1. Introduction
In proprioception, the good functioning of the feedback circuit created by the electrochemical signals propagating along the neuronal fibers is a key and crucial element for the correct management of the spatial movement of the limbs at both a conscious and unconscious level [
1,
2]. At the same time, abnormal ion concentration gradients, mainly Na
+ and K
+, are a determining factor in the production of motor perturbation in conditions of neurodegenerative pathologies such as Parkinson’s disease. As well, the disordered ion concentration gradients promoted by active ion transport, mainly to Na
+ and K
+, are critical for exploring therapeutic interventions for neurodegenerative conditions such as Parkinson’s disease [
2]. Through the activity of voltage-gated ion channels, changes in membrane potential generate action potentials in the form of “trains” of waves that travel along afferent pathways to the central nervous system that modulate the activity of motor neurons, regulate contraction muscle and maintain postural stability through adequate proprioceptive mechanisms [
3,
4]. The modulation of action potentials, influenced by factors such as the dynamics of ion channels and the activity of neurotransmitters, is particularly relevant in the management of muscle tone, and lends itself to insights into the complex interactions between neurophysiological mechanisms and perceptual experiences [
5].
Techniques such as deep brain stimulation (DBS), although very invasive, which involves high-frequency stimulation of specific brain regions, have proven promising in the modulation of neuronal activity in Parkinson’s subjects [
6]. Non-invasive approaches, such as local mechanical vibration (FMV), are based on postural and proprioceptive improvement using local vibrational stimuli capable of strengthening proprioceptive feedback [
7,
8]. Likewise, non-invasive techniques such as thermal blocking of action potentials using infrared sources, direct modulation of the membrane potential using an external electric field and indirect modulation of the membrane potential using an external variable magnetic field have been studied. In particular, with regard to the thermal block of action potentials, the study by Mohit Ganguly et al. [
9] based on Hodgkin–Huxley modelling with the introduction of the localized thermal effect into the model, demonstrated that the increase in temperature accelerates the activation of voltage-gated potassium ion channels, leading to a hyperpolarization condition, causing the action potential block.
An external electric field can modify the membrane potential, which is typically mediated by the extracellular electrolyte fluid, thus influencing the opening or closing of voltage-gated channels. Ions into and out of the cell can be moved directly by the electric field, which influences the concentration of the ions and subsequently alters the dynamics of the channel, as observed in the work of Mortimer and Bhadra on monopolar anodal stimulation [
10]. Speculations about time-dependent magnetic fields have been exploited in practical non-invasive applications of ion cyclotron resonance (ICR), a method that permeates variable magnetic fields within tissues. According to the principles outlined by Abraham R. Liboff [
11], Faraday’s law of induction suggests that a changing magnetic field can induce an electromotive force (EMF) within the electrolyte surrounding a neuron. Liboff’s insights into the interaction of cyclotron resonance with membrane-bound ion channels suggest that this mechanism could influence gating processes that control ion flux.
The literature on the efficacy of non-invasive approaches in the management of Parkinson’s disease (PD) has therefore ranged from local vibrational stimulation and modulation of membrane potential by external electric and magnetic fields [
8,
9,
10,
11]. These approaches have demonstrated significant improvements in motor functions and postural stability, providing a solid basis for further research and development of non-invasive, autonomous and wearable devices, whose mechanism of action is focused—at a hierarchical level—on the most proximal causes of uncontrolled or poorly modulated neurostimulations, as occurs in Parkinson’s patients.
Therefore, this study introduces the use of the Equilibrion® device, a novel non-invasive passive sensor that detects and responds to an input from the physical environment, specifically the concentration of ionic charges in the extracellular fluid of the peripheral neuronal membrane, and responds with an action mechanism based on the dielectric properties of the material preliminary researched with numerical approaches. The Equilibrion® class I medical device, compliant with Regulation (EU) 2017/745, through a capacitive mechanism detailed in this study that has recently received the title of industrial invention patent in Italy and is currently in consideration for EU and USA patent office (still in the pending phase) provides local interaction with the skin to modulate the ionic environment around neuronal membranes. The evaluation has been conducted through a double-blind, double-dummy, randomized controlled observational study (RCOS). This study aimed at systematically collecting clinical sample data on the CE marked device, therefore employing medical device according to indications in a context of normal clinical practice, namely excluding interventions outside of what is claimed in instructions to patients. By measuring and observing improvements in limb mobility and postural stability, we aimed to statistically evaluate the effectiveness of intervention of a non-pharmacological supportive therapy for PD patients. Specifically, this study investigated the enhancement of limb mobility and postural stability in patients with Parkinson’s disease, namely the potential impact on quality of life on individual clinical cases.
2. Materials and Methods
Designed for skin application, the device aims to rebalance ionic charges within the fluid environment of neuronal membranes. Fabricated by A.P.R. Meds srl (Italy), it is a registered Class 1 medical device categorized for neurological physiotherapy, in accordance with European Directive 93/42/EEC–2007/47/EEC, registered with the Ministry of Health under registration number 2434444, protected by industrial patent in Italy, invention patent no. 102022000011933, currently in pending state for extension in USA and EU.
It consists of a 25 mm circular plate (
Figure 1), composed of layers of inert and biocompatible polymeric materials with a specific dielectric constant within the range of relative dielectric constant at 1 MHz, from 2.0 to 5.0.
The mechanism of action of the medical device is strictly correlated to the dielectric properties obtained by assembling 12 layers of specific polymers to achieve the target dielectric constant, based on preliminary studies conducted using the axon numerical model, as depicted below.
A precise dielectric feature is fundamental for the proper interaction with the peripheral nervous system at those target points therapist identify for PD patients based on standardized sensory area targets, as discussed in detail in the
Section 4
.
Regarding the capacitive device’s mechanism of action, it was important for the therapists involved in this study to better understand the key foundations of the device. These foundations were the objectives of the research and development results at the bioengineering and material specialists’ perspectives, just summarized in this work. Therefore, the present work present dual methodological approach:
- (a)
Development of numerical modelling of an axon using the Hodgkin–Huxley differential equation model 12 that allowed developers in selecting property of materials constituting the medical device for the desired interactions with neuronal membranes.
- (b)
Observational study to point out effectiveness level on significative samples with Parkinson Disease patients selected under constraints of study protocol, in the following discuss in depth.
2.1. The Numerical Model
During the development phase, the device was preliminarily studied and numerically validated by developing a numerical model in a MATLAB environment of the baseline Hodgkin–Huxley equivalent circuit (also HH, in the following) (see
Figure 1), subsequently modified with the appropriate insertion of a capacitor or a generator of current, which models the presence of an external element with dielectric behavior (see
Figure 2).
The Hodgkin–Huxley model [
12] is analogous to a simple electrical circuit with batteries, resistors, and capacitors. It describes the flow of ionic currents through the neuron’s membrane, which can be modelled as resistors, and the capacitive properties of the membrane. The model focuses on the movement of sodium (Na
+), potassium (K
+), and leak currents through specific ion channels. When sodium channels open, Na
+ flows into the neuron, causing depolarization and generating an action potential. Subsequently, sodium channels close and potassium channels open, allowing K
+ to exit the cell, leading to repolarization and then hyperpolarization, returning the membrane to its resting potential. The membrane’s conductance is determined by the probability that the ion channels are open, which is voltage dependent. Hodgkin and Huxley developed differential equations to describe these probabilities and their dynamics over time. The conductance of ion channels, such as sodium and potassium, increases in response to specific voltage changes.
In preliminary studies for defining the design of the device capable to interact with electric (in reality, they are electrochemical signals in the real axons) signals with the Hodgkin–Huxley model, a numerical model implemented in MATLAB was constructed to simulate action potential propagation along an axon. Key steps included:
Membrane potential equations: Gating variables, for ion channels opening regulation:
and coefficient
α and
β depending on membrane potential V and fixed parameters. All parameters have been selected considering healthy humans, specifically the maximum conductances of the ion channels (gNa, gK, gL), and the equilibrium potentials (E
Na, E
K, E
L). These values are based on well-established physiological data from classic neuroscience research [
12,
13,
14,
15,
16,
17].
- 4.
MATLAB Simulation Code
Based on the developed model, experiments were conducted to verify that in the presence of certain parameters of the dielectric, the capacitive element placed on the skin is capable of interacting with the ionic environment that envelops neuromembranes, thus modulating neuronal action potentials within the central nervous system.
The axons within this model were divided into discrete segments to allow precise measurement of the threshold block length throughout the axonal model (
Figure 2).
The model incorporated a simulated microcurrent generator to mimic the effect of an external capacitive and dielectric element (
Figure 3), simulated in the Hodgkin–Huxley varied model a current generator), reflecting what produced in dielectric materials due to electrostatic induction realigning polar molecules in response to electric fields.
2.2. Methods and Procedures of Observational Study
A double-blind, double-dummy, randomized controlled observational study investigated the effectiveness of a standard balance training program combined with the use of the wearable proprioceptive stabilizer Equilibrion®. This study did not involve any interventions beyond the normal clinical practice described in the product leaflet. Patients were recruited in accordance with stringent ethical guidelines; informed consent was obtained from all enrolled patients, ensuring that they were fully informed about this study’s purpose. Personal and health data gathering procedures that were processed in compliance with GDPR (EU Regulation 2016/679) and national legislation, ensuring anonymization and secure storage. Participant documents were kept in a secure location, identified only by a code known to researchers. Anonymized data were available for regulatory review and scientific publications. The study protocol was reviewed and approved by the internal Ethics Committee of manufacturer as the body entity responsible of conformity to EU Regulation 745/17 of the Class I medical device, thus ensuring adherence to Good Clinical Practice and Helsinki Declaration principles. The study followed principles of phase III clinical observational study targeting Parkinson’s patients with specific clinical characteristics to evaluate the therapeutic support provided by the passive sensor device, in combination with standard pharmacological and physiotherapy treatments. Moreover, patients were informed that this study aimed to accurately measure the effects of the new treatment compared to existing treatments and they would be randomly assigned to receive either the experimental treatment or a placebo, which contains no active ingredient. As an auxiliary therapy to improve Parkinson’s disease symptoms, posture, movement stability, and reduce the risk of falls, patients did not modify their own standard pharmacological and physiotherapeutic treatments. Unique requirements and the application of the therapeutic wearable passive sensor occurred when the effects of the previously administered pharmacological treatment began to wear off, specifically before the next scheduled dosage. Patients also were informed no risks have been described in the technical documents and material specifications provided by the manufacturer for class I medical devices, except for a skin allergy due to the patch used to apply the device.
This study enrolled 50 patients who were randomized into one of the following groups:
The two groups underwent the following program.
The active group PD/EQ: PD patients were treated with an innovative postural stabilizer device (Equilibrion) for 30 min before postural check with Geit Analysis parameters.
The non-active group PD/P: PD patients treated with a placebo device (similar to Equilibrion but inactive) for 30 min before postural check with Geit Analysis parameters.
Five functional and non-functional devices were applied to the skin of participants in both the active (non-placebo) and non-active (placebo) groups using a medical band, following specific standardized Sensory Area Targets:
Position 1: the 7th cervical vertebra;
Position 2: the bilateral rectus femoris muscle belly on the right leg;
Position 3: the bilateral rectus femoris muscle belly on the left leg;
Position 4: the bilateral gastrocnemius muscle on the right leg;
Position 5: the bilateral gastrocnemius muscle on the left leg.
The application positions follow the protocol developed by the medical therapists at fabricant A.P.R. Meds Srl. This protocol is fundamentally based on identifying areas at the five application points (By most clinical-therapeutic details: 1. C7 (7th cervical vertebra), as it stimulates the head-righting reflex. The cervical region is rich in proprioceptors, which helps to alleviate upper limb adiaconcinesia, improving the kyphotic posture of the upper trunk (camptocormia). 2. Bilateral Rectus Femoris Muscle Belly: Muscle rigidity in patients with extrapyramidal syndrome is a clinically significant symptom that affects the complete extension of the knees, compromising gait and static and dynamic stability. 3. Bilateral Triceps Surae Muscle: Given that plastic rigidity predominates in the muscular system of patients with extrapyramidal syndrome, these points were identified to free the ankle joint, enhancing its flexion–extension, thereby reducing foot dragging, freezing episodes, and the risk of falls). that are functionally and purposefully involved with motor neurons and peripheral nerve bundles. This is because in PD patients, there is a reduction in dopamine production by the putamen. Dopamine is crucial for regulating the transmission of nerve impulses. The decreased dopamine levels result in altered nerve signal transmission, often transmitted as unmodulated bursts of action potentials. In PD patients, these areas are affected by disrupted and spastic signals, which impair posture and mobility. Therefore, the 5 positions are selected as the most critical points for intervention, as the effect of the capacitive wearable device would progressively modulate the action potential trains that are transmitted to key muscles responsible for posture and dynamics in PD patients.
After the application of the wearable skin devices, patients were asked to sit for 30 min to allow for CNS interaction to commence. The T1 motor impairment TUG (Timed Up and Go) test was then initiated for data acquisition.
Inclusion criteria for all patients admitted to the Centro Fisioterapico Medico C.F.M Srl in Treviglio (Italy) were:
Parkinson’s disease stage II–III, evaluated with the H&Y scale,
Presence of postural alterations,
Presence of postural instability
Ability to participate in physiotherapy,
Absence of cognitive impairment (MMSE > 24/30), and
Stable medications.
Exclusion criteria:
Presence of DBS;
Severe cardiac and/or pulmonary disease;
Patients underwent physiotherapy treatment while in an “off medication” state, meaning that the effects of levodopa had worn off.
The study coordinator responsible for statistical analysis was blinded to the group allocation. The therapist providing the interventions was blinded to the allocation of functional and non-functional devices.
Both functional (applied to the active group) and non-functional (applied to non-active group) wearable devices were identical, not distinguished by their external appearance, and the constituent inert materials used for non-functional devices did not target the specific range for the dielectric constant and insulating parameters that allow functional response. Furthermore, as a passive device, neither functional nor non-functional devices caused any recognizable sensory sensation, guaranteeing patients’ blindness.
2.3. Clinical Assessment of Admitted Patients
Baseline data included age, gender, body mass index (BMI), disease duration, anti-PD medications expressed as levodopa daily dose (
Table 1), cognitive status assessed using the Mini-Mental Status Exam (MMSE) or MMSE [
18], and progression of Parkinson’s disease symptoms assessed using the Hoehn and Yahr Scale [
19].
Table 1 summarizes the characteristics and clinical data from the active group (
n = 25) with the placebo group (
n = 25) of patients with Parkinson’s disease (PD). Clinical parameters are: body mass index (BMI), disease duration, Mini-Mental State Examination (MMSE) scores, daily levodopa dose (LED), age, and the Hoehn and Yahr (H&Y) scale. Based on the statistical analysis results:
- -
The mean BMI of the active group was slightly higher than that of the placebo group, but the Mann–Whitney U test yielded a p-value of 0.685, indicating no statistically significant difference between the two groups in terms of BMI.
- -
Disease duration: Both groups had a similar mean disease duration, with a Mann–Whitney U p-value of 0.624, suggesting no significant difference in the duration of PD diagnosis.
- -
The MMSE measures cognitive impairment, with higher scores indicating better functioning. The mean MMSE score of the active group was marginally higher than that of the placebo group, but the p-value of 0.282 from the Mann–Whitney U test suggests no significant cognitive performance difference between the groups.
- -
The daily levodopa dose (LED) was obtained from the pilot study. The average daily levodopa dose was slightly higher in the active treatment group than that in the placebo group. However, a p-value of 0.727 indicated no significant difference in medication dosage between the two groups.
- -
There was a noticeable difference in the mean age between the active (77.56) and placebo groups (67.15). The Mann–Whitney U test yielded a p-value of approximately 0, indicating a statistically significant difference in age between the two groups. This could result in a potentially critical finding in this pilot study, potentially affecting the outcomes and interpretation of treatment effectiveness.
- -
The Hoehn and Yahr scale was used to describe PD progression. The scores were similar between the groups, with no significant differences observed (chi-square p-value of 0.607).
4. Discussion
The non-pharmacological approach employed in the pilot study involved a medical device based on polymeric material multilayers with a specific dielectric constant to passively modulate the repolarization of the neuronal membrane, thus representing a non-invasive therapeutic strategy for a spectrum of neurological conditions. Dielectric materials are substances that do not conduct electricity due to the random orientation of their constituent molecules. However, they can become polarized in response to the presence of electric charges or when immersed in an ionically charged electrolyte, creating internal electric dipoles. Because of their ability to orient these dipoles, dielectric materials are employed in capacitors to store electrical energy and also act as electrical insulators. In fact, inserting a dielectric between the conductors of a capacitor increases its capacitance. This occurs because the polarization of the dielectric reduces the electric field between the conductors, allowing for a greater amount of charge to be stored for the same applied voltage.
Among the characteristics of materials, the dielectric constant is the most relevant parameter that measures a material’s ability to store electrical energy. It indicates the capability of internal dipole orientation under external electric fields. Higher values indicate a greater capacity for energy storage.
When a dielectric material is immersed in a positively charged electrolyte solution, an electric double layer (EDL) [
22,
23] forms at the interface between the dielectric and the solution. This process involves the creation of two distinct layers of opposite charges [
18,
19]:
- -
The Helmholtz Layer (Inner Layer): This is the first layer that forms directly on the surface of the unpolarized dielectric. Positive ions in the electrolyte move towards the dielectric’s surface. In response, negative charges within the dielectric align towards the surface to balance the positive charges. This results in a zone of positive charges in the electrolyte and a zone of negative charges on the dielectric’s surface, forming the initial part of the electric double layer.
- -
Diffusion Layer (Outer Layer): This second layer extends into the electrolyte solution beyond the Helmholtz layer. Here, the positive ions are more weakly bound to the dielectric surface and gradually spread out into the solution. In this layer, the concentration of positive ions decreases with distance from the dielectric surface, forming a gradual distribution of charges.
The establishment of the electric double layer has direct effects on both the dielectric and the electrolyte:
- -
Dielectric: The dielectric molecules polarize, with negative charges aligning towards the surface to balance the positive ions in the electrolyte.
- -
Electrolyte: Positive ions accumulate near the dielectric surface, attracted by the negative charges on the dielectric.
This intrinsic property of dielectric materials and their interactions with electric or electrochemical states is crucial for analyzing the behavior of specific dielectric materials in various applications, including their role in therapeutic medical devices.
Based on above premises, the mechanism of action of the medical device used in this observational study is primarily based on its own dielectric constant, investigated originally with numerical approach, which enables the capacitive device applied to the skin to interact with membrane extracellular electrolytes by exploiting ion charges.
In summary, the interaction between an unpolarized dielectric and a positively charged electrolyte solution leads to the formation of an electric double layer, significantly altering the material’s internal electrical properties through dipole orientation and the external chemical properties of the electrolyte solution. The key therapeutic effect is based on the internal change in the electric properties of the material constituting the device. For a better understanding, we refer to
Figure 6 and
Figure 7. The two schematic views provide a comparison of what happens locally in afferent neuronal cells (i.e., epidermal terminals reporting information to the central nervous system) without (
Figure 6) and with (
Figure 7) the application of the claimed medical device.
Figure 6 illustrates the process of charge distribution and membrane potential changes in an axon during depolarization. Initially, positive charges accumulate on the outer surface of the axon membrane, maintaining the resting potential (
Figure 6a). When a local depolarization is triggered and surpasses a certain threshold, the charge distribution in that specific area inverts (
Figure 6b). This local inversion of charges causes an increase in membrane potential in adjacent areas, which also surpasses the threshold, leading to an increased permeability of the membrane to sodium ions (Na
+) (
Figure 6c). Consequently, a depolarization front moves along the axon. Following this, the membrane’s permeability to potassium ions (K
+) increases while the permeability to Na
+ decreases, returning the potential to negative values (
Figure 6d). This sequence of events allows the depolarization front to propagate along the axon, enabling the transmission of nerve impulses.
Figure 7 depicts the situation of external stimulus and a reversal of the extracellular membrane polarity of the neural cell (i.e., peripheral nerve bundle). The reversal of membrane polarity is responsible for creating the above-threshold potential. However, the local presence in the target area—where the stimulus arises in the receptors—of the functional layer material with a predetermined dielectric constant polarizes passively and attracts more positive ionic charges from the extracellular fluid by electrostatic attraction. This phenomenon of ionic charge attraction in the extracellular fluid around the neural cell membrane is capable of restoring (or at least reducing) the local depolarization of the neural cell’s extracellular membrane. This phenomenon is due to the effective auto polarization of the material constituting the functional layer present in the invention’s medical device: the auto-polarization of the material of the dielectric layers claimed is responsible for the local response (see again
Figure 7), causing the ionic attraction and the restoration or reduction in the action potential trains thus providing the technical effect.
One key technical distinguishing feature is to provide a simple passive function in which the dielectric material layers acts as an additional capacitive element that passively, and thus without a need of an external energy source, attracts additional positive charges from extracellular fluid of neural cell membrane thus restoring polarization or reducing neural cell membrane depolarization within the target areas.
The numerical model demonstrates that introducing an element with capacitive function, capable of interacting with the neuronal membrane (specifically with the extracellular fluid of the neuronal membrane) in critical intervention areas (the SAT points), can lead to signal remodulation and correction. This is evidenced by the modulation of the signal on a cartesian plane (millivolt versus milliseconds) shown in
Figure 4, where the capacitor effectively remodulates the signals.
The technical effect on real cases, in comparison with numerical models, was therefore statistically investigated by the observational clinic study, structured to facilitate the comparison of the effect of an intervention on patients with PD against a control (placebo) group. The statistical tests applied to groups were matched for each data type (continuous vs. categorical). Differences in the mean age between the groups must be addressed in future studies, as it can influence the interpretation of the results and the overall conclusions drawn about the efficacy of the intervention. Except for age, the statistical analyses suggested no significant differences in most PD-related clinical parameters between the two groups.
Statistical analysis of the active (no placebo) group yielded significant insights into the therapeutic effects on various motor functions and risk factors associated with Parkinson disease.
Regarding the performance of the specific tasks in the BBS protocol, by results summarized in
Table 3, we point out the following:
- -
t-statistic: This value measures how much the groups’ means differ in relation to the variability within the groups. A larger absolute t-value indicates a larger difference between the groups.
- -
p-Value: Shows the likelihood that the observed difference happened by chance. A value less than 0.05 indicates a significant difference.
- -
Significance, indicating whether the change is significant based on the p-value.
For the active group, the t-statistic of −3.72 and p-value of 0.0011 indicate a significant difference, meaning the observed changes are unlikely to have occurred by chance. For the non-active group, the t-statistic of 0.55 and p-value of 0.588 suggest that the differences observed are not statistically significant, indicating that they could be due to random variation.
Regarding the performance of the active group during the TUG tests, the Shapiro–Wilk test p-value show that most parameters had approximately normal distributions, with p-values above 0.05, except for P2 and P10.
Table 5 shows the statistical analysis result using the paired
t-test that helps us understand how the treatment affected different aspects of patients’ mobility by comparing measurements taken before (T0) and after (T1) the treatment. Here below are listed the specific parameters that showed significant changes and what they mean for the dynamics of walking:
- -
P1 (Stride Length): Measures the length of each step. An increase in stride length indicates an improvement in the patient’s ability to walk with longer and smoother steps, which is often impaired in Parkinson’s patients.
- -
P2 (Anteroposterior Acceleration—Lift Off): Measures the forward/backward acceleration during the initial phase of standing up. A decrease in P2 suggests better control of lift-off movements, indicating greater efficiency and less effort required to initiate movement from a seated position.
- -
P5 (Trunk Flexion–Extension during Lift-Off): Measures the degree of trunk flexion and extension during lift-off. Improvements in P5 indicate better postural stability and control during lift-off, reducing the risk of falls.
- -
P6 (Flexion Range): Measures the range of trunk flexion movement during lift-off. An increase in flexion range indicates greater flexibility and movement capability, essential for stability and postural control.
- -
P7 (Mid-turn Rotation): Measures the duration of mid-turn rotation. A longer duration may reflect greater caution and control during rotational movements, which are often challenging for Parkinson’s patients.
- -
P10 (Mid-turn Rotation Duration): Measures the duration of mid-turn rotation. An increase in duration suggests that patients are more careful and balanced during turns, reducing the risk of falls.
For the non-active group of 25 patients who did not undergo therapeutic treatment, none of the parameters exhibited significant changes, as indicated by p-values above 0.05. The closest approach was P7 with a p-value of 0.166, suggesting no substantial effect attributable to the placebo.
Summarizing the comparative results between the active and non-active groups, different behaviors in terms of statistical significance and the magnitude of changes can be observed:
- -
Normality Assessment (Shapiro–Wilk Test): Most parameters in both groups followed a normal distribution, indicating the suitability of parametric tests for the analysis.
- -
Paired t-Test Results: The active group showed significant changes in several parameters post-treatment, indicating the effectiveness of the therapeutic intervention. In contrast, the non-active group showed no significant changes, suggesting the placebo had no measurable effect.
- -
The statistical analysis of the placebo group shows the absence of a therapeutic effect from non-active treatment, contrasting with the active group, where certain parameters exhibited significant improvements. This highlights the effectiveness of the therapeutic intervention in improving specific aspects of motor function in patients with Parkinson’s disease. In particular, there is a significant reduction in the risk of falling and improvements in both anteroposterior accelerations and flexion–extension of the shaft, which are crucial for maintaining stability and mobility in patients with Parkinson’s disease.
Ultimately, the application mapping for device placement is key to producing effects on the mobility-related parameters.
As major outcome from patients’ perspectives, one major advantage of the medical device employed in this study is that it does not require to be powered by any internal or external energy source and does not release any substance into the body nor extract drugs, bodily fluids, or any other substances from the body.
5. Conclusions
This study aimed to evaluate the efficacy level of the capacitive autonomous wearable device (i.e., passive sensor, not powered or sourced) in modulating neuromuscular activity in Parkinson’s patients. The device was applied at five critical intervention points, specifically selected by therapists to target key muscles involved in posture and movement. The main advantages over other non-invasive methodologies are:
- ▪
Localized intervention without systemic side effects, improved patient compliance due to the non-invasive nature, and continuous therapeutic effects.
- ▪
Compared with other non-invasive, non-pharmaceutical treatments such as magnetic cyclotron resonance, laser treatments, and anodic/cathodic current applications (all mentioned in the introduction), this wearable solution, characterized by the passive action of the device (no power source needed), allows patients to continue therapeutic treatment autonomously after the initial study and application by a therapist.
- ▪
This targeted modulation helps to progressively correct the aberrant action potential trains responsible for impaired motor function in Parkinson’s patients over prolonged application times of up to 8 h per day.
- ▪
The wearable nature of the device provides continuous therapeutic effects without the need for repeated clinical visits, thus improving patient compliance and overall quality of life.
The results from our observational study demonstrate significant improvements in several key parameters of motor function, including stride length, anteroposterior acceleration, trunk flexion–extension, and mid-turn rotation duration. These improvements were statistically significant in the active treatment group compared to the placebo group, indicating the device’s efficacy in enhancing mobility and reducing the risk of falls.
The wearable passive sensor used in this pilot study represents a non-invasive and non-pharmacological approach for managing Parkinson’s disease, offering a valid alternative to existing therapies. The passive nature of the device ensures minimal interference with the patient’s daily activities, enhancing compliance and quality of life.
The future purpose of our capacitive wearable device is to serve as a complementary therapy for Parkinson’s disease management, not as a substitute for pharmacological treatment. We aim to integrate this device into standard care protocols to enhance motor function and reduce symptoms such as rigidity and bradykinesia. Future developments will focus on optimizing the device for broader clinical use, including refining its design for better comfort and ease of use, and conducting larger-scale clinical trials to further validate its efficacy. Ultimately, we envision this device becoming a standard tool in the therapeutic arsenal against Parkinson’s disease, potentially extending its application to other neuromuscular disorders.
While the initial results are promising, it is important to acknowledge the limitations of this preliminary study, including the sample size. Further large-scale studies are needed to confirm these findings and to solidify the device’s role in the therapeutic landscape for neurodegenerative diseases.