Comparative Analysis of Blood Transfusion Accuracy and Hemolysis Rate of Transfusion Cartridge Set Between Conventional Infusion Pumps and Cylinder-Type Infusion Pumps
Abstract
:1. Introduction
2. Materials and Methods
2.1. Mechanism of Cylinder-Type Infusion Pumps
2.2. Eqiupment and the Packed RBC
2.3. Study Protocols
- i.
- Quantification and error rate according to infusion rate and time;
- ii.
- Degree of hemolysis according to infusion rate.
2.4. Statistical Analysis
3. Results
3.1. Overall Comparison by Infusion Rate of Each Device
3.2. Comparison of Error According to Infusion Rate
3.3. Transfusion Volume Error Rate and Laboratory Temperature for Each Experiment
- i.
- Anyfusion H-100: non-dilution: 3.86%; dilution: 3.60%; overall: 3.77%;
- ii.
- Terufusion TE-LM700: non-dilution: 3.98%; dilution: 4.05%; overall: 4.00%.
3.4. Overall Experiment Error Rate
3.5. Comparison of the Mean Hemolysis Rates of Each Medical Device
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Weinger, M.B.; Kline, A. Reflections on the current state of infusion therapy. Biomed. Instrum. Technol. 2016, 50, 253–262. [Google Scholar] [CrossRef] [PubMed]
- Levine, W.C.; Vernest, K.A. Infusion Pumps. In Anesthesia Equipment: Principles and Applications (Expert Consult: Online and Print); Elsevier: Amsterdam, The Netherlands, 2013; pp. 377–392. [Google Scholar]
- Syroid, N.; Liu, D.; Albert, R.; Agutter, J.; Egan, T.D.; Pace, N.L.; Johnson, K.B.; Dowdle, M.R.; Pulsipher, D.; Westenskow, D.R. Graphical user interface simplifies infusion pump programming and enhances the ability to detect pump-related faults. Anesth. Analg. 2021, 115, 1087–1097. [Google Scholar] [CrossRef] [PubMed]
- Korea Centers for Disease Control and Prevention. Guidelines for Blood Transfusion, 2nd ed.; Korea Centers for Disease Control and Prevention: Cheongju, South Korea, 2016. [Google Scholar]
- Klem, S.A.; Farrington, J.M.; Leff, R.D. Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion. Crit. Care. Med. 1993, 21, 1213–1217. [Google Scholar] [CrossRef] [PubMed]
- Van Gent, J.M.; Clements, T.W.; Cotton, B.A. Resuscitation and care in the trauma bay. Surg. Clin. 2024, 104, 279–292. [Google Scholar] [CrossRef] [PubMed]
- Poder, T.; Boileau, J.C.; Lafrenière, R.; Thibault, L.; Carrier, N.; De Grandmont, M.J.; Beauregard, P. VP189 Hemolysis Induced by Modern Infusion Pumps During Blood Transfusion. Int. J. Technol. Assess. Health Care 2017, 33, 236–237. [Google Scholar] [CrossRef]
- McNaughton, J.; Hughes, J.; Andrews, J.C.; George, T.I.; Bergero, C.; Pike-Grimm, K.; Galel, S.; Gonzalez, C.; Goodnough, L.; Fontaine, M.J. Hemoglobinuria and mechanical hemolysis associated with red blood cell transfusion in pediatric patients. Blood 2012, 120, 3429. [Google Scholar] [CrossRef]
- Capes, D.F.; Dunster, K.R.; Sunderland, V.B.; McMillan, D.; Colditz, P.B.; McDonald, C. Fluctuations in syringe-pump infusions: Association with blood pressure variations in infants. Am. J. Health. Syst. Pharm. 1995, 52, 1646–1653. [Google Scholar] [CrossRef] [PubMed]
