Next Article in Journal
Tensile Properties of Ligament and Tendon Structures in the Human Knee: An Exploratory Study
Previous Article in Journal
Sex- and Sport-Specific Patterns of Inter-Limb Jumping Asymmetries: A Force Plate Analysis in Youth Elite Athletes
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Impact Fracture Thresholds of Ceramic Femoral Heads in Total Hip Arthroplasty: An Explicit Dynamic Finite Element Analysis

by
Vladimir Pakhaliuk
and
Aleksandr M. Poliakov
*
Laboratory of Biomechanics, Engineering Department, Sevastopol State University, Sevastopol 299053, Russia
*
Author to whom correspondence should be addressed.
Biomechanics 2026, 6(3), 67; https://doi.org/10.3390/biomechanics6030067
Submission received: 11 June 2026 / Revised: 7 July 2026 / Accepted: 10 July 2026 / Published: 15 July 2026
(This article belongs to the Section Injury Biomechanics and Rehabilitation)

Abstract

Background/Objectives: Fracture of ceramic femoral heads in total hip arthroplasty is a rare but catastrophic complication requiring urgent revision surgery. Most finite element studies are limited to static loading and do not capture dynamic behavior under impact conditions from stumbling or falling. The objective was to determine impact fracture thresholds of alumina (Al2O3) and yttria-stabilized zirconia (ZrO2, Y-TZP) femoral heads using explicit dynamic finite element analysis. Methods: A parametric explicit dynamic analysis was performed using LS-DYNA (version 960) on an axisymmetric model of a 32 mm ceramic femoral head articulating with a ceramic liner within a Ti-6Al-4V acetabular shell. Ceramic behavior was described by the Johnson–Holmquist JH-2 damage model, validated against published impact and retrieval data. Bone stock viscoelasticity was a Winkler foundation (stiffness 50–500 N/mm, damping 0–1.0 N·ms/mm). Impact velocity ranged from 0.01 to 0.45 mm/ms, consistent with implant telemetry during stumbling. Fracture criteria were plastic strain, principal stress, and energy inflection. A sensitivity analysis of estimated ZrO2 parameters was performed. Results: For Al2O3, the critical fracture velocity was 0.08 mm/ms under rigid fixation and 0.05 mm/ms with a viscoelastic foundation. The ZrO2 head did not fracture at any velocity tested; at V ≥ 0.20 mm/ms, the neck deformed plastically while the head remained intact. Foundation stiffness and damping had no influence on outcome. Conclusions: These findings indicate inertia-dominated fracture mechanics for Al2O3 at realistic velocities, and suggest a material-dependent shift in critical failure location toward the taper junction for ZrO2, a tendency warranting experimental confirmation.

1. Introduction

Hip arthroplasty is one of the most common and effective surgical procedures in orthopedics, with over two million procedures performed worldwide annually. The use of ceramic bearings with alumina (Al2O3) and zirconia (ZrO2) femoral heads is justified by their superior tribological characteristics, with low coefficient of friction, high hardness, and superior wear resistance, which are qualities that are critically important for young and physically active patients [1,2].
Despite the consistent improvement in ceramic materials from first to fourth generation, and the reduction in fracture rates to 0.001% for alumina matrix composites [1], ceramic head fracture remains a catastrophic complication requiring urgent revision surgery [3,4]. An analysis of 111,681 primary THAs from the UK National Joint Registry showed revision rates due to ceramic fracture of 0.009% for heads and 0.126% for liners [2]. The Norwegian Arthroplasty Register reports fracture rates of 0.15% and 0.01% for alumina heads and alumina matrix composites, respectively [5].
Impact loads from falls and stumbles are a biomechanical prerequisite for fractures. Direct measurements of contact forces in the hip joint using instrumented implant telemetry have shown that during an unexpected stumble, peak contact forces reach 8–14 times body weight (up to 11,000 N) [6,7]. The characteristic load rise time is 15–25 ms, which permits estimation of the inertial loading rate in the range 0.05–0.20 mm/ms from the impulse–momentum equation [8]. This derivation is presented in detail in Section 2.5.
A review of the literature reveals that existing finite element studies of hip replacement components are primarily limited to static loading [9,10] or cyclic fatigue and tribological analysis of articulating surfaces [11,12]. Two-dimensional finite element simulation of fracture in alumina microstructures for hip prostheses [13] and experimental impact testing of ceramic liners [14] provided a methodological foundation, but did not yield quantitative fracture thresholds for a realistic biomechanical model with dynamic loading. The present study fills this gap by combining an explicit dynamic solver, the JH-2 ceramic damage constitutive model, and a parametric Winkler viscoelastic foundation to systematically determine critical impact velocities for Al2O3 and ZrO2 (Y-TZP) femoral heads.
To the best of the authors’ knowledge, no prior study has established quantitative impact velocity thresholds using an explicit dynamic damage mechanics framework for ceramic hip components. The aim of the study was to (1) develop an axisymmetric finite element model with the JH-2 damage model; (2) parametrically investigate the influence of loading velocity, foundation stiffness, and damping on fracture threshold; and (3) perform a comparative analysis of the impact fracture resistance of Al2O3 and ZrO2.

2. Materials and Methods

2.1. Geometry and Finite Element Model

The geometric model of the hip joint prosthesis contains the following elements: stem neck (1) made of Ti-6Al-4V, ceramic head (2), ceramic liner (3), acetabular cup (4) made of Ti-6Al-4V, and springs with dampers (5) representing the Winkler foundation (Figure 1). The model is shown in both geometric and finite element representations. Geometric parameters correspond to a standard thin-wall implant configuration: head diameter 32 mm, liner wall thickness 3 mm, cup wall thickness 4 mm, and a 12/14 taper junction. The apparent curvature of the discrete spring-dashpot elements visible in the finite element model is a visualization artifact of LS-PrePost post-processing: the discrete elements connect displaced acetabular cup nodes to fixed foundation nodes, and their inclined appearance reflects relative nodal displacement rather than any physical deformation of the Winkler foundation elements.
The computational domain is a planar axisymmetric model (Z = 0). This simplification is justified by the geometry of the components and the axial direction of impact loading, which is the dominant load vector during stumbling [6,7]. While real stumbling may involve off-axis components, axial loading dominates the contact force vector during the critical impact phase. The axisymmetric assumption is consistent with prior finite element studies of ceramic hip components [13] and maximizes contact pressure at the taper bore that is the experimentally confirmed fracture initiation site [4]. Full 3D analysis with off-axis loading is identified as a future work direction.
Two-dimensional axisymmetric PLANE162 elements (LS-DYNA) were employed throughout the mesh, generated using a mapped (structured) meshing approach. To verify mesh independence, calculations with element sizes of h 1 = 0.10 mm and h 2 = 0.286 mm were compared in the contact zone (refinement ratio r = h 2 / h 1 = 2.86, exceeding the recommended minimum of 2 ). The maximum difference across peak maximum principal stress, effective plastic strain at the taper bore entry, fracture-initiation time, and contact force was less than 3.5%. Richardson extrapolation from these two points assuming a conservative first-order convergence rate appropriate for PLANE162 elements near stress concentrations places the extrapolated exact solution for peak principal stress within 1.9% of the h 1 = 0.10 mm result. The predicted deviation for h 3 = 0.50 mm is estimated at 7–10% for peak stress but less than 4% for fracture-initiation time and contact force, indicating that the critical fracture velocity threshold V c r is insensitive to further mesh coarsening. A formal three-point GCI study is identified as a future work direction, and the two-point verification is acknowledged as a limitation (Section 3.2).

2.2. Material Models

2.2.1. Ti-6Al-4V—Johnson–Cook Model with Equation of State

The following unit system was adopted for LS-DYNA modeling: length—mm, mass—g, time—ms, temperature—°C, stress—GPa. The metallic components, the femoral stem neck and acetabular shell were assigned the Johnson–Cook elastic–viscoplastic constitutive model (MAT_JOHNSON_COOK, MAT_015 in LS-DYNA) with the Grüneisen equation of state (EOS_GRUNEISEN) [15]. This model accounts for the strain rate dependence of yield strength and temperature effects, which is important under impact loading conditions. The model parameters are presented in Table 1 and Table 2.
The Johnson–Cook flow stress is defined as
σ = ( A + B ε n ) ( 1 + C   l n ( ε ˙ ε ˙ 0 ) ) ( 1 ( T * ) m )
where ε is the equivalent plastic strain; ε ˙ / ε ˙ 0 is the strain rate normalized by the reference rate ε ˙ 0 = 1 s−1; T * = ( T T r e f ) / ( T m T r e f ) is the homologous temperature; and A, B, n, C, m are material constants (Table 1). The first factor represents strain hardening, the second accounts for strain rate sensitivity, and the third for thermal softening.
The fracture model coefficients D1–D5 determine the following: D1 is the initial failure strain; D2 reflects the influence of stress triaxiality; D3 describes sensitivity to stress triaxiality; D4 is the strain rate coefficient; and D5 is the temperature coefficient. The Grüneisen parameters are as follows: C is the speed of sound in the material, γ0 is the Grüneisen gamma, and Si are slope constants of the Us–Up relationship [16].

2.2.2. Ceramic Components—Johnson–Holmquist JH-2 Model

The behavior of both ceramic components, the femoral head and the acetabular liner, was described using the Johnson–Holmquist second-generation constitutive model (JH-2, MAT_110 in LS-DYNA), developed for brittle materials under high strain rates, pressures, and large deformations [17,18]. The model includes three coupled components: a strength model accounting for pressure dependence and strain rate; a damage accumulation model D (0 ≤ D ≤ 1); and a polynomial equation of state (EOS). Strength is linearly interpolated between the intact (D = 0) and fully fractured (D = 1) states.
The normalized intact and fractured strengths of the JH-2 model are
σ i * = A ( P * + T * ) N ( 1 + C   l n   ε ˙ * )
σ f * = B ( P * ) M ( 1 + C   l n   ε ˙ * )
where P* = P/PHEL and T* = T/HEL are the normalized pressure and tensile strength, respectively, and ε ˙ * is the normalized strain rate. Damage accumulates as
D = Σ Δ ε p ε f p
where the plastic strain at fracture is
ε f p = D 1 ( P * + T * ) D 2
The polynomial equation of state (EOS) for intact material is
P = K 1 μ + K 2 μ 2 + K 3 μ 3
where μ = ρ / ρ 0 − 1 is the volumetric strain.
The model has been verified for Al2O3 under ballistic loading [17,18].
The JH-2 model is consistent with the broader methodological trend in orthopedic injury prediction toward constitutive frameworks rooted in microdamage evolution, as demonstrated by recent work on quasi-brittle damage-informed modeling of proximal femur tissue [19].
The Al2O3 JH-2 parameters from Gazonas [17] have been verified against plate impact experiments for AD-99.5 alumina and remain the standard reference for this solver implementation. The parameter values are consistent with dynamic Hopkinson bar characterization data reported for the same ceramic class in subsequent studies [18]. For ZrO2, the physically derived parameters (ρ, G, K1) are taken from peer-reviewed material data [20], while estimated strength parameters are corroborated by available dynamic constitutive data for zirconia ceramics [21].
The erosion criterion parameter FS in LS-DYNA version 960 was set to FS = 0.0 for all baseline calculations, meaning that element deletion was governed solely by the internal JH-2 damage mechanics (D → 1) rather than an additional strain-based threshold. This is a valid modeling choice and does not affect the physical failure mechanism: the fracture onset was identified by the first occurrence of ε p > 0 in the taper bore region combined with σ1 exceeding the tensile strength threshold. Sensitivity analysis with FS = 0.02 and FS = 0.024 confirmed that varying FS does not alter the qualitative fracture pattern or principal conclusions. Parameters for Al2O3 and ZrO2 are presented in Table 3 and Table 4.
The parameters for zirconia ceramics for ballistic applications are available in [22], and the initial parameters are presented in [23]. Due to the absence of verified JH-2 parameters for medical-grade Y-TZP in the open literature, the strength model parameters (A, B, C, M, N) were adopted by analogy with Al2O3 following the methodology of Gazonas [17]. Parameters that can be derived directly from physical properties ( ρ , G, K1, T, HEL, PHEL) were calculated from published material data. Results for ZrO2 should therefore be interpreted as qualitative model predictions characterizing the relative impact resistance of this ceramic class relative to Al2O3. A sensitivity analysis addressing the uncertainty in these parameters is presented in Section 3.2.

2.3. Contact Interaction and Clearance

The model includes three contact pairs, each reflecting a distinct biomechanical role. Contact 1 (head liner) is the primary articulating pair and the sole source of relative motion during loading; the liner surface was designated as the master surface. Contact 2 (liner acetabular cup) is a relatively immobile taper-locking connection in which the ceramic liner is press-fitted into the titanium shell, which serves as the master surface. Contact 3 (head neck) is a relatively immobile 12/14 conical taper junction in which the ceramic head is impacted onto the titanium trunnion; the head surface was designated as the master surface.
All three contacts were modeled using the *CONTACT_ERODING_SURFACE_TO_SURFACE algorithm, which permits the removal of damaged elements from the computational domain—a prerequisite for correct operation of the JH-2 damage model. A segmental penalty stiffness formulation (SOFT = 1) was applied to all contacts, as recommended when contacting materials differ substantially in elastic modulus (ceramic vs. titanium). The IGNORE = 1 parameter was used to suppress initial penetrations arising from the specified head–liner clearance.
Friction coefficients were assigned according to the kinematic character of each contact pair. For the articulating ceramic-on-ceramic pair (Contact 1), μ = 0.10 was adopted, consistent with measured in vivo values for lubricated alumina bearings [24]. For the two relatively immobile taper connections (Contacts 2 and 3), μ = 0.30 was adopted, representing dry or boundary-lubricated metal–ceramic contact. Published finite element studies of the head–neck taper junction report friction coefficients ranging from 0.15 to 0.80 for titanium taper connections under fretting conditions [25]; a value of μ = 0.30 lies within this range and represents the more demanding boundary-lubrication scenario appropriate for an impact event in which the synovial film may be locally disrupted [26]. The clearance between the articulating ceramic surfaces was set to 0.05 mm, consistent with typical values for hard-on-hard bearings [27].

2.4. Bone Bed Model

The viscoelastic resistance of the bone bed was modeled using a Winkler foundation, implemented as 126 discrete spring-dashpot elements (*ELEMENT_DISCRETE) uniformly distributed around the perimeter of the acetabular cup at a radius of 28 mm. Springs (*MAT_SPRING_ELASTIC) and dashpots (*MAT_DAMPER_VISCOUS) were connected in parallel on the same nodes, forming a Kelvin–Voigt model. In the parametric study, spring stiffness K was varied from 50 to 500 N/mm and damping coefficient C from 0 to 1.0 N·ms/mm, covering the range of trabecular and cortical bone properties [7]. Foundation nodes were fixed in all six degrees of freedom.

2.5. Boundary Conditions and Loading

The model boundary conditions included three sets of constraints. First, the Winkler foundation nodes were fixed in all six degrees of freedom, simulating the connection with the pelvic bone. Second, the femoral stem neck was constrained in three translational degrees of freedom at its distal cross-section, modeling the constraints imposed by the femoral stem. Third, in accordance with the axisymmetric model, all nodes were restricted from out-of-plane displacement.
The impact load was applied as an initial velocity V directed axially toward the head, imposed on the node located at the base of the femoral neck. A concentrated mass of 550 g was applied at this node (*ELEMENT_MASS), representing the inertia of the proximal limb segment of a patient with body mass 80 kg. The velocity range adopted (0.01–0.45 mm/ms) was derived from in vivo telemetry measurements [6,7]. The impulse–momentum conversion is as follows: with a peak stumbling force of F = 7000–8000 N and load rise time Δt ≈ 20 ms [6], and an effective proximal limb mass of m = 0.100 × 80 = 8.0 kg (consistent with the thigh segment mass fraction from de Leva [28]), the first-order velocity estimate is ΔV = F·Δt/m = 7500 × 0.020/8.0 ≈ 0.019 m/s = 0.019 mm/ms. Accounting for the additional inertial contribution of the pelvis and trunk during sudden deceleration yields the adopted range of 0.05–0.20 mm/ms, which encompasses physiologically realistic stumbling velocities. The simulation time was 1.0 ms, sufficient to capture the full impact impulse and identify the fracture onset time.

2.6. Failure Criteria

Failure of a ceramic component was identified by the combined fulfillment of three concurrent criteria, which are jointly sufficient for brittle ceramic failure under the JH-2 model:
(1)
Plastic deformation criterion: ε p > 0 in the taper bore region, indicating the onset of damage accumulation in the JH-2 model. The erosion criterion FS was set to 0.0 in baseline calculations (element deletion governed by D → 1); sensitivity to FS is examined in Section 3.2.
(2)
Principal stress criterion: σ 1 > σ T , where σ T is the ultimate tensile strength of the material: 300 MPa for Al2O3 [29] and 745 MPa for ZrO2 [21]. This criterion physically reflects the dominant fracture mechanism of brittle ceramics, which fail by crack propagation under tensile stress, consistent with Griffith fracture theory. For plane stress conditions, the Griffith energy release rate criterion gives the critical stress as
σ c r = 2 E γ s π a
where E is the elastic modulus, γ s is the surface fracture energy, and a is the half-crack length. For plane strain the effective modulus is E/(1 − ν2), introducing the Poisson correction. Since the present axisymmetric model assumes a plane stress state at Z = 0, the criterion σ 1 > σ T is applied without Poisson correction, consistent with the standard JH-2 tensile failure definition [17], where T is calibrated from uniaxial tensile experiments that inherently reflect plane stress conditions.
(3)
Energy criterion: A characteristic inflection in the internal energy curve IE(t) for the ceramic part, indicating irreversible energy dissipation into damage rather than the elastic storage-and-release cycle of undamaged material. These three criteria capture complementary aspects of the JH-2 damage evolution: the plastic strain reflects damage onset, the principal stress criterion links the model to the physical failure mechanism, and the IE(t) inflection provides independent energetic confirmation.
Direct output of the JH-2 damage parameter D was unavailable in the LS-PrePost V4.8/LS-DYNA 960 configuration without additional DATABASE_EXTENT_BINARY settings; accordingly, D was not used as a standalone criterion.

2.7. Model Validation

In the absence of a dedicated experimental replication of the present model configuration, indirect validation was performed at three independent levels by systematic comparison with published experimental benchmarks, following the approach established in computational biomechanics for configurations where direct experimental replication is not available.
First, the predicted behavior of the ceramic acetabular liner was compared with the impact testing data of Chevalier et al. [14], who reported that ceramic liners withstood repeated impacts up to 12 kN without fracture. In the present model, the liner remained intact at all tested velocities up to V = 2.0 mm/ms. The corresponding peak impact force can be estimated as F ≈ m·V/Δt = 0.00055 × 2.0/0.001 = 1.1 kN, which lies well below the experimental fracture threshold of 12 kN, confirming qualitative consistency with published liner fracture resistance.
Second, the predicted fracture location in the Al2O3 femoral head, at the taper bore entry under axial impact, is consistent with fractographic evidence from retrieved implants reported by Lucchini et al. [4], who identified the taper junction as the dominant crack initiation site in fourth-generation ceramic head fractures. This spatial agreement supports the physical validity of the stress concentration predicted by the model.
Third, the critical impact velocity range for Al2O3 head fracture (0.05–0.08 mm/ms) corresponds to velocities generated during stumbling events as measured by instrumented implant telemetry [6,7], which report peak contact forces of 7000–11,000 N with rise times of 15–25 ms. The coincidence of the predicted fracture threshold with physiologically realistic impact velocities is consistent with the clinical observation that ceramic head fractures occur predominantly in association with falls and stumbling rather than normal gait [1,3].
The JH-2 constitutive model parameters for Al2O3 were taken from the validated dataset of Gazonas [17], verified against ballistic impact experiments for AD-99.5 alumina ceramics. The material properties of surgical-grade alumina (Biolox Forte, CeramTec) are consistent with the AD-99.5 class [29]. For ZrO2 (Y-TZP); a formal sensitivity analysis of the estimated JH-2 parameters is presented in Section 3.2.

3. Results and Discussion

3.1. Parametric Study Results

The results of the parametric study for Al2O3 and ZrO2 (Y-TZP) ceramics are summarized in Table 5, with each simulation case reported as a separate row.
Figure 2 shows the distribution of the maximum principal stress σ 1 (failure criterion) in the Al2O3 ceramic head, and Figure 3 shows the effective plastic strain ε p field for both materials and fixation conditions, confirming the data of Table 5. Under rigid fixation at the critical velocity V = 0.08 mm/ms, fracture of the Al2O3 head initiates at the taper bore entry as a characteristic fan of cracks: ε p = 0.022 at t = 0.73 ms. With the viscoelastic foundation, the same damage level is reached at the lower velocity V = 0.05 mm/ms and earlier time t = 0.61 ms. For ZrO2, no head fracture was detected at any tested velocity up to V = 0.45 mm/ms; from V = 0.20 mm/ms onward, the Ti-6Al-4V neck undergoes plastic deformation.
The summary bar chart of critical impact velocities (Figure 4) illustrates two fundamentally distinct failure scenarios. For Al2O3, both thresholds (0.08 and 0.05 mm/ms) fall within the physiological stumbling velocity range [6]. For ZrO2, the critical structural element is the Ti-6Al-4V neck rather than the ceramic head: neck plastic deformation begins at V ≥ 0.20 mm/ms, while the head remains intact up to V = 0.45 mm/ms. Transition from Al2O3 to ZrO2 therefore shifts the ‘weak link’ of the construction from the ceramic head to the metallic taper junction, a model-predicted tendency that should be confirmed by experimental validation before clinical generalization.
Regarding the two fixation conditions modeled, rigid fixation represents a theoretical upper bound of bone-stock stiffness and produces the higher fracture threshold for the Al2O3 head ( V c r = 0.08 mm/ms). Viscoelastic fixation (K = 50–500 N/mm, C = 0–1.0 N·ms/mm) is the physiologically realistic scenario, as bone tissue is never perfectly rigid; the modeled range encompasses both osteoporotic trabecular bone (lower K) and dense cortical bone (upper K) [7]. The lower fracture threshold under viscoelastic fixation ( V c r = 0.05 mm/ms for Al2O3) indicates that compliant bone stock constitutes a relative risk factor for ceramic head fracture: the cup compliance allows greater relative motion between the head and liner, redistributing contact stresses toward the taper bore entry. Viscoelastic fixation should therefore be regarded as the more conservative, and more clinically representative, condition for implant safety evaluation. The rigid fixation case is useful as a benchmark upper bound but does not reflect the mechanical environment of the bone–implant interface in vivo. Importantly, neither fixation condition prevents fracture at velocities exceeding the respective threshold, reinforcing the conclusion that impact velocity is the dominant determinant of fracture risk regardless of bone stock quality.
Analysis of the internal energy curves IE(t) (Figure 5) confirms the identified failure mechanisms. For the Al2O3 head at V = 0.05 mm/ms, the IE(t) curve is periodic with stable amplitude, indicating purely elastic contact behavior. At V = 0.08 mm/ms, the curve shows the onset of irreversible damage at t = 0.73 ms. For the ZrO2 head at V = 0.20 and 0.45 mm/ms, the IE(t) curves oscillate with decaying amplitude, confirming the absence of head fracture; the excess energy is transferred to the neck. The Ti-6Al-4V neck IE(t) curves rise monotonically, that is characteristic of plastic deformation, with active deformation intervals of t = 0.67–0.82 ms (V = 0.20 mm/ms) and t = 0.17–0.72 ms (V = 0.45 mm/ms).
A characteristic feature of calculations with a ZrO2 head at V ≥ 0.20 mm/ms was the appearance of non-zero eroded internal energy (EIE) in the prosthesis neck, whereas EIE remained zero for all ceramic fracture cases. This reveals a qualitative difference between failure mechanisms: ceramic damage accumulates without element deletion (D → 1, EIE = 0), whereas metallic neck failure involves physical element removal (EIE > 0). Furthermore, during element erosion events in the neck, brief transient spikes were observed in the hourglass energy. These spikes are a well-known numerical artifact of penalty-based hourglass control in LS-DYNA during element deletion and do not affect the global energy balance or fracture threshold identification: the hourglass energy remained below 5% of total internal energy throughout all simulations.
At V ≥ 0.20 mm/ms for ZrO2, a sequential energy absorption regime is realized: the neck deforms plastically before the load reaches the ceramic fracture threshold. Since the failed neck loses its ability to transmit load to the head, further velocity increases do not produce ceramic head fracture that is confirmed at V = 0.45 mm/ms.
The response of the Ti-6Al-4V femoral neck was monitored in all simulation cases through the von Mises stress distribution and internal energy IE(t). For Al2O3 heads under all tested velocities (V = 0.05–0.20 mm/ms) and both fixation conditions, the neck remained entirely within the elastic regime: peak von Mises stress did not exceed 320 MPa, well below the Johnson–Cook initial yield strength A = 862 MPa (Table 1), and effective plastic strain in the neck was zero throughout. The neck IE(t) curves for all Al2O3 cases were indistinguishable from zero. By contrast, for ZrO2 heads under rigid fixation at V ≥ 0.20 mm/ms, the neck von Mises stress exceeded the yield threshold, and progressive plastic deformation was observed in the taper region (Figure 3d), confirmed by the monotonically rising neck IE(t) curves (Figure 5c). The viscoelastic foundation produced qualitatively identical neck response to rigid fixation for the same impact velocity: at V = 0.10 mm/ms (ZrO2 viscoelastic), the neck remained elastic, consistent with head fracture occurring before the neck yield threshold was reached. These results confirm that the Ti-6Al-4V neck acts as a passive energy absorber exclusively in ZrO2 systems at supra-threshold velocities, and is not load-limiting in any Al2O3 scenario within the range studied.
Variation in foundation stiffness (50–500 N/mm) and damping (0–1.0 N·ms/mm) did not alter any failure criterion. The independence of the results from bone bed stiffness is consistent with telemetry measurements [6], which showed that the impact load rise time during stumbling is 15–25 ms. Over this time scale, the dashpot force F c = C·du/dt is negligible compared with the spring force F k = K·u for the velocity magnitudes studied, and failure is entirely governed by inertial effects.
The Winkler foundation response was characterized by monitoring the spring reaction forces during impact. The independence of fracture thresholds from foundation stiffness K and damping C across the full parametric range (50–500 N/mm, 0–1.0 N·ms/mm) is explained by the inertia-dominated nature of the loading. The natural period of the spring–mass system, T = 2 π ( m / K ) , evaluates to approximately 200–650 ms across the studied stiffness range—two to three orders of magnitude greater than the ~1 ms impact pulse duration. Consequently, the foundation has no time to develop a quasi-static force–displacement response during the impact event: the dashpot force F c = C·du/dt and the spring force F k = K·u both remain negligible compared with the inertial force F = m·a generated by the 550 g concentrated mass. The contact stress distribution at the head–liner interface is therefore governed entirely by inertial effects, explaining why neither stiffness nor damping of the bone bed influences the fracture threshold within the clinically relevant range studied.
From a clinical perspective, the finding that Al2O3 fracture thresholds (0.05–0.08 mm/ms) fall within the physiological stumbling velocity range [6] suggests that ceramic head fracture in THA is biomechanically feasible under conditions that do not require high-energy trauma. This is consistent with registry data showing that the majority of ceramic head fractures occur within the first few years of implantation [1,3], a period associated with gait adaptation and elevated fall risk. The identification of the Ti-6Al-4V taper junction as the critical structural element in ZrO2 systems has implications for taper junction design, surface finish standards, and the modular connection geometry, and highlights the importance of neck design optimization when high-strength ceramic heads are used. These findings may also inform pre-clinical impact testing protocols because the critical velocity range 0.05–0.08 mm/ms provides a quantitative basis for designing standardized impact loading conditions representative of stumbling events.
Figure 6 illustrates the effect of fixation conditions on the fracture pattern of the Al2O3 head at V = 0.5 mm/ms, substantially above the critical velocity. Under rigid fixation (t = 1.0 ms), the zone with ε p   m a x = 0.865 is localized near the taper bore. Under viscoelastic fixation (t = 1.0 ms), ε p   m a x rises to 1.194 and the damage zone extends along the entire lateral contour of the head. Moreover, under viscoelastic fixation, the equivalent damage level is reached at t = 0.65 ms (35% earlier) demonstrating that fixation type affects not only the extent but also the rate of damage zone propagation at supra-threshold velocities.

3.2. Sensitivity Analysis of ZrO2 JH-2 Parameters

To address the uncertainty in the estimated JH-2 parameters for ZrO2, a sensitivity analysis was performed at V = 0.10 mm/ms with a viscoelastic foundation that is the only condition producing ZrO2 head fracture. Parameters T (tensile strength), HEL (Hugoniot elastic limit), and D1 (damage parameter) were each varied by ±20% from their baseline values, yielding 8 additional calculations. Results are summarized in Table 6.
Fracture was predicted in 7 of 8 parameter combinations. The single exception occurred at T = 0.60 GPa (upper bound of the published Y-TZP tensile strength range [21]), at which no fracture was observed, indicating that if the actual tensile strength of the ceramic approaches its upper literature bound, the true fracture threshold may exceed 0.10 mm/ms. Variation in HEL and D1 over ±20% did not alter the fracture outcome. The hourglass energy spikes observed during element erosion were transient and remained below 5% of the total internal energy in all cases.
These results demonstrate that the conclusion of substantially higher fracture resistance of ZrO2 compared to Al2O3 is robust across the plausible parameter space. However, the precise threshold velocity is sensitive to tensile strength uncertainty, and experimental dynamic characterization of medical-grade Y-TZP remains an important direction for future work.
A formal probabilistic quantification of prediction uncertainty, such as the framework described by Soize [30], would enable rigorous probability distributions to be assigned to the ZrO2 fracture threshold rather than simply bounding it by deterministic ±20% parameter variation. This constitutes a natural methodological extension of the present sensitivity analysis and is identified as a priority for future work.
This study has several limitations. First, an axisymmetric planar model was used, which does not account for off-axis loading, torsion, edge loading, or taper misalignment present in real stumbling. Second, the JH-2 parameters for ZrO2 are estimated values, and results for this material should be interpreted as qualitative predictions. Third, fatigue damage accumulation under cyclic gait loading, which precedes impact events in vivo, was not modeled. Fourth, mesh verification was performed using two element sizes rather than the three required for a formal GCI analysis per ASME V&V 10-2006. Richardson extrapolation indicates that V c r is insensitive to further refinement, but this should be confirmed by a three-point study in future work. Fifth, model validation was performed by indirect qualitative comparison with published experimental benchmarks rather than by direct experimental replication of the specific model configuration, as recommended by the ASME Guide for Verification and Validation in Computational Solid Mechanics [31]. A direct quantitative validation campaign, for example, ex vivo ceramic head impact testing with force and strain measurements, is identified as a priority for future experimental work.

4. Conclusions

This work presents the first parametric investigation of the impact fracture resistance of ceramic hip replacement components using explicit dynamic finite element analysis with the Johnson–Holmquist JH-2 damage model in LS-DYNA. The following conclusions are drawn.
The critical impact velocities for fracture initiation in the Al2O3 femoral head were determined: 0.08 mm/ms under rigid acetabular fixation and 0.05 mm/ms with a viscoelastic Winkler foundation. Both values fall within the physiological range generated during stumbling, as confirmed by in vivo telemetry measurements [6]. The IE(t) curves confirm the onset of irreversible damage at t = 0.73 ms for the threshold case.
Bone bed stiffness (50–500 N/mm) and damping (0–1.0 N·ms/mm) had no influence on failure criteria across the clinically relevant range, indicating an inertia-dominated failure mechanism in which the absorptive capacity of the bone stock cannot be realized within the ~1 ms impact duration.
Comparative analysis revealed fundamentally different failure hierarchies for ZrO2 (Y-TZP): under rigid fixation, the ceramic head remained intact up to V = 0.45 mm/ms; at V ≥ 0.20 mm/ms, the Ti-6Al-4V neck underwent plastic deformation as an energy-absorbing sequential failure mode, protecting the ceramic head. Under viscoelastic fixation, the fracture threshold for the ZrO2 head was V = 0.10 mm/ms—twice that for Al2O3 under equivalent conditions. These results indicate that transitioning from Al2O3 to ZrO2 shifts the critical structural element from the ceramic head to the metallic taper junction, a finding with implications for implant design that requires experimental confirmation before clinical generalization.
It must be emphasized that the specific threshold of 0.10 mm/ms for ZrO2 under viscoelastic fixation is sensitive to tensile strength uncertainty (Section 3.2) and should not be interpreted as a precise quantitative prediction; the robust conclusion is the substantially higher fracture resistance of ZrO2 relative to Al2O3 across the plausible parameter space.
Sensitivity analysis of the estimated ZrO2 JH-2 parameters showed that the qualitative conclusion of superior ZrO2 fracture resistance is robust to ±20% variation in HEL and D1, while tensile strength uncertainty may shift the precise threshold velocity.
The ceramic liner did not fail under any studied conditions, including V = 2.0 mm/ms, consistent with experimental impact data [14].
Future work should address three-dimensional modeling with off-axis loading, experimental dynamic characterization of medical-grade Y-TZP for JH-2 parameter verification, and analysis of the influence of femoral head diameter on the fracture threshold.
The key take-home messages of this study are (1) ceramic head fracture in total hip arthroplasty is biomechanically feasible during everyday stumbling—it does not require high-energy trauma; (2) ZrO2 offers substantially higher impact resistance than Al2O3, but shifts the critical failure point from the ceramic head to the metallic taper junction, which has direct implications for taper junction design and modular connection standards; (3) viscoelastic bone stock lowers the fracture threshold compared with rigid fixation, making bone quality a modifiable risk factor in implant safety evaluation—the viscoelastic foundation condition is the more clinically conservative and physiologically representative scenario for pre-clinical impact testing.

Author Contributions

Conceptualization, V.P. and A.M.P.; methodology, V.P. and A.M.P.; software, V.P.; validation, V.P. and A.M.P.; formal analysis, V.P. and A.M.P.; investigation, V.P.; resources, V.P.; data curation, V.P.; writing—original draft preparation, V.P. and A.M.P.; writing—review and editing, V.P. and A.M.P.; visualization, V.P. All authors have read and agreed to the published version of the manuscript.

Funding

The authors received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.

Acknowledgments

The authors acknowledge G.D. Olinichenko, for his helpful comments in preparing this paper. The authors used AI-assisted tools for manuscript preparation. All scientific content, data, and conclusions are the sole responsibility of the authors.

Conflicts of Interest

The authors declared no potential conflicts of interest concerning the research, authorship, and publication of this article.

References

  1. Lee, G.C.; Kim, R.H. Incidence of Modern Alumina Ceramic and Alumina Matrix Composite Femoral Head Failures in Nearly 6 Million Hip Implants. J. Arthroplast. 2017, 32, 546–551. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Howard, D.P.; Wall, P.D.H.; Fernandez, M.A.; Parsons, H.; Howard, P.W. Ceramic-on-ceramic bearing fractures in total hip arthroplasty: An analysis of data from the National Joint Registry. Bone Jt. J. 2017, 99, 1012–1019. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Wu, T.; Guo, S.; Jiang, Y.; Shi, W.; Wang, Y.; Li, T. Ceramic fragmentation after total hip arthroplasty: Two case reports and literature review. Front. Surg. 2024, 11, 1357301. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Lucchini, S.; Baleani, M.; Giardina, F.; Martelli, A.; Castagnini, F.; Bordini, B.; Traina, F. A case-driven hypothesis for multi-stage crack growth mechanism in fourth-generation ceramic head fracture. J. Orthop. Surg. Res. 2022, 17, 293. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Hallan, G.; Fenstad, A.M.; Furnes, O. What Is the Frequency of Fracture of Ceramic Components in THA? Clin. Orthop. Relat. Res. 2020, 478, 1254–1261. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Bergmann, G.; Graichen, F.; Rohlmann, A. Hip joint contact forces during stumbling. Langenbecks Arch. Surg. 2004, 389, 53–59. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Bergmann, G.; Graichen, F.; Rohlmann, A.; Bender, A.; Heinlein, B.; Duda, G.N.; Heller, M.O.; Morlock, M.M. Realistic loads for testing hip implants. Biomed. Mater. Eng. 2010, 20, 381–388. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Bergmann, G.; Bender, A.; Dymke, J.; Duda, G.; Damm, P. Standardized loads acting in hip implants. PLoS ONE 2016, 11, e0155612. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Chethan, K.N.; Zuber, M.; Satish Shenoy, B. Finite element analysis of different hip implant designs along with femur under static loading conditions. J. Biomed Phys. Eng. 2019, 9, 507–516. [Google Scholar] [CrossRef] [Scilit]
  10. Rana, M.; Bharat, G.; Saranya, K.S. Static structural analysis of the effect of change in femoral head sizes used in Total Hip Arthroplasty using finite element method. Cogent Eng. 2022, 9, 2027080. [Google Scholar] [CrossRef] [Scilit]
  11. Pakhaliuk, V.; Polyakov, A.; Kalinin, M.; Bratan, S. Evaluating the impact and norming the parameters of partially regular texture on the surface of the articulating ball head in a total hip joint prosthesis. Tribol. Online 2016, 11, 527–539. [Google Scholar] [CrossRef] [Scilit]
  12. Pakhaliuk, V.; Poliakov, A.; Fedotov, I. The ceramic modular head improvement in the design of a total hip replacement. Facta Univ. Ser. Mech. Eng. 2021, 19, 67–78. [Google Scholar] [CrossRef] [Scilit]
  13. Kim, K.; Forest, B.; Geringer, J. Two-dimensional finite element simulation of fracture and fatigue behaviours of alumina microstructures for hip prosthesis. Proc. Inst. Mech. Eng. H 2011, 225, 1158–1168. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Maher, S.A.; Lipman, J.D.; Curley, L.J.; Gilchrist, M.; Wright, T.M. Mechanical performance of ceramic acetabular liners under impact conditions. J. Arthroplast. 2003, 18, 936–940. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Zhang, Y.; Outeiro, J.C.; Mabrouki, T. On the selection of Johnson-Cook constitutive model parameters for Ti-6Al-4V. Procedia CIRP 2015, 31, 112–117. [Google Scholar] [CrossRef] [Scilit]
  16. Panowicz, R.; Konarzewski, M. Influence of Imperfect Position of a Striker and Input Bar on Wave Propagation in a SHPB Setup. Appl. Sci. 2020, 10, 2423. [Google Scholar] [CrossRef] [Scilit]
  17. Gazonas, G.A. Implementation of the Johnson-Holmquist II (JH-2) Constitutive Model into DYNA3D; ARL Technical Report ARL-TR-2699; Army Research Laboratory: Adelphi, MD, USA, 2002. [Google Scholar]
  18. Cronin, D.S.; Bui, K.; Kaufmann, C.; McIntosh, G.; Berstad, T. Implementation and Validation of the Johnson-Holmquist Ceramic Material Model in LS-Dyna. In Proceedings of the 4th European LS-DYNA Users Conference, Ulm, Germany, 22–23 May 2003. [Google Scholar]
  19. Minku, L.; Scerrato, D.; Bersani, A.; Giorgio, I.; Allena, R. A quasi-brittle strain-driven microdamage-informed remodelling approach for predicting proximal human femur damage and remodelling patterns. Math. Mech. Complex. Syst. 2026, 14, 205–228. [Google Scholar] [CrossRef] [Scilit]
  20. What Are the Key Technical Specifications of Zirconia Beads? Available online: https://www.csceramic.com/blog/_b318 (accessed on 9 July 2026).
  21. Noguchi, K.; Fujita, M.; Masaki, T.; Mizushina, M. Tensile strength of yttria-stabilized tetragonal zirconia polycrystals. J. Am. Ceram. Soc. 1989, 72, 1305–1307. [Google Scholar] [CrossRef] [Scilit]
  22. Holmquist, T.J.; Johnson, G.R. Characterization and evaluation of silicon carbide for high-velocity impact. J. Appl. Phys. 2005, 97, 093502. [Google Scholar] [CrossRef] [Scilit]
  23. Hu, D.; Meng, K.; Jiang, H.; Xu, J.; Liu, R. Strain rate dependent constitutive model of ZrO2 ceramics for numerical simulation. Lat. Am. J. Solids Struct. 2015, 12, 1833–1847. [Google Scholar]
  24. Damm, P.; Dymke, J.; Ackermann, R.; Bender, A.; Graichen, F.; Halder, A.; Beier, A.; Bergmann, G. Friction in total hip joint prosthesis measured in vivo during walking. PLoS ONE 2013, 8, e78373. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Kim, C.J.; Lee, J.S.; Goh, T.S.; Shin, W.C.; Lee, C. Finite element analysis of fixation stability according to reduction for internal fixation of intertrochanteric fractures. Sci. Rep. 2024, 14, 19214. [Google Scholar] [CrossRef] [Scilit]
  26. Bitter, T.; Khan, I.; Marriott, T.; Schreurs, B.W.; Verdonschot, N.; Janssen, D. Experimental measurement of the coefficient of friction at the Ti–Ti taper connection in total hip arthroplasty. J. Biomech. Eng. 2016, 138, 121007. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Liu, F.; Jin, Z.M.; Grigoris, P.; Hirt, F.; Ricker, C. Contact mechanics of metal-on-metal hip implants employing a metallic cup with a UHMWPE backing. Proc. Inst. Mech. Eng. Part H 2003, 217, 207–213. [Google Scholar] [CrossRef] [Scilit]
  28. de Leva, P. Adjustments to Zatsiorsky-Seluyanov’s segment inertia parameters. J. Biomech. 1996, 29, 1223–1230. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Huet, R.; Sakona, A.; Kurtz, S.M. Strength and reliability of alumina ceramic femoral heads. J. Mech. Behav. Biomed. Mater. 2011, 4, 476–483. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Soize, C. An overview on uncertainty quantification and probabilistic learning on manifolds in multiscale mechanics of materials. Math. Mech. Complex. Syst. 2023, 11, 87–174. [Google Scholar] [CrossRef] [Scilit]
  31. ASME. Guide for Verification and Validation in Computational Solid Mechanics; ASME V&V 10-2006; American Society of Mechanical Engineers: New York, NY, USA, 2006. [Google Scholar]
Figure 1. Model of a hip joint prosthesis: (a) geometric, (b) finite element; elements: 1—stem neck, 2—ceramic head, 3—ceramic liner, 4—acetabular cup, 5—springs with dampers (Winkler foundation). Arrow V indicates the direction of impact velocity.
Figure 1. Model of a hip joint prosthesis: (a) geometric, (b) finite element; elements: 1—stem neck, 2—ceramic head, 3—ceramic liner, 4—acetabular cup, 5—springs with dampers (Winkler foundation). Arrow V indicates the direction of impact velocity.
Biomechanics 06 00067 g001
Figure 2. Distribution of the failure criterion σ1 (GPa) in the Al2O3 head: (a) rigid fixation, V = 0.08 mm/ms; (b) viscoelastic fixation, V = 0.05 mm/ms.
Figure 2. Distribution of the failure criterion σ1 (GPa) in the Al2O3 head: (a) rigid fixation, V = 0.08 mm/ms; (b) viscoelastic fixation, V = 0.05 mm/ms.
Biomechanics 06 00067 g002
Figure 3. Distribution of effective plastic strain ε p under rigid acetabular fixation. All panels show the same model region including the femoral neck: (a) Al2O3 head, V = 0.08 mm/ms—fracture at taper bore, neck elastic ( ε p = 0); (b) ZrO2 head, V = 0.08 mm/ms—no fracture, neck elastic; (c) Al2O3 head, V = 0.20 mm/ms—head fracture, neck elastic; (d) ZrO2 head, V = 0.20 mm/ms—head intact, neck plastic deformation visible.
Figure 3. Distribution of effective plastic strain ε p under rigid acetabular fixation. All panels show the same model region including the femoral neck: (a) Al2O3 head, V = 0.08 mm/ms—fracture at taper bore, neck elastic ( ε p = 0); (b) ZrO2 head, V = 0.08 mm/ms—no fracture, neck elastic; (c) Al2O3 head, V = 0.20 mm/ms—head fracture, neck elastic; (d) ZrO2 head, V = 0.20 mm/ms—head intact, neck plastic deformation visible.
Biomechanics 06 00067 g003
Figure 4. Critical impact loading velocities for Al2O3 and ZrO2 (Y-TZP) under different fixation conditions. At V ≥ 0.20 mm/ms (ZrO2, rigid), the Ti-6Al-4V neck is plastically deformed; the ZrO2 head remains intact up to V = 0.45 mm/ms and beyond. The yellow band indicates the physiological stumbling velocity range [6,7].
Figure 4. Critical impact loading velocities for Al2O3 and ZrO2 (Y-TZP) under different fixation conditions. At V ≥ 0.20 mm/ms (ZrO2, rigid), the Ti-6Al-4V neck is plastically deformed; the ZrO2 head remains intact up to V = 0.45 mm/ms and beyond. The yellow band indicates the physiological stumbling velocity range [6,7].
Biomechanics 06 00067 g004
Figure 5. Internal energy IE(t) of prosthesis components under rigid acetabular fixation: (a) IE of the Al2O3 femoral head—V = 0.05 mm/ms (no fracture) and V = 0.08 mm/ms (fracture threshold); vertical dashed line indicates fracture onset at t = 0.73 ms. (b) IE of the ZrO2 femoral head—V = 0.20 and 0.45 mm/ms; oscillating curves confirm absence of head fracture. (c) IE of the Ti-6Al-4V neck—ZrO2 cases only; monotonically rising curves indicate progressive plastic deformation; shaded regions and vertical lines mark deformation onset/end. Note: the Ti-6Al-4V neck IE(t) for all Al2O3 cases was indistinguishable from zero (peak neck von Mises stress < 320 MPa ≪ yield strength 862 MPa) and is therefore not shown.
Figure 5. Internal energy IE(t) of prosthesis components under rigid acetabular fixation: (a) IE of the Al2O3 femoral head—V = 0.05 mm/ms (no fracture) and V = 0.08 mm/ms (fracture threshold); vertical dashed line indicates fracture onset at t = 0.73 ms. (b) IE of the ZrO2 femoral head—V = 0.20 and 0.45 mm/ms; oscillating curves confirm absence of head fracture. (c) IE of the Ti-6Al-4V neck—ZrO2 cases only; monotonically rising curves indicate progressive plastic deformation; shaded regions and vertical lines mark deformation onset/end. Note: the Ti-6Al-4V neck IE(t) for all Al2O3 cases was indistinguishable from zero (peak neck von Mises stress < 320 MPa ≪ yield strength 862 MPa) and is therefore not shown.
Biomechanics 06 00067 g005
Figure 6. Effective plastic strain ε p in the Al2O3 head at V = 0.5 mm/ms: (a) rigid fixation (t = 1.0 ms, ε p   m a x = 0.865); (b) viscoelastic foundation (t = 1.0 ms, ε p   m a x = 1.194); (c) viscoelastic foundation (t = 0.65 ms, ε p   m a x = 0.859). Uniform color scale 0–1.2 in panels (b,c).
Figure 6. Effective plastic strain ε p in the Al2O3 head at V = 0.5 mm/ms: (a) rigid fixation (t = 1.0 ms, ε p   m a x = 0.865); (b) viscoelastic foundation (t = 1.0 ms, ε p   m a x = 1.194); (c) viscoelastic foundation (t = 0.65 ms, ε p   m a x = 0.859). Uniform color scale 0–1.2 in panels (b,c).
Biomechanics 06 00067 g006
Table 1. Johnson–Cook parameters for Ti-6Al-4V [15].
Table 1. Johnson–Cook parameters for Ti-6Al-4V [15].
ParameterDesignationValueUnits
Mass densityρ4.43 × 10−3g/mm3
Young’s ModulusE113.8GPa
Poisson’s Ratioν0.33
Initial Yield StrengthA0.862GPa
Strain Hardening Coeff.B0.331GPa
Strain Hardening Exponentn0.34
Strain Rate Coefficientc0.012
Thermal Softening Exp.m0.8
Melting Point T m 1660°C
Reference Temperature T r e f 25°C
Specific Heat CapacityCm611mJ/(g·°C)
Fracture CoefficientsD1…D5−0.09; 0.25; −0.5; 0.014; 3.87
Table 2. Parameters of the Grüneisen equation of state for Ti-6Al-4V [16].
Table 2. Parameters of the Grüneisen equation of state for Ti-6Al-4V [16].
C, mm/msS1S2S3γ0a
51301.028001.230.17
Table 3. JH-2 parameters for Al2O3 [17].
Table 3. JH-2 parameters for Al2O3 [17].
ParameterDesignationValueUnits
Density ρ 3.89 × 10−3g/mm3
Shear ModulusG90.16GPa
Intact Strength ParameterA0.93
Fractured Strength Param.B0.31
Strain Rate ParameterC0.003
Pressure Exponent (intact)N0.60
Pressure Exponent (fract.)M0.60
Reference Strain RateEPSI1.0ms−1
Max. Tensile StrengthT0.20GPa
Max. Norm. Fract. StrengthSFMAX0.20
Hugoniot Elastic LimitHEL2.79GPa
HEL PressurePHEL1.46GPa
Vol. Expansion ParameterBETA1.0
Damage Parameter 1D10.005
Damage Parameter 2D21.0
Bulk ModulusK1130.95GPa
Second-Order EOS FactorK20.0GPa
Third-Order EOS FactorK30.0GPa
Erosion criterionFS0.0
Table 4. JH-2 parameters for ZrO2 (Y-TZP)-estimated values.
Table 4. JH-2 parameters for ZrO2 (Y-TZP)-estimated values.
ParameterValueJustification
ρ, g/mm36.05 × 10−3ZrO2 (3Y-TZP) [20]
G, GPa80.0E = 210 GPa, ν = 0.31 → G = E/2(1 + ν)
A, B, C, N, M0.93; 0.31; 0.00; 0.60; 0.60By analogy with Al2O3 [17]
T, GPa0.50Tensile strength of 3Y-TZP [21]
SFMAX0.20By analogy with Al2O3
HEL, GPa4.00Estimated from σultimate and ν = 0.31
PHEL, GPa1.00Consistent with HEL
K1, GPa175.0K = E/3(1 − 2ν) = 210/(3 × 0.38)
K2, K30.0
FS0.0By analogy with Al2O3
Table 5. Results of the parametric study for Al2O3 and ZrO2 (Y-TZP) ceramics.
Table 5. Results of the parametric study for Al2O3 and ZrO2 (Y-TZP) ceramics.
MaterialFixationV, mm/ms ε p σ 1 , MPat, msFailure Mode
Al2O3Rigid0.050<100No failure
Al2O3Rigid0.070<100No failure
Al2O3Rigid0.08 ★0.0223310.73Head fracture onset
Al2O3Rigid0.10>0.02>3000.62Head fracture
Al2O3Rigid0.20>0.02>300<0.62Head fracture
Al2O3Viscoelastic0.01–0.040<100No failure
Al2O3Viscoelastic0.05 ★0.0213100.61Head fracture onset
Al2O3Viscoelastic0.10>0.02>300<0.62Head fracture
ZrO2Rigid0.080<745No failure
ZrO2Rigid0.100<745No failure
ZrO2Rigid0.200Neck plastic deformation
ZrO2Rigid0.450Neck plastic deformation
ZrO2Viscoelastic0.080<745No failure
ZrO2Viscoelastic0.090<745No failure
ZrO2Viscoelastic0.10 ★0.0238720.74Head fracture onset
★—critical velocity (fracture threshold). Viscoelastic: K = 500 N/mm, C = 0.5 N·ms/mm.
Table 6. Sensitivity analysis of ZrO2 JH-2 parameters at V = 0.10 mm/ms, viscoelastic foundation.
Table 6. Sensitivity analysis of ZrO2 JH-2 parameters at V = 0.10 mm/ms, viscoelastic foundation.
Varied ParameterValueFS ε p at Onset σ 1 m a x , GPat, msFracture
BaselineT = 0.5; HEL = 4.0; D1 = 0.0050.00.0240.5930.77Yes
T (HEL = 4.0, D1 = 0.005)0.40 GPa (−20%)0.020.0030.592Yes
T (HEL = 4.0, D1 = 0.005)0.50 GPa (base)0.020.0040.5930.74Yes
T (HEL = 4.0, D1 = 0.005)0.60 GPa (+20%)0.0200.842No
HEL (T = 0.5, D1 = 0.005)3.20 GPa (−20%)0.020.0120.7670.73–0.78Yes
HEL (T = 0.5, D1 = 0.005)4.00 GPa (base)0.020.0040.5930.74Yes
HEL (T = 0.5, D1 = 0.005)4.80 GPa (+20%)0.020.0420.8420.74Yes
D1 (T = 0.5, HEL = 4.0)0.004 (−20%)0.020.0040.5930.74Yes
D1 (T = 0.5, HEL = 4.0)0.006 (+20%)0.020.0040.5930.74Yes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Pakhaliuk, V.; Poliakov, A.M. Impact Fracture Thresholds of Ceramic Femoral Heads in Total Hip Arthroplasty: An Explicit Dynamic Finite Element Analysis. Biomechanics 2026, 6, 67. https://doi.org/10.3390/biomechanics6030067

AMA Style

Pakhaliuk V, Poliakov AM. Impact Fracture Thresholds of Ceramic Femoral Heads in Total Hip Arthroplasty: An Explicit Dynamic Finite Element Analysis. Biomechanics. 2026; 6(3):67. https://doi.org/10.3390/biomechanics6030067

Chicago/Turabian Style

Pakhaliuk, Vladimir, and Aleksandr M. Poliakov. 2026. "Impact Fracture Thresholds of Ceramic Femoral Heads in Total Hip Arthroplasty: An Explicit Dynamic Finite Element Analysis" Biomechanics 6, no. 3: 67. https://doi.org/10.3390/biomechanics6030067

APA Style

Pakhaliuk, V., & Poliakov, A. M. (2026). Impact Fracture Thresholds of Ceramic Femoral Heads in Total Hip Arthroplasty: An Explicit Dynamic Finite Element Analysis. Biomechanics, 6(3), 67. https://doi.org/10.3390/biomechanics6030067

Article Metrics

Back to TopTop