Chemical Separation of Uranium and Precise Measurement of 234U/238U and 235U/238U Ratios in Soil Samples Using Multi Collector Inductively Coupled Plasma Mass Spectrometry

1
Environmental Radionuclide Research Group, National Institutes for Quantum and Radiological Science and Technology (QST), 4-9-1 Anagawa, Inage-Ku, Chiba 263-8555, Japan
2
Department of Radiological Sciences, Tokyo Metropolitan University, 7-2-10 Higashiogu, Arakawa-ku, Tokyo 116-8551, Japan
3
Energy Technology Unit, Japan Nus Co. (JANUS), 7-5-25 Nishi-Shinjuku, Shinjuku-ku, Tokyo 160-0023, Japan
*
Author to whom correspondence should be addressed.
Academic Editor: Michael Bolshov
Molecules 2020, 25(9), 2138; https://doi.org/10.3390/molecules25092138
Received: 3 March 2020 / Revised: 17 April 2020 / Accepted: 29 April 2020 / Published: 3 May 2020
A new chemical separation has been developed to isolate uranium (U) using two UTEVA columns to minimize iron and thorium interferences from high background area soil samples containing minerals like monazites and ilmenite. The separation method was successfully verified in some certified reference materials (CRMs), for example, JSd-2, JLk-1, JB-1 and JB-3. The same method was applied for purification of U in Fukushima soil samples affected by the Fukushima dai-ichi nuclear power station (FDNPS) accident. Precise and accurate measurement of 234U/238U and 235U/238U isotope ratios in chemically separated U were carried out using a multi-collector inductively coupled plasma mass spectrometer (MC-ICP-MS). In this mass spectrometric method, an array of two Faraday cups (1011 Ω, 1012 Ω resistor) and a Daly detector were simultaneously employed. The precision of U isotope ratios in an in-house standard was evaluated by replicate measurement. Relative standard deviation (RSD) of 234U/238U and 235U/238U were found to be 0.094% (2σ) and 0.590% (2σ), respectively. This method has been validated using a standard reference material SRM 4350B, sediment sample. The replicate measurements of 234U/238U in SRM shows 0.7% (RSD). This developed method is suitable for separation of U and its isotope ratio measurement in environmental samples. View Full-Text
Keywords: MC-ICP-MS; 234U/238U; 235U/238U; UTEVA resin; soil MC-ICP-MS; 234U/238U; 235U/238U; UTEVA resin; soil
Show Figures

Figure 1

MDPI and ACS Style

Veerasamy, N.; Takamasa, A.; Murugan, R.; Kasar, S.; Aono, T.; Inoue, K.; Fukushi, M.; Sahoo, S.K. Chemical Separation of Uranium and Precise Measurement of 234U/238U and 235U/238U Ratios in Soil Samples Using Multi Collector Inductively Coupled Plasma Mass Spectrometry. Molecules 2020, 25, 2138.

Show more citation formats Show less citations formats
Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Article Access Map by Country/Region

1
Search more from Scilit
 
Search
Back to TopTop