An Ingestible Electronics for Continuous and Real-Time Intraabdominal Pressure Monitoring

1
Department of Trauma and Emergency Surgery, Linkou Chang Gung Memorial Hospital, Chang Gung University, Taoyuan 333423, Taiwan
2
Department of Rehabilitation and Physical Medicine, Linkou Chang Gung Memorial Hospital, Chang Gung University, Taoyuan 333423, Taiwan
3
Department of Chemical Engineering, National United University, Miaoli 360, Taiwan
4
Department of Urology, Chiayi Chang Gung Memorial Hospital, Chang Gung University, Chiayi 613016, Taiwan
*
Author to whom correspondence should be addressed.
J. Pers. Med. 2021, 11(1), 12; https://doi.org/10.3390/jpm11010012
Submission received: 29 November 2020 / Revised: 20 December 2020 / Accepted: 22 December 2020 / Published: 24 December 2020
(This article belongs to the Special Issue The Interface between Human Physiology and Medical Device Development)

Abstract

Abdominal compartment syndrome can be treated through decompressive surgery if intraabdominal hypertension (IAH) can be detected in time. Treatment delays due to manual, conventional intravesical pressure (IVP) monitoring using a Foley catheter have been reported. In this work, we present an innovative gastrointestinal intraluminal pressure (GIP) measurement-based method to monitor and improve pressure-guided relief of intraabdominal pressure (IAP). A novel algorithm for detecting IAH in the gastrointestinal tract of a live porcine model is reported. A wireless pressure-sensing capsule (10 × 13 mm) was developed for absolute measurement. The IAP was estimated during artificial pneumoperitoneum. The pressure waveform-based measurements indicated that the wireless pressure sensor could be used to predict IAP. To enhance GIP monitoring for predicting IAH, the proposed continuous ingestible wireless electronics-based pressure waveform measurement device can be used as a complement to existing modalities. The use of the proposed pressure measurement and communication technology can help provide valuable data for digital health platforms.
Keywords: intraabdominal hypertension; abdominal compartment syndrome; smart capsule; wearable device intraabdominal hypertension; abdominal compartment syndrome; smart capsule; wearable device
Graphical Abstract

Share and Cite

MDPI and ACS Style

Liao, C.-H.; Cheng, C.-T.; Chen, C.-C.; Jow, U.-M.; Chen, C.-H.; Lai, Y.-L.; Chen, Y.-C.; Ho, D.-R. An Ingestible Electronics for Continuous and Real-Time Intraabdominal Pressure Monitoring. J. Pers. Med. 2021, 11, 12. https://doi.org/10.3390/jpm11010012

AMA Style

Liao C-H, Cheng C-T, Chen C-C, Jow U-M, Chen C-H, Lai Y-L, Chen Y-C, Ho D-R. An Ingestible Electronics for Continuous and Real-Time Intraabdominal Pressure Monitoring. Journal of Personalized Medicine. 2021; 11(1):12. https://doi.org/10.3390/jpm11010012

Chicago/Turabian Style

Liao, Chien-Hung, Chi-Tung Cheng, Chih-Chi Chen, Uei-Ming Jow, Chun-Hung Chen, Yen-Liang Lai, Ya-Chuan Chen, and Dong-Ru Ho. 2021. "An Ingestible Electronics for Continuous and Real-Time Intraabdominal Pressure Monitoring" Journal of Personalized Medicine 11, no. 1: 12. https://doi.org/10.3390/jpm11010012

APA Style

Liao, C.-H., Cheng, C.-T., Chen, C.-C., Jow, U.-M., Chen, C.-H., Lai, Y.-L., Chen, Y.-C., & Ho, D.-R. (2021). An Ingestible Electronics for Continuous and Real-Time Intraabdominal Pressure Monitoring. Journal of Personalized Medicine, 11(1), 12. https://doi.org/10.3390/jpm11010012

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.