3.1. Relaxation
The relaxation method is widely recognized as the most reliable OCV measurement approach. This method involves charging or discharging the battery to predetermined SOC levels, followed by rest periods during which the OCV is directly recorded. The overall procedure is illustrated in
Figure 2.
To enable accurate SOC estimation through current integration, the test begins by fully charging the cell to 100% using a constant current–constant voltage (CCCV) protocol. Subsequently, the cell is discharged at a constant current in discrete steps. The accuracy of the OCV–SOC relationship is highly dependent on the number of measurement points, which must be balanced against the total test duration. According to the literature, SOC intervals of 2–5% offer a satisfactory trade-off between resolution and experimental time [
24,
26]. In this study, 5% SOC steps were selected. After the initial full charge, the cell was discharged in 5% SOC increments using a constant current, with the discharge time as the primary control variable. A voltage cut-off was also applied to prevent a deep discharge. Once the minimum voltage was reached, a final rest period marked the end of the discharge phase. The cell was then recharged using the same 5% SOC step protocol and current magnitude to obtain OCV values during charging.
An essential factor in this experiment was the magnitude of the current used for charging and discharging. To avoid excessive heat generation and ensure thermal stability, a relatively low current of 0.1 C was employed. While lower currents could potentially yield more precise results, they would significantly extend the duration of the experiment.
Each rest period was set to 2 h, in line with findings in the literature suggesting that relaxation durations of 2–4 h are generally sufficient for voltage stabilization [
23]. The 2 h interval was chosen to strike a balance between the accuracy and the overall test duration. To evaluate whether the cell reached a relaxed state after 2 h, the voltage difference over the final 5 min of the rest period was calculated. To reduce the impact of measurement noise in this calculation, the voltage difference was not computed directly between two individual points. Instead, each point was represented by the average voltage over a 30 s interval prior to calculating the voltage difference. Therefore, the voltage difference could be calculated as follows:
where
and
are the numbers of samples in the indicated time intervals;
T is the period used to perform the averaging of the signals, which was 30 s;
is the end of the rest period, which was 2 h; and
is the period used to evaluate the cell voltage relaxation, which was 5 min. To be able to directly compare the charge and discharge curves, the absolute value of the voltage difference was considered.
Based on the chosen parameters, the total duration of the test was estimated by summing the active and passive phases. At 0.1 C, the full charge and discharge cycle required approximately 20 h. Given that there were 21 rest periods during discharge and 20 during charge—each lasting 2 h—the total rest time amounted to 82 h. Therefore, the total estimated duration of the experiment was approximately 102 h.
Because the relaxation method is considered the most direct and reliable approach for determining the OCV, the value obtained using the 5% SOC step size and 2 h rest periods will serve as a reference OCV in this paper, and the errors of the other methods will be calculated using this reference. The rationale for choosing either the charging or discharging OCV curve as a reference will be discussed in detail in the following section.
Two key parameters significantly influence both the accuracy and total duration of the OCV measurement experiment: the SOC step size and rest-period duration. In the baseline test configuration, 5% SOC steps and 2 h rest periods were selected.
The baseline test design parameters were selected to ensure the practical feasibility of completing the experiments on three different cells. The 5% SOC step size was adopted to provide sufficient resolution for capturing the key characteristics of the OCV curve. While this choice may limit the ability to resolve fine details at extreme SOC levels, many practical applications predominantly operate cells within the mid-SOC range. Therefore, the selected SOC increment represents a balanced compromise between the measurement resolution and total test duration. A similar rationale was applied in selecting the 2 h rest period. According to the literature, a rest duration of 1–2 h is sufficient to observe the OCV with voltage deviations that are effectively negligible and below the measurement accuracy of the instrumentation. Therefore, a 2 h rest period was chosen to minimize the overall test duration while still ensuring a reliable estimation of the true OCV [
27]. However, because of practical constraints related to the time and energy consumption, this study also investigated the effects of using larger SOC steps and shorter rest durations on the accuracy and reliability of the resulting OCV measurements.
To evaluate the sensitivity of the OCV to the density of SOC points, the measurements were taken at different SOC intervals larger than 5% (i.e., 10%, 15%, and 20%). To find the error of each OCV curve relative to the reference, the obtained OCV of each case was compared with the original OCV measured using the 5% SOC interval. Because data were not directly available for all the points, interpolation was needed. Therefore, a linear interpolation was used to calculate the OCV of each case at the missing SOC points. It is important to note that the error was calculated only at the measurement points.
There are various OCV models (curve expressions) available in the literature to fit an OCV vs. SOC curve. Some are more complex than others, which can make it harder to fit when the number of points is limited. As indicated in [
20], each of these models has different sensitivities to the number of obtained points. A 5th-order polynomial function was selected to interpolate the OCV vs. SOC curve. The accuracies of the OCV estimations using different SOC point densities were also evaluated using a 5th-order polynomial function. The errors in each case were calculated separately for the charge and discharge cycles.
Finally, to evaluate the required relaxation time, Equation (1) was used by varying the
from 5 to 120 min for all the SOC points available. The error could then be expressed as a percentage as follows:
Thus, it was possible to assess how the relaxation time affected the accuracy of the OCV for different SOC points.
Additionally, it should be noted that all of the error and root mean square error (RMSE) values reported in this paper were normalized with respect to the rated voltage. This normalization enabled the results to be readily extended to higher-capacity and higher-voltage energy-storage systems, thereby enhancing their general applicability.
3.2. Low Current Charge/Discharge Curve
This method relies on the principle that, with a low current, the overpotential on the electrodes is minimized, allowing it to be disregarded. Consequently, the voltage recorded during charging or discharging can be considered to be the OCV. To conduct the test, the cell should first be charged in the CCCV mode until it reaches its full charge. Following this, a rest period is required to allow the charges to diffuse in the electrolyte and the cell temperature to reach ambient conditions. In this study, a 0.5 C current was employed, followed by a 2 h rest interval. The procedure for this test is straightforward. After charging the cell, a CCCV discharge is conducted, followed immediately by a CCCV charge without an additional rest period. To perform the test at different C-rates, the fully charged cell was rested for 2 h, and the procedure was restarted with a new current.
It should be noted that using lower current values results in data that more accurately reflect the true OCV because the overpotential at the electrodes is low, although this also exponentially increases the total duration of the test, as shown in
Figure 3. It should also be noted that the estimated time reported does not take into account the time necessary for the CV part of the test, which can differ with the chemistry because of differences in the OCV shape. In this study, three different C-rates (0.025 C, 0.1 C, and 0.2 C) were employed to obtain the OCV curves, enabling a comparative analysis of the trade-off between the measurement accuracy and total test duration. The overall procedure had to be repeated for each current and is illustrated in
Figure 4.
Additionally, it should be noted that the error in the OCV obtained from the low-current test was mainly attributable to neglecting the presence of the ohmic voltage drop, which, although reduced by using a low current, could not be completely eliminated. In addition, residual concentration polarization may also have contributed to the observed error.
3.3. Average at Rated Current
Conducting experiments at the rated current is not a widely recognized method for determining the OCV. Although it is not the most accurate approach due to various inherent limitations and potential inaccuracies, it can provide a rough estimation of the OCV while offering some notable advantages. For instance, this method is considerably faster than alternative testing procedures, and the required profile can be easily obtained in operational systems. In other words, the standard operating conditions of many systems can yield a full charge–discharge cycle at a rated current, making it possible to estimate an OCV profile without extensive modifications. Therefore, it is valuable to briefly examine this method as a potential approach. On the other hand, the use of higher currents can introduce inaccuracies in the measured OCV as a result of the increased influence of internal resistance variations across different SOC levels. Additionally, higher currents can lead to greater temperature fluctuations within the cell, further affecting the accuracy of the OCV measurements.
This testing procedure is also simpler than other methods. The overall procedure is illustrated in
Figure 5. Following a full charge achieved through a CCCV approach, both discharge and charge phases are conducted at the rated current, without any resting interval between them. A diagram of this test procedure is provided in
Figure 5. It is important to note that the resulting charging or discharging curve does not accurately represent the true OCV because significant overpotentials occur at the electrodes. Thus, only the average of the charge and discharge curves is considered. To compute the average of the charge and discharge curves, both curves are first linearly interpolated onto a common SOC vector with a resolution of 0.1%. This step is necessary because the SOC points of the two curves are not necessarily aligned. Owing to the high sampling rate during the test, the error introduced by this interpolation is negligible. Finally, considering that the two curves (i.e., charge and discharge) are obtained at the same current rate, the OCV is obtained as the arithmetic mean of the two interpolated curves.