Abstract
Background/Objectives: Panoramic radiography is an important imaging modality for the evaluation of dentoalveolar and maxillofacial structures in dental diagnosis and treatment planning. The diagnostic quality of panoramic images is closely related to appropriate patient preparation and accurate positioning. Positioning errors may adversely affect image quality, leading to misinterpretation and unnecessary repeat examinations. This study aimed to evaluate the diagnostic quality of panoramic radiographs, determine the frequency of common positioning errors, and identify error types contributing to diagnostically inadequate images. Methods: In this retrospective cross-sectional study, a total of 5155 panoramic radiographs obtained at Mardin Oral and Dental Health Center were evaluated. Patients were divided into four age groups: ≤20, 21–40, 41–60, and ≥61 years. The assessed errors included failure to position the tongue against the palate, slumped patient position, positioning the patient too far forward or backward, failure to bite on the bite block, patient movement, forward or backward head tilt, head rotation to one side, foreign bodies, exposure errors, lip positioning errors, and missing image sections. Radiographic quality was classified as “excellent,” “diagnostically acceptable,” or “unacceptable.” The distribution of error types according to age and sex was evaluated using the chi-square test, with p < 0.05 considered statistically significant. Results: At least one error was detected in 40.1% of the panoramic radiographs. The most frequent errors were failure to position the tongue against the palate (10.3%) and head rotation to one side (9.7%). Of the radiographs, 3.6% were classified as excellent, 95.7% as diagnostically acceptable, and 0.6% as unacceptable. The overall error rate was highest in the ≥61-year age group (53.8%) and lowest in the ≤20 year age group (31%), with a significant increase with age (p < 0.05). No significant association was found between sex and overall error frequency (p > 0.05), although significant sex-related differences were observed for some specific error types. Conclusions: Patient preparation and positioning errors may affect the diagnostic quality of panoramic radiographs. Accurate positioning, effective patient communication, and consideration of age-specific error patterns may contribute to improving image quality.
1. Introduction
In dentistry, the panoramic radiograph is a highly valuable tool for diagnosing various dental and oral conditions and is essential for designing comprehensive treatment plans [1]. Panoramic radiography holds significant clinical importance across several dental specialties, as it allows professionals to comprehensively visualize the dental arches and hard tissues of the maxillofacial region in a single image [2]. It is utilized for the evaluation of jaw trauma, localization and angulation assessment of third molars, diagnosis of developmental anomalies or underlying pathologies, analysis of root angulations during and after orthodontic treatment, assessment of growth and developmental changes in the dentition of children and adolescents, evaluation of the temporomandibular joint (TMJ), and the general assessment of a patient’s dentition [3]. Currently, panoramic radiography is not only widely accessible but also crucial for diagnosing alterations in oral conditions. It provides essential evidence that, when combined with clinical examinations, significantly improves the diagnostic process [4].
Radiographs that are sharp, highly detailed, exhibit minimal distortion, demonstrate correct film framing, lack artifacts, and possess good density and adequate contrast are considered to be of high diagnostic quality. Radiographs failing to meet these criteria indicate that an error occurred during radiographic imaging and/or processing [5]. Research has shown that a considerable number of radiographs exposed in dentistry are of marginal or non-diagnostic quality [6]. Low-quality radiographs can lead to misinterpretation, resulting in incorrect diagnoses and flawed treatment planning [3]. Furthermore, the repetition of panoramic radiography carries an associated risk of inducing cancer, estimated at 0.21 to 1.9 cases per million examinations [7]. Therefore, it is imperative that the imaged structures in panoramic radiographs are of high quality and exhibit minimal distortion [8].
The quality of any radiograph relies heavily on correct patient preparation and positioning, followed by the selection of suitable exposure parameters [9]. To obtain diagnostically useful panoramic radiographs, patients must be properly prepared and their heads carefully positioned within the image layer. The image layer, or “focal trough,” is a three-dimensional curved zone wherein the structures lying within are reasonably well-defined on the final panoramic image [10]. Consequently, correct patient positioning is vital to ensure that all regions of interest appear clearly in the final image [2].
The transition to digital imaging has largely eliminated processing errors, allowing technicians and dentists to focus primarily on mitigating patient and operator errors to obtain more accurate, diagnostically acceptable images [1,6]. Processing errors can be eliminated by the use of digital processing, which has seen increasing in popularity among dentists. However, the appropriate positioning and preparation of the patient is essential for a sharp, accurate, and undistorted image, which is not affected by ghost images [6]. Several studies have emphasized that the difficulty of achieving correct positioning in panoramic radiography should not be underestimated [2]. Literature consistently highlights that positioning errors in panoramic radiographs are common and frequently result in diagnostic difficulties [11].
The aim of this retrospective study was to evaluate the diagnostic quality of panoramic radiographs, determine the prevalence of positioning and technical errors, and identify the error types associated with diagnostically inadequate images using radiographic records retrieved from the archive of the department of oral and maxillofacial radiology.
2. Materials and Methods
2.1. Sample Size and Eligibility Criteria
This study was approved by the Health Sciences Scientific Research Ethics Committee of Sakarya University Faculty of Medicine (Application No. 649; Approval Date: 19 August 2026) and was conducted in accordance with the Declaration of Helsinki. Informed consent was obtained from all subjects involved in the study.
A retrospective cross-sectional study was conducted using panoramic radiographs acquired during routine clinical examinations at Mardin Oral and Dental Health Center between 16 January and 10 March 2026 and retrieved directly from the hospital’s Picture Archiving and Communication System (PACS). When multiple panoramic radiographs were available for the same patient during the study period, only the earliest acquired radiograph was retained based on the acquisition date and time recorded in the PACS. After the removal of repeated radiographs from the same patients, a total of 5167 panoramic radiographs were screened for eligibility. Radiographs with available demographic information, including patient age and sex, were eligible for inclusion. Twelve radiographs were excluded due to missing information regarding patient age and/or sex. No additional exclusion criteria based on positioning errors, technical errors, or diagnostic image quality were applied. Consequently, 5155 panoramic radiographs were included in the final analysis. Patients were categorized into four age groups: ≤20, 21–40, 41–60, and ≥61 years. The reporting of this retrospective cross-sectional study was guided by the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) checklist for cross-sectional studies [12]. Relevant recommendations from the REporting of studies Conducted using Observational Routinely-collected health Data (RECORD) statement [13] were also considered, as the study used routinely collected clinical radiographic data.
Panoramic radiographs were obtained using Castellini X Radius Trio 2D Device (CEFA S.C., Imola, Italy) with settings of 90 kVp, 10 mA. All panoramic radiographs were acquired by two radiology technicians, each with at least five years of experience in dental radiography. Image acquisition was performed according to the standardized panoramic radiography protocol of the center using the same panoramic unit. The procedures were not performed under direct supervision. The radiographs were evaluated under identical conditions without any digital enhancements on a 24 inch, 1920 × 1080 resolution, 8 bit color mode computer monitor (Dell INC, Round Rock, TX, USA), in a room with subdued lighting. To minimize observer fatigue and maintain consistency, approximately 100 panoramic radiographs were evaluated per day. Prior to the main radiographic evaluation, a calibration procedure was performed using 100 randomly selected panoramic radiographs obtained from a separate image pool outside the 5167 radiographs initially screened for the main study. These calibration images were not included in the final study sample or statistical analyses. An experienced dentomaxillofacial radiologist reviewed these images to standardize the application of the predefined criteria for identifying positioning and technical errors. The operational definitions and characteristic radiographic findings used to classify the 13 error categories are described in Section 2.2. Although the calibration procedure was intended to promote consistency in the application of these criteria, it did not constitute an independent validation of the classification system. The repeatability of the radiographic assessments was subsequently examined through an intra-observer reliability analysis, as described in Section 2.3. All images were analyzed by an experienced dentomaxillofacial radiologist, and the findings were recorded.
The assessed errors included: tongue not in contact with the palate, slumped patient position, patients not biting on the bite block, patient movement, head tilted forward, head tilted backward, patient positioned too far forward, patient positioned too far backward, patient’s head turned to one side, presence of foreign bodies in the image, exposure errors, lip positioning errors, and missing image sections (Figure 1, Figure 2, Figure 3 and Figure 4).
Figure 1.
Common patient positioning errors and their radiographic effects in panoramic radiography. (A) The tongue not in contact with the palate: a radiolucent area is superimposed over the apices of the maxillary teeth due to the palatoglossal airspace created by failure of the tongue dorsum to contact the palate. (B) The patient tilts the head forward: a V-shaped smile line, an inverted curvature of the occlusal plane, distortion of the anterior teeth, and apparent narrowing of the anterior structures may be observed. (C) Slumped position: superimposition of the radiopaque shadow of the cervical vertebrae over the mandibular symphysis, resulting in obscuration and reduced visualization of the anterior mandibular region. (D) The patient tilts the head backward: flattening of the occlusal plane, distortion of the mandible, and superimposition of the radiopaque image of the hard palate over the roots of the maxillary teeth are observed.
Figure 2.
Radiographic effects of common patient positioning and movement errors in panoramic radiography. (A) Patient positioned forward: the central incisors are positioned anterior to the bite groove, causing them to appear narrow, elongated, and blurred. The cervical spine is positioned within the focal trough, resulting in its superimposition over the mandible. (B) Patient positioned backward: the anterior teeth are positioned posterior to the bite groove, causing them to appear wider than normal. (C) Patient movement: movement of the patient during image acquisition results in blurring of all or part of the anatomical structures. (D) The head is turned to one side: the structures on the side toward which the head is turned appear reduced in size, whereas those on the opposite side appear enlarged.
Figure 3.
Radiographic effects of common imaging and positioning errors in panoramic radiography. (A) Missing image: anatomical structures such as the condyle, mandibular corpus, or maxillary sinus are not included within the radiographic image. (B) Positioning error of the lips: failure to close the lips during image acquisition results in a radiolucent area over the coronal portions of the anterior teeth. (C) Foreign bodies: jewellery, eyeglasses, and oral appliances produce radiopaque images on panoramic radiographs and may result in contralaterally magnified ghost images. (D) Exposure errors: inadequate exposure, including underexposure or overexposure, results in reduced image quality and may impair the visualization of anatomical structures.
Figure 4.
Radiographic effect of improper positioning on the bite block in panoramic radiography. Patients not biting on the bite block: improper positioning of the teeth results in superimposition and overlapping of the maxillary and mandibular teeth.
2.2. Radiographic Error and Image Quality Assessment
Assessment of positioning and patient preparation errors: Each of the 13 predefined radiographic error types was assessed separately based on observable radiographic characteristics and recorded as either “present” or “absent.” The classification of positioning and technical errors was based on previously described radiographic findings in the literature rather than a newly developed classification system. The operational definitions of the 13 error categories were standardized for the purposes of the present study, and each error was assessed according to its characteristic radiographic appearance [8,14,15]. These errors were identified according to radiographic findings associated with patient positioning, patient preparation, or technical factors during image acquisition, without making definitive assumptions regarding their underlying causes. Although certain positioning and technical errors may be suggested by characteristic radiographic findings, these findings do not necessarily provide definitive evidence of the underlying cause. Therefore, the error categories used in this study represent radiographic interpretations rather than direct observations of patient positioning, operator actions, or image acquisition procedures. When multiple errors were identified in the same radiograph, each error was recorded separately. The presence of an individual radiographic error did not necessarily indicate that the radiograph was diagnostically inadequate. The radiographic characteristics used to identify the 13 error types were defined as follows [8,14,15].
- Tongue not in contact with the palate: Presence of a radiolucent palatoglossal air space superimposed over the maxillary teeth and their periapical regions.
- Head rotated to one side: Asymmetric horizontal magnification of the right and left anatomical structures, particularly the mandibular rami and posterior teeth.
- Foreign bodies: Presence of radiopaque artifacts or superimposed images associated with metallic objects, jewelry, or other external materials.
- Patient movement: Blurring, distortion, or discontinuity of anatomical structures consistent with movement during image acquisition.
