Next Article in Journal
An Effective Dust Collection Tray and Its Performance Optimized for Compact Sweepers Based on CFD-RSM Method
Previous Article in Journal
Robust Maximum Half-Normal Multivariate Control Chart Based on Det-MCD and Fast-MCD Estimators
Previous Article in Special Issue
The Effect of Scatter Radiation on Image Resolution in Gridless Portable X-Ray Imaging: A Monte Carlo Study
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Energy-Resolved CNR Performance in Dense-Breast and Implant X-Ray Mammography Using a CdTe Photon-Counting Detector: A Monte Carlo Study

by
Gerardo Roque
1,*,
Maria Laura Pérez-Lara
2,
Steven Cely
3,
Juan Sebastián Useche Parra
1,
Jesús David Bermúdez
3,
Michael K. Schütz
1,
Michael Fiederle
1,
Carlos Ávila
3 and
Simon Procz
1
1
Freiburg Materials Research Center, Albert-Ludwigs Universität Freiburg, Stefan-Meier St. 21, 79104 Freiburg im Breisgau, Germany
2
UCL Medical Physics and Biomedical Engineering, University College London, Gower Street, London WC1E 6BT, UK
3
Department of Physics, Universidad de Los Andes, Cra. 1 #18ª-12, Bogotá 111711, Colombia
*
Author to whom correspondence should be addressed.
Appl. Sci. 2026, 16(7), 3550; https://doi.org/10.3390/app16073550
Submission received: 23 February 2026 / Revised: 31 March 2026 / Accepted: 2 April 2026 / Published: 5 April 2026

Featured Application

Improved mammography image quality through the use of spectroscopic information to optimize the X-ray tube spectrum and implement energy-based contrast-optimization algorithms.

Abstract

X-ray imaging of dense breasts and breast implants often suffers from reduced lesion visibility because strong attenuation lowers contrast, while conventional rhodium (Rh) K-edge filtering suppresses part of the high-energy spectral tail. This study presents a Monte Carlo framework for spectroscopic mammography using a voxelated 1 mm thick cadmium telluride (CdTe) sensor and a first-order detector interaction model to evaluate energy-dependent image quality. The model reproduces fluorescence and inter-voxel energy redistribution in CdTe, but not the full detector chain, and remains idealized with respect to charge transport, carrier collection, threshold dispersion, and pile-up. Energy-resolved simulations in the 10–50 keV range were used to compute spectroscopic contrast-to-noise ratio (CNR) curves and to form integrated-spectrum (IS) images for four tested spectra. For the dense-breast calcium hydroxyapatite (HA) speck detection task considered here, and under the present simulation assumptions, replacing the standard 28 kVp + 50 μm Rh spectrum with 28 kVp + 1 mm Al increased the simulated IS image CNR by 23.11%, with an approximately 5% increase in estimated primary-incident air kerma at the phantom entrance plane. Preliminary experimental implant-phantom images were included as a qualitative feasibility check, showing a trend consistent with simulations. Within the limits of this task-specific simulation, the results suggest that preserving the transmitted high-energy tail can improve HA speck visibility for the present 1 mm CdTe photon-counting detector, with the 28 kVp + 1 mm Al spectrum outperforming the other tested cases.

1. Introduction

The goal of mammography is the early detection of breast cancer precursors by screening for microscopic calcifications, here referred to as specks, composed of calcium hydroxyapatite (HA) [1,2]. X-ray mammography has demonstrated strong clinical performance for identifying early-stage breast cancer over the past decades [3]. However, the high X-ray absorption due to heterogeneously distributed dense tissue morphologies, or the presence of breast implants, limits image quality and reduces lesion visibility, leading to decreased detection sensitivity. This can lead to both false-positive and false-negative findings and reduce diagnostic performance in women with breast implants [4] or very dense breasts [5].
Imaging of dense breasts remains particularly challenging despite technological advances. This limitation is clinically relevant because women with dense breasts have an elevated cancer risk, and more than 50% of women under 50 years old exhibit high mammographic density [6,7,8,9]. Increased attenuation reduces contrast between lesions and surrounding tissues, often requiring higher radiation exposure to maintain image quality. Alternative modalities, such as digital breast tomosynthesis, MRI, ultrasound, and molecular breast imaging, have been proposed; however, these approaches involve higher costs, longer acquisition times, or limited accessibility [10]. Consequently, ultrasound is often used as an adjunct to mammography to improve detection rates [11,12].
Breast augmentation surgery has also become increasingly common. It is estimated that 35 million women worldwide have breast implants [13], and recent reports indicate continued growth in cosmetic breast procedures [14]. Implants significantly attenuate X-rays due to their high silicon content, obscuring between 22% and 83% of the glandular tissue. Although specialized displacement techniques exist to improve visualization [15], these procedures are technically challenging, increase radiation exposure, and may elevate the risk of implant damage [16]. As a result, screening mammography may miss a substantially higher proportion of cancers in women with implants compared to women without augmentation [17].
Given the large population of patients with dense or augmented breasts, optimizing mammographic imaging for these conditions is of practical importance. Simply increasing exposure is not desirable due to radiation risk. Therefore, alternative detector technologies and spectrum optimization strategies are required to enhance contrast while maintaining similar or lower radiation exposure levels.
From a technological perspective, conventional mammography systems employ charge-integrating flat-panel detectors and K-edge filtering to tailor the emitted spectrum to detector characteristics. However, the high-energy spectral tail can degrade image quality in charge-integrating systems because higher-energy photons contribute disproportionately to the collected charge, reducing dynamic range and lesion contrast. Photon-counting detectors (PCDs) mitigate this challenge by assigning equal weight to each detected photon, independent of energy. When combined with high-efficiency sensor materials, PCDs can exploit higher-energy photons to improve visibility through dense tissue and implants [18]. Spectroscopic photon-counting detectors (SPCDs), such as Timepix3 [19], additionally measure photon energy, enabling energy-resolved imaging and spectral optimization. The spectral grading identified here for the present CdTe SPCD should therefore not be transferred directly to conventional charge-integrating detectors, which weight the spectrum differently.
Previous studies have demonstrated that CdTe-based photon-counting detectors can improve image quality and potentially reduce radiation dose in mammography [20,21,22,23]. Earlier implementations using strip detectors and scanning geometries also indicated dose reductions [24], while simulation studies suggested further improvements through energy-resolved imaging and material decomposition [25]. Other related studies on spectral mammography can be classified in two groups: those that use spectroscopic information to improve image quality and those that use it for material decomposition and classification. Studies have shown that spectroscopic photon-counting systems can benefit from energy weighting, observer-model optimization, task-based beam quality optimization in dual-energy imaging systems, and dose optimizations, while also enabling the classification of breast microcalcifications from energy-resolved measurements [25,26,27,28,29,30,31,32]. These results highlight the potential of energy-resolved photon-counting technology for enhancing mammographic performance and have motivated ongoing industrial development of next-generation systems [33,34,35,36].
While earlier mammography studies with photon-counting detectors have already suggested benefits from harder spectra and energy-resolved imaging [20,21,22,23,24,25], the specific reason why Rh filtration becomes suboptimal in highly attenuating conditions has remained less obvious. Those results still leave open a broad interpretation, namely, that increased spectral hardness or penetration is itself the main reason for improved detectability. The present study aims to narrow that claim, with the energy-resolved CNR(E) analysis showing that, for HA speck detection behind dense-breast tissue or silicone, the relevant mechanism is to selectively keep the high-energy photon interval removed by Rh filtration that still contributes to detectability, and not simply hardening the spectrum through increased tube voltages. The results apply specifically for 1 mm CdTe spectroscopic detectors, for which these higher-energy photons remain efficiently absorbed and can be analyzed spectrally. The same spectral performance cannot therefore be assumed for thinner or lower-Z PCDs, or for conventional charge-integrating detectors, whose signal weighting is fundamentally different [18].
The objective of the present work is to evaluate, for the specific imaging tasks considered here, whether preserving the high-energy tail of the spectrum improves the visibility of high-density structures, and to identify which of the tested spectra performs best within the present simulation framework. The study simulates only the radiation–matter interactions most relevant to image quality and complements the results with a qualitative experimental feasibility check in a challenging implant imaging scenario. Detailed Monte Carlo simulations of a standard mammographic setup were performed in the 10–50 keV range using clinically relevant spectra with aluminum (Al) and rhodium (Rh) filtration. Image quality was assessed using energy-resolved CNR calculations. Dense-breast phantoms were designed using a representative HA speck size within the 49.5 μm to 445.5 μm range reported for accreditation phantoms such as CIRS015 [37]. The aim is not to establish a general mammographic optimization result, but to test the present detector-task hypothesis under controlled simulation conditions.

