Biomechanical Performance of BoneHelix® Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model

by
Laura Leonie Brandes
1,†,
Luis Fernando Nicolini
1,†,
Johannes Greven
1,
Philipp Lichte
1,
Thomas Thaddäus Stopinski
2,
Martin Sattler
3,
Frank Hildebrand
1 and
Miguel Pishnamaz
1,*
1
Department of Orthopedics, Trauma and Reconstructive Surgery, RWTH Aachen University Hospital, Pauwelsstr. 30, 52074 Aachen, Germany
2
Institut für Versuchstierkunde, RWTH Aachen University Hospital, Pauwelsstr. 30, 52074 Aachen, Germany
3
Johannes Wesling Klinikum Minden, Hans-Nolte-Straße 1, 32429 Minden, Germany
*
Author to whom correspondence should be addressed.
Shared first authorship.
Life 2021, 11(11), 1189; https://doi.org/10.3390/life11111189
Submission received: 13 October 2021 / Revised: 31 October 2021 / Accepted: 2 November 2021 / Published: 5 November 2021
(This article belongs to the Special Issue Healing after Trauma)

Abstract

:
Tibial shaft fractures are common injuries in the pediatric and adolescent populations. Elastic stable intramedullary nailing (ESIN) is the treatment of choice for cases that require surgical stabilization. A new intramedullary device, BoneHelix® (BH), may be an alternative for use with fractures that cannot be satisfactorily stabilized with ESIN. This study aimed to assess the biomechanical performance of BH compared with ESIN in a porcine tibia fracture model, observing cyclic fatigue and load to failure. Computed tomography was used to monitor the implant position and to rule out unintended damage. No implant or bone failure occurred during the fatigue testing. An increase in the cumulative plastic displacement was observed in both test groups over the loading cycles applied. Both implant–bone constructs displayed a trend toward closure of the osteotomy gap. During the load-to-failure test, the average loads at failure in specimens instrumented with ESIN and BH were 5364 N (±723) and 4350 N (±893), respectively, which were not statistically significant (p = 0.11). The values of both groups were two to three times higher than the estimated maximal load (2000 N) during physiological weight bearing. The biomechanical results thus indicate equivalent performance and stability by the implants tested.

Graphical Abstract

1. Introduction

Tibial shaft fractures are among the most common skeletal traumas seen in children and adolescents [1]. Isolated tibial fractures with spiral or oblique fracture lines are the most frequent type [2]. The choice of treatment is based on several factors, including age, fracture pattern, associated soft-tissue injuries, and comorbidities [3,4]. Many cases can be managed satisfactorily with nonoperative treatment by closed reduction and casting [3,5,6,7]. These are usually associated with fast healing and a low complication rate [4,8]. However, surgical treatment is required in fractures that cannot be aligned and secured within the limits of spontaneous correction by using conservative methods [6,7,9]. Elastic stable intramedullary nailing (ESIN) is currently considered the gold standard for surgical fracture management.
Since it was first introduced in the early 1980s, ESIN has established itself as the surgical treatment of choice [10,11,12]. It is based on a minimally invasive approach using two titanium or stainless steel elastic nails to form a double arch inside the medullary cavity, providing a three-point-fixation [11,13,14]. Micromovements at the fracture zone are possible due to the flexible character of the nails [15], resulting in secondary bone healing. Nevertheless, ESIN does not come without its flaws. Possible complications include delayed union or malunion, failed fixation, leg length discrepancy, and superficial infection, among others [16,17,18,19,20,21,22,23]. Tibial fracture fixation with ESIN can be demanding, and the achievement of a correct three-point fixation is not always easy to assess intraoperatively.
BoneHelix® (BH) is a helical steel coil spring with limited bending flexibility, designed for the intramedullary stabilization of long bone fractures [24,25,26]. Its multiple coils provide long-segmental contact to the spongiosa and corticalis of the bone, generating internal friction, while a high vertical spring rate prohibits impression and secures axial stability [27]. Its strictly limited flexibility allows micromovements at the fracture zone, which encourage callus formation, while preventing excessive bending [24]. In the pediatric tibia, the helix is inserted in the antegrade direction, sparing the open physis. Fixation of the implant is achieved by traction. Exercise and partial loading with a walker are allowed immediately after surgery [28]. A clinical study shows promising results in adult humeral shaft fractures treated with BH [29].
The present experimental study aimed to investigate and compare the biomechanical performance of BH and ESIN in cyclic fatigue testing, simulating an 18-week postoperative period [30], and also in load to failure, using a pediatric tibial fracture model.

2. Materials and Methods

2.1. Sample Acquisition and Preparation

Experiments were performed on eight pairs of fresh frozen tibiae, collected from juvenile pigs with an average weight of between 30 and 40 kg (Department for Laboratory Animal Science, RWTH Aachen University Hospital). Due to close similarities in anatomy and morphology, porcine bone is a suitable alternative to human bone for investigation of bone implants [31,32]. Specimens were stored at −18 °C and were thawed for over 16 h at 8 °C before testing [33,34]. Each tibia was separated from the knee joint and the fibula, and all soft tissue was removed from the bone. The tibiae were then fixated at their distal end in a 7° valgus alignment in a methyl methacrylate (MMA)-based synthetic material (Technovit® 3040, Kulzer GmbH, Hanau, Germany). The valgus alignment represents the physiological leg axis in adults [30] and is completed in children by the age of 8 to 10 years [35,36]. The fixation was performed with the help of a custom-made rail system, a stencil, and a laser to achieve the correct alignment.

