1. Introduction
Parkinson’s disease (PD) is a progressive neurodegenerative disorder characterized primarily by the degeneration of dopaminergic neurons within the substantia nigra pars compacta. This neuronal loss results in a marked reduction in dopamine synthesis and availability, disrupting basal ganglia circuits critical for motor control. Consequently, patients exhibit a spectrum of motor and non-motor symptoms [
1]. Impaired ability to walk and maintain balance is among the most impactful motor disorders [
1]. In PD patients, walking is often slow, with small, dragging steps, which impair the individual’s postural capacity and increase their risk of falling [
1].
As the disease progresses, additional symptoms may occur. One of the most debilitating is Freezing of Gait (FoG), defined by sudden episodes of cessation or inability to start walking [
2]. This phenomenon can occur in a variety of situations, such as at the beginning of the walking task, in overcoming narrow spaces or obstacles, during turning, or during dual-tasking [
3,
4]. This significantly impacts patients’ mobility, further increasing the risk of falling and reducing their quality of life [
2,
5]. Targeted rehabilitation interventions focusing on specific gait parameters have demonstrated potential to alleviate FoG in PD [
6,
7]. Thus, detailed kinematic characterization of gait in patients with FoG, compared to patients without FoG and healthy controls, is crucial for developing tailored rehabilitation strategies and objectively assessing treatment effectiveness [
1].
The analysis of the duration and length of the stride, its elevation, speed, cadence, and distribution of the different phases of the stride can objectively assess patients’ motor impairment when compared to healthy individuals [
8]. Measuring these parameters also allows for monitoring the progression of the disease, assessing the effectiveness of therapies, and developing more precise, patient-focused clinical treatment plans. The number of studies performing this quantitative analysis is rapidly growing, leveraging advanced approaches and technologies such as wearable sensors and machine learning to better understand the gait characteristics of Parkinsonian patients and FoG [
8,
9,
10,
11,
12,
13,
14,
15,
16,
17,
18,
19,
20,
21,
22,
23,
24,
25]. In 2024, Tang et al. [
25] analyzed gait characteristics in Parkinsonian patients using wearable sensors. In their study, the researchers compared a group of PD patients who were exclusively in the OFF state of dopaminergic therapy, that is, in the absence of drug administration, with a sample of healthy individuals. The results showed reduced motor abilities of PD patients compared with healthy subjects. Although FoG was mentioned in the results, no specific data on patients who presented with the symptom were included. Additionally, among the spatiotemporal gait parameters measured, foot clearance, which plays a very important role in fall risk, is not present. Similarly, in 2017, Schlachetzki et al. [
10] performed a gait analysis, using two inertial measurement units (IMUs) placed on the side of the foot, measuring several spatiotemporal parameters. In this instance, clearance was also measured, and it was found to have a reduced amplitude compared to the healthy control group. However, the study’s primary focus did not center on subjects with FoG.
Nevertheless, a few studies have made comparisons between patients with FoG and those without. In 2024, Lin et al. [
21] employed a Kinect to ascertain the spatiotemporal parameters of patients with FoG, with the objective of evaluating their responsiveness to dopaminergic therapy. The findings indicated that patients with FoG exhibited a higher prevalence of motor complications and diminished parameters in comparison to subjects without FoG. This discrepancy is mitigated by the administration of the pharmaceutical agent. However, the issue of clearance remains unaddressed in their work. A comparable study by Zhao et al. [
22] proceeded to analyze parameters specifically in the interictal period of FoG patients, obtaining similar results to those of Lin et al., again with no consideration for clearance.
The analysis of spatiotemporal parameters has been demonstrated to play a crucial role in the classification of patients with FoG, as evidenced by the studies conducted by [
15,
20]. Zampogna et al. [
20] conducted a continuous analysis of PD patients in their homes for a period of approximately one week. This approach facilitated the acquisition of information regarding activities of daily living. The objective of this study was to differentiate between patients with and those without FoG. However, the results pertaining to spatiotemporal parameters demonstrated no statistically significant differences between the two groups. In contrast, Bouchouras et al. [
15] employed a machine learning algorithm, a Random Forest, to differentiate between FoG subjects both in the OFF and ON states of therapy. The algorithm demonstrated a higher degree of efficacy in classifying FoG during the OFF state compared to the ON one. However, in both cases, the authors do not consider clearance in their analysis. This observation underscores a conspicuous need for the analysis of this parameter.
Furthermore, it is imperative to obtain validation of the systems employed that deviate from the gold standard, such as IMUs, to ensure the accuracy of the measurements made. This is particularly salient for parameters that are closely related to fall risk factors, such as clearance, which represents the elevation of the foot during ambulation. Despite the numerous validations conducted on spatiotemporal parameters associated with walking [
26,
27,
28,
29], to the best of the authors’ knowledge, clearance has not been a consideration in any of these studies. In 2024, He et al. [
26] and, in 2025, Boutaayamou et al. [
27] independently validated IMUs for measuring spatiotemporal parameters, excluding clearance, in healthy subjects. For patients with PD, similar research was conducted by Alberto et al. [
29] in 2021, who validated a system comprising 15 IMUs for use in both supervised and unsupervised environments. Additionally, Yang et al. [
28] in 2022 validated a system consisting of 5 IMUs for detecting early-stage PD during turns. Consistent with prior research on healthy individuals, neither study validated IMUs for foot clearance. Indeed, this parameter is frequently validated on systems using combinations of multiple techniques or sensors to maximize accuracy [
18,
30]. In 2017, Arami et al. [
30] validated a clearance measurement system that used an IMU to support an infrared sensor for measuring clearance within a range of 4 to 30 cm. Conversely, the IMU’s primary function was to identify the foot inclination. In contrast, in 2024, Fehr et al. [
18] validated a clearance measurement using a single IMU, tasked with measuring the parameter, combining it with a 3D scan of the foot that reconstructed its geometry. Consequently, this method enabled the identification of the lowest point of the foot during the swing phase. However, these studies do not utilize only a single IMU, a choice that would offer significant advantages in terms of comfort and simplicity when monitoring real-life situations. This has led to the necessity of investigating the accuracy of IMUs for clearance measurement.
This study describes the analysis of several spatiotemporal parameters obtained using IMUs, as an extension of a previous study [
1], and pursues two main objectives.
The first objective is to assess motor function in individuals diagnosed with PD by comparing them with a healthy control group. Emphasis was placed on the phenomenon of FoG, with the subdivision of the patient sample into two distinct groups: experiencing and non-experiencing FoG as a PD symptom. An analysis of the differences in the parameters among healthy subjects, Parkinsonian subjects without FoG, and those with FoG was conducted to ascertain further motor impairments caused by this symptom. An experimental campaign of gait task acquisitions was conducted, both during the OFF state of therapy, i.e., in the absence of dopaminergic drug administration, and the ON state, i.e., during the effect of the dopaminergic drug. A series of spatiotemporal parameters was extracted from kinematic data acquired through IMUs and compared across different groups through a statistical study. This study aims to broaden knowledge regarding the phenomenon of FoG by including clearance, which is often not considered in overall analyses of spatiotemporal parameters. It also aims to make a comparison between different groups that include both healthy subjects and different classifications of PD, going on to distinguish between subjects with FoG and without FoG in OFF and ON states.
The second objective is to introduce and validate a methodology for measuring foot clearance using IMUs, building upon the parameter validation performed by Mileti et al. [
24]. A comprehensive assessment of the accuracy of the IMU-based foot clearance estimation is provided, including a comparison with measurements obtained through a gold-standard optoelectronic motion capture system. This approach contributes to the validation of inertial systems by leveraging an IMU placed on the foot, thereby addressing the limitations identified in previous studies on spatiotemporal parameters. Additionally, it eliminates variables associated with the integration of additional sensors or systems that support the IMU, ensuring a more streamlined and comprehensive methodology.
2. Materials and Methods
2.1. Subjects
A total of 30 subjects diagnosed with PD were enrolled in this study (25 males and 5 females; mean age: 69.8 ± 8.9 years). Of these, 15 participants did not exhibit FoG (13 males and 2 females; mean age: 69.4 ± 9.8 years), while the remaining 15 participants demonstrated FoG (12 males and 3 females; mean age: 70.1 ± 8.2 years). The duration of the disease and its severity, as identified by clinical scales, were not subject to any restrictions. Additionally, patients who did not have overly debilitating gait changes were selected.