2. Materials and Methods

2.1. Contrast-to-Noise Ratio (CNR): Definition and Analytical Model

The CNR was selected as the figure of merit for image quality evaluation, as it incorporates both HA contrast, governed by material attenuation properties, and the statistical noise associated with background photon detection. A reduction in CNR directly corresponds to reduced speck visibility. CNR is defined in terms of the normalized contrast ( C ), and the coefficient of variation ( C v ):
C N R = C C v   with   C = I B I S I B , C v = σ B I B
where I S and I B are the mean pixel intensities in the signal and background regions, respectively, and σ B is the standard deviation of background pixel intensities.
An analytical model for CNR was constructed as a theoretical reference for validating Monte Carlo results. The model expresses I S and σ B using the Beer–Lambert law [38], and assumes Poisson statistics for photon detection with an ideal noiseless detector. For an intensity, measured in incident number of photons I 0 over the sample entrance surface, the expected number of photons detectable behind the background region of the sample is as follows:
I B = I 0 · e μ B · t B
where t B is the breast tissue thickness and μ B is the corresponding linear attenuation coefficient. For a region containing an HA inclusion of thickness t H and attenuation coefficient μ H , the expected number of detectable photons in the signal region becomes the following:
I S = I 0 e μ B t B e ( μ H μ B ) t H
Assuming Poisson-limited noise, σ B = I B and C v = 1 I B . Combining these expressions leads to an energy-dependent CNR in the mono-energetic limit:
C N R ( E ) = I 0 ( E ) ( e μ B ( E ) t B 2 ) 1 e μ H ( E ) μ B ( E ) t H
Here, “spectroscopic” means that the CNR is expressed as a function of incident photon energy and that detection (and imaging) can be decomposed into energy bins that discretize the detected continuous polychromatic spectrum. Material attenuation coefficients were computed using homogeneous compound mixing based on mass-attenuation coefficients and densities from the NIST (Gaithersburg, MD, USA) X-ray attenuation database [39].
The model does not include readout electronics and assumes that quantum noise dominates. This is a reasonable approximation for photon-counting detectors operating with energy thresholds that substantially suppress electronic noise and at count rates below the pile-up regime [22]; it does not include threshold dispersion, charge sharing/K-fluorescence, charge-transport losses, or long-term CdTe polarization drift. In addition, flat-field corrections compensate for fixed-pattern and structural gain variations between pixels [40], so quantum (Poisson) noise becomes the dominant noise source. Accordingly, the mathematical model assumes σ 2 I for the background counts.
When applied to a pixelated detection system, the CNR defined in Equation (4) is proportional to the area ratio between the signal and background regions, as each region detects only a fraction of the total number of photons I 0 . This consideration also becomes important when estimating the minimum required number of simulated photons to achieve high-quality results.

2.2. Mammography X-Ray Spectra

Four theoretical spectra were calculated and used for the main simulation in this study (Figure 1). These input spectra cover relevant energy ranges for mammography imaging, and were recreated using the tool SpekPy, version 2.0.8 [41,42]. A tungsten (W) anode X-ray source was taken as the reference, and four tube voltage/filter combinations were then applied:
  • Standard mammography spectrum ( W 28 k R h ): 28 kVp tube voltage, 50 µm Rh filter;
  • Filter-modified spectrum ( W 28 k A l ): with 1 mm Al filter instead;
  • Voltage-modified spectrum ( W 50 k R h ): 50 kVp tube voltage instead;
  • Filter-Voltage-modified spectrum ( W 50 k A l ): 50 kVp tube voltage, 1 mm Al filter.
The spectral set was intentionally limited to four cases to isolate the effect of preserving the high-energy interval removed by Rh filtration, rather than to perform a global optimization over all mammography filter materials. All input spectra were normalized to their maximum value. Supplementary Monte Carlo simulations were performed to validate the SpekPy spectra, yielding virtually identical results. These simulations can be found in Appendix B.1.

2.3. Mammography Simulations and Analysis

All simulations were performed using GATE (Geant4 Application for Tomographic Emission) version 9.2, a Monte Carlo framework designed to simulate physical processes in medical imaging modalities, including X-ray imaging, PET, and CT [43,44]. GATE provides a command interpreter that enables straightforward system configuration through manual command entry or batch scripts that define the simulation geometry and execution sequence. The standard emstandard_opt4 electromagnetic physics list was used for all simulations. No charge transport or carrier collection was modeled, and all results are based solely on radiation–matter interactions and the corresponding energy depositions in a voxelated sensor matrix. Accordingly, the detector should be interpreted as a first-order interaction model. It reproduces photon transport and deposited energy in CdTe, but does not include charge drift and diffusion, carrier trapping, incomplete carrier collection, threshold dispersion, or pile-up. The simulated energy response is therefore narrower than in a real detector, and the reported CNR values should be interpreted as idealized with respect to detector electronics. For a detailed profile of the simulation stack, please refer to Appendix A.
Figure 2 shows the simulation setup schematic in three sections: the imaging geometry, the X-ray source definition, and the sample geometry and segmentation scheme.

2.3.1. Imaging Geometry

The simulated imaging configuration was replicated from the geometrical parameters used in the Hologic, Inc. (Marlborough, MA, USA) Selenia Dimensions AWS5000 digital mammography system [45]. The detector was modeled as a sensor voxel array with dimensions based on the physical characteristics of the SPCD Timepix3: a 1 mm thick CdTe semiconductor array with 55 µm pixel pitch. Each simulated voxel corresponds to one detector pixel. This sensor thickness is sufficient to achieve close to total photon absorption efficiency at mammographic X-ray energies [23]. This supports the CdTe-specific implementation, while not excluding other high-absorption sensor materials or thicknesses capable of achieving near-total attenuation. It also matches the sensor thickness used in previous experimental studies by this research group [20].
To optimize simulation time, a small 61 × 61 voxel matrix was defined, enough to confirm the feasibility of the proposed methods and their underlying physical principles. The detector was used for local CNR analysis, and no large-detector-size convergence study was performed, so absolute photon scatter is likely underestimated relative to a full-field breast geometry. The scatter contribution is nevertheless expected to be very low and was therefore not considered in this model. The center of the detector’s front surface was located at the position (0, 0, 70) cm, as shown in Figure 2a.

2.3.2. X-Ray Source

The geometry of the source is shown in Figure 2b. For ease of analysis, the X-ray source was modeled as an ideal divergent cone-beam point source with uniform intensity, allowing source-related effects such as penumbra to be decoupled from the detected intensity [46]. It was located at the world origin, with an opening angle θ = 0.39 and radius R = 2.382   mm chosen to cover the entire detector area while avoiding unnecessary computations for wide-angle photons that do not reach the detector. Using an ideal point source should have only a limited effect on the results, and is justified because the expected geometric unsharpness in typical mammography systems is on the order of 30 µm, below the typical detector pixel pitches of more than 70 µm. It was adopted to decouple spectral effects from source penumbra, which yields optimistic spatial resolution. Therefore, the present results do not represent full-system-level image sharpness performance.

2.3.3. Phantom

Breast density depends on the proportion of fibroglandular to adipose tissue, with denser breasts having a higher fibroglandular-to-adipose tissue ratio. The Breast Imaging Reporting and Data System (BI-RADS) [47,48] is a widely used framework that reduces variability in diagnostic imaging reports among radiologists. Breast composition is classified into four categories based on the visual assessment of fibroglandular content: A, almost entirely fatty; B, scattered areas of fibroglandular density; C, heterogeneously dense, which may obscure small masses; and D, extremely dense, while foregoing quantitative classification. Previous editions classified images by fibroglandular content. Categories A–D corresponded approximately to <25%, 25–50%, 50–75%, and 75–100%, respectively.
To simulate a dense breast, a block measuring 3.5 mm × 3.5 mm × 5 cm was defined, composed of 85% fibroglandular tissue and 15% adipose tissue (BI-RADS category D, 4th Edition). Its center was positioned at (0, 0, 63) cm, resulting in a geometric magnification of M = d γ d / d γ s = 1.111 . The 5 cm thickness was selected based on values reported in mammographic studies of dense breasts [49]. Length and width were chosen to be large enough for the projected image to completely cover the detector area.
The phantom contains 13 cubic HA specks embedded in the center of the tissue block to emulate malignant lesions within the breast. They were arranged laterally in a spaced checkerboard pattern (Figure 2c), with sufficient separation to observe each speck’s edge behavior. This number of specks provided a good compromise between sufficient statistical sampling of the attenuation signal and a reasonably short simulation time.
The speck side length t c was defined such that the projected shadows aligned with the voxels’ entrance surface, avoiding intensity drops caused by partial voxel coverage. t c = 346.5   μm was selected, which projected a shadow of 385 μm (7 voxels). This speck size is in the mid-range of those found in accreditation phantoms such as the CIRS011A [50] and CIRS015 [37], represents a small feature size, and should be readily detectable by mammography systems. Because a 2-pixel border was excluded from the CNR calculations, the measured signal was taken from the central region of the speck’s projection. The reported CNR is therefore expected to be only weakly sensitive to subpixel offsets, whereas smaller calcifications would be more affected by partial voxel coverage. The voxel-aligned placement defines a controlled reference case for the selected object size. In general, subpixel offsets and smaller or irregular inclusions would reduce the measured contrast.
Materials were defined using the same compound homogeneous material mixing method [39] employed in the analytical CNR(E) model. GATE handles the calculations internally, requiring the user only to specify the constituent elements and their weight fractions. Both adipose and glandular tissues were defined as indicated in Table 1.

2.3.4. Simulation Statistics

According to the Rose criterion, developed to evaluate the detectability of a signal in a noisy image, a CNR of 5 allows a signal to be detected with less than one part-per-million uncertainty [51,52]. To this end, Equation (4) can be inverted to calculate the number of detected photons required to achieve a target CNR at a specific energy for a defined set of materials and morphology:
I 0 ,   min ( E ) = C N R 2 e μ B ( E ) t B 1 e μ H ( E ) μ B ( E ) t H 2
Using Equation (5), the minimum photon intensity required to satisfy the Rose criterion can be estimated by taking 15 keV as a guide for a photon-starved low-energy bin, yielding I 0 ,   min ( 15   keV ) = 3.7 × 10 4 photons / voxel . For the defined detector (3721 voxels), the minimum number of photons required is 1.4 × 10 8 . Thus, 10 10 emitted photons were simulated per energy bin to ensure high-statistics results with negligible quantum noise.
Equal simulation statistics were applied to flat-field calibration images. Flat-field calibration is a well-known, easily implemented standard procedure used in X-ray imaging with digital detectors [40]. It is applied in the standard manner: M FFC = M RAW M FF . The operator denotes Hadamard (matrix element-wise) division. It can be mathematically proven that the final expression for Equation (4) remains invariant under a flat-field correction, provided the correction images have high statistics.
In physical detectors, flat-field images map variations in detector electronics between pixels and sensor defects, appearing as image inhomogeneities and visual artifacts. Although the simulated detector contains no defects, this standard procedure was retained to mirror standard X-ray imaging workflows, allowing methodological comparisons with experimental X-ray imaging pipelines.

2.3.5. Spectroscopic Analysis

The main polychromatic X-ray mammography simulation results were derived from a set of individual high-statistics mono-energetic simulations, performed at discrete energies E i , over the range E m i n = 10 keV to E m a x = 50 keV in 0.5 keV increments, resulting in a batch of 81 spectral flat-field-corrected images I M FFC ( E i ) . The polychromatic mammography spectra were then introduced post-simulation by weighting the intensity of each mono-energetic image with the corresponding spectrum’s mono-energetic intensity at each energy step, W kVp f i l t e r ( E i ) . The final polychromatic images were obtained by summing all weighted mono-energetic images to form the integrated-spectrum (IS) image I S kVp filter (Equation (6)), which corresponds to the type of images recorded by a simple PCD without spectroscopic information:
I S kVp filter = i = 1 81 W kVp filter ( E i ) · I M FFC ( E i )
The CNR values for the simulated IS images were calculated using Equation (1), with the signal and background voxel regions as depicted in Figure 2c, whereas the simulated CNR(E) curves were calculated for each energy bin from the corresponding individual spectral image, namely C N R E i = C ( E i ) / C v ( E i ) . The analytical results were obtained by applying Equation (4), using the mammography spectra, the material thickness, and NIST attenuation data. All curves were normalized to allow for a direct shape comparison, revealing energy ranges of importance. CNR values calculated from these IS images quantify image quality after spectral integration, whereas the spectroscopic CNR(E) curves are used only to identify the energy intervals that govern the behavior of the IS images.

2.4. Monte Carlo Validation Setup

In real-life detectors, energy deposited by a single incident photon may be shared between multiple pixels. During detection, partial energy absorption occurs at the primary interaction site, while secondary photons and electrons are generated through local scattering processes within the sensor material. If the range of these particles extends beyond the primary pixel pitch, a fraction of the deposited energy will be recorded in neighboring pixels and manifest in the image as multi-pixel clusters of false low-energy photons, thereby reducing contrast and blurring edges. This effect is called charge sharing. The extent of charge sharing depends on the incident photon energy, the proximity of the primary interaction site to pixel boundaries, and the pixel size.
Sensor X-ray fluorescence is a particular form of charge sharing in high-Z sensor materials such as CdTe, which preferentially absorb photons at K-edge energies (26.7 keV for Cd, 31.8 keV for Te). Subsequent fluorescence photon emission occurs across the 22–32 keV energy range, with α , β , and other emission lines from both materials appearing as relaxation by-products of sensor ionization [53], and with an expected fluorescence yield above 80% [54]. With a mean free path of around 100 µm in CdTe, these photons can travel beyond the primary pixel boundary in pixels smaller than this scale.
It is important to confirm that sensor fluorescence is properly simulated with the defined sensor geometry and to assess its impact on the results, in a manner analogous to charge sharing in real detectors, where this effect can degrade spatial resolution and spectral fidelity if not properly accounted for [23,55]. Although no charge-transport dynamics are simulated, lateral energy redistribution through secondary particle generation in the sensor is expected due to the high-aspect-ratio detector elements and the mean free path of Cd K-fluorescence photons in CdTe. Therefore, this study goes beyond an attenuation comparison, because the model geometrically includes CdTe spectral effects that shift the detector response away from the ideal case.
The first validation simulation consisted of a 5 mm thick Pb slab with a square, 5-voxel-wide pinhole aperture that collimated an ideal point source, thereby limiting primary photon lateral spread. The collimator was positioned a few mm away from the sensor surface. The sensor was voxelated and sized the same as for the main mono-energetic simulation set. Two simulations were performed using the unfiltered polychromatic W emission spectra at 28 kVp ( W 28 k none ) and 50 kVp ( W 50 k none ), obtained from supplementary Monte Carlo simulations (see Appendix B, Figure A1).
The second verification test was conducted using the same imaging setup as the main mono-energetic simulation set. It was performed to investigate intense halo artifacts appearing around specks and detector edges. Two separate mono-energetic simulations were executed at 26 keV and 28 keV, surrounding the Cd K-edge energy. Detected events were classified according to their creation vertex as either primary interactions within the initial voxel or secondary particles escaping to neighboring voxels.

2.5. Experimental Validation—Breast Implant X-Ray Imaging

To test the study hypothesis experimentally, the extreme case of breast-implant mammography imaging was selected. The recorded sample was a CIRS015 mammography accreditation phantom [37] with a 250 cm3 silicone implant partially covering the region of interest (Alumina specks, diameter 540 µm). The X-ray setup was configured for mammography imaging at the Uniandes High Energy Physics Laboratory [20]. The detector used was a Medipix3RX, supplied by the Medipix Collaboration at CERN (Geneva, Switzerland) [21], bump-bonded to a 1 mm thick CdTe sensor, sourced from Acrorad Co., Ltd. (Okinawa, Japan), and processed at the Freiburg Materials Research Center (Freiburg, Germany). This imaging geometry replicates that of the Hologic, Inc. Selenia Dimensions AWS5000 digital mammography system [45], consistent with the mammography simulations shown in Figure 2, but using a Hamamatsu Photonics K.K. (Hamamatsu, Japan) µFocus polychromatic X-ray source. This imaging system had been previously dose-calibrated for a W 28 k R h spectrum, to allow for direct comparisons with commercial mammography systems.
Three experimental images were acquired using two different spectra. The reference image was taken in the Hologic, Inc. system, using the W 28 k R h spectrum, delivering a standard Entrance Surface Dose (ESD) of 1.5 mGy to the sample. The second image was acquired in the laboratory at 5× the standard ESD in an attempt to increase speck visibility. The third image was also acquired in the laboratory using the W 28 k A l spectrum and the same exposure time required to achieve one standard ESD with the W 28 k R h spectrum. This allows for dose comparability across all images without the practical difficulties associated with system recalibration. Mean glandular dose was not used for comparability because, under the laboratory imaging conditions, the absence of a dedicated breast-dosimetry model would introduce an additional level of model dependence beyond the scope of the current study.
For the vendor system, the default CC view protocol was used, without implant displacement or access to vendor-proprietary post-processing, besides flat-field correction; for the lab system, only flat-field correction was applied. This experiment was designed as a qualitative feasibility check under implant attenuation conditions. It was not intended as a controlled cross-system comparison between the laboratory setup and the vendor system.

3. Results

3.1. Monte Carlo Validation Results

Figure 3 shows the results of the spectrum-level simulation tests. Figure 3a illustrates the interaction of CdTe with X-rays, including its K-edges and fluorescence peaks, and contains an inset showing the simulation setup. Figure 3b,c present the detected spectra for both setups, along with the corresponding intensity images. Fluorescence is confirmed by the detected spectrum from the W 50 k none simulation, which shows Cd and Te fluorescence peaks that are absent in the W 28 k none case. There is also a notable increase in low-energy hits, as energy is spatially redistributed.
The inset in Figure 3c shows detected events extending roughly seven voxels beyond the collimator pinhole. Their intensities are close to 10% of the primary beam intensity and decrease progressively toward zero. Notably, the image inset in Figure 3b also shows a few detections along the collimator edge, 2–3 voxels away, with less than 1% of the intensity measured at the pinhole. This indicates slight activation of Cd fluorescence, caused by the small fraction of photons in the W 28 k none spectrum above 26.7 keV.
Figure 4 shows the results of the two monochromatic simulation tests. The top and bottom rows correspond to the 26 keV and 28 keV monochromatic simulations, respectively. The left-column images were generated by counting all detected hits, whereas the right-column images were generated after excluding secondary interactions. The halo artifacts were completely absent in the 26 keV photon images but clearly appeared in the 28 keV photon images, around the specks, as well as on voxels near the sensor edge (Figure 4c). They disappear when secondary interactions are filtered out (Figure 4d).
Figure 4e shows the deposited energy histogram for the 28 keV photon simulation test. Interactions are separated according to their creation vertex location: blue for the primary voxel, red for secondary voxels. The histogram shows redistribution of the primary-photon energy among several discrete values: 2 keV, 5 keV and 28 keV in the primary voxel, and below 1 keV, 23 keV, and 26 keV in secondary voxels.
The energy histogram in Figure 4e shows the reason for the halo artifacts and confirms sensor fluorescence as the cause. In approximately 42% of the cases, the primary photon deposits all its energy in the primary voxel. In the remaining 58% of the cases, a fluorescence interaction occurs and the energy is spatially redistributed within the sensor through low-energy processes such as Auger-electron emission and the photoelectric effect. This is also visible as an increase in the recorded hits in Figure 4c compared with Figure 4a,b. Fluorescence photons also escape the sensor volume, as seen in the reduced hit counts in edge voxels in Figure 4c, closer to the number of primary hits in Figure 4d.

3.2. Mammography Simulation Results

Figure 5a shows two datasets: continuous analytical spectroscopic CNR curves and simulation-based spectroscopic CNR data points. The CNR(E) plots exhibit peaks at different energies. The W 28 k R h and W 28 k A l curves peak at 22 keV and display nearly identical shapes up to the Rh K-edge energy of 23.3 keV, beyond which the W 28 k R h curve drops sharply due to a lack of photons. The W 50 k R h curve peaks at the filter’s K-edge (23.2 keV), with a secondary peak at around 31 keV. The W 50 k A l curve peaks at 24 keV and maintains a continuous shape up to the maximum photon energy of the input spectrum.
Figure 5b shows the IS images resulting from the mammography simulations. The scale limits were defined by the minimum (zero) and maximum (one) intensity values across all images, thereby allowing direct visual comparison. The top-row images exhibit high contrast, but also show a fluctuating background. Conversely, the bottom-row images display lower contrast and a flatter background. Of the four tested spectra, image I S 28 k A l yields the highest CNR value, 23.11% higher than the standard I S 28 k R h . The CNR values obtained for images I S 50 k R h and I S 50 k A l were lower than that of the standard image by 1.60% and 4.92%, respectively. All IS image CNR values were calculated using the definition from Equation (1), and the image segmentation shown in Figure 2c (see Section 2.3.5). As a supplementary comparison, a W 28 k A g spectrum with a 53 µm thick silver (Ag) filter was applied to the same mono-energetic image set, yielding an IS image CNR 16.21% higher than the standard, falling between the I S 28 k R h and I S 28 k A l results (see Appendix B.2).

3.3. Breast Implant Imaging Results

Given the limited scope of the experimental imaging, the images in Figure 6 are interpreted only qualitatively. Figure 6a shows a flat gray region without discernible specks, even after adjustment of the brightness and contrast parameters. An equivalent result was initially obtained in the laboratory, with a subsequent 5× standard ESD image (Figure 6b) showing no improvement behind the implant. Figure 6c was acquired using the W 28 k A l spectrum, as described in Section 2.5. The specks were clearly visible behind the implant, without post-processing beyond the standard flat-field correction.

4. Discussion

4.1. Modified Mammography X-Ray Spectra

By analyzing the relative intensities of the recorded spectra, the energy ranges contributing most significantly to image formation can be identified. As a practical guideline, high-quality images generally require photon counts on the order of several thousand per energy bin. Energy intervals in which the CNR falls below approximately 30% of its maximum value can therefore be considered negligible, as only a minimal number of photons are detected in these regions. In all cases presented, photon energies below 14 keV contribute negligibly to image formation.
In the case of K-edge-filtered spectra, photons with energies above the filter’s K-edge can also be neglected, because any image structure that might be formed by detecting these photons is effectively lost. This has direct consequences for dense-breast and implant mammography imaging, as confirmed by both simulated and experimental results: increased CNR on simulated dense-breast material and increased microcalcification detectability through the implant. In both cases, the effect is driven by transmitted high-energy photons.
Previous studies already suggested that harder spectra and energy-resolved imaging can improve photon-counting mammography, but those results remain broad, implying that improved penetration alone is the main cause of the gain. The current results give a more precise interpretation: if beam hardening alone were responsible, the 50 kVp spectra would be expected to perform at least as well as the 28 kVp + 1 mm Al case, but they do not. This suggests that, for the present task, the optimum spectrum should at least include the high-energy tail removed by the Rh filter, which still carries useful HA speck contrast, but simply increasing photon energy up to 50 keV is actually detrimental. Therefore, there is an optimal upper photon energy, above which imaging performance deteriorates.
Because the detector is modeled as a 1 mm thick CdTe sensor, the preserved high-energy photons are still efficiently absorbed and contribute to the measured CNR. The Monte Carlo results also show that their recorded contribution is modified by Cd/Te fluorescence and inter-voxel energy redistribution, so the effect is not described by attenuation alone [23,56]. The result is therefore specific to the present task and detector, and should not be transferred directly to conventional charge-integrating mammography detectors without a separate analysis.
Supplementary simulations confirm that the attenuation of 2 cm of silicone closely matches that of 5 cm dense-breast tissue, supporting transferability of the result within the specific task–detector–material combinations (see Appendix B, Figure A1).

4.2. Mammography Simulations – Analysis

As seen in Figure 5a, CNR is a spectrally resolved quantity. The simulated CNR(E) results are in close agreement with the analytical model (Equation (4)), confirming that the simulation behaves as expected and serves as a reliable tool for spectral analysis in the high-statistics limit, despite the absence of charge-transport calculations.
The residual differences are dominated by photon scattering, K-shell fluorescence emission with partial escape of characteristic X-rays, and inter-voxel energy redistribution caused by secondary particle generation and reabsorption. Although scattering has a low yield in the energy range used in this study, fluorescence has been shown to play an important role, and is therefore the primary source of the discrepancy. The halo artifacts persist in all monochromatic images above the Cd K-edge energy. By carefully segmenting the sample and leaving edge margins, this effect was largely mitigated in the spectroscopic results; however, its impact on spatial resolution remains evident because object edges appear blurred.
The analytical CNR(E) model should be interpreted only as a photon-statistics estimate. A full detectability analysis would require MTF/NPS or model-observer methods that account for charge transport, focal-spot blur, and subpixel object placement, all of which reduce the capture of high spatial frequencies and would lower the visibility of smaller or subpixel microcalcifications.
This is especially relevant for the simulation CNR results. Given the projected HA speck shadow sizes (385 μm, 7 image pixels), and the edge-pixel exclusions from CNR calculations (Figure 2c), the reported CNR is less sensitive to edge blur than it would be for calcifications closer to the pixel size. These conditions support the relevance of the results for the chosen speck size, but they do not show that a full-system MTF treatment would preserve the relative grading of the tested spectra: the omitted detector- and source-level effects would reduce the absolute CNR values, but their effect on the relative grading between spectra was not quantified here. The results should therefore be interpreted as task-specific first-order spectral comparisons for the selected object size, geometry, and detector, as the study was not intended to probe the spatial-resolution limits of the system.
The simulation-based CNR(E) curves obtained represent an idealized first-order detector response limit. In practice, electron cloud broadening, threshold dispersion, charge-transport properties, bulk defects, fluorescence and residual undetected charge sharing below the threshold, all contribute to broadening, biasing and shifting the measured spectrum [55]. For Medipix3RX CdTe/CZT systems, reported energy resolution around 27 keV is on the order of a few keV FWHM rather than sub-keV [57]. Under a realistic detector response, the CNR(E) curves in Figure 5a would therefore be smoother, with reduced local structure, and the apparent optimal spectral intervals would become detector-dependent, particularly near K-edges and in the fluorescence-affected region. However, spectral broadening redistributes detected counts across neighboring bins; it does not restore photons removed from photon-starved energy bins. Thus, the broader distinction between the Al- and Rh-filtered cases is expected to remain qualitatively similar for the present task, whereas the exact local maxima and optimal spectral energy ranges should not be interpreted as detector-invariant.
CNR curves show an interplay between image noise, contrast, and material attenuation. For example, the standard spectrum produced images with higher contrast but also comparatively higher noise, whereas the W 50 k A l spectrum yielded lower-contrast I S 50 k A l images because HA becomes more transparent at higher energies, even when the background noise is reduced by the likewise greater transparency of the breast tissue. In other words, noise reduction does not always directly translate into improved object visibility.
Among the IS images obtained from Equation (6), the I S 28 k A l case had the highest CNR, whereas both I S 50 k performed worse. It can therefore be said that imaging performance is not monotonic with beam hardness, with the improvement coming from the retention of the transmitted photons removed by the Rh filter, combined with the high absorption efficiency of the 1 mm CdTe sensor. Conversely, simply adding higher and higher energy photons becomes detrimental above some energy value, which in this case lies below 50 keV. Finally, although the standard I S 28 k R h image exhibits the highest contrast, the background noise is also the highest. This is a direct consequence of removing the high-energy tail of the spectrum, between 23.2 keV and 28 keV.

