Round 1
Reviewer 1 Report
Comments and Suggestions for AuthorsThis study offers a unique contribution to the use of raw (unprocessed) 3D data obtained from OCT devices in the diagnosis of keratoconus. Eliminating the dependence on preprocessed data and providing analysis independent of device and software updates is an innovative approach addressing a significant gap in the literature.
In addition, the development of an end-user GUI brings the engineering contribution to clinical practice.
The shortcomings of the study are outlined below:
1. Approximately two-thirds of the 15,457 files were eliminated, and 6,141 were included in the model. However, the criteria used for elimination and the impact of this elimination on data distribution and generalization capacity should be discussed more thoroughly.
2. The structure of the raw 3DV format is not adequately described. (3D volume? Is the number of OCT slices fixed?)
3. Although the extension of EfficientNet-B0 to 16 channels is described, the dimensions of the additional fully connected layer used, its activation functions, and the total number of parameters are not provided.
4. Although the training and validation losses are plotted, the overfitting analysis is not discussed.
5. The ESI score is generated by the device's own algorithm. The error margin or compatibility of this score with the gold standard needs to be discussed.
6. Detailed screenshots of the GUI can be found in the supplementary material.
7. Figures 1 and 2 are not high-resolution.
8. The findings are compared with the literature, but statistical significance tests (e.g., McNemar, Bland–Altman) are not available.
Author Response
Please see the attachment.
Author Response File:
Author Response.pdf
Reviewer 2 Report
Comments and Suggestions for AuthorsDear authors,
You present a novel deep-learning–based method aimed at differentiating between keratoconus and non-keratoconus/suspect keratoconus eyes using raw data from the CASIA2 device. Furthermore, you have developed a user-friendly graphical interface. Overall, I find the idea of the study very useful; however, I have several questions and concerns. Please note that I am not familiar with deep-learning methods.
-
Is the ESI one of the parameters normally displayed by the CASIA2, or is it available only through analysis of the raw data? Please add this information to the Methods section.
-
You chose an ESI threshold of 30 to differentiate between keratoconus and non-keratoconus/suspect keratoconus eyes. From a clinical point of view, however, differentiation between suspect keratoconus and healthy eyes is more important, as manifest keratoconus usually does not present a diagnostic challenge. Many clinicians struggle more with identifying suspect keratoconus among apparently healthy eyes. This might be an important issue to address, or at least to consider for future studies.
-
Please explain clearly what the MAE value represents. Is a value of 6.65 considered high or low in this context?
-
Please explain abbreviations such as PPV (line 229) when they are first mentioned.
-
You describe your method as useful for analyzing large datasets in an “automated” way. However, the user must manually specify whether the eye is right or left. How can this method be efficiently applied to large datasets given this limitation?
-
The first sentence of the Discussion contains two abbreviations and a numerical value that may not be familiar to readers without a background in deep learning. I suggest rewriting this sentence in a more descriptive and accessible manner.
-
Lines 379–380: Is it still possible to compare the degree of ectasia if the predicted classification changes (for example, from non-keratoconus/suspect keratoconus to keratoconus)?
Author Response
Please see the attachment.
Author Response File:
Author Response.pdf
Reviewer 3 Report
Comments and Suggestions for AuthorsThis manuscript presents a technically well-executed and carefully designed deep learning approach for estimating an ectasia score directly from raw anterior segment OCT data. The authors should be commended for the clear description of the model architecture, the robust training strategy, and the strong overall performance metrics, including high discrimination ability and stable agreement with the Ectasia Screening Index (ESI). The attempt to bypass vendor-derived preprocessing and software-dependent indices is timely and conceptually appealing, particularly in the context of reproducibility and long-term consistency.
That said, several conceptual and clinical considerations limit the current impact of the study. First, while the authors frame their work as moving away from a vendor-dependent “black-box” algorithm, the proposed deep learning model remains a black box from the clinical end-user’s perspective. In this sense, the notion of black-box avoidance primarily reflects software independence rather than interpretability. Incorporating a minimal explainability analysis, such as Grad-CAM visualizations, would better align the methodology with the stated motivation of the study.
Second, the use of ESI as the sole reference standard constrains the interpretation of the proposed score, as the model is ultimately trained to approximate an existing vendor-derived index rather than a clinical ground truth. While acceptable for a proof-of-concept study, this limitation should be more explicitly acknowledged when discussing the clinical significance of the results.
Third, the manuscript suggests potential utility of the continuous ectasia score for longitudinal monitoring(line 380-382); however, the study design is entirely cross-sectional. Claims regarding temporal trends or disease progression should therefore be tempered or clearly framed as hypotheses for future longitudinal validation.
Finally, the absence of external validation limits the ability to assess whether the proposed score offers meaningful utility beyond reproducing ESI within a single device and institutional setting. Demonstrating consistent performance on raw OCT data acquired from different devices or independent institutions would substantially strengthen the argument that this approach provides a robust alternative to vendor-specific indices. If such validation is not feasible at present, this limitation should be clearly discussed to appropriately contextualize the contribution of the study.
Author Response
Please see the attachment.
Author Response File:
Author Response.pdf
Round 2
Reviewer 1 Report
Comments and Suggestions for Authors--
Author Response
Thank you for reviewing the manuscript
Reviewer 3 Report
Comments and Suggestions for AuthorsThe inclusion of Grad-CAM visualizations is appreciated and addresses the request for a minimal explainability analysis.
However, for illustrative purposes, the current examples—showing two cases with similarly high ESI values—provide limited insight into how the model’s attention differs across disease states.
To improve interpretability, it would be more informative to present representative Grad-CAM examples from a normal eye (low ESI) and an ectatic eye (high ESI), rather than two high-ESI cases. This comparison would more clearly demonstrate how the model’s attention changes between non-ectatic and ectatic corneas, particularly given that the Grad-CAM maps are overlaid on averaged OCT images.
Author Response
Please see the attachment.
Author Response File:
Author Response.pdf