Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement

1
Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria
2
Faculty of Physics, University of Vienna, Strudelhofgasse 4, 1090 Vienna, Austria
3
Ingenieurbüro für technische Physik, Speichmühlengasse 1, 2380 Perchtoldsdorf, Austria
4
Biomechanische Forschungs-Gesellschaft m.b.H., 1040 Vienna, Austria
*
Author to whom correspondence should be addressed.
Sensors 2019, 19(17), 3751; https://doi.org/10.3390/s19173751
Received: 12 July 2019 / Revised: 19 August 2019 / Accepted: 26 August 2019 / Published: 30 August 2019
(This article belongs to the Special Issue Sensors for Biomechanics Application)
Wear is a common cause for aseptic loosening in artificial joints. The purpose of this study was to develop an automated diagnostical method for identification of the number and size distribution of wear debris. For this purpose, metal debris samples were extracted from a hip simulator and then analyzed by the electrospray method combined with a differential mobility analyzer, allowing particle detection ranging from several nanometers up to 1 µm. Wear particles were identified with a characteristic peak at 15 nm. The electrospray setup was successfully used and validated for the first time to characterize wear debris from simulated total joint replacement. The advantages of this diagnostic method are its time- and financial efficiency and its suitability for testing of different materials. View Full-Text
Keywords: debris; particles; wear; implants; prosthesis debris; particles; wear; implants; prosthesis
Show Figures

Figure 1

MDPI and ACS Style

von Skrbensky, G.; Mühlbacher, K.; Benca, E.; Kolb, A.; Windhager, R.; Reischl, G.; Reinisch, G. Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement. Sensors 2019, 19, 3751. https://doi.org/10.3390/s19173751

AMA Style

von Skrbensky G, Mühlbacher K, Benca E, Kolb A, Windhager R, Reischl G, Reinisch G. Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement. Sensors. 2019; 19(17):3751. https://doi.org/10.3390/s19173751

Chicago/Turabian Style

von Skrbensky, Gobert; Mühlbacher, Karoline; Benca, Emir; Kolb, Alexander; Windhager, Reinhard; Reischl, Georg; Reinisch, Georg. 2019. "Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement" Sensors 19, no. 17: 3751. https://doi.org/10.3390/s19173751

Find Other Styles
Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Article Access Map by Country/Region

1
Search more from Scilit
 
Search
Back to TopTop