The Lead Tip Beyond the Myocardial Contour: A Narrative Review of Cardiac Implantable Electronic-Device Perforation Detected by Computed Tomography

,
,
,
and
1
West China Hospital Sichuan University, Meishan Hospital, Meishan 620010, China
2
Department of Cardiovascular Medicine, Meishan People’s Hospital, Meishan 620010, China
*
Author to whom correspondence should be addressed.

Abstract

Computed tomography (CT) frequently shows the tip of a cardiac implantable electronic device (CIED) lead, or the fixation tine of a leadless pacemaker, beyond the myocardial contour in patients without symptoms or hemodynamic compromise. What such an image means is a separate question the literature has not settled. This narrative review sets out the definitions in use, the three CT cohorts behind the argument, the observed natural history and the management choices that follow, keeping anatomical observations apart from clinical events. Reported penetration was 15/100 conventional-device recipients, 26/463 conventional leads (25/271 patients) and 15/88 leadless recipients; a conference abstract reported 12 of 31 leadless recipients (38.7%). Sampling, observation unit, device structure and imaging interval differ across these sources, so none supports an incidence figure or a device comparison. No clinical perforation developed over 2.7 years among 25 patients with conventional-lead findings, and no related adverse event occurred over a mean of 12.4 months among 15 leadless recipients; both figures concern findings already silent at CT, neither cohort sized the residual risk, and imaging accuracy measured in recognized perforation does not carry over to screening. Implantation guidance sets no screening or intervention criteria. What the evidence supports is individual assessment, not routine CT screening, universal intervention or fixed warning thresholds. Assessment criteria, a reporting checklist, and emerging imaging and automation tools are summarized.

1. Introduction

A CIED lead can cross the outer myocardial contour on CT while the patient reports nothing and device interrogation looks satisfactory. Three readings of that image are still available: an injury that calls for intervention, an anatomical variant that can be watched, or an artefact of uncertain meaning. Mechanisms, clinical presentations, diagnosis and treatment of lead perforation are covered in general reviews [1,2]. CT-detected subclinical penetration is a narrower problem and needs separate treatment, because both its definition and its clinical consequences differ from those of symptomatic perforation.
In practice, the question is seldom whether to act at all, but which of three defensible responses fits: reassurance, structured surveillance, or revision [1,2,3]. Each carries its own risk profile, and no evidence yet tells the clinician which one to choose.
Perforation that reaches clinical attention is uncommon but consequential. Waddingham et al. identified 70 perforations across three UK centres; 51 had been implanted locally, and the reported rate of 0.50% refers to these local implants counted against 10,631 procedures (all 70 cases against that denominator would give 0.66%) [3]. Presentation is often delayed and non-specific: the median time to diagnosis was 9 days (range 0–989), chest pain was the commonest symptom, at 46%, and lead-parameter abnormalities appeared in 86% of the whole cohort and 98.6% of perforations diagnosed more than 24 h after implantation [3]. Tamponade complications were 17% and carried higher mortality [3]. Systematic reviews confirm that estimates shift with device type and with how completely events are ascertained [4]. Events like these are not directly comparable with anatomical findings in selected patients imaged by CT for other indications.
Hirschl et al., Lim et al. and Lee et al. supply the central evidence for this review [5,6,7]. Their definitions, denominators, imaging indications and follow-up come first, followed by what can reasonably be inferred about diagnosis and management. What the literature lacks is equally specific. No cohort defines a protrusion depth or effusion pattern that predicts progression. No study follows asymptomatic CT findings with a prespecified protocol. No guideline sets a threshold for action [8,9,10,11,12]. This review, therefore, does four things: it keeps anatomical findings and clinical events apart; it carries per lead and per patient denominators separately; it states the assessment criteria the cohorts actually used; and it condenses them into a reporting checklist for initial evaluation, labelled as unvalidated. The iceberg analogy conveys incomplete clinical visibility; it does not measure the occult reservoir, and no pathway from silent penetration to tamponade has been quantified.
Observations along these lines now extend to leadless devices [7], yet current guidance on implantation and lead management sets no criteria for CT-detected penetration in recipients without symptoms [8,9,10,11]. The scope here is explicit: transvenous atrial and ventricular leads, tine-based leadless pacemakers, and findings detected at any time after implantation. Helix-fixation leadless devices and conduction-system pacing leads generate the same CT question but have no published cohorts, and are not considered further. Whether the finding occurs is settled; what a clinician should do about it is not.

2. Methods

The synthesis here is narrative and purposive, not an exhaustive systematic review. Three CT cohorts, chosen as core evidence against the criteria below, sit within a wider frame of broader reviews, series of clinically recognized perforation, device registries and lead-management guidance. Bibliographic identifiers were checked against Crossref and PubMed in October 2026, and the database search was rerun in November 2026 to record screening and selection counts (Table 1). We registered no protocol and performed no meta-analysis or formal risk-of-bias scoring; the counts in Table 1 record a narrative selection process, rather than a PRISMA flow.
Table 1. Search and selection flow for this narrative review.
PubMed and Europe PMC were searched from January 2000 to September 2026; extending the window earlier would add only isolated case descriptions, which this review treats as illustrative material identified through reference lists. The full strategy combined three concepts with Boolean operators: (“cardiac implantable electronic device”[tiab] OR CIED[tiab] OR pacemaker[tiab] OR “implantable cardioverter defibrillator”[tiab] OR “leadless pacemaker”[tiab]) AND (perforation[tiab] OR penetration[tiab] OR “lead perforation”[tiab]) AND (“computed tomography”[tiab] OR “cardiac CT”[tiab]). The same concepts were searched in Europe PMC over titles and abstracts. Reference lists of the retrieved reports were screened for additional relevant records. Core evidence comprised studies that reported CT-detected myocardial crossing by a transvenous lead tip or a leadless pacemaker fixation tine in a population described as asymptomatic, with an explicit numerator, denominator and observation unit. Studies reporting only clinically recognized perforation, pericardial effusion, tamponade, or composite device complications were treated as contextual evidence, rather than as core evidence. Cohorts that reported both CT-defined subclinical findings and a small number of clinically recognized events were retained as core evidence, with those events described, but not pooled. No language restriction was imposed a priori, although no non-English report meeting the eligibility criteria was identified.
Each eligible cohort was charted for the clinical reason for imaging, population and device type, the definition of penetration, the interval from implantation to CT, and reported follow-up. Where available, published text and tables were read; bibliographic records and abstracts served for additional checks. Case reports appear here only to illustrate unusual or delayed presentations, never to estimate incidence or the probability of progression. General device complications, pericardial effusion, tamponade and anatomically defined perforation were kept as distinct outcomes.
Study populations, device structures, CT protocols and observation times all differ, so proportions are given descriptively, and are not pooled. The cohorts do not even share an observation unit, with per lead in one and per recipient in another, which puts the numerators beyond pooling, as well. Electrical changes and management considerations are read within the populations that reported them. The research priorities proposed at the end are our own reading of the gaps, not validated screening rules or a consensus statement.

3. Definitions and Evidence from CT Cohorts

3.1. Anatomical Findings and Clinical Phenotypes

The comparison this review asks forneeds to be anatomical. A conventional passive-fixation lead carries short tines near its tip that hook into trabeculae; an active-fixation lead carries a metal helix that is screwed into the myocardium, and its tip sits at the endocardial surface after implantation [1,2]. A tine-based leadless pacemaker is a capsular device whose four tines grip the trabeculae, with the device body remaining inside the ventricular cavity [1,2]. The myocardial wall between endocardium and epicardium measures only a few millimetres over the right ventricle, thinner toward the apex, and the outer contour on CT borders epicardial fat before the parietal pericardium is reached. Crossing the outer contour, therefore, has device-specific meaning. A lead tip that passes it has traversed the wall. A leadless tine that passes it represents tine penetration while the body stays intracardiac. Displacement of the device body is a third event again. Estimates that mix these structures measure different things.
In conventional systems, the core CT definition is traversal of the lead tip beyond the outer myocardial layer, with no symptom or sign attributable to perforation. Lim et al. required the finding on at least two CT views [6]. For the tine-based leadless device that Lee et al. studied, subclinical perforation meant fixation tines extending past the outer myocardial contour without symptoms, pericardial effusion, hemodynamic instability or malfunction [7]. A protruding tine is not the same event as penetration by a conventional lead tip, and neither is displacement of a leadless device body. Every reported estimate should, therefore, carry its device-specific definition.
Clinical status and electrical function add two further dimensions. A patient may present with symptoms or circulatory compromise, may stay asymptomatic despite an electrical abnormality, or may stay asymptomatic with stable electrical parameters. These are useful clinical descriptions; they are not established stages of a single disease process. Whether a larger population of smaller, undetected micro-perforations exists is unquantified in the available cohorts and cannot be treated as an observed category.

3.2. Selected Cohorts and Observation Units

Hirschl et al. read chest CT studies from 100 consecutive patients carrying conventional pacemakers or defibrillators; images were acquired on multidetector CT, with gating not reported, and were reviewed by two cardiothoracic radiologists with a third reader for disagreements [5]. Fifteen patients had a lead classified as perforated. The counts were, per lead: nine of 61 right atrial leads and six of 100 right-ventricular leads, and the atrial proportion exceeded the ventricular one (chi-squared test, p < 0.05) [5]. Electrical parameters, available for 79 patients, did not differ significantly between perforated and non-perforated leads [5]. The sample shows that an anatomical finding can be electrically silent, though it describes only patients who underwent imaging in the first place.
Lim et al. studied 271 patients with 463 leads, all imaged by cardiac CT for cardiovascular anatomy or coronary assessment; symptomatic patients and inadequate studies were excluded, and images were read by one radiologist and two cardiologists [6]. Penetration was detected in 26/463 leads (5.6%), involving 25/271 patients (9.2%). Screw-in ventricular leads carried a higher detected proportion than tined ventricular leads—14/105 (13.3%) against 5/153 (3.3%)—and none was found among 56 septal ventricular leads [6]. No significant difference appeared between atrial and ventricular leads, screw-in and tined atrial leads, pacing and defibrillation ventricular leads, or MR-conditional and MR-unsafe screw-in ventricular leads [6]. The per lead analysis did not adjust for covariates or for the clustering of several leads within one patient, and the all-lead proportion spans atrial, ventricular and left-ventricular leads, so it is not a fourth independent comparison group.
The earliest CT series made that distinction explicit, well before the cohorts discussed above. One early ECG-gated multidetector CT series examined 52 right-ventricular and 35 right-atrial leads. using CT criteria, five of 17 apical and one of 35 non-apical right ventricular leads crossed the myocardium; none of the 35 atrial leads did. Two leads protruding 5 and 15 mm beyond the outer myocardial border carried pericardial effusions, and were repositioned. Four apparent radiographic perforations, protruding an average of 1.5 ± 0.5 mm, were not [13]. The authors determined that the clinical significance and natural history of leads with only the appearance of perforation were uncertain, and that point is still the central interpretive problem for the cohorts reviewed here.
Lee et al. included 88 consecutive leadless pacemaker recipients who had undergone CT of the heart, chest or abdomen for other indications; two cardiologists and one radiologist assessed tine position [7]. The mean interval from implantation to CT was 8.6 ± 9.2 months. Fifteen patients (17.0%) had fixation tines beyond the outer myocardial contour, and the commonest implant site was the junction between the septum and the right-ventricular free wall [7]. A conference abstract from the same group reported a higher figure in a smaller series: 12 of 31 recipients (38.7%), with two of the twelve showing moderate pericardial effusion; setting those two aside, still leaves ten of thirty-one recipients (32%) with protrusion alone [14]. Their exact 95% confidence intervals (9.9–26.6% and 21.8–57.8%) overlap; overlap alone does not establish that the two cohorts share an underlying rate, and the spread shows how far small, selected cohorts can move [7,14]. This is a selected post-implant imaging cohort, rather than a prospective screening sample of all leadless implants. Table 2 sets the three studies side by side; Figure 1 keeps per lead and per recipient estimates apart.
Table 2. Core CT cohorts of anatomically defined subclinical penetration.
Figure 1. CT-detected subclinical penetration in selected cohorts. (A) Per lead estimates from Lim 2021; the overall all-lead estimate is separated from the ventricular fixation subgroups. (B) Per recipient estimates for conventional devices from Hirschl 2007 [5] and Lim 2021 [6]. (C) Per recipient estimate for the tine-based leadless device from Lee 2026 [7]. Points and labels give the counts reported in each study and the corresponding proportions; error bars are exact binomial (Clopper–Pearson) 95% confidence intervals, calculated for this review from those counts, and they describe sampling uncertainty within each study. Selection, CT timing and device-specific definitions differ across the three studies, so visual proximity carries no implication of comparative risk.

