Next Article in Journal
Data-Driven Quantitation of Movement Abnormality after Stroke
Next Article in Special Issue
FAST (Flexible Acetylcholine Sensing Thread): Real-Time Detection of Acetylcholine with a Flexible Solid-Contact Potentiometric Sensor
Previous Article in Journal
An Inverse Method to Determine Mechanical Parameters of Porcine Vitreous Bodies Based on the Indentation Test
Previous Article in Special Issue
Multifunctional Biosensing Platform Based on Nickel-Modified Laser-Induced Graphene
 
 
Article
Peer-Review Record

Evaluation of Polymer-Coated Carbon Nanotube Flexible Microelectrodes for Biomedical Applications

Bioengineering 2023, 10(6), 647; https://doi.org/10.3390/bioengineering10060647
by Chethani Ruhunage 1, Vaishnavi Dhawan 2, Chaminda P. Nawarathne 1, Abdul Hoque 1, Xinyan Tracy Cui 2,* and Noe T. Alvarez 1,*
Reviewer 1:
Reviewer 2:
Bioengineering 2023, 10(6), 647; https://doi.org/10.3390/bioengineering10060647
Submission received: 18 April 2023 / Revised: 19 May 2023 / Accepted: 24 May 2023 / Published: 26 May 2023
(This article belongs to the Special Issue Recent Advances of Biosensors for Biomedical Applications)

Round 1

Reviewer 1 Report

The submitted paper (Biocompatible ...) can be viewed from two discrete experimental research angles: 1)the materials (HNBR-coated CNTs) and  2)their biocompatibility.

The first is wonderful, but the second is, actually, unfinished, and this part should be reconsidered after major revision. The main (experimental, biocompatibility) issue(s) arrive from the a)restricted electrochemical characterization and the b)inadequate implementation of (reduced time of 10s, only) the Cu-electrodeposition. The WE (Working Electrode of the electrochemical cells) is highly anisotropic (say, structures like wood) having dimensionalities' issues similar to the scientific  issues discussed in the  (bibliography of the) '70s about complex mechanisms of the  HOPG-intercalation compounds.

So, all about the electrochemical (and biocompatibility, Long-Term Stability, in Vitro Cytotoxicity of HNBR Coating) tests and effects are interpreted as semi-isotropic WE (and microelectrodes, HNBR-coated CNT fibers), are producing misleading 'results and discussion' in EIS (3.4.1. EIS Analysis), CV (Cyclic Voltammetry), the anisotropic Cu-electrodeposition (Figure S1), and the (anisotropic)  aging (3.4.3. Aging Studies of HNBR-Coated CNT Fiber Electrodes).

Also, the (EIS) impedance Z_amplitude spectra, only, is not creditable; so, it is strongly recommended to add the Z_phase spectra (or/and the Nyquist plot) inside Figure 5(A).

Author Response

 

 

Author Response File: Author Response.pdf

Reviewer 2 Report

In this paper, the morphology, chemical nature, and insulating properties of polymer-coated carbon nanotube fibers fabricated by the continuous dip coating method were investigated. In addition, aging studies were conducted at elevated temperatures to evaluate the stability. This paper is very interesting and meaningful for long-lasting implantable neural electrodes. I recommend that this paper is published after minor revisions according to a following comments.

Comments:

1.     (L295): (boiling point: -39.6 °C)---> (boiling point: 39.6 °C)

 2.     (Figure 6G and 6H): The authors need a description of Electrode 1-3, in particular, it should be explained that Electrode 3 has a large increase in impedance.

Author Response

Please see attachment

Author Response File: Author Response.pdf

Round 2

Reviewer 1 Report

The manuscript of this version (v2) is better. However, it is recommended to include the word 'studies' in their title. A proposal title is: Biocompatibility studies of Polymer-Coated Carbon Nanotube Fiber Flexible
Microelectrodes.

Also, the misleading (and unsupported) value of the 'specific impedance' must be deleted, at least, from the Abstract (Line 023).

Author Response

Thanks the reviewer for this valuable suggestions. Our replies are in black to reviewer's questions in green:

it is recommended to include the word 'studies' in their title. A proposal title is: Biocompatibility studies of Polymer-Coated Carbon Nanotube Fiber Flexible

The title has been changed as suggested

Also, the misleading (and unsupported) value of the 'specific impedance' must be deleted, at least, from the Abstract (Line 023).

The specific impedance is calculated at 1.0 kHz and corresponds to measurements in Fig 5A. The details about the calculation are provided in section 3.4.1. Therefore we have clarified this information in the Abstract so we can maintain the reported impedance values

Back to TopTop