- Patient positioned too far forward: Anterior teeth appearing horizontally narrowed and blurred due to their position relative to the focal trough.
- Lip positioning error: Presence of a radiolucent band over the anterior teeth, particularly the crowns, consistent with an open-lip position during image acquisition.
- Head tilted backward: Flattening or reversal of the occlusal plane, potentially accompanied by superimposition of the hard palate over the maxillary teeth.
- Patient positioned too far backward: Horizontal enlargement and blurring of the anterior teeth due to their position relative to the focal trough.
- Head tilted forward: An excessively pronounced curvature of the occlusal plane, often accompanied by altered visualization of the mandibular condyles.
- Slumped posture: Presence of a radiopaque cervical spine shadow superimposed over the anterior mandibular region.
- Failure to bite on the bite block: Abnormal separation or positioning of the anterior teeth, with radiographic findings consistent with failure to maintain the intended bite-block relationship.
- Exposure error: Excessive or insufficient radiographic exposure resulting in abnormal image brightness, density, or contrast and reduced visualization of anatomical structures.
- Missing image: Incomplete visualization or truncation of anatomical structures resulting in the absence of part of the expected panoramic image.
Assessment of overall radiographic quality: Overall radiographic quality was evaluated independently of the recording of individual errors, using the three-category quality assessment approach. Unlike the assessment of individual positioning and technical errors, overall radiographic quality reflects the combined influence of several factors, including image contrast, sharpness, visibility of relevant anatomical structures, and image acquisition conditions. These factors may affect diagnostic image quality even in the absence of the predefined positioning or technical errors. This assessment focused on the extent to which the radiograph allowed adequate visualization of relevant anatomical structures and supported diagnostic interpretation.
Radiographs exhibiting optimal overall image quality, without positioning or technical errors or other image-quality limitations, were classified as “excellent.” Radiographs containing errors or minor image-quality limitations that did not substantially compromise diagnostic utility were classified as “diagnostically acceptable,” whereas radiographs in which errors or image-quality limitations substantially impaired the visualization of relevant anatomical structures or diagnostic interpretation were classified as “unacceptable.” Therefore, the assessment of positioning and technical errors and the evaluation of overall radiographic quality represent two distinct but related aspects of radiographic assessment. The absence of predefined positioning or technical errors does not necessarily indicate optimal overall image quality, just as the presence of such errors does not necessarily imply diagnostic inadequacy.
2.3. Statistical Analysis
The observer evaluated all images once, and a randomly selected sub-sample comprising 19.5% of the images (1000 of 5155 images) was re-evaluated after a 2-weeks interval to assess intra-examiner reliability. Intra-observer agreement was assessed using the Cohen’s kappa coefficient for binary error variables and linearly weighted Cohen’s kappa coefficient for the ordinal radiographic quality classification, with 95% confidence intervals reported for the weighted kappa estimates. The overall kappa statistic for intra-observer agreement in the assessment of positioning errors ranged from 0.812 to 0.888, demonstrating high repeatability of the observer’s examinations (Supplementary Table S1). For the three-category radiographic quality assessment, the linearly weighted Cohen’s kappa was 0.849 (95% CI: 0.772–0.927), likewise indicating a high intra-observer agreement. The collected data were recorded digitally and analyzed using Microsoft Excel 2021 (Microsoft Corporation, Redmond, WA, USA) and IBM SPSS Statistics (version 20.0; IBM Corp., Armonk, NY, USA). The associations between panoramic radiographic error types and age groups or gender were evaluated using the chi-square test of independence. Multivariable binary logistic regression analyses were performed to evaluate the independent associations of age group and sex with each panoramic radiographic error. Each error type was included as a separate binary dependent variable (presence or absence), while age group (≤20, 21–40, 41–60, and ≥61 years) and sex were entered simultaneously as independent variables. Patients aged ≤20 years and female patients were used as the reference categories. The results were expressed as adjusted odds ratios (ORs) with 95% confidence intervals (CIs). The overall significance of age group was assessed using the Wald test. A p-value < 0.05 was considered statistically significant.
3. Results
Of the 5167 panoramic radiographs initially screened, 12 (0.2%) were excluded due to missing demographic information regarding patient age and/or sex. Consequently, 5155 radiographs (99.8%) were included in the final analysis. A reference sample-size calculation was performed based on the prevalence of panoramic radiographic positioning errors reported in a previous study. Using an expected error prevalence of 64% [16], a 95% confidence level, and a 1.5% margin of error, the minimum required sample size was calculated as 3934 panoramic radiographs. The final sample of 5155 radiographs exceeded this reference requirement. The mean age of the patients was 29.23 ± 16.49 years, with an age range of 2–90 years. The study population comprised 2745 females (53.2%; mean age, 29.26 ± 15.73 years) and 2410 males (46.8%; mean age, 29.20 ± 17.32 years).
3.1. Prevalence of Panoramic Radiographic Errors
Overall, 2069 (40.1%) of the 5155 panoramic radiographs exhibited at least one positioning or technical error, whereas 3086 (59.9%) showed no errors. The prevalence of at least one radiographic error differed significantly across age groups (p < 0.001). The proportions of radiographs with at least one error were 31.0% (524/1688) in patients aged ≤20 years, 44.1% (935/2120) in those aged 21–40 years, 43.4% (476/1098) in those aged 41–60 years, and 53.8% (134/249) in those aged ≥61 years. Regarding sex, at least one error was identified in 39.8% (1093/2745) of radiographs from female patients and 40.5% (976/2410) of those from male patients, with no statistically significant difference (p = 0.619).
Multivariable binary logistic regression analysis demonstrated a significant association between age group and the presence of at least one radiographic error after adjustment for sex (overall p < 0.001). Compared with patients aged ≤20 years, the adjusted odds of having at least one radiographic error were higher among patients aged 21–40 years (OR = 1.760, 95% CI: 1.539–2.013; p < 0.001), 41–60 years (OR = 1.702, 95% CI: 1.454–1.993; p < 0.001), and ≥61 years (OR = 2.584, 95% CI: 1.973–3.384; p < 0.001). Sex was not significantly associated with overall radiographic error presence after adjustment for age group (OR = 1.051, 95% CI: 0.939–1.177; p = 0.384) (Table 1).
