Open AccessFeature PaperReview

On the Corrosion Behaviour of Low Modulus Titanium Alloys for Medical Implant Applications: A Review

1
Faculty of Materials and Metallurgical Engineering, Semnan University, Semnan 35131-19111, Iran
2
School of Mechanical Engineering, University of Adelaide, Adelaide, SA 5000, Australia
3
College of Science and Engineering, Flinders University, Clovelly Park, SA 5042, Australia
*
Author to whom correspondence should be addressed.
Metals 2019, 9(8), 878; https://doi.org/10.3390/met9080878
Received: 2 July 2019 / Revised: 30 July 2019 / Accepted: 6 August 2019 / Published: 9 August 2019
(This article belongs to the Special Issue Failure Analysis of Biometals)
  |  
PDF [3163 KB, uploaded 19 August 2019]
  |  

Abstract

The corrosion behaviour of new generation titanium alloys (β-type with low modulus) for medical implant applications is of paramount importance due to their possible detrimental effects in the human body such as release of toxic metal ions and corrosion products. In spite of remarkable advances in improving the mechanical properties and reducing the elastic modulus, limited studies have been done on the electrochemical corrosion behaviour of various types of low modulus titanium alloys including the effect of different beta-stabilizer alloying elements. This development should aim for a good balance between mechanical properties, design features, metallurgical aspects and, importantly, corrosion resistance. In this article, we review several significant factors that can influence the corrosion resistance of new-generation titanium alloys such as fabrication process, body electrolyte properties, mechanical treatments, alloying composition, surface passive layer, and constituent phases. The essential factors and their critical features are discussed. The impact of various amounts of α and β phases in the microstructure, their interactions, and their dissolution rates on the surface passive layer and bulk corrosion behaviour are reviewed and discussed in detail. In addition, the importance of different corrosion types for various medical implant applications is addressed in order to specify the significance of every corrosion phenomenon in medical implants. View Full-Text
Keywords: low modulus titanium alloys; corrosion; medical implants; young’s modulus low modulus titanium alloys; corrosion; medical implants; young’s modulus
Figures

Figure 1

This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited (CC BY 4.0).
SciFeed

Share & Cite This Article

MDPI and ACS Style

Afzali, P.; Ghomashchi, R.; Oskouei, R.H. On the Corrosion Behaviour of Low Modulus Titanium Alloys for Medical Implant Applications: A Review. Metals 2019, 9, 878.

Show more citation formats Show less citations formats

Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Related Articles

Article Metrics

Article Access Statistics

1

Comments

[Return to top]
Metals EISSN 2075-4701 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert
Back to Top