Enhanced Force Systems for Efficient Maxillary Anterior Diastema Closure Using Clear Aligners: Evaluation of Attachments and Enhanced Structural Modifications
Abstract
:1. Introduction
2. Materials and Methods
2.1. Forces/Moments Measurement Apparatus Design and Fabrication
2.2. Design Modification and Fabrication of the Experimental CA for Mesial Bodily Shift in Maxillary Left Central Incisor
2.3. Statistical Analysis
3. Results
3.1. Forces
3.1.1. Fx (Labiolingual Force, (+): Labial Force, (−): Lingual Force)
3.1.2. Fy (Mesiodistal Force, (+): Mesial Force, (−): Distal Force)
3.1.3. Fz (Vertical Force, (+): Extrusion Force, (−): Intrusion Force)
3.2. Moments
3.2.1. Mx (Mesiodistal Angulation Moment, (+): Distal Tipping Moment, (−): Mesial Tipping Moment)
3.2.2. My (Labiolingual Inclination Moment, (+): Labial Inclination Moment, (−): Lingual Inclination Moment)
3.2.3. Mz (Rotational Moment, (+): Mesial (Mesial-In) Rotational Moment, (−): Distal (Mesial-Out) Rotational Moment)
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
CA | Clear Aligners |
CAD | Computer-Aided Design |
CAM | Computer-Aided Manufacturing |
PETG | Polyethylene terephthalate glycol |
References
- Gu, J.F.; Tang, J.S.Y.; Skulski, B.; Fields, H.W.; Beck, F.M.; Firestone, A.R.; Kim, D.G.; Deguchi, T. Evaluation of Invisalign treatment effectiveness and efficiency compared with conventional fixed appliances using the Peer Assessment Rating index. Am. J. Orthod. Dentofac. Orthop. 2017, 151, 259–266. [Google Scholar]
- Gao, M.; Yan, X.; Zhao, R.; Shan, Y.; Chen, Y.; Jian, F.; Long, H.; Lai, W. Comparison of pain perception, anxiety, and impacts on oral health related quality of life between patients receiving clear aligners and fixed appliances during the initial stage of orthodontic treatment. Eur. J. Orthod. 2021, 43, 353–359. [Google Scholar] [PubMed]
- Zhao, R.; Huang, R.; Long, H.; Li, Y.; Gao, M.; Lai, W. The dynamic of the oral microbicome and oral health among patients receiving clear aligner orthodontic treatment. Oral Dis. 2020, 26, 473–483. [Google Scholar]
- Robertson, L.; Kaur, H.; Fagundes, N.C.F.; Romanyk, D.; Major, P.; Flores, M.C. Effectiveness of clear aligner therapy for orthodontic treatment: A systematic review. Orthod. Craniofac. Res. 2020, 23, 133. [Google Scholar]
- Papadimitriou, A.; Mousoulea, S.; Gkantidis, N.; Kloukos, D. Clinical effectiveness of Invisalign® orthodontic treatment: A systematic review. Prog. Orthod. 2018, 19, 37. [Google Scholar]
- Simon, M.; Keilig, L.; Schwarze, J.; Jung, B.A.; Bourauel, C. Forces and moments generated by removable thermoplastic aligners: Incisor torque, premolar derotation, and molar distalization. Am. J. Orthod. Dentofac. Orthop. 2014, 145, 728–736. [Google Scholar]
- Kurth, J.R.; Kokich, V.G. Open gingival embrasures after orthodontic treatment in adults: Prevalence and etiology. Am. J. Orthod. Dentofac. Orthop. 2001, 120, 116–123. [Google Scholar] [CrossRef]
- Fan, Y.; Lv, X.; Wang, B.; An, L. Initial forces acting on the maxillary teeth with aligner activation of canine distalization on the basis of different initial canine tips in vitro. Am. J. Orthod. Dentofac. Orthop. 2023, 163, 867–873. [Google Scholar]
- Yokoi, Y.; Arai, A.; Kawamura, J.; Uozumi, T.; Usui, Y.; Okafuji, N. Effects of Attachment of Plastic Aligner in Closing of Diastema of Maxillary Dentition by Finite Element Method. J. Healthc. Eng. 2019, 3, 1075097. [Google Scholar] [CrossRef]
- An, S.S.; Choi, Y.J.; Kim, J.Y.; Chung, C.J.; Kim, K.H. Risk factors associated with open gingival embrasures after orthodontic treatment. Angle Orthod. 2018, 88, 267–274. [Google Scholar] [CrossRef]
- Jin, X.; Tian, X.; Lee Zhi Hui, V.; Zheng, Y.; Song, J.; Han, X. The effect of enhanced structure in the posterior segment of clear aligners during anterior retraction: A three-dimensional finite element and experimental model analysis. Prog. Orthod. 2024, 25, 3. [Google Scholar] [CrossRef] [PubMed]
