Pressure Sensors for Measuring Tibiofemoral Contact Mechanics in Meniscal Root Repair: A Systematic Review

,
,
,
,
,
and
1
Department of Biomedical Engineering, University of Birmingham, Birmingham B15 2TT, UK
2
Leicester Academic Knee Unit, University Hospitals of Leicester NHS Trust, Leicester LE1 5WW, UK
3
Department of Trauma and Orthopaedics, Morriston Hospital, Swansea SA6 6NL, UK
4
Department of Trauma & Orthopaedics, University Hospitals of Birmingham NHS Trust, Birmingham B15 2GW, UK
This article belongs to the Section Biomedical Sensors

Abstract

Background: Tibiofemoral contact mechanics (TFCM) is an accepted biomechanical metrics for evaluating the meniscus in its intact, torn, and repaired states. Pressure sensors are increasingly used, with accuracy and repeatability influenced by test conditions, their design, and their properties. To identify factors optimising performance, we performed a systematic review of the literature on their use for measuring TFCM in posterior meniscal root tears. Methods: The Cochrane Controlled Register of Trials, PubMed, and Embase were used to perform a systematic review using the PRISMA criteria. As laboratory and surgical setup can influence sensor performance, we collected data on specimen preparation, repair techniques, hardware use, and biomechanical testing parameters. Results: 24 biomechanical studies were included. Specimen preparations were similar across studies with respect to femoral and tibial mounting. Single axial compressive forces were applied between 100 and 1800 N at varying flexion angles (0–90°). Tekscan (Boston, MA, USA) was the commonest sensor used to measure TFCM, followed by digital capacitive sensors and Fujifilm (Tokyo, Japan). Factors influencing their performance included fluid exposure, lack of adequate fixation, non-specific calibration protocols, load saturation exceeding calibration, damaged sensels and inappropriate pre-test conditioning. Conclusions: Understanding potential factors influencing pressure sensors may improve accuracy, area, and pressure distribution measurements.

1. Introduction

The posterior roots of the menisci play a critical role in preserving meniscal function and the distribution of axial load across the tibial plateau [1,2,3]. Posterior meniscal root tears (PMRTs) compromise the circumferential integrity of the meniscus and prohibit the generation of hoop stress. In doing so, the chondral surface is susceptible to injury upon loading, and can subsequently degenerate [4,5,6,7]. Therefore, it is strongly recommended to repair these tears in appropriately indicated patients [8,9]. Tibiofemoral contact mechanics (TFCM)—in particular, contact area and pressure—provides a form of measurement for chondral injury, and is an accepted biomechanical metric for the evaluation of the menisci in its intact, torn, and repaired state [10,11]. Knowledge of the patterns of TFCM in these conditions will determine the approach one takes to restore knee joint function [12]. Pressure sensors are increasingly being used to quantify contact area and pressure [13,14,15,16,17,18,19,20,21,22], with importance placed on appropriate and correct utilisation, as such outcomes in biomechanical research can affect clinical practice and patient care. For instance, the performance of the sensors depends on the test setup and the laboratory conditions in which they operate. This includes their exposure to fluid, temperature settings, and humidity; the ease of insertion within the joint; their calibration sequence; and how the tibiofemoral joint is mounted to ensure the sensors are securely fitted for analysis [11,13,15,17,19,20,21,22]. The sensor accuracy, to include repeatability and bias, is influenced by its inherent design and properties. Therefore, to properly interpret studies using these techniques, it is crucial to understand the behaviour and limitations of the sensors within this context. In this respect, we performed a systematic review of the literature on the use and applicability of pressure sensors for the measurement of tibiofemoral contact mechanics following repair of posterior meniscal root tears.

2. Methods

2.1. Literature Search

A systematic review was performed and reported according to the standards of the PRISMA (Preferred Reporting Items for Systematic Review and Meta-Analyses) criteria [23]. Searches of Cochrane Controlled Register of Trials, PubMed (MEDLINE), and Embase were conducted from the inception of the databases to 15 January 2024, and repeated on 10 November 2024 for an update of the literature. Both dates are illustrated in Tables S1 and S2. The Boolean search items included (‘posterior horn’ OR ‘root’ OR ‘radial’) AND (‘meniscus’ or ‘meniscal) AND (‘tibiofemoral contact’ OR ‘contact pressure’ OR ‘contact area’). No restrictions were made on language, and efforts were made to obtain translated versions of all included studies. Restrictions were not placed on the date of publication or the journal. All relevant articles and reviews were examined for further relevant citations.

2.2. Eligibility Criteria and Outcome Measures

All biomechanical studies on cadaveric or animal knee joints that used pressure sensors to measure tibiofemoral contact mechanics (to include contact area and pressure) following repair of posterior meniscal roots were included. No limitations were placed on the type of repair (i.e., transtibial pull-through (TPR), suture anchor repair (SAR), side-to-side repair, all-inside repair), and studies with treatment for concomitant injuries were also included. This includes, but is not limited to, anterior cruciate ligament reconstruction (ACLR), meniscotibial ligament tenodesis, and open-wedge high tibial osteotomy (OWHTO). Whilst these additional techniques could potentially confound the analyses from the sensors, this review ensured interpretation of the sensors was confined to the sub-groups where only a root repair was performed. Exclusion criteria included computational-based studies, editorial letters, case reports, technical notes, and expert consensuses and abstract-only studies.

2.3. Study Selection and the Assessment of Quality of Studies

Two authors (Authors 1 and 2) independently reviewed the titles and abstracts, after which relevant papers were reviewed in full by each author independently. Those that met the eligibility criteria were chosen, with discrepancies highlighted and reviewed by senior authors (Authors 5, 6, and 7). The same 2 authors independently assessed the methodological quality of the biomechanical studies using the Methodological Index for Non-randomised studies (MINORS) tool (Table S3) [24]. The risk of bias of the included biomechanical studies was assessed and reported by the same 2 authors in accordance with the Risk of Bias in Nonrandomized Studies of Interventions tool (ROBINS) [25] Each item was judged according to high, moderate, low, or unclear risk of bias. Studies were deemed to have the highest risk of bias if they scored a high or unclear risk of bias.
Furthermore, the final level of certainty of evidence across all studies was assessed using the Grading of Recommendations, Assessment, Development, and Evaluation (GRADE) tool. Five domains were assessed to include risk of bias, inconsistency, imprecision, indirection, and publication bias [26]. Rating of evidence was high, moderate, low, or very low. High certainty of evidence implies the evidence is very likely to represent the truth, and low certainty and below indicates that the evidence is unlikely to represent the truth.
As laboratory and surgical setup can potentially influence the performance of pressure sensors, we collected data on specimen preparation, repair techniques, hardware use, biomechanical testing parameters, and the conditions. Where possible, further data were collected on the design and properties of the sensor.

