Efficacy and Safety of 6.3 Fr Versus 7.5 Fr Single-Use Flexible Ureteroscopes for Upper Urinary Tract Stones: A Systematic Review and Meta-Analysis of Randomized Controlled Trials

,
,
,
,
,
and
1
Faculty of Medicine, Arabian Gulf University, Manama 329, Bahrain
2
Kuwait Institute for Medical Specializations, Kuwait City 12050, Kuwait
3
Department of Surgery, Al Amiri Hospital, Kuwait City 12050, Kuwait
4
Faculty of Medicine, Jordan University of Science and Technology, Irbid 22110, Jordan
This article belongs to the Section Urology & Nephrology

Abstract

Background and Objectives: Retrograde intrarenal surgery (RIRS) is a cornerstone in managing upper urinary tract stones, with a growing trend towards instrument miniaturization. The introduction of the ultra-slim 6.3 Fr single-use flexible ureteroscope presents a potential advancement over the standard 7.5 Fr device, but clinical evidence remains scarce. This systematic review and meta-analysis aims to synthesize data from randomized controlled trials (RCTs) to compare the efficacy and safety of the 6.3 Fr versus the 7.5 Fr ureteroscope. Materials and Methods: A systematic search of PubMed, Scopus, CENTRAL, and Web of Science was conducted for RCTs published up to September 2025. The primary outcomes were the stone-free rate and procedural success rate. Secondary outcomes included operation duration and postoperative complications. Risk ratios (RRs) for dichotomous data and mean differences (MDs) for continuous data, with 95% confidence intervals (CIs), were pooled for analysis. Results: Three RCTs involving 140 patients were included. There was no significant difference between both groups in stone-free rates (RR: 1.06, 95% CI [0.96, 1.18], p = 0.22) or success rates (RR: 1.06, 95% CI [0.97, 1.16], p = 0.17). However, the 6.3 Fr ureteroscope was associated with a significantly shorter operation duration (MD: −6.66 min, 95% CI [−11.29, −2.03], p < 0.001). No significant differences were found in laser operating time (MD: −1.46 min, 95% CI [−3.93, 1.01], p = 0.25), length of hospital stay (MD: −0.09 days, 95% CI [−0.23, 0.05], p = 0.19), or postoperative complications (Clavien I: RR 0.86, 95% CI [0.30, 2.43], p = 0.77; Clavien II–III: RR 0.67, 95% CI [0.12, 3.84], p = 0.65). Conclusions: Based on low-certainty evidence, the 6.3 Fr ureteroscope does not significantly improve stone-free rates but may reduce overall operation duration compared to the 7.5 Fr scope, with a comparable safety profile. These findings are limited by the small number of available studies, highlighting a clear need for larger, high-quality RCTs to confirm these preliminary results.

1. Introduction

Retrograde intrarenal surgery (RIRS) is the cornerstone of minimally invasive management for upper urinary tract stones []. Since the first fiber-optic ureteroscope was developed in the late 1980s [], the field has undergone significant technological advancements. Digital, disposable devices have increasingly replaced older fiber-optic instruments, though reusable scopes remain widely utilized depending on economic and institutional factors []. This key update in procedure has been facilitated by multiple benefits, including the eradication of cross-contamination risk, the elimination of expensive maintenance and sterilization expenses, and the assurance of stable, top-tier performance with every new device []. This progress sets the stage for today’s cutting-edge technology and the persistent pursuit of improvement.
A rising trend in endourology is the gradual reduction in size of surgical instruments []. This trend is clinically justified because smaller ureteroscopes are meant to decrease ureteral injury and improve navigation through constricted or unstented ureters, and may help to reduce postoperative complications []. Despite these significant benefits, their extent can be influenced by several factors such as surgeon experience []. Additionally, while usage trends may vary by institution and region, the 7.5 Fr flexible ureteroscope remains a widely used standard, offering reliable performance in a smaller size. Still, introducing the ultra-slim 6.3 Fr single-use flexible ureteroscope represents the latest advancement []. This development presents a crucial clinical question: Does this considerable reduction in diameter lead to observable improvements in patient outcomes, while maintaining essential aspects, such as image quality, deflection, and long-term reliability?
Although some early randomized controlled trials (RCTs) comparing the new 6.3 Fr scope to the standard 7.5 Fr model have been released, the data are still scarce, warranting further evidence synthesis to guide clinical practice [,,]. Therefore, the primary objective of this study is to conduct the first systematic review and meta-analysis of RCTs to rigorously compare the 6.3 Fr and 7.5 Fr single-use flexible ureteroscopes. We will assess efficacy, primarily measured by stone-free rates, and safety, measured by postoperative complication rates.

2. Materials and Methods

2.1. Protocol Registration

Registration of this systematic review was pre-completed through the (CRD420251152225) in the International Prospective Register of Systematic Reviews (PROSPERO). The procedures for this systematic review and meta-analysis complied with the PRISMA guidelines [] and the Cochrane Handbook for Systematic Reviews of Interventions [].

2.2. Data Sources & Search Strategy

On 16 September 2025, a literature search was systematically conducted by (A. M. H.) across the following electronic databases: PubMed, Scopus, CENTRAL, and Web of Science. The search strategy utilized a string of the following keywords: (ureteroscopy OR ureteroscop* OR RIRS OR “retrograde intrarenal surgery” OR “flexible ureteroscope*”) AND (“6.3 Fr” OR “6.3 Fr” OR “7.5 Fr” OR “7.5 Fr” OR “ultra-slim” OR “ultraslim” OR “small diameter”). The search terms and database results are comprehensively summarized in (Table S1). No language restrictions or other search filters were used. We also manually reviewed the reference sections of eligible trials to ensure comprehensive coverage and to avoid overlooking any relevant studies.

2.3. Eligibility Criteria

RCTs were included if they followed the following Population, Intervention, Control, and Outcome (PICO) framework:
  • Population (P): adult patients with upper urinary tract stones (kidney or proximal ureteral) of small-to-moderate size (≤2 cm) who are candidates for flexible ureteroscopy.
  • Intervention (I): RIRS performed using a 6.3 Fr single-use flexible ureteroscope.
  • Control (C): RIRS performed using a 7.5 Fr single-use flexible ureteroscope.
  • Outcomes (O): The primary outcome was the stone-free rate (as defined by the individual trial protocols) and success rate, defined as the successful completion of the planned stone clearance, including the ability to advance the ureteroscope to the target stone and not converting to open surgery. Secondary outcomes included operation duration, laser operating time, length of hospital stay (LoS), and Clavien complications.
The following criteria were used to exclude studies: quasi-randomized studies, conference abstracts or proceedings, study protocols, and observational studies or reviews.

2.4. Study Selection

Using the Covidence online platform, two reviewers (S.A.A. and A.A.H.) independently assessed the eligibility of the retrieved records. Once the duplicates were removed automatically, the remaining unique articles were screened in two stages. We screened titles and abstracts, then assessed the full text of potentially eligible studies. Conflicts between reviewers were resolved through discussion, resulting in a consensus.

2.5. Data Extraction

The data extraction was conducted independently by two reviewers (M. S. A. and S. H.). Discrepancies were resolved through discussion and in consultation with a senior author. The data extraction process involved creating an Excel spreadsheet, which was tested in a pilot phase before the final extraction. The extraction form was categorized into three primary sections:
Study characteristics: study ID, country, study design, total number of patients, intervention group details, control group details, main inclusion criteria, primary outcome, and follow-up duration.
  • Participant baseline characteristics: age, body mass index (BMI), gender, stone size, stone volume, stone number, hydronephrosis, and pre-procedural stenting.
  • Outcome data: stone-free rate, success rate, operation duration, laser operating time, LoS, and Clavien complications.