Table 1.
Prevalence of at least one panoramic radiographic error according to age group and sex, and multivariable logistic regression results.
The most frequently observed error was failure to position the tongue against the palate, which was detected in 533 radiographs (10.3%), followed by the patient’s head being rotated to one side, observed in 500 radiographs (9.7%). Other relatively frequent errors included foreign bodies (224, 4.3%), patient movement (180, 3.5%), and positioning the patient too far forward (173, 3.4%). Less frequent errors included lip positioning errors (132, 2.6%), backward head tilt (124, 2.4%), positioning the patient too far backward (110, 2.1%), forward head tilt (89, 1.7%), and slumped posture (58, 1.1%). The least frequent errors were failure to bite on the bite block (31, 0.6%), exposure errors (23, 0.4%), and missing image sections (22, 0.4%). The frequency distribution of the evaluated panoramic radiographic errors is presented in Table 2.
Table 2.
Prevalence of panoramic radiographic errors, ranked by frequency of occurrence.
When the prevalence of panoramic radiographic errors was evaluated according to age group, several differences in the distribution of error types were observed (Table 3). Failure to position the tongue against the palate was observed in 7.6%, 12.2%, 11.0%, and 10.0% of patients aged ≤20, 21–40, 41–60, and ≥61 years, respectively. Head rotation to one side was observed in 5.3%, 11.0%, 12.3%, and 16.9% of the respective age groups. Foreign body artifacts increased progressively with age, from 2.6% in the 0–20-year group to 3.8%, 6.7%, and 10.4% in the 21–40, 41–60, and ≥61-year groups, respectively.
Table 3.
Prevalence of panoramic radiographic errors according to age groups.
Patient movement was most frequent among patients aged 0–20 years (5.0%), followed by those aged ≥61 years (4.8%), 21–40 years (2.7%), and 41–60 years (2.4%). Positioning the patient too far forward was observed in 1.3%, 4.2%, 4.3%, and 6.4% of the respective age groups, whereas positioning the patient too far backward was observed in 1.2%, 2.9%, 2.3%, and 0.8%. Lip positioning errors were most frequent in the 0–20-year group (3.4%) and decreased with increasing age, with frequencies of 2.5%, 1.6%, and 0.8% in the 21–40, 41–60, and ≥61-year groups, respectively. Backward head tilt was observed in 1.4%, 3.0%, 2.6%, and 3.6% of the respective age groups, whereas forward head tilt was observed in 2.5%, 2.0%, 0.4%, and 0%. Slumped posture was observed in 0.5%, 1.4%, 1.4%, and 2.0%, respectively. Failure to bite on the bite block occurred in 0.7%, 0.5%, 0.5%, and 1.2% of the respective age groups. Missing image sections were observed in 0.2%, 0.4%, 0.7%, and 0.4%, whereas exposure errors were observed in 0.6%, 0.6%, 0.1%, and 0%, respectively.
When panoramic radiographic errors were evaluated according to gender, significant differences were observed for several error types (Table 4). Failure to position the tongue against the palate was more frequent in males than females (11.7% vs. 9.1%, p = 0.002), as was head rotation to one side (11.0% vs. 8.6%, p = 0.004). In contrast, foreign body artifacts were substantially more frequent among females than males (7.0% vs. 1.3%, p < 0.001). Positioning the patient too far forward was also more frequent among males than females (4.2% vs. 2.6%, p = 0.003). Exposure errors were observed in 0.3% of females and 0.7% of males (p = 0.020), while missing image sections were observed in 0.2% of females and 0.7% of males (p = 0.010). No statistically significant gender differences were observed for patient movement (p = 0.100), lip positioning errors (p = 0.400), backward head tilt (p = 0.365), positioning the patient too far backward (p = 0.700), forward head tilt (p = 0.600), slumped posture (p = 0.600), or failure to bite on the bite block (p = 0.100).
Table 4.
Prevalence of panoramic radiographic errors according to sex.
Multivariable binary logistic regression analysis revealed significant associations between age group and 10 of the 13 radiographic error types (Table 5). Compared with patients aged ≤20 years, those aged ≥61 years had the highest adjusted odds of being positioned too far forward (OR = 5.123, 95% CI: 2.649–9.905), foreign body artifacts (OR = 4.813, 95% CI: 2.873–8.063), and slumped posture (OR = 4.279, 95% CI: 1.388–13.188). Sex was significantly associated with six error types after adjustment for age. Male patients had higher odds of tongue positioning errors, head rotation, forward positioning errors, exposure errors, and missing images, but lower odds of foreign body artifacts (OR = 0.171, 95% CI: 0.117–0.250). No significant age-group associations were observed for failure to bite on the bite block, exposure errors, or missing images.
Table 5.
Multivariable binary logistic regression analysis of factors associated with panoramic radiographic errors.
3.2. Overall Radiographic Quality Assessment
Regarding overall radiographic quality, 188 radiographs (3.6%) were classified as “excellent,” 4934 (95.7%) as “diagnostically acceptable,” and 33 (0.6%) as “unacceptable” (Table 6). Overall radiographic quality differed significantly across age groups (χ2 = 160.940, p < 0.001). The proportion of excellent radiographs was highest in patients aged ≤20 years (8.2%) and decreased to 2.0%, 0.5%, and 0.4% in the 21–40, 41–60, and ≥61-year age groups, respectively. When radiographic quality was evaluated according to sex, 111 radiographs from female patients (4.0%) and 77 from male patients (3.2%) were classified as “excellent.” The proportion of diagnostically acceptable radiographs was 95.3% among females and 96.1% among males, whereas 0.6% and 0.7%, respectively, were classified as “unacceptable.” No statistically significant difference was observed between females and males in terms of overall radiographic quality (p = 0.264).
Table 6.
Distribution of overall diagnostic quality of panoramic radiographs according to age group and sex.
4. Discussion
Panoramic radiographs are a widespread extra-oral imaging modality that allows clinicians to visualize the lower third of the maxillary sinus, both TMJ structures, and the dentoalveolar complex on a single image. Typically, panoramic radiographs exhibit a magnification factor of between 10% and 30%. Consequently, errors in patient positioning may result in disproportionate and uneven magnification [14]. To ensure the diagnostic quality of panoramic radiography remains adequate, strict attention must be paid to the correct preparation and positioning of the patient [8]. Proper positioning requires aligning the patient’s head accurately along the Frankfort horizontal plane, the canthomeatal line, and the midsagittal plane [14].
Radiographic errors are a frequent occurrence in panoramic dental radiography, stemming from either processing faults or patient preparation and positioning mistakes. However, because digital processing techniques have largely minimized the incidence of processing errors, operators must focus heavily on patient preparation and positioning to yield diagnostically adequate radiographs [17]. Panoramic radiography is a highly technique-sensitive imaging modality in which errors during patient preparation, positioning, or image acquisition may adversely affect and, in some cases, substantially compromise the diagnostic quality of the resulting radiographs. Therefore, assessing the prevalence of errors associated with this widely used technique in dentistry, together with evaluating their potential impact on image quality and the need for repeat examinations, is essential for identifying potential areas for improvement, optimizing radiographic practice, and enhancing the training and competency of dental professionals [2].
Our findings reveal that 40.1% of the evaluated panoramic radiographs contained one or more positioning or preparation errors. In comparison, previously conducted studies have reported much higher fault rates, typically ranging from 89% to 92% [18]. Rushton et al. [4] and Dhillon et al. [6]. reported error rates exceeding 85% and reaching 89%, respectively. Similarly, Schiff et al. [18] reported an error prevalence of 80%, while Suparno et al. [19], Kattimani et al. [10] (82.8%) and Mohite et al. [7] (95%), reported rates of 98.78%, 82.8%, and 95%, respectively. Lower but still relatively high prevalences were reported by Subbulakshmi et al. [16] (64%) and Granlund et al. [20]. (79%). In addition, Akarslan et al. [15] reported that positioning faults accounted for more than 62% of the observed errors. The differences in error prevalence between our findings and previous studies may be associated with variations in patient populations, radiographic equipment, image acquisition protocols, and the criteria used to identify and classify radiographic errors. Patient comprehension and adherence to positioning instructions, as well as communication-related factors during image acquisition, may also contribute to these differences [1]. To the best of our knowledge, the present study represents one of the largest series evaluating panoramic radiographic errors in terms of the number of radiographs assessed.
In the present study, the prevalence of at least one radiographic error differed significantly across age groups, with the highest prevalence observed among patients aged ≥61 years (53.8%). Multivariable logistic regression analysis confirmed that patients in this age group had significantly higher odds of exhibiting at least one radiographic error compared with those aged ≤20 years (adjusted OR = 2.584, 95% CI: 1.973–3.384; p < 0.001). Furthermore, the associations between age and radiographic errors varied according to error type, indicating that different radiographic findings exhibited distinct age-related patterns. These results highlight the importance of considering age-related differences when evaluating the prevalence and distribution of panoramic radiographic errors.
The prevalence of at least one radiographic error differed significantly across age groups, with the highest prevalence observed among patients aged ≥61 years (53.8%). Multivariable analysis demonstrated that this age group had significantly higher odds of presenting at least one radiographic error compared with patients aged ≤20 years (adjusted OR = 2.584). These findings indicate that age group is significantly associated with the prevalence of radiographically identifiable errors, although the direction and magnitude of these associations vary according to the specific error type.
In our study, the most common error was the presence of a palatoglossal airspace shadow (10.3%), caused by the patient failing to raise their tongue against the palate. This aligns with a vast body of literature indicating that improper tongue positioning is the most frequently encountered error [21]. The highest patient positioning error rates in studies by Suparno et al. [19] (49.68%) and Marsha et al. [21] 47.07%) were also attributed to this specific fault. Furthermore, Subbulakshmi et al. [16] (33%), Dhillon et al. [6] (55.7%), and Akaraslan et al. [15] (46.3%) all identified the failure to position the tongue against the palate during exposure as their most common error, which strongly supports our findings. The occurrence of this radiographic finding may be associated with difficulties in understanding or maintaining the required tongue position during image acquisition, potentially influenced by patient-related factors and the clarity of positioning instructions. The consequence of this error is the superimposition of a radiolucent air band at the apical level of the maxillary teeth, which degrades the diagnostic quality of the dental roots and surrounding structures [22]. Some patients may experience difficulty understanding or maintaining the required tongue position throughout image acquisition. In many cases, patients misunderstand the prompt and only place the tip of their tongue on the palate, or simply lose focus during the exposure. It is a challenging error for operators to eliminate because, despite clear instructions, patients do not always maintain the correct posture. Numerous studies corroborate that tongue misplacement is the leading positional error in panoramic radiography [19]. Technicians may struggle to instruct patients effectively on how to swallow and hold their tongue against the roof of the mouth. This incorrect tongue positioning significantly impedes the diagnosis of apical periodontitis and compromises the assessment of root anatomy and resorption [16].
In the systematic review by Fairozekhan et al., the second most common positioning error was head rotation to one side (17.9%). On the radiograph, this error was observed as asymmetry of the condyles, with the ramus appearing larger on one side than the other [23]. Similarly, in our study, head rotation to one side was the second most common positioning error (9.7%), although Mohite et al. [7] observed it as their absolute most common error (33.8%). Conversely, the least frequent positioning error in our dataset was the failure to bite the bite-block, a finding that correlates directly with Akarslan et al. [15], who reported this fault at a remarkably low 2.39%.
