Next Article in Journal
Class-Shared SparsePCA for Few-Shot Remote Sensing Scene Classification
Next Article in Special Issue
Evaluation of the Spatial Representativeness of In Situ SIF Observations for the Validation of Medium-Resolution Satellite SIF Products
Previous Article in Journal
Comparing Coarse-Resolution Land Surface Temperature Products over Western Australia
Previous Article in Special Issue
Absolute Radiometric Calibration of an Imaging Spectroradiometer Using a Laboratory Detector-Based Approach
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Band-Averaged Response Sensitivity Study of an Imaging Spectrometer for the CLARREO Pathfinder Mission

1
Science Systems and Applications, Inc., Lanham, MD 20706, USA
2
Sciences and Exploration Directorate, Goddard Space Flight Center, NASA, Greenbelt, MD 20771, USA
3
Lincoln Laboratory, Massachusetts Institute of Technology (MIT), Lexington, MA 02421, USA
*
Author to whom correspondence should be addressed.
Remote Sens. 2022, 14(10), 2302; https://doi.org/10.3390/rs14102302
Submission received: 28 February 2022 / Revised: 28 April 2022 / Accepted: 7 May 2022 / Published: 10 May 2022
(This article belongs to the Special Issue Hyperspectral Remote Sensing Data Calibration and Validation)

Abstract

:
Prelaunch absolute, SI-traceable radiometric calibration of satellite-based sensors is key to ensuring the utility of imaging spectrometer-based data products. The development of detector-based calibration techniques leads to the feasibility of meeting the 0.3% uncertainty level needed to provide climate quality data sets. Detector-based calibration is a method in which a well-understood and stable transfer radiometer is calibrated in a standards laboratory to SI-traceable standards, and transported to a facility calibrating a sensor of interest. The transfer radiometer provides the calibration of the source used in the radiometric calibration. A detector-based calibration approach is part of the prelaunch calibration of the CLARREO (Climate Absolute Radiance and Refractivity Observatory) Pathfinder (CPF) sensor with the Goddard Laser for Absolute Measurement of Radiance (GLAMR) system. The SI-traceability of GLAMR is through the electric watt as part of the absolute radiometric calibration of the detectors at the National Institute of Standards and Technology using the Primary Optical Watt Radiometer. The current work uses GLAMR data collected with a visible and near-infrared imaging spectrometer calibration demonstration system to develop a source/sensor modeled calibration data set as part of a sensitivity study to evaluate uncertainties from the spectral sampling and processing methods that accompany the GLAMR calibration process. The spectral “supersets” include realistic instrumental features as well as effects from the GLAMR source. The methods needed to ensure that spurious sensor and GLAMR data are excluded are described. Results are given from the sensitivity study related to GLAMR spectral sampling and signal-to-noise ratio (SNR) effects, sensor integration time, and frame averaging of the imaging spectrometer data. The study shows that the 6 nm bandwidth sensor simulation requires a 1 nm spectral sampling of the GLAMR source with a radiance level that provides an in-band peak SNR > 200 to ensure that climate quality accuracies can be achieved. The results are also used to refine the test plan for the independent calibration for the CLARREO Pathfinder sensor calibration to optimize test time while meeting the required accuracy levels.

1. Introduction

Prelaunch laboratory measurements are critical to the success of all satellite-based imaging spectroscopy missions. A key component is an absolute, SI-traceable radiometric calibration achieved by the sensor, using viewing sources for which the output is known through a well-defined traceability path to an SI standard such as the electric watt. Fusion of data from multiple sensors calibrated to SI is much simpler, and allows for comprehensive sets of long-term, consistent data and products. Reducing the absolute uncertainty in the radiometric calibration to <0.3% (k = 2) in the reflected solar part of the spectrum leads to climate quality data sets [1].
Laboratory-based, absolute radiometric calibrations in the reflected solar portion of the spectrum (350 to 2500 nm) typically rely on sources of spectral radiance based on melting-point blackbodies, lamp-illuminated spherical integrating sources, or lamp-illuminated diffuser plaques [2,3,4]. The state-of-the-art uncertainties for spectral radiance products that rely on calibrated source transfer standards are at best 3.2% (k = 2) for laboratory-based instrumentation [5] and 5.0% (k = 2) for on-orbit sensors [6]. Reductions in absolute uncertainty are achieved by producing reflectance-based products rather than spectral radiance, and state-of-the-art uncertainties have been shown to be <3.6% (k = 2) [6,7].
Clearly, significant improvements are needed to achieve <0.3% reflectance uncertainties for climate-quality data. One approach developed for the Climate Absolute Radiance and Refractivity Observatory (CLARREO) is that of the ratio radiometer. Reflectance is retrieved by taking the ratio of earth scene measurements to those of the incident solar irradiance. This approach is similar to that implemented for the on-orbit solar diffuser monitors for the Moderate Resolution Imaging Spectroradiometer [8,9] and the Visible Infrared Imaging Radiometer Suite [10]. The ratio radiometer approach is being demonstrated by the Hyperspectral Imager for Climate Science (HySICS) as part of the CLARREO Pathfinder (CPF) mission to the International Space Station (ISS). The absolute uncertainty requirement for CPF HySICS is 0.6% (k = 2) due to limitations of the frequency at which the sensor can perform solar views because of ISS operations.
The CPF requirement of 0.6% uncertainty is still a factor of five more stringent than the state of the art. Demonstrating that this requirement is met for an on-orbit sensor cannot be accomplished through typical methods, such as intercomparisons with other sensors. Thus, an additional calibration method, which follows the traditional path of absolute prelaunch calibration with a transfer to orbit, is being undertaken as part of CPF’s independent calibration. The independent calibration achieves a significant improvement in uncertainties by making use of detector-based calibration of HySICS [11]. A detector-based calibration is one in which a well-understood and stable detector package is calibrated in a standards laboratory to SI-traceable standards. The detector package is transported to the facility calibrating the sensor of interest and the package then provides the calibration of the source being used in the radiometric calibration. The approach is different from the source-based method described above, in which the source is calibrated and transported. The detector-based method for CPF Independent calibration relies on the Goddard Laser for Absolute Measurement of Radiance (GLAMR) system [12]. The SI-traceability of GLAMR is through the electric watt as part of the absolute radiometric calibration of the detectors at the National Institute of Standards and Technology (NIST) using the Primary Optical Watt Radiometer (POWR). The transfer radiometers are absolutely calibrated with reference to an irradiance source with an uncertainty <0.3% (k = 2) in the VNIR and 0.3–3% in other spectral regions. The transfer radiometers’ calibration, along with their use with an extended radiance source, are the dominant uncertainties in the absolute radiometric calibration of HySICS. The overall uncertainty for the HySICS GLAMR calibration that makes use of the GLAMR sphere source is <1.2% (k = 2) in radiance [12].
One challenge for the CPF independent calibration is ensuring that the GLAMR calibration process, coupled with an imaging spectrometer, provides the expected uncertainties, and that the sampling and processing methods do not lead to unexpected systematic errors. The independent calibration has made use of GLAMR data collected as part of the evaluation of the Solar, Lunar Absolute Reflectance Imaging Spectroradiometer (SOLARIS) that was developed as a calibration demonstration system for the original CLARREO mission [13,14]. The data were collected to understand the spectral sampling strategies and expected test time needed to ensure an accurate HySICS calibration while minimizing impact on the project’s critical path.
The current work presents the results of a sensitivity study undertaken to evaluate the appropriate GLAMR wavelength sampling strategy and radiance levels needed to ensure that the radiometric calibration uncertainty is dominated by the GLAMR source absolute uncertainty. The study relies on modeled GLAMR spectral radiance coupled with modeled absolute spectral response (ASR) for a simulated imaging spectrometer. The next section gives an overview of a GLAMR calibration approach, and a description of the GLAMR and SOLARIS data collections that are part of the modeling efforts. Section 3 provides a description of how the GLAMR and SOLARIS data are converted to the spectral “supersets” needed for the sensitivity study. The advantage of using real data is that they include both realistic instrumental features and effects from the GLAMR source as part of the evaluation. Section 3 also includes a discussion of the methods used to ensure that spurious SOLARIS and GLAMR data are not included in the modeled system. In Section 4, results are given from the sensitivity study; these results are related to GLAMR spectral sampling and signal to noise ratio (SNR) effects in the sensor being tested due to GLAMR radiance levels, sensor integration time, and frame averaging of the imaging spectrometer data. This is followed by discussions of the results and the impact on planned testing for HySICS specifically, as well as for more general imaging spectrometer testing. The key result from the effort is that a 1 nm spectral sampling is needed with a radiance level that provides an in-band peak SNR >200 to ensure that the CPF requirements are met.

2. Detector-Based Calibration

The GLAMR detector-based calibration that is the basis for the current work relies on a tunable near-monochromatic laser light source coupled with a spherical integrating source (SIS). The narrower linewidth of the laser source compared to grating-based monochromators, and the large spectral range that can be covered by the tunable laser, lead to a higher fidelity calibration. The use of detector standards as part of the process allows both a relative and absolute spectral response measurement. The use of the SIS allows for a full focal plane evaluation of the imaging spectrometer if the exit port of the SIS overfills both the entrance pupil of the sensor and the full field. Current error budget estimates for a GLAMR-based radiometric calibration show that the uncertainties are dominated by the accuracy of the detector standards and the SIS uniformity [12].
This basic approach used by GLAMR is similar to that of NIST for the Spectral Irradiance and Radiance Responsivity Calibrations using Uniform Sources (SIRCUS) facility, which is designed to lower the radiance and irradiance responsivity calibrations uncertainty down to a 0.1% level [15]. The radiance and irradiance calibrations made by SIRCUS are traceable to POWR that is the US standard for optical power. The current GLAMR systems are based on a portable version of SIRCUS that was loaned to the Goddard Space Flight Center as part of its early CLARREO efforts.
The GLAMR light source and detectors can cover wavelengths in the spectral range from 340 to 2500 nm, which is a range suitable for calibrating instruments operating in the reflected solar portion of the spectrum. The main part of the light source consists of two tunable optical parametric oscillators (OPOs). Both OPOs use lithium triborate (LBO) nonlinear crystals pumped with a green light (532 nm) from frequency doubling of a neodymium doped yttrium vanadate laser at a 80 MHz repetition rate. The fundamental signal for these OPOs covers 680–1000 nm and 1100–1500. The use of an idler extends the achievable wavelengths upwards to 2500 nm and the second harmonic signal covers the UV and visible region of 340–680 nm. Figure 1 shows a simplified schematic of the optical path used to achieve the 550 to 700 nm spectral range. The removal of the LBO2 crystal shown in Figure 1 provides the 1050 to 1400 nm spectral range. Redirection of the optical beam to a second harmonic generator, or including the idler mentioned above after mirror m2, provides the other spectral regions needed to obtain full spectral coverage.
The monochromatic, broadband, and high-intensity output from the parametric conversion in the source laser is coupled with a multi-mode fiber that is directed to the GLAMR integrating sphere as illustrated in Figure 2. The illuminated integrating sphere has a large exit port which can fill the full aperture and field of view of most sensors. A set of radiometers are used to measure the absolute radiance during the GLAMR calibration. The radiometers include a silicon-based trap detector radiometer, a single-detector indium gallium arsenide (InGaAs) radiometer, and an extended InGaAs detector coupled with an integrating sphere collector; these combine to cover the 340–2500 nm range generated by the GLAMR source. The radiometers are permanently attached to the integrating sphere and are regularly calibrated using transfer radiometers that are in turn calibrated using POWR. The absolute uncertainty level for the transfer radiometers is 0.09% when calibrated in irradiance mode [15]. Work undertaken to evaluate the effects on the sphere uniformity and angular effects leads to the final GLAMR uncertainty of 0.6% (k = 2) for the radiance calibration of a test sensor for which both the field of view and entrance pupil are overfilled [12].
A GLAMR-based radiometric calibration consists of tuning the GLAMR source across the spectral range of interest for the sensor under examination. The wavelength tuning is fully automated within each of the spectral ranges through computer control of the LBO temperature and the tuning prism stages that are shown in Figure 1. The tuning process takes approximately 30 s per wavelength, allowing the GLAMR teams to measure several hundred calibration points in a single day, and the entire spectral range of the OPO can be scanned over several days.
Sensor calibration is accomplished by aligning the sensor being tested with the SIS to ensure full illumination of the instrument’s field and entrance aperture. The collection sequence begins with the shutter shown in Figure 1 in the closed position while the GLAMR source is tuned to a desired wavelength. The shutter is opened to illuminate the sensor. The tuning process takes approximately 30 s and the system is held at a given wavelength for at least 30 s to allow the source to stabilize. As mentioned above, radiometers monitor the sphere output, and the radiometer outputs are coupled with a pair of crossed polarizers as part of the power stabilizer shown in Figure 1. The shutter remains open until a sufficient number of samples is obtained by the sensor under examination. The shutter is closed, and the process is repeated until the spectral range is covered.
GLAMR telemetry includes wavelength, SIS radiance output, shutter state, and collection time. The sensor under examination typically collects during the entire tuning process in the multispectral sensor case, or the shutter state is used to trigger the collection of a sufficient number of frames/images in the imaging spectrometer case. Signal levels, shutter state telemetry, and time of collections are all used to pair the appropriate SIS absolute radiance with the sensor data. One of the improvements that has been made since the original GLAMR configurations is the inclusion of a time synchronization package that allows the data acquisition of the system being tested to be synchronized with the clocks of the sensor’s acquisition software and GLAMR’s control software. The synchronization simplifies matching of stabilized source signals with the sensor measurements.
Processing of the detector-based calibration consists of three basic steps: (1) pre-processing of the GLAMR and test sensor data; (2) conversion of sensor data to absolute spectral response (ASR); and (3) calculation of band-averaged calibration parameters.
Pre-processing includes the conversion of raw GLAMR telemetry to time-stamped absolute radiances for the sphere monitors. Also included in this step are the dark subtraction of illuminated test data using shutter data and the normalization of sensor data based on integration time and gain settings. ASR for each detector in the test sensor is calculated from
A S R ( λ n , i , j ) = D N ( λ n , i , j ) L ( λ n )
where λn is the tuned GLAMR wavelength,   i , j specify the detector number, A S R ( λ n , i , j ) the absolute spectral response at λ n for detector (i,j), DN(λn,i,j) is the dark corrected sensor output for that detector, and L(λn) is the radiance from the GLAMR SIS at the tuned wavelength. Band-averaged spectral response for detector (i,j), R(i,j), can be calculated from numerical integration of the ASR over all N GLAMR tuned wavelengths [16].
R ( i , j ) = n = 2 N A S R ( λ n , i , j ) + A S R ( λ n 1 , i , j ) 2 ( λ n λ n 1 )
The band-averaged center wavelength for detector (i,j), λc(i,j) is calculated from
λ c ( i , j ) = n = 2 N λ n A S R ( λ n , i , j ) ( λ n λ n 1 ) n = 2 N A S R ( λ n , i , j ) ( λ n λ n 1 )
Figure 3 shows sample results from a single spectral band of a GLAMR-based calibration of a multispectral, filter-based, non-imaging radiometer [5]. The results shown include the GLAMR source radiance output as a function of wavelength as determined from the monitoring radiometer of the GLAMR sphere source (Figure 3a). Of note in the radiance curve is the sharp change in the spectral shape at 895 nm, which is due to a change in the laser configuration of the GLAMR system at that wavelength. Such a change in the spectral output of the source is not an issue for the sensor calibration since the radiance is well known from the monitoring radiometers in the sphere source.
The second graph (Figure 3b) gives the sensor output for the GLAMR source radiance and wavelengths shown. The sensor output and source radiance are used to determine the ASR as a function of wavelength according to Equation (1); these are shown in the lower graphs (Figure 3c,d). The two ASR graphs show the same data, but Figure 3a shows the full spectral range, and 3d shows a subset of that range to highlight individual GLAMR wavelength samples. One feature of the automated tuning system developed for GLAMR is that the wavelength interval is not constant over the collection. The impact, for the most part, is not noticeable in the full spectral range but can be seen in the subset spectral range. The spectral sampling effect is further shown in Figure 4, which gives the GLAMR tuned wavelength interval for collection shown in Figure 3.
The band-averaged spectral response of the sensor is calculated through numerical integration of the ASR data shown in Figure 3 using Equation (2); this process is sensitive to spectral sampling. The impact of the tuning methodology is that multiple calibrations will lead to different results simply because the wavelengths sampled from one calibration to another will change, thus sampling different portions of the ASR curve leads to variations in the resulting band-averaged response. The effect is small for well-behaved ASRs, but the calibration requirements for CPF are stringent enough that the variability in retrieved band-averaged response needs to be <<0.1% for this effect to ensure that the impact is negligible on the overall HySICS calibration uncertainty.
Using actual GLAMR calibration data sets to understand the impacts of spectral sampling effects is non-trivial because real measurements will include noise from the test sensor. The approach chosen in the current work is to develop a set of modeled GLAMR outputs at a high spectral resolution much finer than would be needed for a calibration collection. A sensor spectral response model is developed at the same high spectral resolution as the GLAMR source model. The combination of the two allows for simulations of realistic GLAMR calibration outputs, such as those shown in Figure 3, which can be used to evaluate the impact of data-set-to-data-set variability as well as impacts from the sampling interval. The GLAMR calibration model data set is described in the next section and the sensitivity studies conducted with it are presented in Section 4.

3. Spectral Superset Models

The most accurate GLAMR-based calibration would ideally be one that spectrally samples the test sensor spectral response with both very high spectral resolution and very small sampling intervals. GLAMR has the capability to accomplish both, but scanning the full spectral range of a test sensor coupled with the limits that are typically placed on sensor testing schedules makes it impractical to conduct arbitrarily fine sampling. Spectral sampling effects are further complicated by the uncertainties in individual ASR values caused by noise in the test sensor. Determining a spectral sampling resolution that reduces uncertainties while optimizing test time is one of the motivating factors for the current work. The sensitivity study shown here is an approach that helps develop a GLAMR calibration test plan for an imaging spectrometer that reduces systematic uncertainties originating from the GLAMR sampling. The resulting test plan will provide parameters such as wavelength sampling, GLAMR source radiance levels, GLAMR source stability, test sensor integration time and gain, and number of test sensor samples. An additional benefit is that a set of metrics can be developed that provide quality checks on the calibration data set as the data are being collected.
A straightforward method to assess spectral sampling effects is to collect a GLAMR calibration at a very high spectral resolution and then subsample that test data set randomly and multiple times with varying spectral sampling. A spectral resolution of 0.05 nm would provide 20 times the number of samples for the current 1 nm GLAMR sampling strategy employed for imaging spectrometers. The challenge is that collecting such a data set over the spectral range of HySICS would take more than one year of daily calibration work, and such a data set would still suffer the effects of noise from the sensor under examination.
Thus, a model-based GLAMR calibration data set is the most suitable approach to understand the impacts of spectral sampling effects on the band-averaged spectral response as computed in Equation (2). The modeling approach allows the current work to concentrate on the impacts of spectral sampling while avoiding effects from sensor variability and avoiding the schedule impacts on the GLAMR Team to collect a high-spectral-sampling data set. The term “superset” is used to denote the high spectral resolution sampling of the modeled source and ASR at 0.05 nm spectral intervals. While the emphasis of the current work is to assess the uncertainties expected for the GLAMR-based radiometric calibration of CPF HySICS, the model-based approach here has applicability to a wider range of sensors, as long as there is a reasonable understanding of the general shape of the test sensor’s relative spectral response and the SNR expected for typical GLAMR radiance levels.
The modeled supersets described here are based on a set of GLAMR calibration data collected over several months as part of studies for the SOLARIS instrument. The two key elements from those collections are the GLAMR source radiance levels for the data sets, and the SOLARIS instrument ASRs derived from the measurements at the GLAMR source wavelengths. The spectral sampling of the multiple GLAMR calibrations was 1 nm or larger from 400 to 1100 nm at the 0.05 nm sampling interval, with 20–30 SOLARIS image frames collected for each shutter open state. The advantage to using real data from SOLARIS is that the sensitivity studies will include realistic instrumental features.

