Abstract
Background/Objective: Artificial intelligence (AI) has been investigated for the radiographic recognition of existing complications and prediction of subsequent adverse outcomes after total hip arthroplasty (THA). We synthesized the reported diagnostic and prognostic performance and examined testing strategies, calibration, and methodological limitations. Methods: We conducted a systematic review reported in accordance with Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020. Eligible English-language full-text studies published between 1 January 2010 and 31 August 2026, evaluated Artificial intelligence and Machine Learning (AI/ML) models using radiographic information, alone or with clinical data, to detect an existing post-THA complication or predict a subsequent complication or implant failure. Mixed-population studies required separately extractable THA-specific results. The final search was executed on 21 October 2026. Eligible models incorporated radiographic images or image-derived features; clinical-only models were not eligible. A structured narrative synthesis was used because of clinical and methodological heterogeneity. Results: Twelve studies met the eligibility criteria for qualitative synthesis after the exclusion of studies focused on pre-arthroplasty osteoarthritis progression or short-term need for THA, which were outside the pre-defined review scope. Across the included post-THA studies, AI models were applied to tasks including aseptic loosening detection, multiclass failure classification, dislocation risk prediction, periprosthetic fracture as a label within the broader THA complication recognition, and longitudinal prediction of prosthesis failure. Several studies reported high apparent discrimination, particularly in internally validated single-center datasets, while some studies evaluated performance in additional institutional cohorts. Registry linkage was not treated as evidence of external evaluation. However, the evidence base was limited by substantial heterogeneity in populations, outcome definitions, model designs, and validation strategies, and most studies were judged to be at high overall risk of bias. Conclusions: Current evidence suggests that AI-based radiographic assessment of selected post-THA complications is promising, but the principal limitations are incomplete independent external evaluation, sparse calibration and clinical-utility evidence, and methodological concerns. The reported retrospective performance does not establish improved implant survivorship or patient outcomes. A prospective multicenter evaluation, transparent reporting, human–AI workflow assessment, and clinical-impact studies are needed before routine adoption.
1. Introduction
THA is one of the most successfully performed and cost-effective orthopedic procedures worldwide, which restores mobility, and improves quality of life for patients with severe hip joint diseases. Annual primary THA volumes in high-income countries alone are projected to exceed 500,000 procedures per year by 2030, growing at a ~5.2% rate and continue rising thereafter [1,2,3,4]. Even though low- and middle-income countries lack complete national registry, global estimates still show hundreds of thousands of primary THRs performed worldwide each year, with global volumes expected to surpass one million procedures annually in the coming decade by 2030. The numbers for total hip and knee arthroplasties vary but are often cited in the range of 1–2 million annually in older projections (e.g., pre-2020 data), though more recent analysis suggests higher volumes [2]. Despite the high success rates of THA, complications such as implant malposition, periprosthetic fracture, implant breakage, aseptic loosening, leg-length or offset discrepancy, malalignment, and dislocation may necessitate revision surgery [2,4,5,6]. Revision surgeries drive the substantial healthcare costs and morbidity rates, with projections indicating primary THA volumes could rise by 129–174% by 2030 in the U.S., amplifying the need for better complication management [4,5]. The conventional assessment of these complications relies heavily on radiographic interpretation, longitudinal follow-up, and clinical judgment, all of which may be affected by subjectivity, inter-observer variability, and delayed recognition of subtle failure patterns. This has encouraged interest in artificial intelligence (AI), machine learning (ML), and deep-learning (DL) methods as tools for automated image interpretation, failure detection, and outcome prediction in arthroplasty care. Convolutional neural networks and related architectures have been investigated for loosening detection, multiclass complication classification, fracture identification, and prognostic modeling using radiographs alone or combined with clinical variables [7]. The potential improvements in speed, consistency, or decision support remain as hypotheses requiring comparative clinical evaluation; improved rehabilitation or implant longevity cannot be inferred from retrospective discrimination alone. Enthusiasm for these methods must be balanced against important methodological limitations in the current literature, including retrospective single-center designs, selected cohorts, inconsistent outcome definitions, incomplete calibration reporting, and limited external validation [7,8]. These issues directly affect the reliability, transportability, and clinical applicability of reported performance metrics. Kim et al. [7] synthesized five hip/knee loosening studies identified through September 2022 and pooled diagnostic accuracy. Stuani et al. [8] synthesized ten hip/knee radiographic loosening or mechanical-failure studies identified on November 26, 2025. Stuani et al. already examined internal versus external testing, PROBAST risk of bias, longitudinal imaging, calibration, and clinical implementation; these topics are not unique contributions of the present review. The remaining question addressed here is how radiographic AI evidence maps to distinct THA decisions across existing-complication diagnosis and subsequent-event prognosis, including current-dislocation recognition and future-dislocation prediction. Our contribution is a THA-specific, task- and evaluation-unit-based synthesis through 31 August 2026, rather than a first review of radiographic AI failure detection. The primary question was: in patients with THA, how accurately do AI/ML models incorporating radiographic information identify an existing complication or predict a subsequent complication or failure, and how credible and applicable are those estimates? Complications include dislocation, periprosthetic fracture, infection, wear, osteolysis, and component loosening. The specific objectives were to: (1) summarize the AI/ML/DL models and input data used for post-THA complication assessment; (2) describe reported diagnostic and prognostic performance for failure-related outcomes; and (3) evaluate study design quality, validation strategies, and sources of bias that may affect generalizability and clinical usefulness.
2. Materials and Methodology
2.1. Study Design and Reporting Framework
This systematic review is reported in accordance with the PRISMA 2020 statement [9]. The methodology was defined in a protocol specifying eligibility criteria, information sources, search strategies, screening processes, data extraction items, risk of bias assessment, and synthesis plans. The review protocol was not registered in PROSPERO or another platform. The PRISMA checklist is in the Supplementary Materials (Supplementary File S1).
2.2. Eligibility Criteria
Eligible studies evaluated patients with primary or revision THA and reported diagnostic or prognostic performance for a defined post-THA complication or failure-related outcome. Radiographic information was required, either as images or as image-derived features; clinical variables could be additional input. Pre-primary-THA imaging was eligible only when used to predict a subsequent post-THA outcome. Images obtained before revision surgery were classified as post-implantation diagnostic inputs when used to identify an existing complication. Mixed-joint or non-THA populations were eligible only when THA-specific sample information and relevant performance results were separately extractable. Eligible learned methods included support vector machines, tree-based methods, neural networks, deep-learning architectures, and hybrid/ensemble models. Conventional logistic regression was treated as a comparator. Eligible designs included model development or evaluation studies in human clinical populations. The publication dates, language, and full-text requirements were 1 January 2010–31 August 2026, English, and full-text reports, respectively.
2.3. Exclusion Criteria
The excluded reports comprised non-human studies, case reports or series with fewer than 10 patients, reviews and non-primary commentaries, and reports without an AI/ML/DL component or post-THA outcome. Studies of native-hip osteoarthritis progression or prediction of whether a patient would receive a first THA were excluded. Isolated landmark detection, segmentation, implant identification, or alignment measurement without an eligible complication/failure endpoint was outside our scope. General femoral-fracture studies without separately extractable THA results were excluded. Conference abstracts without full text were ineligible; where available, full conference papers were considered. Preprints and unpublished/under-review reports were excluded.
2.4. Information Source and Search Strategy
A systematic search was performed in PubMed, IEEE Xplore, Scopus, and Web of Science Core Collection, with a supplementary JBJS journal search. The period for publication eligibility was 1 January 2010–31 August 2026; the final search execution date was 21 October 2026. Search concepts combined THA/THR, radiographic information, AI/ML/deep learning, and complication/failure outcomes using source-specific Boolean syntax. Supplementary Material File S2 provides complete strategies, search fields, platforms, execution dates, limits, and source-specific retrieval counts. The publication cutoff and search execution date represent different dates and are not interchangeable.
2.5. Study Selection
All records retrieved from the searches were imported into a reference manager, and duplicates were removed using automated and manual procedures. Title and abstract screening were performed independently by two reviewers using pre-defined eligibility criteria. Articles deemed potentially relevant by either reviewer proceeded to full-text review. Full-text articles were then assessed independently by the same reviewers. Discrepancies at both stages were resolved through discussion or consultation with a third reviewer. Reasons for exclusion at the full-text stage were recorded.
2.6. Data Extraction
Two reviewers independently extracted data, with disagreements resolved by consensus. Extraction distinguishes unique patients, operated hips, examinations, and images; development, tuning, internal test, temporal test, and external evaluation datasets; existing diagnostic outcomes; and specific model configurations. The calibration, predictive values, uncertainty intervals, clinical-utility measures, preprocessing, missing-data handling, and explainability are recorded as ‘reported’, or ‘not reported’, or ‘unclear’. An absence from a summary table is not evidence of absence from the primary article. Metric definitions, positive classes, averaging schemes, and evaluation units were retained separately. Where a primary report supplied an explicit confusion matrix or complete class-specific counts for the same evaluation, derived metrics were labelled as reconstructed and distinguished from the published summary. The source-reported values that could not be reconciled were retained with an explicit qualification.
2.7. Risk of Bias and Applicability Assessment
The risk of bias was assessed using the original Prediction model Risk of Bias Assessment Tool (PROBAST-2019), covering participants, predictors, outcome, and analysis [10]. Two reviewers independently performed the assessments, with disagreements resolved by discussion. We retained the original framework actually used for this review, rather than presenting the existing assessments as a retrospective PROBAST+AI reassessment. The original PROBAST covers diagnostic models estimating a current condition and prognostic models estimating a future event; the diagnostic reference standard and outcome timing/follow-up must therefore be considered separately. PROBAST+AI provides updated, distinct model-development and model-evaluation assessments and additional AI-relevant guidance [11]. Retaining PROBAST-2019 limits the appraisal and does not imply an equivalence to the updated tool. Table 3 presents the existing study-summary judgments; each distinct relevant model and development/evaluation component requires an auditable assessment. A high-risk domain yields high overall risk; low overall risk requires all domains to be low, and an unclear domain without any high domain yields unclear overall risk. Applicability is distinct from a risk of bias. Cohort size, multicenter recruitment, an absence of external evaluation, or class imbalance alone is not a sufficient explanation for a domain judgment.
2.8. Effect Measures
Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, F1 score, and AUC were extracted for diagnostic classifications where reported. Prognostic discrimination included area under the receiver operating characteristic curve (AUC) or a time-to-event C-index, with the prediction horizon and event definition retained. Calibration plots, calibration intercept/slope, Brier score, decision-curve net benefit, and workflow or patient outcomes were extracted when available. AUC and C-index were not treated as interchangeable estimates across tasks. Missing metrics were not imputed or reconstructed from incompatible cohorts or thresholds. Confusion-matrix-derived estimates were identified explicitly and were not mixed with differently averaged or separately evaluated source summaries. Reported recall was not automatically relabeled as sensitivity for the failure-positive class.
2.9. Synthesis Methods
A structured narrative synthesis grouped binary diagnosis, multiclass/multilabel complication recognition, future-dislocation prediction, and longitudinal failure prediction. Only eligible THA-specific evidence contributes to the primary synthesis. Heterogeneity in population, reference standards, outcomes, prevalence, sampling, and testing design precluded a defensible, pooled estimate or cross-study ranking. Internal testing denotes evaluation within the development source using held-out data or resampling; external evaluation uses participants not used for model development, with geographic/institutional independence described explicitly. A temporal evaluation uses a later, non-overlapping cohort and is distinguished from a prospective workflow evaluation. A paper’s use of the word “external” is retained as author terminology when training or tuning independence is unclear.
3. Results
Following the PRISMA methodology [9], the reported selection pathway identified 387 records: Scopus 119, Web of Science 116, PubMed 104, IEEE Xplore 45, and the supplementary JBJS journal search 3. The removal of 203 records before screening left 184 records for title/abstract screening; 156 were excluded. All 28 reports sought were retrieved and assessed, and 16 full-text reports were excluded, leaving 12 reports representing 12 included studies. Two pre-arthroplasty osteoarthritis/first-THA reports, by Chen et al. [12] and Hidaka et al. [13], were excluded during title/abstract screening and did not contribute to the primary synthesis or appraisal. Figure 1 presents the final selection pathway.
Figure 1.
PRISMA flow diagram for study selection process. Publication eligibility: 1 January 2010–31 August 2026; Final Search: 21 October 2026.
Table 1 summarizes cohort characteristics and testing design; Table 2 links model configurations, clinical tasks, and reported performance. These are descriptive study-level summaries, not a ranking of models across incompatible populations.
Table 1.
Summary of characteristics, sampling and evaluation strategies of included studies.
Table 2.
Task-specific models and reported performance.
The studies ranged from small selected diagnostic samples to registry-linked THA cohorts. Imaging inputs included AP or AP/lateral radiographs, serial examinations, and radiographic information combined with clinical variables. The following synthesis compares tasks and testing settings rather than selecting a cross-study performance winner.
3.1. Comparative Synthesis According to Clinical Task
Diagnostic studies identified existing complications, whereas prognostic studies estimated subsequent events from earlier available information. Binary diagnosis, multilabel recognition, and future-event prediction differed in reference standards, prevalence, and decision purpose and were synthesized separately. Imaging before a revision surgery should not be confused with imaging before primary implantation [14,15,16,17,18,19,20,21,22,23,24,25].
Wei et al. [25] adds a distinct diagnostic task: identifying a dislocation already visible on a THA radiograph. Its external Hospital B and Internet samples contain approximately 50% dislocated images. By contrast, Rouzrokh et al. [22] and Khosravi et al. [21] estimate subsequent dislocation in low-event-rate cohorts. Their AUCs, sensitivities and predictive values address different clinical questions and should not be ranked against Wei et al.’s selected diagnostic cohorts.
Loosening studies differed in sampling and comparator definitions [16,17,18,19,23]. In Guo et al., the internal dataset comprised 443 pre-revision radiographs with complication labels, not an unselected surveillance population. A case without the loosening label was not necessarily an uncomplicated well-fixed hip [17]. Binary loosening results therefore address a different clinical spectrum from routine screening.
Guo et al. reported mAP 34.6% and F1 42.9% for the ResNet18 + GFS + CFS + MC-CSRA multilabel configuration. Separate binary loosening experiments used ResNet18 + GFS + CFS, with internal accuracy of 88.1% and F1 91.2%, and accuracy of 92.9% and F1 94.7% on a separate dataset described by the authors as external [17]. These results are not interchangeable, and dataset reuse does not by itself establish frozen-model external validation without verifying the training/tuning procedure.
3.2. Performance in Internal and External Testing
Muscato et al. reported an internal validation accuracy of 95.8% and source-reported external balanced accuracy of 86.1% [16]; these are not presented as a like-for-like decline in ordinary accuracy. The primary report’s external metric summaries are not fully reconcilable under standard single-confusion-matrix definitions, as noted in Table 2. Wu et al.’s AUC was 0.908 internally and 0.900 externally for the extracted loosening task [15]. Masciulli et al.’s external evaluation included 14 patients [14]. The source reports accuracy of 0.786, precision of 0.85, and recall of 0.786, while describing the correct classification of all seven failed patients and four of seven non-failed patients. The precision and recall are retained as source-reported summaries with averaging unspecified; they are compatible with averaging across the two balanced classes and should not be interpreted as failure-positive PPV and sensitivity. These heterogeneous evaluation settings do not establish a ranking across architectures. Wei et al. [25] developed six ImageNet-pretrained CNN architectures using 357 AP THA radiographs from Hospital A, split in a 70/10/20% ratio for training/tuning/internal testing. Model configurations were chosen by tuning-set accuracy before evaluation on Hospital B (49 images) and Internet data (100 images). Five architectures achieved Hospital B AUC 1.000 (reported 95% CI 1.000–1.000); Internet AUCs ranged from 0.9832 to 0.9992. Wide-ResNet-101 had Hospital B AUC 1.000 and Internet AUC 0.999 (95% CI 0.994–1.000), with an Internet sensitivity of 100%, specificity of 98%, PPV of 98.04%, and NPV of 100%.
3.3. Diagnostic Models for Loosening and Broader Complications
Loppini et al. [18] reported high internal performance for broader hip-prosthesis failure classification, whereas Rahman et al. [19] evaluated aseptic loosening. Muscato et al. [16] evaluated a hybrid imaging-feature/classifier pipeline in developmental and additional cohorts. Guo et al.’s multilabel and binary experiments are reported separately in Section 3.1 [17]. High performance in these settings does not establish accuracy in an unselected postoperative population.
Diagnostic Detection of an Existing THA Dislocation
Wei et al. [25] evaluated current dislocation detection rather than risk prediction. Class activation maps were localized to THA components, but no controlled human–AI comparison, calibration assessment, prospective time-to-diagnosis evaluation, or patient-outcome benefit was reported. Thus, the study supports retrospective recognition feasibility but does not demonstrate emergency-department triage effectiveness.
3.4. Multiclass and Multilabel Complication Recognition
Wu et al. investigated several THA failure etiologies using AP and lateral radiographs in a multicenter design, with author-reported external AUC of approximately 0.900 [15]. Guo et al. evaluated coexisting complication labels in selected revision cases [17]. These are different recognition tasks and should not be merged into one multiclass endpoint.
3.5. Prognostic Models for Subsequent THA Complications and Failure
Masciulli et al. used sequential radiographs for failure prediction and included a small additional evaluation cohort; an absence of external testing is therefore not an accurate blanket description [14]. Bulloni et al. used evolving radiographic indices [20]. Khosravi et al. combined imaging and clinical variables; the multimodal survival model achieved a held-out-test C-index of 0.74 versus 0.64 for the clinical-only model (p = 0.02) [21]. Rouzrokh et al. used 92,584 eligible postoperative AP radiographs from a larger assembled THA cohort, with multiple images per patient and patient-ID partitioning [22]. Abul et al. evaluated general-purpose multimodal systems in a single-center cohort [24]; performance must be attributed to the specific system and prompt/threshold protocol rather than assigned collectively to all evaluated systems.
3.6. Explainability and Model Behavior
Explainability outputs included image heatmaps, feature-importance methods, and multimodal explanations [15,16,19,21,22]. These techniques should be attributed to the corresponding model rather than implying every study used both Grad-CAM and SHAP. Anatomically plausible heatmaps were not treated as proof of faithful explanations or improved clinician performance. In particular, Muscato et al. used SHAP, Rahman et al. used Score-CAM, and Khosravi et al. combined image saliency with SHAP for the survival-model features; these methods were not treated as interchangeable [16,19,21].
Wei et al. [25] used class activation mapping (CAM), not a reported formal clinical explanation-validation study. Heatmaps emphasized THA components in located and dislocated images. This qualitative localization does not establish explanation faithfulness, robustness, or improved reader decisions.
Taken together, these studies suggest that THA-related failure patterns may be detectable by AI models with high apparent discrimination in enriched datasets. However, because most studies in this group were retrospective, single-center, and at high overall risk of bias, the results should be interpreted as proof-of-concept evidence rather than definitive estimates of real-world diagnostic performance.
3.7. Risk-of-Bias Assessment
The existing PROBAST-2019 summary assigns high overall risk of bias to 10 of the 12 included studies and low overall risk to the studies by Wu et al. [15] and Khosravi et al. [21] (Table 3). These labels summarize the current appraisal rather than a PROBAST+AI reassessment. These risk-of-bias labels concern the assessed model-performance evidence and do not constitute judgments of clinical readiness, transportability, or demonstrated patient benefit.
Table 3.
Model-evaluation domain judgments from available evidence using PROBAST-2019.
4. Discussion
4.1. Comparative Interpretation by Clinical Task
Table 4 organizes the findings by clinical task and links the evidence to interpretation and implementation requirements. It separates diagnosis from prognosis and distinguishes the observed limitations from questions requiring further evaluation.
Table 4.
Clinical-task-based synthesis of AI applications for THA failure: evidence limitations and implications for implementation.
Relationship to Previous Systematic Reviews
Seven of the present 12 studies also appear among Stuani et al.’s ten formally included studies: Guo et al. [17], Loppini et al. [18], Masciulli et al. [14], Muscato et al. [16], Rahman et al. [19], Shah et al. [23], and Wu et al. [15]. Thus, the present synthesis includes all seven THA-containing reports in Stuani et al.’s [8] included set while excluding its three knee-only reports (Corti et al. [26], Kim et al. [27], and Lau et al. [28]). The five present reports which are absent from that formal included set—Bulloni et al. [20], Khosravi et al. [21], Rouzrokh et al. [22], Abul et al. [24], and Wei et al. [25]. Bulloni et al. [20]—was nevertheless cited and discussed by Stuani et al. [8]; it is an addition to their formal synthesis, not previously unrecognized literature. Four of these five reports predate Stuani et al.’s [8] search, so the difference cannot be attributed solely to a later cutoff. Only Abul et al. [24] (2026) postdates that search in the present five-report difference. Three of the present 12 studies overlap with Kim et al.’s systematic review [7]: Loppini et al. [18], Rahman et al. [19], and Shah et al. [23]. Lau et al.’s knee-only report [28], included in that review, is outside the present THA scope. The study-set differences are summarized in Table 5.
Table 5.
Comparison with prior systematic reviews and scope of the present synthesis.
The analytical extension is the explicit separation of current-dislocation recognition (Wei et al. [25]) from future-dislocation prediction (Rouzrokh et al. [22] and Khosravi et al. [21]), and of binary loosening experiments from multilabel complication recognition (Guo et al. [17]). Selected approximately balanced diagnostic datasets and low-event-rate prognostic cohorts answer different questions: their PPV, alert workload, prediction horizon, and image-versus-patient evaluation unit cannot be combined into a model ranking. Longitudinal methods (Bulloni et al. [20] and Masciulli et al. [14]) and multimodal systems (Khosravi et al. [21] and Abul et al. [24]) are interpreted within those task-specific requirements. These are additional applications of a clinical-decision framework, not claims that earlier reviews omitted external validation, calibration, or clinical utility.
Kim et al. demonstrated pooled sensitivity of 0.92 (95% CI 0.84–0.97) and specificity of 0.95 (95% CI 0.93–0.96), with substantial reported heterogeneity [7]. Those pooled diagnostic estimates are not benchmarks for the prognostic outcomes included here. Like Stuani et al., we do not pool incompatible clinical tasks [8]. The present review complements these syntheses by extracting THA-specific evidence from 12 reports and separately interpreting existing-complication diagnosis, future-dislocation prediction, longitudinal failure modelling, and multimodal-system evaluation. Its contribution is a task-specific interpretation of evaluation design and decision consequences, rather than novelty in the underlying overlapping studies; it does not establish improved patient outcomes.
4.2. Interpretation of the Evidence
High retrospective discrimination is a signal of technical feasibility, not evidence that a model improves outcomes in routine care. Geographic or institutional testing may support an evaluation beyond the development source, but prospective workflow effectiveness, calibration at relevant thresholds, and consequences for patient management remain distinct requirements. Small evaluation samples constrain precision; enriched sampling constrains applicability and predictive values; neither should be conflated automatically with a specific risk-of-bias judgment. The review does not identify a “best” architecture or current real-world performance estimate. Its contribution is the task-based interpretation of reported performance alongside evaluation design and implementation gaps [14,15,16,17,18,19,20,21,22,23,24,25].
4.2.1. Discrimination, Predictive Values, and Clinical Usefulness
High sensitivity is not in itself a defect. Rouzrokh et al.’s sensitivity of 89.0%, specificity of 48.8%, and PPV of 3.3% illustrate that a sensitivity-oriented operating point may entail substantial false-positive workload [22]. Approximately 96.7% of positive image classifications were false positives, distinct from the approximately 51.2% false-positive rate among negative examples. These are rounded fold summaries, not a reconstructed patient-level confusion matrix. AUC describes ranking, not calibration or net benefit.
The diagnostic/prognostic distinction is illustrated by Wei et al. [25]. Very high predictive values in approximately balanced current-dislocation datasets do not contradict the low PPV of the future-dislocation prediction in Rouzrokh et al. [22]. Sampling prevalence, event timing, clinical spectrum and operating thresholds differ. Near-perfect empirical AUCs and reported degenerate AUC intervals in a small, selected test sample do not imply zero uncertainty in other populations. Likewise, a non-significant difference between internal and external AUC is not an equivalence or non-inferiority demonstration.
Rouzrokh et al. weighted the dislocation class, oversampled training and tuning subsets, and selected models with sensitivities of at least 80%; test subsets retained imbalance [22]. Rare-event prevalence, model selection, and operating threshold jointly affect predictive values. A high NPV does not establish that follow-up or precautions can safely be reduced. Evaluation should retain the prediction horizon and image-versus-patient unit and report on calibration, precision–recall behavior, clinically selected thresholds, and net benefit.
4.2.2. Dataset Composition, Missingness, and Experimental Bias
Failure-enriched diagnostic cohorts and rare-dislocation prognostic cohorts represent different clinical spectra. PPV from a balanced or enriched sample cannot be transferred directly to routine surveillance. Excluding difficult images or incomplete records may narrow applicability; the direction and magnitude of any bias require study-specific assessment. Repeated images and bilateral hips require patient-level partitioning to prevent correlated observations from crossing development and test sets. Rouzrokh et al. explicitly partitioned them by patient ID [22]; leakage should not be alleged merely because multiple images were available.
Wei et al. [25] used previously de-identified clinical data and public Internet images; patient age and sex distributions were unavailable, although skeletal maturity was visually confirmed. One image per patient was reported, reducing within-patient repetition, but the independent provenance and potential cross-source overlap of Internet images were not established. Training and tuning images were augmented approximately 24-fold; the resulting 6863 images do not represent 6863 independent participants. The roughly balanced samples limit the direct extrapolation of PPV/NPV to routine emergency-department or surveillance prevalence.
4.2.3. Multimodal and Longitudinal Models
Khosravi et al. used an EfficientNet-B4/Swin-B hybrid classifier to learn imaging features, which were subsequently combined with clinical variables in a survival XGBoost model [21]. The final multimodal survival model showed higher discrimination than the clinical-only model on the held-out test set (C-index 0.74 vs 0.64; p = 0.02). The binary image-classifier metrics were evaluated in the 721 test participants with definite five-year outcomes, whereas the survival analysis retained participants with censoring. These distinct analysis sets should not be conflated. This within-study improvement does not establish its external transportability, calibration, or clinical benefit. Longitudinal methods [14,20] require prior examinations, consistent chronology, and an approach to irregular intervals and missing visits. Clinical usefulness depends on these practical requirements as well as discrimination.
4.3. Engineering Implementation and Clinical Translation
Beyond predictive performance and generalizability, clinical implementation depends on the reliability of the complete imaging-to-decision pipeline and the integration of model outputs into an actionable clinical workflow. Relevant failure points include image retrieval and view recognition, prosthesis localization, cropping, image-quality assessment, inference, and communication of an actionable result. Rouzrokh et al., for example, used a separate object-detection model before dislocation classification, illustrating the dependence of downstream predictions on preprocessing [22]. An evaluation should therefore include processing failures and end-to-end performance, not only results from successfully curated inputs. A deployment assessment should additionally examine workflow integration, alert burden, performance across institutions and implant designs, and monitoring for changes in acquisition or patient populations.
For current-dislocation triage, Wei et al. [25] identifies a plausible workflow use but does not measure worklist reprioritization, time-to-diagnosis/reduction, clinician workload, or patient outcomes. The pipeline converts images to JPEG/PNG, normalizes channels, and resizes inputs; deployment must therefore assess end-to-end performance on native clinical image flows, failed preprocessing, acquisition/view variation, alert burden and escalation rules. Model accuracy on selected images is not sufficient evidence that triage delay is reduced.
The included studies use saliency, Grad-CAM, SHAP, attention, or feature-importance outputs, but these are not equivalent explanations or evidence of clinical validity. Rouzrokh et al. presented anatomically plausible saliency maps [22]; this does not establish its stability, faithfulness, or benefit to clinicians. Explanation quality should be evaluated in controlled human–AI studies alongside performance, confidence, and decision consequences.
4.4. Methodological Limitations and Limitations of This Review
Studies varied in clinical tasks—diagnosis of existing complications versus prediction of subsequent outcomes—and in sampling, imaging inputs, reference standards, outcome definitions, prediction horizons, and evaluation procedures. Their performance estimates were therefore not directly comparable, supporting a structured narrative synthesis rather than pooling or ranking model architectures.
Several studies used retrospective, selected, or outcome-enriched datasets and excluded inadequate images or incomplete records, potentially limiting its applicability to consecutive postoperative populations and routine imaging. Although some included institutional or Internet-derived external datasets, retrospective testing did not establish prospective workflow effectiveness. Small evaluation samples also constrain precision despite high reported discrimination.
Calibration and clinical usefulness could not be assessed consistently across models. High AUC or accuracy alone does not establish reliable individual risk estimates, acceptable false-positive workload, or improved clinical decisions. Predictive values also depend on sample prevalence and composition. Consequently, reported performance does not demonstrate improved implant survivorship, shorter treatment delays, or better patient outcomes. Masciulli et al. explicitly identify supplementary calibration plots [14]; this does not provide a consistent calibration assessment across the full evidence set. Some published performance summaries use different averaging conventions or are internally unreconciled. Such values were identified as source-reported and were not used to reconstruct clinical operating characteristics or rank models across studies.
The review was restricted to English-language full-text publications within the specified period, potentially excluding relevant evidence in other languages or formats. The unregistered protocol limits the independent verification of prespecified methodological decisions. Its appraisal also depended on the completeness and clarity of published reporting, with insufficient information leaving some judgments uncertain. These limitations warrant a cautious interpretation. The evidence does not support firm recommendations for routine clinical adoption or firm conclusions about patient benefit. The use of PROBAST-2019 rather than PROBAST+AI and the condensed study-level appraisal are additional limitations, particularly for publications containing multiple models or distinct development and evaluation components.
4.5. Clinical Implications
At present, the most realistic clinical role for these systems is as investigational decision-support rather than as autonomous clinical tools. In principle, validated high-sensitivity models could support radiographic triage, help prioritize suspicious cases for specialist review, or contribute to a structured risk assessment when combined with clinical information. However, such use remains premature for most published models because evidence of transportability, calibration, clinical workflow integration, and patient-level benefit is still limited.
4.6. Future Directions
4.6.1. Prospective, Multicenter Evaluation and Calibration
Large, prospective multicenter studies are needed to evaluate model performance across institutions, imaging protocols, and implant designs, with pre-registered analysis plans and harmonized outcome definitions. External validation should report not only AUC and accuracy but also calibration metrics, decision thresholds, and clinically meaningful measures such as PPV, NPV, and net benefit on decision-curve analysis.
4.6.2. Standardized Reporting and Benchmarks
Adherence to AI-specific reporting frameworks should be encouraged, including explicit detailing of data flow, class imbalance handling, missing-data strategies, and ground-truth adjudication processes. Establishing shared, de-identified benchmark datasets could facilitate reproducibility and a fair comparison of competing architectures.
4.6.3. Multimodal and Longitudinal Modeling
Building on Masciulli F et al. [14], Bulloni M et al. [20], Khosravi B et al. [21], the future models should systematically integrate temporal radiographic data with clinical variables, implant metadata, and possibly other imaging modalities to better capture the multifactorial nature of failure and progression. Architectures that explicitly model trajectories may yield more clinically interpretable risk estimates over fixed time horizons.
4.6.4. Explainability, Human–AI Interaction, and Fairness
Explainability methods need to be evaluated beyond visual appeal, via controlled user-studies assessing whether saliency maps or feature-importance tools meaningfully improve clinician accuracy, confidence, and calibration. Research should examine how to best present risk estimates and heatmaps within PACS or reporting interfaces to minimize cognitive load. Subgroup analyses by age, sex, BMI, implant design, and comorbidities are needed to detect and mitigate performance disparities, particularly given that most current datasets have skewed demographics.
4.6.5. Clinical Impact and Health-Economic Evaluation
Randomized or quasi-experimental implementation studies will be required to determine whether deploying these models improves patient outcomes, reduces unnecessary imaging or clinic visits, and is cost-effective compared with standard care.
5. Conclusions
Artificial intelligence models incorporating radiographic information show technical promise for identifying selected existing complications and predicting subsequent adverse outcomes after total hip arthroplasty. However, reported predictive performance should not be equated with reliable clinical usefulness. Most included studies were judged to have high overall risk of bias, although a minority received low-risk judgments under the assessment framework used. These judgments concern the methodological credibility of the evaluated performance estimates and do not, by themselves, establish clinical readiness or patient benefit. Heterogeneity in clinical tasks, populations, outcome definitions, and evaluation methods limits direct comparison across studies. Independent evaluation, calibration at clinically relevant thresholds, transparent reporting, and prospective assessment of human–AI workflows remain priorities. The current evidence supports further investigation of AI-assisted post-THA imaging assessment but does not establish improved implant survivorship, better patient outcomes, or readiness for widespread routine adoption.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics16233969/s1, Supplementary Material File S1—PRISMA checklist; Supplementary Material File S2—Database and supplementary journal search strategies.
Author Contributions
N.K. contributed to conceptualization, data curation, analysis, investigation, methodology, project administration, resources, supervision, validation, visualization, drafting, and revision. S.P. contributed to analysis, methodology, validation, and revision. R.V. contributed to conceptualization, data curation, analysis, methodology, supervision, validation, and revision. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not Applicable.
Data Availability Statement
The study-level characteristics, extracted performance measures, and appraisal summaries used in this review are presented in the manuscript tables. No new patient-level dataset was generated.
Conflicts of Interest
Author Netaji Khot was employed by the company Orchid Orthopedic Solutions. The remaining authors declare that they have no relevant financial or non-financial interests to disclose in this research review. No author has any proprietary or commercial interest in any materials, methods, software, or devices discussed in this review. No company or external sponsor had any role in the design, conduct, analysis, or reporting of this review.
Abbreviations
| mAP | Mean average precision |
| MC-CSRA | Multi-Coefficient Class-Specific Residual Attention |
| ML | Machine Learning |
| NPV | Negative predictive value |
| OA | Osteoarthritis |
| PACS | Picture Archiving and Communication System |
| PJI | Periprosthetic joint infection |
| PNG | Portable Network Graphics |
| PPF | Periprosthetic fracture |
| PPV | Positive predictive value |
| Prec | Precision |
| PRISMA | Preferred Reporting Items for Systematic Reviews and Meta-Analyses |
| PROBAST | Prediction model Risk Of Bias Assessment Tool |
| PROSPERO | International Prospective Register of Systematic Reviews |
| RF | Random Forest |
| ROC | Receiver operating characteristic |
| Sens | Sensitivity |
| SHAP | SHapley Additive exPlanations |
| Spec | Specificity |
| SVM | Support vector machine |
| THA/THR | Total hip arthroplasty/total hip replacement |
| Val | Validation |
| XAI | Explainable artificial intelligence |
| AI | Artificial intelligence |
| Acc | Accuracy |
| AP | Anteroposterior |
| AUC | Area under the receiver operating characteristic curve |
| BMI | Body Mass Index |
| CAM | class activation mapping |
| CFS | Channel feature stream |
| CI | Confidence interval |
| CNN | Convolutional neural network |
| CV | Cross validation |
| Dev | Development |
| DL | Deep learning |
| Ext | External |
| F | Female |
| F1 | F1 score (harmonic mean of precision and recall) |
| GFS | Global feature stream |
| Grad-CAM, Score-CAM | Gradient-weighted/score-based class activation mapping methods for visual explainability |
| GRU | Gated recurrent unit |
| IEEE | Institute of Electrical and Electronics Engineers |
| Int | Internal |
| JBJS | The Journal of Bone and Joint Surgery |
| JPEG | Joint Photographic Experts Group |
| Lat | Lateral |
References
- Shichman, I.; Roof, M.; Askew, N.; Nherera, L.; Rozell, J.C.; Seyler, T.M.; Schwarzkopf, R. Projections and Epidemiology of Primary Hip and Knee Arthroplasty in Medicare Patients to 2040–2060. JBJS Open Access 2023, 8, e22.00112. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stubnya, B.G.; Schulz, M.; Váncsa, S.; Szilágyi, G.S.; Szatmári, A.; Bejek, Z. Global Trends in Joint Arthroplasty: A Systematic Review and Future Projections. J. Clin. Med. 2025, 14, 8214. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lisacek-Kiosoglous, A.B.; Powling, A.S.; Fontalis, A.; Gabr, A.; Mazomenos, E.; Haddad, F.S. Artificial intelligence in orthopaedic surgery. Bone Jt. Res. 2023, 12, 447–454. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kurtz, S.; Ong, K.; Lau, E.; Mowat, F.; Halpern, M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J. Bone Jt. Surg. Am. 2007, 89, 780–785. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Longo, U.G.; De Salvatore, S.; Piccolomini, A.; Ullman, N.S.; Salvatore, G.; D’HOoghe, M.; Saccomanno, M.; Samuelsson, K.; Papalia, R.; Pareek, A. Revolutionizing total hip arthroplasty: The role of artificial intelligence and machine learning. J. Exp. Orthop. 2025, 12, e70195. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zou, W.; Gao, B.; Liu, Y.; Zhang, X. Predicting Dislocation Risk in THA Squat Using Musculoskeletal Modeling and Random Forest: The Role of Acetabular Cup Positioning. In Proceedings of the 2025 6th International Conference on Computer Science and Management Technology ICCSMT ‘25, Xiamen, China, 26–28 December 2025; pp. 103–107. [Google Scholar] [CrossRef] [Scilit]
- Kim, M.S.; Kim, J.J.; Kang, K.H.; Lee, J.H.; In, Y. Detection of Prosthetic Loosening in Hip and Knee Arthroplasty Using Machine Learning: A Systematic Review and Meta-Analysis. Medicina 2023, 59, 782. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stuani, R.; Di Maio, M.; Di Matteo, V.; Chiappetta, K.; Grappiolo, G.; Loppini, M. Performance of Artificial Intelligence Models in Radiographic Image Analysis for Predicting Hip and Knee Prosthesis Failure: A Systematic Review. Bioengineering 2026, 13, 122. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Page, M.J.; McKenzie, J.E.; Bossuyt, P.M.; Boutron, I.; Hoffmann, T.C.; Mulrow, C.D.; Shamseer, L.; Tetzlaff, J.M.; Akl, E.A.; Brennan, S.E.; et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ 2021, 372, n71. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wolff, R.F.; Moons, K.G.; Riley, R.; Whiting, P.F.; Westwood, M.; Collins, G.S.; Reitsma, J.B.; Kleijnen, J.; Mallett, S.; for the PROBAST Group. PROBAST: A Tool to Assess the Risk of Bias and Applicability of Prediction Model Studies. Ann. Intern Med. 2019, 170, 51–58. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Moons, K.G.M.; A A Damen, J.; Kaul, T.; Hooft, L.; Navarro, C.A.; Dhiman, P.; Beam, A.L.; Van Calster, B.; Celi, L.A.; Denaxas, S.; et al. PROBAST+AI: An updated quality, risk of bias, and applicability assessment tool for prediction models using regression or artificial intelligence methods. BMJ 2025, 388, e082505. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chen, C.-C.; Wu, C.-T.; Chen, C.P.C.; Chung, C.-Y.; Chen, S.-C.; Lee, M.S.; Cheng, C.-T.; Liao, C.-H. Predicting the Risk of Total Hip Replacement by Using A Deep Learning Algorithm on Plain Pelvic Radiographs: Diagnostic Study. JMIR Form. Res. 2023, 7, e42788. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hidaka, R.; Matsuda, K.; Igari, T.; Takeuchi, S.; Imoto, Y.; Yagi, S.; Kawano, H. Development and accuracy of an artificial intelligence model for predicting the progression of hip osteoarthritis using plain radiographs and clinical data: A retrospective study. BMC Musculoskelet. Disord. 2024, 25, 893. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Masciulli, F.; Corti, A.; Lindemann, A.; Chiappetta, K.; Loppini, M.; Corino, V.D.A. Hip prosthesis failure prediction through radiological deep sequence learning. Int. J. Med. Inform. 2025, 196, 105802. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wu, L.; Wang, B.; Lin, B.; Li, M.; Wu, Y.; Si, H.; Zeng, Y.; Lu, L.; Gao, L.; Chen, Z.; et al. A Deep Learning-Based Clinical Classification System for the Differential Diagnosis of Hip Prosthesis Failures Using Radiographs. J. Bone Jt. Surg. 2025, 107, 1798–1809. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Muscato, F.; Corti, A.; Manlio Gambaro, F.; Chiappetta, K.; Loppini, M.; Corino, V.D.A. Combining deep learning and machine learning for the automatic identification of hip prosthesis failure: Development, validation and explainability analysis. Int. J. Med. Inform. 2023, 176, 105095. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Guo, S.; Zhang, J.; Li, H.; Zhang, J.; Cheng, C.K. A multi-branch network to detect post-operative complications following hip arthroplasty on X-ray images. Front. Bioeng. Biotechnol. 2023, 11, 1239637. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Loppini, M.; Gambaro, F.M.; Chiappetta, K.; Grappiolo, G.; Bianchi, A.M.; Corino, V.D.A. Automatic Identification of Failure in Hip Replacement: An Artificial Intelligence Approach. Bioengineering 2022, 9, 288. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rahman, T.; Khandakar, A.; Islam, K.R.; Soliman, M.; Islam, M.T.; Elsayed, A.; Qiblawey, Y.; Mahmud, S.; Rahman, A.; Musharavati, F.; et al. HipXNet: Deep Learning Approaches to Detect Aseptic Loos-ening of Hip Implants Using X-Ray Images. IEEE Access 2022, 10, 53359–53373. [Google Scholar] [CrossRef] [Scilit]
- Bulloni, M.; Gambaro, F.M.; Chiappetta, K.; Grappiolo, G.; Corino, V.; Loppini, M. AI-based hip prosthesis failure prediction through evolutional radiological indices. Arch. Orthop. Trauma Surg. 2024, 144, 895–907. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Khosravi, B.; Rouzrokh, P.; Kremers, H.M.; Larson, D.R.; Johnson, Q.J.; Faghani, S.; Kremers, W.K.; Erickson, B.J.; Sierra, R.J.; Taunton, M.J.; et al. Patient-specific Hip Arthroplasty Dislocation Risk Calculator: An Explainable Multimodal Machine Learning–based Approach. Radiol. Artif. Intell. 2022, 4, e220067. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rouzrokh, P.; Ramazanian, T.; Wyles, C.C.; Philbrick, K.A.; Cai, J.C.; Taunton, M.J.; Kremers, H.M.; Lewallen, D.G.; Erickson, B.J. Deep Learning Artificial Intelligence Model for Assessment of Hip Dislocation Risk Following Primary Total Hip Arthroplasty From Postoperative Radiographs. J. Arthroplast. 2021, 36, 2197–2203.e3. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shah, R.F.; Bini, S.A.; Martinez, A.M.; Pedoia, V.; Vail, T.P. Incremental inputs improve the automated detection of implant loosening using machine-learning algorithms. Bone Jt. J. 2020, 102-B, 101–106. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Abul, M.S.; Sevim, Ö.F.; Kılıç, N.C.; Agir, M.; Tuncay, I. Diagnostic Accuracy of Artificial Intelligence Systems in Early Prediction of Dislocation and Aseptic Loosening After Total Hip Arthroplasty. J. Arthroplast. 2026, in press. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wei, J.; Li, D.; Sing, D.C.; Yang, J.; Beeram, I.; Puvanesarajah, V.; Della Valle, C.J.; Tornetta, P.; Fritz, J.; Yi, P.H. Detecting total hip arthroplasty dislocations using deep learning: Clinical and Internet validation. Emerg. Radiol. 2022, 29, 801–808. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Corti, A.; Galante, S.; Rauch, R.; Chiappetta, K.; Corino, V.; Loppini, M. Leveraging transfer learning for predicting total knee arthroplasty failure from post-operative radiographs. J. Exp. Orthop. 2024, 11, e70097. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kim, M.S.; Cho, R.K.; Yang, S.C.; Hur, J.H.; In, Y. Machine Learning for Detecting Total Knee Arthroplasty Implant Loosening on Plain Radiographs. Bioengineering 2023, 10, 632. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lau, L.C.M.; Chui, E.C.S.; Man, G.C.W.; Xin, Y.; Ho, K.K.W.; Mak, K.K.K.; Ong, M.T.Y.; Law, S.W.; Cheung, W.H.; Yung, P.S.H. A novel image-based machine learning model with superior accuracy and predictability for knee arthroplasty loosening detection and clinical decision making. J. Orthop. Transl. 2022, 36, 177–183. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.