- Pelletier, N.P.; Fisher, A. Infusion pump in UH60L/M flight in Afghanistan: Why failures occur. Air. Med. J. 2017, 36, 248–257. [Google Scholar] [CrossRef] [PubMed]
- Décaudin, B.; Dewulf, S.; Lannoy, D.; Simon, N.; Secq, A.; Barthélémy, C.; Debaene, B.; Odou, P. Impact of multi-access infusion devices on in vitro drug delivery during multi-infusion therapy. Anesth. Analg. 2009, 109, 1147–1155. [Google Scholar] [CrossRef] [PubMed]
- Kan, K.; Levine, W.C. Infusion Pumps. In Anesthesia Equipment; WB Saunders: Philadelphia, PA, USA, 2021; pp. 351–367. [Google Scholar]
- Min, J.J.; Kim, D.K.; Hong, K.Y.; Choi, J.W.; Choi, K.Y. Comparison of operator workloads associated with the single-unit Anyfusion® pump and the changeover from a syringe pump to an infusion pump. J. Korean. Med. Sci. 2019, 34, e314. [Google Scholar] [CrossRef] [PubMed]
- Kessler, C. Priming blood transfusion tubing: A critical review of the blood transfusion process. Crit. Care. Nurse 2013, 33, 80–84. [Google Scholar] [CrossRef] [PubMed]
- McCurdy, D.E.; Arnold, M.T. Development and implementation of a pediatric/neonatal iv syringe pump delivery system. Neonatal. Netw. 1995, 14, 9–18. [Google Scholar] [PubMed]
- Golpaygani, T.A.; Movahedi, M.M.; Reza, M.; Hassani, K.A. Study on performance and safety test of infusion pump devices. Biomed. Res. India 2017, 28, 5179–5181. [Google Scholar]
- Ilfeld, B.M.; Morey, T.E.; Enneking, F.K. The delivery rate accuracy of portable infusion pumps used for continuous regional analgesia. Anesth. Analg. 2002, 95, 1331–1336. [Google Scholar] [CrossRef] [PubMed]
- Mohseni, M.; Ebneshahidi, A. The flow rate accuracy of elastomeric infusion pumps after repeated filling. Anesthesiol. Pain Med. 2014, 4, E14989. [Google Scholar] [CrossRef] [PubMed]
- Calkins, J.M.; Vaughan, R.W.; Cork, R.C.; Barberii, J.; Eskelson, C. Effects of dilution, pressure, and apparatus on hemolysis and flow rate in transfusion of packed erythrocytes. Anesth. Analg. 1982, 61, 776–780. [Google Scholar] [CrossRef] [PubMed]
- Herrera, A.J.; Corless, J. Blood transfusions: Effect of speed of infusion and of needle gauge on hemolysis. J. Pediatr. 1981, 99, 757–758. [Google Scholar] [CrossRef] [PubMed]
- Kusahara, D.M.; Pedreira, M.; Avelar, A.M.; Peterlini, M.A. Performance of a linear peristaltic infusion pump during red blood cells administration and the influence of infusion rates. Intensive Care Med. Exp. 2015, 3, A548. [Google Scholar] [CrossRef]
- Wilson, A.M.M.M.; Peterlini, M.A.S.; Pedreira, M.D.L.G. Infusion pumps and red blood cell damage in transfusion therapy: An integrative revision of the academic literature. Rev. Latino. Am. Enfermagem. 2016, 24, e2763. [Google Scholar] [CrossRef] [PubMed]
- Kim, M.S.; Kim, J.S.; Lim, E.; Park, H.; Hwang, S.H.; Oh, H.B.; Ko, D.H. Performance Evaluation of Infusion Systems for Red Blood Cell Transfusion. J. Lab. Med. Qual. Assur. 2019, 41, 161–165. [Google Scholar] [CrossRef]