4.3. Breast Implant X-Ray Imaging – Analysis

According to the estimated primary-incident air kerma obtained using SpekPy, the dose rates associated with the standard W 28 k R h spectrum and the modified W 28 k A l spectrum differ by approximately 5%. This is reflected in the corresponding spectra plots (Figure 1 and Figure A1), where the high-energy tail of the W 28 k A l represents a relatively small area under the curve, while the remainders of both spectra closely overlap.
Changing the filtration to 1 mm Al produces measurable image quality gains with only a small increase in estimated primary-incident air kerma: it changed the outcome from no visible specks to clear speck visibility. The W 28 k A l spectrum performed best among the four spectra studied for the chosen task, in agreement with the trend seen in the simulations. However, this result should be interpreted as a feasibility test only, as the experimental comparison was not designed as a controlled quantitative study of CNR optimization or overall system performance.
Figure 6b also confirms that the most important factor for visibility behind dense objects in the breast is the emitted spectrum, as even a fivefold increase in emitted fluence using the standard spectrum was insufficient to achieve speck visibility with a Medipix3RX 1 mm CdTe PCD.

5. Conclusions

A Monte Carlo simulation framework for detailed spectroscopic analysis in mammography imaging was successfully implemented by constructing a voxelated detector matrix that reproduces the most relevant physical phenomena. The virtual setup simulates a standard mammography system and allows the application of any energy spectrum of interest with a flexibility that is difficult to achieve in a real clinical setting.
This flexibility enabled investigation of the spectroscopic nature of the CNR and of the additional information it can provide for image quality enhancement. These curves serve an additional purpose: they can be combined with post-processing techniques in real systems with spectroscopic capabilities to further enhance object contrast and detectability without increasing radiation exposure. Techniques such as energy weighting and material decomposition have been demonstrated to be effective analytically, in simulations, and experimentally [22,25,56,58,59]. Yet, the spectroscopic CNR curves were used only as an analysis tool to explain the behavior of the IS images. Any post-processing methods remained outside the scope of this study.
The simulations also confirmed CdTe sensor fluorescence, with photons above 26.72 keV generating clusters of lower-energy hits. This explains the notable discrepancies between the analytical and simulation results. In real measurements, photon clustering techniques combined with SPCDs such as Timepix3 or HEXITEC MHz allow recovery of the undistorted energy spectrum [59], a basic post-processing step required when implementing spectroscopic techniques.
These results show that simulation-based spectral optimization is a promising approach for improving image quality in X-ray mammography of dense breasts. Among the four spectra studied, under the present simulation assumptions, the I S 28 k A l image had the highest CNR for HA speck detection in the task considered here, with a 23.11% increase relative to the standard I S 28 k R h image, while both I S 50 k images had worse performance. The improvement is therefore not due simply to generic beam hardening. It reflects retention of the transmitted energy interval removed by Rh filtration, together with the high absorption efficiency of a 1 mm thick CdTe sensor. The result also suggests an optimal upper photon energy above which imaging performance degrades. This conclusion is specific to the selected object size, geometry, and detector. The present study did not quantify whether full-system MTF degradation would preserve the same relative performance between spectra.
Simulation-informed spectral optimizations under laboratory conditions resulted in promising qualitative improvements in implant X-ray imaging. However, the experimental evidence in the present study is limited in scope and does not establish a general optimization result for mammography. These findings should be interpreted as a feasibility check only and not as a controlled quantitative study of CNR optimization or full-system performance.
Future work can extend this study in several directions. Subsequent studies can extend the present comparison by incorporating MTF/NPS or model-observer analysis, using a more realistic detector response, and systematically exploring tube potentials and filtration using different materials, including clinically common mammography filters and other candidate materials for dense-breast imaging. They should also consider sample size, shape, and subpixel-position sweeps in heterogeneous breast backgrounds.
The new capabilities of SPCDs open additional possibilities for X-ray imaging optimization by enabling the study of appropriate filter/spectrum combinations and post-processing algorithms that leverage the available spectroscopic information, potentially increasing image quality and/or reducing radiation exposure. Furthermore, SPCDs will be crucial for the continued development and refinement of medical X-ray imaging as spectroscopic information becomes increasingly available in clinical settings.

Author Contributions

Conceptualization, G.R., S.P., C.Á. and M.L.P.-L.; methodology, M.L.P.-L. and G.R.; software, G.R. and M.L.P.-L.; validation, G.R., M.L.P.-L., J.D.B., J.S.U.P. and S.C.; formal analysis, G.R., M.L.P.-L., S.C., J.D.B. and C.Á.; investigation, M.L.P.-L., S.C., G.R., S.P. and C.Á.; resources, C.Á., S.P. and M.F.; data curation, G.R., M.L.P.-L., S.C., C.Á., J.S.U.P., J.D.B., S.P. and M.K.S.; writing—original draft preparation, G.R.; writing—review and editing, G.R., M.L.P.-L., S.C., C.Á., S.P., J.S.U.P., J.D.B., M.K.S. and M.F.; visualization, G.R., S.C., M.L.P.-L. and J.S.U.P.; supervision, C.Á., S.P. and M.F.; project administration, C.Á., S.P. and G.R.; funding acquisition, C.Á., S.P. and G.R. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the following entities: BMBF, Germany [project number 01DN21005]; MINCIENCIAS, Colombia [project number 873-2020]; Universidad de los Andes, Faculty of Science, Colombia [project number INV-2025-213-3462].

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors on request.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
CNRContrast-to-Noise Ratio
RhRhodium
CdTeCadmium Telluride
AlAluminum
ESDEntrance Surface Dose
HACalcium Hydroxyapatite
MRIMagnetic Resonance Imaging
PCDPhoton-Counting Detector
SPCDSpectroscopic Photon-Counting Detector
NISTNational Institute of Standards and Technology (USA)
WTungsten
GATEGeant4 Application for Tomographic Emission
PETPositron Emission Tomography
CTComputed Tomography
BI-RADSBreast Imaging Reporting and Data System
ISIntegrated-Spectrum

Appendix A. Simulation Validation Table

The following list summarizes items to be considered when carrying out Monte Carlo studies. It is adapted from the AAPM Task Group 268 report on simulation studies (Table A1) [60]. Specifically, we verified the geometry, event tallies, and physics lists by reproducing transmission through reference slabs. We compared the results with the Beer–Lambert law using the NIST μ/ρ tables employed elsewhere in the study, and the simulated tube spectra with analytical spectra generated using SpekPy version 2.0.8. Finally, we confirmed Cd/Te K-edge activation and escape-range-consistent halos in CdTe above 26.7/31.8 keV, which are directly related to the results shown in Figure 3 and Figure 4. Items not in the table are described explicitly in the text.
Table A1. AAPM TG 268 recommended checklist for results validation in Monte Carlo studies.
Table A1. AAPM TG 268 recommended checklist for results validation in Monte Carlo studies.
Checklist ItemDetails
Monte Carlo SoftwareGate version 9.2. Based on GEANT4, version 11.0.3
Simulation Times
  • Between 5 min and 6 h per bin, depending on the number of photons arriving at the detector
  • Single full execution completes 162 mono-energetic simulations in 5 days (81 energy bins with corresponding flat fields)
Hardware
  • CPUs: Ryzen9 3950X/Ryzen9 5950X
  • RAM: 32GB/32GB
  • Custom parallel job management on Ubuntu 20.04
Physics and Transportemstandard_opt4 physics list
Travel Cut-offsGamma, Electrons:
  • World, 5 mm
  • Phantom, 0.25 mm
  • Sensor, 0.1 mm
Digitizer Energy Threshold3 keV
Number of primaries, all simulations1010 primary photons per energy bin. 1.62 × 1012 total photons generated