3.1. GLAMR Source Superset Model

The key element of the modeled data set is an assumed truth data set for the GLAMR source radiance at 0.05 nm spectral intervals. The GLAMR source superset model has been developed at 0.05 nm spectral sampling from 400–1100 nm. Each sample consists of 20 radiance samples at each wavelength to allow the radiance variability of the GLAMR source to be included as part of the superset model. The source variability, while small, affects the retrieved sensor ASR and will be present at approximately the same levels regardless of the spectral sampling and resolution chosen.
The first step in the GLAMR source superset generation was to determine the average source radiance and variability for each wavelength of the SOLARIS GLAMR calibration data sets. GLAMR radiance levels for all wavelengths of the superset are generated based on interpolation of the average radiance at surrounding wavelengths. Twenty radiance levels were determined for each of the modeled superset spectral samples based on the standard deviation of the SOLARIS GLAMR calibration source values from the nearest wavelength to the superset wavelength. The 20 superset radiance levels were generated based on random sampling of a gaussian distribution. The mean of the gaussian distribution at a modeled superset wavelength was determined by linear interpolation of the mean radiance measured at the two nearest wavelengths from the SOLARIS GLAMR calibration dataset. The variance of the gaussian distribution is the standard deviation of the measured radiance values from the SOLARIS GLAMR wavelength nearest to the modeled superset wavelength.
Figure 5a shows the GLAMR source superset for the full spectral range of the model. Both the original and modeled data are shown, though the two are not discernible at the scale of the full range. Figure 5b shows the same data except over a limited spectral range to highlight the individual data points at each sample, as well as the original and modeled data. The data in the figure clearly demonstrate the spectral variability of the GLAMR source. Note that the data from 800 to 900 nm follow a similar shape as that shown in Figure 3b. Numerous sharp changes in the spectral shape are due to changes in the laser configuration of the GLAMR system by the operator. These changes will vary from collection to collection based on the operator, power output, and configurations of the tuning optics. The spectral changes in the source radiance should not be a dominant uncertainty source since the source output is assumed known from the radiance levels reported by the monitoring radiometers in the sphere source.
Figure 5b shows the modeled GLAMR source superset data for 11 wavelengths between 668.2 and 668.7 nm. Also shown are the original GLAMR calibration source data between 668.2 and 668.7 nm. The change in radiance over the 20 samples is evident in the figure and can be seen to be <0.1% peak to peak from the measurements. The figure shows the 20 modeled data points. The gaussian assumption for the variation of the modeled source radiances overestimates the model peak-to-peak variability on average relative to the measured values. The larger variability is not viewed as an issue for the sensitivity study since the overestimation is the more conservative situation.

3.2. Imaging Spectrometer Superset Model

As mentioned, the imaging spectrometer superset is based on the SOLARIS imaging spectrometer that was developed as part of the original CLARREO project. The sensor is an Offner-Chrisp spectrometer that currently operates with a commercially-available 2160 × 2560 pixel, silicon-based sCMOS detector array. SOLARIS is a key element of the CPF independent calibration enabling evaluation of the GLAMR approaches planned for HySICS without affecting the program’s critical path. Work with SOLARIS has included performance characterization such as component level tests of optics, detector, and grating. Characterization has included determining the relative spectral response, noise, and temperature sensitivity while performing both laboratory measurements and field measurements [17,18,19]. The same instrument characteristics that make it a suitable surrogate for independent calibration activities also makes it a suitable instrument for developing a spectrometer superset model to evaluate GLAMR calibration uncertainties.
The imaging spectrometer superset consists of modeled ASR values at the same 0.05 nm spacing as the GLAMR source superset described in the previous section. The superset ASRs are created for each SOLARIS instrument center wavelength at the center of the instrument’s field of view. The modeled ASRs are based on the ASR values from the multiple SOLARIS GLAMR calibration data sets. The use of the measured values provides a model with more realistic instrument line shapes. The challenge in developing the model is maintaining the realistic nature of the spectral sensitivity simulations while excluding SOLARIS noise artifacts from the GLAMR calibrations that would affect the conclusions derived from the sensitivity study.
A screening process was developed to remove anomalous data points from the measured ASR results using a median absolute deviation (MAD) approach. The process was applied to the SOLARIS output from each tuned GLAMR source wavelength for all the SOLARIS calibration data sets. Each SOLARIS frame at a GLAMR wavelength consists of a dark-corrected output for each spectral detector (SOLARIS band) and spatial detector (field of view angle). The peak signal is determined for each frame and used to normalize the signal for each detector. The average normalized signal, the standard deviation of that average, and the relative error (defined as the ratio of the standard deviation to the normalized average) are determined and reported for the 20 samples for each detector. The relative error is essentially an inverse SNR for each detector for each GLAMR tuned wavelength.
Outlier filtering is performed on 100 spectral detector groupings at a time centered on the peak average signal detector ±50 spectral detectors. The median absolute deviation for the relative error is found at each field of view (spatial detector row) using
M A D = b × m e d i a n   ( | x M e a n ( x ) | )
where x represents the relative error of the pixel, and b is a distribution-dependent constant which is equal to b = 1.4826 for a relative error with a normal distribution. Data points considered to be outliers are those with relative error >3 MAD from the median value and these are removed from ASR determination. An entire 20-sample data set is removed if the number of outlier pixels is >1% in this spectral region subset. Detectors that regularly fail the MAD filtering are marked and removed from all ASR calculations.
The end result is a set of SOLARIS data for which poor SNR data points within 50 spectral samples of a SOLARIS band center are removed. A smoothing spline is applied to the derived ASR values from the outlier-removed data set, and the spline fit is used to determine the ASR for each wavelength of the superset. The result is a set of modeled ASRs at 0.05 nm sampling for each of the 1300 SOLARIS spectral bands; these become the reference used for all comparisons in the sensitivity study. The modeled ASR is the imaging spectrometer superset. A set of reference band-averaged spectral response values are computed using Equation (2) and based on the full imaging spectrometer superset.

4. Sensitivity Studies

The motivating factor for a sensitivity study of spectral sampling of the GLAMR-based calibration is the stringent absolute uncertainties of CPF. The 0.3% (k = 1) absolute uncertainty requirement means that effects from typically ignored aspects of the radiometric calibration process can now play an important role. An additional factor is that radiometric testing for remote sensing flight projects is time constrained due to budget and overall project schedules. Such constraints are even more important for smaller-budget missions such as CPF.
The goal of the CPF independent calibration team is to characterize as large a spectral range of HySICS as possible, while still achieving an absolute radiometric uncertainty for the GLAMR calibration that is dominated by the SI-traceability of the GLAMR system. The outcome is that effects such as numerical integration techniques must cause minimal uncertainties while spectral sampling is as coarse as possible to limit schedule impacts. The supersets are used here to evaluate the impacts of GLAMR spectral sampling, source variability, and sensor noise on the accuracy of absolute radiometric calibration. The sensitivity results here need to have <0.1% (k = 1) effects to ensure that the absolute uncertainty of the GLAMR source that is traceable to SI is the dominant term in the error budget.
The result of a given sensitivity analysis is an uncertainty based on the standard deviation of a set of calibration simulations. A single calibration simulation is defined as a simulated ASR retrieval for a full range of GLAMR wavelengths. The simulation relies on spectral sampling simulating the GLAMR wavelength variability seen in Figure 3 and Figure 4 based on a starting wavelength and sampling interval, Δλ. Subsequent wavelengths in the list are generated by incrementing from the initial wavelength by Δλ ± ε, where ε is a randomly generated value based on a uniform distribution with values ±0.1 nm. The wavelength list is refined to the nearest GLAMR source superset value.
The GLAMR source radiance for the calibration simulation is determined for each wavelength by randomly selecting from the 20 GLAMR source superset radiance values at each wavelength. The source radiance for the calibration simulation is combined with the imaging spectrometer superset ASR for each simulated detector to create an imaging spectrometer frame image for all detectors in the simulation. Noise can be added to the frame values at this point as needed. Repeating the process for multiple frames and then further repeating for each simulated wavelength creates a set of simulated imaging spectrometer frames corresponding to a simulated full calibration collection by the imaging spectrometer using the GLAMR system. Processing these data as described in Section 2 provides a retrieved set of ASRs, center wavelengths, and band-averaged spectral responses.

4.1. Spectral Sampling

The sensitivity of band-averaged response to spectral sampling increment was assessed using the above approach to create the GLAMR calibration simulations based on the superset data. Impacts from the wavelength sampling increment and the uneven spectral sampling are important features to understand the impact on uncertainties, due to the trade between spectral increments and test time for a GLAMR-based radiometric calibration. The issue is complicated by the fact that there is not a one-to-one relationship between the sample interval and the length of time to collect, due to the automated tuning approach that is used for GLAMR. For instance, increasing the sampling from 1.0 nm to 0.5 nm intervals triples the amount of time taken to collect the data, due to the need for manual inputs by the operator when tuning to small increments. Sensor noise in this case is assumed to be zero to isolate the impact of spectral sampling only.
The gain derived from the highest spectral resolution (Δλ = 0.05 nm) of the superset is considered the ‘true’ ASR for the instrument. This baseline gain of each band can be calculated as described in Section 2. Band-averaged spectral response values were computed for several sampling intervals for multiple cases at each interval. The changes in response for the multiple cases at a given spectral interval were found to be negligible for Δλ < 2.5 nm, which is expected since the modeled ASR curves are reasonably smooth, and the bandwidths are approximately 5 nm.
Figure 6 shows the band responsivity ratios derived for a single calibration simulation for values of Δλ = 1.0, 1.5, and 2.0 nm. The data for Δλ = 0.05 nm is also provided as the reference. Larger spectral sampling intervals show greater variability in responsivity ratio across all bands as expected. The ratio of the band responsivity for the case of Δλ = 2.0 nm shows that the error from integration is >0.3% for a large number of bands but is <1% in all cases. The total uncertainty limit for the CPF independent calibration is 0.3% for a GLAMR-based calibration, thus the 2 nm case would lead to unacceptable uncertainties. Lowering the wavelength increment to Δλ = 1.5 nm results in errors less than 0.2% for nearly all bands. A spectral sampling of Δλ = 1.0 nm results in errors much less than 0.1% for all bands, which is sufficient to ensure that the spectral sampling would not be a major error source in the CPF independent calibration 0.3% absolute uncertainty.
It is apparent in Figure 6 that the uncertainty changes in a non-linear fashion with changing spectral sampling. The variability in the responsivity ratio is more than halved when going from 2 nm sampling to 1.5 nm, and likewise going from 1.5 nm to 1 nm. There is minimal change in uncertainty when going from 1 to 0.05 nm. The effect is the interplay between the spectral sampling of the GLAMR-based calibration and the spectral bandwidth of the imaging spectrometer being tested. Larger spectral sampling intervals lead to an inadequate retrieval of the instrument line shape that goes into the numerical integration in Equation (2), leading to large uncertainties. The reduction in uncertainties becomes limited once the line shape is adequately sampled.
Work related to other projects in addition to CPF has started to evaluate this effect in a more quantitative manner with additional sensitivity runs with varying bandwidths, band shapes, and spectral intervals. The results of that work are not critical to the current work since, as mentioned above, the goal here is to determine the coarsest level of sampling that would provide the accuracy needed for the HySICS calibration. There is also the added factor that using sampling intervals that are the same as those used in past GLAMR calibrations simplifies the test procedure development for HySICS testing. The spectral intervals shown in Figure 6 all correspond to intervals already used in previous GLAMR calibrations for various projects. The fact that the 1 nm sampling case matches previously used testing procedures, provides satisfactory levels of uncertainties, and will fit within the HySICS testing schedule, meant that it was not necessary to explore fully the entire spectral sampling trade space.

4.2. Sensor Noise

The sensitivity to sensor noise of retrieved band-averaged spectral response was investigated using two noise models. The first uses the standard deviation of the imaging spectrometer signal multiplied by a random number in the range of −1 to 1 following a uniform probability distribution:
N 1 ( i , j ,   λ n ) = D N I S ( i , j ,   λ n ) σ I S ( i , j ,   λ n ) × r a n d ( 1 , 1 )
where N1(i,j,λn) is the noise from model 1 for imaging spectrometer detector (i,j) at wavelength λn; DNIS is the calibration simulation signal from the superset modeling; and σIS is the standard deviation from the imaging spectrometer superset for that detector and wavelength. The second model, N2, is generated from a term proportional to the inverse of the maximum signal to noise ratio (SNR) for each ASR curve and a second term related to the noise floor of the modeled sensor. Both terms are also modified by a uniform random number:
N 2 ( i , j ,   λ n ) = D N I S ( i , j ,   λ n ) r a n d ( 0.5 , 0.5 ) S N R ( i , j ,   λ n ) m a x + N f l o o r × r a n d ( 0 , 1 ) × D N I S ( i , j ,   λ m a x )
where Nfloor is the noise floor term set here to a value of 10−6; SNR(i,j,λn)max is the maximum SNR at a given wavelength and detector that is set to 200 for the work here; and DNIS(i,j,λmax) is the imaging spectrometer superset model signal for the wavelength that has the maximum SNR. The choices for the limits on the random number generators in Equation (6) are to allow variability of the signal to noise about the average signal, while ensuring that signals remain above zero for cases where the noise floor dominates.
The noise models were applied to the calibration simulations and the retrieved band-averaged spectral responses were compared with those from the no-noise case. Spectral sampling intervals of 1.0 and 1.5 nm were examined. The scenario included 30 imaging spectrometer frames at each source wavelength for the calibration simulation. Figure 7 shows the results for both cases and noise models. The key feature to note in the graphs is that the inclusion of the two sensor noise models does not significantly impact the retrieved response for the case where SNR > 200 for the spectral detectors most closely matching the tuned GLAMR wavelength. The combination of 1 nm sampling coupled with the 30 frames of data collected for each tuned wavelength minimizes errors to the point that the noise floor for the second noise model dominates. That effect is larger for larger radiance levels and this is the cause of the increase in the ratio at 700 nm.

4.3. Sensor Sampling

The last consideration discussed here is the impact that the number of sensor samples or frames has on retrieval accuracy. Increasing the number of imaging spectrometer data frames while viewing the GLAMR source is similar to increasing detector integration time but with the advantage of providing additional information on measurement repeatability. The end result of multiple frames will be an improvement in SNR for systems with random signal noise. The trade is once again collection time, but an additional consideration for imaging spectrometer applications is the volume of data that will be collected for larger numbers of frames. Ultimately, it becomes an optimization problem balancing between the data collection burden and the required quality of the uncertainty.
Multiple spectral sampling cases related to frame averaging as well as both noise models were examined. The results for the 1 nm sampling case are as expected, in that the overall uncertainty for all scenarios of noise models and sensor sampling led to uncertainties well below the threshold needed for the HySICS testing. Likewise, the results at larger spectral sampling intervals led to unacceptably high uncertainties even in the case where 50 frames of data were simulated.
Figure 8 shows the results of the sensor sampling study for the 1.5 nm spectral sampling case for the N2 noise model. The choice of noise model is made to avoid overly weighting the effects due to noise in the out-of-band region. The 1.5 nm spectral sampling case is the same as that shown in Figure 7 above and helps to highlight how including multiple frames can help to reduce uncertainties below the 0.1% uncertainty for GLAMR sampling as part of the CPF independent calibration. Three frame-averaging cases, 1, 10, and 50 frames, are shown in Figure 8. The unsurprising result for N = 1 shows that collecting a single frame of data will exceed 0.5% errors for a large number of bands. The N = 1 case corresponds to the similar 1.5 nm sampling case shown in Figure 6, except that now there is a noise factor in the data set that leads to even larger uncertainties. Reasonable uncertainties generally less than 0.3% occur for N = 10, while uncertainties are <0.2% for N = 50 with a large fraction of the uncertainties being <0.1%.

5. Discussion

The primary motivation for the sensitivity studies shown here has been prompted by the absolute uncertainty needed for the GLAMR-based calibration of the HySICS sensor as part of the CPF independent calibration. Sensitivity tests using the spectral supersets described in Section 3 demonstrate that selection of the GLAMR spectral sampling increment is a critical factor in band responsivity uncertainty. The sensor SNR also plays an important role, due to sensor noise, numbers of frames of data, or GLAMR source radiance level. A spectral sampling of 1 nm is shown to be necessary for cases where the SNR = 200 for the spectral detectors matching the tuned wavelength of the GLAMR source. Such an SNR value ensures that there is sufficient signal to evaluate the shoulders of the ASR of the sensor while also providing reasonable signal levels for the out-of-band region. A sampling interval of 1.5 nm is also shown to be suitable but is more sensitive to noise because of the poor sampling of the ASR shape coupled with the variability of the sensor output.
The expected behavior of the HySICS instrument is similar to that modeled in the imaging spectrometer superset, and the results shown here led to a test plan relying on a spectral interval of 1.0 nm and a collection of 30 frames at each GLAMR wavelength. Radiance levels will be set to ensure the highest SNR in each band without saturation for the standard operating 15 Hz frame rate of the sensor. The importance of the results shown here is that a 1.5 nm sampling would be sufficient to achieve the required accuracy given HySICS’s 6 nm nominal bandpass. Fortunately, the allotted test time can tolerate the 1.0 nm sampling. The result will be both improved accuracy for the HySICS calibration, and an improved understanding of the detector-to-detector relative calibration and isolation of the dominant noise sources.
The results can be extended to more general cases as well. The 1 nm sampling that was found to be suitable for the imaging spectrometer superset here is partially driven by the bandwidth of the imaging spectrometer bands. Sensors with larger bandwidths than the 5 nm case used here for the imaging spectrometer superset can tolerate larger sampling intervals, so long as there are >4 spectral samples with sufficient SNR covering the spectral range of the full width at half maximum of the ASR shape. A smaller spectral sampling interval would not only provide better absolute uncertainty but would also provide better band-to-band calibration, commonly referred to as flat-fielding. Additionally, the necessary parameters for sensors with narrower bandpasses and sharply peaked ASRs can be modeled. The imaging spectrometer superset for such a case would allow for the determination of sampling strategies to mitigate the non-uniform wavelength grid on the instrument profile retrieval. Likewise, optimized sampling strategies can be developed that would use varying wavelength intervals to sample the spectral shape at a denser sampling in sections that have sharper instrument features or are tailored to portions of the spectrum for which the instrument profiles might have different bandwidths.

6. Conclusions

Instrument modeling for imaging spectrometry as shown here is not unique [20], but the importance of such models increases as uncertainty requirements continue to become more stringent. A key lesson learned from the CLARREO and subsequent CLARREO Pathfinder projects has been that climate quality accuracies lead to the need to understand the impacts of the prelaunch radiometric calibration test equipment on sensor calibration, including effects from the processing approaches applied to the test data. It is well known that radiometric calibration of a flight sensor takes place later in the test program, and, as such, is often subject to changing schedule constraints. The use of the approach shown here allows the test team to assess the accuracy impacts of possible shifts in schedule and reduced testing.
The use of a detector-based calibration approach coupled with a large aperture SIS offers numerous advantages for the radiometric calibration of imaging spectrometers. The detector-based calibration using the GLAMR system is reasonably efficient when coupled with an imaging spectrometer, since all spatial and spectral detectors can be calibrated with one scan through the spectrum. The work shown here gives confidence that limiting features of the GLAMR system related to spectral sampling and source radiance variations with wavelength can be overcome. The results obtained with the current spectral superset model show that CPF independent calibration uncertainties can be achieved if the in-band sensor SNR exceeds 200 through appropriate use of GLAMR source radiance levels, frame averaging, and sensor integration time or frame rate. Sampling the sensor ASR at 1 nm spectral intervals will ensure that the dominant uncertainty is the uncertainty of the SI-traceable radiance exiting the GLAMR SIS.
Future work will include extending the spectral superset model to the shortwave infrared portion of the spectrum. Modeling the GLAMR source is straightforward in this case, since data from past calibrations of multispectral sensors can be used. Plans are also underway to include additional imaging spectrometer supersets based on other sensor designs to look for unexpected effects in the GLAMR calibration approach. Sensitivity studies with the current model will continue to refine the test plan for CPF independent calibration of HySICS, and will include updated information on the HySICS sensor based on the prelaunch characterization data. The results are expected to confirm that a 0.3% absolute uncertainty is feasible, and give greater confidence that climate-quality, top-of-atmosphere reflectance will be achieved by the CPF mission.