3.3. What the Proportions Can and Cannot Show

Across the selected CT studies, apparent anatomical penetration coexists with satisfactory clinical status. Prevalence in unselected CIED recipients is a different quantity, and one these studies do not supply. Referral for CT, exclusion of symptomatic cases or poor-quality images, survival until imaging and the time elapsed after implantation all shape the observed sample [5,6,7]. Selection by itself does not tell us whether the resulting estimate runs high or low, but one bias runs one way: where an acquisition was not ECG-gated and the lead was not the imaging target, an artefact could push the observed proportion up. Hirschl et al. did not report gating, and Lee et al. drew on studies requested for other indications [5,7]. Five CT reports of right-ventricular perforation in one 12-month audit proved artefactual, on further investigation [15]; of 35 leads called perforating on non-gated CT before extraction, 29 were shown by right ventriculography to lie within the right ventricle [16]; and in a pre-extraction cohort of 143 patients with 348 leads, protrusion of at least 5 mm affected only 13 patients (9.1%), against 46% with protrusion of any depth [17]. The 5.6% per lead and 17.0% per recipient figures, therefore, support neither a claim of device difference at any magnitude nor a lower bound on true biological prevalence.
The 11.2% asymptomatic fraction in Liu et al. rests on 153 participants assembled from case reports and case series of recognized perforation [18]. Reported perforation cases form its denominator, not all device recipients, and not all CT-positive lesions. Data of this sort describe clinical presentations; they do not measure the hidden part of an iceberg. Anatomical findings, recognized clinical cases, and hypothetical undetected lesions need to stay distinct at every step of interpretation.

4. Observed Natural History

What follow-up does exist offers some reassurance, and no more. Among patients with CT-detected subclinical findings, Lim et al. saw no clinical cardiac perforation over 2.7 years (IQR 1.1–4.5) [6]. None of the 15 CT-positive leadless recipients in Lee et al. had a related adverse event during a mean 12.4 ± 12.3 months [7]. Observations like these count against the assumption that every anatomical penetration ends in a clinically significant event. They count just as much against the opposite assumption, that the finding is harmless respecting the life of a device.
Both cohorts are small, and were assembled retrospectively around patients who happened to undergo imaging, and both excluded symptomatic patients or those with effusion at the time of CT. The reassuring follow-up, therefore, applies only to findings that were already silent and stable when imaged; it says nothing about lesions that come to attention because of symptoms. Anyone who had an early major complication may never appear in a later asymptomatic CT cohort, and follow-up was not uniform. Treatment choices, competing illness and the sensitivity of outcome ascertainment bear on the reading, as well. Zero observed events do not exclude a risk that is clinically important, yet uncommon. With no event among the 25 Lim patients with subclinical findings, the one-sided exact 95% upper bound on the event proportion is approximately 11%; with the 15 CT-positive Lee recipients, the corresponding bound is approximately 18% [6,7]. Both bounds assume complete follow-up, which neither cohort formally demonstrated. These bounds inform without reassuring.
Delayed perforation is documented months or years after implantation, including isolated very late presentations [19,20,21,22,23]. Possibility is settled by these reports; rate is not. A delayed event may come from a penetration present and unrecognized since implantation, or from mechanical change arising later. The case literature does not separate the two mechanisms reliably. The iceberg analogy describes clinical visibility and nothing more; it is not a validated natural-history model.

5. Recognition and Diagnostic Limitations

5.1. Electrical Interrogation

Pacing threshold, sensing amplitude, and impedance may shift when perforation is clinically recognized. Waddingham et al. recorded abnormal lead parameters in 86% of their perforation cohort and in 98.6% of the perforations diagnosed more than 24 h after implantation [3]; Rajkumar et al. found electrical abnormalities in 16 of 18 recognized cases [24]. Both are proportions within selected cases, not prospectively validated sensitivities for screening an unselected population. A parameter change has other explanations too, dislodgement and other lead faults among them, and its predictive value depends on the clinical setting.
The CT cohorts answer a different question. Hirschl et al. found no significant difference in electrophysiological parameters between perforated and non-perforated leads [5]. Lim et al. reported the same for sensing, stimulation threshold, and impedance, in subclinical perforation [6]. Normal interrogation, therefore, leaves anatomical penetration in play, while an abnormal reading leaves it unconfirmed. Table 3 sets out how to read these results, without proposing numerical warning thresholds for general use.
Table 3. Electrical findings and their limits in evaluating lead perforation.

5.2. Imaging in Suspected Clinical Perforation

Chest pain of new onset, dyspnea, diaphragmatic stimulation, loss of capture, or circulatory deterioration after implantation call for evaluation of device complications, perforation among them. Echocardiography identifies pericardial effusion and gauges its hemodynamic consequences. CT settles lead position when the anatomical diagnosis is still in doubt [3,24,25]. A chest radiograph or echocardiogram that shows no protruding lead does not, by itself, exclude perforation.
Rajkumar et al. reported CT sensitivity and specificity of 100% and 85.7%, against 41.2% and 84.2% for transthoracic echocardiography and 27.7% and 94.4% for chest radiography [24]. Waddingham et al. put CT sensitivity at approximately 97% [3]. The denominators behind these percentages are small, and differ by modality: the Rajkumar series drew 18 recognized perforations from 426 extraction procedures; CT was performed in 14 patients, transthoracic echocardiography in 17, and chest radiography in all 18; and the accuracy comparison came from a blinded review of 30 anonymized datasets, 15 with, and 15 without, perforation, read by three imaging experts [24]. The report gives the percentages, but not the counts behind them, so the intervals cannot all be reconstructed. Where the stated denominators do constrain the arithmetic, precision is limited: CT was reported to detect every case in which it was performed, so the sensitivity of 100% rests on no more than 14 positive studies, and, even at that most favourable denominator, the exact 95% interval is 76.8–100%; the specificity of 85.7% was measured against a comparably short list of control scans [24]. A ranking built on a few dozen blinded reads is a hypothesis awaiting external testing, not a calibrated benchmark. Supplementary Figure S1 gathers the diagnostic estimates for recognized perforation, the clinical profile of the Waddingham cohort, the multivariable predictors reported by Celikyurt et al. and the management procedures in Rajkumar. Experience such as this, small and clinically selected, supports CT once perforation is suspected; whether the same performance holds in asymptomatic screening is untested.

5.3. CT Technique and Artefact

What a CT study can show about a lead depends on how it was acquired and read. Cardiac and lead motion degrade non-gated studies, and metal streak or blooming artefact can displace the apparent tip position by millimetres [13,26]. ECG-gated acquisition with multiplanar reformation addresses both: Pang et al. used a 320-row gated scanner, and were able to measure protrusion depth directly [13], and Balabanoff et al. found good interobserver agreement for perforation versus non-perforation on gated non-contrast cardiac CT (kappa 0.71), though agreement on a five-point likelihood scale was only moderate (weighted kappa 0.54) [26]. Chest radiography performed far worse in the same study, with 15% average sensitivity against CT and no observer agreement (kappa −0.1) [26]. Gating is not a complete remedy: Rajkumar et al. found a hard-beam artefact from the leads in every protocol they reviewed, judged that iterative reconstruction and ECG gating did little to reduce it, and, nevertheless, adopted an ECG-gated non-contrast acquisition with slices of at least 2.5 mm as their standard practice [24]. Reader teams differed across the cohorts: two cardiothoracic radiologists with arbitration in Hirschl et al. [5], one radiologist and two cardiologists in Lim et al. [6], two cardiologists and one radiologist in Lee et al. [7], and two independent reviewers in Pang et al. [13]. Lim et al. reconstructed images at mid-diastole, required a short-axis series, accepted traversal only when it appeared on at least two views, and excluded studies with severe metal artefact [6]. Protrusion depth gives the reading its granularity. The four apparent perforations in Pang et al. protruded 1.5 ± 0.5 mm and had no effusion or parameter change, while the two clinically perforated leads protruded 5 and 15 mm and were repositioned [13]. Bhatia et al. graded protrusion as below or at least 5 mm, and found the larger protrusions in 13 of 143 patients (9.1%) [17].
Two boundaries deserve separate mention. A tip that crosses the outer myocardial contour usually sits in epicardial fat; whether it has also crossed the parietal pericardium is a different question, and pericardial effusion on the same study is the practical marker of transpericardial perforation [1,2,13]. An aArtefact, in the other direction, can create the appearance of perforation where none exists. Five reported right-ventricular perforations in one 12-month audit were all artefactual [15], and 29 of 35 CT-called perforations were intracardiac on ventriculography [16]. The reviewed cohorts protected themselves differently: gated acquisition in Pang et al. [13], exclusion of degraded studies in Lim et al. [6], and no stated protection in Hirschl et al. [5]. Photon-counting CT with dedicated metal-artifact reconstruction improves lead visualization in phantom work [27], but none of these techniques has been validated against clinical outcomes in asymptomatic recipients.

5.4. CT Interpretation in Asymptomatic Recipients

Interpreting a protruding tip or tine in a patient without symptoms starts with what the finding is called. The cohorts reviewed here used anatomical definitions: traversal of the tip [5,6] or of the tines [7] beyond the outer myocardial contour, with no attributable symptom or sign. A clinically recognized perforation is a different entity, defined by symptoms, an effusion, hemodynamic compromise, electrical abnormalities or confirmation at revision, as in the recognized-perforation series discussed above [3,24,28]. The three core cohorts adopted the CT reading itself as the reference standard, so no sensitivity or specificity can be derived from them; accuracy estimates exist only where an independent standard was available, and those cohorts enrolled patients already suspected of perforation [24,25]. Metallic streak or blooming, cardiac motion, and a thin myocardial contour can all obscure where a lead tip or fixation tine actually sits [13,26]. For silent penetration, the literature offers neither validated fluoroscopic cutoffs nor a universal CT grading scale [3,6,24,25]. Routine CT screening would bring radiation, contrast exposure where contrast is used, incidental findings and downstream investigation, with no demonstrated net clinical benefit.

7. Management Under Uncertainty

7.1. Conventional Transvenous Leads

Clinical compromise sets urgency before anything else. Tamponade or hemodynamic deterioration needs prompt specialist management, with drainage and, where indicated, assessment of the responsible lead. Whether to revise, reposition, extract, or operate depends on the anatomical finding, clinical stability, dwell time, operator expertise and available backup [3,12,24,35]. A large effusion alone does not settle the extraction technique, and CT-confirmed protrusion does not remove the need for hemodynamic assessment.
Intervention series show that the procedures are feasible in selected patients; they leave the choice between observation and intervention for silent CT findings exactly where it was. Issa et al. reported 54 subacute or delayed perforations; 31 patients were symptomatic, and the 23 asymptomatic patients were identified from incidental imaging findings or anomalous device data; all leads were removed or repositioned percutaneously, without major periprocedural complications [28]. Elbatran et al. managed 46 recognized perforations at a single referral centre largely percutaneously, with computed tomography used selectively and cardiac surgery held in reserve [36]. Al-Maisary et al. described 32 perforations over 25 years, with simple traction successful in 23 (71.9%) and no in-hospital mortality [37]. In Rajkumar et al., 17 patients underwent transvenous lead extraction (TLE) and one a hybrid procedure, with complete procedural success reported in the TLE group [24]. None of this shows that every incidental CT-positive lesion needs removal, or that routine observation is the safer course.
With no comparative data to go on, surveillance in selected asymptomatic recipients is an inference from the absence of observed progression in small cohorts, rather than a studied strategy. Where it is chosen, careful imaging review and individual electrophysiology assessment should come first. Stable interrogation, no clinically important or increasing effusion, certainty about the CT finding, pacing dependence, new symptoms, and the risks of a procedure all bear on the decision. No comparative prospective study in the literature reviewed here fixes a follow-up interval, an electrical trigger, or a CT protrusion threshold for general use. A good outcome after a particular corrective intervention in a case report is not evidence for observation alone [38].
Anticoagulation deserves separate attention, having been associated with tamponade among clinically recognized perforation cases (OR 21.7; 95% CI 1.7–275.5; p = 0.018) [3]. A conference abstract in leadless recipients reported tine protrusion more often in patients taking non-vitamin K antagonist oral anticoagulants, without adjusted analysis [14]. Starting or continuing anticoagulation calls for a weighing of thromboembolic risk against bleeding risk, and for some confidence in the anatomical diagnosis; it does not, in itself, indicate extraction. CT may enter procedural planning as a way of judging extraction complexity, but evidence on general extraction risk is not a perforation-specific estimate of treatment success [12,39].

7.2. Leadless Pacemakers

The leadless situation brings a different structure and a different removal procedure with it. In Lee et al. the finding was penetration by fixation tines in patients free of malfunction or effusion, and no related adverse events occurred during the reported follow-up [7]. Decisions made for a symptomatic conventional lead crossing the myocardium do not transfer mechanically to this setting. Removing a functioning device for no good reason carries procedural risk; keeping a confirmed abnormality means accepting uncertainty about what comes next.
Early retrieval experience in the published work shows percutaneous removal to be feasible in selected cases [40]. It sets no universal deadline after which removal becomes impossible, and no optimal timing for an asymptomatic CT finding. Reports of serious leadless perforation and post-marketing adverse events concern complications that were clinically recognized; the Micra five-year registry describes broader device outcomes [41,42,43]. Neither kind of source stands in for follow-up of a cohort whose subclinical tine penetration was adjudicated in advance. Shared decisions should reflect those differences and the retrieval or surgical backup actually available.

8. Limitations and Research Priorities

This review is selective, and exhaustive coverage and freedom from publication or language bias are not claims it can make. The search was targeted, rather than exhaustive, and no formal appraisal of study quality was performed. The core cohorts are retrospective and differ in CT indication, device type, image-quality criteria and timing [5,6,7]. Case series of clinically recognized perforation and diagnostic studies address other target populations [3,24,25,28]. Pooling, a population-prevalence estimate, and a reliable progression rate are all out of reach, on this basis.
Standardizing what is measured comes first. Conventional lead-tip traversal, fixation-tine protrusion, device-body penetration, a clinically important effusion, and symptomatic perforation each need a definition of its own. CT protocols should state reconstruction, artifact handling, multiplanar criteria, reader blinding, adjudication and interobserver agreement. Analysis should keep both patient and lead denominators, and allow for more than one lead in the same patient.
Prospective follow-up would then tie a confirmed anatomical finding to prespecified outcomes: new symptoms, electrical instability, progressive effusion, tamponade, corrective procedures, and the burden patients report. Imaging and clinical schedules belong in the record alongside treatment changes, anticoagulation exposure, competing mortality, and losses to follow-up. Registries of overt device complications give context, but progression has to be studied in dedicated subclinical cohorts, rather than inferred from late-case reports.
Screening and management studies should, finally, be judged on net benefit. Detecting more abnormalities is not the same as doing better. Radiation, contrast exposure where applicable, incidental findings, anxiety, extra visits, and complications of corrective procedures all need weighing against whatever cardiac events are prevented. Expert consensus can harmonize definitions and nominate questions, and there its authority ends. An optimal screening schedule or intervention threshold is beyond what consensus can supply. Existing pacing and lead-management guidance provides the clinical frame in which this new evidence would be judged [8,9,10,11,12]. No reviewed guidance document sets a screening or intervention threshold for CT-defined subclinical penetration. What those documents supply is the assessment and management language in which such a threshold would have to be written.
The search was extended to Europe PMC for this revision, with screening counts reported in Table 1; a title-and-abstract strategy of this kind remains vulnerable to missing reports that do not name perforation in their title or that surface through report-level audits, rather than literature searches, as two relevant studies identified through other routes show [15,39]. The limitation is stated, rather than resolved.

9. Practical Considerations and the Limits of Current Inference

Nothing summarized above will carry a screening strategy, an intervention threshold or a fixed warning criterion. What it will carry is a structured approach to the patient whose lead tip or fixation tine has been found on CT beyond the myocardial contour [5,6,7,24]. Contemporary device guidance covers complications of lead implantation and management without setting screening or intervention thresholds for CT-detected penetration in the absence of symptoms [8,9,10,11,12]. The points below condense the criteria the reviewed cohorts actually used; they are a reporting aid, not a validated instrument.
The first step is to reconcile the finding with the implantation record, the device-interrogation history, and any earlier imaging. CT review has to address reconstruction plane and artifact handling, and to separate lead-tip traversal from displacement of the device body or of a fixation tine. The cohorts reviewed here applied different definitions of traversal, so a second reading against one agreed set of criteria is a precondition for any clinical decision [25,28].
  • Confirm the finding on at least two multiplanar reconstructions, and record whether the study was ECG-gated and lead-focused [6,13].
  • Grade image degradation from streak or blooming artefact, and treat a degraded study as uninformative, rather than negative [6,26].
  • Measure protrusion depth beyond the outer contour, using the millimetre conventions of the source literature [13,17].
  • State whether the tip or tine sits in epicardial fat or appears to cross the parietal pericardium, and record any pericardial effusion [13].
  • Compare device interrogation with baseline and repeat it, since single normal readings do not exclude penetration [5,6].
  • Review symptoms and examine for tamponade physiology, before any decision about revision [3,24].
  • Record the reasoning and the uncertainty in the notes, and make the decision jointly with the patient [28,38].
Diagnostic accuracy belongs to the population it was measured in. The sensitivity and specificity figures cited in Section 5.2 come from patients already under investigation for suspected perforation, and they do not carry across to asymptomatic screening [24]. Asymptomatic detection brings risks of its own: radiation, contrast exposure where applicable, incidental findings, and the consequences of intervening on an anatomical finding whose clinical significance is unknown [29,39]. Electrical findings give context, and not a rule, as Table 3 sets out; trends across more than one interrogation say more than any isolated measurement [5,29,35]. Device design changes the situation outright: retrieval options for transvenous leads depend on fixation type and lead age, and leadless devices encapsulate progressively, which narrows percutaneous options, with time [39,40].
The gap should be named plainly. As Section 7 sets out, existing series describe both conservative follow-up and active percutaneous management in asymptomatic patients with stable parameters and no effusion or very little. No randomized comparison of the two strategies exists [28,38]. Individualized assessment, with the reasoning written down and the uncertainty acknowledged, can be defended; protocol-driven intervention for an anatomical finding cannot.
Counselling should be framed around uncertainty. Patients deserve to hear that CT-detected penetration is commoner than clinically recognized perforation, that the long-term natural history is unknown, and that no prospective data on progression exist [6,7,43].

10. Future Perspectives

The assessments reviewed here rest on manual reading by small teams with variable agreement [13,26]. Automation is moving on two fronts. Deep-learning detectors already identify CIEDs and characterize their MRI safety status on chest radiographs [44]. Prototype tools detect leads and in-paint metal artefact on cardiac CT [45], and photon-counting systems with dedicated metal-artifact reconstruction improve lead visualization at the source [27]. Applied to this problem, the same methods could triage scans for lead position, measure protrusion depth reproducibly, and standardize the reading that currently varies between observers. None of it has been validated for asymptomatic perforation detection. Validation would require annotated cohorts, an agreed reference standard and linkage to clinical outcomes. Until then, automated readings belong in the same category as the reporting checklist in Section 9: promising instruments, not evidence.

11. Conclusions

In selected CT cohorts, conventional lead tips and leadless fixation tines extend beyond the myocardial contour, without symptoms or device malfunction. The proportions reported there are anatomical findings within selected imaging populations, not comparable estimates of population incidence. Follow-up was short. The Lim cohort showed no progression to clinical perforation and the CT-positive Lee cohort no related adverse events, which leaves long-term risk open. What the current evidence supports is careful confirmation of the finding and individual assessment; it does not support routine CT screening, universal extraction, or fixed electrical warning thresholds. The reporting checklist and assessment criteria assembled here organize what is known, and the imaging and automation work now under way may eventually test it. Prospective studies should establish which findings matter to patients, and whether detecting or treating them improves outcomes. Given how selective the evidence base is, these conclusions delimit what current data can bear, rather than summarize established practice.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics16233963/s1, and is reproduced here for review. Supplementary Figure S1 gathers the diagnostic, clinical, risk and management profile of clinically recognized perforation, a population distinct from the asymptomatic CT findings that form the core of this review.

Author Contributions

Conceptualization, Y.J. and H.D.; methodology, Y.J. and C.G.; formal analysis, Y.J., P.L., Y.L., and H.D.; investigation, P.L. and C.G.; data curation, Y.J.; visualization, Y.J. and C.G.; supervision, H.D.; writing—original draft preparation, Y.J.; writing—review and editing, all authors. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable. This narrative review analysed only previously published data, and did not involve any new studies with human or animal subjects.

Data Availability Statement

All data discussed in this review are contained within the cited published articles. No new data were generated.

Acknowledgments

OpenAI Codex was used for assistance with language and structure, checking the correspondence between cited evidence and manuscript claims, and preparing code for figures drawn from published numerical results. The revised manuscript was further restructured in response to peer review with AI assistance (WorkBuddy v5.11.4, GLM-5.4), including literature verification searches, recalculated confidence intervals and regenerated figures. This assistance did not generate new patient data or establish a clinical recommendation. The authors are responsible for verifying the underlying sources, approving the interpretations, and reviewing the final text and figures. The figures summarize published results and were generated with Matplotlib in Python version 3.14.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

BMIBody mass index
CIConfidence interval
CIEDCardiac implantable electronic device
CTComputed tomography
ICDImplantable cardioverter-defibrillator
IQRInterquartile range
LVEFLeft-ventricular ejection fraction
OROdds ratio
RVRight ventricle
TLETransvenous lead extraction
TTETransthoracic echocardiography