2.2. Surgical Groups and Implantation Procedure

Each pair of bones was randomly instrumented with a gliding nail system for ESIN (Königsee Implantate GmbH, Allendorf, Germany) or a BH device (H&R Medizintechnik GmbH & Co. KG, Lennestadt, Germany), in either the left or the right tibia.
For ESIN, titanium nails with wire diameters of 2.0 and 2.5 mm and a length of 450 mm with flattened ends were selected in accordance with the bone size and the expected diameter of the intramedullary canal (Figure 1). The nails were implanted to fill 80% of the intramedullary canal isthmus [9,13].
BH implants are made from surgical long-term implant steel alloy (1.4441). The pediatric sizes C1 and C2, measuring an outer diameter of 7.0 or 8.0 mm, respectively, were used in accordance with the bone size (Figure 1). The aim is for the diameter of the selected helix to be 1 mm less than the narrowest point of the tibial diaphysis [28]. Implants intended for children have a standard length of 300 mm and a wire gauge of 2.5 mm.
For ESIN, two prebent C-shaped nails were introduced into the medullary cavity through openings medial and lateral to the tibial tuberosity using an inserter. The aim was to position the bent nails in a double-arch configuration within the medullary cavity. Their final position was secured by applying hammer taps to the pliers holding the nails.
The chosen BH was inserted into the medullary cavity via a guiding rod through a hole cranial to the tibial tuberosity. The helix was turned further in until resistance was noted, indicating the correct position inside the distal metaphysis. The construct was considered stable when at least six helical coils were positioned inside the bone.
The positions of both implants were checked with computed tomography (CT) (Figure 2).
The implants were trimmed to the minimum, and the proximal ends of the specimens were embedded. It was ensured that the implant ends had no contact with the embedding material so that the compressive load was transferred to the implant via the bone, and no force was acting directly on the implants during testing under compressive loads. A transverse midshaft fracture (classified as a 42t-D/4.1 fracture by PCCF [37] or a 4.2.s.3.0 fracture by LiLa [38]) was created 7 cm from the distal epiphysis using a handheld saw. A gap of 3 mm was kept between the fracture fragments to establish an unstable situation [39].

2.3. Biomechanical Tests

All biomechanical tests were performed at room temperature. A servo-pneumatic material testing machine (DYNA-MESS Prüfsysteme GmbH, Stolberg, Germany) was used for the cyclic compressive fatigue testing (Figure 3). Each implant–bone construct was exposed to three series of increased sinusoidal cyclic loading, each including 40,000 test cycles. In total, 120,000 load cycles at a frequency of 2 Hz were performed. The chosen load protocol (Figure 4) followed a protocol by Tschegg et al. [30], which was designed to simulate physiological loading conditions on the tibia.
One series of 40,000 cycles simulates normal weight bearing for a 6-week postoperative period [30]; therefore, with the present experimental design, about 18 weeks of physiological loading was simulated. Applied loads were chosen according to the average weights of children in various age groups [40]. In the first test series, the load oscillated from a middle compressive load of 175 N, with an amplitude of 75 N, simulating the full body weight (BW) of children aged 4–8 years. The second test series applied 250 N at middle load, with a 150 N amplitude, simulating full BW at 8–12 years. In the third series, loading was increased to 350 N at middle load, with an amplitude of 250 N, simulating full BW at 12–16 years (Figure 4).
If 120,000 cycles of fatigue loading were tolerated without failure, the bone–implant construct was exposed to axial compression testing. A material testing machine (modernized with RetroLine testControl II, ZwickRoell GmbH & Co. KG, Ulm, Germany) was used to determine the individual maximal load until failure occurred for each bone–implant construct (Figure 5). Tests were started with a preload of 100 N and a testing speed of 1 mm/s until a predefined upper force limit was reached, or until a reduction of 10% of the applied load was detected, implying specimen failure (force shutdown threshold at 10%). Upper force limits (Table 1) were chosen according to maximal peak forces expected on the tibia in a subject of 60 kg (approximately girls at 16 years and boys at 15 years [40]) during normal daily activities.
Peak forces reach 2 to 2.5 times the BW during level walking and 3.5 times the BW when descending stairs [41]. A maximal peak force of about 1200–1500 N for walking and of about 2000 N for descending stairs can be derived from this consideration. Load-to-failure tests started with an upper force limit of 1000 N (approximately 50% of expected peak load) and were continued with 1500 N (75%) and 2000 N (100%) if no signs of failure were observed. Loading continued in steps of 1000 N until failure in the form of visual fracturing, implant failure, or embedding failure occurred.

2.4. Data Acquisition and Analysis

Load and axial displacements were recorded by a load cell and a displacement sensor at a frequency of 100 Hz in both fatigue and quasi-static compression tests. The sensors have a maximum error of 1% with respect to the target value. The measured axial displacements correspond to the reduction in specimen size. The first 10 cycles were used as preconditioning for adjusting the machine to the specimens. The definitions of elastic displacement (ED), plastic/permanent displacement (PD), cumulative plastic displacement, and stiffness are shown in Figure 6. ED describes reversible displacement that is reduced to zero after load removal [42]. PD is defined as displacement that persists after unloading the specimen [30], and it is associated with a reduction of the fracture site during testing. These variables were calculated by considering the displacement of the piston of the machine at the time of maximal and minimal loading of a loading cycle at the beginning (10th test cycle), at an early stage of testing (2000th test cycle), and at the end (40,000th test cycle) for each loading sequence (Figure 6). The ED was calculated by subtracting the PD from the total displacement. Stiffness was calculated by taking the slope considering the force–displacement datapoints at the maximum and minimum force. The cumulative plastic displacement represents the permanent displacement that occurred in the specimen from the beginning of the test until a certain cycle.
The fracture gap size was additionally measured with a digital caliper before and after cyclic testing. The caliper had an accuracy of ±0.2 mm and a resolution of 0.1 mm. CT images were captured by a 2 × 128-slice SOMATOM Definition Flash CT scanner (Siemens Healthcare AG, Zurich, Switzerland). Specimens were scanned after implant instrumentation to confirm the correct position inside the bone, after the cyclic fatigue tests, and again after the quasi-static compression tests were completed. Images were viewed and analyzed with the software IntelliSpace PACS Enterprise (version 4.4, Philips GmbH Market DACH Health System, Hamburg, Germany).
Statistical analyses were performed in Microsoft Excel (version 16.44, Microsoft Corporation, Redmond, DC, USA). The means of both test groups were compared using two-sample two-tailed t-tests or paired t-tests, respectively, when looking at values within the same test groups at different times of testing. All statistical tests were based on a 5% alpha error. Results with p values < 0.05 were considered statistically significant.
For the load-to-failure tests, the shortening of the specimen was analyzed from a load of 150 N to a maximum value representing a physiological situation (e.g., 1000 N). This load of 150 N served to reduce to zero the gap between the parts of the machine (e.g., interface of the specimen with the mechanical parts of the machine) so that the obtained displacement would be only from the specimen. The software G*Power (version 3.1.9.6) [43] was used to perform a power analysis at the point of discussing the maximal load tolerated at the point of failure.