2.4. Data Synthesis

Continuous-variable data were reported as mean ± standard deviation from the mean unless otherwise specified. Categorical-variable data were reported as frequency and percentage. Of note, when data were presented, it was ensured the statistical tests for the included studies were adequately justified; that is, parametric tests were performed for studies comparing mean differences between two groups (e.g., two-tailed independent t-test), where normal distribution of data with equal variance was assumed, and non-parametric tests (e.g., Mann–Whitney U test) were performed for studies in which non-normal data distribution with unequal variance was assumed. In addition, studies comparing mean differences between three or more groups were analysed to ensure parametric tests (e.g., ANOVA) were only performed when the assumptions of normal data distribution with homogeneity of variances were met, with pairwise post hoc comparison tests (e.g., Tukey) to identify groups that differed from one another. Furthermore, it was ensured that non-parametric tests (e.g., Kruskal–Wallis) were only performed when no assumptions of the data distribution were made, with adequate post hoc comparison tests (e.g., Dunns method). A value of p < 0.05 was considered statistically significant unless otherwise noted.

3. Results

Thirty-seven full-text articles were reviewed. Thirteen studies were excluded based on the eligibility criteria (Figure 1). Twenty-four papers were therefore included in this systematic review [1,3,21,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47]. The mean rating of all studies using the MINORS tool was 20.5 ± 1.35 points (of a maximum of 24 points, 85.4%; range, 19–22 points) (Table S4). A common reason for point deductions was item 5 (unbiased assessment of the study endpoint), as the observers were unblinded to the study endpoints and therefore may have influenced the results. This can potentially distort the true effect being measured, reducing the validity of the impact of the exposure being studied. The same reason was also noted following ROBINS assessment (Figures S1 and S2), whilst the certainty of evidence using the GRADE criteria was high across all studies.
Figure 1. PRISMA flow diagram for study selection.

3.1. Biomechanical Characteristics

In total, there were 235 testing specimens, of which 187 were in cadaveric and 48 in porcine knees. Seventeen studies measured tibiofemoral contact mechanics following the repair of a medial meniscal posterior root tear (MMPRT) [1,3,21,27,28,29,30,31,35,36,37,39,40,41,42,44,45], of which two performed additional treatment for concomitant injuries to include meniscotibial ligament (MTL) tenodesis and OWHTO [31,41]. The remaining seven studies were for lateral meniscal posterior root tears (LMPRTs) [32,33,34,38,43,46,47], and two studies performed either an ACL reconstruction or created in addition a meniscofemoral ligament (MFL) injury [32,33]. The baseline characteristics are described in Table 1.
Table 1. Biomechanical characteristics of included studies. ACL, anterior cruciate ligament; ATPR, anatomic transtibial pull-through; HCT, horizontal cleavage tear; LFC, lateral femoral condyle; LM, lateral meniscus; LMPHT, lateral meniscal posterior horn tear; LMPRT, lateral meniscus posterior root tear; MAT, meniscal allograft transplantation; MD, mean difference; MM, medial meniscus; MMPH, medial meniscus posterior horn; MMPRT, medial meniscus posterior root tear; MCL, medial collateral ligament; MFC, medial femoral condyle; MFL, menisco-femoral ligament; MTL, meniscotibial ligament; MTM, material-testing machine; NATPR, non-anatomic transtibial pull-through; PCL, posterior cruciate ligament; PMMA, polymethylmethacrylate; TFCM, tibiofemoral contact mechanics; TPR, transtibial pull through repair. Biomechanical characteristics of included studies. ASA, all suture anchor; MMA, modified Mason–Allen; TSS, two simple stitch.
Bone and specimen preparation were similar across most studies, including storage, dissection, preservation of knee-stabilising structures, and mounting on the testing machine. Despite the application of a single axial load across most studies, its range varied between 100 N and 1800 N in cadaveric knees, with 1000 N being commonly used. For porcine knees, this ranged between 200 N and 1500 N. Transosseous pull-through repair (TPR) was the most common repair performed for posterior meniscus root tears. Flexion angles that were commonly used were 0°, 30°, 60°, and 90°. Although this can affect the pressure on the joint surfaces, the consensus in most studies (twenty-one) was that the surgical repair of root tears restored tibiofemoral contact mechanics to that of the intact meniscus at most of these angles, and these results were similar when compared across studies [1,3,21,28,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46].

3.2. Pressure Sensor Application and Insertion

Eighteen studies used the Tekscan pressure sensor (Boston, MA, USA) to measure tibiofemoral contact mechanics. The remaining studies used either Fuji Prescale films (Fujifilm, Tokyo Japan) [1,28,39] or digital capacitive sensors (Munich, Germany) [29,32,35]. The application and insertion of the pressure sensors are described in Table 2.
Table 2. Pressure sensor application and insertion. N/A, not applicable; NR, not reported; kPA, kilo Pascals.
Capsular arthrotomies in the anterior and posterior meniscotibial ligaments were generally undertaken to position the sensors beneath the meniscus. Only three studies placed the sensors above the meniscus [27,35,44]. All but one study preserved the cruciate and collateral ligaments. This individual study sectioned the medial collateral ligament (MCL) for Tekscan insertion [27]. Fourteen studies secured the sensors within the knee, either with suture anchors [30,33,37,38,40,46], with screws [31,41,43,45], or securing them to the meniscocapsular junction [22,28,44,47]. Sensor position was confirmed during biomechanical testing either by disarticulating the femoral osteotomy initially created [30,31,33,34,36,38,40,47] or by cross-referencing its position with a standardised reference point [1,32,46].
Fujifilm was packed in a fluid-proof packet, preventing exposure to saline before and during testing. Tekscan was soaked in saline prior to testing in only six studies [33,37,38,40,41,45]. However, most soaked the sensor during testing to prevent shear forces and soft-tissue desiccation [21,27,30,33,34,36,37,38,40,41,44,45,47]. Five studies provided precautions to avoid measurement error in the presence of saline [30,33,37,38,40].
To deal with potential sensor damage or wrinkling, a new Tekscan sensor was generally used in the relevant studies [21,27,30,33,34,37,38,40,43,45,46]. Only four of these studies accounted for potential damage from repetitive loading on the sensors. This was by replacing false measures by means of the surrounding sensels [33,37,38,40]. However, none of these studies analysed the specific mechanism of damage.
A new Fujifilm was used for every test, whilst the same capacitive sensor was used throughout testing.
Five studies with Tekscan performed pre-tensioning [21,27,44,45,47], and two with capitative sensors [29,35]. For accurate Tekscan readings, two studies reported on their maximum detection limit [3,36], and three provided meticulous calibration to establish a baseline reference of 0 [21,44,45]. This ensured all sensors were activated and ready for data collection. Tekscan was frequently re-calibrated, either between tests [21,35,42,43,44,47] or with a new knee specimen [27,30,33,37,38,40,41,45,46] No re-calibration was required with the capacitive or the Fujifilm sensors.