- Kohda, N.; Iijima, M.; Muguruma, T.; Brantley, W.A.; Ahluwalia, K.S.; Mizoguchi, I. Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances. Angle Orthod. 2013, 83, 476–483. [Google Scholar] [PubMed]
- Hahn, W.; Dathe, H.; Fialka-Fricke, J.; Fricke-Zech, S.; Zapf, A.; Kubein-Meesenburg, D.; Sadat-Khonsari, R. Influence of thermoplastic appliance thickness on the magnitude of force delivered to a maxillary central incisor during tipping. Am. J. Orthod. Dentofac. Orthop. 2009, 136, 12.e1–12.e7. [Google Scholar]
- Badawi, H.M.; Toogood, R.W.; Carey, J.P.; Heo, G.; Major, P.W. Three-dimensional orthodontic force measurements. Am. J. Orthod. Dentofac. Orthop. 2009, 136, 518–528. [Google Scholar]
- Bichu, Y.M.; Alwaf, A.; Liu, X.; Andrews, J.; Ludwig, B.; Bichu, A.Y.; Zou, B. Advances in orthodontic clear aligner materials. Bioact. Mater. 2023, 22, 384–403. [Google Scholar]
- Choi, Y.-K.; Kim, S.-H.; Park, H.R.; Kim, S.-S.; Kim, Y.-I. Analysis of the Forces and Moments in Canine Bodily Movement with Different Clear Aligners’ Extraction Space Designs. Appl. Sci. 2024, 14, 7619. [Google Scholar] [CrossRef]
- Baik, J.C.; Choi, Y.K.; Cho, Y.; Baek, Y.; Kim, S.H.; Kim, S.S.; Park, S.B.; Kim, K.B.; Kim, Y.I. Evaluation of different designs of 3D printed clear aligners on mandibular premolar extrusion using force/moment measurement devices and digital image correlation method. Korean J. Orthod. 2024, 54, 359–373. [Google Scholar] [CrossRef]
- Grassia, V.; Fiori, A.; Diodati, F.; Sayahpour, B.; Jamilian, A.; Armogida, N.G.; d’Apuzzo, F.; Nucci, L. Clear aligners: A network and bibliometric analysis of 50 pivotal articles. Dent. Med. Probl. 2025, 62, 161–171. [Google Scholar]
- Cheng, Y.; Liu, X.; Chen, X.; Li, X.; Fang, S.; Wang, W.; Ma, Y.; Jin, Z. The three-dimensional displacement tendency of teeth-depending on incisor torque compensation with clear aligners of different thicknesses in cases of extraction: A finite element study. BMC Oral Health 2022, 22, 499. [Google Scholar]
- Al-Thomali, Y.; Basha, S.; Mohamed, R.N. Pendulum and modified pendulum appliances for maxillary molar distalization in Class II malocclusion a systematic review. Acta Odontol. Scand. 2017, 75, 394–401. [Google Scholar]
- Ravera, S.; Castroflorio, T.; Garino, F.; Daher, S.; Cugliari, G.; Deregibus, A. Maxillary molar distalization with aligners in adult patients: A multicenter retrospective study. Prog. Orthod. 2016, 17, 12. [Google Scholar] [PubMed]
- Caruso, S.; Nota, A.; Ehsani, S.; Maddalone, E.; Ojima, K.; Tecco, S. Impact of molar teeth distalization with clear aligners on occlusal vertical dimension: A retrospective study. BMC Oral Health 2019, 19, 182. [Google Scholar]
- Nanda, R.S.; Tosun, Y.S. Biomechanics in Orthodontics; Princ Pract. Hanover Park Quintessence Publ. Co.: Hanover Park, IL, USA, 2010. [Google Scholar]
- Yassir, Y.A.; Nabbat, S.A.; McIntyre, G.T.; Bearn, D.R. Clinical effectiveness of clear aligner treatment compared to fixed appliance treatment: An overview of systematic reviews. Clin. Oral Investig. 2022, 26, 2353–2370. [Google Scholar] [PubMed]
- Elshazly, T.M.; Salvatori, D.; Elattar, H.; Bourauel, C.; Keilg, L. Effect of trimming line design and edge extension of orthodontic aligners on force transmission: A 3D finite element study. J. Mech. Behav. Biomed. Mater. 2023, 140, 105741. [Google Scholar]
- Cowley, D.P.; Mah, J.; O’Toole, B. The effect of gingival margin design on the retention of thermoformed aligners. J. Clin. Orthod. 2012, 46, 697. [Google Scholar]
- Elshazly, T.M.; Keilig, L.; Salvatori, D.; Chavanne, P.; Aldesoki, M.; Bourauel, C. Effect of trimming line design and edge extension of orthodontic aligners on force transmission: An in vitro study. J. Dent. 2022, 125, 104276. [Google Scholar] [CrossRef]
- Gomez, J.P.; Peña, F.M.; Martínez, V.; Giraldo, D.C.; Cardona, C.I. Initial force systems during bodily tooth movement with plastic aligners and composite attachments: A three-dimensional finite element analysis. Angle Orthod. 2015, 85, 454–460. [Google Scholar]
- Guo, R.; Lam, X.Y.; Zhang, L.; Li, W.; Lin, Y. Biomechanical analysis of miniscrew-assisted molar distalization with clear aligners: A three-dimensional finite element study. Eur. J. Orthod. 2024, 46, cjad077. [Google Scholar]
- Lu, T.Y.; Ahmad, M.A.; Hassan, W.N.W.; Hariri, F. The stress and deformation effect of novel rapid maxillary expanders by finite element analysis. J. Med. Biol. Eng. 2022, 42, e397–e403. [Google Scholar]
- Rossini, G.; Parrini, S.; Castroflorio, T.; Deregibus, A.; Debernardi, C.L. Efficacy of clear aligners in controlling orthodontic tooth movement: A systematic review. Angle Orthod. 2015, 85, 881–999. [Google Scholar] [CrossRef]
- Palone, M.; Longo, M.; Arveda, N.; Nacucchi, M.; Pascalis, F.D.; Spedicato, G.A.; Siciliani, G.; Lombardo, L. Micro-computed tomography evaluation of general trends in aligner thickness and gap width after thermoforming procedures involving six commercial clear aligners: An in vitro study. Korean J. Orthod. 2021, 51, 135–141. [Google Scholar] [PubMed]
- Bucci, R.; Rongo, R.; Levate, C.; Michelotti, A.; Barone, S.; Razionale, A.V.; D’Anto, V. Thickness of orthodontic clear aligners after thermoforming and after 10 days of intraoral exposure: A prospective clinical study. Prog. Orthod. 2019, 20, 36–43. [Google Scholar] [PubMed]
Component | Definition | Sign Convention | |
---|---|---|---|
Force (X) | Buccolingual | (+) Labial | (−) Lingual |
Force (Y) | Mesiodistal | (+) Mesial | (−) Distal |
Force (Z) | Occlusogingival | (+) Occlusal | (−) Gingival |
Moment (X) | Angulation | (+) Distal tipping | (−) Mesial tipping |
Moment (Y) | Inclination | (+) Labial | (−) Lingual |
Moment (Z) | Rotation | (+) Mesial rotation | (−) Distal rotation |
Measurements (Total 80 Sample) | Group 1 (Median, IQR) | Group 2 (Median, IQR) | Group 3 (Median, IQR) | Group 4 (Median, IQR) | p-Value |
---|---|---|---|---|---|
Fx (N) | −2.30, 0.29 a | −1.74, 0.39 b | −3.01, 1.29 a | −2.03, 1.23 ab | H = 22.813 <0.001 *** |
Fy | 0.26, 0.19 a | 0.16, 0.12 b | 0.31, 0.12 a | 0.27, 0.39 a | H = 19.708 <0.001 *** |
Fz | 0.31, 0.16 a | −0.01, 0.20 b | 0.39, 0.09 a | −0.19, 0.57 b | H = 47.129 <0.001 *** |
Mx (Nmm) | −10.70, 1.69 a | −12.43, 1.83 b | −8.98, 6.46 ac | −7.91, 4.54 c | H = 29.465 <0.001 *** |
My | 2.27, 3.88 a | −2.27, 3.55 b | 2.97, 5.21 a | 0.97, 8.01 a | H = 17.114 <0.001 *** |
Mz | −23.21, 4.20 a | −23.36, 3.87 a | −18.87, 26.17 a | −12.50, 20.24 b | H = 17.725 <0.001 *** |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Jee, M.-J.; Kim, Y.-I.; Kim, S.-H.; Kim, S.-S.; Park, S.-B.; Choi, Y.-K. Enhanced Force Systems for Efficient Maxillary Anterior Diastema Closure Using Clear Aligners: Evaluation of Attachments and Enhanced Structural Modifications. Appl. Sci. 2025, 15, 3672. https://doi.org/10.3390/app15073672
Jee M-J, Kim Y-I, Kim S-H, Kim S-S, Park S-B, Choi Y-K. Enhanced Force Systems for Efficient Maxillary Anterior Diastema Closure Using Clear Aligners: Evaluation of Attachments and Enhanced Structural Modifications. Applied Sciences. 2025; 15(7):3672. https://doi.org/10.3390/app15073672
Chicago/Turabian StyleJee, Min-Jeong, Yong-Il Kim, Sung-Hun Kim, Seong-Sik Kim, Soo-Byung Park, and Youn-Kyung Choi. 2025. "Enhanced Force Systems for Efficient Maxillary Anterior Diastema Closure Using Clear Aligners: Evaluation of Attachments and Enhanced Structural Modifications" Applied Sciences 15, no. 7: 3672. https://doi.org/10.3390/app15073672
APA StyleJee, M.-J., Kim, Y.-I., Kim, S.-H., Kim, S.-S., Park, S.-B., & Choi, Y.-K. (2025). Enhanced Force Systems for Efficient Maxillary Anterior Diastema Closure Using Clear Aligners: Evaluation of Attachments and Enhanced Structural Modifications. Applied Sciences, 15(7), 3672. https://doi.org/10.3390/app15073672