Next Article in Journal
Fault Detection, Isolation, Identification and Recovery (FDIIR) Methods for Automotive Perception Sensors Including a Detailed Literature Survey for Lidar
Previous Article in Journal
Performance Analysis of Cooperative and Non-Cooperative Relaying over VLC Channels
Previous Article in Special Issue
Long-Term Evaluation and Calibration of Low-Cost Particulate Matter (PM) Sensor
Open AccessArticle

Laboratory Evaluations of Correction Equations with Multiple Choices for Seed Low-Cost Particle Sensing Devices in Sensor Networks

1
Research Center for Environmental Changes, Academia Sinica, Nangang, Taipei 115, Taiwan
2
Department of Atmospheric Sciences, National Taiwan University, Taipei 106, Taiwan
3
Institute of Environmental Health, National Taiwan University, Taipei 106, Taiwan
*
Author to whom correspondence should be addressed.
Sensors 2020, 20(13), 3661; https://doi.org/10.3390/s20133661
Received: 15 May 2020 / Revised: 21 June 2020 / Accepted: 24 June 2020 / Published: 30 June 2020
(This article belongs to the Special Issue Air Quality and Sensor Networks)
To tackle the challenge of the data accuracy issues of low-cost sensors (LCSs), the objective of this work was to obtain robust correction equations to convert LCS signals into data comparable to that of research-grade instruments using side-by-side comparisons. Limited sets of seed LCS devices, after laboratory evaluations, can be installed strategically in areas of interest without official monitoring stations to enable reading adjustments of other uncalibrated LCS devices to enhance the data quality of sensor networks. The robustness of these equations for LCS devices (AS-LUNG with PMS3003 sensor) under a hood and a chamber with two different burnt materials and before and after 1.5 years of field campaigns were evaluated. Correction equations with incense or mosquito coils burning inside a chamber with segmented regressions had a high R2 of 0.999, less than 6.0% variability in the slopes, and a mean RMSE of 1.18 µg/m3 for 0.1–200 µg/m3 of PM2.5, with a slightly higher RMSE for 0.1–400 µg/m3 compared to EDM-180. Similar results were obtained for PM1, with an upper limit of 200 µg/m3. Sensor signals drifted 19–24% after 1.5 years in the field. Practical recommendations are given to obtain equations for Federal-Equivalent-Method-comparable measurements considering variability and cost. View Full-Text
Keywords: air pollutant sensor; sensor evaluation; aerosol sensors; particle correction equation; chamber evaluation; interclass correlations air pollutant sensor; sensor evaluation; aerosol sensors; particle correction equation; chamber evaluation; interclass correlations
Show Figures

Figure 1

MDPI and ACS Style

Wang, W.-C.V.; Lung, S.-C.C.; Liu, C.H.; Shui, C.-K. Laboratory Evaluations of Correction Equations with Multiple Choices for Seed Low-Cost Particle Sensing Devices in Sensor Networks. Sensors 2020, 20, 3661.

Show more citation formats Show less citations formats
Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.