2.6. Risk of Bias and Certainty of Evidence

The revised Cochrane Collaboration’s Risk of Bias tool (ROB 2) [] was employed to assess the methodological quality and bias risk for each RCT. Two reviewers (S. H. and L. T. A) independently evaluated each study across domains such as selection bias, performance bias, reporting bias, and attrition bias. Disagreements were resolved by consensus. Additionally, the overall certainty of the evidence was assessed using the Grading of Recommendations Assessment, Development, and Evaluation (GRADE) approach [,]. This framework considers risk of bias, inconsistency, indirectness, imprecision, and publication bias. Each factor was carefully assessed, and the rationale for each judgment was documented, with any discrepancies resolved through discussion.

2.7. Statistical Analysis

Statistical analyses were conducted using Stata MP version 17 (Stata Corp., College Station, TX, USA). The risk ratio (RR) was computed for dichotomous outcomes, and the mean difference (MD) was calculated for continuous outcomes, both presented alongside the corresponding 95% confidence intervals (CI). Heterogeneity was evaluated using the chi-squared (χ2) test and the I2 statistic. A p-value less than 0.1 for the χ2 test or an I2 value of 50% or higher indicated significant heterogeneity. The default analysis model was fixed-effects due to the limited number of included trials, which limits the precision of between-study variance estimation in random-effects models. A random-effects model was applied only if significant heterogeneity was observed, defined as a Chi-squared p < 0.1 or an I2 > 50%. An assessment for publication bias was not performed, as all analyzed outcomes included fewer than 10 RCTs [].

3. Results

3.1. Search Results and Study Selection

A total of 380 records were retrieved from the initial literature search. Following the automated exclusion of 273 irrelevant records, the titles and abstracts of the remaining 107 articles underwent screening. Consequently, 99 studies were excluded for failing to satisfy the inclusion criteria. Thus, eight articles were evaluated for eligibility by screening their full texts. Five studies were excluded for various reasons (Table S2). Ultimately, three RCTs [,,] were included in the qualitative and quantitative synthesis (Figure 1).
Figure 1. PRISMA flow chart of the screening process.

3.2. Characteristics of Included Studies

This review included three RCTs, with a total of 140 patients [,,]. The included population consisted of adult patients with small-to-moderate upper urinary tract stones, with maximum size criteria ranging from <1.5 cm to ≤2 cm. Some differences in the procedure methodology were noted in the use of a ureteral access sheath (UAS): one trial used a sheathless technique for the intervention group [], another used a suction UAS for all subjects [], and the third utilized a standard UAS selectively in 47% of the cases []. Further key details about the study designs are summarized in (Table 1). Also, information about the included patients’ baseline data is summarized in (Table 2).
Table 1. Summary characteristics of the included RCTs.
Table 2. Baseline characteristics of the participants.

3.3. Risk of Bias and Certainty of Evidence

Ding et al. showed an overall low risk of bias [], while the other two trials showed some concerns [,] (Figure 2). Geavelte et al. showed some concerns about selection bias due to the absence of details of the randomization procedure []. Krajewski et al. showed some concerns of detection bias as the stone-free rate was assessed intraoperatively and subjectively by the unblinded operating surgeon []. Additionally, the certainty of evidence assessment is outlined in (Table 3).
Figure 2. Quality assessment of risk of bias in the included trials. The upper panel presents a schematic representation of risks (low = green, unclear = yellow) for specific types of biases of the studies in the review. The lower panel presents risks (low = green, unclear = yellow, and high = red) for the subtypes of biases of the combination of studies included in this review [, , ].
Table 3. GRADE evidence profile.

3.4. Primary Outcomes

There was no significant difference between the two ureteroscope sizes regarding the incidence of stone-free rate, which was assessed either intraoperatively or at one-month post-surgery, (RR: 1.06, 95% CI [0.96, 1.18], p = 0.22) (Figure 3A) or procedure success rate (RR: 1.06, 95% CI [0.97, 1.16], p = 0.17) (Figure 3B). Pooled studies were homogeneous in both stone-free rate (I2 = 0%) and procedure success rate (I2 = 12.11%).
Figure 3. Forest plots of the primary outcomes, CI: confidence interval [,,].

3.5. Secondary Outcomes

3.5.1. Secondary Procedural Outcomes

The 6.3 Fr ureteroscope was associated with a significantly shorter operation duration (MD: −6.66 min, 95% CI [−11.29, −2.03], p < 0.001) (Figure 4A); however, there was no significant difference between both groups in laser operating time (MD: −1.46 min, 95% CI [−3.93, 1.01], p = 0.25) (Figure 4B) or LoS (MD: −0.09 days, 95% CI [−0.23, 0.05], p = 0.19) (Figure 4C). Pooled studies were homogeneous in operation duration (I2 = 0%), laser operating time (I2 = 0%), and LoS (I2 = 0%).
Figure 4. Forest plots of the secondary procedural outcomes, CI: confidence interval [,,].

3.5.2. Secondary Safety Outcomes

There was no significant difference between the two strategies regarding the incidence of Clavien I complications (RR: 0.86, 95% CI [0.30, 2.43], p = 0.77) (Figure 5A), Clavien II–III complications (RR: 0.67, 95% CI [0.12, 3.84], p = 0.65) (Figure 5B), or Clavien IV–V complications (RR: 1.00, 95% CI [0.11, 9.37], p = 1.00) (Figure 5C). Pooled studies were homogeneous in Clavien I complications (I2 = 0%), Clavien II–III complications (I2 = 0%), and Clavien IV–V complications (I2 = 0%).
Figure 5. Forest plots of the safety outcomes, CI: confidence interval [,,].

4. Discussion

After synthesizing three RCTs and 140 patients, there was no difference in 6.3 Fr and 7.5 Fr ureteroscope sizes in terms of stone-free rate, procedural success rate, laser operating time, LoS, and complications. Still, patients treated with a 6.3 Fr ureteroscope showed a significantly shorter operation duration compared to those treated with a 7.5 Fr ureteroscope. Additionally, the evidence remains uncertain due to the limited number of trials, underscoring the need for further investigation.
In current clinical practice, the selection of ureteroscope sizes often depends on the surgeon’s experience, patient-specific ureteral conditions, and stone characteristics []. However, robust evidence-based guidance to inform this decision is lacking, which highlights the need for a comprehensive evidence synthesis. Accordingly, our primary efficacy outcome, the stone-free rate, showed no statistically significant difference between the two ureteroscopes. This result is difficult to interpret, mainly because the studies used different ways to define and measure SFR, which our GRADE assessment considered a “serious indirectness”. Stone-free status was evaluated in two trials using a postoperative non-contrast computed tomography scan, which is the most reliable method [,]. Conversely, one study utilized a subjective, unblinded intraoperative visual evaluation conducted by the operating surgeon, which is highly susceptible to detection bias, as numerous studies have demonstrated that surgeons’ intraoperative estimates of stone clearance are often overly optimistic and correlate poorly with postoperative imaging [,].
Despite this critical variance in stone-free rate evaluation, statistical heterogeneity was not observed, likely reflecting the meta-analysis’s low statistical power to detect heterogeneity with only three small studies, rather than a truly consistent treatment effect []. Still, the observed positive trend in SFR with the 6.3 Fr scope could theoretically be attributed to improved irrigation dynamics, which was supported by previous preclinical studies [,,]. When used with a UAS, the smaller scope results in a larger sheath-to-scope ratio. This may improve irrigation outflow, lower intrarenal pressure, and potentially facilitate more efficient passive evacuation of stone dust and small fragments, contributing to improved overall clearance [,,].
Moreover, the 6.3 Fr ureteroscope showed a statistically significant advantage, reducing the mean operation duration by 6.66 min, though the evidence supporting this is of low certainty. This effect was primarily driven by Ding et al., who had the greatest impact on the pooled analysis. They utilized a “sheathless” RIRS for the 6.3 Fr group, while the 7.5 Fr group was treated with a conventional UAS approach. The reduction in operation time in the 6.3 Fr arm can be attributed to the complete elimination of the UAS placement step. This step was time-consuming and caused three patients in the 7.5 Fr cohort to experience access failure, thereby extending the procedure duration []. However, other potential confounders, such as surgeon experience [], stone density [], and preoperative stenting [] may also influence procedural duration and were not uniformly controlled across all trials.
Beyond the sheathless approach, the availability of the 6.3 Fr ureteroscope opens new possibilities for sheath selection. Although the included studies utilized standard 10/12 Fr sheaths, the ultra-slim scope can theoretically permit the use of smaller access sheaths (9.5/11.5 Fr). While not tested in the current analysis, future research could investigate if this combination mitigates the risk of ureteral injury [,], as this method may reduce the risk of ureteral injury during sheath insertion, offering a substantial advantage for patients with narrow ureters []. However, this strategy presents a trade-off, as employing a smaller sheath could diminish the gap between the scope and the sheath, which might impede irrigation outflow and obstruct the passive elimination of stone fragments, potentially compromising the effectiveness of stone clearance []. Geavlete et al. highlighted this notion; using the 6.3 Fr scope within a standard 10/12 Fr sheath created a more favorable scope-to-sheath ratio (63%) compared to a 7.5 Fr scope (75%), thereby improving the available working space for irrigation and fragment removal [].
Accordingly, the reduced diameter of the ultra-slim scope offers a notable operational benefit, allowing for a more practical sheathless approach, especially in cases of unstented or narrowed ureters, which subsequently facilitates the procedure and reduces the operation duration []. Still, the non-significant findings for laser operating time and LoS are justified. Laser time is primarily dependent on stone characteristics, such as volume and density, rather than scope diameter []. At the same time, LoS for uncomplicated RIRS is already minimal, leaving little room for a clinically meaningful or measurable difference between the groups [,].
Furthermore, the safety profiles of both ureteroscopes were excellent, with no severe complications (Clavien-Dindo grade IV–V) reported in any of the 140 patients across the three trials. Still, our pooled analysis revealed a non-significant trend toward fewer minor complications with the 6.3 Fr scope, including Clavien-Dindo grade I (RR 0.86) and grades II–III (RR 0.67) events. This is consistent with the core clinical justification for using smaller instruments in endourology: ideally, a smaller instrument will place less mechanical strain on the ureteral wall, resulting in less injury to the lining, fewer complications after surgery, and a decreased risk of long-term issues, such as ureteral strictures [,]. However, the analysis lacked sufficient power to identify a real difference in safety results.

Strengths & Limitations

To our knowledge, this meta-analysis is the first to address this specific clinical question. It was conducted according to rigorous PRISMA and Cochrane methodologies, including only RCTs to minimize selection bias, and providing a transparent, formal GRADE assessment of the certainty of the evidence for each outcome.
Despite these strengths, the conclusions are limited by a few limitations. First, there is a low number of included trials, with only three small trials and 140 patients. Hence, the analysis lacks the statistical power to provide precise effect estimates for most outcomes, particularly for rare events like complications. This imprecision is a primary reason for downgrading the certainty of evidence. Second, the analysis was challenged by clinical heterogeneity, which comprises the inconsistent use of a UAS, as previously discussed. To clarify, this issue is most prominent in the analysis of operation duration, where the pooled result was mainly driven by the Ding et al. study [], which accounts for half of the patients in this meta-analysis. In that trial, the 6.3 Fr group underwent a sheathless procedure while the 7.5 Fr group did not, meaning the comparison was between two different surgical techniques, not just two instruments. Third, the variable definitions and assessment methods for the stone-free rate, as previously discussed, introduce serious methodological heterogeneity and a high risk of detection bias in one trial.
Fourth, another significant source of heterogeneity was the inconsistent use of preoperative stenting. The included trials reported different rates: 100% of patients were pre-stented in the study by Geavlete et al. [], 56.7% in the Krajewski et al. trial [], and none in the Ding et al. study [], where pre-stenting was an exclusion criterion. Preoperative stenting is known to passively dilate the ureter, facilitating easier instrument access and potentially reducing procedural difficulty; this represents a critical confounding factor []. This difference probably affected some outcomes, such as procedural success, as the only instances of access failure noted in our analysis occurred in the non-stented 7.5 Fr cohort reported by Ding et al. []. Finally, the overall certainty of the evidence ranged from low to very low, reflecting the inherent limitations that warrant careful clinical interpretation of our results.

5. Conclusions

The 6.3 Fr ureteroscope did not significantly enhance clinical outcomes, including the primary outcome of stone-free rate, which was assessed at one month postoperatively or intraoperatively. Additionally, the 6.3 Fr ureteroscope significantly reduced the operation duration and demonstrated a favorable safety profile. However, these results are based on an overall low to very low certainty of evidence according to the GRADE assessment, reflecting the limited number of trials and their methodological inconsistencies. Therefore, current evidence is insufficient to recommend a routine shift to 6.3 Fr ureteroscopes in clinical practice. Further high-quality, multicenter RCTs are required to confirm these preliminary findings.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/medicina61122103/s1, Table S1: Search Strategy; Table S2: Excluded records in full-text screening.

Author Contributions

Conceptualization, A.M.A. and S.A.A.; methodology, A.M.A., A.A.H. and M.S.A.; software, S.H. and H.A.; validation, L.T.A. and H.A.; formal analysis, A.A.H. and M.S.A.; investigation, S.A.A., L.T.A. and H.A.; resources, M.S.A. and S.H.; data curation, A.M.A. and A.A.H.; writing—original draft preparation, A.M.A. and S.A.A.; writing—review and editing, A.A.H., M.S.A., S.H., L.T.A. and H.A.; visualization, L.T.A. and H.A.; supervision, A.M.A.; project administration, A.M.A. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Inoue, T.; Okada, S.; Hamamoto, S.; Fujisawa, M. Retrograde intrarenal surgery: Past, present, and future. Investig. Clin. Urol. 2021, 62, 121–135. [Google Scholar] [CrossRef] [PubMed]
  2. Bagley, D.H.; Huffman, J.L.; Lyon, E.S. Flexible ureteropyeloscopy: Diagnosis and treatment in the upper urinary tract. J. Urol. 1987, 138, 280–285. [Google Scholar] [CrossRef]
  3. Şener, T.E.; Pietropaolo, A.; Talso, M.; Somani, B.; Tanıdır, Y.; Young Academic Urologists (YAU) “Endourology and Urolithiasis” Working Party. An Overview of the Advantages of Digital Flexible Ureteroscopes. A Review by Young Academic Urologists Endourology and Urolithiasis Working Party of the European Association of Urology. J. Urol. Surg. 2020, 7, 76–82. [Google Scholar] [CrossRef]
  4. Huang, F.; Zhang, X.; Cui, Y.; Zhu, Z.; Li, Y.; Chen, J.; Zeng, F.; Li, Y.; Chen, Z.; Chen, H. Single-Use vs. Reusable Digital Flexible Ureteroscope to Treat Upper Urinary Calculi: A Propensity-Score Matching Analysis. Front. Surg. 2022, 8, 778157. [Google Scholar] [CrossRef]
  5. Talyshinskii, A.; Hameed, B.M.Z.; Naik, N.; Guliev, B.; Zhanbyrbekuly, U.; Khairley, G.; Juliebø-Jones, P.; Somani, B.K. Miniaturization of flexible ureteroscopes: A comparative trend analysis of 59 flexible ureteroscopes. Urolithiasis 2024, 52, 16. [Google Scholar] [CrossRef] [PubMed]
  6. Da Silva Ferreira, D.; Bhattu, A.S.; Adam, A. Miniaturizing the approach to upper tract renal calculi: Is smaller always better? A narrative review. Curr. Urol. 2023, 17, 280–285. [Google Scholar] [CrossRef]
  7. Kılınç, M.F.; Doluoğlu, Ö.G.; Karakan, T.; Dalkılıç, A.; Sönmez, N.C.; Yücetürk, C.N.; Reşorlu, B. The effect of ureteroscope size in the treatment of ureteral stone: 15-year experience of an endoscopist. Turk Urol. Derg. 2016, 42, 64–69. [Google Scholar] [CrossRef]
  8. Xu, X.; Xu, F.; Mao, K.; Zhong, W.; Zhang, M.; Huang, W.; Cheng, X.; Lu, Z.; Wei, G. First clinical application of the novel 6.3 Fr disposable digital flexible ureteroscope for treatment of lower calyx renal stones in anticoagulated patient. Urol. Case Rep. 2025, 60, 103015. [Google Scholar] [CrossRef]
  9. Ding, T.; Xu, Y.; Chen, Y.; Xiao, B.; Li, J.; Jianxing, L. Efficacy of 6. 3Fr disposable digital flexible ureteroscope versus 7. 5Fr disposable digital flexible ureteroscope in the treatment of upper urinary tract stones < 1.5 cm: A randomized controlled trial. World J. Urol. 2025, 43, 2–7. [Google Scholar] [CrossRef]
  10. Krajewski, W.; Nowak, Ł.; Tomczak, W.; Molik, K.; Ostrowski, T.; Łaszkiewicz, J.; Chorbińska, J.; Małkiewicz, B.; Szydełko, T. Feasibility and safety of 6.3 Fr vs. 7.5 Fr digital disposable ureteroscopes in retrograde intrarenal surgery: A prospective randomised trial. World J. Urol. 2025, 43, 510. [Google Scholar] [CrossRef]
  11. Geavlete, P.A.; Bulai, C.A.; Multescu, R.D.; Ene, C.V.; Militaru, A.; Punga, A.M.A.; Geavlete, B.F. Early clinical outcomes using a 6.3 Fr single use ureteroscope compared to a 7.5 Fr device. Sci. Rep. 2025, 15, 15362. [Google Scholar] [CrossRef]
  12. Page, M.J.; McKenzie, J.E.; Bossuyt, P.M.; Boutron, I.; Hoffmann, T.C.; Mulrow, C.D.; Shamseer, L.; Tetzlaff, J.M.; Akl, E.A.; Brennan, S.E.; et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. Syst. Rev. 2021, 10, 89. [Google Scholar] [CrossRef]
  13. Higgins, J.P.T.; Thomas, J.; Chandler, J.; Cumpston, M.; Li, T.; Page, M.J.; Welch, V.A. (Eds.) Cochrane Handbook for Systematic Reviews of Interventions; Wiley: Hoboken, NJ, USA, 2023. [Google Scholar]
  14. Sterne, J.A.C.; Savović, J.; Page, M.J.; Elbers, R.G.; Blencowe, N.S.; Boutron, I.; Cates, C.J.; Cheng, H.-Y.; Corbett, M.S.; Eldridge, S.M.; et al. RoB 2: A revised tool for assessing risk of bias in randomised trials. BMJ 2019, 366, l4898. [Google Scholar] [CrossRef]
  15. Guyatt, G.H.; Oxman, A.D.; Kunz, R.; Vist, G.E.; Falck-Ytter, Y.; Schünemann, H.J. Rating Quality of Evidence and Strength of Recommendations: What is “quality of evidence” and why is it important to clinicians? BMJ Br. Med. J. 2008, 336, 995. [Google Scholar] [CrossRef]
  16. Guyatt, G.H.; Oxman, A.D.; Vist, G.E.; Kunz, R.; Falck-Ytter, Y.; Alonso-Coello, P.; Schünemann, H.J. Rating Quality of Evidence and Strength of Recommendations: GRADE: An emerging consensus on rating quality of evidence and strength of recommendations. BMJ Br. Med. J. 2008, 336, 924. [Google Scholar] [CrossRef]
  17. Lin, L.; Chu, H. Quantifying publication bias in meta-analysis. Biometrics 2018, 74, 785–794. [Google Scholar] [CrossRef] [PubMed]
  18. Kim, H.J.; Daignault-Newton, S.; DiBianco, J.M.; Conrado, B.; Mohammad Jafri, S.; Seifman, B.; Konheim, J.; Dauw, C.A.; Ghani, K.R. Real-world Practice Stone-free Rates After Ureteroscopy: Variation and Outcomes in a Surgical Collaborative. Eur. Urol. Focus 2023, 9, 773–780. [Google Scholar] [CrossRef]
  19. Ulvik, Ø.; Harneshaug, J.R.; Gjengstø, P. What Do We Mean by “stone Free,” and How Accurate Are Urologists in Predicting Stone-Free Status following Ureteroscopy? J. Endourol. 2021, 35, 961–966. [Google Scholar] [CrossRef]
  20. Higgins, J.P.; Thompson, S.G. Quantifying heterogeneity in a meta-analysis. Stat. Med. 2002, 21, 1539–1558. [Google Scholar] [CrossRef] [PubMed]
  21. Auge, B.K.; Pietrow, P.K.; Lallas, C.D.; Raj, G.V.; Santa-Cruz, R.W.; Preminger, G.M. Ureteral Access Sheath Provides Protection against Elevated Renal Pressures during Routine Flexible Ureteroscopic Stone Manipulation. J. Endourol. 2004, 18, 33–36. [Google Scholar] [CrossRef] [PubMed]
  22. Maccraith, E.; Yap, L.C.; Elamin, M.; Patterson, K.; Brady, C.M.; Hennessey, D.B. Evaluation of the impact of ureteroscope, access sheath, and irrigation system selection on intrarenal pressures in a porcine kidney model. J. Endourol. 2021, 35, 512–517. [Google Scholar] [CrossRef] [PubMed]
  23. Xu, F.; Zhong, R.; Zhong, W.; Ye, S.; Lu, Z.; Xiong, L.; Xu, X. First application of a 6.3 Fr flexible ureteroscope in the treatment of bilateral staghorn calculi: A case report. Medicine 2025, 104, e43524. [Google Scholar] [CrossRef]
  24. Jeong, J.Y.; Cho, K.S.; Jun, D.Y.; Moon, Y.J.; Kang, D.H.; Jung, H.D.; Lee, J.Y. Impact of Preoperative Ureteral Stenting in Retrograde Intrarenal Surgery for Urolithiasis. Medicina 2023, 59, 744. [Google Scholar] [CrossRef]
  25. De Coninck, V.; Somani, B.; Sener, E.T.; Emiliani, E.; Corrales, M.; Juliebø-Jones, P.; Pietropaolo, A.; Mykoniatis, I.; Zeeshan Hameed, B.M.; Esperto, F.; et al. Ureteral Access Sheaths and Its Use in the Future: A Comprehensive Update Based on a Literature Review. J. Clin. Med. 2022, 11, 5128. [Google Scholar] [CrossRef] [PubMed]
  26. Yildirim, Ü.; Uslu, M.; Ezer, M.; Erdoğdu, Ç.; Kocaaslan, R.; Sarica, K. Comparison of flexible ureteroscopic suction techniques: Efficacy and safety of flexible and navigable access sheath (FANS) vs. direct in-scope suction (DISS) in the management of 2–3 cm lower pole renal stones. Urolithiasis 2025, 53, 75. [Google Scholar] [CrossRef] [PubMed]
  27. Abdelrasol, W.F.; Assem, A. Flexible ureteroscopy with and without ureteral access sheath in treatment of large renal stones: A randomized controlled study. Afr. J. Urol. 2024, 30, 62. [Google Scholar] [CrossRef]
  28. Güzel, Ö.; Balcı, M.; Tuncel, A.; Asfuroğlu, A.; Aykanat, İ.C.; Aslan, Y. Ureteral Stricture Outcomes Using Small Ureteral Access Sheath During Retrograde Intrarenal Surgery. J. Urol. Surg. 2023, 10, 321–325. [Google Scholar] [CrossRef]
  29. Traxer, O.; Thomas, A. Prospective evaluation and classification of ureteral wall injuries resulting from insertion of a ureteral access sheath during retrograde intrarenal surgery. J. Urol. 2013, 189, 580–584. [Google Scholar] [CrossRef]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.