Point of Care Quantitative Assessment of Muscle Health in Older Individuals: An Investigation of Quantitative Muscle Ultrasound and Electrical Impedance Myography Techniques
Abstract
:1. Introduction
2. Materials and Methods
2.1. Research Cohort
2.2. Investigational Techniques
2.2.1. Quantitative Muscle Ultrasonography (QMUS)
2.2.2. Electrical Impedance Myography (EIM)—Standard Equipment (sEIM)
2.2.3. Electrical Impedance Myography (hEIM)-Handheld Device
2.2.4. Standard Clinical Measures
2.2.5. Lower Extremity Strength and Physical Function Testing
2.2.6. Imaging Studies
2.3. Data Analysis
3. Results
3.1. Intra- and Inter-Rater Reliability
3.1.1. QMUS
3.1.2. sEIM
3.1.3. hEIM
3.2. Correlations
3.2.1. QMUS
3.2.2. sEIM
3.2.3. hEIM
3.2.4. DEXA Correlation with Strength and Functional Measures
4. Discussion
4.1. QMUS
4.2. sEIM
4.3. hEIM
4.4. Study Limitations
5. Conclusions
Author Contributions
Funding
Acknowledgments
Conflicts of Interest
References
- Lauretani, F.; Russo, C.R.; Badinelli, S.; Bartali, B.; Cavazzini, C.; Di Iorio, A.; Corsi, A.M.; Rantanen, T.; Guralnik, J.M.; Ferrucci, L. Age-associated changes in skeletal muscles and their effect on mobility: An operational diagnosis of sarcopenia. J. Appl. Physiol. 2003, 95, 1851–1860. [Google Scholar] [CrossRef] [PubMed]
- Hirani, V.; Blyth, F.; Naganathan, V.; Le Couteur, D.G.; Seibel, M.G.; Waite, M.J.; Handelsman, D.J.; Cumming, R.G. Sarcopenia is associated with incident disability, institutionalization, and mortality in community-dwelling older men: The Concord Health and Ageing in Men Project. J. Am. Med. Dir. Assoc. 2015, 16, 607–613. [Google Scholar] [CrossRef] [PubMed]
- Visser, M.; Goodpaster, B.H.; Kritchevsky, S.B.; Newman, A.B.; Nevitt, M.; Rubin, S.M.; Simonsick, E.M.; Harris, T.B. Muscle mass, muscle strength, and muscle fat infiltration as predictors of incident mobility limitations in well-functioning older persons. J. Gerontol. A Biol. Sci. Med. Sci. 2005, 60, 324–333. [Google Scholar] [PubMed]
- Han, A.; Bokshan, S.L.; Marcaccio, S.E.; DePasse, J.M.; Daniels, A.H. Diagnostic criteria and clinical outcomes in sarcopenia research: A literature review. J. Clin. Med. 2018, 7, 70. [Google Scholar] [CrossRef] [PubMed]
- Goodpaster, B.H.; Thaete, F.L.; Kelley, D.E. Composition of skeletal muscle evaluated with computed tomography. Ann. N. Y. Acad. Sci. 2000, 904, 18–24. [Google Scholar] [CrossRef] [PubMed]
- Pillen, S.; Tak, R.O.; Zwarts, M.J.; Lammens, M.M.; Verrijp, K.N.; Arts, I.M.; van der Laak, J.A.; Hoogerbruge, P.M.; van Engelen, B.G.; Verrips, A. Skeletal muscle ultrasound: Correlation between fibrous tissue and echo intensity. Ultrasound Med. Biol. 2009, 35, 443–446. [Google Scholar] [CrossRef] [PubMed]
- Pillen, S.; van Dijk, J.P.; Weijers, G.; Raijmann, W.; de Korte, C.L.; Zwarts, M.J. Quantitative gray-scale analysis in skeletal muscle ultrasound: A comparison study of two ultrasound devices. Muscle Nerve 2009, 39, 781–786. [Google Scholar] [CrossRef]
- Arts, I.M.; Pillen, S.; Schelhaas, H.J.; Overeem, S.; Zwarts, M.J. Normal values for quantitative muscle ultrasonography in adults. Muscle Nerve 2010, 41, 32–41. [Google Scholar] [CrossRef]
- Parry, S.M.; El-Ansary, D.; Cartwright, M.S.; Sarwal, A.; Berney, S.; Koopman, R.; Annoni, R.; Puthucheary, Z.; Gordon, I.R.; Morris, P.E.; et al. Ultrasonography in the intensive care setting can be used to detect changes in the quality and quantity of muscle and is related muscle strength and function. J. Crit. Care 2015, 30, e9–e14. [Google Scholar] [CrossRef]
- Puthucheary, Z.A.; Phadke, R.; Rawal, J.; McPhail, M.J.; Sidhu, P.S.; Rowlerson, A.; Moxham, J.; Harridge, S.; Hart, N.; Montgomery, H.E. Qualitative ultrasound in acute critical illness muscle wasting. Crit. Care Med. 2015, 43, 1603–1611. [Google Scholar] [CrossRef]
- Fukumoto, Y.; Ikezoe, T.; Yamada, Y.; Tsukagoshi, R.; Nakamura, M.; Mori, N.; Kimura, M.; Ichihashi, N. Skeletal muscle quality assessed from echo intensity is associated with muscle strength of middle-aged and elderly persons. Eur. J. Appl. Physiol. 2012, 112, 1519–1525. [Google Scholar] [CrossRef] [PubMed]
- Mirón Mombiela, R.; Facal de Castro, F.; Moreno, P.; Borras, C. Ultrasonic echo intensity as a new noninvasive in vivo biomarker of frailty. J. Am. Geriatr. Soc. 2017, 65, 2685–2690. [Google Scholar] [CrossRef] [PubMed]
- Sanchez, B.; Rutkove, S.B. Electrical impedance myography and its applications in neuromuscular disorders. Neurotherapeutics 2017, 14, 107–118. [Google Scholar] [CrossRef]
- Shefner, J.M.; Rutkove, S.B.; Caress, J.B.; Benatar, M.; David, W.S.; Cartwright, M.C.; Macklin, E.A.; Bohoroquez, J.L. Reducing sample size requirements for future ALS clinical trials with a dedicated electrical impedance myography system. Amyotroph. Lateral Scler. Frontotemporal Degener. 2018, 28, 1–7. [Google Scholar] [CrossRef] [PubMed]
- Rutkove, S.B.; Kapur, K.; Zaidman, C.M.; Wu, J.S.; Paternak, A.; Madabusi, L.; Yim, S.; Pacheck, A.; Szelag, H.; Harrington, T.; et al. Electrical impedance myography for assessment of Duchenne muscular dystrophy. Ann. Neurol. 2017, 81, 622–632. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Zaidman, C.M.; Wang, L.L.; Connolly, A.M.; Florence, J.; Wong, B.L.; Parsons, J.A.; Apkon, S.; Goyal, N.; Williams, E.; Escolar, D.; et al. Electrical impedance myography in Duchenne muscular dystrophy and healthy controls: A multicenter study of reliability and validity. Muscle Nerve 2015, 52, 592–597. [Google Scholar] [CrossRef] [PubMed]
- Ishida, H.; Suehiro, T.; Suzuki, K.; Watanabe, S. Muscle thickness and echo intensity measurements of the rectus femoris muscle of healthy subjects: Intra and interrater reliability of transducer tilt during ultrasound. J. Bodyw. Mov. Ther. 2018, 22, 657–660. [Google Scholar] [CrossRef]
- Toledo, D.O.; Silva, D.C.L.E.; Santos, D.M.D.; Freitas, B.J.; Dib, R.; Cordioli, R.L.; Figueiredo, E.J.A.; Piovacari, S.M.F.; Silva, J.M., Jr. Bedside ultrasound is a practical measurement tool for assessing muscle mass. Revista Brasileira de Terapia Intensiva 2017, 29, 476–480. [Google Scholar] [CrossRef] [Green Version]
- Hadda, V.; Khilnani, G.C.; Kumar, R.; Dhungana, A.; Mittal, S.; Khan, M.A.; Madan, K.; Mohan, A.; Guleria, R. Intra- and inter-observer reliability of quadriceps muscle thickness measured with bedside ultrasonography by critical care physicians. Indian J. Crit. Care Med. 2017, 21, 448–452. [Google Scholar] [CrossRef]
- Thoirs, K.; English, C. Ultrasound measures of muscle thickness: Intra-examiner reliability and influence of body position. Clin. Physiol. Funct. Imaging 2009, 29, 440–446. [Google Scholar] [CrossRef]
- Tillquist, M.; Kutsogiannis, D.J.; Wischmeyer, P.E.; Kummerlen, C.; Leung, R.; Stollery, D.; Karvellas, C.K.; Preiser, J.C.; Bird, N.; Kozar, R.; et al. Bedside ultrasound is a practical and reliable measurement tool for assessing quadriceps muscle layer thickness. JPEN J. Parenter. Enteral Nutr. 2014, 38, 886–890. [Google Scholar] [CrossRef] [PubMed]
- Lopez, P.; Wilhelm, E.N.; Rech, A.; Minozzo, F.; Radaelli, R.; Pinto, R.S. Echo intensity independently predicts functionality in sedentary older men. Muscle Nerve 2017, 55, 9–15. [Google Scholar] [CrossRef] [PubMed]
- Rech, A.; Radaelli, R.; Goltz, F.R.; da Rosa, L.H.; Schneider, C.D.; Pinto, R.S. Echo intensity is negatively associated with functional capacity in older women. Age 2014, 36, 9708. [Google Scholar] [CrossRef]
- Aaron, R.; Esper, G.J.; Shiffman, C.A.; Bradonjic, K.; Lee, K.S.; Rutkove, S.B. Effects of age on muscle as measured by electrical impedance myography. Physiol. Meas. 2006, 27, 953–959. [Google Scholar] [CrossRef]
- Kortman, H.G.; Wilder, S.C.; Geisbush, T.R.; Narayanaswami, P.; Rutkove, S.B. Age- and gender-associated differences in electrical impedance values of skeletal muscle. Physiol. Meas. 2013, 34, 1611–1622. [Google Scholar] [CrossRef]
- McLester, C.N.; Dewitt, A.D.; Rooks, R.; McLester, J.R. An investigation of the accuracy and reliability of body composition assessed with a handheld electrical impedance myography device. Eur. J. Sport Sci. 2018, 18, 763–771. [Google Scholar] [CrossRef]
- Geisbush, T.R.; Visyak, N.; Madabusi, L.; Rutkove, S.B.; Darras, B.T. Inter-session reliability of electrical impedance myography in children in a clinical trial setting. Clin. Neurophysiol. 2015, 126, 1790–1796. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- O’Brien, T.G.; Cazares Gonzalez, M.L.; Ghosh, P.S.; Mandrekar, J.; Boon, A.J. Reliability of a novel ultrasound system for gray-scale analysis of muscle. Muscle Nerve 2017, 56, 408–412. [Google Scholar] [CrossRef]
Male/Female | 19 (70%)/8 (30%) |
Age (years) | 72.6 ± 5 (range 65–82 years) |
Height (cm) | 172.2 ± 11 (range 152–200 cm) |
Weight (kg) | 83.3 kg ± 19 (range 52–131 kg) |
BMI (kg/m2) | 28.1 |
Ultrasound Measurements | |
Fat thickness (mm) | 8.8 ± 7 (range 0.6–23.5) |
RF thickness (mm) | 13.7 ± 6 (range 1.2–23.5) |
VI thickness (mm) | 13.2 ± 6 (range 0.9–22.7) |
Echointensity RF | 94.16 ± 20 (range 68.8–167.5) |
VI | 62.6 ± 32 (range 10.3–121.2) |
Test | Intra-Rater | Inter-Rater | |
---|---|---|---|
Examiner 1 | Examiner 2 | ||
QMUS | |||
SQ thickness | 0.98 * | 0.99 * | 0.99 * |
RF thickness | 0.98 * | 0.98 * | 0.98 * |
VI thickness | 0.98 * | 0.99 * | 0.97 * |
RF echointensity | 0.89 * | 0.87 * | 0.94 * |
VI echointensity | 0.93 * | 0.93 * | 0.96 ** |
sEIM | |||
50 kHz R | 1.0 * | 1.0 * | 1.0 * |
50 kHz Xc | 1.0 * | 1.0 * | 0.99 * |
50 kHz θ | 1.0 * | 0.99 * | 0.99 * |
200 kHz R | 1.0 * | 1.0 * | 1.0 * |
200 kHz Xc | 1.0 * | 0.99 * | 0.99 * |
200 kHz θ | 0.97 * | 0.99 * | 0.98 * |
hEIM | |||
Fat % | 0.99 * | 0.99 * | 0.98 * |
Muscle Quality | 0.99 * | 0.99 * | 0.98 * |
Test | Isometric Normalized Peak Torque | Isokinetic Normalized Peak Torque | Timed Up and Go (TUG) |
---|---|---|---|
QMUS | |||
SQ thickness | −0.56, p = 0.002 | −0.50, p = 0.08 | --- |
RF thickness | --- | --- | −0.37, p = 0.06 |
VI thickness | --- | --- | --- |
RF echointensity | −0.48, p = 0.01 | --- | --- |
VI echointensity | --- | --- | --- |
sEIM | |||
50 kHz R | −0.46, p = 0.016 | --- | --- |
50 kHz Xc | --- | --- | --- |
50 kHz θ | −0.45, p = 0.02 | --- | 0.35, p = 0.07 |
200 kHz R | −0.57, p = 0.01 | −0.53, p = 0.01 | 0.45, p = 0.04 |
200 kHz Xc | --- | --- | --- |
200 kHz θ | −0.54, p = 0.01 | −0.51, p = 0.02 | 0.49, p = 0.02 |
50/200 kHz θ Ratio | --- | --- | --- |
200/50 kHz Phase Ratio | 0.50, p = 0.01 | 0.44, p = 0.027 | --- |
hEIM | |||
Fat % | −0.49, p = 0.009 | --- | --- |
Muscle Quality | --- | --- | --- |
DEXA | |||
Thigh muscle mass | --- | --- | --- |
Thigh fat mass | −0.52, p = 0.005 | −0.39, p = 0.04 | --- |
Test | DEXA-Measured Right Thigh Fat Mass | DEXA-Measured Right Thigh Muscle Mass |
---|---|---|
QMUS | ||
SQ thickness | 0.81, p < 0.0001 | --- |
RF thickness | --- | 0.53, p = 0.0045 |
VI thickness | --- | 0.54, p = 0.004 |
RF echointensity | 0.35, p = 0.07 | −0.33, p = 0.09 |
VI echointensity | --- | −0.52, p = 0.006 |
sEIM | ||
50 kHz R | 0.65, p = 0.0003 | --- |
50 kHz Xc | --- | --- |
50 kHz θ | 0.57, p = 0.002 | −0.37, p = 0.06 |
200 kHz R | 0.70, p < 0.0001 | −0.26, p = 0.06 |
200 kHz Xc | --- | −0.46, p = 0.016 |
200 kHz θ | 0.72, p < 0.001 | −0.34, p = 0.09 |
50/200 kHz θ Ratio | --- | --- |
200/50 kHz θ Ratio | −0.41, p = 0.041 | --- |
hEIM | ||
Fat % | 0.74, p < 0.0001 | −0.38, p = 0.0492 |
Muscle Quality | −0.72, p < 0.0001 | --- |
© 2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Hobson-Webb, L.D.; Zwelling, P.J.; Pifer, A.N.; Killelea, C.M.; Faherty, M.S.; Sell, T.C.; Pastva, A.M. Point of Care Quantitative Assessment of Muscle Health in Older Individuals: An Investigation of Quantitative Muscle Ultrasound and Electrical Impedance Myography Techniques. Geriatrics 2018, 3, 92. https://doi.org/10.3390/geriatrics3040092
Hobson-Webb LD, Zwelling PJ, Pifer AN, Killelea CM, Faherty MS, Sell TC, Pastva AM. Point of Care Quantitative Assessment of Muscle Health in Older Individuals: An Investigation of Quantitative Muscle Ultrasound and Electrical Impedance Myography Techniques. Geriatrics. 2018; 3(4):92. https://doi.org/10.3390/geriatrics3040092
Chicago/Turabian StyleHobson-Webb, Lisa D, Paul J Zwelling, Ashley N Pifer, Carrie M Killelea, Mallory S Faherty, Timothy C Sell, and Amy M Pastva. 2018. "Point of Care Quantitative Assessment of Muscle Health in Older Individuals: An Investigation of Quantitative Muscle Ultrasound and Electrical Impedance Myography Techniques" Geriatrics 3, no. 4: 92. https://doi.org/10.3390/geriatrics3040092
APA StyleHobson-Webb, L. D., Zwelling, P. J., Pifer, A. N., Killelea, C. M., Faherty, M. S., Sell, T. C., & Pastva, A. M. (2018). Point of Care Quantitative Assessment of Muscle Health in Older Individuals: An Investigation of Quantitative Muscle Ultrasound and Electrical Impedance Myography Techniques. Geriatrics, 3(4), 92. https://doi.org/10.3390/geriatrics3040092