1. Introduction
Optical fibre-based beam loss monitors (oBLMs) consist of one or more optical fibres running alongside a beamline, where at least one end of each fibre is monitored by a photodetector.
Beam losses are detected through monitoring secondary particle showers from beam loss events, produced when high-energy beam particles strike an obstacle—such as limiting apertures within beam pipes, the jaws of a beam collimator, or an inserted scintillation screen. Relativistic particles in the shower that pass through the optical fibre produce a pulse of Cherenkov radiation inside it ([
1],
Figure 1), with the Cherenkov photons emitted within the fibre acceptance angle being captured in the fibre. The captured radiation propagates along the fibre and exits at the ends, where the photodetectors convert it into a voltage signal. The beam loss location can then be reconstructed from the arrival times of the Cherenkov pulse at each end of the fibre [
2], or through comparison with a master clock if using a single photodetector [
3].
oBLMs were first developed as a low-cost beam loss monitoring solution, with the capability for continuous accelerator coverage and approximately 1 m precision in the reconstruction of beam loss locations [
4,
5]. A step-index silica fibre, with inner core diameter above 200 μm, is one type of optical fibre that is effective for this purpose.
The work in this paper aims to apply oBLMs to energy recovery LINACs (ERLs), a type of accelerator on which these detectors have not previously been implemented.
ERLs achieve higher energy efficiencies than conventional linear accelerators (LINACs) by recycling the kinetic energy of the beam [
6,
7,
8]. This is performed using a recirculating configuration where the beam is returned to the LINAC on the decelerating RF phase, depositing the kinetic energy of the beam back into the RF field. The recovered energy is then used to accelerate the following beam.
The beam energy of an ERL increases and decreases in discrete steps, set by the energy gain or decrease applied by a single pass of the beam through the LINAC. The simplest ERLs are single-turn designs, which use one accelerating pass and one decelerating pass as shown in
Figure 2. Multi-turn designs can be used to attain higher beam energies via additional passes through the LINAC, and often include separate sets of return arcs for each beam energy.
However, the recirculating configuration necessary for ERLs increases their susceptibility to destabilising collective effects such beam break-up [
9] and the microbunching instability [
10,
11], which leaves them at risk of high levels of beam loss. The continuous machine coverage of oBLMs therefore makes them a good prospect for diagnosing beam losses and managing collective effects across the entirety of an ERL.
An important feature for successful application of the oBLM to ERLs is beam loss energy estimation.
Many ERLs use multi-turn designs [
12,
13,
14], which feature simultaneously circulating multiple-energy bunched beams interleaved to form a bunch train. Beam losses in multi-turn ERLs may occur from all bunches at once, or just from bunches on specific turn numbers (the number of completed circuits of the main loop). To enable tracking of beam losses of ERL bunches throughout their full journey in the machine, it is necessary to be able to determine both the locations of beam losses and the turn numbers of the bunches that caused them.
Due to the discrete energy steps between turns, each beam energy in an ERL corresponds to specific turn numbers. Therefore, one option to determine the turn number of a bunch that produced a beam loss is to estimate its energy using the oBLM signal intensity—the total number of photons in the output Cherenkov pulse.
The total Cherenkov photon yield
for a single particle of charge
q, which is travelling through a medium of refractive index
per unit path length
in the medium over the spectral range defined by wavelengths
and
, is given in Equation (
1) [
15], derived from the work of Frank and Tamm [
16]:
where
is the fine structure constant and
is the particle’s speed factor
.
For a highly relativistic particle
, Equation (
1) becomes constant and the photon yield depends only on the path length. Assuming that
particles are the main contributors and that refractive index
is approximately constant over the wavelength interval, the Cherenkov photon yield for a shower is therefore expected to be directly proportional to the total integrated path length of the shower—the total distance travelled by all shower particles in the fibre.
This, along with the fact that the total track length of a shower is also directly proportional to the energy of the instigating particle [
17], determines that the Cherenkov photon yield for a particle shower should vary linearly with the instigating particle’s energy. This has seen successful usage in optical fibre Cherenkov-based calorimetry [
18] and suggests that the bunch energy can, in principle, be estimated from the intensity of the Cherenkov signal in the oBLM.
For the case of an ERL, the energy resolution only needs to be sufficient to distinguish energies from adjacent turns. A resolution equal to or better than half the energy change between turns is likely to be acceptable.
Some ERL beam energies correspond to two turn numbers—one with an accelerating beam and one with a decelerating beam. To aid in distinguishing these, this method could be augmented using a comparison of the photon detection time with the nominal bunch positions expected from the RF clock and the filling pattern [
19], which allocates the bunches from each turn into specific RF timing buckets.
However, for beam energy estimation with the oBLM to be feasible, the effect of the beam energy must be separated from other influences on the signal—such as the total number of particles that collide with the obstacle in a beam loss event (the loss intensity). This has been observed to also exhibit a linear relationship with the oBLM signal intensity [
1,
2,
20,
21], which presents a confounding factor to this approach.
The effect of photon attenuation in the fibre on the signal intensity is another necessary consideration. Since the attenuation depends on the photon travel distances in the fibre, changes in the beam loss position will change the expected attenuation level. However, when investigating the energies of beam loss signals originating from the same location, the photon intensity reduction from attenuation is expected to be constant and the effect may be neglected.
Furthermore, fibre Cherenkov calorimetry methods involve direct beam impact in the fibre, using arrays of fibres to increase the fraction of the particle shower captured within the detector. This is not the case for an oBLM, which captures a slice of the shower over the small solid angle around the accelerator covered by the fibre. Hence, it is important to also investigate how the positioning of the fibre relative to the beam loss location and ensuing shower may affect the relationship between the oBLM signal and beam energy.
The Cherenkov signal observed in an oBLM depends on the properties of the charge shower as it passes through the fibre; such as its energy, intensity, angular distribution, and temporal structure. These are difficult to estimate analytically: the initial properties of the shower are expected to depend on a combination of beam loss properties, such as beam energy [
18], loss intensity, and the cross-section for the beam to interact with the scattering target. The ensuing shower evolution, such as attenuation by accelerator components and deflection in magnetic fields, depends on the specific component geometry, materials, and magnetic fields present around the beam loss location [
22].
In light of this, it is necessary to investigate the expected properties of the oBLM signal by simulating the radiation environment with Monte Carlo tools in order to determine the feasibility of this method.
The simulations and experiments presented here investigate the relationship between the beam energy and the oBLM signal, comparing the effects of the energy and loss intensity on the oBLM, and considering potential options for distinguishing them. The effect of fibre placement and obstacle thickness on the relationship between the oBLM signal and the beam energy is also investigated as a consideration for this method.
2. Geant4 Simulations
Simulation studies of the oBLM system were conducted using the Geant4 simulation programme [
23,
24,
25]: a Monte Carlo particle-tracking code that uses random sampling to simulate the physics of particles passing through matter.
2.1. Simulation Model
All simulations used a simplified straight-line model of an oBLM set-up (
Figure 3), similar to existing simulation studies of oBLMs [
3,
26,
27,
28]. The beamline was modelled as a 2 mm thick copper beam pipe with a 2 cm inner radius, based on the beam pipe dimensions of CLEAR (CERN Linear Electron Accelerator for Research)—the accelerator used to help validate the simulation results. Beam losses were generated by a cylindrical target of solid material placed in the path of the electron beam.
The optical fibre was modelled as two concentric cylinders of fused silica glass, positioned parallel to the beam axis at a distance of 45 cm. Different sets of simulations used different fibre diameters and lengths; this will be discussed alongside the simplifications and modelling assumptions performed for each of these.
Cylinders of fused silica 1 mm thick were placed at each end of the fibre to serve as idealised photodetectors with refractive index-matched fibre connections. To prevent photon number inaccuracies from Cherenkov radiation production inside the photodetector volumes, any charged particles striking them were deleted. Captured photons travelling in the same direction as the beam were recorded by the “downstream” detector, while the “upstream” detector recorded those travelling in the opposite direction.
The refractive index (RI) and attenuation data for the fibre were based on the Thorlabs “FG200LEA” step-index multimode optical fibre [
29], the model used for the oBLM system installed on CLEAR.
The RI of the fibre core
was modelled as pure fused silica, using data from [
30,
31]; while the RI of the cladding
was calculated from the fibre numerical aperture
and the values of
using Equation (
2) [
32]:
Data for the photon attenuation coefficient
(in
) was taken from the Thorlabs website and extrapolated using fitted functions to match its wavelength range to that of the available refractive index data. For wavelengths under 400 nm, the extrapolation function followed a
proportionality based on Rayleigh scattering and UV absorption [
33], while longer wavelengths (above 1000 nm) used an
proportionality based on IR absorption [
34]. Further details of the Geant4 version and physics settings are discussed in
Appendix A.
For the purposes of this simulation, any beam particle that hits the obstacle is considered a “lost particle”; in keeping with the previous definition of “loss intensity” in this paper. Due to the layout of the model, all particles generated at the beginning of the simulation (primary particles) will hit the target and become “lost particles”.
The simulation output consisted of the positions, momenta, and energies recorded for all charged particles that passed through the fibre (the “fibre hits”), as well as all Cherenkov photons that reached the photodetector volumes (the “photon hits”).
2.2. Preliminary Parameter Scan
For a preliminary investigation into the relationship between beam energy and the oBLM signal, a square grid search was performed over beam energies (
) of 7–500 MeV and loss intensities (
) of 1–10 million particles (equivalent to bunch charges of 16–160 fC). This simulation was presented at the IBIC 2025 conference [
35].
In this version of the simulation, the fibre was modelled at actual size (200 μm core diameter), with a 1 cm thick cylindrical copper target that acted as a thick scattering obstacle. The fibre and beam pipe both extended 5 m upstream and downstream of the target for a total length of 10 m.
The beam energy range was chosen to match that of the 6-turn configuration of the under-construction ERL PERLE (Powerful Energy Recovery LINAC for Experiments) [
12,
36], serving as an initial test scenario for the beam loss energy estimation method applied to an ERL energy range. Computing resources constraints restricted the use of a
square grid search incorporating all the PERLE energies, as it was not possible to maintain sufficient loss intensities to ensure measurable photon statistics. A
square grid search was chosen instead, as this was sufficient for the feasibility study investigation of oBLM signal behaviour in this energy range.
The analysis of this data focused on the physical output signal of the oBLM system: the total number of photons exiting the fibre (
). The values of
, with the upstream and downstream photons combined, were plotted against beam energy for all values of loss intensity (
Figure 4, left), as well as against loss intensity for all values of beam energy (
Figure 4, right).
To estimate the growth rate of the photon number with both beam energy and loss intensity, straight-line fits were performed on each grouped data set and their gradients were compared.
The average values of the reduced chi-squared statistic , where is the number of degrees of freedom, were approximately 1.8 for the fits of photon number against beam energy and approximately 1.1 for the fits of photon number against loss intensity. Since all values were within the range , commonly regarded as indicative of an acceptable fit quality, this provides evidence that the expected linear trends are present in the simulation.
The gradients of
against beam energy, in
, had a range of 0.09–0.97 photons/MeV, while the gradients of
against loss intensity had a range of
–
. These results are plotted in
Figure 5.
Since is observed to vary with approximately linear relationships with respect to both beam energy and loss intensity, a sufficiently high loss intensity change is expected to be able to mask the effect of an energy change. This is hereafter referred to as the loss intensity change threshold.
Since the vertical axis intercepts are close to zero for both sets of linear fits in
Figure 4, an equivalent percentage change in either variable is expected to produce an identical change in
. To verify this, example loss intensity change thresholds capable of completely masking the effect of an energy change were estimated via interpolation of the linear fits from
Figure 4.
At each simulated energy value, the fits to
against
were used to estimate the loss intensity change that produced the same increase in
as an energy increase to the next simulated value. For greater consistency of the expected changes in
with energy, these values were estimated using the fits to
against
. The resulting estimates of the loss intensity change thresholds are shown in
Table 1 as percentage changes.
Since no photons were detected for simulations with beam energies of 7 MeV, the gradient of against could not be reliably obtained for this energy; estimations for the 7 MeV → 130 MeV energy change were therefore not performed.
From this data, it is apparent that the percentage change in loss intensity required to mask an energy change is, as expected, both approximately independent of the initial loss intensity value and close to the percentage change in energy. Therefore, for the thick-target beam loss case modelled in this simulation, the loss intensity change is expected to fully mask an energy change if the percentage loss intensity change is approximately equal to or greater than the percentage energy change.
In the case of an ERL, an energy resolution of half the energy change per ERL pass (or better) is likely to be sufficient in distinguishing beam losses from different ERL passes. To achieve this, the effect of the loss intensity must remain below half of the effect of the energy change. However, the percentage energy change decreases for the higher-energy passes. Therefore, the minimum allowable loss intensity change in this case is expected to be the half-masking loss intensity change threshold for the energy change between the two highest-energy passes. This is approximately equal to half of the percentage energy change between those passes.
To further exemplify this, consider the energies of PERLE. With an initial energy of 7 MeV and an energy gain per pass of 82 MeV, the percentage energy change between the two highest-energy passes (418 MeV and 500 MeV) is approximately 19.6%. Therefore, the oBLM is expected to be capable of distinguishing beam losses from these two passes if the loss intensity changes are less than approximately 10%.
Notably, the sensitivity to loss intensity is expected to be lower for ERLs with fewer passes, since the energy gain per pass is larger. The 3-turn variant of PERLE has a maximum energy of 250 MeV and an energy change of approximately 49% between its two highest-energy passes. In this case, the maximum allowable loss intensity change would be approximately 24%.
If a sufficiently low loss intensity change between beam losses at each energy is not achievable in practice, distinguishing losses from different beam energies is then best supported by a complementary means of determining the loss intensity. One example of this is the installation of localised beam loss monitors (BLMs) such as ionisation chambers in key beam loss regions. In this case, the oBLM and the localised BLMs would both need to be calibrated for a range of loss intensities at all energies present in the ERL beam [
37,
38]. Non-invasive beam current monitoring can also be utilised for this [
21], such as with an integrating current transformer (ICT). In an ERL, ICTs could be placed in the mono-energetic recirculating arcs to monitor the beam current changes at each energy. However, this method may be less effective if the change in beam current due to beam losses is close to the resolution limit of the ICT.
2.3. Investigations into the Effects of Fibre Position and Screen Thickness
For verification of the results from the preliminary parameter scan, beam loss signals were recorded from the oBLM system installed on the CLEAR accelerator (discussed fully in
Section 3). However, the properties of the beam target and oBLM set-up available for this experiment exhibited some significant differences to the model used in the parameter scan, necessitating adjustment to the simulations:
Firstly, the yttrium aluminium garnet (YAG) scintillation screen used as the beam target in the experiment is approximately fifty times thinner and half as dense as the copper target used for the simulation. Secondly, the screen is located near the end of the beamline. Due to practical constraints on the oBLM installation in this region imposed by the experimental area and beam dump, the fibre bends away from the beamline immediately downstream of the screen and subsequently exits the beam area. As a result, the effective fibre length available downstream of the screen is reduced in comparison to the parameter scan.
Simulation results from [
26], which investigate the interaction of beam losses with the oBLM system on CLEAR, show that the majority of photons detected in the oBLM are expected to be produced in the section of fibre downstream of the screen. This suggests that a shorter effective downstream length of fibre is expected to reduce the numbers of observed photons.
Additionally, the region where the beam loss shower intersects with the fibre is expected to shift further downstream of the loss position as the beam energy increases, due to the increasing forward momentum of the beam. This has the potential to cause a portion of the particle shower to miss the fibre, with this proportion thought to increase with the beam energy. The thinner and lower-density screen is also expected to cause less deflection of the beam due to its smaller cross-section, which would again shift the region where the fibre intercepts the particle shower to be further downstream. Both of these differences may cause changes to the relationship between the oBLM signal and the beam energy, necessitating investigation into their effects.
2.3.1. Simulation Methods
A suite of simulations were performed to investigate the effects of thinner obstacles and reduced effective downstream fibre length on the relationship of the oBLM signal with energy, as well as to assess the potential for energy estimation with the oBLM in potentially non-optimal situations (such as those found in the experiment on CLEAR).
Seven beam energies were simulated, spanning the 60 MeV to 220 MeV achievable by CLEAR without incurring the cost to beam quality for energies below 60 MeV [
39].
The fibre length downstream of the scattering target was varied as a simplified representation of the reduced effective length caused by the fibre turning away from the beam as it does on CLEAR. Downstream fibre lengths of m, 5 m, and 10 m were simulated; where the 5 m fibre represented the model used in the preliminary parameter scan, and the m fibre aimed to represent the set-up on CLEAR. The 10 m fibre was included to catch a larger proportion of the particle showers, representing more ideal conditions for observing a linearly increasing trend between oBLM signal and beam energy.
Note the m fibre is an overestimation, since the actual downstream distance on CLEAR is closer to 1 m. However, the rate of detected photons produced in simulations with a 1 m downstream fibre length was too low to obtain reliable results within the simulation runtime constraints: a simulation run of 5 million 220 MeV primary electrons (≈ pC) using the 1 m model produced zero Cherenkov photon detections. The m fibre was, therefore, considered an sufficient compromise for the purposes of this feasibility study.
The target material was changed to YAG [
40] to more closely match the properties of the screen on CLEAR. Thicknesses from 0.1mm to 4.0 mm were tested, incorporating thinner screens similar to the experiment at CLEAR as well as thicker screens that could produce more ideal conditions for observing the aforementioned linearly increasing trend in the oBLM signal.
Also, the fibre diameter was increased by a factor of 10 to increase the overall photon statistics and reduce the uncertainty on the results in an effort to reduce the level of statistical fluctuations compared to the preliminary parameter scan. The photon number increase resulting from this change is expected to be a constant scaling factor based on the fibre core radius
R, although there is currently a lack of consensus on the correct formulation for this factor [
20,
41]. Regardless, it is not expected to change the overall shape of the oBLM signal-beam energy relationship.
Therefore, while these changes mean the following simulations are not reliable for absolute calibration of the photon numbers and quantitative comparisons with experiment, they are beneficial for more robust qualitative analysis and identification of trends in the relationship between oBLM signal and beam energy, which are necessary for feasibility analysis. To this end, 5 million primary particles were used per simulation run (equivalent to an pC bunch charge). The upstream and downstream photons were also analysed separately in case of differences in the signal trends between each fibre end.
To estimate additional sources of statistical Monte Carlo uncertainty not considered by the assumption of Poisson statistics used in the preliminary parameter scan; uncertainties for the fibre hits and photon numbers were obtained using a grouped averaging method. The results from each simulation run were split into five equal groups, analysed separately, and the mean and its standard error calculated from the resulting values, taking advantage of the fact that each primary particle in Geant4 occurs independently.
The difference between the percentage uncertainty of this method compared to Poisson statistics is minimal at the lower photon numbers observed in the preliminary parameter scan (approximately 1% difference), but is very noticeable at the higher photon numbers observed in the following simulations (more than 50% difference). While the grouped averaging method came at the cost of a reduction in the effective loss intensity values by a factor of five, it was considered an appropriate compromise for the improved uncertainty estimation.
A second parameter scan simulation over beam energy and bunch charge was also performed to produce data for immediate comparison with CLEAR. The scan used a downstream fibre length of m and a YAG screen thickness of mm, and the same seven beam energies were used for the investigations of the effects of thinner obstacles and reduced effective downstream fibre length.
The four simulated bunch charge (loss intensity) values ranged from 80 to 800 fC; these were chosen due to simulation runtime constraints. After dividing the data into groups for analysis, these bunch charge values increase from 16 to 160 fC.
To make this simulation more similar to the conditions at CLEAR, the detected photons were weighted by the corresponding value of the photon detection efficiency (PDE) for the silicon photomultipliers (SiPMs) used as photodetectors for the experiment. These weights were summed to obtain expected numbers of photon detections
NphPDE, which could then be more easily compared with the experimental results (
Section 3.3). The PDE values were estimated from a spline fit to a graph from the SiPM data sheet [
42], scaled linearly to match the lower fill factor (percentage of detector area sensitive to photons) of the SiPMs in the experiment. Photons outside the sensitive range of the detectors (200 nm to 900 nm) were given a weighting of zero and excluded from the analysis. This 200 nm to 900 nm photon bandwidth limitation was also applied to the simulations investigating fibre downstream length and obstacle thickness, facilitating comparison between the two sets of results.
The results from this simulation are discussed in
Section 3.3.4, along with the results from the experiment at CLEAR.
2.3.2. Simulation Results
The results of the investigations into the effects of changing fibre downstream length and obstacle thickness are presented below.
The smoothing splines in the figures in this section, used to display the shape of the data, were chosen because the simulated behaviour is expected to be continuous between the data points. Since the smoothing spline weights were estimates of the data point standard deviations, the recommended range of the smoothing parameter is suggested by Dierckx [
43] to be
, where
m is the length of the data set. The smoothing parameter was chosen as
m, the midpoint of this range.
Figure 6 presents the effects of changing fibre downstream length. It can be seen that a difference in the downstream fibre length can cause a significant change in the relationship between the beam energy and photon number, with three main patterns observed: decreasing, increasing, and peaked.
These shapes can be explained by
Figure 7, which shows the longitudinal (
z) positions of the fibre hits for each of the tested fibre lengths up to 10 m downstream of the beam loss position. The intensity of the shower is observed to rise after the screen position, reaches a peak, then decreases towards the horizontal axis with a long, slowly decaying tail.
As the beam energy increases, the peak is seen to widen and shift further downstream (
Figure 8), and rises slightly as well. The rise suggests an increasing overall shower intensity, in agreement with the literature, while the widening and downstream shifting suggests the expected narrowing of the transverse angle distribution of the particle shower due to the increasing forward momentum of the beam. As the graph demonstrates, for shorter downstream fibre lengths, the intensity peak can shift well beyond the end of the fibre, leading to much of the shower simply missing it.
Due to this, the m fibre shows a decrease in both the numbers of photon and fibre hits as beam energy increases, since its length is short enough for most of the shower to miss it. Conversely, the 10 m fibre shows increase of both the photon and fibre hits with beam energy that is close to linear, since the shower intensity peak is well within the bounds of the fibre and effects of the greater shower intensity at higher energy becomes more evident. It should be noted that the 10 m fibre data is not always close to linear—this is because the screen is sufficiently thin for the proportion of the shower that still misses the fibre at higher energies to cause the gradient to flatten.
The 5 m fibre results are different, showing a peak around 140 MeV. At this point, the increasing effect from the shower intensity and the decreasing effect from the angular distribution are found to balance out.
A similar effect can be seen when the screen thickness decreases—the shower intensity peak widens, lowers, and shifts further downstream (
Figure 9).
For the thinnest screens in the
m fibre data, shown in
Figure 10a,b, the fibre hits and photon numbers show decreasing relationships with energy—but as the screen thickness increases, the graphs transition first to the peaked shape and then tend towards an upper limit at screen thicknesses of approximately
mm and above. For similar reasons to the 10 m fibre data in
Figure 6, this upper limit shape is not quite a linear increase—but this time due to the shorter fibre length instead of the screen thickness.
In the 10 m fibre data in
Figure 11, the upper limit of the graph shape is much closer to being linear than at
m, although there is a slight steepening of the downstream photon numbers above 140 MeV. It also reaches this upper limit at a slightly lower screen thickness—approximately
mm and above.
In general, the greater proportion of the shower that is intercepted by the fibre, the closer the relationship between the photon number (oBLM signal) and the beam energy approaches to a near-linear increase. More specifically, when the screen (or other obstacle) is either sufficiently thick or there is sufficient fibre length downstream of it, the shape of the photons–energy relationship approaches the linear increase with energy expected from the literature.
In these simulations, the cases with YAG screens thicker than 2 mm and downstream fibre lengths of at least 10 m display the photons–energy relationships that are most similar to a linear increase.
Linear fits to these five data sets, each comprising an upstream and downstream data, produced values in the range of 0.4 to 6.0. Of these, two sets of downstream data exhibited values outside of the range indicative of an acceptable fit quality.
These results suggest that even under idealised conditions, the relationship between photon number and beam energy is not strictly linear and a description of “near-linear” may be considered more accurate. Nevertheless, the generally acceptable fit quality observed across the other data sets indicates that the dependence in these idealised cases may be approximated as linear to a good degree.
In other cases, where the downstream fibre length is effectively shorter or the obstacle is thinner, and the beam energy can follow a highly non-linear relationship and may exhibit a peaked or decreasing shape.
The maximum beam energy resolution of the oBLM is expected to depend on the gradient of the relationship between photon number
and energy. Assuming this relationship is piecewise linear over a small
range, the beam energy
is estimated from
as:
where
m is the local gradient around the measured value of
, and
is the intercept with the
axis when extrapolating this gradient. A higher gradient will result in a lower uncertainty on
from a given uncertainty on
. For this feasibility study, the measurement uncertainty
is assumed to be small and therefore the piecewise linear relationship in Equation (
3) can be assumed to be valid. Beyond this, however, more work is needed.
The largest overall gradients are found in the situations that produce a near-linear increase of photon number with beam energy, which suggests these conditions are likely to be the most useful for energy estimation with the oBLM. It may also be possible to perform energy estimation in situations that produce decreasing relationships of photon number with beam energy, as long as no photon number values correspond to multiple energies. However, the energy resolution may be lower in these situations due to their overall flatter gradients.
Some common obstacles in beam loss situations, such as beam collimation systems, accelerator magnets, or the beam pipe wall, are typically 1 mm or greater in thickness and made of materials denser than YAG, such as steel, copper, or other metals [
44,
45,
46,
47]. This suggests that beam losses in practice may exhibit an near-linear increase of oBLM signal with energy, but local geometry and shielding are also expected to play significant roles and may disrupt this relationship. It is also important to note that some accelerator facilities use lower-density beam pipe materials such as aluminium [
13]; a near-linear increasing relationship of photon number with energy is less likely to appear in these situations.
In some of the observed photons–energy relationships, such as the peaked shapes, there are degenerate photon intensity values which correspond to multiple energies. This prevents energy estimation directly from the Cherenkov photon intensity in these cases, but it may be possible to take advantage of the differences between the photons–energy graph shapes at the upstream and downstream ends of the fibre and use their ratio as an alternative energy estimation method.
Figure 12 presents results from a preliminary investigation into the energy-dependence of the upstream/downstream photon ratio. It can be seen that for many of the situations with clearly degenerate photon intensities (
m fibre, 0.3 mm to 1.0 mm obstacle thickness), the shape of the downstream photon graphs tend to decrease more strongly with energy than their upstream equivalents. This is because the majority of the detected downstream photons are expected to be produced further downstream on the fibre than the detected upstream photons; so as the particle shower shifts further downstream (e.g., with increasing beam energy), the downstream photon yield will drop faster than the upstream photon yield. In these cases, the upstream/downstream photon ratio increases with the beam energy and may be a useful indicator for future beam energy estimation options.
However, as the photons–energy graph shapes tend towards their upper limits and a larger proportion of the shower is caught in the fibre, the ratio tends towards a relationship that is roughly constant with energy (
Figure 12, ≥
mm obstacle thickness). This suggests the potential use of the upstream/downstream ratio for energy estimation may be limited to the specific types of situations described.
4. Discussion
The work presented here partly demonstrates feasibility of a novel potential use case of the oBLM for extracting higher-detail beam loss information from ERL beam—such as determining the turn number of a bunch that produced a given beam loss—through estimating the bunch energy from the intensity of its Cherenkov signal in the oBLM.
The results show that the behaviour of the oBLM signal intensity in response to changing beam energy is complex and there are many additional effects on the signal intensity, including a linear effect from the loss intensity and non-linear effects from the fibre positioning and obstacle thickness (the latter of which roughly corresponds to changes to the beam loss mode).
The simulations suggest a near-linear increase of oBLM signal with beam energy can appear on a straight section of beamline when the beam strikes a thick obstacle (such as a few of steel) and that there is sufficient length of fibre downstream of the loss position to capture a large proportion of the loss shower at all beam energies.
Outside of these situations, the non-linear effects dominate and the signal–energy relationship may adopt a variety of shapes, such as a flattening increase, a peaked shape, or a decrease, depending on the specifics of the fibre positioning and screen thickness. Even in the narrow parameter space investigated, there is a strong effect on the energy response.
These results are supported by data from the oBLM system at CLEAR, CERN, in which beam losses were generated from an mm thick scintillation screen near the end of the beamline. The fibre is required to curve away from the beamline and eventually leave it altogether in this region due to practical restrictions regarding the user experiment area and beam dump following the screen, which reduced the effective length of fibre downstream of the screen.
A decreasing relationship of oBLM signal with energy was observed, showing qualitative agreement with the general trends observed in the simulations for shorter downstream fibre lengths and thinner obstacles.
A peak belonging to an early beam loss was also identified in the oBLM signal, which introduced an underestimation of approximately 10% in the measured charge values that was unfortunately not possible to correct shot-to-shot with the data available. However, since the loss signal from this peak was independent of the beam energy, it was determined to be unlikely to be affecting the shape of the relationship between the oBLM signal and beam energy in the main peak.
These results suggest that while the situations that produce the near-linear increase in the oBLM signal with beam energy have potential to align with certain beam loss conditions on linear accelerators, this is often not the case and the linear increase cannot always be assumed. The signal–energy relationship should be expected to vary depending on the beam loss position on the accelerator, as this drives changes in the discussed non-linear effects.
An example of how changes in the effective downstream fibre length may occur on ERLs and other types of accelerators can be seen in the consideration of beam losses occurring shortly before bends.
Standard oBLM set-ups aim to deliver loss location performance by maintaining the fibre at a constant transverse distance from the beam path. When a beam loss occurs in a straight section of the accelerator shortly before a bend, the fibre curves away from the path of the loss shower as it follows the beam around the bend. This reduces the effective downstream length of the fibre, dependent on the beam loss position. A beam loss within a few f the start of a bend could follow similar energy relationships to the simulated m fibre, while a beam loss further upstream could behave more similarly to the simulated 5 m and 10 m fibres.
Extending the oBLM fibres out into the space past the bends could produce more suitable conditions for observing a linearly increasing relationship between the oBLM signal and beam energy. In an effort to avoid reductions to the loss location performance, the oBLM system could potentially be split up such that the straight and curved beamline sections are monitored by separate fibres. However, the additional fibres and detectors required for such an installation would increase the cost of the system, and the practicality of this idea ultimately depends on spatial constraints around the accelerator.
An alternative option could be to install the oBLM fibre parallel to the beam for as much of the beamline as possible, and determine what type of relationship between the oBLM signal and the beam energy can be expected at different beam loss positions through the use of simulations.
However, the understanding of oBLM signal behaviour for more realistic beamline scenarios is not complete, since the simulations presented here considered only “drift” sections of beamline—straight sections with no magnets or other components. Further simulations moving forward will focus on the benchmarking and development of higher-detail models to investigate the effects of fibre curvature and accelerator magnets on the relationship between oBLM signal, beam energy, and loss position, as well as shielding from other accelerator components.
Another use case for simulations is the investigation of the non-linear effects to obtain extra metrics for the oBLM signal intensity. For example, the simulation results also suggest that the upstream/downstream photon ratio has potential to be useful for situations where direct energy estimation from the oBLM signal intensity is not possible. An example of this is the peaked distributions observed in
Figure 6b and
Figure 10b, which are degenerate in their correspondence of signal integral to beam energy. However, since the uncertainty on these ratios obtained from both the experimental measurements and the simulations were high, future studies will perform more precise measurements of this relationship to extract more detailed conclusions.
In general, the results suggest that beam loss energy estimation with an oBLM may be feasible, but any use of this method as a diagnostic will require detailed, accelerator-specific simulations of the radiation environment around the beamline, including effects such as magnetic fields and local shielding, to guide measurements of the oBLM signal intensity at different loss locations.
Other considerations were also identified: firstly, multi-shot measurements of oBLM signal intensity are recommended due to potential for significant statistical fluctuations to occur in single-shot measurements, such as the up to 60% fluctuation observed in the upstream data from CLEAR.
Secondly, loss intensity is another important factor; its effect on the oBLM signal intensity is capable of masking the effect from the beam energy if the percentage change in loss intensity between two beam losses is greater than half the corresponding percentage change in beam energy. A complementary method of determining the loss intensity is recommended for best performance, such as beam current monitoring or the use of localised BLMs (such as ionisation chambers) placed in key beam loss regions, similar to the BLM set-up described in [
54]. This will require calibration of the oBLM signal intensity against loss intensity—a practice already necessary for signal intensity measurements with the oBLM, since differences in shielding depending on loss location are expected to change the gradient of the oBLM signal intensity response [
38]. The choice of loss intensity correction method is also generally expected to be heavily application-driven and influenced by the diagnostics available at the accelerator facility where the oBLM is to be installed.
Further uncertainty considerations for this method that were not investigated in this feasibility study include differences in local geometry and shielding with loss position, suggested in the existing literature to produce location-dependent signal intensity variations of up to 20% [
21]. Additionally, oBLM signal intensities are reduced by fibre attenuation, with the magnitude of this effect also depending on loss position. At 420 nm, the maximum-PDE wavelength of the SiPMs in the experiment at CLEAR, the attenuation of the fibre is expected to cause signal intensity variations of ∼20% over the full length of the CLEAR accelerator. Calibration of the oBLM signal intensity response for a range of loss intensities and loss positions for all available beam energies is expected to be necessary to mitigate these uncertainties.
Finally, the high repetition rates of ERLs, which can exceed 300 MHz, present an additional challenge for the implementation of an oBLM-based energy estimation method for ERL bunch identification. Since the Cherenkov signals in the oBLM are expected to repeat at the same rate as the beam, ERL repetition rates are expected to saturate the SiPMs. This then becomes a feasibility constraint, since the integration of oBLM signal peaks requires the Cherenkov light pulses from individual beam losses to be distinguishable in the detector signal.
Some mitigation options for this include the use of faster photodetectors with pulse widths of approximately 1 ns or less [
63,
64,
65]; the use of multiple shorter fibres (instead of a single longer fibre) to reduce the effects of pulse broadening; and peak-finding methods that are more robust in the case of overlapping signals, such as least-squares fitting- or Fourier-based deconvolution [
1,
66,
67]. Gated acquisition methods [
68] may also facilitate the extraction of beam loss peaks from specific ERL bunches. The efficacy and implementation of these mitigation options are factors that will need to be considered for a full realisation of this method on an ERL, and will be a main focus of future work.