When analyzing errors by age group, consistent with research by both Marsha et al. [21] and Kumar et al. [17]. Our data demonstrate that the prevalence of patient movement was most frequent in the youngest age group (≤20 years, 5.0%), followed closely by patients aged ≥61 years (4.8%). Multivariable analysis showed significantly lower odds of patient movement in the 21–40 and 41–60 age groups compared with those aged ≤20 years, whereas no significant difference was observed in patients aged ≥61 years. Researchers attribute this to the inherent difficulty children face when required to remain perfectly still for the duration of the exposure (typically 13–20 s). This lack of stability also explains why vertical and horizontal alignment errors are more prevalent in children [20,24,25]. For example, a child’s chin may slip out of the correct position, leading to distorted images of the mandible. Additionally, the developing dentition of children means their teeth may not be fully erupted, altering how they interact with the bite block and positioning them incorrectly within the focal trough [21]. Patient movement was also relatively frequent among individuals aged ≥61 years (4.8%), suggesting that both younger and older patients may require additional attention during panoramic image acquisition.
The findings reported by Marsha et al. [21] are broadly consistent with those of the present study, particularly with respect to the association between incorrect tongue positioning and age. In the study by Marsha et al. [21] the frequency of incorrect tongue positioning was highest among children and progressively decreased with advancing age. In contrast, the present study showed that this error was observed in 7.6% of individuals aged 0–20 years, 12.2% of those aged 21–40 years, 11.0% of those aged 41–60 years, and 10.0% of those aged ≥61 years. Thus, the lowest prevalence was observed in the youngest age group, whereas the highest prevalence occurred in the 21–40-year group. Importantly, this pattern did not demonstrate a consistent linear increase or decrease with age. This discrepancy may be attributable to age-related declines in motor coordination and the ability to perform complex voluntary movements, as well as to anatomical changes associated with aging. Such factors may adversely affect patient positioning, particularly among elderly individuals with neurological disorders such as Parkinson’s disease, which is characterized by motor impairment [26,27]. Incorrect tongue positioning may also be associated with inadequate communication between the operator and the patient. In particular, insufficient or unclear instructions regarding swallowing and positioning the tongue against the palate may contribute to the occurrence of this error. Furthermore, some patients may misinterpret the instructions and position only the tip of the tongue against the palate, whereas others may fail to adequately comprehend or follow the instructions provided by the operator [14,22].
Lingam et al. [1] reported a statistically significantly higher prevalence of forward head tilting and failure to position the tongue against the palate among individuals aged 19–59 years compared with those in other age groups. In our study, age-related differences varied according to the specific error type. In the study by Lingam et al. [1]., the incidence of foreign bodies was higher in children, whereas in our study, foreign body artifacts increased progressively with age, from 2.6% in the 0–20-year group to 10.4% among individuals aged ≥61 years. This difference may be associated with age-related factors, including the presence of removable dental prostheses, hearing aids, jewelry, or other external objects; however, this explanation should be interpreted cautiously unless the specific types of foreign bodies were systematically recorded.
Previous studies have investigated sex-related differences in panoramic radiographic errors and image quality, reporting no significant associations between sex and overall error prevalence or radiographic acceptability [1,8]. Consistent with these findings, the present study found no significant association between sex and the overall presence of radiographic errors after adjustment for age group. However, multivariable logistic regression analysis revealed significant associations between sex and six of the thirteen individual error types after adjustment for age group. Male patients had higher odds of tongue positioning errors, head rotation, forward positioning errors, exposure errors, and missing images, whereas female patients had higher odds of foreign body artifacts. These findings suggest that although overall error prevalence may not differ significantly between sexes, the distribution of specific radiographic error types may vary.
In the present study, we aimed to determine the prevalence of thirteen different radiographically identifiable positioning and technical errors in panoramic radiographs and to evaluate overall radiographic quality. Although 59.9% of the radiographs showed none of the predefined errors, only 3.6% were classified as excellent, while 95.7% were considered diagnostically acceptable and 0.6% were classified as unacceptable. The difference between the proportion of error-free and excellent radiographs may be explained by the broader criteria used for overall image quality assessment, which included image contrast, sharpness, and anatomical visibility in addition to the predefined errors. Therefore, the absence of identifiable positioning or technical errors did not necessarily indicate optimal image quality. Conversely, although errors were identified in 40.1% of the radiographs, only 0.6% of all images were classified as diagnostically unacceptable. This finding suggests that the presence of radiographic errors does not necessarily compromise diagnostic adequacy. Minor positioning errors may affect the visualization of specific anatomical structures without substantially impairing overall diagnostic interpretation. Accordingly, radiographs containing errors that did not significantly affect diagnostic assessment were classified as diagnostically acceptable. These findings emphasize the importance of distinguishing between the occurrence of radiographic errors, overall image quality, and their impact on diagnostic acceptability.
The guidelines on radiological standards for primary dental care establish quality criteria for dental radiography and recommend that the proportion of unacceptable radiographs should not exceed 10%. Furthermore, a target of at least 70% excellent (error-free) radiographs has been suggested in these guidelines [7]. A limited number of studies have investigated the image quality of panoramic radiographs. Kaplan et al. [8] reported that 18.4% of the radiographs evaluated were perfect, 45.0% were diagnostically acceptable, and 36.6% were unacceptable. Similarly, Belgin et al. [14] found that 18.4% of the images were perfect, 50.3% were diagnostically acceptable, and 31.3% were unacceptable. Mayil et al. reported that 18.7% of the radiographs were obtained under ideal conditions, 78.0% were diagnostically acceptable, and 3.3% were diagnostically unacceptable [28]. Dhillon et al. reported corresponding rates of 11.0%, 64.1%, and 24.9%, respectively [6]. Yasa et al. found that the proportion of diagnostically unacceptable radiographs ranged from 2.6% to 11.3%, depending on the observer [29].
In the present study, only 0.6% of the radiographs were classified as unacceptable. This proportion was considerably lower than those reported by Mohite et al. [7] (10.8%), Brezden and Brooks [30] (18.2%), and Rushton et al. [4] (33%), suggesting that diagnostically adequate panoramic radiographs were predominantly produced in our department. Furthermore, Kaplan et al. [8] reported no statistically significant association between sex and the acceptability of panoramic radiographs, which is consistent with the findings of the present study. However, differences were observed in the proportions of diagnostically acceptable and ideal radiographs compared with previous studies. These discrepancies may be partly attributable to differences in sample size among studies. In addition, variations in observer experience, undergraduate dental education, radiographic training, and assessment criteria across institutions and countries may have contributed to the differences observed between studies [29].
Patient positioning and preparation errors in panoramic radiography may adversely affect the accuracy and reliability of diagnostic evaluations across different dental specialties. Failure to position the tongue against the palate produces a radiolucent airspace shadow that may be superimposed on the periapical regions of the maxillary teeth, making it difficult to assess anatomical structures and pathological lesions in these regions, particularly during endodontic evaluation [2,5]. Similarly, head rotation may cause asymmetric visualization of the mandibular rami and condyles, complicating the accurate assessment of existing anatomical asymmetries or creating the false impression of asymmetry, potentially leading to misinterpretation of radiographic findings in orthodontic and maxillofacial evaluations [5]. Furthermore, patient positioning errors have been reported to increase measurement inaccuracies in panoramic radiographs [14]. Such inaccuracies may lead to incorrect assessments of the locations, dimensions, and spatial relationships of anatomical structures, particularly during dental implant planning and other surgical procedures. In addition, ghost images resulting from the failure to remove metallic objects, such as jewelry, during patient preparation may occasionally be mistaken for odontomas or other radiopaque pathological lesions, potentially leading to diagnostic misinterpretation [16]. Therefore, recognizing patient positioning and preparation errors in panoramic radiography and considering their potential effects during image interpretation are essential for supporting accurate diagnostic evaluation and reliable treatment planning.
Several limitations should be considered when interpreting the findings of this study. First, radiographic errors and overall image quality were evaluated by a single radiologist. Although intra-observer reliability was assessed in 19.5% of the sample and demonstrated high agreement, inter-observer reliability could not be evaluated because a second independent observer was not included. Second, positioning and technical errors were identified retrospectively based solely on their radiographic appearances. Although certain radiographic patterns are suggestive of specific positioning errors, similar findings may arise from different patient-related, operator-related, or technical factors. Therefore, the actual positioning of the patient, the actions of the radiology technician, and the precise causes of the observed errors could not be independently verified. Consequently, the findings should be interpreted as the prevalence of radiographic features suggestive of positioning and technical errors rather than definitive evidence of specific operator or patient related errors. Furthermore, as this was a single-center study, the findings may not be fully generalizable to other clinical settings with different patient populations, imaging equipment, or radiographic acquisition protocols.
5. Conclusions
In this study of 5155 panoramic radiographs, at least one positioning or technical error was identified in 40.1% of the images. The most frequent errors were failure to position the tongue against the palate (10.3%) and head rotation (9.7%). The prevalence of radiographic errors differed significantly across age groups, with the highest proportion observed in patients aged ≥61 years (53.8%). Multivariable analysis confirmed that age group was significantly associated with overall error presence, with patients aged ≥61 years showing the highest adjusted odds compared with those aged ≤20 years (OR = 2.584). Despite the observed errors, 95.7% of the radiographs were classified as diagnostically acceptable, while only 0.6% were considered unacceptable. Overall radiographic quality also differed significantly across age groups, whereas no significant difference was observed between sexes. These findings indicate that radiographic errors are relatively common, but their presence does not necessarily compromise diagnostic adequacy, and that age is an important factor associated with radiographic error prevalence.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics16223791/s1, Table S1: Intra-observer agreement for individual panoramic radiographic error categories; STROBE Statement checklist for cross-sectional studies; RECORD Statement extension checklist.
Author Contributions
Conceptualization: F.C. and F.K. methodology: F.C. and F.K.; software: F.C.; validation: E.H.; formal analysis: F.K. and E.H.; investigation: F.C., F.K. and E.H.; resources: F.C.; data curation: F.C. and F.K.; writing—original draft preparation: F.C., F.K., and E.H.; writing—review and editing: F.K., E.H., and S.B.D.; visualization: F.C.; supervision: E.H. and S.B.D. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
The study was conducted in accordance with the Declaration of Helsinki, and approved by the Institutional Review Board of the Sakarya University, Health Sciences Scientific Research Ethics Committee (Approval date/number: 19 August 2026/649).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Lingam, A.S.; Koppolu, P.; Abdulsalam, R.; Reddy, R.L.; Anwarullah, A.; Koppolu, D. Assessment of common errors and subjective quality of digital panoramic radiographs in dental institution, Riyadh. Ann. Afr. Med. 2023, 22, 49–54. [Google Scholar] [CrossRef] [Scilit]
- Costa, E.D.; Cral, W.G.; Murad, F.P.; Oliveira, M.L.; Ambrosano, G.M.B.; Freitas, D.Q. Prevalence of errors and number of retakes in panoramic radiography: Influence of professional training and patient characteristics. Int. J. Odontostomatol. 2021, 15, 719–726. [Google Scholar] [CrossRef] [Scilit]
- White, S.C.; Pharoah, M.J. Oral Radiology: Principles and Interpretation, 4th ed.; Mosby: St. Louis, MO, USA, 2000; pp. 535–553. [Google Scholar]
- Rushton, V.; Horner, K. The use of panoramic radiology in dental practice. J. Dent. 1996, 24, 185–201. [Google Scholar] [CrossRef] [Scilit]
- Rondon, R.H.N.; Pereira, Y.C.L.; do Nascimento, G.C. Common positioning errors in panoramic radiography: A review. Imaging Sci. Dent. 2014, 44, 1–6. [Google Scholar] [CrossRef] [Scilit]
- Dhillon, M.; Raju, S.M.; Verma, S.; Tomar, D.; Mohan, R.S.; Lakhanpal, M.; Krishnamoorthy, B. Positioning errors and quality assessment in panoramic radiography. Imaging Sci. Dent. 2012, 42, 207. [Google Scholar] [CrossRef] [Scilit]
- Khator, A.M.; Motwani, M.B.; Choudhary, A.B. A study for determination of various positioning errors in digital panoramic radiography for evaluation of diagnostic image quality. Indian J. Dent. Res. 2017, 28, 666–670. [Google Scholar] [CrossRef] [Scilit]
- Kaplan, B.; Katı, E. Evaluation of error types seen in digital panoramic radiographs. Dicle Dent. J. 2024, 25, 73–78. [Google Scholar] [CrossRef] [Scilit]
- Khan, S.Q.; Ashraf, B.; Mehdi, H. Evaluation of patient preparation and positioning errors on digital panoramic radiographs. Pak. Oral Dent. J. 2015, 35, 1. [Google Scholar]
- Kattimani, S.; Kempwade, P.; Ramesh, D.; Byatnal, A.; Nasreen, S.; Subashani, T. Determination of different positioning errors in digital panoramic radiography: A retrospective study. J. Med. Radiol. Pathol. Surg. 2019, 6, 5–8. [Google Scholar] [CrossRef] [Scilit]
- Altındağ, A.; Yalın, H.; Ertürk, M.; Altındağ, Ö. Error Types in Panoramic Imaging. Van Diş Hekim. Derg. 2025, 6, 14–23. [Google Scholar]
- von Elm, E.; Altman, D.G.; Egger, M.; Pocock, S.J.; Gøtzsche, P.C.; Vandenbroucke, J.P. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) Statement: Guidelines for Reporting Observational Studies. PLoS Med. 2007, 4, e296. [Google Scholar]
- Benchimol, E.I.; Smeeth, L.; Guttmann, A.; Harron, K.; Moher, D.; Petersen, I.; Sorensen, H.T.; von Elm, E.; Langan, S.M. The REporting of studies Conducted using Observational Routinely-collected health Data (RECORD) statement. PLoS Med. 2015, 12, e1001885. [Google Scholar]
- Belgin, C.A.; Serindere, G. Evaluation of error types and quality on panoramic radiography. Int. Dent. Res. 2019, 9, 99–104. [Google Scholar] [CrossRef] [Scilit]
- Akarslan, Z.Z.; Erten, H.; Güngör, K.; Çelik, L. Common errors on panoramic radiographs taken in a dental school. J. Contemp. Dent. Pract. 2005, 4, 24–34. [Google Scholar]
- Subbulakshmi, A.C.; Mohan, N.; Thiruneervannan, R.; Naveen, S.; Gokulraj, S. Positioning errors in digital panoramic radiographs: A study. J. Orofac. Sci. 2016, 8, 22–26. [Google Scholar] [CrossRef] [Scilit]
- Kumar, N. Assessment of common errors and subjective quality of digital panoramic radiographs in a dental institution. Dent. Med. Res. 2020, 8, 23–26. [Google Scholar] [CrossRef] [Scilit]
- Schiff, T.; D’Ambrosio, J.; Glass, B.J.; Langlais, R.P.; McDavid, W.D. Common positioning and technical errors in panoramic radiography. J. Am. Dent. Assoc. 1986, 113, 422–426. [Google Scholar] [CrossRef] [Scilit]
- Suparno, N.; Faizah, A.; Nafisah, A. Assessment of Panoramic Radiograph Errors: An Evaluation of Patient Preparation and Positioning Quality at Soelastri Dental and Oral Hospital. Open Dent. J. 2023, 17, e187421062309120. [Google Scholar] [CrossRef] [Scilit]
- Granlund, C.M.; Lith, A.; Molander, B.; Gröndahl, K.; Hansen, K.; Ekestubbe, A. Frequency of errors and pathology in panoramic images of young orthodontic patients. Eur. J. Orthod. 2012, 34, 452–457. [Google Scholar] [CrossRef] [Scilit]
- Marsha, E.A.; Kiswanjaya, B.; Bachtiar-Iskandar, H.H. Comparison of positioning errors on panoramic radiographs in Indonesian children, adults, and elderly patients. J. Stomatol. 2023, 76, 175–181. [Google Scholar] [CrossRef] [Scilit]
- Bissoon, A.K.; Whaites, E.; Moze, K.; Naidu, R. Evaluation of common operator errors in panoramic radiography in Trinidad and Tobago: A comparison of formally vs informally trained operators. West Indian Med. J. 2012, 61, 733–738. [Google Scholar] [CrossRef]
- Fairozekhan, A.T.; Mohammed, F.; Mubarak, S. Positional Errors in Orthopantomograph—A Comparative Systematic Review with analysis of 1014 additional cases. Eur. J. Mol. Clin. Med. 2020, 7, 1711–1723. [Google Scholar]
- Ekströmer, K.; Hjalmarsson, L. Positioning errors in panoramic images in general dentistry in Sörmland County, Sweden. Swed. Dent. J. 2014, 38, 31–38. [Google Scholar]
- Peretz, B.; Gotler, M.; Kaffe, I. Common errors in digital panoramic radiographs of patients with mixed dentition and patients with permanent dentition. Int. J. Dent. 2012, 2012, 584138. [Google Scholar] [CrossRef] [Scilit]
- Paknahad, M.; Hasani, M.; Hasani, M.; Evazpour, A. Evaluation of image magnification in digital panoramic radiography. J. Dentomaxillofac. Radiol. Pathol. Surg. 2016. [Google Scholar] [CrossRef] [Scilit]
- Salemi, F.; Shokri, A.; Falah, K.S. Common errors on conventional and digital panoramic radiographs. Avicenna J. Dent. Res. 2014, 6, 1–4. [Google Scholar] [CrossRef] [Scilit]
- Mayil, M.; Keser, G.; Pekiner, F.N. Clinical image quality assessment in panoramic radiography. Clin. Exp. Health Sci. 2014, 4, 126–132. [Google Scholar] [CrossRef] [Scilit]
- Yasa, Y.; Çoban, D.; Özbey, F.; Tuna, T.; Yılmaz, B.N.G.; Erzurumlu, Z.Ü.; Sadık, E. Evaluation of common diagnostic errors in panoramic radiographs and interobserver agreement in error identification. Med. Rec. 2025, 7, 114–119. [Google Scholar] [CrossRef] [Scilit]
- Brezden, N.A.; Brooks, S.L. Evaluation of panoramic dental radiographs taken in private practice. Oral Surg. Oral Med. Oral Pathol. 1987, 63, 617–621. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.