Appendix B. Supplementary Mammography Spectra and Filter Comparisons

Appendix B.1. Monte Carlo Mammography Spectra

An unfiltered W-anode X-ray spectrum was obtained from a detailed GEANT4 simulation of the Hamamatsu µFocus X-ray source model L10101, with an energy step of 0.1 keV. The simulation was configured to reproduce realistic properties of the X-ray tube, including the 150 µm Beryllium (Be) window, the 3D model of the housing, an approximation of the anode size, and the 41° emission angle. For reference, an unfiltered spectrum is shown, using the X-ray tube bias voltage at 28 kVp, labeled W 28 k none .
Subsequent photon transmission spectra were calculated using the Beer–Lambert law, combined with mass-attenuation coefficients obtained from NIST tables [39]. To achieve a finer energy step size than that provided by the tabulated attenuation data, the coefficients were interpolated between existing points while accounting for the presence of K-edges.
Figure A1. Transmission spectra for dense-breast and implant mammography imaging. The red curves show the simulated unfiltered spectrum W 28 k none . Subsequent curves were obtained from NIST tables. Dashed plots represent sample materials present in mammography: 85% glandular tissue breast, silicone, and the combination of both. (a) 50 µm Rh filtering. (b) 1 mm Al filtering.
Figure A1. Transmission spectra for dense-breast and implant mammography imaging. The red curves show the simulated unfiltered spectrum W 28 k none . Subsequent curves were obtained from NIST tables. Dashed plots represent sample materials present in mammography: 85% glandular tissue breast, silicone, and the combination of both. (a) 50 µm Rh filtering. (b) 1 mm Al filtering.
Applsci 16 03550 g0a1
The transmission through the sample materials shows that photons below 10 keV are negligible. The Rh-filtered spectra have the expected K-edge jump. Above 16 keV, 5 cm of dense-breast tissue has attenuation similar to that of 2 cm of silicone, which makes the results comparable between materials.

Appendix B.2. Supplementary 28 kVp Filter Comparison

To test whether the trend observed in the main four-spectrum comparison extends beyond the original W 28 k R h versus W 28 k A l pair, one supplementary was evaluated. A 53 µm thick Ag filter was selected (K-edge at 25.5 keV), generating a W 28 k A g spectrum that is close to the standard W 28 k R h case while testing a different K-edge filter. Ag filter thickness was chosen by visual inspection, to peak at the same energy as the other two spectra, close to 20 keV. The corresponding I S 28 k A g image and CNR were calculated from the same mono-energetic image set and methods described in Section 2.3.5, using Equations (1) and (6) and image segmentation shown in Figure 2c. Table A2 lists the CNR for all three W 28 k cases.
Table A2. Supplementary 28-kVp filter comparison obtained from the same mono-energetic simulation set used in the main study.
Table A2. Supplementary 28-kVp filter comparison obtained from the same mono-energetic simulation set used in the main study.
Unfiltered SpectrumW-anode, Tube Voltage 28 kVp
Filter50 µm Thick Rh53 µm Thick Ag1 mm Thick Al
IS image CNR102.0118.5125.5
% relative to W 28 k R h 0%16.21%23.11%

References

  1. Li, X.; Weaver, O.; Desouki, M.M.; Dabbs, D.; Shyum, S.; Carter, G.; Zhao, C. Microcalcification Is an Important Factor in the Management of Breast Intraductal Papillomas Diagnosed on Core Biopsy. Am. J. Clin. Pathol. 2012, 138, 789–795. [Google Scholar] [CrossRef] [PubMed]
  2. Wilkinson, L.; Thomas, V.; Sharma, N. Microcalcification on Mammography: Approaches to Interpretation and Biopsy. Br. J. Radiol. 2017, 90, 20160594. [Google Scholar] [CrossRef] [PubMed]
  3. Dance, D.R.; Christofides, S.; Maidment, A.D.A.; McLean, I.D.; Ng, K.H. Diagnostic Radiology Physics—A Handbook for Teachers and Students; International Atomic Energy Agency: Vienna, Austria, 2014; ISBN 978-92-0-131010-1. [Google Scholar]
  4. Bantick, G.; Taggart, I. Mammography and Breast Implants. Br. J. Plast. Surg. 1995, 48, 49–52. [Google Scholar] [CrossRef]
  5. Siu, A. Screening for Breast Cancer: U.S. Preventive Services Task Force Recommendation Statement. Ann. Intern. Med. 2016, 164, 279–296. [Google Scholar] [CrossRef] [PubMed]
  6. McCormack, V.A.; dos Santos Silva, I. Breast Density and Parenchymal Patterns as Markers of Breast Cancer Risk: A Meta-analysis. Cancer Epidemiol. Biomark. Prev. 2006, 15, 1159–1169. [Google Scholar] [CrossRef]
  7. Boyd, N.F.; Martin, L.J.; Bronskill, M.; Yaffe, M.J.; Duric, N.; Minkin, S. Breast Tissue Composition and Susceptibility to Breast Cancer. JNCI J. Natl. Cancer Inst. 2010, 102, 1224–1237. [Google Scholar] [CrossRef]
  8. Vachon, C.M.; van Gils, C.H.; Sellers, T.A.; Ghosh, K.; Pruthi, S.; Brandt, K.R.; Pankratz, V.S. Mammographic Density, Breast Cancer Risk and Risk Prediction. Breast Cancer Res. 2007, 9, 217. [Google Scholar] [CrossRef]
  9. Sprague, B.L.; Gangnon, R.E.; Burt, V.; Trentham-Dietz, A.; Hampton, J.M.; Wellman, R.D.; Kerlikowske, K.; Miglioretti, D.L. Prevalence of Mammographically Dense Breasts in the United States. JNCI J. Natl. Cancer Inst. 2014, 106, dju255. [Google Scholar] [CrossRef]
  10. Jackson, V.P.; Hendrick, R.E.; Feig, S.A.; Kopans, D.B. Imaging of the Radiographically Dense Breast. Radiology 1993, 188, 297–301. [Google Scholar] [CrossRef]
  11. Scheel, J.R.; Lee, J.M.; Sprague, B.L.; Lee, C.I.; Lehman, C.D. Screening Ultrasound as an Adjunct to Mammography in Women with Mammographically Dense Breasts. Am. J. Obstet. Gynecol. 2015, 212, 9–17. [Google Scholar] [CrossRef]
  12. Rebolj, M.; Assi, V.; Brentnall, A.; Parmar, D.; Duffy, S.W. Addition of Ultrasound to Mammography in the Case of Dense Breast Tissue: Systematic Review and Meta-analysis. Br. J. Cancer 2018, 118, 1559–1570. [Google Scholar] [CrossRef] [PubMed]
  13. Sieber, D.A.; Adams, J.; William, P. What’s Your Micromort? A Patient-Oriented Analysis of Breast Implant-Associated Anaplastic Large Cell Lymphoma (BIA-ALCL). Aesthetic Surg. J. 2017, 37, 887–891. [Google Scholar] [CrossRef] [PubMed]
  14. Kalaaji, A. ISAPS International Survey on Aesthetic/Cosmetic Procedures Performed in 2024. International Society of Aesthetic Plastic Surgery; West Lebanon, NH, USA, 2025; Available online: https://www.isaps.org/media/oogpzodr/isaps-global-survey_2024.pdf (accessed on 18 February 2026).
  15. Eklund, G.; Busby, R.; Miller, S.; Job, J. Improved Imaging of the Augmented Breast. Am. J. Roentgenol. 1988, 151, 469–473. [Google Scholar] [CrossRef] [PubMed]
  16. Smalley, S.M. Breast Implants and Breast Cancer Screening. J. Midwifery Women’s Health 2003, 48, 329–337. [Google Scholar] [CrossRef]
  17. Miglioretti, D.L.; Rutter, C.M.; Geller, B.M.; Cutter, G.; Barlow, W.E.; Rosenberg, R.; Weaver, D.L.; Taplin, S.H.; Ballard-Barbash, R.; Carney, P.A.; et al. Effect of Breast Augmentation on the Accuracy of Mammography and Cancer Characteristics. JAMA—J. Am. Med. Assoc. 2004, 291, 442–450. [Google Scholar] [CrossRef]
  18. Berger, N.; Marcon, M.; Saltybaeva, N.; Kalender, W.A.; Alkadhi, H.; Frauenfelder, T.; Boss, A. Dedicated Breast Computed Tomography with a Photon-Counting Detector: Initial Results of Clinical In Vivo Imaging. Investig. Radiol. 2019, 54, 409–418. [Google Scholar] [CrossRef]
  19. Poikela, T.; Plosila, J.; Westerlund, T.; Campbell, M.; De Gaspari, M.; Llopart, X.; Gromov, V.; Kluit, R.; van Beuzekom, M.; Zappon, F.; et al. Timepix3: A 65K Channel Hybrid Pixel Readout Chip with Simultaneous ToA/ToT and Sparse Readout. J. Instrum. 2014, 9, C05013. [Google Scholar] [CrossRef]
  20. Roque, G.; Avila, C.; Pérez-Lara, M.L.; Mendoza, L.; Procz, S. Study of Contrast-to-Noise Ratio Performance of a Medipix3RX CdTe Detector for Low Dose Mammography Imaging. Nucl. Instrum. Methods Phys. Res. Sect. A 2021, 992, 165000. [Google Scholar] [CrossRef]
  21. Ballabriga, R.; Alozy, J.; Blaj, G.; Campbell, M.; Fiederle, M.; Frojdh, E.; Heijne, E.H.M.; Llopart, X.; Pichotka, M.; Procz, S.; et al. The Medipix3RX: A High Resolution, Zero Dead-time Pixel Detector Readout Chip Allowing Spectroscopic Imaging. J. Instrum. 2014, 8, C02016. [Google Scholar] [CrossRef]
  22. Ballabriga, R.; Alozy, J.; Campbell, M.; Frojdh, E.; Heijne, E.H.M.; Koenig, T.; Llopart, X.; Marchal, J.; Pennicard, D.; Poikela, T.; et al. Review of Hybrid Pixel Detector Readout ASICs for Spectroscopic X-ray Imaging. J. Instrum. 2016, 11, P01007. [Google Scholar] [CrossRef]
  23. Procz, S.; Roque, G.; Avila, C.; Racedo, J.; Rueda, R.; Santos, I.; Fiederle, M. Investigation of CdTe, GaAs, Se and Si as Sensor Materials for Mammography. IEEE Trans. Med. Imaging 2020, 39, 3766–3778. [Google Scholar] [CrossRef]
  24. Hemdal, B.; Herrnsdorf, L.; Andersson, I.; Bengtsson, G.; Heddson, B.; Olsson, M. Average Glandular Dose in Routine Mammography Screening using a Sectra Microdose Mammography Unit. Radiat. Prot. Dosim. 2005, 114, 436–443. [Google Scholar] [CrossRef]
  25. Ghammraoui, B.; Glick, S.J. Investigating the Feasibility of Classifying Breast Microcalcifications Using Photon-counting Spectral Mammography: A Simulation Study. Med. Phys. 2017, 44, 2304–2311. [Google Scholar] [CrossRef] [PubMed]
  26. Cahn, R.N.; Cederström, B.; Danielsson, M.; Hall, A.; Lundqvist, M.; Nygren, D. Detective quantum efficiency dependence on X-ray energy weighting in mammography. Med. Phys. 1999, 26, 2680–2683. [Google Scholar] [CrossRef] [PubMed]
  27. Cederström, B.; Fredenberg, E.; Lundqvist, M.; Ericson, T.; Åslund, M. Observer-model optimization of X-ray system in photon-counting breast imaging. Nucl. Instrum. Methods Phys. Res. Sect. A 2011, 648, S54–S57. [Google Scholar] [CrossRef]
  28. Cederström, B.; Fredenberg, E. The influence of anatomical noise on optimal beam quality in mammography. Med. Phys. 2014, 41, 121903. [Google Scholar] [CrossRef]
  29. Berglund, J.; Johansson, H.; Lundqvist, M.; Cederström, B.; Fredenberg, E. Energy weighting improves dose efficiency in clinical practice: Implementation on a spectral photon-counting mammography system. J. Med. Imaging 2014, 1, 031003. [Google Scholar] [CrossRef]
  30. Kim, H.; Lee, M.; Kim, D.; Lee, D.; Kim, H.-J. Evaluation of photon-counting spectral mammography for classification of breast microcalcifications. Radiat. Phys. Chem. 2019, 162, 39–47. [Google Scholar] [CrossRef]
  31. Ghammraoui, B.; Gkoumas, S.; Glick, S.J. Characterization of a GaAs photon-counting detector for mammography. J. Med. Imaging 2021, 8, 033504. [Google Scholar] [CrossRef]
  32. Cely, E.; Roque, G.; Useche, S.; Ávila, C.; Procz, S. CNR enhancement in breast implant phantoms via spectral X-ray imaging. Phys. Med. Biol. 2025, 71, 015013. [Google Scholar] [CrossRef]
  33. Siemens Healthineers AG. Photon-Counting CT—A Quantum Leap in Computed Tomography. 2025. Available online: https://www.siemens-healthineers.com/computed-tomography/ct-technologies-and-innovations/photon-counting-ct (accessed on 18 February 2026).
  34. Canon Medical Systems Corporation. Photon Counting CT—A New Era in Computed Tomography. 2025. Available online: https://global.medical.canon/products/computed-tomography/pcct (accessed on 18 February 2026).
  35. Zhan, X.; Zhang, R.; Niu, X.; Hein, I.; Budden, B.; Wu, S.; Markov, N.; Clarke, C.; Qiang, Y.; Taguchi, H.; et al. Comprehensive Evaluations of a Prototype Full Field-of-view Photon Counting CT System through Phantom Studies. Phys. Med. Biol. 2023, 68, 175007. [Google Scholar] [CrossRef] [PubMed]
  36. Koninklijke Philips, N.V. See Beyond the Conventional to the Next Generation of Multi-Energy CT. Available online: https://www.philips.com.my/healthcare/sites/spectral-ct-learning-center (accessed on 18 February 2026).
  37. CIRS, Inc. Mammographic Accreditation Phantom—Model 015; Computerized Imaging Reference Systems—CIRS: Norfolk, VA, USA, 2013. [Google Scholar]
  38. Swinehart, D.F. The Beer-Lambert Law. J. Chem. Educ. 1962, 39, 333. [Google Scholar] [CrossRef]
  39. Hubbell, J.; Seltzer, S. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients from 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest; NIST: Gaithersburg, MD, USA, 2008. [CrossRef]
  40. Jakubek, J. Data Processing and Image Reconstruction Methods for Pixel Detectors. Nucl. Instrum. Methods Phys. Res. Sect. A 2007, 576, 223–234. [Google Scholar] [CrossRef]
  41. Poludniowski, G.; Omar, A.; Bujila, R.; Andreo, P. Technical Note: SpekPy v2.0—A Software Toolkit for Modeling X-ray Tube Spectra. Med. Phys. 2021, 48, 3630–3637. [Google Scholar] [CrossRef]
  42. Vorbau, R.; Poludniowski, G. Technical Note: SpekPy Web—Online X-ray Spectrum Calculations using an Interface to the SpekPy Toolkit. J. Appl. Clin. Med. Phys. 2024, 25, e14301. [Google Scholar] [CrossRef]
  43. Sarrut, D.; Arbor, N.; Baudier, T.; Borys, D.; Etxebeste, A.; Fuchs, H.; Gajewski, J.; Grevillot, L.; Jan, S.; Kagadis, G.C.; et al. The OpenGATE Ecosystem for Monte Carlo Simulation in Medical Physics. Phys. Med. Biol. 2022, 67, 184001. [Google Scholar] [CrossRef]
  44. Allison, J.; Amako, K.; Apostolakis, J.; Arce, P.; Asai, M.; Aso, T.; Bagli, E.; Bagulya, A.; Banerjee, S.; Barrand, G.; et al. Recent Developments in Geant4. Nucl. Instrum. Methods Phys. Res. Sect. A 2016, 835, 186–225. [Google Scholar] [CrossRef]
  45. Hologic, Inc. Selenia Dimensions with Acquisition Workstation 5000; Hologic, Inc.: Marlborough, MA, USA, 2011. [Google Scholar]
  46. Harms, A.A.; Zeilinger, A. A New Formulation of Total Unsharpness in Radiography. Phys. Med. Biol. 1977, 22, 70–80. [Google Scholar] [CrossRef]
  47. Spak, D.A.; Plaxco, J.S.; Santiago, L.; Dryden, M.J.; Dogan, B.E. BI-RADS® fifth edition: A summary of changes. Diagn. Interv. Imaging 2017, 98, 179–190. [Google Scholar] [CrossRef]
  48. Destounis, S.V.; Friedewald, S.M.; Grimm, L.J.; Poplack, S.P.; Sung, J.S. Mammography. In ACR BI-RADS® v2025 Manual; Newell, M.S., Destounis, S.V., Eds.; American College of Radiology: Reston, VA, USA, 2025. [Google Scholar]
  49. Oliveira, A.; Meguerian, B.A.; Mesquita, C.T. Influence of Breast Characteristics in Myocardial Scintigraphy through the Monte Carlo Method. Arq. Bras. Cardiol. 2011, 96, 8–12. [Google Scholar] [CrossRef]
  50. CIRS, Inc. Tissue-Equivalent Phantom for Mammography—Model 011A; Computerized Imaging Reference Systems—CIRS: Norfolk, VA, USA, 2013. [Google Scholar]
  51. Rose, A. Vision—Human and Electronic, 1st ed.; Springer: New York, NY, USA, 1973; ISBN 978-1-4684-2039-5. [Google Scholar] [CrossRef]
  52. Burgess, A.E. The Rose Model, Revisited. J. Opt. Soc. Am. A 1999, 16, 633–646. [Google Scholar] [CrossRef] [PubMed]
  53. Bearden, J.A. X-Ray Wavelengths. Rev. Mod. Phys. 1967, 39, 78–124. [Google Scholar] [CrossRef]
  54. Krause, M.O. Atomic Radiative and Radiationless Yields for K and L Shells. J. Phys. Chem. Ref. Data 1979, 8, 307–327. [Google Scholar] [CrossRef]
  55. Koenig, T.; Zuber, M.; Hamann, E.; Cecilia, A.; Ballabriga, R.; Campbell, M.; Ruat, M.; Tlustos, L.; Fauler, A.; Fiederle, M.; et al. How spectroscopic X-ray imaging benefits from inter-pixel communication. Phys. Med. Biol. 2014, 59, 6195–6213. [Google Scholar] [CrossRef]
  56. Anton, G.; Durst, J.; Firsching, M.; Giersch, J.; Korn, A.; Mitschke, M.; Niederlöhner, D.; Pfeiffer, F. Imaging Theory for X-ray Pixel Detectors. Nucl. Instrum. Methods Phys. Res. Sect. A 2006, 563, 116–123. [Google Scholar] [CrossRef]
  57. Marsh, J.F.; Jorgensen, S.M.; Rundle, D.S.; Vercnocke, A.J.; Leng, S.; Butler, P.H.; McCollough, C.H.; Ritman, E.L. Evaluation of a photon counting Medipix3RX cadmium zinc telluride spectral X-ray detector. J. Med. Imaging 2018, 5, 043503. [Google Scholar] [CrossRef]
  58. Giersch, J.; Niederlöhner, D.; Anton, G. The Influence of Energy Weighting on X-ray Imaging Quality. Nucl. Instrum. Methods Phys. Res. Sect. A 2004, 531, 68–74. [Google Scholar] [CrossRef]
  59. Roque, G.; Schütz, M.K.; Fey, J.; Useche, J.S.; Fiederle, M.; Procz, S. Sub-pixel Energy-weighting Techniques for Metallic Contaminant Highlighting in a Pharmaceutical Hard Capsule using a Timepix3 CdZnTe Hybrid Pixel Detector. J. Instrum. 2022, 17, P10030. [Google Scholar] [CrossRef]
  60. Sechopoulos, I.; Rogers, D.W.O.; Bazalova-Carter, M.; Bolch, W.E.; Heath, E.C.; McNitt-Gray, M.F.; Sempau, J.; Williamson, J.F. RECORDS: Improved Reporting of montE CarlO RaDiation transport Studies: Report of the AAPM Research Committee Task Group 268. Med. Phys. 2017, 45, e1–e5. [Google Scholar] [CrossRef]
Figure 1. Numerically generated normalized input spectra for four distinct cases, used in combination with the main mammography Monte Carlo simulation, to calculate image quality for HA speck detection. Spectra generated with the tool SpekPy [41,42].
Figure 1. Numerically generated normalized input spectra for four distinct cases, used in combination with the main mammography Monte Carlo simulation, to calculate image quality for HA speck detection. Spectra generated with the tool SpekPy [41,42].
Applsci 16 03550 g001
Figure 2. Simulated imaging setup schematic. (a) Imaging geometry, replicating the Hologic, Inc. Selenia Dimensions AWS5000 mammography system [45]. (b) Divergent point source geometry. (c) Phantom with cubic specks (gray), showing segmentation for calculations (blue: signal/red: background). A 2-voxel margin along all borders was excluded from the calculations.
Figure 2. Simulated imaging setup schematic. (a) Imaging geometry, replicating the Hologic, Inc. Selenia Dimensions AWS5000 mammography system [45]. (b) Divergent point source geometry. (c) Phantom with cubic specks (gray), showing segmentation for calculations (blue: signal/red: background). A 2-voxel margin along all borders was excluded from the calculations.
Applsci 16 03550 g002
Figure 3. Spectrum-level simulation tests. (a) CdTe material mass-attenuation coefficient and XRF peaks, with inset showing the simulation geometry. Simulated detector response for two polychromatic spectra: (b) W 28 k none ; (c) W 50 k none .
Figure 3. Spectrum-level simulation tests. (a) CdTe material mass-attenuation coefficient and XRF peaks, with inset showing the simulation geometry. Simulated detector response for two polychromatic spectra: (b) W 28 k none ; (c) W 50 k none .
Applsci 16 03550 g003
Figure 4. Monochromatic simulation tests. (a) Image generated by 26 keV photons. (b) Image generated by removing secondary interactions in (a). (c) Image generated by 28 keV photons. (d) Image generated by removing secondary interactions in (c). (e) Deposited energy histogram for (c,d).
Figure 4. Monochromatic simulation tests. (a) Image generated by 26 keV photons. (b) Image generated by removing secondary interactions in (a). (c) Image generated by 28 keV photons. (d) Image generated by removing secondary interactions in (c). (e) Deposited energy histogram for (c,d).
Applsci 16 03550 g004
Figure 5. CNR(E) and imaging results from the main mono-energetic simulation set. (a) CNR(E) analytical curves and simulated data points. (b) IS images for the four spectra, with corresponding CNR values.
Figure 5. CNR(E) and imaging results from the main mono-energetic simulation set. (a) CNR(E) analytical curves and simulated data points. (b) IS images for the four spectra, with corresponding CNR values.
Applsci 16 03550 g005
Figure 6. CIRS015 phantom + 250 cm3 implant images. (a) Selenia Dimensions system, W 28 k R h spectrum, 1× standard ESD. (b) Uniandes X-ray imaging setup, W 28 k R h spectrum, 5× standard ESD (specks circled in red). (c) Uniandes X-ray imaging setup, W 28 k A l spectrum, 1/5 the exposure time of image (b) (specks circled in black).
Figure 6. CIRS015 phantom + 250 cm3 implant images. (a) Selenia Dimensions system, W 28 k R h spectrum, 1× standard ESD. (b) Uniandes X-ray imaging setup, W 28 k R h spectrum, 5× standard ESD (specks circled in red). (c) Uniandes X-ray imaging setup, W 28 k A l spectrum, 1/5 the exposure time of image (b) (specks circled in black).
Applsci 16 03550 g006
Table 1. Material definitions to simulate breast with an 85%/15% glandular-to-adipose ratio.
Table 1. Material definitions to simulate breast with an 85%/15% glandular-to-adipose ratio.
MaterialAdipose TissueGlandular Tissue
Density [ g / c m 3 ]0.921.02
ElementWeight Fraction
Hydrogen0.1200.106
Carbon0.6400.332
Nitrogen0.0080.030
Oxygen0.2290.527
Sodium00.001
Sulfur00.002
Chlorine00.001
Phosphor0.0020.001
Calcium0.0010
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Roque, G.; Pérez-Lara, M.L.; Cely, S.; Useche Parra, J.S.; Bermúdez, J.D.; Schütz, M.K.; Fiederle, M.; Ávila, C.; Procz, S. Energy-Resolved CNR Performance in Dense-Breast and Implant X-Ray Mammography Using a CdTe Photon-Counting Detector: A Monte Carlo Study. Appl. Sci. 2026, 16, 3550. https://doi.org/10.3390/app16073550

