Next Article in Journal
Performance Evaluation of Wearable Sensor Systems: A Case Study in Moderate-Scale Deployment in Hospital Environment
Previous Article in Journal
A Robust Reweighted L1-Minimization Imaging Algorithm for Passive Millimeter Wave SAIR in Near Field
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

A Novel Scheme and Evaluations on a Long-Term and Continuous Biosensor Platform Integrated with a Dental Implant Fixture and Its Prosthetic Abutment

1
Graduate Institute of Mechanical and Electrical Engineering, National Taipei University of Technology, Taipei 10608, Taiwan
2
Department of Mechanical Engineering, National Taipei University of Technology, Taipei 10608, Taiwan
*
Author to whom correspondence should be addressed.
Sensors 2015, 15(10), 24961-24976; https://doi.org/10.3390/s151024961
Submission received: 26 June 2015 / Revised: 4 September 2015 / Accepted: 10 September 2015 / Published: 25 September 2015
(This article belongs to the Section Biosensors)

Abstract

:
A miniature intra-oral dental implant system including a built-in biosensor device is proposed in this article. The dental implant system, or platform, is replaced over maxilla and allows relatively non-invasive procedures for a novel biosensing scheme for human blood analysis. Due to placement of the implant fixture, periodontal ligaments and the pulp structure, which are regarded as the main origin of pain, are thus removed, and long-term, continuous blood analysis and management through maxillary bone marrow becomes achievable through the dental implant platform. The new pathway of biological sensing is for the first time presented to realize an accurate and painless approach without injections. The dental implant system mainly consists of an implant fixture and a prosthetic abutment, a biosensor module, a bluetooth 4.0 wireless module and a dc button cell battery. The electrochemical biosensor possesses three electrodes, including working, reference and counter ones, which are arranged to pass through the titanium implant fixture below the biosensor module. The electrodes are exposed to the blood pool inside the maxillary bone marrow and perform oxidation/reduction reactions with the coating of biosensing enzyme. To prove the proposed platform, the immobilization process of glucose oxidase (GOD) enzyme and in vitro detections of glucose levels are successfully carried out, and proven sensitivity, linearity and repeatability of the glucose biosensor system are obtained. Moreover, a preliminary canine animal model adopting the new pathway shows significant consistency with the traditional method through dermal pricks for blood sugar detection. Despite the prospective results, further challenges in engineering implementation and clinical practices are addressed and discussed. In brief, the novel biosensing pathway and intra-oral biosensor platform may increasingly reveal their promising value and feasibilities in current bio-medical analysis, diagnosis, drug release and even healthcare technologies.

1. Introduction

Along with the increased average life expectancy, diagnostic management for chronic and critical diseases is becoming more important in the early twenty-first century. Clinical or home healthcare analysis from online monitoring data is thus considered indispensable for the further diagnosis and other medical procedures. In addition, the concept of long-term and continuous bio-supervision or bio-monitoring turns out to be more essential in modern medicine for both healthcare and diagnostic procedures than ever. These monitoring approaches are classified into two categories: invasive and non-invasive procedures. For instance, blood pressure (BP), respiration rate (RR), and heart rate (HR) measurements are categorized as non-invasive procedures, while most of blood analysis procedures are still considered invasive due to inevitable dermal or mucosal irritations. Blood analysis offers plenty of bio-chemistry data that would have great influence on differential diagnosis. For example, the blood urea nitrogen (BUN) and serum creatinine (Cr) value in renal function evaluation, The AST/ALT ratio (aspartate transaminase–alanine transaminase ratio or aspartate aminotransferase–alanine aminotransferase ratio) corresponding to the hepatic cell damage, and the cardiac marker such as Myoglobin, creatine kinase (CK), Cardiac Troponin I in acute myocardiac infraction (AMI) diagnosis, are all proven to be significant markers in diagnosis [1]. On the other hand, due to its invasive attribution, frequent and continuous blood monitoring is highly limited as dermal destruction process is pain-arising and only performed in critical or emergent patients with indwelling needles or catheters in vein and artery [2]. A well-known example of metabolic disorder is Diabetes Mellitus (DM), leading to relative renal and cardiovascular diseases in hundreds of millions of people. Due to high blood sugar value and relative ischemic status, patients will also remain in a high risk of infection. In the later stage, DM patients may be regularly arranged for four invasive blood sugar detection and subcutaneous injections of insulin to control blood sugar levels each day [3]. Such detection procedures may bring mental stress and physical discomfort at the same time. Currently, the most popular method for blood sugar detection is the dermal or mucosal irritations via an invasive penetration for blood sampling, and the biosensor operates mainly through glucose oxidation reaction using glucose oxidase (GOD) catalysis [4]. Despite the stress and irritations, however, recurrent blood analysis and monitoring is a routine practice in most medical procedures for patients in hospitals, especially for the elderly with chronic or other critical diseases. Recently, though there have been several non-invasive methods or techniques such as subcutaneous glucose monitoring developed to measure the bio-medical parameters, these methods are still immature and unable to be effective diagnostic tools for clinic applications [5]. As a result, it is very urgent and vital to find an alternative pathway to complete bio-medical analysis, and, in the meantime, minimize the discomfort during these invasive medical procedures so as to provide the patients with improved quality of life.
Dental implants have been widely accepted as an advanced dental procedure for tooth reconstructions since Dr. Brånemark first introduced the concept of osseointegration in dental implant in the 1970s [6]. Today this kind of surgical techniques in dental implantology along with the cylindrical pure titanium implant body develops into a standard medical procedure and broadly applied in clinical practice [7]. The implant surgery is mainly performed under local anesthesia and divided into two stages. In the first stage, the jawbone is exposed after flap design, and the titanium implant fixture is carefully placed into the created bony socket with initial stability. After a period of 3~6 months for mature bone healing, the cortical bone structure will closely contact the implant surface, which is referred as osseointegration. In the second stage, the surgery can be arranged to expose beneath implant fixture [8], and then the prosthetic abutment is jointed above the fixture. Finally, under proper impression and other relative dental procedures, the prosthetic crown or bridge is delivered over the abutment. In the whole dental procedure, the cylinder-type fixture is about 8~15 mm in length and 4~5 mm in diameter depending on the jawbone morphology; while the prosthetic crown in bilateral posterior molar area is in the average dimensions of 15 mm × 10 mm [9].
From the anatomic point of view, the origins of the tooth sensation are mainly from the pulp structures and the periodontal ligaments (PDL) [10]. With the artificial implant fixtures placement into jawbones, these paths of major painful sensation are both removed and this fact thus proposes a potential method for relatively painless blood analysis and monitoring techniques notwithstanding the risk of inflammation or peri-implantitis. Further on the anatomic structure of jawbones, due to the structural difference between maxilla (the upper jawbone) and mandible (the lower jawbone), the spongy and more chancellors bone structure on maxilla can offer the rich blood supply for diagnostic analysis and assure the biosensor’s electrodes to be inserted to the blood pool [11]. In addition, its chancellors structure may work as the cushion to lower down the discomfort from the surrounded pressure receptors. Furthermore, the intra-bony blood pool may also offer another prospective application in continuous and painless drug release [12].
Inspired by the above anatomic facts and currently mature intra-oral implant techniques, we herein present for the first time a novel concept of miniature bio-sensing platform embedded into the maxillary implant fixture for a long-term, continuous and painless blood analysis technique in this article. As shown in Figure 1, after removal of the pulp structure (A) and the periodontal ligament (B), the artificial titanium implant has a lack of pain receptors and will lead to painless status (C). The principal design of the bio-medical sensing platform as well as GOD enzyme coating and voltammetric calibration procedures of the biosensors are first described. Next, followed are the implementations of the transducing circuit for blood sugar detection and the wireless transmission measurement of the bluetooth 4.0 module. In addition to these engineering verifications, a proven canine experiment demonstrates efficient insulin release and successful detection of blood sugar levels. These preliminarily positive results provide us commercial and technical feasibilities for a new biosensing way in the near future. Finally, concluding remarks and the outlook for the proposed bio-medical platform are presented. More improvement works on the novel platform are currently under way in our laboratory.
Figure 1. The intra-oral environment over maxilla and the tooth structures. The pain origins are mainly from the pulp structures (A), and the periodontal ligament (B). After artificial implant restoration (C), they are both removed.
Figure 1. The intra-oral environment over maxilla and the tooth structures. The pain origins are mainly from the pulp structures (A), and the periodontal ligament (B). After artificial implant restoration (C), they are both removed.
Sensors 15 24961 g001

2. Design, Methods and Experiments

2.1. Design Structure of the Implant-Based Biosensor

The proposed platform is basically designed for various sensing arrangements for blood monitoring as long as they are small enough for installation. Figure 2 illustrates the unique design of the intra-oral implant-supported system. The implant platform mainly consists of a prosthetic abutment and a pure titanium implant fixture, which is directly connected to the maxillary bone structure. This prototype is 4.0 cm × 2.5 cm in size, which is about twice the target dimensions (2 cm × 1.2 cm) for preliminary engineering evaluation. The preliminary design of the electrochemical biosensor contains a set of tri-electrodes, a transducing circuit board along with button cells as power supply, and a wireless bluetooth 4.0 module to transfer the measured data to an external recipient server, such as cell phones, personal computers or other machines with micro control units (MCUs). All of the above biosensor components are inserted into the prosthetic abutment where the intra-oral electrochemical biosensor can prevent the occlusal loading of teeth. Besides, two individual canals or channels in the implant fixture 1 mm in diameter allow three sensing electrodes to pass through the surface of the implant fixture and immerse in the blood pool inside the maxillary bone marrow. A set of tri-electrodes includes working, reference and counter electrodes, which are isolated from each other to circumvent short circuit and current leakage.
The implant fixture and prosthetic abutment with the tri-electrodes and channels is required to place into the patient’s maxilla in the beginning. Three sensing electrodes in the abutment will extend out of the fixture afterward through the channels to expose themselves to the maxillary blood pool outside and then directly contact the surrounding bone marrow. Finally, in the post-sensing phase, wireless data transmission is very important in such biomedical sensor application. Due to the properties of low energy consumption, a higher transmission rate and a wider range, the bluetooth 4.0 module is adopted as our wireless communication unit [13], as it is able to obtain the complete set of data regarding analyte concentrations in blood measured from the biomedical sensing system.
Figure 2. The design of the intra-oral implant-supported biosensor over maxilla. (A) marks each components of the module, including a set of the electrodes; and (B) shows the prototype biosensor with electrodes exposed to outside blood pool as denoted.
Figure 2. The design of the intra-oral implant-supported biosensor over maxilla. (A) marks each components of the module, including a set of the electrodes; and (B) shows the prototype biosensor with electrodes exposed to outside blood pool as denoted.
Sensors 15 24961 g002

2.2. GOD Immobilization Procedure and Glucose Calibration

Diabetes Mellitus is a well-known metabolic disorder, which will lead to relative renal and cardiovascular diseases, and patients with such a disease will usually lead to regularly invasive procedures with both blood sugar monitoring and insulin delivery eventually, which can avoid relative renal failure and other cardiovascular complications. Thus, proper and regular management of blood sugar levels is rather important for DM control. By using the usual finger pricking approach, one of the common methods for monitoring the blood glucose levels is to apply glucose oxidation reactions with glucose oxidase (GOD) catalysis, and the chemical equations are as follows [4,14,15,16]:
βDGlucose +  O 2  +  H 2 O     glucose oxidase      DGluconic acid +  H 2 O 2
Basically, GOD is a dimer structural protein with molecular weight equals to 126,000. Thus it is important to immobilize GOD onto the electrode surface with proper chemically binding modifications, and these immobilization techniques may be different from materials or protocols in laboratories. The materials and protocols employed in this work are as follows:
Materials:
Pure gold electrode with 0.5 mm in diameter, Crosslinker agent as Stock 1, two kinds of coupling agent as Stocks 2 and 3 (store at −20 °C), GOD enzyme as Stock 4 (store at 4 °C), 100% Ethanol, and Phosphate buffer saline (PBS).
Protocol:
(1)
Gold electrode washing in distilled water and 100% ethanol.
(2)
Gold electrode immersing Stock 1 in the ethanol solution for 24 h.
(3)
Gold electrode washing in distilled water and 100% ethanol again.
(4)
Coupling agent preparation by mixing 10 μL each of Stocks 2 and 3.
(5)
Incubation for 1 hour after adding the coupling agent onto the gold electrode surface.
(6)
After PBS washing, add 20 μL of GOD enzyme solution as Stock 4, and then incubate at 4 °C for 2 h.
(7)
Then the enzyme-modified electrode is ready for cyclic voltammetric operation.
Finally, the glucose solutions in our experiments are prepared in 50 mg/dL, 100 mg/dL, 200 mg/dL, 300 mg/dL, 400 mg/dL, etc. to fit the measurement range for clinical applications of the biosensor system.

2.3. Design and Verification of the Transducing Circuit

Figure 3 shows a brief schematic of the electrochemical sensor measurement circuit in our experiments. Electrochemical sensors work by allowing fluid to diffuse into the sensor and interacting with the working electrode (WE). The sensor reference electrode (RE) provides feedback to maintain a constant potential with the WE terminal by varying the voltage at the counter electrode (CE). The direction of the current at the WE terminal depends on whether the reaction occurring is oxidation or reduction. In the case of carbon monoxide, oxidation takes place; therefore, the current flows into the working electrode, which requires the counter electrode to be at a negative voltage (typically 300 mV to 400 mV) with respect to the working electrode. In our experiment, the current into the WE terminal is less than 100 nA per ppm of the liquid concentration. Thus, converting this current into an output voltage requires a transimpedance amplifier with a very low input bias current. The amplifier (AD8500 Operational Amplifier Analog Devices, Cambridge, MA, United States) has CMOS inputs with maximal input bias current of 1 pA at room temperature, making this amplifier very appropriate for the application. Furthermore, the 2.5 V power source (LM385) establishes the pseudo-ground reference for the circuit, which allows for single-supply operation while consuming very little quiescent current.
Figure 3. The brief circuit diagram in the glucose sensing device with two amplifiers, the working (WE), counter (CE), and reference (RE) electrodes for glucose sensing.
Figure 3. The brief circuit diagram in the glucose sensing device with two amplifiers, the working (WE), counter (CE), and reference (RE) electrodes for glucose sensing.
Sensors 15 24961 g003
Amplifier sinks enough current from the CE terminal to maintain the zero potential voltage between the WE and RE terminals of the sensor. The RE terminal is connected to the inverting input, which stops any current flow passing through it. This results in all the current passing through the WE terminal, which leads the current to show a linear response to the contact glucose concentration. Thus, transimpedance amplifier converts the current through the working electrode into the corresponding voltage proportional to the glucose concentration in liquid.

2.4. Design and Manipulation of Transducing Circuit

The output voltage of the transimpedance amplifier is described below:
V 0 = 0.7 V + I W E × R F
where IWE is the current into the WE terminal, and RF is the transimpedance feedback resistor. The maximum response of the sensor is 58.8 nA/(mg/dL), and its maximum input range is 400 mg/dL of carbon monoxide. This results in a maximum output current of 23.5 μA, and a maximum output voltage determined by the transimpedance resistor, as shown in Equations (2) and (3):
V 0 = 0.7 V + 400 mg/dL × 58.8 nA mg/dL × R F
V 0 = 0.7 V + 23.5 μA × R F
By the circuit with a 3 V power supply will result in a usable range of 0.7~3 V at the output of the transimpedance amplifier. By selecting a 100 kΩ resistor for the transimpedance feedback resistor, we can get a maximal output voltage of 3 V, which allows for approximately 8% of over range capability. The resistor (RF) keeps the noise gain at a reasonable level. Selecting the value of this resistor is a compromise between the magnitude of the noise gain and the error of sensor settling time when exposed to high glucose concentrations. For this example, RF = 330 Ω, which results in a noise gain (NG) of 304, as shown in Equation (4):
N G = 1 + 100 ( R F ) 330 Ω 304
The input noise of the transimpedance amplifier appears at the output amplified by the noise gain. For this circuit, we are only interested in the low frequency noise because that from the sensor operation maintains at low level. In Equation (5), the AD8500 amplifier has an input voltage noise of 6 μVp-p from 0.1 to 10 Hz in frequency, which results in the output voltage of 1.82 mVp-p.
V o u t = 6 μV × N G = 1.82 mV p-p
It is hard to filter out the noise because it appears at very low frequency. However, the sensor response corresponding to the noise is also at the very low frequency level. Thus we choose a low-pass filter (resistor and capacitance) for very low frequency bandwidth and it is with a cutoff frequency of 0.16 Hz for the noise. Here it should be noted that, compared with the 30 s of response time for the sensor, the effect on the sensor response time is negligible even with such a low frequency filter.
Finally, one of the important characteristics among these electrochemical sensors should be their long response time compared with other types of sensors. Basically, after system power on, the biosensor may take 1~2 min to settle down for its final output value. When exposed to a specific concentration of the target liquid, it may require from 25 to 30 s for the sensor output to reach 90% of its final value. Thus the response time, on average, for the biosensor is estimated as 2 min. If the voltage between the RE and WE terminals has an abrupt change in magnitude, it may take several minutes for the sensor current output to settle.

2.5. The Bluetooth Module and Data-Logging Interface

The bluetooth 4.0 module shown in Figure 4A is Model CC2540 with 2.4 GHz bluetooth low energy system-on-chip produced by Texas Instruments Corp. Major specifications of the wireless module includes: a 128/512 Kb flash memory and 8 Kb SRAM inside, a 21 universal I/O interface, 2 USB 2.0 ports, an advanced 8 channel 8–12 bit analog to digital (AD) converter, low energy consumption and maximum +97 dB link budge [13]. The bluetooth 4.0 module is obviously smaller than the USB recipient device, as shown on the right hand side. The control program for data logging/transmission was implemented by Visual C# in Microsoft Visual Studio 2010. The complete interface shown in Figure 4B can record and display the transmitted digital data for glucose sensing procedures from the bluetooth 4.0 module.
Figure 4. (A) A bluetooth 4.0 module and its USB recipient device; (B) The interface of the bluetooth control panel completed by Visual C# and its function description.
Figure 4. (A) A bluetooth 4.0 module and its USB recipient device; (B) The interface of the bluetooth control panel completed by Visual C# and its function description.
Sensors 15 24961 g004

2.6. Animal Model Experiment

The one-year-old canine’s bilateral maxillary first premolar was extracted and prepared for dental implantation. Two of the 6 × 5 mm Bicon© titanium fixtures were implanted over bilateral maxilla on premolar area under barbiturate anesthesia. After 1 month of healing, the fixtures were exposed under local anesthesia, and the drug delivery syringe (DDS) was placed on maxillary bone marrow for insulin release on the right hand side. The integrated biosensor fixture will be placed on the implant fixture on the left hand side to detect the change of blood sugar levels.
Initially, before the insulin releasing procedure, the blood sugar level of the canine was measured twice to show the canine’s baseline value. Five units of the Neutral Protamine Hagedorn (NPH), also known as rapid onset insulin, was then loaded on the DDS and released into the maxilla. After insulin release, the blood sugar values were detected every 5 min by both the maxillary implant biosensor and an ACCU-CHEK® Performa blood sugar detector (Roche company, Basel, Switzerland), as shown in Figure 5, representing the preliminary canine model experiment. The ACCU-CHEK® detector is responsible for the control purpose of the subcutaneous blood sugar levels. Figure 5A shows the surgical wound after first titanium implant placement surgery, Figure 5B represents the proposed biosensor platform and the DDS to monitor blood sugar and deliver insulin, respectively. Finally, Figure 5C,D represents the blood glucose calibration operations over the canine’s maxilla and anterior limb by ACCU-CHEK® blood sugar detector.
Figure 5. The canine model for blood sugar monitoring. (A) The wound exposure after 1 month for further implant placement; (B) The DDS and biosensor module for insulin releasing and blood sugar monitoring through the maxillary pathway; (C) Blood sugar calibration via the maxillary pathway by ACCU-CHEK® blood sugar detector; (D) Blood sugar detection via the anterior limb.
Figure 5. The canine model for blood sugar monitoring. (A) The wound exposure after 1 month for further implant placement; (B) The DDS and biosensor module for insulin releasing and blood sugar monitoring through the maxillary pathway; (C) Blood sugar calibration via the maxillary pathway by ACCU-CHEK® blood sugar detector; (D) Blood sugar detection via the anterior limb.
Sensors 15 24961 g005

3. Results

The main results in our experiments can be divided into three parts. First of all, we clarify the relationship between logged bluetooth data and the output voltage of the biosensor, which corresponds to the resistance caused by different glucose concentrations. Next, we evaluate the quality of the GOD immobilization by cyclic voltammetry techniques, and investigate the proportional relationships between the glucose concentrations and the reactive currents. We also monitored solutions with different glucose concentration by the biosensor module, and analyzed bluetooth data. Finally, we perform the canine animal model for blood sugar monitoring to evaluate validity of the new detection pathway and repeatability of the biosensor module.

3.1. Resistance Modulation and Bluetooth Transmission Verification

Bluetooth module is responsible for signaling transfer in the biosensor. Thus calibrations at first are important. During cyclic voltammetry (CV) sweeping, the applied voltage is around −5 to +5 V, and the resistance changes may correspond to various glucose concentrations. Figure 6A shows voltage changes due to various input resistances over the working electrode in the biosensor circuit, while Figure 6B represents the relationship between the input resistances and bluetooth signal data. Both relationship patterns may not totally follow a linear response to each other, but they move linearly toward each other at intervals. This will allow for the evaluation of glucose concentrations. It should be noted that since the biosensor demonstrates a period of time delay in measurement, each of the transmitted measurement is required to last for one minute to confirm its validity, as shown in Figure 7. The bluetooth signal data from distilled water as a negative control is around 50 bluetooth units and can be regarded as background noise. On the other hand, e.g., the signal triggered from a 400 mg/dL glucose solution, exhibits a clear peak within the first 10 s, and gradually degrades to the background baseline beyond about 50 s. Therefore, the peak value will represent the bluetooth signal from a specific glucose concentration in solution.
Figure 6. The calibration process for the glucose sensor and bluetooth 4.0 module. (A) The relationship between transducing voltage vs. input resistance in the circuit, which may simulate different glucose concentrations in solutions; (B) The logged bluetooth signal vs. input resistance standing for different glucose solutions in the transducing circuit.
Figure 6. The calibration process for the glucose sensor and bluetooth 4.0 module. (A) The relationship between transducing voltage vs. input resistance in the circuit, which may simulate different glucose concentrations in solutions; (B) The logged bluetooth signal vs. input resistance standing for different glucose solutions in the transducing circuit.
Sensors 15 24961 g006
Figure 7. The temporal change of logged bluetooth signal data sets with respect to distilled water and a 400 mg/dL glucose solution.
Figure 7. The temporal change of logged bluetooth signal data sets with respect to distilled water and a 400 mg/dL glucose solution.
Sensors 15 24961 g007

3.2. Cyclic Voltammetry and Calibration of the Glucose Biosensor System

With a proper GOD immobilization procedure, the current through the working electrode shows highly positive correlations to glucose concentrations in cyclic voltammetric analysis. Figure 8A shows the typical cyclic voltammograms illustrating the responses among current, applied potential and specific solutions with different glucose concentrations due to electrochemical interactions. After rearrangement of the peak values of electrode currents, as shown in Figure 8B, it is observed that the sensor demonstrates approximately linear behavior between the glucose concentration and sensor current. The slope of the linearly fitted curve, equivalent to sensitivity of the basic biosensor, is found to be about 240 nA/(mg/dL). Hence, it can be inferred that a 4 mg/dL increase in the glucose concentration is able to boost the current higher for about 1 μA. The CV sweeping experiment is very helpful to evaluate the GOD immobilization effect over working electrodes and offer the calibration basis for further wireless signal translations in blood sugar monitoring.
Figure 8. Cyclic voltammograms and calibration plot of the intra-oral biosensor module. (A) The cyclic voltammetric sweep shows the electrochemical relationships among current and applied potential with respect to different glucose concentrations (0~500 mg/dL); (B) The sensitivity expression and calibration plot of the amperometric biosensor.
Figure 8. Cyclic voltammograms and calibration plot of the intra-oral biosensor module. (A) The cyclic voltammetric sweep shows the electrochemical relationships among current and applied potential with respect to different glucose concentrations (0~500 mg/dL); (B) The sensitivity expression and calibration plot of the amperometric biosensor.
Sensors 15 24961 g008

3.3. Repeatibility Test of the Glucose Biosensor Platform

To ensure the constant accuracy and repeatability of the biosensing system, it is vital to operate the biosensor at baseline point, 50, 100, 200, 300, and 400 mg/dL in glucose solutions with an interval of 30 min, as depicted in Figure 9. According to the transmitted bluetooth peak signals, the biosensor exhibits better repeatability at lower concentrations under 200 mg/dL than that at higher concentrations, and the errors of detection among different time-interval measurements are verified to remain less than 10%, even at a higher concentration of 400 mg/dL. Based on our completed experiments, it is estimated that the repeatability of the biosensor may last for about 8~10 h, until the immobilized GOD enzyme degrades in the end.
Figure 9. The repeatability test results of the glucose biosensor.
Figure 9. The repeatability test results of the glucose biosensor.
Sensors 15 24961 g009

3.4. Insulin Release and Blood Sugar Detection over Canine Model

After primary osteointegration for one month, the canine received five units of NPH loading throughout the DDS system over maxilla. The initial blood sugar level detected by our biosensor over canine’s maxillary implant was 112 mg/dL, and, for comparison, the control value measured by an ACCU-CHEK® blood sugar detector was about 118 mg/dL. The blood sugar detection was performed every 5 min subsequently after NPH loading. Figure 10 illustrates a brief temporal change of blood sugar levels with our biosensing implant system and an ACCU-CHEK® blood sugar detector. It is clearly shown that the blood glucose levels, after five minutes of initial NPH loading, was rapidly decreased in both biosensors due to the NPH loading effect. However, the glucose levels then rebounded from 18 mg/dL, and more promptly approached to the stable level about 80 mg/dL after about half an hour. Here the rebounding effect of blood sugar level for the canine may be justified as the physiological feedback due to the reactive glucogen secretion after a rapid decrease of blood sugar level.
Figure 10. The temporal variation of blood sugar levels measured by the proposed intra-oral implant system and ACCU-CHEK® blood sugar detector.
Figure 10. The temporal variation of blood sugar levels measured by the proposed intra-oral implant system and ACCU-CHEK® blood sugar detector.
Sensors 15 24961 g010

4. Discussions and Outlook

The purpose of the intra-oral implant-based biosensor platform is for long-term and continuous blood analysis and monitoring, and the resultant advantages of this sensor platform are as follows:
(1)
Compared to other implanted biosensors, the proposed sensor is fixed by surrounding bone tissue, which will reduce the risk of component loosening and internal bleeding problems.
(2)
The use of the implant fixture and prosthetic abutment will result in several benefits including biosensor module replacement without any additional surgery, regular sterilization at dental clinic in a non-invasive way, and the easy substitution of the system power supply with a dc button cell battery.
(3)
By mature dental implant procedures, further maintenance of the biosensing system may go along under apical X-ray films with minimized dosage. Other periodontal curettage may also be arranged to maintain the sensor in a workable condition if needed.
In this preliminary evaluation, we design the intra-oral biosensor prototype, which is about twice the size of a comparable unit currently used in practice for clinical applications, to reduce the existing difficulties such as specific design and manufacture for compact components, GOD enzyme immobilization on fine electrodes, and system assembly. On cyclic voltammetry and its resultant bluetooth signals, the current shows reasonable linearity with the glucose concentration. Thus every 2 mg/dL glucose concentration will increase the current to about 1 μA and correspond to 0.8 reading unit of bluetooth signal. However, the ideal linearity is evaluated in pure glucose solution in vitro, and the noise may become more significant when it applies to real blood due to its complicated components. Besides, as shown in Figure 9, the GOD enzyme exhibits good repeatability and accuracy in laboratory in pure glucose solutions. In spite of these characteristics, it is noted that GOD enzyme tends to degrade within 48 h and result in lower signal response with time. This issue thus becomes challenging if the biosensing platform is expected to be employed as a long-term application, e.g., valid for 30 days. Therefore, further improvement of GOD coating or the use of temporal calibration are needed for clinical practice. Moreover, the 5 V button cell battery may last for about one month under the operation frequency of continuous detection every 5 min, which satisfactorily meets the requirements of the system applications.
For further clinical applications, one should carefully deal with the dental procedures under infection control and other safety problems. Fortunately, mature dental implant procedures have provided hints of long-term care for the intra-oral biosensor platform. At first, a tight connection to the fixture and abutment will isolate the inner bone marrow structure from outside contaminations. This tight connection will also prevent significant blood loss during continuous blood monitoring. Further, regular dental appointments with the replacement of battery and biosensor module after proper H2O2 irrigation will also cut down the risk of infection. Moreover, the blood-rich marrow structure will also minimize the probability of thrombosis compared to other intra-venous analyzing procedures. To sum up, the intra-oral biosensor platform is required to fulfill the following facts or requirements:
(1)
Prevention from the occlusal loading forces to circumvent unexpected saliva micro-leakage, which may result in further contaminations, sensor damage, and chronic marrow infections.
(2)
GOD or other enzyme coating improvements maybe based on proper polymerization to maintain continuous sensor operation at least for one month.
(3)
The power supply scheme for the platform in both safety control and long-term reliability. In addition to the miniature battery solution, the emerging wireless charging technique may also be taken into consideration in the near future.
(4)
Further minimization of the biosensor module is extremely necessary as the smaller size it will be, the less discomfort and side effect it may cause.
(5)
More in vivo animal experiments are expected, as long-term histology evaluations will result in improvement of the safety requirements.
On limitations of the biosensor, it is noted that the proposed platform can be mainly suitable for patients with the requirements of regular invasive procedures. For example, type I diabetes mellitus (DM) patients need to receive the punches to monitor blood sugar four times a day, and patients receiving hemodialysis may face a similar situation. The proposed device may minimize patients’ sufferings when they face frequently invasive arrangements. For healthy and fully dentulous people, it seems unnecessary to place this class of bio-sensing device unless they are willing to do so for the reason of future health care. In addition, the intra-oral implant-based biosensor also includes the following restrictions:
(1)
The biosensor is specifically designed for some patient groups with the demands for long-term and regular blood monitoring, especially for the elderly with chronic and/or critical diseases. However, elderly people might mostly have edentulous ridge.
(2)
The fabricated cost of this intra-oral implanted device would be much more expensive compared to other biosensing devices or approaches.
(3)
With the more complicated biosensor, the risks of infection and peri-implant issues may increase, and the user thus needs more dental appointments for maintenance procedure, which may not be an economical process in some countries.
Finally, this intra-oral implant-supported analysis platform can be regarded as a kind of novel biosensor that can directly achieve accurate blood analyzing functions with minimal painful sensation. Without the need of dermal destruction or injection, various classes of developed electrochemical biosensors or those in development can be integrated with the proposed system. More importantly, this new approach could be exceedingly superior to the non-invasive blood analysis techniques such as optic or transdermal detection methods, particularly in sensitivity and accuracy. In this work, we select blood sugar as the analyte because it is the most often seen in daily life. Other feasible analytes in blood analysis items with this biosensor system may also be considered and customize in specific medical procedures. For example, continuous monitoring of cardiovascular markers toward the patients with high risk factors may decrease stroke possibilities or avoid delayed administrations. As modern medicine evolves, geriatrics has become much more important than ever before, so that the quality of life should also be taken into account. Since this work is a preliminary study, more advanced module designs, animal model tests, and clinical practice assessment are needed for system advances and reliability assessment.

5. Conclusions

We report a new class of an intra-oral biosensor platform embedded in a dental implant fixture and its prosthetic abutment, which may meet the needs of long-term, continuous and accurate blood analysis through a relatively painless pathway. The newly proposed method of biological sensing is presented for the first time to realize a non-invasive blood monitoring way in a painless manner without injections. As a novel and miniature biosensing platform, in addition to blood sugar levels, biochemistry data such as CK, Troponin I, and so on can be selected and developed by the use of replaceable modules. In this study, a number of works including design and assembly of the implant-based biosensor platform, implementation and communication of the transducing circuit and wireless bluetooth 4.0 module, biochemical synthesis of GOD enzyme and immobilization verification, performance tests of the biosensor platform, and a preliminary canine model experiment on blood sugar level detection are successfully carried out. In summary, these experimental results indicate very high feasibility of success for the proposed biosensor system. Nevertheless, the integrated biosensing platform needs constant efforts or improvements in system miniaturization, the infection control routine, advanced or smart control on automatic analysis, enhanced reliability, increased lifetime of the biosensor, and even business model assessment. More importantly, the realization of this new biosensing platform may have a significant influence on current medical diagnosis and therapy, especially for chronic diseases and elderly patients with critical diseases.

Acknowledgments

Special thanks to Yen-Ping Kuo and Associate Hao-Hueng Chang of Graduate Institute of Clinical Dentistry, National Taiwan University, for their clinical advice and canine experimental arrangements. Valuable assistance from Jung-Tang Huang and Bo-Wen Chang of National Taipei University of Technology is also acknowledged. This research received financial support under grants MOST 103-2221-E-027-017 and MOST 103-2218-E-027-012, Taiwan.

Author Contributions

Chih-Cheng Lu led the team and conducts the research work. Yu-Jung Li prepared the manuscript and contributed to the practical studies and experiments. Chih-Cheng Lu improved the quality of publication and completed revision.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Newby, L.K.; Goldmann, B.U.; Ohman, E.M. Troponin: an important prognostic marker and risk-stratification tool in non-ST-segment elevation acute coronary syndromes. J. Am. Coll. Cardiol. 2003, 41, 31–36. [Google Scholar] [CrossRef]
  2. David, C.; McGee, M.D.; Michael, K.G. Preventing complications of central venous catheterization. N. Engl. J. Med. 2003, 12, 1123–1133. [Google Scholar]
  3. Weng, J.; Li, Y.; Xu, W. Effect of intensive insulin therapy on β-cell function and glycaemic control in patients with newly diagnosed type 2 diabetes: A multicentre randomised parallel-group trial. Lancet 2008, 371, 1753–1760. [Google Scholar] [CrossRef]
  4. Li, D.; He, Q.; Cui, Y.; Duan, L.; Li, J. Immobilization of glucose oxidase onto nanoparticles with enhanced thermostability. Biochem. Biophys. Res.Commun. 2007, 355, 488–493. [Google Scholar] [CrossRef] [PubMed]
  5. Joseph, W. Glucose biosensors: 40 years of advances and challenges. Electroanalysis 2001, 13, 983–988. [Google Scholar]
  6. Yasuyuki, S.; Masaki, K.; Junichiro, T.; Tomoko, A.; Maho, M.; Masahiro, U.; Takahide, K. Analysis of 472 Brånemark system TiUnite implants: A retrospective study. Kobe J. Med. Sci. 2009, 55, 73–81. [Google Scholar]
  7. Kelly, M.; Hom-Lay, W. Implant surgery complications: Etiology and treatment. Implant Dent. 2008, 17, 159–168. [Google Scholar]
  8. Mavrogenis, A.F.; Dimitriou, R.; Parvizi, J.; Babis, G.C. Biology of implant osseointegration. J. Musculoskelet. Neuronal Interact. 2009, 9, 61–71. [Google Scholar] [PubMed]
  9. Mijiritsky, E.; Mazor, Z.; Lorean, A.; Levin, L. Implant diameter and length influence on survival: Interim results during the first 2 years of function of implants by a single manufacturer. Implant Dent. 2013, 22, 394–398. [Google Scholar] [CrossRef] [PubMed]
  10. Nusstein, J.M.; Reader, A.; Melissa, D. Local Anesthesia strategies for the patient with a “hot” tooth. Dent. Clin. North Am. 2010, 54, 237–247. [Google Scholar] [CrossRef] [PubMed]
  11. Tina, C.; Sanjay, V.G.; Ameet, V.R.; Abhay, K.J. Quantitative assessment of interradicular bone density in the maxilla and mandible: Implications in clinical orthodontics. Prog. Orthod. 2013, 14, 38. [Google Scholar]
  12. Lu, C.C.; Li, Y.J.; Tsai, W.L.; Chang, C.K. An Implantable and Painless Drug Delivery Device for Long-Term Operation. Taiwan Patent I488620, 21 June 2015. [Google Scholar]
  13. Sandeep, K.; Joakim, L. Measuring bluetooth® low energy power consumption. In Application Note AN092; Texas Instruments: Dallas, TX, USA, 2012. [Google Scholar]
  14. Gerd, W.; Svetlana, T.; Jasmina, Z.; Draginja, P.; Vladimir, L. The chemical mechanism of action of glucose oxidase from Aspergillus niger. Mol. Cell. Biochem. 2004, 260, 69–83. [Google Scholar]
  15. Syamasri, G.; Lana, M.; Moulik, S.P. Kinetics in microemulsion V. Glucose oxidase catalyzed oxidation of β-D-glucose in aqueous, micellar and water-in-oil microemulsion media. Indian J. Biochem. Biophys. 2003, 40, 340–349. [Google Scholar]
  16. Julio, R.; Horacio, A.M. Glucose oxidase as an analytical reagent. Crit. Rev. Anal. Chem. 1995, 25, 1–42. [Google Scholar]

Share and Cite

MDPI and ACS Style

Li, Y.-J.; Lu, C.-C. A Novel Scheme and Evaluations on a Long-Term and Continuous Biosensor Platform Integrated with a Dental Implant Fixture and Its Prosthetic Abutment. Sensors 2015, 15, 24961-24976. https://doi.org/10.3390/s151024961

AMA Style

Li Y-J, Lu C-C. A Novel Scheme and Evaluations on a Long-Term and Continuous Biosensor Platform Integrated with a Dental Implant Fixture and Its Prosthetic Abutment. Sensors. 2015; 15(10):24961-24976. https://doi.org/10.3390/s151024961

Chicago/Turabian Style

Li, Yu-Jung, and Chih-Cheng Lu. 2015. "A Novel Scheme and Evaluations on a Long-Term and Continuous Biosensor Platform Integrated with a Dental Implant Fixture and Its Prosthetic Abutment" Sensors 15, no. 10: 24961-24976. https://doi.org/10.3390/s151024961

Article Metrics

Back to TopTop