The control group consisted of 20 healthy individuals matched for age and sex distribution (14 males and 6 females; mean age: 69.9 ± 7.2 years) who were recruited. The number of subjects was selected following a preliminary power analysis. A one-way ANOVA and a Fisher’s exact test were performed to determine the homogeneity of the groups in terms of age and sex proportions.
Patients diagnosed with PD were evaluated both in the pharmacological OFF state, defined as at least 12 h after the withdrawal of dopaminergic therapy, and in the pharmacological ON state, assessed one hour after the administration of 150% of their usual L-Dopa dosage. The study was conducted according to the guidelines of the Declaration of Helsinki and was approved by the Ethics Committee of Sapienza University of Rome (protocol code 0372/2022 and date of approval 4 May 2022).
Table 1 reports a summary of the demographic and clinical characteristics of the subjects.
2.2. Experimental Setup
The experimental setup involved the use of 2 MTw triaxial IMUs (Xsens, Movella Inc., Henderson, NV, USA) placed on the feet through elastic Velcro bands. Data from the IMUs were collected using an Awinda station connected to a PC, with a sampling rate of 100 Hz via the MT Manager 4.6 software (Movella Inc., Henderson, NV, USA). All tests were conducted along a straight, unobstructed corridor approximately 20 m in length, allowing the acquisition of 40 test strides per participant. A schematic representation of the experimental setup is provided in
Figure 1.
2.3. Experimental Protocol
For PD patients, a brief interview and preliminary health assessments were conducted prior to testing, to evaluate their ability to safely perform the walking test in the OFF state. Only those who met the safety criteria proceeded with the full experimental protocol. This preliminary screening step was not necessary for healthy control subjects.
IMUs were positioned on the instep of each foot and affixed through elastic Velcro bands. Subsequently, the sensors were connected to the acquisition software via the Awinda station. The participant was escorted to the commencement of the corridor and requested to traverse it at a self-selected speed.
The test protocol entailed a direct trajectory traversing the entire 20 m length of the designated corridor. A visual marker was strategically placed at the end of the corridor to indicate the turning point. The subject was required to walk around this marker without interruption and subsequently return to the initial starting position. This mode resulted in a total straight walk length of approximately 40 m per subject. Only strides from the straight walk were utilized for analysis. For patients diagnosed with Parkinson’s disease, a physician was present to ensure their safety in the event of falls during the test. The experiment was replicated four times for each subject, who was requested to turn around twice to the left and twice to the right.
The trial was then replicated in a similar manner in the ON state of therapy, with a total of eight trials conducted for each patient. At the conclusion of each trial, data comprising accelerometer, gyroscope, magnetometer, and quaternion values were stored on the computer at a sampling rate of 100 Hz in a .txt format for subsequent analysis.
2.4. Data Analysis
The data collected from the IMUs were processed using MATLAB™ 2025a software (MathWorks Inc., Natick, MA, USA). Gyroscope data underwent filtration with a Butterworth filter of order 2, with a cutoff frequency of 3 Hz. Similarly, accelerometer data was filtered with a Butterworth filter of order 2, with a cutoff frequency of 10 Hz. Ten central strides of the straight walk were selected for going and returning, with a total of 20 strides per trial. This avoided issues related to acceleration and deceleration that occur during the commencement and cessation of the walk and the turning. To analyze the differences between patients with FoG and those without in their daily lives, regardless of the occurrence of the symptom, trials that presented FoG events were excluded from the analysis.
The following subsections delineate the methodology employed in deriving the spatiotemporal parameters examined in this study. For each spatiotemporal parameter, the mean and standard deviation were calculated for each of the following groups: healthy control (HC), overall parkinsonian patients in the OFF (PDoff) state, overall parkinsonian patients in the ON (PDon) state, parkinsonian patients without FoG in the OFF (NOFOGoff) state, parkinsonian patients without FoG in the ON (NOFOGon) state, parkinsonian patients with FoG in the OFF (FOGoff) state, and parkinsonian patients with FoG in the ON (FOGon) state.