No. | Infusion Speed (cc/hr) | Duration (hrs) | Anyfusion H-100 | Terumo TE-LM700 | ||||
---|---|---|---|---|---|---|---|---|
Total Infusion Volume (cc) | Error Rate (%) | Temperature (°C) | Total Infusion Volume (cc) | Error Rate (%) | Temperature (°C) | |||
1 | 10 | 3 | 30.9 | 4.39% | 25.1 | 28.8 | 5.00% | 24.5 |
2 | 15 | 2 | 29.5 | 3.71% | 24.9 | 29.6 | 3.99% | 24.4 |
3 | 30 | 1 | 31.2 | 4.28% | 24.8 | 29 | 3.33% | 24.5 |
4 | 60 | 1 | 58.9 | 3.92% | 24.8 | 57.2 | 4.65% | 24.6 |
5 | 120 | 1 | 118.9 | 4.12% | 25.4 | 120.2 | 3.33% | 24.7 |
6 | 180 | 1 | 181.6 | 2.78% | 25.4 | 179.4 | 3.55% | 24.7 |
No | Infusion Speed (cc/hr) | Duration (hrs) | Anyfusion H-100 | Terumo TE-LM700 | ||
---|---|---|---|---|---|---|
Types of Errors | Number of Occurrences (n, %) | Types of Errors | Number of Occurrences (n, %) | |||
1 | 10 | 3 | Air | 2 (33.3) | Air | 1 (16.7) |
Clot | 2 (33.3) | Clot | 2 (33.3) | |||
Sample ineligibility | 0 (0) | Sample ineligibility | 0 (0) | |||
2 | 15 | 2 | Air | 1 (14.3) | Air | 1 (14.3) |
Clot | 2 (28.6) | Clot | 2 (28.6) | |||
Sample ineligibility | 0 (0) | Sample ineligibility | 0 (0) | |||
3 | 30 | 1 | Air | 1 (16.7) | Air | 1 (16.7) |
Clot | 0 (0) | Clot | 0 (0) | |||
Sample ineligibility | 1 (16.7) | Sample ineligibility | 1 (16.7) | |||
4 | 60 | 1 | Air | 1 (16.7) | Air | 1 (16.7) |
Clot | 0 (0) | Clot | 0 (0) | |||
Sample ineligibility | 1 (16.7) | Sample ineligibility | 2 (33.3) | |||
5 | 120 | 1 | Air | 1 (16.7) | Air | 0 (0) |
Clot | 0 (0) | Clot | 0 (0) | |||
Sample ineligibility | 1 (16.7) | Sample ineligibility | 0 (0) | |||
6 | 180 | 1 | Air | 2 (33.3) | Air | 1 (16.7) |
Clot | 0 (0) | Clot | 0 (0) | |||
Sample ineligibility | 0 (0) | Sample ineligibility | 0 (0) | |||
Occurrence rate out of 36 cases | Air | 8 (21.6) | Air | 5 (13.5) | ||
Clot | 4 (10.8) | Clot | 4 (10.8) | |||
Sample ineligibility | 4 (10.8) | Sample ineligibility | 3 (8.1) | |||
Error | 1 (2.7) | Error | 0 (0) |
No | Anyfusion H-100 | Terumo TE-LM700 | No | Anyfusion H-100 | Terumo TE-LM700 | ||||
---|---|---|---|---|---|---|---|---|---|
Error Rate (%) | Temperature (°C) | Error Rate (%) | Temperature (°C) | Error Rate (%) | Temperature (°C) | Error Rate (%) | Temperature (°C) | ||
1 | 4.00 | 25.4 | −4.00 | 24.3 | 30 | −4.25 | 25.1 | −1.41 | 25.1 |
2 | 6.33 | 25.6 | −8.30 | 24.8 | 31 | 5.91 | 25 | −3.16 | 24.8 |
3 | −4.00 | 25 | 3.00 | 24.9 | 32 | 2.10 | 26.3 | −3.30 | 24.9 |
4 | 5.67 | 24.9 | −3.00 | 24.6 | 33 | 2.33 | 26.2 | −2.67 | 25 |
5 | 2.67 | 25.3 | −7.30 | 24.3 | 34 | −1.22 | 24.3 | −3.44 | 24 |
6 | 3.67 | 24.2 | −4.33 | 24.1 | 35 | Error | −4.33 | 24.6 | |
7 | −2.33 | 25.1 | 4.00 | 24.2 | 36 | −3.77 | 24.9 | 3.78 | 24.6 |
8 | 1.67 | 25.3 | −3.60 | 24.6 | 37 | 4.50 | 25.3 | 3.78 | 24.9 |
9 | −7.00 | 24.6 | −4.67 | 24.8 | D1 | −4.50 | 24.8 | 3.25 | 24.3 |
10 | −5.67 | 24.8 | −3.33 | 24.3 | D2 | −2.58 | 24.9 | 0.42 | 24.5 |
11 | 1.67 | 24.9 | 4.33 | 24.3 | D3 | −3.17 | 25 | −4.58 | 24.3 |
12 | 3.33 | 25.1 | 2.00 | 24.3 | D4 | 4.25 | 25.1 | −3.41 | 24.5 |
13 | −4.33 | 24.7 | −6.00 | 24.5 | D5 | −5.08 | 24.7 | −5.17 | 24.6 |
14 | −0.06 | 24.9 | −3.00 | 24.1 | D6 | 0.67 | 24.6 | 6.33 | 24.1 |
15 | 4.33 | 25 | −5.33 | 24.6 | D7 | 4.58 | 24.8 | 7.00 | 24.2 |
16 | 5.67 | 24.4 | −2.67 | 24.7 | D8 | 2.25 | 24.9 | 2.08 | 25.1 |
17 | 3.33 | 25 | −4.00 | 24.6 | D9 | −4.83 | 25.2 | 1.41 | 24.8 |
18 | 7.33 | 24.9 | −1.33 | 24.4 | D10 | 2.58 | 24.7 | −1.75 | 24.6 |
19 | 5.00 | 24.5 | −3.67 | 24.4 | D11 | 3.67 | 24.3 | −5.33 | 24.8 |
20 | 1.33 | 25.4 | −4.67 | 24.4 | D12 | 0.75 | 24.6 | 3.42 | 24.4 |
21 | −4.83 | 24.7 | −3.67 | 24.9 | D13 | −4.33 | 24.5 | 6.92 | 25 |
22 | −3.17 | 25.3 | −5.50 | 25.1 | D14 | −2.50 | 25 | −3.17 | 24.6 |
23 | −4.50 | 24.3 | −5.10 | 24.2 | D15 | −4.00 | 24.7 | −3.83 | 24.7 |
24 | 5.00 | 24.4 | −4.80 | 24.4 | D16 | 5.25 | 25.2 | 5.25 | 24.9 |
25 | −4.67 | 24.4 | −4.16 | 24.5 | D17 | 6.50 | 24.4 | −0.92 | 24.3 |
26 | −2.75 | 26.2 | 5.08 | 24.2 | D18 | −3.83 | 24.7 | 5.41 | 24.6 |
27 | −4.25 | 25.8 | 3.50 | 24.9 | D19 | −2.67 | 24.3 | −6.00 | 24.2 |
28 | −3.67 | 25.1 | 1.90 | 24.3 | D20 | 4.17 | 25.3 | 5.25 | 24.8 |
29 | 3.91 | 25.4 | −4.91 | 25.1 | Average | 3.77 | 25.0 | 4.00 | 24.6 |
Medical Device | Air Trapping (n, %) | Clot (n, %) | Normal (n, %) | X2 | p-Value |
---|---|---|---|---|---|
Anyfusion H-100 | 9 (15.8) | 4 (7.0) | 44 (77.2) | 0.294 | 0.863 |
Terumo TE-LM700 | 7 (12.3) | 4 (7.0) | 46 (80.7) |
Medical Device | Values (%, Mean ± SD) | t | p-Value |
---|---|---|---|
Anyfusion H-100 | 0.566 ± 0.095 | 0.294 | 0.863 |
Terumo TE-LM700 | 0.518 ± 0.126 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Lee, H.-Y.; Kim, S.-J.; Lee, K.-H.; Park, I.-H.; Jeon, H.-J.; Youk, H. Comparative Analysis of Blood Transfusion Accuracy and Hemolysis Rate of Transfusion Cartridge Set Between Conventional Infusion Pumps and Cylinder-Type Infusion Pumps. Biomedicines 2024, 12, 2421. https://doi.org/10.3390/biomedicines12112421
Lee H-Y, Kim S-J, Lee K-H, Park I-H, Jeon H-J, Youk H. Comparative Analysis of Blood Transfusion Accuracy and Hemolysis Rate of Transfusion Cartridge Set Between Conventional Infusion Pumps and Cylinder-Type Infusion Pumps. Biomedicines. 2024; 12(11):2421. https://doi.org/10.3390/biomedicines12112421
Chicago/Turabian StyleLee, Hee-Young, Sun-Ju Kim, Kang-Hyun Lee, Il-Hwan Park, Hyeok-Jin Jeon, and Hyun Youk. 2024. "Comparative Analysis of Blood Transfusion Accuracy and Hemolysis Rate of Transfusion Cartridge Set Between Conventional Infusion Pumps and Cylinder-Type Infusion Pumps" Biomedicines 12, no. 11: 2421. https://doi.org/10.3390/biomedicines12112421
APA StyleLee, H.-Y., Kim, S.-J., Lee, K.-H., Park, I.-H., Jeon, H.-J., & Youk, H. (2024). Comparative Analysis of Blood Transfusion Accuracy and Hemolysis Rate of Transfusion Cartridge Set Between Conventional Infusion Pumps and Cylinder-Type Infusion Pumps. Biomedicines, 12(11), 2421. https://doi.org/10.3390/biomedicines12112421