3. Results

3.1. Cyclic Loading Fatigue Testing

All 16 specimens survived 120,000 cycles of increased fatigue loading without showing signs of implant or bone failure. This statement is supported by the data collected during testing and the CT imaging after the experiments were completed.

3.1.1. Axial Displacements

ED was significantly reduced in the ESIN specimens in all loading series between the 10th and 40,000th test cycle (paired t-tests: first: p = 0.0028, second: p = 0.0032, third: p = 0.0036). For the BH specimens, the reduction of ED was not verified to be statistically significant (paired t-tests: first: p = 0.058, second: p = 0.42, third: p = 0.39). Both implants showed a decreasing trend in the development of ED in all test series. However, the change in ED within each series was negligible (Figure 7). For the first loading series, it was 0.059 mm (±0.042) for ESIN and 0.023 mm (±0.032) for BH.
An increase in the cumulative PD was observed in both test groups over the loading cycles applied. The PD was greatest during the first 2000 loading cycles for each sequence when the force was increased, presenting an asymptotic behavior (Figure 8). For the first series, in the 2000th cycle it reached values of 0.92 and 0.58 mm for ESIN and BH, respectively (Figure 7). In the 40,000th cycle, the cumulative PD was 1.35 mm for ESIN and 0.93 mm for BH. The ESIN specimens showed a significantly greater cumulative PD than the BH specimens (t-test; p = 0.014).
Cumulative PDs of 2.71 mm (±0.71) in the ESIN and of 2.12 mm (±0.58) in the BH specimens were seen after 120,000 cycles of fatigue loading (t-test; p = 0.092). These results display an equivalent accumulated PD after simulated physiological weight-bearing modelling for an 18-week period. The manually measured reduction in the fracture gap (ESIN: 2.0 mm (±0.5), BH: 1.3 mm (±0.7)) showed the same trend toward closure, analogous to the increase in PD.

3.1.2. Axial Stiffness

An increase in stiffness was observed for both test groups in relation to the increasing number of test cycles (Table 2). The increase was confirmed to be statistically significant for the ESIN group in all test sequences (paired t-test: first: p = 0.00016, second: p = 0.00028, third: p = 0.00081). However, the BH group only showed a significant increase in stiffness in the first and third sequences, but not in the second (first: p = 0.0088, second: p = 0.19, third: p = 0.0024).
Early stiffness was significantly higher in BH specimens in the second and third test series (t-test: second: p = 0.013, third: p = 0.026), while in the first series, no difference was revealed in early stiffness between groups (p = 0.91). This indicates a higher capacity for deformation, allowing more movement for specimens operating with ESIN under increased loading in comparison with BH. Late stiffness was equivalent for the test groups in all three test sequences (t-test: first: p = 0.89, second: p = 0.37, third: p = 0.57).

3.2. Load-to-Failure Testing

3.2.1. Location and Type of Failure

Neither of the test groups showed signs of implant failure during loading or in the following CT. A homologous fracture pattern was seen in both test groups. Fractures or fissures were located exclusively in the upper bone fragment, mostly in relation to the previously created transverse fracture site in the middle third of the tibia (Figure 9). An exception was seen in one ESIN specimen, where a fissure started at the lateral drill hole created for the implant insertion. Longitudinal stress fissures were displayed in three ESIN and two BH specimens. Incomplete, displaced, simple longitudinal fractures were seen in one ESIN and two BH specimens. In the BH group, fatigue loading resulted once in a complete, longitudinal, displaced multifragment fracture with blasted-out bone fragments.

3.2.2. Load at Failure

Loadings of 50% (1000 N), 75% (1500 N), and 100% (2000 N) of the expected peak load were tolerated by all bone–implant constructs in both test groups. The mean maximal load detected at failure in both test groups was 4801 N (±885). This equals about 240% of the expected peak load of 2000 N. ESIN specimens tolerated 5364 N (±723) on average, while BH specimens were loaded with 4350 N (±893) before bone failure occurred (268% vs. 218% of the expected peak load). The maximal loads at failure did not differ significantly between test groups (t-test; p = 0.11). The original data for the load-to-failure tests are available as Supplementary Material in FigShare at DOI 10.6084/m9.figshare.16910611 (last accessed on 1 November 2021).

3.2.3. Axial Displacement at Maximal Load

Load–displacement graphs display a linear, proportional, and statistically relevant increase in axial displacement according to increased loading in both test groups (Figure 10).
Displacement data at 1000, 1500, 2000, and 3000 N were compared for each group. In the ESIN specimens, displacement increased significantly from 0.55 mm (±0.18) at 1000 N to 0.75 mm (±0.14) at 1500 N (paired t-test; p = 0.025), to 0.99 mm (±0.18) at 2000 N (p = 0.001), and to 1.52 mm (±0.30) at 3000 N (p = 0.0034). In the BH specimens, displacement increased from 0.75 mm (±0.19) at 1000 N to 1.09 mm (±0.25) at 1500 N (p = 0.00020), to 1.40 mm (±0.27) at 2000 N (p = 0.000028), and to 2.32 mm (±0.57) at 3000 N (p = 0.0035).
No relevant difference of displacements between test groups was observed for the upper force limits of 1000 N (t-test; p = 0.17) and 1500 N (p = 0.050). In contrast, BH specimens displayed a greater axial displacement when loaded until 2000 N (p = 0.037) and 3000 N (p = 0.041) (Figure 10). BH specimens apparently suffered greater displacement than ESIN specimens when exposed to greater loading.

