Geometry Alteration of Orthodontic Microimplants Following In Vivo Loading: Microanalysis on 3D Scans—A Cross-Sectional Study
Abstract
1. Introduction
2. Materials and Methods
2.1. Ethical Approval and Reporting Standards
2.2. Specimen Selection and Preparation
2.3. Preclinical Processing and Clinical Procedure
2.4. Surgical Procedure and Loading Protocol
2.5. 3D Analysis and Data Processing
3. Results
4. Discussion
5. Conclusions
Supplementary Materials
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
| MI | mini-implants, microimplants |
| TAD | Temporary anchorage device |
| Spec. | Specimen |
References
- Cope, J.B. Temporary anchorage devices in orthodontics: A paradigm shift. Semin. Orthod. 2005, 11, 3–9. [Google Scholar] [CrossRef]
- Gainsforth, B.; Higley, L. A study of orthodontic anchorage possibilities in basal bone. Am. J. Orthod. Oral Surg. 1945, 31, 406–417. [Google Scholar] [CrossRef]
- Creekmore, T.D.; Eklund, M.K. The possibility of skeletal anchorage. J. Clin. Orthod. 1983, 17, 266–269. [Google Scholar]
- Kanomi, R. Mini-implant for orthodontic anchorage. J. Clin. Orthod. 1997, 31, 763–767. [Google Scholar] [PubMed]
- Costa, A.; Raffaini, M.; Melsen, B. Miniscrews as orthodontic anchorage: A preliminary report. Int. J. Adult Orthod. Orthognath. Surg. 1998, 13, 201–209. [Google Scholar]
- Melsen, B.; Costa, A. Immediate loading of implants used for orthodontic anchorage. Clin. Orthod. Res. 2000, 3, 23–28. [Google Scholar] [CrossRef]
- Park, H.S. The skeletal cortical anchorage using titanium microscrew implants. Korean J. Orthod. 1999, 29, 699–706. [Google Scholar]
- Antoszewska-Smith, J.; Sarul, M.; Łyczek, J.; Konopka, T.; Kawala, B. Effectiveness of orthodontic miniscrew implants in anchorage reinforcement during en-masse retraction: A systematic review and meta-analysis. Am. J. Orthod. Dentofac. Orthop. 2017, 151, 440–455. [Google Scholar] [CrossRef] [PubMed]
- Chaimanee, P.; Suzuki, B.; Suzuki, E.Y. “Safe zones” for miniscrew implant placement in different dentoskeletal patterns. Angle Orthod. 2011, 81, 397–403. [Google Scholar] [CrossRef]
- Crismani, A.G.; Bertl, M.H.; Celar, A.G.; Bantleon, H.P.; Burstone, C.J. Miniscrews in orthodontic treatment: Review and analysis of published clinical trials. Am. J. Orthod. Dentofac. Orthop. 2010, 137, 108–113. [Google Scholar] [CrossRef]
- Eliades, T.; Zinelis, S.; Papadopoulos, M.A.; Eliades, G. Characterization of retrieved orthodontic miniscrew implants. Am. J. Orthod. Dentofac. Orthop. 2009, 135, 10.e1–10.e7. [Google Scholar] [CrossRef]
- Patil, P.; Kharbanda, O.P.; Duggal, R.; Das, T.K.; Kalyanasundaram, D. Surface deterioration and elemental composition of retrieved orthodontic miniscrews. Am. J. Orthod. Dentofac. Orthop. 2015, 147, S88–S100. [Google Scholar] [CrossRef]
- Walter, A.; Wendl, B.; Ploder, O.; Mojal, S.; Puigdollers, A. Stability determinants of bone-borne force-transmitting components in three RME hybrid expanders—An in vitro study. Eur. J. Orthod. 2017, 39, 76–84. [Google Scholar] [CrossRef]
- Jedliński, M.; Janiszewska-Olszowska, J.; Mazur, M.; Grocholewicz, K.; Suárez Suquía, P.; Suárez Quintanilla, D. How does orthodontic mini-implant thread minidesign influence stability? Systematic review with meta-analysis. J. Clin. Med. 2022, 11, 5304. [Google Scholar] [CrossRef]
- Stefaniak, K.; Jedliński, M.; Mazur, M.; Janiszewska-Olszowska, J. Fracture and deflection of orthodontic miniscrews—A systematic review. Appl. Sci. 2024, 14, 5577. [Google Scholar] [CrossRef]
- Jiman, P.A.; Prodan, D.; Moldovan, M.; Muntean, A.; Sarosi, C.; Tarmure, V.; Baciut, G.; Popa, C.; Pop, A.S. New and recovered temporary anchorage devices: In vitro assessment of structural and surface properties. Materials 2021, 14, 6271. [Google Scholar] [CrossRef] [PubMed]
- Ranjan, A.; Shetty, P.; Deshpande, R.; Biradar, A.; Khan, W.; Kulshrestha, R. A comparative surface evaluation of orthodontic mini-implants before and after en masse retraction—A SEM study. J. Orthod. Sci. 2023, 12, 15. [Google Scholar] [CrossRef] [PubMed]
- Hergel, C.A.; Acar, Y.B.; Ates, M.; Kucukkeles, N. In vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of miniscrews. Eur. Oral Res. 2019, 53, 25–31. [Google Scholar] [CrossRef] [PubMed]
- Gross, J.M.; Nascimento, G.G.; Araújo, V.C.; Bönecker, M.; Furuse, C. Mini-implants for orthodontic anchorage: Surface analysis after redrilling and sterilization—An in vitro study. J. Contemp. Dent. Pract. 2016, 17, 300–305. [Google Scholar] [CrossRef]
- Mattos, C.T.; Ruellas, A.C.; Sant’Anna, E.F. Effect of autoclaving on the fracture torque of mini-implants used for orthodontic anchorage. J. Orthod. 2011, 38, 15–20. [Google Scholar] [CrossRef]
- Alavi, S.; Asadi, F.; Raji, S.A.H.; Samie, S. Effect of steam and dry heat sterilization on the insertion and fracture torque of orthodontic miniscrews. Dent. Res. J. 2020, 17, 219–224. [Google Scholar] [CrossRef]
- Velasco-Ortega, E.; Jos, A.; Cameán, A.M.; Pato-Mourelo, J.; Segura-Egea, J.J. In vitro evaluation of cytotoxicity and genotoxicity of a commercial titanium alloy for dental implantology. Mutat. Res. 2010, 702, 17–23. [Google Scholar] [CrossRef] [PubMed]
- Vasoglou, M.; Chrysomali, E.; Zinelis, S.; Bitsanis, I.; Haralambakis, N.; Makou, M.; Eliades, G. Retrieval Analysis of Immediately Loaded Orthodontic Mini-Implants: Material and Tissue Characterization. Eur. J. Orthod. 2014, 36, 683–689. [Google Scholar] [CrossRef][Green Version]
- Kim, H.J.; Park, H.S. Long-Term Evaluation of Factors Affecting Removal Torque of Microimplants. Prog. Orthod. 2021, 22, 42. [Google Scholar] [CrossRef]
- Yun, S.D.; Choi, S.H.; Cha, J.Y.; Yu, H.S.; Kim, K.M.; Kim, J.; Hwang, C.J. Effects of recycling on the biomechanical characteristics of retrieved orthodontic miniscrews. Korean J. Orthod. 2017, 47, 238–247. [Google Scholar] [CrossRef]
- Chung, C.J.; Jung, K.Y.; Choi, Y.J.; Kim, K.H. Biomechanical characteristics and reinsertion guidelines for retrieved orthodontic miniscrews. Angle Orthod. 2014, 84, 878–884. [Google Scholar] [CrossRef] [PubMed]
- Morais, L.S.; Serra, G.G.; Muller, C.A.; Andrade, L.R.; Palermo, E.F.; Elias, C.N.; Meyers, M. Titanium alloy mini-implants for orthodontic anchorage: Immediate loading and metal ion release. Acta Biomater. 2007, 3, 331–339. [Google Scholar] [CrossRef] [PubMed]
- Büchter, A.; Wiechmann, D.; Gaertner, C.; Hendrik, M.; Vogeler, M.; Wiesmann, H.P.; Piffko, J.; Meyer, U. Load-related bone modelling at the interface of orthodontic microimplants. Clin. Oral Implant. Res. 2006, 17, 714–722. [Google Scholar] [CrossRef]
- Ma, J.; Zhang, W.; Wang, L.; Zhao, C.; Chen, W. Stability and bone response of immediately loaded microimplants in beagle dogs. Int. J. Oral Maxillofac. Implant. 2008, 23, 885–890. [Google Scholar]
- Chen, Y.; Lee, J.W.; Cho, W.H.; Kyung, H.M. Potential of self-drilling orthodontic microimplants under immediate loading. Am. J. Orthod. Dentofac. Orthop. 2010, 137, 496–502. [Google Scholar] [CrossRef]
- Sousa-Santos, P.; Sousa-Santos, S.; Oliveira, A.C.; Queirós, C.; Mendes, J.; Aroso, C.; Mendes, J.M. Evaluation of orthodontic mini-implants stability based on insertion and removal torques: An experimental study. Bioengineering 2025, 12, 549. [Google Scholar] [CrossRef]
- Pithon, M.M.; Nojima, M.G.; Nojima, L.I. In vitro evaluation of insertion and removal torques of orthodontic mini-implants. Int. J. Oral Maxillofac. Surg. 2011, 40, 80–85. [Google Scholar] [CrossRef]
- Michalczyk, K. Analysis of helical compression spring support influence on its deformation. Arch. Mech. Eng. 2020, 67, 349–362. [Google Scholar] [CrossRef]
- Teixeira, A.B.; Shimano, A.C.; Macedo, A.P.; Valente, M.L.; dos Reis, A.C. Influence of torsional strength on different types of dental implant platforms. Implant Dent. 2015, 24, 281–286. [Google Scholar] [CrossRef]
- Wilmes, B.; Panayotidis, A.; Drescher, D. Fracture resistance of orthodontic mini-implants: A biomechanical in vitro study. Eur. J. Orthod. 2011, 33, 396–401. [Google Scholar] [CrossRef] [PubMed]
- Scribante, A.; Montasser, M.A.; Radwan, E.S.; Bernardinelli, L.; Alcozer, R.; Gandini, P.; Sfondrini, M.F. Reliability of orthodontic miniscrews: Bending and maximum load of different Ti-6Al-4V titanium and stainless steel temporary anchorage devices. Materials 2018, 11, 1138. [Google Scholar] [CrossRef] [PubMed]
- Smith, A.; Hosein, Y.K.; Dunning, C.E.; Tassi, A. Fracture resistance of commonly used self-drilling orthodontic mini-implants. Angle Orthod. 2015, 85, 26–32. [Google Scholar] [CrossRef]
- Reicheneder, C.; Rottner, K.; Bokan, I.; Mai, R.; Lauer, G.; Richter, G.; Gedrange, T.; Proff, P. Mechanical loading of orthodontic miniscrews—Significance and problems: An experimental study. Biomed. Tech. 2008, 53, 242–245. [Google Scholar] [CrossRef] [PubMed]
- Al Mallah, M.R.; Salih Al Qassar, S.S.; Badran, I.B. Assessing the relationship between mini-screw implant shape and primary stability: An experimental study in orthodontics. J. Orthod. Sci. 2026, 15, 3. [Google Scholar] [CrossRef]
- Mohan, S.; Srinivasan, D.; Arumugam, E.; Devasahayam, D.; Kannan, R. Effect of Placement Angle, Diameter, Length and Bone Density on the Pull-Out Strength of Orthodontic Mini-Implants: An In Vitro Study. J. Orthod. 2022, 49, 143–150. [Google Scholar] [CrossRef]
- Sousa-Santos, C.; Sousa-Santos, S.; Mendes, J.; Coelho, C.; Aroso, C.; Sousa-Santos, P.; Mendes, J.M. The Influence of the Diameter of Orthodontic Mini-Implants on Primary Stability: Bending Tests-An In Vitro Study. Materials 2024, 17, 3149. [Google Scholar] [CrossRef]


| Specimen Number | No. of Reference Points | Min | Q1 | Median | Q3 | Max | Mean | SD | p | CI (95%) |
|---|---|---|---|---|---|---|---|---|---|---|
| Spec. 1. | 4435 | −0.1378 | −0.02278 | −0.008904 | 0.007099 | 0.1528 | −0.005774 | 0.02933 | <0.001 | (−0.006638, −0.004911) |
| Spec. 2. | 4781 | −0.1264 | −0.01928 | −0.005052 | 0.009212 | 0.1653 | −0.00333 | 0.02636 | <0.001 | (−0.004077, −0.002583) |
| Spec. 3. | 4562 | −0.09681 | −0.03597 | −0.02269 | −0.009591 | 0.1301 | −0.0224 | 0.01898 | <0.001 | (−0.02296, −0.02185) |
| Spec. 4. | 4751 | −0.1018 | −0.03665 | −0.01977 | −0.004005 | 0.0683 | −0.01983 | 0.02486 | <0.001 | (−0.02054, −0.01912) |
| Spec. 5. | 4945 | −0.3927 | −0.01648 | −0.002804 | 0.009341 | 0.1407 | −0.004424 | 0.0275 | <0.001 | (−0.005191, −0.003657) |
| Spec. 6. | 4372 | −0.08651 | −0.02259 | −0.009163 | 0.001356 | 0.05522 | −0.01077 | 0.01863 | <0.001 | (−0.01132, −0.01022) |
| Spec. 7. | 4629 | −0.05906 | −0.01588 | −0.005759 | 0.002989 | 0.08069 | −0.006809 | 0.01419 | <0.001 | (−0.007218, −0.006401) |
| Spec. 8. | 4621 | −0.05987 | −0.01982 | −0.008374 | 0.004529 | 0.05058 | −0.007739 | 0.01756 | <0.001 | (−0.008245, −0.007232) |
| Spec. 9. | 4511 | −0.06839 | −0.01327 | −0.003153 | 0.007005 | 0.05946 | −0.003354 | 0.01605 | <0.001 | (−0.003822, −0.002885) |
| Spec. 10. | 4747 | −0.2136 | −0.03484 | −0.01851 | −0.003814 | 0.1746 | −0.01787 | 0.0305 | <0.001 | (−0.01874, −0.017) |
| Spec. 12. | 4996 | −0.225 | −0.02323 | −0.00492 | 0.01186 | 0.08322 | −0.006521 | 0.02556 | <0.001 | (−0.00723, −0.005812) |
| Spec. 13. | 4392 | −0.3834 | −0.01361 | −0.001316 | 0.008818 | 0.1421 | −0.004458 | 0.02542 | <0.001 | (−0.00521, −0.003706) |
| Spec. 14. | 4909 | −0.188 | −0.02046 | −0.006909 | 0.006137 | 0.1389 | −0.006822 | 0.023 | <0.001 | (−0.007466, −0.006179) |
| Spec. 15. | 4536 | −0.1101 | −0.03381 | −0.01531 | 0.002447 | 0.06948 | −0.01547 | 0.02685 | <0.001 | (−0.01625, −0.01469) |
| Spec. 16. | 4682 | −0.2897 | −0.01788 | −0.001998 | 0.01565 | 0.1436 | −0.002334 | 0.02714 | <0.001 | (−0.003112, −0.001557) |
| Spec. 17. | 4828 | −0.09419 | −0.03645 | −0.01423 | 0.009269 | 0.1754 | −0.01105 | 0.03709 | <0.001 | (−0.01209, −0.01) |
| Spec. 18. | 4590 | −0.09305 | −0.03006 | −0.01516 | −0.001474 | 0.1474 | −0.01561 | 0.02241 | <0.001 | (−0.01626, −0.01496) |
| Spec. 19. | 4639 | −0.2509 | −0.03116 | −0.01717 | −0.001655 | 0.144 | −0.01684 | 0.02469 | <0.001 | (−0.01755, −0.01613) |
| Spec. 20. | 4639 | −0.2159 | −0.02576 | −0.01247 | 0.002193 | 0.1437 | −0.01186 | 0.02712 | <0.001 | (−0.01264, −0.01108) |
| All | 88,565 | −0.3927 | −0.02505 | −0.009668 | 0.004574 | 0.1754 | −0.01015 | 0.02575 | <0.001 | (−0.01032, −0.009981) |
| Author, Year [Ref. No] | Number of Subjects or Specimens | Type of Study | Miniscrew Brand and Dimensions | Insertion Site |
|---|---|---|---|---|
| Vasoglu et al., 2014 [23] | 70 | In vivo study (retrieval analysis) | Aarhus ø1.5 mm AbsoAnchor ø1.3 mm | Maxilla, mandible |
| Jiman et al., 2021 [16] | 15 | In vitro study | AbsoAnchor ø1.5 mm | Maxilla, mandible, buccal area |
| Ranjan et al., 2023 [17] | 50 | In vivo study (retrieval analysis) | AbsoAnchor ø1.8 mm | Maxilla, buccal area |
| Kim and Park, 2021 [24] | 327 | In vivo study (retrieval analysis) | AbsoAnchor ø1.3 mm, ø1.4 mm, ø1.5 mm | Maxilla, buccal area of posterior teeth |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Stefaniak, K.; Jedliński, M.; Tandecka, K.; Janiszewska-Olszowska, J. Geometry Alteration of Orthodontic Microimplants Following In Vivo Loading: Microanalysis on 3D Scans—A Cross-Sectional Study. Appl. Sci. 2026, 16, 5044. https://doi.org/10.3390/app16105044
Stefaniak K, Jedliński M, Tandecka K, Janiszewska-Olszowska J. Geometry Alteration of Orthodontic Microimplants Following In Vivo Loading: Microanalysis on 3D Scans—A Cross-Sectional Study. Applied Sciences. 2026; 16(10):5044. https://doi.org/10.3390/app16105044
Chicago/Turabian StyleStefaniak, Katarzyna, Maciej Jedliński, Katarzyna Tandecka, and Joanna Janiszewska-Olszowska. 2026. "Geometry Alteration of Orthodontic Microimplants Following In Vivo Loading: Microanalysis on 3D Scans—A Cross-Sectional Study" Applied Sciences 16, no. 10: 5044. https://doi.org/10.3390/app16105044
APA StyleStefaniak, K., Jedliński, M., Tandecka, K., & Janiszewska-Olszowska, J. (2026). Geometry Alteration of Orthodontic Microimplants Following In Vivo Loading: Microanalysis on 3D Scans—A Cross-Sectional Study. Applied Sciences, 16(10), 5044. https://doi.org/10.3390/app16105044