4. Discussion

The purpose of this systematic review was to help researchers in making informed decisions regarding the applicability and limitations of pressure sensors in measuring tibiofemoral contact mechanics following repair of the meniscal root. Our review demonstrated that, qualitatively, surgical repair recreates the function of an intact state, and that a variety of pressure sensors can detect such change.
The piezoresistive Tekscan Pressure Sensor (Tekscan, Boston, MA, USA) was commonly used in the included studies. It is widely used in pressure and force detection and has been validated in several studies [48,49,50]. It allows for continuous data collection throughout several load configurations and dynamic simulation of joint movements in vitro. However, potential limitations to data accuracy involve several factors. These include crinkling or damage to individual sensels following repetitive manipulation and shear loading, load saturation exceeding calibration, output changes related to its placement and liquid exposure, and whether it is appropriately conditioned prior to testing [21,40]. During liquid exposure, the load output diminishes with time, with 90% of this decline mitigated by sensor exposure to saline for 48 h [51]. Therefore, saturating the sensor prior to loading improves the accuracy of the output recorded over time, reducing the likelihood of requiring post hoc data collection [51]. Six studies provided evidence of pre-saturation [33,37,38,40,41,45]. However, despite this, linear decline in mean total load outputs over several data captures for each knee is possible. Therefore, the acquired data should be normalised with the use of a measured linear rate of decline. This was undertaken in five studies [30,33,37,38,40]. Alternatively, one can prevent the output drift by isolating the sensor with a plastic film or Teflon tape [18,19,22,51]. Teflon, however, can potentially triple the thickness of the original 0.1 mm-thick sensors, inducing a 10–26% increase of contact pressure. Therefore, modifications that increase the thickness should be avoided [52,53]. However, polyurethane plastic film is 0.01 mm thick [18,52], which may potentially have a negligible effect on TFCM following repair of the meniscal root. This requires further investigation. However, a potential solution in dealing with the spurious results from sensor thickness, regardless of the technological design (resistive or capacitive), is to pre-tension the sensor first. This will remove the compressive pressure caused by its thickness prior to application of real compressive force. This was performed in seven of the twenty-one suitable studies [21,27,29,35,44,45,47].
Measurement accuracy relies on the secure attachment of the sensors. Anderson et al. reported that not all measured parameters could be evaluated with shifts in sensor position [54]. Fourteen studies in this review prevented this from occurring through securing the sensor on the tibia. Overestimation of the contact area and stress can confound the results when a shearing force is experienced within the joint. This was limited through keeping the sensors moist within the experiments and preventing the knee from over-constraining within the biomechanical setup.
Though the Tekscan sensor is reusable, its lifetime is greatly influenced by repetitive manipulation, testing, and subsequent wear following load. Damage to individual sensels can lead to erroneous contract stress and area measurements, identified from false pressure readings during no load. This can be minimised through periodic calibration checks, changing sensors, or replacing the erroneous sensel value with the mean of values reported by the surrounding sensels, as has been done in previous studies [17]. This was reported and performed in four of the included studies [33,37,38,40].
A potential factor that may affect sensor accuracy is the extent of knee disarticulation and soft-tissue dissection. Martens et al. reported that femoral condyle osteotomy for sensor visualisation and coronary ligament incision for sensor placement had no effect on the TFCM of the knee [55]. The minimal differences between the osteotomised joint and the osteotomised joint with an incised coronary ligament indicated that the ligament does not involve the transmission of compressive loads in the joint, at least not loads that are detectable with the range of Fujifilm used and the methods employed [55]. Even if ligament transection were to render the meniscus unstable and alter the contact area and pressure, the effect would be to decrease the contact area and increase the pressure, tending to minimise the potential difference between the intact, torn, and repaired knee menisci. However, care is required to keep the main stabilising elements of the knee, the cruciate and collaterals, intact. Otherwise, this creates rotational instability within the knee, affecting the pressure readings of the sensor. Only one included study incised the collateral for pressure sensor insertion [27]. Interestingly, this was one of the very few that did not restore TFCM following meniscal root repair.
Frequent calibration checks can minimise pressure measurement errors. Where described, Tekscan sensor studies had specific calibration protocols. This included having a maximum detection limit, an activation load for data collection, and tests being performed immediately after calibration. The latter is particularly significant in optimising sensitivity for force detection [56,57]. However, the calibration process, though simple, can be exhaustive and repetitive.
Capacitive sensors do not require re-calibration [29,32,35], and thus, they may be more useful for complex and serial experiments where several testing conditions are under investigation. Furthermore, capacitive sensors are shown to better conform to shaped surfaces and have a lower detection error [58]. In addition, their higher repeatability means they are less likely to be affected by sensor re-position compared to the Tekscan system. However, their 1 mm dielectric thickness results in more uniform redistribution of loading, which may not represent the actual loading of the joint being measured [12].
Fujifilm (Tokyo, Japan) is static, passive technology whereby its colour intensity is proportional to the applied load (colour-based sensors). Pressure can be applied to the films in one of two ways: (a) gradually increasing the pressure to the required level in two minutes and maintaining that pressure for another two minutes, which is known as continuous pressure, or (b) applying pressure for five seconds and maintaining pressure for another five seconds, which is known as momentary pressure. For ±10% precision, the recommended temperature and room humidity range is 20–35°C and 35–80%, respectively. After loading, the imprinted Fujifilm is usually photographed and imported into a customised MATLAB programme (Mathworks, Natwick, Massachusetts) based on the optical density of the scanned Fujifilm and a fifth-order polynomial developed from Fuji film calibration data. The calibration data are provided in the manufacturer guidelines, with each pressure film having its own pixel values and colour density. A written coded file converts this intensity to stress, providing an estimate of the local contact stress and area. Although it only provides a measure of TFCM at one time under one set of circumstances [58,59,60], its use with instantaneous loads can help to avoid the redistribution of fluid in articular cartilage and, in turn, pressure distributions that occur when testing a joint with a static load [61]. There are various ranges of films available (super low–super high), all of which operate under different load levels to ensure pressure maps can be created without the film becoming fully saturated. This may limit the acquisition of data based on the load applied. However, the super-low-range film (0.5–2.5 MPa) has greater pressure resolution compared to its counterparts [62], and Martens et al. noted a load of 1.75× body weight provides accurate pressure maps under these circumstances [55]. This was further underlined in the three Fujifilm studies in this review [1,28,39]. All provided appropriate pressure maps and contact area with the super-low film at loads between 1000 and 1800 N, reflecting the calculated target described by Martens et al. [55]. Harris et al. reported that although Fujifilm underestimated the contact area compared to Tekscan sensors across all flexion angles [63], this difference was reduced with the super-low films, particularly at 1000 N, the commonest load for biomechanical testing in this review. However, its relatively lower level of accuracy is suggestive in the use of the alternative sensors described. Yet despite this, underestimation of TFCM with Fujifilm occurred across all testing conditions, therefore minimising the potential difference between the intact, torn, and repaired meniscal root. In fact, all three studies showed similar conclusions to those of Tekscan and capacitive sensors [1,28,39], whereby they detected TFCM to restore to the intact state following root repair. In this context, super-low Fujifilm, both cheap and readily accessible, may be a viable alternative to the more desirable options in situations where simpler, more static, and preliminary experiments are undertaken.

Limitations

This review is not without limitations. The presence of a variety of surgical techniques, loading cycles, and performance loads may have altered the true TFCM and its detection by the sensors, thereby confounding the biomechanical outcomes. However, similar conclusions in many of the included studies suggest these effects were minimal. Furthermore, we included porcine knees, which do not represent the consequences of a posterior meniscal root tear and repair to that of a human knee. Nevertheless, they are considered a valid alternative in the orthopaedic field, especially for meniscal root studies, due to the analogous function between the model and the anatomy of the human knee with persistent material properties [14,64,65]. Thus, their structural and mechanical properties were unlikely to have affected the pressure sensors. Finally, as all were time-zero studies, the pressure sensors could not accurately reflect the true behaviour of the meniscus, particularly once it has healed. Care should therefore be taken to extrapolate such findings in clinical practice, particularly as the sensors do not account for biological factors such as soft-tissue status, muscle contractions, proprioception, healing, and cartilage status. However, this limitation is more in keeping with the study’s methodological design than the sensor itself. There is no study to date comparing the common sensors described in this review on posterior meniscal root tears. Whilst one study in knee arthroplasty has compared Tekscan and Fujifilm [63], demonstrating more reliable and reproducible measurements with the Tekscan system, it is likely these results would correlate in root repair experiments. However, this requires further investigation to truly identify their differences in performance comparison to select the most suitable sensor.
Whilst most of the studies (21 out of 24) showed root repair to restore TFCM to that of the intact state, none were able to provide an analysis regarding how the repair directly influences the way the sensor collects the data. This requires further exploration.
The effect these laboratory data have on clinical implications is yet to be explored. Whilst wearable inertial measurement units and force-sensing sensors are used in practice to estimate joint kinematics and reaction force [66,67], introducing Tekscan, Fujifilm, or capacitive sensors within a real-life knee joint for dynamic simulation and loading is practically impossible. However, from a biomechanical perspective, the sensors do indirectly provide surgeons with evidence to argue for or against performing such repair techniques in clinical practice, particularly as the overall certainty of evidence from this review is high.
Further limitations are that majority of the studies did not provide in-depth analysis regarding how the sensors were able to detect pressure gradients, the computing of contact force, and the movement of the centre of pressure in flexion. Although this may have been beyond the scope of description for each individual study, such information would provide the reader an idea of the simplicity or complexity of the mathematical algorithm and coding sequences required to provide outcome data for the tibiofemoral contact mechanics.

5. Conclusions

Piezoresistive Tekscan and digital capacitive sensors are preferred for the collection of data during dynamic simulation of joint movements in intact, torn, and repaired posterior meniscal roots. Understanding the potential factors that influence such variables may improve accuracy, area, and pressure distribution measurements. The colour-based Fujifilm sensor is a passive technology that is ideal in simpler experiments with fewer cycles of load application.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/s25051507/s1, Table S1. Search strategy for Pubmed (Medline), Embase and Cochrane Central Register of Controlled Trials (15 January 2024). Table S2. Search strategy for Pubmed (Medline), Embase and Cochrane Central Register of Controlled Trials (10 November 2024). Table S3. MINORS Quality Appraisal Tool. NOTE. Each item was assigned a score of 0 points (not reported), 1 point (reported but inadequate), or 2 points (reported and adequate). MINORS, Methodological Index for Non-randomized Studies. Table S4. Quality of Included Studies Assessed by MINORS Quality Appraisal Tool. NOTE. Each item was assigned a score of 0 points (not reported), 1 point (reported but inadequate), or 2 points (reported and adequate), with a maximum possible score of 24. Item 1 indicates a clearly stated aim; item 2, inclusion of consecutive patients; item 3, prospective collection of data; item 4, endpoints appropriate to the aim of the study; item 5, unbiased assessment of the study endpoint; item 6, follow-up period appropriate to the aim of the study; item 7, loss to follow-up less than 5%; item 8, prospective calculation of the study size; item 9, an adequate control group; item 10, contemporary groups; item 11, baseline equivalence of groups; and item 12, adequate statistical analyses. Figure S1. Risk of bias summary. Red circle, high risk of bias; yellow circle, moderate risk of bias; green circle, low risk of bias. D1: Bias due to confounding data (selection bias), D2: bias in selection of participants into the study (selection bias), D3: bias in classification of interventions (information bias), D4: bias due to deviations from intended interventions (performance bias), D5: bias due to missing data (attrition data), D6: bias in measurement of outcomes (detection bias), D7: bias in selection of the reported result (outcome reporting bias). Figure S2. Risk of Bias graph.

Author Contributions

Conceptualisation, K.B., D.E.T.S., D.M.E. and T.B.; methodology, K.B., B.B., A.G., R.A. and T.B.; validation, K.B., B.B. and R.A.; writing—original draft, K.B.; writing—review and editing, K.B., D.E.T.S., D.M.E. and T.B.; supervision, D.E.T.S., D.M.E. and T.B. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the joint ORUK-RCSEd 2023 research fellowship.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created or analysed in this study.

Conflicts of Interest

The authors declare no conflict of interest.

Abbreviations

The following abbreviations are used in this manuscript:
ACLRanterior cruciate ligament reconstruction
ATPRanatomical transtibial pull-through repair
HCThorizontal cleavage tear
LFClateral femoral condyle
LMlateral meniscus
LMPHTlateral meniscus posterior horn tear
LMPRTlateral meniscus posterior root tear
MATmeniscal allograft transplantation
MCLmedial collateral ligament
MFCmedial femoral condyle
MFLmeniscofemoral ligament
MTMmaterials testing machine
MMmedial meniscus
MMAmodified Mason–Allen
MMPHmedial meniscus posterior horn
MMPRTmedial meniscus posterior root tear
MTLmeniscotibial ligament
NATPRnon-anatomic transtibial pull-through repair
OWHTOopen-wedge high tibial osteotomy
PCLposterior cruciate ligament
PMMApolymethylmethacrylate
PMRTposterior meniscus root tear
PRISMAPreferred Reporting Items for Systematic Review and Meta-Analyses
SARsuture anchor repair
TFCMtibiofemoral contact mechanics
TPRtranstibial pull-through repair
TSStwo simple stitch

References

  1. Allaire, R.; Muriuki, M.; Gilbertson, L.; Harner, C.D. Biomechanical consequences of a tear of the posterior root of the medial meniscus: Similar to total meniscectomy. J. Bone Jt. Surg. Am. 2008, 90, 1922–1931. [Google Scholar] [CrossRef] [PubMed]
  2. Hein, C.N.; Gurske-Deperio, J.; Ehrensberger, M.T.; Marzo, J.M. Effects of medial meniscal posterior horn avulsion and repair on meniscal displacement. Knee 2011, 18, 189–192. [Google Scholar] [CrossRef] [PubMed]
  3. Marzo, J.M.; Gurske-DePerio, J. Effects of medial meniscus posterior horn avulsion and repair on tibiofemoral contact area and peak contact pressure with clinical implications. Am. J. Sports Med. 2009, 37, 124–129. [Google Scholar] [CrossRef]
  4. Fairbank, T. Knee joint changes after meniscectomy. J. Bone Jt. Surg. Br. Vol. 1948, 30, 664–670. [Google Scholar] [CrossRef]
  5. Hunter, D.J.; Zhang, Y.Q.; Niu, J.B.; Tu, X.; Amin, S.; Clancy, M.; Guermazi, A.; Grigorian, M.; Gale, D.; Felson, D.T. The association of meniscal pathologic changes with cartilage loss in symptomatic knee osteoarthritis. Arthritis Rheum. 2006, 54, 795–801. [Google Scholar] [CrossRef]
  6. Moatshe, G.; Chahla, J.; Slette, E.; Engebretsen, L.; Laprade, R.F. Posterior meniscal root injuries. Acta Orthop. 2016, 87, 452–458. [Google Scholar] [CrossRef] [PubMed]
  7. Feucht, M.J.; Grande, E.; Brunhuber, J.; Rosenstiel, N.; Burgkart, R.; Imhoff, A.B.; Braun, S. Biomechanical evaluation of different suture materials for arthroscopic transtibial pull-out repair of posterior meniscus root tears. Knee Surg. Sports Traumatol. Arthrosc. 2015, 23, 132–139. [Google Scholar] [CrossRef]
  8. Bhatia, S.; LaPrade, C.M.; Ellman, M.B.; LaPrade, R.F. Meniscal root tears: Significance, diagnosis, and treatment. Am. J. Sports Med. 2014, 42, 3016–3030. [Google Scholar] [CrossRef]
  9. Furumatsu, T.; Kodama, Y.; Kamatsuki, Y.; Hino, T.; Okazaki, Y.; Ozaki, T. Meniscal extrusion progresses shortly after the medial meniscus posterior root tear. Knee Surg. Relat. Res. 2017, 29, 295–301. [Google Scholar] [CrossRef]
  10. Frank, J.M.; Moatshe, G.; Brady, A.W.; Dornan, G.J.; Coggins, A.; Muckenhirn, K.J.; Slette, E.L.; Mikula, J.D.; LaPrade, R.F. Lateral meniscus posterior root and meniscofemoral ligaments as stabilizing structures in the ACL-deficient knee: A biomechanical study. Orthop. J. Sports Med. 2017, 5, 2325967117695756. [Google Scholar] [CrossRef]
  11. Krych, A.J.; Bernard, C.D.; Kennedy, N.I.; Tagliero, A.J.; Camp, C.L.; Levy, B.A.; Stuart, M.J. Medial versus lateral meniscus root tears: Is there a difference in injury presentation, treatment decisions, and surgical repair outcomes? Arthrosc. J. Arthrosc. Relat. Surg. 2020, 36, 1135–1141. [Google Scholar] [CrossRef] [PubMed]
  12. Martinelli, L.; Hurschler, C.; Rosenbaum, D. Comparison of capacitive versus resistive joint contact stress sensors. Clin. Orthop. Relat. Res. 2006, 447, 214–220. [Google Scholar] [CrossRef] [PubMed][Green Version]
  13. Becher, C.; Huber, R.; Thermann, H.; Paessler, H.H.; Skrbensky, G. Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: A biomechanical study. Knee Surg. Sports Traumatol. Arthrosc. 2008, 16, 56–63. [Google Scholar] [CrossRef] [PubMed]
  14. Flanigan, D.C.; Harris, J.D.; Brockmeier, P.M.; Siston, R.A. The effects of lesion size and location on subchondral bone contact in experimental knee articular cartilage defects in a bovine model. Arthrosc. J. Arthrosc. Relat. Surg. 2010, 26, 1655–1661. [Google Scholar] [CrossRef]
  15. Hofer, J.K.; Gejo, R.; McGarry, M.H.; Lee, T.Q. Effects of kneeling on tibiofemoral contact pressure and area in posterior cruciate-retaining and posterior cruciate-sacrificing total knee arthroplasty. J. Arthroplast. 2012, 27, 620–624. [Google Scholar] [CrossRef]
  16. Kock, N.B.; Smolders, J.M.; van Susante, J.L.; Buma, P.; van Kampen, A.; Verdonschot, N. A cadaveric analysis of contact stress restoration after osteochondral transplantation of a cylindrical cartilage defect. Knee Surg. Sports Traumatol. Arthrosc. 2008, 16, 461–468. [Google Scholar] [CrossRef]
  17. Lee, S.J.; Aadalen, K.J.; Malaviya, P.; Lorenz, E.P.; Hayden, J.K.; Farr, J.; Kang, R.W.; Cole, B.J. Tibiofemoral contact mechanics after serial medial meniscectomies in the human cadaveric knee. Am. J. Sports Med. 2006, 34, 1334–1344. [Google Scholar] [CrossRef]
  18. Ohori, T.; Mae, T.; Shino, K.; Fujie, H.; Hirose, T.; Tachibana, Y.; Yoshikawa, H.; Nakata, K. Different effects of the lateral meniscus complete radial tear on the load distribution and transmission functions depending on the tear site. Knee Surg. Sports Traumatol. Arthrosc. 2021, 29, 342–351. [Google Scholar] [CrossRef]
  19. Ostermeier, S.; Holst, M.; Hurschler, C.; Windhagen, H.; Stukenborg-Colsman, C. Dynamic measurement of patellofemoral kinematics and contact pressure after lateral retinacular release: An in vitro study. Knee Surg. Sports Traumatol. Arthrosc. 2007, 15, 547–554. [Google Scholar] [CrossRef]
  20. Prisk, V.R.; Imhauser, C.W.; O’Loughlin, P.F.; Kennedy, J.G. Lateral ligament repair and reconstruction restore neither contact mechanics of the ankle joint nor motion patterns of the hindfoot. J. Bone Jt. Surg. Am. 2010, 92, 2375–2386. [Google Scholar] [CrossRef]
  21. Seo, J.H.; Li, G.; Shetty, G.M.; Kim, J.H.; Bae, J.H.; Jo, M.L.; Kim, J.S.; Lee, S.J.; Nha, K.W. Effect of repair of radial tears at the root of the posterior horn of the medial meniscus with the pullout suture technique: A biomechanical study using porcine knees. Arthrosc. J. Arthrosc. Relat. Surg. 2009, 25, 1281–1287. [Google Scholar] [CrossRef]
  22. von Lewinski, G.; Hurschler, C.; Allmann, C.; Wirth, C.J. The influence of pre-tensioning of meniscal transplants on the tibiofemoral contact area. Knee Surg. Sports Traumatol. Arthrosc. 2006, 14, 425–436. [Google Scholar] [CrossRef]
  23. Liberati, A.; Altman, D.G.; Tetzlaff, J.; Mulrow, C.; Gøtzsche, P.C.; Ioannidis, J.P.; Clarke, M.; Devereaux, P.J.; Kleijnen, J.; Moher, D. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: Explanation and elaboration. Bmj 2009, 339, b2700. [Google Scholar] [CrossRef] [PubMed]
  24. Slim, K.; Nini, E.; Forestier, D.; Kwiatkowski, F.; Panis, Y.; Chipponi, J. Methodological index for non-randomized studies (minors): Development and validation of a new instrument. ANZ J. Surg. 2003, 73, 712–716. [Google Scholar] [CrossRef] [PubMed]
  25. Sterne, J.A.; Hernán, M.A.; Reeves, B.C.; Savović, J.; Berkman, N.D.; Viswanathan, M.; Henry, D.; Altman, D.G.; Ansari, M.T.; Boutron, I.; et al. ROBINS-I: A tool for assessing risk of bias in non-randomised studies of interventions. BMJ 2016, 355, i4919. [Google Scholar] [CrossRef] [PubMed]
  26. Brignardello-Petersen, R.; Guyatt, G.H. Assessing the Certainty of the Evidence in Systematic Reviews: Importance, Process, and Use. Am. J. Epidemiol. 2024, 31, kwae332. [Google Scholar] [CrossRef]
  27. Amano, Y.; Ozeki, N.; Matsuda, J.; Nakamura, T.; Nakagawa, Y.; Sekiya, I.; Koga, H. Augmentation of a Nonanatomical Repair of a Medial Meniscus Posterior Root Tear with Centralization Using Three Knotless Anchors May Be Associated with Less Meniscal Extrusion and Better Compressive Load Distribution in Mid-Flexion Compared with Non-Anatomical Root Repair Alone in a Porcine Knee Model. Arthrosc. J. Arthrosc. Relat. Surg. 2023, 39, 2487–2498. [Google Scholar]
  28. Baratz, M.E.; Fu, F.H.; Mengato, R. Meniscal tears: The effect of meniscectomy and of repair on intraarticular contact areas and stress in the human knee. A preliminary report. Am. J. Sports Med. 1986, 14, 270–275. [Google Scholar] [CrossRef]
  29. Chung, K.S.; Choi, C.H.; Bae, T.S.; Ha, J.K.; Jun, D.J.; Wang, J.H.; Kim, J.G. Comparison of Tibiofemoral Contact Mechanics After Various Transtibial and All-inside Fixation Techniques for Medial Meniscus Posterior Root Radial Tears in a Porcine Model. Arthrosc. J. Arthrosc. Relat. Surg. 2018, 34, 1060–1068. [Google Scholar] [CrossRef]
  30. Daney, B.T.; Aman, Z.S.; Krob, J.J.; Storaci, H.W.; Brady, A.W.; Nakama, G.; Dornan, G.J.; Provencher, M.T.; LaPrade, R.F. Utilization of Transtibial Centralization Suture Best Minimizes Extrusion and Restores Tibiofemoral Contact Mechanics for Anatomic Medial Meniscal Root Repairs in A Cadaveric Model. Am. J. Sports Med. 2019, 47, 1591–1600. [Google Scholar] [CrossRef]
  31. Doan, K.; Dey Hazra, M.E.D.; Brown, J.R.; Hollenbeck, J.F.M.; Dey Hazra, R.O.; Johnson, D.; Fossum, B.; Vidal, A. Biomechanical Analysis of Meniscotibial Ligament Tenodesis to Treat Meniscal Extrusion in the Setting of Posterior Medial Meniscus Root Repair. Am. J. Sports Med. 2023, 51, 3204–3210. [Google Scholar] [CrossRef] [PubMed]
  32. Forkel, P.; Herbort, M.; Sprenker, F.; Metzlaff, S.; Raschke, M.; Peterson, W. The Biomechanical Effect of a Lateral Meniscus Posterior Root Tear with and Without Damage to the Meniscofemoral Ligament: Efficacy of Different Repair Techniques. Arthrosc. J. Arthrosc. Relat. Surg. 2014, 30, 833–840. [Google Scholar] [CrossRef]
  33. Geeslin, A.G.; Civitarese, D.; Turnbull, T.L.; Dornan, G.J.; Fuso, F.A.; LaPrade, R.F. Influence of lateral meniscal posterior root avulsions and the meniscofemoral ligaments on tibiofemoral contact mechanics. Knee Surg. Sports Traumatol. Arthrosc. 2016, 24, 1469–1477. [Google Scholar] [CrossRef]
  34. Gupta, A.; Sanchez, M.; Storaci, H.W.; Rohde, M.S.; Sherman, S.L.; Shea, K.G. Transtibial Repair of Lateral Meniscus Posterior Root Tears Improves Contact Biomechanics in Pediatric Cadavers. Arthrosc. Sports Med. Rehabil. 2022, 4, E2011–E2018. [Google Scholar] [CrossRef] [PubMed]
  35. Kim, J.G.; Lee, Y.S.; Bae, T.S.; Ha, J.K.; Lee, D.H.; Kim, Y.J.; Ra, H.J. Tibiofemoral contact mechanics following posterior root of medial meniscus tear, repair, meniscectomy, and allograft transplantation. Knee Surg. Sports Traumatol. Arthrosc. 2013, 21, 2121–2125. [Google Scholar] [CrossRef] [PubMed]
  36. Koh, J.L.; Yi, S.J.; Ren, Y.; Zimmerman, T.A.; Zhang, L.Q. Tibiofemoral Contact Mechanics with Horizontal Cleavage Tear and Resection of the Medial Meniscus in the Human Knee. J. Bone Jt. Surg. Am. 2016, 98, 1829–1836. [Google Scholar] [CrossRef]
  37. LaPrade, C.M.; Foad, A.; Smith, S.D.; Turnbull, T.L.; Dornan, G.J.; Engebretsen, L.; Wijdicks, C.A.; LaPrade, R.F. Biomechanical consequences of a nonanatomic posterior medial meniscal root repair. Am. J. Sports Med. 2015, 43, 912–920. [Google Scholar] [CrossRef]
  38. LaPrade, C.M.; Jansson, K.S.; Dornan, G.; Smith, S.D.; Wijdicks, C.A.; LaPrade, R.F. Altered tibiofemoral contact mechanics due to lateral meniscus posterior horn root avulsions and radial tears can be restored with in situ pull-out suture repairs. J. Bone Jt. Surg. Am. 2014, 96, 471–479. [Google Scholar] [CrossRef]
  39. Muriuki, M.G.; Tuason, D.A.; Tucker, B.G.; Harner, C.D. Changes in tibiofemoral contact mechanics following radial split and vertical tears of the medial meniscus an in vitro investigation of the efficacy of arthroscopic repair. J. Bone Jt. Surg. Am. 2011, 93, 1089–1095. [Google Scholar] [CrossRef]
  40. Padalecki, J.R.; Jansson, K.S.; Smith, S.D.; Dornan, G.J.; Pierce, C.M.; Wijdicks, C.A.; Laprade, R.F. Biomechanical consequences of a complete radial tear adjacent to the medial meniscus posterior root attachment site: In situ pull-out repair restores derangement of joint mechanics. Am. J. Sports Med. 2014, 42, 699–707. [Google Scholar] [CrossRef]
  41. Park, H.J.; Chang, M.J.; Cho, H.J.; Hong, E.; Kim, T.W.; Chang, C.B.; Bae, T.S.; Kwak, D.S.; Kang, S.B. Medial meniscus posterior root repair restores contact pressure and contact area to its native state even after opening-wedge high tibial osteotomy: A cadaveric biomechanical study. Arthrosc. J. Arthrosc. Relat. Surg. 2023, 39, 638–646. [Google Scholar] [CrossRef] [PubMed]
  42. Pasic, N.; Storaci, H.; Guzman, R.; Debaun, M.; Maruyama, M.; Hall, K.; Salazar, B.P.; Dragoo, J.L. A Biomechanical Comparison of All-Inside Versus Transtibial Meniscus Root Repair Techniques. Am. J. Sports Med. 2023, 51, 2366–2373. [Google Scholar] [CrossRef]
  43. Perez-Blanca, A.; Espejo-Baena, A.; Trujillo, D.A.; Nóvoa, M.P.; Espejo-Reina, A.; López, C.Q.; Juanco, F.E. Comparative Biomechanical Study on Contact Alterations After Lateral Meniscus Posterior Root Avulsion, Transosseous Reinsertion, and Total Meniscectomy. Arthrosc. J. Arthrosc. Relat. Surg. 2016, 32, 624–633. [Google Scholar] [CrossRef]
  44. Saengpetch, N.; Noowan, S.; Boonrod, A.; Jaruwanneechai, K.; Sumanont, S.; Vijittrakarnrung, C. Comparison of medial tibiofemoral joint mechanics between all-suture anchors and transtibial pullout technique for posterior medial meniscal root tears. J. Orthop. Surg. Res. 2023, 18, 591. [Google Scholar] [CrossRef] [PubMed]
  45. Saltzman, B.M.; Habet, N.A.; Rao, A.J.; Trofa, D.P.; Corpus, K.T.; Yeatts, N.C.; Odum, S.M.; Garcia, J.C.C.; Varkey, D.T.; Piasecki, D.P.; et al. Biomechanical Evaluation of an All-Inside Posterior Medial Meniscal Root Repair Technique Via Suture Fixation to the Posterior Cruciate Ligament. Arthrosc. J. Arthrosc. Relat. Surg. 2020, 36, 2488–2497. [Google Scholar] [CrossRef] [PubMed]
  46. Schillhammer, C.K.; Werner, F.W.; Scuderi, M.G.; Cannizaro, J.P. Repair of Lateral Meniscus Posterior Horn Detachment Lesions A Biomechanical Evaluation. Am. J. Sports Med. 2012, 40, 2604–2609. [Google Scholar] [CrossRef]
  47. Zhang, Z.Z.; Luo, H.; Zhang, H.Z.; Zhou, Y.F.; Chen, Z.; Jiang, C.; Song, B.; Li, W.P. H-Plasty Repair Technique Improved Tibiofemoral Contact Mechanics After Repair for Adjacent Radial Tears of Posterior Lateral Meniscus Root: A Biomechanical Study. Arthrosc. J. Arthrosc. Relat. Surg. 2021, 37, 2204–2216. [Google Scholar] [CrossRef]
  48. Wirz, D.; Becker, R.; Li, S.F.; Friederich, N.F.; Müller, W. Validation of the Tekscan system for statistic and dynamic pressure measurements of the human femorotibial joint. Biomed. Tech. Biomed. Eng. 2002, 47, 195–201. [Google Scholar] [CrossRef]
  49. Chen, T.; Pekmezian, A.; Leatherman, E.R.; Santner, T.J.; Maher, S.A. Tekscan analysis programs (TAP) for quantifying dynamic contact mechanics. J. Biomech. 2022, 136, 111074. [Google Scholar] [CrossRef]
  50. Wilson, D.R.; Apreleva, M.V.; Eichler, M.J.; Harrold, F.R. Accuracy and repeatability of a pressure measurement system in the patellofemoral joint. J. Biomech. 2003, 36, 1909–1915. [Google Scholar] [CrossRef]
  51. Jansson, K.S.; Michalski, M.P.; Smith, S.D.; LaPrade, R.F.; Wijdicks, C.A. Tekscan pressure sensor output changes in the presence of liquid exposure. J. Biomech. 2013, 46, 612–614. [Google Scholar] [CrossRef]
  52. Wang, H.; Chen, T.; Gee, A.O.; Hutchinson, I.D.; Stoner, K.; Warren, R.F.; Rodeo, S.A.; Maher, S.A. Altered regional loading patterns on articular cartilage following meniscectomy are not fully restored by autograft meniscal transplantation. Osteoarthr. Cartil. 2015, 23, 462–468. [Google Scholar] [CrossRef] [PubMed][Green Version]
  53. Wu, J.Z.; Herzog, W.; Epstein, M. Effects of inserting a pre-sensor film into articular joints on the actual contact mechanics. J. Biomech. Eng. 1998, 120, 655–659. [Google Scholar] [CrossRef]
  54. Anderson, I.A.; MacDiarmid, A.A.; Harris, M.L.; Gillies, R.M.; Phelps, R.; Walsh, W.R. A novel method for measuring medial compartment pressures within the knee joint in-vivo. J. Biomech. 2003, 36, 1391–1395. [Google Scholar] [CrossRef] [PubMed]
  55. Martens, T.A.; Hull, M.L.; Howell, S.M. An in vitro osteotomy method to expose the medial compartment of the human knee. J. Biomech. Eng. 1997, 119, 379–385. [Google Scholar] [CrossRef] [PubMed]
  56. Agins, H.J.; Harder, V.S.; Lautenschlager, E.P.; Kudrna, J.C. Effects of sterilization on the Tekscan digital pressure sensor. Med. Eng. Phys. 2003, 25, 775–780. [Google Scholar] [CrossRef] [PubMed]
  57. Wilharm, A.; Hurschkler, C.; Dermitas, T.; Bohnsack, M. Use of Tekscan K-scan sensors for retropatellar pressure measurement avoiding errors during implantation and the effects of shear forces on the measurement precision. BioMed Res. Int. 2013, 829171. [Google Scholar] [CrossRef]
  58. Bertsch, C.; Rosenbaum, D.; Claes, L. Intra-articular and plantar pressure distribution of the ankle joint complex in relation to foot position. Unfallchirurg 2001, 104, 426–433. [Google Scholar]
  59. Chen, Z.; Zhang, H.; Luo, H.; Yang, R.; Zhang, Z.; Jiang, C.; Hou, J.; Zhou, Y.; Xu, Y.; Song, B.; et al. Contact mechanics after mattress suture repair of medial meniscus vertical longitudinal tear: An in vitro study. Arch. Orthop. Trauma Surg. 2020, 140, 1221–1230. [Google Scholar] [CrossRef]
  60. Rosenbaum, D.; Eils, E.; Hillmann, A. Changes in talocrural joint contact stress characteristics after simulated rotationplasty. J. Biomech. 2003, 36, 81–86. [Google Scholar] [CrossRef]
  61. Cederlund, A.A.; Aspden, R.M. Walking on water: Revisiting the role of water in articular cartilage biomechanics in relation to tissue engineering and regenerative medicine. J. R. Soc. Interface 2022, 19, 20220364. [Google Scholar] [CrossRef] [PubMed]
  62. Liggins, A.B.; Hardie, W.R.; Finlay, J.B. The spatial and pressure resolution of fuji pressure-sensitive film. Exp. Mech. 1995, 35, 166–173. [Google Scholar] [CrossRef]
  63. Harris, M.L.; Morberg, P.; Bruce, W.J.; Walsh, W.R. An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: A comparison of K-scan sensor and Fuji film. J. Biomech. 1999, 32, 951–958. [Google Scholar] [CrossRef]
  64. Barber, F.A.; Herbert, M.A. All-suture anchors: Biomechanical analysis of pullout strength, displacement, and failure mode. Arthrosc. J. Arthrosc. Relat. Surg. 2017, 33, 1113–1121. [Google Scholar] [CrossRef]
  65. Wu, S.H.; Yeh, T.T.; Hsu, W.C.; Wu, A.T.H.; Li, G.; Chen, C.H.; Lee, C.H.; Wu, J.L. Biomechanical comparison of four tibial fixation techniques for meniscal root sutures in posterior medial meniscus root repair: A porcine study. J. Orthop. Transl. 2020, 24, 144–149. [Google Scholar] [CrossRef] [PubMed]
  66. Renggli, D.; Graf, C.; Tachatos, N.; Singh, N.; Meboldt, M.; Taylor, W.R.; Stieglitz, L.; Schmid-Daners, M. Wearable Inertial Measurement Units for Assessing Gait in Real-World Environments. Front. Physiol. 2020, 11, 90. [Google Scholar] [CrossRef]
  67. Crescini, D.; Sardini, E.; Serpelloni, M. Design and test of an autonomous sensor for force measurements in human knee implants. Sens. Actuators A Phys. 2011, 166, 1–8. [Google Scholar] [CrossRef]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.