4. Discussion

This is the first study that compares the biomechanical effects of fracture stabilization by BH or ESIN on a porcine tibial fracture model simulating a pediatric trauma with 18 weeks of postsurgical mobilization. BH is an innovative treatment option for the treatment of long bone fractures, which are common skeletal trauma patterns. A clinical study on adult humeral shaft fractures treated with BH presented promising results [29]. However, limited data are available regarding the fatigue and load-to-failure performance of BH, which could contribute to clinical findings, and there is no analysis from the biomechanical point of view of the possibility of using BH as an alternative to ESIN.
Several studies attest that ESIN has a low complication rate and positive outcomes after treatment, and they promote it as an effective and reliable method for the surgical management of pediatric tibial shaft fractures [21,44,45,46,47].
However, possible complications, such as delayed union, malunion, and failed fixation, can occur [16,17,18,19,20,21,22,23]. The main reasons for the failure of ESIN are errors during the insertion process or, in particular, insufficient intraosseous three-point fixation. This problem can occur because of complex fracture scenarios (oblique or multifragmentary) or due to adverse angles of the insertion axis of ESIN in the long bone, resulting in a weak bracing of the wires in the bone shaft [13,20,48,49,50,51,52]. The rate of complications, such as malunion or delayed union, is also higher in heavier adolescent patients [9,16,17,18,19], although greater weight alone does not seem to predict a higher rate of malunion [16,46,53]. These limitations indicate the need to develop alternative surgical treatment options for cases in which ESIN does not provide sufficient stabilization for adequate bone healing.
Stabilization of the fracture is an important requirement for tibial implants. The implants need to withstand forces occurring at the knee joint under normal weight bearing during activities of daily life, since they must compensate for the temporary loss of function of the traumatized bone. The absence of signs of implant or bone failure as analyzed by visual inspection or with CT scans indicates that simulated physiological loading after surgical treatment of a tibial shaft fracture with ESIN or BH in the pediatric and adolescent age group is tolerated without causing failure of implants or of the osteosynthesis–bone construct.
This demonstrates that both types of implants are capable of fixing and keeping the bone parts together while the bone–implant construct is submitted to physiological loads adapted to the bone healing process of a child [54].
The increase in cumulative PD and the reduction of fracture gap present in both groups indicate that the load exceeded the yield strength of the bone–implant construct, causing it to compress. This effect led to an increase in the axial stiffness of the system when the results of the last series were compared against those of the first series. The cumulative PD tended to converge to a specific value until the external load was increased in our testing protocol, suggesting that the amount of damage tends to stabilize.
The results demonstrated that the bone–implant constructs tested can withstand loading beyond physiological conditions with no statistical differences between groups, while the magnitude of the load is limited to the weight tolerance of the fractured tibial bone structure.
For both implants, the load at failure was two to three times higher than the expected peak load experienced by the tibia of a subject of 60 kg performing daily activities, such as walking and descending stairs. Additionally, collapse of the bone or the interface with the embedding material was observed during load to failure; however, there were no signs of implant damage. This suggests that the weakest structure is the bone volume close to the fracture site, and therefore, from a purely mechanical point of view, further improvement of the instrumentation to resist higher forces would require a reduction of the forces exerted on the tibia close to the fracture site. The implant itself does not seem to be the problem in resisting physiological stress inside the tibia since none of the implants failed. Moreover, the bone instrumented with BH had an equivalent load tolerance to that instrumented with ESIN.
This study has some limitations. The in vitro testing differs from the in vivo condition due to the missing vasculature and soft tissue envelope, and also in terms of the loading condition. This was also the reason why no tests were performed with regard to the rotational stability of the two implant types, as the surrounding soft tissue plays a decisive role here. More complex tests must be carried out to simulate further physiological parameters and loads and to investigate whether or not BH is superior to ESIN in certain scenarios. The physiological forces acting on the tibia adapt to a complex system of muscular forces that change to control movements and to keep the body in equilibrium when external loads, such as the gravity, are acting. We approximated this configuration by using a compressive force that acted along the tibia shaft. However, several authors perform simplification of the loads [30,55,56,57,58], and this is frequently necessary to make the testing reliable. Furthermore, we used pig bones instead of human cadaver specimens. However, animal specimens do represent a suitable alternative to human bones because of their low variance in material properties, low cost, and availability [32].
In summary, all test results implied equivalent biomechanical performance by the two types of implants investigated, with BH providing higher stability in the early weight loading period, paired with greater stiffness than ESIN. This might be an advantage in the stabilization of more complicated fractures or in heavier adolescent patients in whom fractures need a longer time to form callus and are prone to develop complications [16,17,18,19]. Our model, specifically designed to address the issue of axial load parameters, provides the basis for further investigation of BH in regard to rotational stability. Nevertheless, for the first time we obtained results for axial load experiments to attest to BH’s satisfactory biomechanical properties for use in pediatric and adolescent tibial shaft fractures, which could compensate for the possible limitations of ESIN.

Supplementary Materials

The following are available online in https://www.mdpi.com/article/10.3390/life11111189/s1 and FigShare at DOI 10.6084/m9.figshare.16910611 (last accessed on 1 November 2021). Figure S1: Images of specimens with ESIN implants after load to failure tests. Arrows indicate fractures; Figure S2: Images of specimens with BoneHelix implants after load to failure tests. Arrows indicate fractures; Table S1: Experimental data after load-to-failure tests of BoneHelix; Table S2: Experimental data after load-to-failure tests of ESIN.

Author Contributions

Conceptualization: J.G., F.H., L.F.N. and M.S.; methodology: L.L.B. and L.F.N.; formal analysis: L.F.N. and L.L.B.; investigation: J.G., L.F.N., L.L.B. and P.L.; resources: T.T.S. and F.H.; data curation: L.F.N. and L.L.B.; writing—original draft preparation: L.L.B. and L.F.N.; writing—review and editing: J.G. and M.P.; supervision: F.H.; funding acquisition: M.P., J.G., and L.F.N. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by H&R Medizintechnik GmbH & Co. KG, Elsper Straße 36, 57368 Lennestadt, UST-ID: DE 813 100 566.

Institutional Review Board Statement

The bones of the pigs were taken from the officially registered animal experiment with the number 81-02.04.2020.A215. It is registered at the State Office for Nature, Environment and Consumer Protection North Rhine-Westphalia (LANUV).

Data Availability Statement

The data are available online in FigShare at DOI 10.6084/m9.figshare.16910611 (last accessed on 1 November 2021).

Acknowledgments

We would like to thank Marx Ribeiro for his valuable knowledge inputs in terms of mechanical engineering, which helped to improve this publication.

Conflicts of Interest

The authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

References

  1. Shannak, A.O. Tibial fractures in children: Follow-up study. J. Pediatr. Orthop. 1988, 8, 306–310. [Google Scholar] [CrossRef]
  2. Yang, J.P.; Letts, R.M. Isolated fractures of the tibia with intact fibula in children: A review of 95 patients. J. Pediatr. Orthop. 1997, 17, 347–351. [Google Scholar] [CrossRef] [PubMed]
  3. Mashru, R.P.; Herman, M.J.; Pizzutillo, P.D. Tibial shaft fractures in children and adolescents. J. Am. Acad. Orthop. Surg. 2005, 13, 345–352. [Google Scholar] [CrossRef] [PubMed]
  4. Patel, N.K.; Horstman, J.; Kuester, V.; Sambandam, S.; Mounasamy, V. Pediatric Tibial Shaft Fractures. Indian J. Orthop. 2018, 52, 522–528. [Google Scholar] [CrossRef]
  5. Briggs, T.W.; Orr, M.M.; Lightowler, C.D. Isolated tibial fractures in children. Injury 1992, 23, 308–310. [Google Scholar] [CrossRef]
  6. Gordon, J.E.; O’Donnell, J.C. Tibia fractures: What should be fixed? J. Pediatr. Orthop. 2012, 32, S52–S61. [Google Scholar] [CrossRef]
  7. Heinrich, S. Fractures of the shaft of the tibia and fibula. In Rockwood & Wilkins’ Fractures in Children, 5th ed.; Beaty, J.H., Kasser, J.R., Eds.; Lippincott Williams & Wilkins: Philadelphia, PA, USA, 2001; pp. 1077–1118. [Google Scholar]
  8. Setter, K.J.; Palomino, K.E. Pediatric tibia fractures: Current concepts. Curr. Opin. Pediatr. 2006, 18, 30–35. [Google Scholar] [CrossRef]
  9. Flynn, J.M.; Skaggs, D.; Sponseller, P.D.; Ganley, T.J.; Kay, R.M.; Leitch, K.K. The operative management of pediatric fractures of the lower extremity. J. Bone Joint Surg. 2002, 84A, 2288–2300. [Google Scholar] [CrossRef]
  10. Ligier, J.; Métaizeau, J.; Prévot, J. L’embrochage élastique stable à foyer fermé en traumatologie infantile. Chir. Pédiatrique 1983, 24, 383–385. [Google Scholar]
  11. Ligier, J.N.; Metaizeau, J.P.; Prévot, J.; Lascombes, P. Elastic stable intramedullary nailing of femoral shaft fractures in children. J. Bone Joint Surg. Br. 1988, 70, 74–77. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  12. Metaizeau, J.; Ligier, J. Le traitement chirurgical des fractures des os longs chez l’enfant. J. Chir. 1984, 121, 527–537. [Google Scholar]
  13. Lascombes, P.; Haumont, T.; Journeau, P. Use and abuse of flexible intramedullary nailing in children and adolescents. J. Pediatr. Orthop. 2006, 26, 827–834. [Google Scholar] [CrossRef] [PubMed]
  14. Rether, J.R. Intramedulläre Stabilisierung von Schaftfrakturen im Wachstumsalter. Trauma Und Berufskrankh. 2005, 7, 112–117. [Google Scholar] [CrossRef]
  15. Swindells, M.G.; Rajan, R.A. Elastic intramedullary nailing in unstable fractures of the paediatric tibial diaphysis: A systematic review. J. Child Orthop. 2010, 4, 45–51. [Google Scholar] [CrossRef] [Green Version]
  16. Deakin, D.E.; Winter, H.; Jain, P.; Bache, C.E. Malunion following flexible intramedullary nails for tibial and femoral fractures in adolescents. J. Child Orthop. 2010, 4, 571–577. [Google Scholar] [CrossRef] [Green Version]
  17. Goodwin, R.C.; Gaynor, T.; Mahar, A.; Oka, R.; Lalonde, F.D. Intramedullary flexible nail fixation of unstable pediatric tibial diaphyseal fractures. J. Pediatr. Orthop. 2005, 25, 570–576. [Google Scholar] [CrossRef]
  18. Gordon, J.E.; Gregush, R.V.; Schoenecker, P.L.; Dobbs, M.B.; Luhmann, S.J. Complications after titanium elastic nailing of pediatric tibial fractures. J. Pediatr. Orthop. 2007, 27, 442–446. [Google Scholar] [CrossRef]
  19. Griffet, J.; Leroux, J.; Boudjouraf, N.; Abou-Daher, A.; El Hayek, T. Elastic stable intramedullary nailing of tibial shaft fractures in children. J. Child Orthop. 2011, 5, 297–304. [Google Scholar] [CrossRef] [Green Version]
  20. Illing, P.; Rapp, M.; Strack, A. Unterschenkelfraktur im Kindesalter Vorstellung der AWMF S1-Leitlinie 006-131. Med.-Thopädische Tech. 2015, 2, 14–17. [Google Scholar]
  21. Vallamshetla, V.R.; De Silva, U.; Bache, C.E.; Gibbons, P.J. Flexible intramedullary nails for unstable fractures of the tibia in children. An eight-year experience. J. Bone Joint. Surg. Br. 2006, 88, 536–540. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  22. Pandya, N.K. Flexible Intramedullary Nailing of Unstable and/or Open Tibia Shaft Fractures in the Pediatric Population. J. Pediatr. Orthop. 2016, 36, S19–S23. [Google Scholar] [CrossRef] [PubMed]
  23. Pandya, N.K.; Edmonds, E.W.; Mubarak, S.J. The incidence of compartment syndrome after flexible nailing of pediatric tibial shaft fractures. J. Child Orthop. 2011, 5, 439–447. [Google Scholar] [CrossRef] [Green Version]
  24. Labitzke, R. Bone adaptation to dynamic osteosynthetic implants. In Wolff’s Law and Connective Tissue Regulation; Regling, G., Ed.; de Gruyter: Berlin, Germany, 1993. [Google Scholar]
  25. Labitzke, R. Atraumatisch-biologische Markraumstabilisierung der Humerusschaftfraktur durch die Endo-Helix. Aktuelle Traumatol. 2000, 30, 195–196. [Google Scholar]
  26. Labitzke, R.; Stiebeiner, M.; Wirtz, W.; Frotscher, M. The ‘BoneHelix’—A dynamic intramedullary stabilisator for shaft fractures. In Proceedings of the 1st Emirati-German Health Congress, Dubai, United Arab Emirates, 4–6 March 2013. [Google Scholar]
  27. Labitzke, R.; Stiebeiner, M.; Wirtz, W.; Frotscher, M. Der limitiert-flexible Markraumstabilisator BoneHelix® bei Humerusschaftfrakturen. Chir. Allg. 2013, 14, 1–8. [Google Scholar]
