Reviewer 1: Anonymous Reviewer 2: Anonymous

Round 1

Reviewer 1 Report

The authors  Yang Yang  et al wanted to know in how far the selection of respective material and the optimization of the side holes of catheters could be of importance  for hemodynamic properties and  stimulation of coagulation. The paper is technically thorough giving an abundance of detailed information on the topic of the authors’ work.

I have only a few remarks:

(1)    Some sort of statement or hypothesis should be given at the end of the introduction: What did the authors test or wanted to know specifically in their project ?

(2)    The description of the methods is enriched with a lot of details and technical information. The authors should somewhere give a short overview of this section.

(3)    Recirculation test bench:  As far as I understand the setup is not used for an in vivo evaluation. The catheters are still not tested in patients. Furthermore, the description of the test bench should be integrated into methods.

(4)    The authors should start the discussion with a short comprehensive statement on the major findings, the consequences of these findings and – if possible - the impact that these findings could have in everyday dialysis care.

Author Response

Please see the attachment

Author Response File: Author Response.pdf

Reviewer 2 Report

 

The Methods section does not describe how experiments were done.

 

How many experiments were performed? Please provide statistical analysis.

 

Figure legends need more descriptions.

 

Fig. 10: How was RR(%) calculated?

 

I presume “4. In-vivo experiment for recirculation” should be “4. In-vitro experiment for recirculation”.

 

The Abstract section needs to include more specific descriptions of what were found in this study. For example, just by reading the Abstract, readers would not know what “The Lower PLI was seen with the GDK3 and GDK4 catheters.” Since the abstract does not describe what GDK3 and GDK4 are.

Author Response

please see the attachment

Author Response File: Author Response.pdf

Round 2

Reviewer 2 Report

Please provide means +/- SEM values in the results, perform appropriate statistical analysis, and then make conclusions.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Round 3

Reviewer 2 Report

It would be easier for the readers to see if the authors indicate SEM values on the bars of Fig. 9 with appropriate notations of significant differences.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf