Activity Evaluation Methodology for the Disposed Medical Linear Accelerators

Appl. Sci. 2023, 13(1), 528; https://doi.org/10.3390/app13010528
by Na Hye Kwon 1,2,†, Young Jae Jang 3,4,†, Kum Bae Kim 4,5, Ji Hyun Yu 6, Jinsung Kim 1,2,7, Dong Wook Kim 1,2,* and Sang Hyoun Choi 4,5,*
Reviewer 1:
Appl. Sci. 2023, 13(1), 528; https://doi.org/10.3390/app13010528
Submission received: 22 November 2022 / Revised: 23 December 2022 / Accepted: 27 December 2022 / Published: 30 December 2022

Round 1

Reviewer 1 Report

1.    What are the challenges for the disposed medical linear accelerators?

2.    Comparing with other activity evaluation methodologies, what are the advantages and disadvantages of the activity evaluation methodology in this paper?

3.    How feasible is the activity evaluation methodology?

4.    Please add the diagrams of structures and the related materials of each components of the accelerators.

Author Response

We are deeply thanks for the reviewer's comments to improve the quality of the paper. Please see the attachment.

Author Response File: Author Response.docx

Reviewer 2 Report

A good review of the English language is needed. Pay attention to the use of possessive case.

Add more references in the introduction, as the following "Recommendations for the selection of in situ measurement techniques for radiological characterization in nuclear/radiological installations under decommissioning and dismantling processes"

I think that you nedd to justify better the choice of HPGe in your case. HPGe have a lot of good characteristics, but they are more expensive and require more maintenance with respect to other scintillator crystals

Did you checked the temperature when using the HPGe? The detector itself has no problems , but the associated electronics could have. Keep constant the envirnoment temperature colud be helpfull or in any cse a lot  of attention as to be payed to the calibration phase

See paper for all the comments

 

Comments for author File: Comments.pdf

Author Response

We are deeply thanks for the reviewer's comments to improve the quality of the paper. Please see the attachment.

Author Response File: Author Response.docx

Round 2

Reviewer 1 Report

Accept