  28. H&R Medizintechnik GmbH & Co., KG. BoneHelix® after Prof. Dr. Labitzke—Instruction Manual: Instructions for Cleaning and Sterilization. Surgical Instructions for Humerus, Tibia and Femur Fractures in Adults and Children. Available online: http://www.bonehelix.com/fileadmin/user_upload/downloads/190426-BONEHELIX-ANLEITUNG-FREIGABEVERSIONmit_CE_ENG_ANSICHT.pdf (accessed on 16 February 2021).
  29. Mihalic, R. Die EndoHelix nach Labitzke bei Humerusschaftfrakturen. Operationsverfahren, Nachuntersuchungen und Ergebnisse einer Multi-Center-Studie; Universität Witten/Herdecke: Witten, Germany, 2007. [Google Scholar]
  30. Tschegg, E.K.; Herndler, S.; Weninger, P.; Jarnek, M.; Stanzi-Tschegg, S.; Redl, H. Stiffness analysis of tibia-implant system under cyclic loading. Mater. Sci. Eng. C-Biomimetic Supramol. Syst. 2008, 28, 1203–1208. [Google Scholar] [CrossRef]
  31. Thorwarth, M.; Schultze-Mosgau, S.; Kessler, P.; Wiltfang, J.; Schlegel, K.A. Bone regeneration in osseous defects using a resorbable nanoparticular hydroxyapatite. J. Oral Maxillofac. Surg. 2005, 63, 1626–1633. [Google Scholar] [CrossRef]
  32. Wancket, L.M. Animal Models for Evaluation of Bone Implants and Devices: Comparative Bone Structure and Common Model Uses. Vet. Pathol. 2015, 52, 842–850. [Google Scholar] [CrossRef] [Green Version]
  33. Beckmann, A.; Herren, C.; Nicolini, L.F.; Grevenstein, D.; Oikonomidis, S.; Kobbe, P.; Hildebrand, F.; Stoffel, M.; Markert, B.; Siewe, J. Biomechanical testing of a polycarbonate-urethane-based dynamic instrumentation system under physiological conditions. Clin. Biomech. 2019, 61, 112–119. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  34. Beckmann, A.; Nicolini, L.F.; Grevenstein, D.; Backes, H.; Oikonomidis, S.; Sobottke, R.; Kobbe, P.; Hildebrand, F.; Stoffel, M.; Markert, B.; et al. Biomechanical In Vitro Test of a Novel Dynamic Spinal Stabilization System Incorporating Polycarbonate Urethane Material Under Physiological Conditions. J. Biomech. Eng. 2020, 142, 011005. [Google Scholar] [CrossRef]
  35. Salenius, P.; Vankka, E. The development of the tibiofemoral angle in children. J. Bone Joint Surg. Am. 1975, 57, 259–261. [Google Scholar] [CrossRef]
  36. Westhoff, B.; Jäger, M.; Krauspe, R. Kindliche Beinachsen. Der. Orthopäde 2007, 36, 485–500. [Google Scholar] [CrossRef]
  37. Slongo, T.F.; Audigé, L. Fracture and dislocation classification compendium for children: The AO pediatric comprehensive classification of long bone fractures (PCCF). J. Orthop. Trauma 2007, 21, S135–S160. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  38. Schneidmüller, D.; Röder, C.; Kraus, R.; Marzi, I.; Kaiser, M.; Dietrich, D.; von Laer, L. Development and validation of a paediatric long-bone fracture classification. A prospective multicentre study in 13 European paediatric trauma centres. BMC Musculoskelet. Disord. 2011, 12, 89. [Google Scholar] [CrossRef] [Green Version]
  39. Kaspar, K.; Schell, H.; Seebeck, P.; Thompson, M.S.; Schütz, M.; Haas, N.P.; Duda, G.N. Angle stable locking reduces interfragmentary movements and promotes healing after unreamed nailing. Study of a displaced osteotomy model in sheep tibiae. J. Bone Joint Surg. Am. 2005, 87, 2028–2037. [Google Scholar] [CrossRef]
  40. Neuhauser, H.; Schienkiewitz, A.; Rosario, A.S.; Dortschy, R.; Kurth, B.-M. Referenzperzentile für anthropometrische Maßzahlen und Blutdruck aus der Studie zur Gesundheit von Kindern und Jugendlichen in Deutschland (KiGGS); Robert Koch-Institut: Berlin, Germany, 2013. [Google Scholar]
  41. Kutzner, I.; Heinlein, B.; Graichen, F.; Bender, A.; Rohlmann, A.; Halder, A.; Beier, A.; Bergmann, G. Loading of the knee joint during activities of daily living measured in vivo in five subjects. J. Biomech. 2010, 43, 2164–2173. [Google Scholar] [CrossRef]
  42. Schüller, M.; Herndler, S.; Weninger, P.; Jamek, M.; Redl, H.; Tschegg, E.K. Stiffness and permanent deformation of extra-articular distal tibia fractures treated with unreamed small diameter intramedullary nailing. Mater. Sci. Eng. C 2008, 28, 1209–1216. [Google Scholar] [CrossRef]
  43. Faul, F.; Erdfelder, E.; Lang, A.G.; Buchner, A. G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav. Res. Methods 2007, 39, 175–191. [Google Scholar] [CrossRef] [PubMed]
  44. Kubiak, E.N.; Egol, K.A.; Scher, D.; Wasserman, B.; Feldman, D.; Koval, K.J. Operative treatment of tibial fractures in children: Are elastic stable intramedullary nails an improvement over external fixation? J. Bone Joint Surg. Am. 2005, 87, 1761–1768. [Google Scholar] [CrossRef] [PubMed]
  45. O’Brien, T.; Weisman, D.S.; Ronchetti, P.; Piller, C.P.; Maloney, M. Flexible titanium nailing for the treatment of the unstable pediatric tibial fracture. J. Pediatr. Orthop. 2004, 24, 601–609. [Google Scholar] [CrossRef]
  46. Sankar, W.N.; Jones, K.J.; David Horn, B.; Wells, L. Titanium elastic nails for pediatric tibial shaft fractures. J. Child Orthop. 2007, 1, 281–286. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  47. Srivastava, A.K.; Mehlman, C.T.; Wall, E.J.; Do, T.T. Elastic stable intramedullary nailing of tibial shaft fractures in children. J. Pediatr. Orthop 2008, 28, 152–158. [Google Scholar] [CrossRef]
  48. Lascombes, P. Flexible Intramedullary Nailing in Children: The Nancy University Manual; Springer Science & Business Media: Berlin, Germany, 2009. [Google Scholar]
  49. Parikh, S.N.; Jain, V.V.; Denning, J.; Tamai, J.; Mehlman, C.T.; McCarthy, J.J.; Wall, E.J.; Crawford, A.H. Complications of elastic stable intramedullary nailing in pediatric fracture management. J. Bone Joint Surg. Am. 2012, 94, e184. [Google Scholar] [CrossRef]
  50. Schalamon, J.; Ainödhofer, H.; Joeris, A.; Knorr, P.; Lieber, J.; Höllwarth, M. Elastic stable intramedullary nailing (ESIN) in lower leg fractures. Eur. J. Trauma 2005, 31, 19–23. [Google Scholar] [CrossRef]
  51. Slongo, T.F. Complications and failures of the ESIN technique. Injury 2005, 36 (Suppl. 1), A78–A85. [Google Scholar] [CrossRef]
  52. Sommerfeldt, D.W.; Schmittenbecher, P.P. Elastic stable intramedullary nailing (ESIN) in the adolescent patient-perils, pearls, and pitfalls. Eur. J. Trauma Emerg. Surg. 2014, 40, 3–13. [Google Scholar] [CrossRef]
  53. Pennock, A.T.; Huang, S.G.; Pedowitz, J.M.; Pandya, N.K.; McLaughlin, D.C.; Bastrom, T.P.; Ellis, H.B. Risk Factors for Adverse Radiographic Outcomes After Elastic Stable Intramedullary Nailing of Unstable Diaphyseal Tibia Fractures in Children. J. Pediatr. Orthop. 2020, 40, 481–486. [Google Scholar] [CrossRef] [PubMed]
  54. Silva, M.; Eagan, M.J.; Wong, M.A.; Dichter, D.H.; Ebramzadeh, E.; Zionts, L.E. A comparison of two approaches for the closed treatment of low-energy tibial fractures in children. J. Bone Joint Surg. Am. 2012, 94, 1853–1860. [Google Scholar] [CrossRef] [PubMed]
  55. Gueorguiev, B.; Wähnert, D.; Albrecht, D.; Ockert, B.; Windolf, M.; Schwieger, K. Effect on Dynamic Mechanical Stability and Interfragmentary Movement of Angle-Stable Locking of Intramedullary Nails in Unstable Distal Tibia Fractures: A Biomechanical Study. J. Trauma Acute Care Surg. 2011, 70, 358–365. [Google Scholar] [CrossRef] [PubMed]
  56. Kaiser, M.M.; Wessel, L.M.; Zachert, G.; Stratmann, C.; Eggert, R.; Gros, N.; Schulze-Hessing, M.; Kienast, B.; Rapp, M. Biomechanical analysis of a synthetic femur spiral fracture model: Influence of different materials on the stiffness in flexible intramedullary nailing. Clin. Biomech. 2011, 26, 592–597. [Google Scholar] [CrossRef] [PubMed]
  57. Kaiser, M.M.; Zachert, G.; Wendlandt, R.; Eggert, R.; Stratmann, C.; Gros, N.; Schulze-Hessing, M.; Rapp, M. Increasing stability by pre-bending the nails in elastic stable intramedullary nailing. J. Bone Jt. Surgery. Br. Vol. 2012, 94-B, 713–718. [Google Scholar] [CrossRef]
  58. Schüller, M.; Weninger, P.; Tschegg, E.; Jamek, M.; Redl, H.; Stanzl-Tschegg, S. Micromotion at the Fracture Site After Tibial Nailing With Four Unreamed Small-Diameter Nails—A Biomechanical Study Using a Distal Tibia Fracture Model. J. Trauma Acute Care Surg. 2009, 66, 1391–1397. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Implants compared in biomechanical testing: titanium gliding nails for ESIN with 2.0 mm diameter (above) and BH C1 implant (below); measuring tape is displayed in cm.
Figure 1. Implants compared in biomechanical testing: titanium gliding nails for ESIN with 2.0 mm diameter (above) and BH C1 implant (below); measuring tape is displayed in cm.
Life 11 01189 g001
Figure 2. Specimens with ESIN (a) or BH (b) after instrumentation before fracture creation. The CT scans are shown on the right side of each subfigure.
Figure 2. Specimens with ESIN (a) or BH (b) after instrumentation before fracture creation. The CT scans are shown on the right side of each subfigure.
Life 11 01189 g002
Figure 3. DYNA-MESS machine used for fatigue tests of the specimens. The specimen is a porcine bone used for pilot tests.
Figure 3. DYNA-MESS machine used for fatigue tests of the specimens. The specimen is a porcine bone used for pilot tests.
Life 11 01189 g003
Figure 4. Three sequences of cyclic loading were applied: (a) load according to weight at 4–8 years, (b) load according to weight at 8–12 years, (c) load according to weight at 12–16 years. Own figure, based on Tschegg et al. [30].
Figure 4. Three sequences of cyclic loading were applied: (a) load according to weight at 4–8 years, (b) load according to weight at 8–12 years, (c) load according to weight at 12–16 years. Own figure, based on Tschegg et al. [30].
Life 11 01189 g004
Figure 5. ZwickRoell machine used for load-to-failure tests. The specimen is a porcine bone used for pilot tests.
Figure 5. ZwickRoell machine used for load-to-failure tests. The specimen is a porcine bone used for pilot tests.
Life 11 01189 g005
Figure 6. Definitions of elastic displacement and plastic displacement calculated for one cycle (loading and unloading), and cumulative plastic displacement calculated until the end of the 10th cycle. Stiffness was calculated by taking the slope considering the force–displacement datapoints at the maximum and minimum force (100 N) of the half cycle. Source: adapted from Tschegg et al. [30].
Figure 6. Definitions of elastic displacement and plastic displacement calculated for one cycle (loading and unloading), and cumulative plastic displacement calculated until the end of the 10th cycle. Stiffness was calculated by taking the slope considering the force–displacement datapoints at the maximum and minimum force (100 N) of the half cycle. Source: adapted from Tschegg et al. [30].
Life 11 01189 g006
Figure 7. Elastic displacement (ED) and cumulative permanent displacement (PD) for the (A) beginning (10th test cycle), (B) early stage (2000th test cycle), and (C) end (40,000th test cycle) of the first loading sequence. ESIN specimens showed significantly greater cumulative PD than BH specimens (t-test; p = 0.014).
Figure 7. Elastic displacement (ED) and cumulative permanent displacement (PD) for the (A) beginning (10th test cycle), (B) early stage (2000th test cycle), and (C) end (40,000th test cycle) of the first loading sequence. ESIN specimens showed significantly greater cumulative PD than BH specimens (t-test; p = 0.014).
Life 11 01189 g007
Figure 8. Development of cumulative PD over 120,000 test cycles with increased loading at 40,000th and 80,000th cycles. PD was greatest during the first 2000 cycles of each test series. ESIN specimens showed a significantly greater PD during the first loading sequence (t-test; p = 0.014). No statistically relevant difference in the development of cumulative PD was observed in the subsequent test series (t-test: second: p = 0.98, third: p = 0.13).
Figure 8. Development of cumulative PD over 120,000 test cycles with increased loading at 40,000th and 80,000th cycles. PD was greatest during the first 2000 cycles of each test series. ESIN specimens showed a significantly greater PD during the first loading sequence (t-test; p = 0.014). No statistically relevant difference in the development of cumulative PD was observed in the subsequent test series (t-test: second: p = 0.98, third: p = 0.13).
Life 11 01189 g008
Figure 9. Implants after failure loading (examples): (a) ESIN specimen displaying a simple, displaced fracture after testing (left) and in CT (right); (b) BH specimen displaying multifragment, displaced fracture after testing (left) and in CT (right). Arrows indicate fractures. More examples are available as supplementary material in FigShare at DOI 10.6084/m9.figshare.16910611.
Figure 9. Implants after failure loading (examples): (a) ESIN specimen displaying a simple, displaced fracture after testing (left) and in CT (right); (b) BH specimen displaying multifragment, displaced fracture after testing (left) and in CT (right). Arrows indicate fractures. More examples are available as supplementary material in FigShare at DOI 10.6084/m9.figshare.16910611.
Life 11 01189 g009
Figure 10. Load–displacement graphs of quasi-static compressive loading until 2000 N: development of axial displacement in ESIN (dotted lines) and BH specimens (continuous lines). Displacement was significantly greater in BH specimens (p = 0.037).
Figure 10. Load–displacement graphs of quasi-static compressive loading until 2000 N: development of axial displacement in ESIN (dotted lines) and BH specimens (continuous lines). Displacement was significantly greater in BH specimens (p = 0.037).
Life 11 01189 g010
Table 1. Load protocol of load-to-failure testing. First, peak loads of 50%, 75%, and 100% were applied. Loads were increased in steps of 1000 N until failure was observed. An amount of 2000 N was expected to be the tibial peak load in a subject of 60 kg, representing 3.5 times the body weight when descending stairs. A 100% peak load was estimated to cover the forces occurring in everyday activities.
Table 1. Load protocol of load-to-failure testing. First, peak loads of 50%, 75%, and 100% were applied. Loads were increased in steps of 1000 N until failure was observed. An amount of 2000 N was expected to be the tibial peak load in a subject of 60 kg, representing 3.5 times the body weight when descending stairs. A 100% peak load was estimated to cover the forces occurring in everyday activities.
TestUpper Force Limit (N)Percentage of Expected Peak Load
1100050%
21500 (walking)75%
32000 (stairs)100% (3.5 × 60 kg)
4–n3000–n150%–n
Table 2. Development of axial stiffness in cyclic loading fatigue testing. A significant increase in stiffness was observed in ESIN specimens for all test series. Increase in stiffness in BH specimens was verified to be significant in the first and third test series only.
Table 2. Development of axial stiffness in cyclic loading fatigue testing. A significant increase in stiffness was observed in ESIN specimens for all test series. Increase in stiffness in BH specimens was verified to be significant in the first and third test series only.
SeriesCyclesStiffness ESIN (N/mm)Stiffness BH (N/mm)
110
40,000
1044 (±606)
1685 (±733)
1070 (±191)
1639 (±513)
240,010
80,000
911 (±279)
1406 (±336)
1433 (±434)
1562 (±339)
380,010
120,000
1122 (±246)
1709 (±339)
1594 (±477)
1824 (±435)
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Brandes, L.L.; Nicolini, L.F.; Greven, J.; Lichte, P.; Stopinski, T.T.; Sattler, M.; Hildebrand, F.; Pishnamaz, M. Biomechanical Performance of BoneHelix® Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model. Life 2021, 11, 1189. https://doi.org/10.3390/life11111189

AMA Style

Brandes LL, Nicolini LF, Greven J, Lichte P, Stopinski TT, Sattler M, Hildebrand F, Pishnamaz M. Biomechanical Performance of BoneHelix® Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model. Life. 2021; 11(11):1189. https://doi.org/10.3390/life11111189

Chicago/Turabian Style

Brandes, Laura Leonie, Luis Fernando Nicolini, Johannes Greven, Philipp Lichte, Thomas Thaddäus Stopinski, Martin Sattler, Frank Hildebrand, and Miguel Pishnamaz. 2021. "Biomechanical Performance of BoneHelix® Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model" Life 11, no. 11: 1189. https://doi.org/10.3390/life11111189

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop