Next Article in Journal
A Novel Scheme for High-Accuracy Frequency Estimation in Non-Contact Heart Rate Detection Based on Multi-Dimensional Accumulation and FIIB
Previous Article in Journal
Application of MWD Sensor System in Auger for Real-Time Monitoring of Soil Resistance During Pile Drilling
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
This is an early access version, the complete PDF, HTML, and XML versions will be available soon.
Article

Dynamic Calibration of Quartz Flexure Accelerometers

Institute of Systems Engineering, China Academy of Engineering Physics, Mianyang 621999, China
*
Authors to whom correspondence should be addressed.
Sensors 2025, 25(16), 5096; https://doi.org/10.3390/s25165096 (registering DOI)
Submission received: 16 July 2025 / Revised: 10 August 2025 / Accepted: 13 August 2025 / Published: 16 August 2025
(This article belongs to the Section Electronic Sensors)

Abstract

The dynamic behavior of quartz flexure accelerometers remains a subject of ongoing investigation, particularly in areas such as theoretical modeling, standardization, calibration methodology, and performance evaluation. To address the limitation of conventional static calibration models in accurately representing accelerometer responses under dynamic acceleration excitation, a dynamic calibration model is proposed. A mathematical model is first developed based on the physical mechanism of the accelerometer, characterizing its intrinsic dynamic response. Simulation-based analysis demonstrates that the proposed dynamic model offers significantly improved accuracy compared to traditional static approaches. Furthermore, a dynamic calibration method leveraging a dual-axis precision centrifuge is designed and validated. The results confirm that the proposed approach enables the precise calibration of quartz flexure accelerometers in accordance with the dynamic model. The calibration of the dynamic parameter yields a relative standard deviation of −0.048%.
Keywords: dynamic behavior; quartz flexure accelerometers; calibration method; dual-axis precision centrifuge dynamic behavior; quartz flexure accelerometers; calibration method; dual-axis precision centrifuge

Share and Cite

MDPI and ACS Style

Sheng, X.; Wang, X.; Chen, W.; Shu, Y.; Zhang, K. Dynamic Calibration of Quartz Flexure Accelerometers. Sensors 2025, 25, 5096. https://doi.org/10.3390/s25165096

AMA Style

Sheng X, Wang X, Chen W, Shu Y, Zhang K. Dynamic Calibration of Quartz Flexure Accelerometers. Sensors. 2025; 25(16):5096. https://doi.org/10.3390/s25165096

Chicago/Turabian Style

Sheng, Xuan, Xizhe Wang, Wenying Chen, Yang Shu, and Kai Zhang. 2025. "Dynamic Calibration of Quartz Flexure Accelerometers" Sensors 25, no. 16: 5096. https://doi.org/10.3390/s25165096

APA Style

Sheng, X., Wang, X., Chen, W., Shu, Y., & Zhang, K. (2025). Dynamic Calibration of Quartz Flexure Accelerometers. Sensors, 25(16), 5096. https://doi.org/10.3390/s25165096

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop