Next Article in Journal
Effect of an Individualised Training Programme on Hamstrings and Change Direction Based on Tensiomyography in Football Players
Previous Article in Journal
Research and Application of Combined Reactive Power Compensation Device Based on SVG+SC in Wind Power Gathering Area
Previous Article in Special Issue
Numerical and Experimental Investigations of Humeral Greater Tuberosity Fractures with Plate Fixation under Different Shoulder Rehabilitation Activities
 
 
Article
Peer-Review Record

Development and Finite Element Analysis of a Novel Bent Bone Plate

Appl. Sci. 2022, 12(21), 10900; https://doi.org/10.3390/app122110900
by Joyceline Kurniawan, Shen-Yung Lin * and Wen-Teng Wang
Reviewer 1:
Reviewer 2: Anonymous
Reviewer 3:
Appl. Sci. 2022, 12(21), 10900; https://doi.org/10.3390/app122110900
Submission received: 5 September 2022 / Revised: 14 October 2022 / Accepted: 19 October 2022 / Published: 27 October 2022
(This article belongs to the Special Issue Recent Advance in Finite Elements and Biomechanics)

Round 1

Reviewer 1 Report

Title: Development and Finite Element Analysis for a Novel Bent Bone Plate

Author: Joyceline Kurniawan, Shen-Yung Lin and Wen-Teng Wang

After reviewing the manuscript, there are some mistakes. The detail comments are as follows:

1. In Table 2, the unit of Yield Strength and Tensile Strength should be MPa instead of GPa, please revise. In addition, are the references [3] and [16] or the references [17] and [18] used for the mechanical Properties of Titanium Alloy 6Al-4V presented in Table 2? Please unify.

2. In Table 3 and Figure 27, please give the calculation formula of the average displacement.

3. In reference [17], plte should be plate.

4. The quality of the figures are poor, and need to be redrawn.

Author Response

Please see attachment

Author Response File: Author Response.docx

Reviewer 2 Report

This paper shows high sigificance of content and high quality of presentation. 

I think

I think this paper can be published.

Author Response

No comments should be revised.

Reviewer 3 Report

1. In this paper, the authors develop a new bone plate implant design with middle bending. The explanation of traditional flat plate features is not enough, and more previous research on flat plate or bending plate should be given in the introduction section.

2. The authors have tested plates using a FEA software, so the details of COMSOL Multiphysics need to be given, such as the version number, company name and so on.

3. In this manuscript, the bending angle is used in different plates, and what is the methodology based on and are there any relevant studies?

4. In the “result summary” part, it said that “It would be essential for a bone plate to not be easily broken, especially in the fixation spot, in order to ensure a successful recovery.” Please clarify the more reliable basis, like other previous studies.

5. The manuscript may need to add a discussion part, and this part will make the results more useful for other researchers.

Author Response

Please see the attachment

Author Response File: Author Response.docx

Round 2

Reviewer 1 Report

The manuscript has been revised according to reviewer's comments, and can be accepted in this journal.

Back to TopTop