References

  1. Bolaji, O.A.; Markson, F.E.; Kwaku, K.F.; Schaller, R.D. Lead Perforation in Cardiac Implantable Electronic Devices: Incidence, Diagnosis, and Management. Pacing Clin. Electrophysiol. 2025, 48, 874–885. [Google Scholar] [CrossRef] [Scilit]
  2. Awashra, A.; AbuBaha, M.; Salameh, H.; Jallad, H.; Milhem, A.; Zahran, A.; Badwan, A.; Milhem, F.; Shubietah, A. Cardiac lead perforation: Mechanisms, detection, and therapeutic approaches. Heart Rhythm O2 2026, 7, 156–170. [Google Scholar] [CrossRef] [Scilit]
  3. Waddingham, P.H.; Elliott, J.; Bates, A.; Bilham, J.; Muthumala, A.; Honarbakhsh, S.; Ullah, W.; Hunter, R.J.; Lambiase, P.D.; Lane, R.E.; et al. Iatrogenic cardiac perforation due to pacemaker and defibrillator leads: A contemporary multicentre experience. Europace 2022, 24, 1824–1833. [Google Scholar] [CrossRef] [Scilit]
  4. Vamos, M.; Erath, J.W.; Benz, A.P.; Bari, Z.; Duray, G.Z.; Hohnloser, S.H. Incidence of Cardiac Perforation with Conventional and With Leadless Pacemaker Systems: A Systematic Review and Meta-Analysis. J. Cardiovasc. Electrophysiol. 2017, 28, 336–346. [Google Scholar] [CrossRef] [Scilit]
  5. Hirschl, D.A.; Jain, V.R.; Spindola-Franco, H.; Gross, J.N.; Haramati, L.B. Prevalence and Characterization of Asymptomatic Pacemaker and ICD Lead Perforation on CT. Pacing Clin. Electrophysiol. 2007, 30, 28–32. [Google Scholar] [CrossRef] [Scilit]
  6. Lim, Y.M.; Uhm, J.S.; Kim, M.; Kim, I.S.; Jin, M.N.; Yu, H.T.; Kim, T.H.; Lee, H.J.; Kim, Y.J.; Joung, B.; et al. Subclinical cardiac perforation by cardiac implantable electronic device leads detected by cardiac computed tomography. BMC Cardiovasc. Disord. 2021, 21, 346. [Google Scholar] [CrossRef] [Scilit]
  7. Lee, Y.S.; Uhm, J.S.; Park, J.W.; Park, S.J.; Kim, Y.C.; Kim, I.S.; Lee, H.J.; Kim, M.H.; Park, H.; Kim, D.; et al. Impact of Leadless Pacemaker Implantation Position on Subclinical Right Ventricular Perforation. Pacing Clin. Electrophysiol. 2026, 49, 1447–1454. [Google Scholar] [CrossRef] [Scilit]
  8. Glikson, M.; Nielsen, J.C.; Kronborg, M.B.; Michowitz, Y.; Auricchio, A.; Barbash, I.M.; Barrabés, J.A.; Boriani, G.; Braunschweig, F.; Brignole, M.; et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur. Heart J. 2021, 42, 3427–3520. [Google Scholar] [CrossRef] [Scilit]
  9. Kusumoto, F.M.; Schoenfeld, M.H.; Barrett, C.; Edgerton, J.R.; Ellenbogen, K.A.; Gold, M.R.; Goldschlager, N.F.; Hamilton, R.M.; Joglar, J.A.; Kim, R.J.; et al. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation 2019, 140, e382–e482. [Google Scholar] [CrossRef] [Scilit]
  10. Bongiorni, M.G.; Burri, H.; Deharo, J.C.; Starck, C.; Kennergren, C.; Saghy, L.; Rao, A.; Tascini, C.; Lever, N.; Kutarski, A.; et al. 2018 EHRA expert consensus statement on lead extraction: Recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: Endorsed by APHRS/HRS/LAHRS. Europace 2018, 20, 1217–1233. [Google Scholar] [CrossRef] [Scilit]
  11. Zeppenfeld, K.; Tfelt-Hansen, J.; de Riva, M.; Winkel, B.G.; Behr, E.R.; Blom, N.A.; Charron, P.; Corrado, D.; Dagres, N.; de Chillou, C.; et al. 2022 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Eur. Heart J. 2022, 43, 3997–4126. [Google Scholar] [CrossRef] [Scilit]
  12. Kusumoto, F.M.; Schoenfeld, M.H.; Wilkoff, B.L.; Berul, C.I.; Birgersdotter-Green, U.M.; Carrillo, R.; Cha, Y.M.; Clancy, J.; Deharo, J.C.; Ellenbogen, K.A.; et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm 2017, 14, e503–e551. [Google Scholar] [CrossRef] [Scilit]
  13. Pang, B.J.; Lui, E.H.; Joshi, S.B.; Tacey, M.A.; Alison, J.; Seneviratne, S.K.; Cameron, J.D.; Mond, H.G. Pacing and Implantable Cardioverter Defibrillator Lead Perforation as Assessed by Multiplanar Reformatted ECG-Gated Cardiac Computed Tomography and Clinical Correlates. Pacing Clin. Electrophysiol. 2014, 37, 537–545. [Google Scholar] [CrossRef] [Scilit]
  14. Lee, Y.; Uhm, J.S.; Kim, M.H.; Park, J.W.; Kim, D.; Yu, H.T.; Kim, T.H.; J, B.; Pak, H.N.; Lee, M.H. Incidence and prognosis of subclinical cardiac perforation after leadless pacemaker implantation. Europace 2024, 26, euae102.520. [Google Scholar] [CrossRef] [Scilit]
  15. Morgan, H.; Williams, C.; Bleasdale, R.A. Computed tomography artefact finding of pacing lead perforation. Br. J. Cardiol. 2020, 27, 97–99. [Google Scholar] [CrossRef] [Scilit]
  16. Asada, S.; Nishii, N.; Shinya, T.; Miyoshi, A.; Morimoto, Y.; Miyamoto, M.; Nakagawa, K.; Nakamura, K.; Morita, H.; Ito, H. Usefulness of right ventriculography compared with computed tomography for ruling out the possibility of lead perforation before lead extraction. PLoS ONE 2021, 16, e0245502. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Bhatia, P.; Chiou, T.; Svennberg, E.; Khoche, S.; Jacobs, K.; Pollema, T.; Pretorius, V.; Birgersdotter-Green, U. Clinical significance of incidentally detected lead perforations by computed tomography. Pacing Clin. Electrophysiol. 2021, 44, 936–942. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Liu, Y.; Liu, H.; Zhu, B.; Li, H.; Cui, N.; Ye, F.; Gao, J.; Zhang, X.; Wu, Y. Study on the Identification and Treatment Strategy of Cardiac Perforation Caused by Pacing Lead: A Systematic Review and Meta-Analysis. Cardiology 2025. ahead of print. [Google Scholar] [CrossRef] [Scilit]
  19. Hamada, Y.; Sakaki, M.; Watanabe, Y.; Hata, S.; Kimura, K.; Sakagoshi, N. Delayed Right Ventricular Pacemaker Lead Perforation 9 Years After Implantation. Am. J. Case Rep. 2023, 24, e940291. [Google Scholar] [CrossRef] [Scilit]
  20. Yamamoto, A.; Takahashi, S. Delayed right ventricular lead perforation by a pacemaker lead 2-year post-implantation. Clin. Case Rep. 2022, 10, e05760. [Google Scholar] [CrossRef] [Scilit]
  21. Nakao, T.; Fujiu, K.; Uehara, M.; Komuro, I. Asymptomatic Long-term Cardiac Perforation by a Pacemaker Lead. Intern. Med. 2019, 58, 2409–2410. [Google Scholar] [CrossRef] [Scilit]
  22. Gupta, A.; Agstam, S.; Agarwal, T.; Verma, S. An Interesting Case of Acute Asymptomatic Lead Perforation of a Permanent Cardiac Pacemaker. Cureus 2021, 13, e13334. [Google Scholar] [CrossRef] [Scilit]
  23. Ko, K.Y.; Kim, S.E.; Hwang, J.W.; Namgung, J.; Cho, S.W. Delayed Cardiac Perforation Caused by Pacemaker Lead Dislodgement During Chiropractic Therapy. JACC Case Rep. 2025, 30, 104192. [Google Scholar] [CrossRef] [Scilit]
  24. Rajkumar, C.A.; Claridge, S.; Jackson, T.; Behar, J.; Johnson, J.; Sohal, M.; Amraoui, S.; Nair, A.; Preston, R.; Gill, J.; et al. Diagnosis and management of iatrogenic cardiac perforation caused by pacemaker and defibrillator leads. Europace 2017, 19, 1031–1037. [Google Scholar] [CrossRef] [Scilit]
  25. Zhang, X.; Zheng, C.; Wang, P.; Wang, D.; Huang, B.; Li, G.; Hu, H.; Yang, Z.; Duan, X.; Zheng, S.; et al. Assessment of Cardiac Lead Perforation: Comparison Among Chest Radiography, Transthoracic Echocardiography and Electrocardiography-gated Contrast-enhanced Cardiac CT. Eur. Radiol. 2019, 29, 963–974. [Google Scholar] [CrossRef] [Scilit]
  26. Balabanoff, C.; Gaffney, C.E.; Ghersin, E.; Okamoto, Y.; Carrillo, R.; Fishman, J.E. Radiographic and electrocardiography-gated noncontrast cardiac CT assessment of lead perforation: Modality comparison and interobserver agreement. J. Cardiovasc. Comput. Tomogr. 2014, 8, 384–390. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Schalla, S.; Wildberger, J.E.; Ramaekers, M.J.F.G.; Mihl, C.; McDermott, M.C.; Stammen, L.; Vernooy, K.; Flohr, T.G. Imaging of cardiovascular implantable electronic device leads with photon-counting detector computed tomography. Investig. Radiol. 2025, 60, 616–625. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Issa, Z.F.; Issa, T.Z. Feasibility and Safety of Percutaneous Lead Revision for Subacute and Delayed Cardiac Device Lead Perforation. JACC Clin. Electrophysiol. 2021, 7, 26–35. [Google Scholar] [CrossRef] [Scilit]
  29. Celikyurt, U.; Acar, B.; Yavuz, S.; Agacdiken, A.; Vural, A. Predictors of the right ventricular perforation caused by active-fixation pacing and defibrillator leads: A single-centre experience. J. Cardiovasc. Electrophysiol. 2024, 35, 399–405. [Google Scholar] [CrossRef] [Scilit]
  30. Sterlinski, M.; Przybylski, A.; Maciag, A.; Syska, P.; Pytkowski, M.; Lewandowski, M.; Kowalik, I.; Firek, B.; Kolsut, P.; Religa, G.; et al. Subacute cardiac perforations associated with active fixation leads. Europace 2009, 11, 206–212. [Google Scholar] [CrossRef] [Scilit]
  31. Cano, Ó.; Andrés, A.; Alonso, P.; Osca, J.; Sancho-Tello, M.J.; Olagüe, J.; Martínez-Dolz, L. Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: A single-centre experience with over 3800 implanted leads. Europace 2017, 19, 96–102. [Google Scholar] [CrossRef] [Scilit]
  32. Moore, K.; Ganesan, A.; Labrosciano, C.; Heddle, W.; McGavigan, A.; Hossain, S.; Horton, D.; Hariharaputhiran, S.; Ranasinghe, I. Sex Differences in Acute Complications of Cardiac Implantable Electronic Devices: Implications for Patient Safety. J. Am. Heart Assoc. 2019, 8, e010869. [Google Scholar] [CrossRef] [Scilit]
  33. Güzel, T.; Demir, M.; Aktan, A.; Kılıç, R.; Arslan, B.; Günlü, S.; Altıntaş, B.; Karahan, M.Z.; Özbek, M.; Aslan, B.; et al. The effect of body mass index on complications in cardiac implantable electronic device surgery. Pacing Clin. Electrophysiol. 2024, 47, 292–299. [Google Scholar] [CrossRef] [Scilit]
  34. Piccini, J.P.; Cunnane, R.; Steffel, J.; El-Chami, M.F.; Reynolds, D.; Roberts, P.R.; Soejima, K.; Steinwender, C.; Garweg, C.; Chinitz, L.; et al. Development and validation of a risk score for predicting pericardial effusion in patients undergoing leadless pacemaker implantation: Experience with the Micra transcatheter pacemaker. Europace 2022, 24, 1119–1126. [Google Scholar] [CrossRef] [Scilit]
  35. Younis, A.; Awashra, A.; Matteo, M.; Hussein, A.A.; Demian, J.; Santangeli, P.; Callahan, T.; Martin, D.O.; Mdaihly, M.; Nakhla, S.; et al. Management of Transvenous Leads in Patients with Iatrogenic Lead Perforation. J. Cardiovasc. Electrophysiol. 2025, 36, 2449–2451. [Google Scholar] [CrossRef] [Scilit]
  36. Elbatran, A.I.; Akhtar, Z.; Bajpai, A.; Leung, L.W.M.; Li, A.; Pearse, S.; Zuberi, Z.; Kaba, R.; Saba, M.M.; Norman, M.; et al. Percutaneous management of lead-related cardiac perforation with limited use of computed tomography and cardiac surgery. Pacing Clin. Electrophysiol. 2021, 44, 614–624. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Al-Maisary, S.; Karck, M.; Guzman-Ruvalcaba, M.J.; Arif, R.; Romano, G. Cardiac implantable electronic device lead perforation: A 25-year single-center experience. J. Clin. Med. 2026, 15, 2705. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Caiati, C.; Pollice, P.; Truncellito, L.; Lepera, M.E.; Favale, S. Minimal Cardiac Perforation by Lead Pacemaker Complicated with Pericardial Effusion and Impending Tamponade: Optimal Management with No Pericardiocentesis Driven by Echocardiography. Diagnostics 2020, 10, 191. [Google Scholar] [CrossRef] [Scilit]
  39. Svennberg, E.; Jacobs, K.; McVeigh, E.; Pretorius, V.; Birgersdotter-Green, U. Computed Tomography–Guided Risk Assessment in Percutaneous Lead Extraction. JACC Clin. Electrophysiol. 2019, 5, 1439–1446. [Google Scholar] [CrossRef] [Scilit]
  40. Afzal, M.R.; Daoud, E.G.; Cunnane, R.; Mulpuru, S.K.; Koay, A.; Hussain, A.; Omar, R.; Wei, K.K.; Amin, A.; Kidwell, G.; et al. Techniques for successful early retrieval of the Micra transcatheter pacing system: A worldwide experience. Heart Rhythm 2018, 15, 841–846. [Google Scholar] [CrossRef] [Scilit]
  41. Hauser, R.G.; Gornick, C.C.; Abdelhadi, R.H.; Tang, C.Y.; Kapphahn-Bergs, M.; Casey, S.A.; Okeson, B.K.; Steele, E.A.; Sengupta, J.D. Leadless pacemaker perforations: Clinical consequences and related device and user problems. J. Cardiovasc. Electrophysiol. 2022, 33, 154–159. [Google Scholar] [CrossRef] [Scilit]
  42. Hauser, R.G.; Gornick, C.C.; Abdelhadi, R.H.; Tang, C.Y.; Casey, S.A.; Sengupta, J.D. Major adverse clinical events associated with implantation of a leadless intracardiac pacemaker. Heart Rhythm 2021, 18, 1132–1139. [Google Scholar] [CrossRef] [Scilit]
  43. El-Chami, M.F.; Garweg, C.; Clementy, N.; Al-Samadi, F.; Iacopino, S.; Martinez-Sande, J.L.; Roberts, P.R.; Tondo, C.; Johansen, J.B.; Vinolas-Prat, X.; et al. Leadless pacemakers at 5-year follow-up: The Micra transcatheter pacing system post-approval registry. Eur. Heart J. 2024, 45, 1241–1251. [Google Scholar] [CrossRef] [Scilit]
  44. Kim, U.H.; Kim, M.Y.; Park, E.A.; Lee, W.; Lim, W.H.; Kim, H.L.; Oh, S.; Jin, K.N. Deep learning-based algorithm for the detection and characterization of MRI safety of cardiac implantable electronic devices on chest radiographs. Korean J. Radiol. 2021, 22, 1918–1928. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. McKeown, T.; Gach, H.M.; Hao, Y.; An, H.; Robinson, C.G.; Cuculich, P.S.; Yang, D. ICD lead and primary metal artifact detection and inpainting in cardiac CT images. Med. Phys. 2025, 52, e17947. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.