2.4.1. Stride Time Detection
Angular velocity in the sagittal plane (i.e., calculated on the body’s frontal axis, corresponding to the y-axis of the sensor) was used to identify key gait events, including heel strike (HS), toe strike (TS), heel off (HO), and toe off (TO). Following the approach described by Mileti et al. [
31], HS and TO events were derived by identifying the second and first maximum points of a stride cycle, respectively. Meanwhile, TS and HO were derived by identifying the instants within the stance phase at which the absolute value of angular velocity crossed the 30°/s threshold. Accordingly, the stride time (ST) for each stride, calculated in seconds, could be derived by calculating the time elapsed between two consecutive HS events.
Figure 2 shows a representative angular velocity trend, with the identified gait events marked.
2.4.2. Gait Phases Detection
Based on the instants identified in the previous analysis, the different gait phases (GPs) were segmented. A four-phase model was adopted, including loading response (LR), defined as the interval between the HS and the subsequent TS; midstance (MS), defined as the flat foot (FF) period from TS to HO; pre-swing (PS), defined as the interval between HO and TO; and swing (SW), defined as the period between TO and the subsequent HS that initiates the next stride. Each phase was expressed as a percentage of the total stride time (ST) using the formula:
where
identifies the time of the gait phase studied and
is the total stride time.
2.4.3. Cadence Detection
Cadence (CA) was calculated as the number of strides, determined from successive HS events by the total gait duration (in minutes). The result was rounded down to the nearest integer.
2.4.4. Gait Phases Quality Index Detection
The Gait Phases Quality Index (GPQI) is an index initially introduced by Mileti et al. [
32] that provides a summary quantification of gait quality. It quantifies the discrepancy between the subject’s GP sequence and a reference sequence obtained from the mean values of healthy subjects. GPQI was computed as
where
identifies the foot (right or left),
,
,
and
are the GPs of the subject, and
,
,
and
are the GPs of the reference sequence described previously in
Section 2.4.2.
A lower GPQI indicates a gait phase distribution more similar to the reference, and therefore, higher gait quality.
2.4.5. Stride Length Detection
The stride length (SL), in meters, was obtained by the Zero Velocity Update (ZUPT) algorithm. Initially, the acceleration data obtained from the accelerometer trials were rotated in the global coordinate system, and the component due to gravitational acceleration was removed. The rotation matrix was derived from the quaternions. Subsequently, the intervals relative to the flat foot, obtained by measuring the period between one TS and the subsequent HO, were set to zero. Then, the velocity was computed by integrating the acceleration on the three axes within the specified interval between HO and TS, employing the following formula:
where
and
are the velocity and the acceleration in the global coordinate system. Subsequently, a linear detrend was implemented on the velocity to eradicate any drift that might have ensued from the integration of acceleration.
To obtain the subject’s stride length, the modulus of horizontal velocity was calculated with the following formula:
where
is the horizontal velocity and
and
are the velocity components in the x and y axes belonging to the horizontal plane.
Once the horizontal velocity of the subject was obtained, the flat foot intervals were again set to zero and the integration of the horizontal velocity was repeated to obtain the horizontal displacement:
where
is the horizontal displacement and
is the horizontal velocity.
The stride length is defined as the maximum value of the horizontal displacement obtained between two successive instants of HS.
An example of the process to obtain stride length is shown in
Figure 3.
2.4.6. Speed Detection
Walking speed (SP), calculated in m/s, was obtained by dividing the SL of each individual stride by the corresponding ST.
2.4.7. Clearance Detection
The clearance (CL), calculated in mm, is defined as the vertical displacement of the foot in the sagittal plane. Its estimation followed a similar approach to that of SL, but only the vertical component of the acceleration signal was considered. Due to the unknown initial height of the foot from the ground, clearance was obtained as a relative vertical displacement trend.
The key parameters extracted from the CL signal were the first local maximum (Max1), the second local maximum (Max2), and the point of local minimum (Min) that lies between Max1 and Max2. These events represent the foot’s trajectory during swing. A graphical example is shown in
Figure 4.