AMA Style

Roque G, Pérez-Lara ML, Cely S, Useche Parra JS, Bermúdez JD, Schütz MK, Fiederle M, Ávila C, Procz S. Energy-Resolved CNR Performance in Dense-Breast and Implant X-Ray Mammography Using a CdTe Photon-Counting Detector: A Monte Carlo Study. Applied Sciences. 2026; 16(7):3550. https://doi.org/10.3390/app16073550

Chicago/Turabian Style

Roque, Gerardo, Maria Laura Pérez-Lara, Steven Cely, Juan Sebastián Useche Parra, Jesús David Bermúdez, Michael K. Schütz, Michael Fiederle, Carlos Ávila, and Simon Procz. 2026. "Energy-Resolved CNR Performance in Dense-Breast and Implant X-Ray Mammography Using a CdTe Photon-Counting Detector: A Monte Carlo Study" Applied Sciences 16, no. 7: 3550. https://doi.org/10.3390/app16073550

APA Style

Roque, G., Pérez-Lara, M. L., Cely, S., Useche Parra, J. S., Bermúdez, J. D., Schütz, M. K., Fiederle, M., Ávila, C., & Procz, S. (2026). Energy-Resolved CNR Performance in Dense-Breast and Implant X-Ray Mammography Using a CdTe Photon-Counting Detector: A Monte Carlo Study. Applied Sciences, 16(7), 3550. https://doi.org/10.3390/app16073550

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop