2. Materials and Methods
2.1. Reagents, Media, and Buffer Solutions
Amplex Red reagent (N-acetyl-3,7-dihydroxyphenoxazine, lot 13489098) and horseradish peroxidase (HRP, lot 1138650A) were obtained from Invitrogen/Thermo Fisher Scientific (Waltham, MA, USA). Ferric chloride (FeCl3), ferrous chloride tetrahydrate (FeCl2·4H2O), aluminum chloride hexahydrate (AlCl3·6H2O), potassium dichromate (K2Cr2O7), phosphate-buffered saline (PBS, 10× solution, pH 7.2–7.6 at 25 °C, cat. no. P5493-1L), sodium chloride, and other analytical-grade chemicals were purchased from Sigma-Aldrich Chemie (Steinheim, Germany). The 10× PBS contained 1.37 M NaCl, 27 mM KCl, 100 mM Na2HPO4, and 18 mM KH2PO4. Before use, it was diluted tenfold with ultrapure water to obtain 1× PBS containing 137 mM NaCl, 2.7 mM KCl, 10 mM Na2HPO4, and 1.8 mM KH2PO4.
HB1 buffer was prepared to contain 136 mM NaCl, 5 mM KCl, 2 mM MgCl2, 2 mM CaCl2, 10 mM HEPES, and 10 mM glucose, adjusted to pH 7.4. Complete cell culture medium consisted of Dulbecco’s Modified Eagle’s Medium (DMEM, cat. no. D5546, Sigma-Aldrich Chemie, Steinheim, Germany) supplemented with 9% fetal bovine serum, 1% L-glutamine solution, 100 U/mL penicillin, and 100 µg/mL streptomycin. Distilled water was used as a low-ionic-strength and low-conductivity control medium.
A concentrated Amplex Red stock solution was prepared and protected from light. A fresh Amplex Red/HRP working solution was prepared before each experiment by diluting the Amplex Red stock solution in the corresponding assay medium and adding HRP to the same solution. The Amplex Red/HRP working solution added to the microplate wells containing 100 µM Amplex Red and 0.2 U/mL HRP, resulting in final assay concentrations of 50 µM Amplex Red and 0.1 U/mL HRP after mixing with the remaining reaction components. All Amplex Red-containing solutions were protected from light, prepared freshly on the day of each experiment, and maintained at room temperature.
2.2. Electric Pulse Treatment and Electrode Specifications
Electric pulse treatment was performed to simulate electroporation-related conditions. For pulse-exposure experiments, 50 µL aliquots of test medium or Amplex Red/HRP-containing medium were placed between two flat parallel stainless-steel electrodes separated by a 2 mm gap. The electrodes were prepared from stainless-steel plates measuring 50 mm × 10 mm × 1 mm. The stainless steel used for the electrodes consisted mainly of iron, chromium, nickel, manganese, titanium, cobalt, copper, and molybdenum and corresponded to an austenitic 300-series stainless steel [
12].
Square-wave electric pulses were applied using a BTX ECM 2001 Electro Cell Manipulator (Harvard Apparatus, Holliston, MA, USA). Depending on the experiment, the pulse duration ranged from 0.1 to 2 ms and the pulse amplitude ranged from 40 to 400 V, corresponding to electric field strengths of 0.2–2.0 kV/cm. Pulse shape and parameters were monitored using a TBS 1072B-EDU oscilloscope (Tektronix U.K. Ltd., Berkshire, UK). After pulse application, treated samples were collected immediately and used for fluorescence measurements.
For direct metal-ion addition experiments, the pulse-treatment step was omitted in order to isolate the effect of selected metal ions on the Amplex Red/HRP/H2O2 fluorescence readout independently of pulse-induced electrochemical reactions.
2.3. Amplex Red Fluorescence Assays
Fluorescence assays were performed to evaluate the influence of medium composition, electric pulse exposure, pulse parameters, pH conditions, and electrode-relevant metal ions on the Amplex Red/HRP fluorescence readout. Reactions were performed in black 96-well plates using freshly prepared Amplex Red/HRP working solution, and fluorescence intensity was measured immediately at room temperature using a spectrofluorimeter Genios Pro (Tecan Austria GmbH, Salzburg, Austria). Excitation and emission wavelengths were 535 nm and 590 nm, respectively. Experiments were performed on three separate days using freshly prepared media and reagents.
To evaluate the effect of medium composition on the Amplex Red/HRP/H2O2 assay, reaction mixtures were prepared in PBS, HB1 buffer, complete DMEM, or distilled water. Each well contained 10 µL of H2O2 solution, 40 µL of the respective medium, and 50 µL of Amplex Red/HRP working solution, giving a final volume of 100 µL. The final assay concentrations were 0.4 µM H2O2, 50 µM Amplex Red, and 0.1 U/mL HRP.
To evaluate fluorescence formation after electric pulse treatment in different media, Amplex Red/HRP-containing solutions were prepared in PBS, HB1 buffer, complete DMEM, or distilled water. For each condition, 50 µL of the corresponding Amplex Red/HRP-containing medium was placed between stainless-steel electrodes and exposed to 1 or 3 rectangular electric pulses of 100 V amplitude and 2 ms duration. Non-pulsed samples served as controls. Immediately after treatment, samples were transferred to a 96-well plate, and fluorescence intensity was measured.
To examine the effect of pulse parameters on the Amplex Red fluorescence readout, Amplex Red/HRP-containing medium was exposed to single rectangular electric pulses of different amplitudes and durations. Pulse amplitudes of 100–400 V and pulse durations of 0.5 and 2 ms were applied, and fluorescence was measured immediately after pulse treatment. This experiment was used to determine whether an increasing electrical load produced a monotonic increase in fluorescence or whether signal suppression occurred under stronger pulse conditions.
The effect of pH on the Amplex Red/HRP/H2O2 fluorescence response was assessed using pH-adjusted citrate, phosphate, borate, and glycine buffer systems. H2O2 dilutions were prepared in the corresponding pH-adjusted buffers and mixed with Amplex Red/HRP working solution in a 1:1 ratio in 96-well plates. The final H2O2 concentrations ranged from 0.02 to 2 µM. This experiment was performed to determine whether pH-dependent changes in the assay environment affected resorufin fluorescence formation.
For metal-ion interference experiments, Fe2+, Fe3+, Al3+, and Cr6+ solutions were added directly to the Amplex Red/HRP/H2O2 reaction mixture. Ferric chloride and ferrous chloride tetrahydrate were dissolved separately in distilled water to prepare Fe3+ and Fe2+ stock solutions, respectively. Aluminum chloride hexahydrate and potassium dichromate were dissolved in 0.9% NaCl to prepare Al3+ and Cr6+ stock solutions. Working solutions were then prepared freshly from these stocks to obtain the required final concentrations.
For each metal-ion assay, the wells contained 10 µL of metal-ion solution, 10 µL of H2O2 solution, 30 µL of PBS or the corresponding assay medium, and 50 µL of Amplex Red/HRP working solution, giving a final reaction volume of 100 µL. Control wells contained the corresponding vehicle solution instead of metal ions. Two final H2O2 concentrations were tested: 0.4 µM and 10 µM. The 10 µM H2O2 concentration was selected as a higher peroxide-availability condition to evaluate whether metal-ion-mediated fluorescence suppression persists when the assay is less limited by H2O2 availability. The tested final concentration ranges were 0.01–5.0 mM for Fe2+, Fe3+, and Al3+, and 0.1–1000 µM for Cr6+.
2.4. Electrical Conductivity Measurement of Media
The electrical conductivity of the tested media was determined by conductometric analysis. Measurements were performed for distilled water, HB1 buffer, PBS, and cell culture medium using platinum electrodes. Because the exact electrode surface area and distance between electrodes were not known, the cell constant was determined using a 100 mM KCl standard solution. The resistance of the KCl standard and each tested medium was measured using an LCR E7-11 conductometer, and the specific electrical conductivity was calculated from the measured resistance and the determined cell constant. Conductivity values were expressed as S/m.
2.5. Determination of Nickel and Chromium Release from Stainless-Steel Electrodes
To evaluate the release of nickel and chromium ions from stainless-steel electrodes during pulse exposure, filtered sterile 0.9% NaCl solution was placed in a stainless-steel cuvette and exposed to rectangular electric pulses of 2 ms duration. The electric field strength was varied from 0.2 to 1.2 kV/cm. After pulse treatment, the solution was collected from the cuvette, and nickel and chromium ion concentrations were determined by atomic absorption spectrophotometry using a Perkin Elmer (Waltham, MA, USA) Zeeman 30/30 instrument. These measurements were performed to assess whether pulse-treated media could contain electrode-derived contaminants at concentrations relevant to analytical interference.
2.6. Data Presentation and Analysis
Data are presented as averages of three independent experiments (n = 3), each performed on a separate day with freshly prepared media and reagents. Fluorescence values were expressed either as absolute fluorescence intensity or as values normalized to the corresponding control condition, as indicated in the figure legends. For medium-comparison analyses, PBS was used as the reference condition and set to 100%. For metal-ion experiments, fluorescence intensity was expressed relative to control wells without added metal ions. Raw data were organized in Microsoft Excel (Microsoft Corporation, Redmond, WA, USA), and graphs were prepared using SigmaPlot 11.0 (Systat Software Inc., San Jose, CA, USA).
4. Discussion
The present study examined how electric pulse parameters, electrical conductivity, medium composition, pH conditions, and as well as electrode-relevant metal ions influence the Amplex Red/HRP/H2O2 fluorescence readout under conditions relevant to high-voltage electric pulse treatment. The results show that the assay response is highly context-dependent.
Phosphate-buffered saline, HB1 buffer, complete DMEM, and distilled water were treated with electrical pulses and the formation of hydrogen peroxide as a result of this treatment was evaluated by using Amplex Red, which is a fluorogenic probe for H
2O
2 detection [
8,
16]. It turned out that the fluorescence of resorufin, which is the product of the reaction of Amplex Red with H
2O
2 catalyzed by the horseradish peroxidase (HRP), depends not only on the parameters of electric treatment, such as, pulse amplitude, duration, and number. The intensity of fluorescence differed markedly between PBS, HB1 buffer, complete DMEM, and distilled water (
Figure 1).
Main factors affecting the fluorescence of Amplex Red/HRP/H
2O
2 system in electric pulse treated media, such as, the electical conductivity of the medium, its composition and pH and as well as various metal ions, which can be released from the electrodes during high-voltage pulses, have been studied. Because the intensity of electrochemical reactions during electric pulse treatment depends strongly on electric current density, which is determined by the electrical conductivity of the medium, it was expected that the intensity of the hydrogen peroxide formation would depend on the specific conductance of each medium. To ascertain whether this assumption is correct, the electrical conductivities of the tested media were determined and compared with their corresponding pulse-induced fluorescence responses (
Figure 2).
The results obtained provide some explanation for the medium-dependent pulse responses albeit incomplete one. Electrical conductivities of the media used in this study followed the order PBS > HB1 buffer > complete DMEM > distilled water (
Figure 2a), which broadly paralleled the pulse-induced fluorescence trend (
Figure 2b). This agrees with the expectation that electrochemical reactions at the electrode–solution interface depend on current density and therefore on medium conductivity. However, conductivity did not fully explain the fluorescence response. HB1 had conductivity close to PBS but produced substantially lower fluorescence, while complete DMEM also showed a response that could not be predicted from conductivity alone. Therefore, pulse-induced Amplex Red fluorescence reflects at least two overlapping processes: electrochemical oxidant generation during pulse exposure and medium-dependent performance of the Amplex Red/HRP/H
2O
2 system. Buffer composition, pH stability, ionic strength, peroxide-scavenging components, amino acids, and serum-derived constituents in complete DMEM may all contribute to the final fluorescence output.
Even when identical nominal H
2O
2 concentrations were used, fluorescence intensity differed markedly between PBS, HB1 buffer, complete DMEM, and distilled water, which were not treated with electric pulses (
Figure 3). In addition, electric pulse-treated samples showed strong dependence on medium conductivity (
Figure 2) and pulse conditions (
Figure 1), while electrode-relevant metal ions directly suppressed the apparent fluorescence signal (
Figure 5,
Figure 6,
Figure 7 and
Figure 8). These findings demonstrate that Amplex Red fluorescence in pulse-treated media cannot be interpreted as a direct and interference-free measure of H
2O
2 without appropriate controls.
The medium-comparison experiments demonstrated a pronounced matrix dependence of the Amplex Red/HRP/H
2O
2 fluorescence response, with PBS producing the highest signal and distilled water the lowest. This finding is important because Amplex Red-based H
2O
2 estimation usually relies on comparison with a calibration curve. If calibration standards and experimental samples are prepared in different media, the calculated H
2O
2 concentration may be substantially biased. The observed medium dependence suggests that the chemical matrix affects HRP-catalyzed Amplex Red oxidation, resorufin fluorescence, peroxide availability, or a combination of these factors. This interpretation is consistent with previous reports showing that quantitative use of the Amplex Red assay depends on pH, H
2O
2 concentration range, and reaction conditions [
19] and that reducing compounds such as NADH and reduced glutathione can interfere with the signal [
23]. Komlódi et al. also showed that Amplex UltraRed-based H
2O
2 flux measurements can be affected by medium-specific background fluorescence and H
2O
2-independent artifacts, reinforcing the need for medium-specific controls [
15].
The different fluorescence responses observed in PBS, HB1 buffer, complete DMEM, and distilled water can be explained by the different chemical environments provided by these media. The Amplex Red assay depends on HRP-catalyzed oxidation of Amplex Red by H
2O
2 to form fluorescent resorufin; therefore, any factor that affects HRP activity, H
2O
2 availability, Amplex Red oxidation, or resorufin fluorescence can alter the final signal [
16,
19,
20,
23]. PBS is a relatively simple, well-buffered salt solution and therefore provides favorable and reproducible conditions for the HRP/Amplex Red reaction. HB1 buffer also provides buffering capacity and physiological salts, but its different ionic composition and the presence of glucose may alter the enzyme microenvironment, peroxide stability, or fluorescence yield compared with PBS. Complete DMEM is substantially more chemically complex because it contains inorganic salts, amino acids, vitamins, glucose, phenol red, serum-derived proteins and other components. These constituents may scavenge H
2O
2, act as reducing or redox-active species, interact with HRP, contribute to background absorbance/fluorescence, or partially suppress resorufin fluorescence [
8,
23,
24]. Future studies using resorufin-spiking or post-reaction dilution controls will be needed to distinguish whether the lower signal in complete DMEM mainly reflects reduced HRP-mediated Amplex Red oxidation, altered H
2O
2 availability, optical/background effects, or direct effects on resorufin fluorescence. Distilled water, in contrast, has very low ionic strength and essentially no buffering capacity, which may reduce HRP reaction efficiency and make resorufin fluorescence more sensitive to changes in pH.
Thus, the observed order of fluorescence intensity, PBS > HB1 > complete DMEM > distilled water, should not be interpreted as reflecting differences in the nominal H
2O
2 concentration alone. Instead, it indicates that the Amplex Red/HRP/H
2O
2 readout is strongly matrix-dependent. This is particularly important in electric pulse experiments, where electrochemical reactions at the electrode surface may further change local pH [
25], generate redox-active species [
8], and release metal ions [
4,
5,
6,
7,
9,
10,
11,
12,
13,
14]. Such ions may consume H
2O
2 through Fenton-type reactions or interfere directly with Amplex Red/resorufin fluorescence [
15,
26]. Therefore, medium-matched calibration curves and spike-recovery controls are necessary when Amplex Red is used for quantitative H
2O
2 assessment in electroporation-related media.
The pulse-parameter experiment provides a key methodological warning. Under short-pulse conditions, fluorescence increased with voltage, which is consistent with the expectation that stronger electrical input increases electrochemical activity. However, under the longer pulse condition, the fluorescence signal no longer increased monotonically; a stronger pulse condition produced a lower signal than a weaker pulse condition. This signal inversion indicates that stronger electrical treatment does not necessarily result in higher measured Amplex Red fluorescence. Possible contributors include electrode-derived products, local pH shifts near the electrode surfaces, peroxide consumption, HRP inhibition, resorufin quenching, or direct redox interactions with Amplex Red. Additional pulse-related effects, such as gas bubble formation, local heating, electrolysis-driven acidification or alkalinization, may also contribute [
6,
7]. Degradation or redox conversion of Amplex Red/resorufin at the electrode interface may additionally affect the fluorescence response [
15,
19,
23]. However, bulk pH was not measured immediately after each pulse condition, and local pH changes near the electrodes cannot be excluded. Similar concerns have been raised for Amplex-type assays, where fluorescence may be influenced by background oxidation or interfering redox reactions rather than H
2O
2 alone [
15,
19,
23]. Therefore, in pulse-treated systems, a lower fluorescence signal should not automatically be interpreted as lower H
2O
2 formation.
The detection of nickel and chromium ions in pulse-treated 0.9% NaCl confirms that stainless-steel electrodes are not chemically inert under the tested pulse conditions. Both Ni and Cr concentrations increased with electric field strength, showing that electrode-derived contaminants can accumulate in the treated medium. This agrees with earlier studies showing that stainless-steel electrodes can release metal ions during electroporation or pulsed electric field exposure [
5,
9,
10,
11,
27]. The recent study by Maček Lebar et al. provides particularly relevant support, showing that electrode dissolution during electroporation releases metal byproducts into the medium and that stainless-steel electrodes, especially under longer pulse protocols, can increase iron, chromium, and nickel concentrations [
26]. The present study extends this issue to analytical fluorescence measurements by showing that electrode-relevant metal ions can directly suppress the Amplex Red/HRP/H
2O
2 fluorescence signal. Thus, corrosion products may lead to underestimation or distortion of apparent H
2O
2 formation in pulse-treated media.
Among the tested ions, Fe
2+ produced the strongest fluorescence suppression, particularly at a low H
2O
2 concentration. At 0.4 µM H
2O
2, Fe
2+ reduced fluorescence by up to 92%, followed by Fe
3+, Cr
6+, and Al
3+ (
Figure 7). This result is chemically plausible because Fe
2+ can react with H
2O
2 through Fenton reactions, consuming peroxide and generating hydroxyl radicals [
24,
28]:
In the context of the Amplex Red assay, H2O2 consumption would reduce substrate availability for HRP-mediated resorufin formation, while secondary radical products may further affect Amplex Red or resorufin. Although the present experiments do not distinguish these mechanisms directly, the stronger Fe2+ effect at 0.4 µM H2O2 than at 10 µM H2O2 suggests that competition for or depletion of available H2O2 through Fenton-type reactions is likely an important contributor, while direct redox effects on Amplex Red/resorufin or fluorescence quenching cannot be excluded. Fe3+ also reduced fluorescence, although less strongly than Fe2+, suggesting that iron-mediated interference is not restricted to the reduced form and may involve redox cycling, complexation, or indirect effects on assay components. The suppressive effects of Al3+ and Cr6+ were weaker but still evident, showing that non-iron electrode-relevant ions can also distort the assay. Importantly, the relative suppressive effects were lower at 10 µM H2O2 than at 0.4 µM H2O2, indicating that low-peroxide measurements are especially vulnerable to interference.
The pH-dependent fluorescence measurements further show that the Amplex Red readout is sensitive to the pH and buffer compositions (
Figure 8). This agrees with Towne et al., who reported that HRP-catalyzed oxidation of dihydroxyphenoxazine derivatives depends strongly on pH and H
2O
2 concentration [
19]. This sensitivity is particularly relevant to electric pulse experiments because water electrolysis at electrode surfaces can generate H
+ near the anode and OH
− near the cathode [
7,
25]. In unbuffered or weakly buffered media, such as distilled water, these local pH shifts may substantially alter the assay environment. Even in buffered media, the local pH near the electrode surface may deviate from the bulk pH during pulse exposure [
25]. Therefore, pH and buffer capacity provide an additional explanation for why conductivity alone does not predict the fluorescence response across media.
In addition, the appearance of the ions of iron and aluminum in the medium changes its pH. This is because these ions behave as a Lewis acid and undergo spontaneous hydrolysis reactions, e.g., [
12]:
Because the fluorescence of Amplex Red/HRP/H2O2 system strongly depends on the medium pH (the lower the pH the weaker fluorescence is), appearance of iron or aluminum ions in the medium leads to the reduction of the fluorescence not only because H2O2 can be consumed as a result of Fenton reactions but because of decrease in the medium pH caused by these ions.
The present findings do not mean that Amplex Red is unsuitable for electroporation studies. Amplex Red and related ADPH-based systems remain useful for ROS-responsive fluorescence detection, as illustrated by their continued use in biological imaging and controlled analytical systems [
17,
18]. However, the present data show that in pulse-treated media, Amplex Red fluorescence cannot be treated as an interference-free standalone measure of H
2O
2. For reliable use of the assay in electroporation-related experiments, several controls are necessary. First, calibration curves should be prepared in the same medium as the experimental samples. Second, spike-recovery experiments should be performed to determine whether known H
2O
2 additions are recovered accurately after pulse exposure. Third, metal-ion controls should be included when stainless-steel or other dissolvable electrodes are used. Fourth, pH and conductivity should be monitored or controlled, particularly in weakly buffered media. Fifth, where quantitative hydrogen peroxide determination is required, Amplex Red fluorescence should ideally be supported by complementary approaches, such as catalase controls, electrochemical H
2O
2 detection [
29], or other orthogonal ROS-detection methods [
30]. For practical H
2O
2 measurements in electroporated cell suspensions prepared in complete DMEM using stainless-steel electrodes, we consider the minimum control set to include medium-matched H
2O
2 calibration, non-pulsed cell-containing controls, pulse-treated medium-only controls, H
2O
2 spike-recovery controls under the same medium/electrode conditions, and catalase-treated controls to confirm H
2O
2-dependent signal contribution.
Several limitations should be acknowledged. The present experiments do not distinguish between direct quenching of resorufin fluorescence, Fenton-mediated H
2O
2 consumption, HRP inhibition, and degradation of Amplex Red or resorufin by secondary radicals. Mechanistic experiments using pre-formed resorufin would help clarify whether metal ions suppress fluorescence by quenching the fluorescent product directly or by interfering with its formation. Because chromium release was detected from stainless-steel electrodes [
9], Cr
6+ was included in the present interference experiment as a soluble and chemically well-defined chromium-containing redox-active model species to assess whether chromium-related compounds may interfere with the Amplex Red/HRP/H
2O
2 readout. Chromium speciation in pulse-treated media may vary depending on electrode material, pulse parameters, pH, oxygen availability, and medium composition [
9,
10,
11]. Because Cr
3+ differs from Cr
6+ in solubility, hydrolysis, complexation, and precipitation behavior, especially in phosphate- or protein-containing media [
31], its effect on the assay may differ and would require separate controlled experiments.
In addition, although Ni and Cr release from stainless-steel electrodes was detected, Ni
2+ was not directly tested in the Amplex Red assay. Furthermore, the present study evaluated metal ions individually and did not examine mixed-ion systems, although pulse-treated media may contain several electrode-derived ions simultaneously, and these mixtures may produce additive, antagonistic, or synergistic effects on the fluorescence readout. Therefore, future studies should evaluate mixed-ion combinations that more closely reproduce the chemical composition of pulse-treated electroporation media. The tested metal-ion concentrations, particularly for iron, also exceed the bulk concentrations typically measured after pulse exposure [
4,
5,
9,
12], although localized concentrations near the electrode surface may be higher. Finally, the pH experiments were performed in buffer systems rather than during real-time pulse exposure, so the combined effects of pulse-induced pH shifts, metal-ion release, and Amplex Red chemistry remain to be investigated.
In summary, the Amplex Red/HRP/H2O2 fluorescence readout in electroporation-relevant systems is shaped by multiple interacting factors, including medium composition, conductivity, pulse parameters, pH/buffer conditions, and electrode-derived metal ions. The practical implication is that Amplex Red should not be used as a standalone, interference-free H2O2 probe in pulse-treated media. Instead, the fluorescence signal should be interpreted as an assay-dependent readout that requires appropriate medium-specific, pulse-condition, pH, and metal-ion controls.