2.4.8. Statistical Analysis
A statistical analysis was conducted to ascertain whether there were any substantial disparities in spatiotemporal parameters among the various groups. The groups compared are HC vs. PDoff, HC vs. PDon, HC vs. NOFOGoff, HC vs. NOFOGon, HC vs. FOGoff, HC vs. FOGon, NOFOGoff vs. FOGoff, and NOFOGon vs. FOGon.
Initially, the normality of the parameters was ascertained through a Shapiro-Wilk test. ST, GP, CA, GPQI, and SL exhibited non-normal distribution, while SP and CL (Max1, Max, and Min) demonstrated normal distribution. Subsequently, for parameters that did not conform to a normal distribution, a nonparametric Mann–Whitney U test was performed for each comparison, and the relative p-value was obtained. Conversely, for the determination of standard parameters, an independent-samples t-test was conducted for each comparison, yielding the respective p-value.
2.5. Clearance’s Accuracy Validation
The validation of the measurement of the spatial parameter of SL from IMU data has been previously addressed in the work of Mileti et al. [
33]. In this study, we focused on validating the CL measurement using a similar methodology.
2.5.1. Subjects
To validate the IMU-based CL estimation, a group of 3 healthy subjects (2 males and 1 female; mean age: 32 ± 8.2 years) was recruited. The sole stipulation was the absence of gait-related complications.
2.5.2. Experimental Setup
The validation was executed using the same IMUs employed for spatiotemporal parameter analysis. A Vicon Nexus optoelectronic system (Vicon Motion Systems Ltd., Oxford, UK) with six Vicon IR Vero 2.2 cameras was utilized as a reference for validation. In the present study, inertial measurement units (IMUs) were strategically positioned on the plantar surfaces of both feet. Two reflective markers were placed on the lateral parts of the IMUs to follow the same vertical trajectory in the sagittal plane of the inertial unit. Calibration of the optoelectronic system yielded a capture volume with a surface area of 6 × 2 m. The data provided by the IMUs were acquired using the proprietary MVN Analyze Pro 2025.0 software, while the data from the cameras were acquired using the proprietary Vicon Nexus 2.17 software. The acquisitions were synchronized via a trigger cable.
2.5.3. Experimental Protocol
Each subject performed a straight-line walk of approximately six meters, repeated for ten trials. During each trial, IMU data, including accelerometer, gyroscope, magnetometer, and quaternion signals, were recorded, along with the spatial coordinates of the reflective markers tracked by the optoelectronic system. All data were stored on a PC for post-processing in MATLAB 2025a.
2.5.4. Data Analysis
The collected data were processed using MATLAB 2025a software. Vertical displacement of the foot-mounted IMUs was derived from the vertical component of acceleration in the global coordinate system. This estimation was performed using a Zero Velocity Update (ZUPT) algorithm, as previously described (
Figure 5).
To evaluate the accuracy of the IMU-based clearance measurements, individual strides were segmented and compared against those obtained from the optoelectronic system. Therefore, the clearance trend from the optoelectronic system was transformed into a displacement difference by subtracting its absolute minimum value from the vector. For each trial of each subject, three strides were extracted for a total of 30 strides per subject.
Figure 6 presents an example of the average clearance trends obtained from the two systems for one subject.
Once the strides were divided, the clearance points of Max1, Max2, and Min were calculated for both systems. For each point, the absolute error and absolute relative percentage error were calculated according to the following formulas:
where
is the absolute error of the single stride,
is the absolute relative percentage error,
is the point of clearance measured by the optoelectronic system, and
is the same point measured by the IMU.
Then the mean values of both errors were calculated for each subject among all strides. Finally, the mean absolute error and mean absolute relative percentage error among all subjects with relative standard deviations were calculated.
To assess the agreement between the two measurement methods, a Bland–Altman plot with a 95% confidence interval for each clearance parameter was generated.
4. Discussion
This section discusses the main findings of the study considering existing literature and clinical implications. The discussion is structured in two parts: the first focuses on the interpretation of spatiotemporal gait alterations observed in Parkinson’s disease (PD) and Freezing of Gait (FoG), while the second addresses the validation and clinical relevance of foot clearance estimation using wearable IMUs.
4.1. Interpretation of Spatiotemporal Gait Alterations in PD and FoG
Main findings reported in
Table 2 can be summarized as follows:
- -
Significant differences in toe strike (TS) and cadence (CA) between healthy control (HC) and PD patients in the ON state. More pronounced alterations in FOG patients. No significant difference in the OFF state.
- -
Greater variability in gait phase organization is seen in PD, especially in FOG. Significant changes in loading response (LR), midstance (MS), and swing (SW) phases.
- -
GPQI is significantly higher in all PD groups vs. HC, especially in FOG groups.
- -
Significant reduction in stride length (SL) and speed (SP) in PD patients vs. HC. Most pronounced in FOG.
- -
Significant decrease in foot elevation (Max1, Max2) in PD, particularly severe in FOG.
Temporal parameters, such as stride time (ST) and cadence (CA), were generally consistent with previous literature [
25,
34,
35]. Interestingly, significant differences in ST and CA were observed between healthy controls (HC) and PD patients in the ON state, particularly among those with FOG, in contrast with [
36]. Contrary to expectations, no differences were found with the OFF state. This could be due to compensatory mechanisms implemented by PD patients, who normally have shorter strides and take more steps as a result. These alterations may be more prevalent in FOG patients as they are usually in a more advanced stage of the disease. The presence of these alterations in the ON state rather than the OFF state could be due to motor difficulties. In the OFF state, patients are unable to perform these less evident compensatory mechanisms.
Gait phase (GP) analysis revealed more nuanced differences. HC showed a well-defined and consistent phase distribution, while PD patients, especially those with FoG, exhibited greater variability, particularly in the midstance (MS) and swing (SW) phases, aligning with previous findings [
25,
35]. These alterations suggest a compensatory strategy aimed at increasing stability by prolonging the stance phase and reducing swing time. The comparison between NOFOGoff and FOGoff further confirmed the destabilizing effect of FoG, particularly during the LR and MS phases, which are critical for load transfer and balance. Notably, dopaminergic therapy partially restored gait phase organization in NOFOG patients, but had a limited effect in those with FoG, indicating a more persistent motor deficit in this subgroup.
These findings were further supported by the GPQI study, with values significantly elevated in all PD groups compared to HS, with the highest values observed in FoG patients.
Regarding spatial parameters, stride length (SL) and walking speed (SP) expectedly presented a marked reduction in PD subjects compared with HC, in accordance with [
25,
34] but with lower mean values than [
21]. Particularly, the FOGoff group showed a significant reduction, consistent with the established phenomenon of hypokinesia in PD patients. It is noteworthy that drug therapy resulted in a reduction in the disparities between HC and PD, particularly among the NOFOGon group. However, this did not result in a complete restoration of performance in FoG subjects, where SL and SP remain significantly lower than in HC. This finding underscores the role of FoG in the deterioration of ambulatory performance, even in the presence of dopaminergic therapy.
The analysis of CL parameters, particularly Max1 and Max2, revealed substantial reductions in PDs, with the most pronounced deficits observed in the FOG group. Furthermore, discrepancies were identified between NOFOG and FOG in both OFF and ON states during the Max2 study. Max2 demonstrates substantial disparities across all comparisons between HC and the patient groups (PD, NOFOG, and FOG), except for the NOFOGon comparison, unveiling a pervasive deficiency in foot elevation capability during the latter half of the swing phase, in accordance with [
10,
37]. The reduction, particularly pronounced in patients with FoG, is crucial in terms of the increased risk of falling in PD patients. However, the minimum clearance does not exhibit substantial variation across the designated groups, except for NOFOGoff vs. FOGoff. This outcome is most likely attributable to the substantial variability inherent in the data sets of the respective groups under analysis. Furthermore, the limitations due to the extraction of spatial parameters from the double integration of the accelerometer signals must be taken into consideration, which, although compensated by the application of the ZUPT algorithm, still might lead to complications.
However, to ensure a comprehensive interpretation of the findings, it is important to acknowledge that the parameters under investigation did not consider any potential gait asymmetries. Indeed, no distinction was made between the right and left foot, as the emphasis was placed on evaluating the general disparities among the various groups, which could be more readily discerned and identified. This finding, however, may be indicative of greater or lesser variability between the groups, a matter that requires further investigation. It may also be of interest to evaluate the parameters by comparing them with the patients’ clinical scales to identify any patterns that could help in the early discrimination of the disease or its follow-up.
4.2. Validation of Clinical Relevance of Clearance Estimation
The validation study compared IMU-based clearance measurements to those from an optoelectronics system. While Max1 and Max2 showed good agreement and low bias, Min exhibited higher error and poor consistency, as evidenced by Bland–Altman plots and relative error metrics. These results suggest that IMU-based measurements are reliable for detecting maximum clearance points but less accurate for minimum clearance, a limitation that should be addressed in future algorithmic improvements.
Indeed, an examination of the two measurement methods reveals a negative average bias of approximately 5 mm, suggesting a substantial underestimation of IMUs in comparison to the optoelectronic system. The range of variability is also very high, which agrees with the analysis of absolute error and absolute relative percentage error values. This suggests that there is poor agreement between the two methods, and that the measurement of minimum clearance through IMU data is less reliable. The plots of Max1 and Max2 values provide more optimistic results, with smaller bias, especially for Max1, and variability ranges smaller than those of Min. Yet, the overall performance of IMUs in measuring spatiotemporal parameters is encouraging. Unlike many validation studies that omit clearance analysis [
26,
28,
29] or rely on hybrid sensor systems [
18,
30], this study demonstrates that standalone IMUs can provide clinically relevant data, supporting their use as a viable alternative to laboratory-grade optoelectronic systems [
10,
25,
30,
37,
38].
Importantly, the data indicate that FoG contributes to a generalized decline in motor function, even when freezing episodes are not actively triggered. This suggests that spatiotemporal gait parameters could serve as early indicators of FoG, aiding clinicians in diagnosis and treatment planning.
Future research should focus on expanding the sample size and improving cohort homogeneity in terms of gender, disease severity, and FoG status. Integrating additional modalities, such as electromyography, could offer deeper insights into neuro-muscular impairments. Moreover, the application of machine learning techniques for automated detection and prediction of FoG holds promise for enhancing fall risk assessment and personalized therapy.
5. Conclusions
The present study analyzed the spatiotemporal parameters in individuals with Parkinson’s disease (PD) using wearable inertial measurement units (IMUs), with a particular focus on the impact of Freezing of Gait (FoG). A cohort of 30 patients, divided into FoG and non-FoG subgroups, was analyzed under both OFF and ON dopaminergic therapy conditions and compared with 20 healthy controls. The results confirmed a significant deterioration in motor performance among PD patients, especially those with FoG. Notable impairments were observed in gait phase distribution, stride length, and foot clearance. While dopaminergic therapy improved several parameters, particularly in non-FoG patients, it did not fully restore gait performance in those with FoG, highlighting the persistent nature of this symptom. Interestingly, patients in the ON state exhibited unexpected differences in stride time and cadence compared to healthy controls, suggesting possible compensatory mechanisms. These findings warrant further investigation to clarify their clinical significance. Furthermore, possible gait asymmetry was not considered and therefore requires further study and could be addressed in future developments.
The validation of IMU-based clearance measurements demonstrated good accuracy for maximum clearance points (Max1 and Max2), with mean absolute relative errors below 10%. However, the estimation of minimum clearance proved less reliable due to higher variability and systematic underestimation. These results emphasize the need for further refinement of the signal processing algorithms used to extract clearance metrics from IMU data.
Overall, this study supports the use of wearable IMUs as a reliable, low-cost alternative to laboratory-based systems for gait analysis in PD, in consideration of the unlimited workspace. The findings also underscore the potential of IMU-derived parameters for early detection and monitoring of FoG, with implications for personalized treatment and fall prevention strategies. Future work should focus on expanding the sample size for spatiotemporal parameter analysis and for clearance validation, incorporating additional physiological signals (e.g., EMG), and exploring machine learning approaches for predictive modeling of FoG episodes.