The Reliability and Validity of an Isometric Knee Strength Measurement Device in Older Adult Individuals

,
,
,
,
and
1
Regional Industrial Innovation Department, Korea Institute of Industrial Technology, Cheonan-si 31056, Republic of Korea
2
Department of Physical Therapy, Konyang University, Daejeon 35365, Republic of Korea
3
Department of Physical Therapy, Sahmyook University, Seoul 01795, Republic of Korea
*
Author to whom correspondence should be addressed.
This article belongs to the Collection Sensors in Biomechanics

Abstract

This study aims to evaluate the reliability and validity of the Leg Strength Analyzer (IB-LS) in assessing isometric knee flexion and extension strength in elderly adults, compare its performance with that of the CSMI dynamometer, and examine its agreement with isokinetic knee strength measurements. A total of 21 elderly participants (mean age: 65.10 ± 4.56 years) were recruited. Participants underwent knee flexion and extension strength assessments using both the IB-LS and CSMI devices, with isokinetic strength at 60°/s measured in a follow-up session at least one day later. The IB-LS demonstrated high test–retest reliability in elderly adults (ICC = 0.856–0.987). The validity analysis comparing IB-LS isometric peak torque with CSMI isometric peak torque showed moderate to high validity (ICC = 0.826–0.946). Furthermore, IB-LS isometric peak torque and CSMI isokinetic 60°/s peak torque demonstrated high agreement (ICC = 0.775–0.881), demonstrating its strong association with isokinetic strength assessments. Bland–Altman analysis revealed that mean differences between IB-LS and CSMI isometric peak torque values ranged from 13.2 to 93.5 Nm, with limits of agreement (LoA) spanning from −55.8 to 192.5 Nm. When comparing IB-LS isometric peak torque with CSMI isokinetic 60°/s peak torque, mean differences ranged from 31.0 to 56.0 Nm, with LoA from −28.5 to 138.9 Nm. The IB-LS is a reliable and valid tool for evaluating isometric knee strength in elderly adults. Its strong agreement with the CSMI dynamometer and close correlation with isokinetic strength measurements indicate that IB-LS can be a feasible alternative for assessing knee strength in clinical and research settings focused on elderly populations.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.