Author Contributions

Conceptualization and methodology, F.S. and K.T.; software and formal analysis, F.S., K.T. and R.L.; validation, Z.W.; writing—original draft preparation, F.S., K.T. and B.N.W.; writing—review and editing, K.T., B.N.W., R.L. and Z.W.; visualization, F.S.; supervision, project administration, and funding acquisition, K.T. All authors have read and agreed to the published version of the manuscript.

Funding

The work shown here is supported by funds provided by the CLARREO Pathfinder project.

Data Availability Statement

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Acknowledgments

The authors acknowledge the efforts and dedication of the NASA/GSFC GLAMR team for the development and operation of the GLAMR system and their support in the collection of the data used in this manuscript.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Wielicki, B.A.; Young, D.F.; Mlynczak, M.G.; Thome, K.J.; Leroy, S.; Corliss, J. Achieving climate change absolute accuracy in orbit. Bull. Amer. Meteor. Soc. 2013, 94, 1519–1539. [Google Scholar] [CrossRef] [Green Version]
  2. Ono, A.; Sakuma, F.; Arai, K.; Yamaguchi, Y.; Fujisada, H.; Slater, P.N.; Thome, K.J.; Palluconi, F.; Kieffer, H. Preflight and in-flight calibration plan for ASTER. J. Atmos. Ocean. Technol. 1996, 13, 321–335. [Google Scholar] [CrossRef] [Green Version]
  3. Knight, E.J.; Kvaran, G. Landsat-8 Operational Land Imager Design, Characterization and Performance. Remote Sens. 2014, 6, 10286–10305. [Google Scholar] [CrossRef] [Green Version]
  4. Biggar, S. Calibration of a visible and near-infrared portable transfer radiometer. Metrologia 1998, 35, 701–706. [Google Scholar] [CrossRef]
  5. Thome, K.; Czapla-Myers, J.; Wenny, B.; Anderson, N. Calibration and use of an ultra-portable field transfer radiometer for automated vicarious calibration. In Earth Observing Systems XXII; International Society for Optics and Photonics: Washington, DC, USA, 2017; p. 104020L. [Google Scholar] [CrossRef] [Green Version]
  6. Markham, B.; Barsi, J.; Kvaran, G.; Ong, L.; Kaita, E.; Biggar, S.; Czapla-Myers, J.; Mishra, N.; Helder, D. Landsat-8 Operational Land Imager Radiometric Calibration and Stability. Remote Sens. 2014, 6, 12275–12308. [Google Scholar] [CrossRef] [Green Version]
  7. Xiong, X.; Angal, A.; Barnes, W.L.; Chen, H.; Chiang, V.; Geng, X.; Li, Y.; Twedt, K.; Wang, Z.; Wilson, T.; et al. Updates of Moderate Resolution Imaging Spectroradiometer on-orbit calibration uncertainty assessments. J. Appl. Rem. Sens. 2018, 12, 34001. [Google Scholar] [CrossRef]
  8. Slater, P.N.; Palmer, J.M. Solar-diffuser panel and ratioing radiometer approach to satellite sensor on-board calibration. In Calibration of Passive Remote Observing Optical and Microwave Instrumentation; International Society for Optics and Photonics: Washington, DC, USA, 1991. [Google Scholar] [CrossRef]
  9. Xiong, X.; Chiang, K.; Esposito, J.; Guenther, B.; Barnes, W. MODIS on-orbit calibration and characterization. Metrologia 2003, 40, 89–92. [Google Scholar] [CrossRef]
  10. Fulbright, J.; Lei, N.; Efremova, B.; Xiong, X. Suomi-NPP VIIRS Solar Diffuser Stability Monitor Performance. IEEE Trans. Geosci. Remote Sens. 2016, 54, 631–639. [Google Scholar] [CrossRef]
  11. Aytac, Y.; Thome, K.J.; Wenny, B.N.; Angal, A.; Shuman, T.M.; McAndrew, B. Detector based calibration of a portable imaging spectrometer for CLARREO Pathfinder Mission. In Earth Observing Systems XXIV; International Society for Optics and Photonics: Washington, DC, USA, 2019; p. 111270A. [Google Scholar] [CrossRef] [Green Version]
  12. Thome, K.; Aytac, Y. Independent calibration approach for the CLARREO Pathfinder mission. In Imaging Spectrometry XXIII: Applications, Sensors, and Processing; International Society for Optics and Photonics: Washington, DC, USA, 2019; p. 111300B. [Google Scholar] [CrossRef]
  13. Thompson, P.L. and Hill, P.C. Conceptual optical design and system engineering of the CLARREO/RS (reflected solar) instrument suite. In Imaging Spectrometry XVII; International Society for Optics and Photonics: Washington, DC, USA, 2012; p. 85150N. [Google Scholar] [CrossRef]
  14. Thome, K.; McCorkel, J.; Hair, J.; McAndrew, B.; Daw, A.; Jennings, D.; Rabin, D. Test plan for a calibration demonstration system for the reflected solar instrument for the climate absolute radiance and refractivity observatory. In Remote Sensing System Engineering IV; International Society for Optics and Photonics: Washington, DC, USA, 2012; p. 851602. [Google Scholar] [CrossRef] [Green Version]
  15. Brown, S.W.; Eppeldauer, G.P.; Lykke, K.R. Facility for spectral irradiance and radiance responsivity calibrations using uniform sources. Appl. Opt. 2006, 45, 8218–8237. [Google Scholar] [CrossRef] [PubMed]
  16. Barnes, R.A.; Brown, S.W.; Lykke, K.R.; Guenther, B.; Butler, J.J.; Schwarting, T.; Turpie, K.; Moyer, D.; DeLuccia, F.; Moeller, C. Comparison of two methodologies for calibrating satellite instruments in the visible and near-infrared. Appl. Opt. 2015, 54, 10376–10396. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  17. Angal, A.; McCorkel, J.; Thome, K. Evaluation of GLAMR-based calibration for SI-traceable field reflectance retrievals. In Earth Observing Systems XXI; International Society for Optics and Photonics: Washington, DC, USA, 2016; p. 99721U. [Google Scholar] [CrossRef]
  18. Thome, K.J.; McCorkel, J.; McAndrew, B. Demonstrating the error budget for the Climate Absolute Radiance and Refractivity Observatory through solar irradiance measurements. In Earth Observing Systems XX; International Society for Optics and Photonics: Washington, DC, USA, 2015; p. 96071C. [Google Scholar] [CrossRef]
  19. Wang, Z.; Thome, K.; Lockwood, R.; Wenny, B.N. Absolute radiometric calibration of an imaging spectroradiometer using a laboratory detector-based approach. Remote Sens. 2022, 14, 2245. [Google Scholar] [CrossRef]
  20. Goodenough, A.A.; SBrown, S.D. DIRSIG5: Next-Generation Remote Sensing Data and Image Simulation Framework. IEEE J. Sel. Top. Appl. Earth Obs. Remote Sens. 2017, 10, 4818–4833. [Google Scholar] [CrossRef]
Figure 1. Simplified schematic of the GLAMR optical path used to achieve the 550 to 700 nm spectral range. Mirror elements in optical paths are labeled m1 to m8. Power stabilization is achieved through an active feedback detector coupled with a set of cross polarizers.
Figure 1. Simplified schematic of the GLAMR optical path used to achieve the 550 to 700 nm spectral range. Mirror elements in optical paths are labeled m1 to m8. Power stabilization is achieved through an active feedback detector coupled with a set of cross polarizers.
Remotesensing 14 02302 g001
Figure 2. Diagram illustrating the test configuration for a GLAMR-based calibration using the spherical integrating source to illuminate the full field and aperture of the sensor under examination. GLAMR optical bench shown is that appearing in Figure 1.
Figure 2. Diagram illustrating the test configuration for a GLAMR-based calibration using the spherical integrating source to illuminate the full field and aperture of the sensor under examination. GLAMR optical bench shown is that appearing in Figure 1.
Remotesensing 14 02302 g002
Figure 3. Sample GLAMR calibration data from a single band of a multispectral radiometer [5]. The upper graphs show the GLAMR source radiance (a) and the radiometer’s output signal (b) for those source radiances. The lower graphs show the retrieved ASR over the full range of the GLAMR radiance (c), as well as a smaller spectral range (d), to show the variable wavelength sampling of the GLAMR system.
Figure 3. Sample GLAMR calibration data from a single band of a multispectral radiometer [5]. The upper graphs show the GLAMR source radiance (a) and the radiometer’s output signal (b) for those source radiances. The lower graphs show the retrieved ASR over the full range of the GLAMR radiance (c), as well as a smaller spectral range (d), to show the variable wavelength sampling of the GLAMR system.
Remotesensing 14 02302 g003
Figure 4. Spectral interval for the GLAMR-tuned wavelengths of the data shown in Figure 3. Small wavelength intervals < 0.2 nm are the result of the GLAMR operator repeating measurements for cases with intervals > 0.7 nm.
Figure 4. Spectral interval for the GLAMR-tuned wavelengths of the data shown in Figure 3. Small wavelength intervals < 0.2 nm are the result of the GLAMR operator repeating measurements for cases with intervals > 0.7 nm.
Remotesensing 14 02302 g004
Figure 5. Graph showing the measured GLAMR source radiance values from the SOLARIS calibration work, and the modeled GLAMR source superset values as a function of wavelength. The data for the full spectral range are shown in (a). The spectral variation in radiance is, as detailed in the text, the result of the tunable laser methodology of GLAMR. (b) shows a narrower 0.5 nm spectral range that helps to highlight the measured versus modeled data. The first (668.2 nm) and last (668.7 nm) groups of 20 samples are the measured values from the original SOLARIS calibration data collection. All other values are the 20 GLAMR superset model radiance values for each superset wavelength. The twenty modeled values at a given wavelength are obtained from interpolation of the two measured mean values and a gaussian-based random sampling using the nearest-wavelength standard deviation of those measured radiance values.
Figure 5. Graph showing the measured GLAMR source radiance values from the SOLARIS calibration work, and the modeled GLAMR source superset values as a function of wavelength. The data for the full spectral range are shown in (a). The spectral variation in radiance is, as detailed in the text, the result of the tunable laser methodology of GLAMR. (b) shows a narrower 0.5 nm spectral range that helps to highlight the measured versus modeled data. The first (668.2 nm) and last (668.7 nm) groups of 20 samples are the measured values from the original SOLARIS calibration data collection. All other values are the 20 GLAMR superset model radiance values for each superset wavelength. The twenty modeled values at a given wavelength are obtained from interpolation of the two measured mean values and a gaussian-based random sampling using the nearest-wavelength standard deviation of those measured radiance values.
Remotesensing 14 02302 g005
Figure 6. Ratios of band responsivity for various wavelength increments for the case of no signal noise. Three sampling intervals are shown in addition to a reference data set at ∆λ = 0.05 nm.
Figure 6. Ratios of band responsivity for various wavelength increments for the case of no signal noise. Three sampling intervals are shown in addition to a reference data set at ∆λ = 0.05 nm.
Remotesensing 14 02302 g006
Figure 7. Results from superset-based calibration simulations including two sensor noise models, sampling intervals ∆λ = 1.5 nm (a) and ∆λ = 1.0 nm (b), and 30 simulated frames for each source wavelength. Results indicate that sampling at 1 nm intervals with an in-band SNR = 200 ensures that the uncertainty is dominated by the noise floor in the out-of-band and sensitivity to sampling interval is negligible.
Figure 7. Results from superset-based calibration simulations including two sensor noise models, sampling intervals ∆λ = 1.5 nm (a) and ∆λ = 1.0 nm (b), and 30 simulated frames for each source wavelength. Results indicate that sampling at 1 nm intervals with an in-band SNR = 200 ensures that the uncertainty is dominated by the noise floor in the out-of-band and sensitivity to sampling interval is negligible.
Remotesensing 14 02302 g007aRemotesensing 14 02302 g007b
Figure 8. Effect of responsivity sensitivity to frame averaging for wavelength sampling Δλ = 1.5 nm. Results show that 50 frames are necessary to achieve the similar level of uncertainties as found for the 1.5 nm sampling interval shown in Figure 6.
Figure 8. Effect of responsivity sensitivity to frame averaging for wavelength sampling Δλ = 1.5 nm. Results show that 50 frames are necessary to achieve the similar level of uncertainties as found for the 1.5 nm sampling interval shown in Figure 6.
Remotesensing 14 02302 g008
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Salehi, F.; Thome, K.; Wenny, B.N.; Lockwood, R.; Wang, Z. Band-Averaged Response Sensitivity Study of an Imaging Spectrometer for the CLARREO Pathfinder Mission. Remote Sens. 2022, 14, 2302. https://doi.org/10.3390/rs14102302

AMA Style

Salehi F, Thome K, Wenny BN, Lockwood R, Wang Z. Band-Averaged Response Sensitivity Study of an Imaging Spectrometer for the CLARREO Pathfinder Mission. Remote Sensing. 2022; 14(10):2302. https://doi.org/10.3390/rs14102302

Chicago/Turabian Style

Salehi, Fatholah, Kurtis Thome, Brian N. Wenny, Ronald Lockwood, and Zhipeng Wang. 2022. "Band-Averaged Response Sensitivity Study of an Imaging Spectrometer for the CLARREO Pathfinder Mission" Remote Sensing 14, no. 10: 2302. https://doi.org/10.3390/rs14102302

APA Style

Salehi, F., Thome, K., Wenny, B. N., Lockwood, R., & Wang, Z. (2022). Band-Averaged Response Sensitivity Study of an Imaging Spectrometer for the CLARREO Pathfinder Mission. Remote Sensing, 14(10), 2302. https://doi.org/10.3390/rs14102302

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop