September 2025).
- Wen, Y.; Liu, Q.; Zeng, H.; Lyu, L.; He, X.; Zhang, X.; Lyu, W.; Chen, W.; Xiao, Y. Age-specific reference intervals for plasma amino acids and their associations with nutrient intake in the Chinese pediatric population. iMeta 2025, 4, e70051. [Google Scholar] [CrossRef] [Scilit]
- Carnitines, Serum-Mayo Clinic Laboratories Extended Catalog. Available online: https://isj.testcatalog.org/show/CARNS (accessed on 10 February 2026).
- Acylcarnitines, Quantitative, Serum-Mayo Clinic Laboratories Extended Catalog. Available online: https://www.testcatalog.org/show/ACRNS (accessed on 18 September 2025).
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.