Next Article in Journal
Validation of a UPLC-MS/MS Method for Quantifying Intracellular Olaparib Levels in Resistant Ovarian Cancer Cells
Previous Article in Journal
Biological Activities of Stachys rupestris, Development of S. rupestris Extract-Loaded Alginate Films as Wound Dressing
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Drug Delivery Device Design and Compatibility with Nitrogen Dioxide Gas Sterilization

1
Noxilizer, Inc., 1334 Ashton Road, Suite E, Hanover, MD 21076, USA
2
Department of Materials Science and Engineering, University of Maryland, College Park, MD 20742, USA
*
Authors to whom correspondence should be addressed.
Pharmaceuticals 2025, 18(12), 1869; https://doi.org/10.3390/ph18121869
Submission received: 23 October 2025 / Revised: 21 November 2025 / Accepted: 22 November 2025 / Published: 8 December 2025
(This article belongs to the Section Pharmaceutical Technology)

Abstract

Polymeric materials have become important components in prefilled syringes, drug delivery systems, and advanced medical devices. Background/Objectives: Nitrogen dioxide gas is used for the terminal sterilization of drug delivery systems. For the implementation of sterilization methods, compatibility with materials must be demonstrated such that the materials maintain product requirements and specifications after sterilization and at the time of use (i.e., product shelf life). Methods: Commonly used polymers were selected based on their chemical structures to provide insight into the nature of reactions that occur at the temperature and NO2 concentration levels used in the sterilization process. After exposure to the NO2 process, materials were evaluated for chemical, mechanical, and biocompatibility properties. Results: In this paper, we demonstrated the compatibility of polymers comprising carbonyl, unsaturated ester, and ketone groups which have been used in medical devices sterilized with NO2. No significant chemical or physical changes were observed upon the treatment of Amorphous Polyester, Polysulfone (PSU), Polycarbonate (PC), PolyEtherEtherKetone (PEEK), PolyArylEtherKetone (PAEK), and Polypropylene (PP) with NO2 at a sterilization temperature of 20 °C. At this relatively low sterilization temperature, the reactions of NO2 with the polymer do not typically occur because the activation energies of these reactions require much higher temperatures. Conclusions: Not all materials will be compatible with NO2 sterilization, and even with the established data, many devices will need to have their polymers evaluated for compatibility before moving to NO2 sterilization. These results will provide guidance to device designers selecting materials for new drug delivery devices and to regulators that review the safety and efficacy of these devices.

Graphical Abstract

1. Introduction

Biologic therapeutics, including monoclonal antibodies, recombinant proteins, and cell- or gene-based therapeutics, are often delivered via drug delivery devices such as prefilled syringes and autoinjectors [1]. To meet the requirements for patient safety, some drug delivery devices require terminal sterilization after aseptic filling [2]. This terminal sterilization targets the surfaces of the drug delivery device and is not a sterilization of the biologic contained within. The need for terminal sterilization comes from the criticality of the application. For example, prefilled syringes used for ophthalmic injections (intravitreal injections) are terminally sterilized after filling and packaging to render the external surfaces of the syringes sterile to reduce the risk of intravitreal infection [3]. Additionally, autoinjectors may undergo terminal sterilization where microbial contamination in the fluid path would pose a risk to immunocompromised patients.
The sterilization method used for the drug delivery devices must be selected with consideration for the therapeutic agent in the filled drug delivery device and for the sterilization compatibility of the materials used in the construction and packaging of the device. Candidate sterilization methods include radiation sterilization (gamma, X-ray, or e-beam) and gas sterilization, including ethylene oxide (EO), hydrogen peroxide (H2O2), and nitrogen dioxide (NO2). There are many papers regarding the radiation sterilization of polymers but very few about gas sterilization’s compatibility with polymers [4,5,6]. However, terminal sterilization using ionizing radiation will degrade pharmaceutical and biologic agents contained in the drug delivery devices and is typically not suitable for terminal sterilization [4]. Therefore, there is a need for detailed analyses of the nature of chemical reactions observed with the gas sterilization of polymers [4,5,6].
Some gas sterilization methods can cause degradation of the therapeutic agent within the drug delivery device. The mechanisms of degradation with gas sterilization may result from the elevated temperature of the sterilization process and from the gas sterilant passing through the primary container closure and contaminating the biologics within [7].
Regarding the temperature degradation of thermolabile biologics, the bioactivity of biologics can be reduced by elevated temperatures during processing, storage, and transit, as aggregation, denaturation, and oxidative degradation render therapeutic molecules ineffective [8]. EO sterilization will use a process temperature that is greater than 40 °C and H2O2 sterilization will require process temperatures greater than 30 °C, whereas NO2 sterilization has been validated at as low as 18 °C. The potential degradation caused by elevated temperature sterilization processes will follow the Arrhenius equation, meaning that increasing process temperature will increase the degree of degradation [9]. Therefore, the relatively low process temperature of the NO2 sterilization process will be less degradative for biologics.
The degradation of the biologics caused by contamination with the sterilant gas entering the primary container has been found to occur with both EO and H2O2 sterilization [7]. Conversely, the NO2 sterilization process did not result in this degradation of proteins. The reason for this is that NO2 does not pass through the bulk of polymeric materials like radiation nor permeate through materials as easily as EO [10,11]. The permeability of hydrogen peroxide through materials has also been reported, which should be taken into consideration as a potential mode for degradation of the biologics [12].
In addition to permeability, the magnitude of the pressure fluctuations during the EO and H2O2 sterilization processes span from ambient pressure to less than 100 mbar and as low as 1 mbar for H2O2, causing the piston in a prefilled syringe to move, which may contribute to contamination of the syringe contents and degradation of the proteins in the reported studies [13,14]. Given the lower process temperature and ability to use a process with a small magnitude of pressure fluctuations, NO2 sterilization has gained acceptance as a gas sterilization method for drug delivery device terminal sterilization [15].
Another consideration for the design of drug delivery devices is the material compatibility of the drug delivery device with the intended terminal sterilization method. Prior published data on the chemical reactions observed with polymeric materials exposed to NO2 considered the exposure of such polymeric materials to atmospheric pollutants, including NO2 [16]. The concentration of NO2 in the environment is low (less than 1 ppm), and as such, the rates of observed reactions with polymers are slow. To accelerate the nature of potential interactions between polymers and atmospheric levels of NO2, the published data reported on reactions observed using elevated temperatures and elevated NO2 concentration levels so that the reactions are amplified compared to true atmospheric exposure [16]. The reported results from these studies are instructive but do not portray the conditions that polymeric materials will encounter during an NO2 terminal sterilization process. The objective of this report is to share data on polymer compatibility with NO2 sterilization and to describe the types of chemical reactions that can occur when polymers are exposed to NO2 at the sterilant-relevant temperature and concentration levels. An understanding of these reactions will further facilitate the design of drug delivery devices that use NO2 for terminal sterilization.

1.1. NO2 Sterilization Process

NO2 is a gaseous sterilant which, when combined with humid air, can sterilize medical devices [17]. The NO2 process has been validated, and marketed drug delivery devices use the NO2 terminal sterilization method [18]. In support of the validation, the biological indicators used for the NO2 sterilization process are shown to exhibit predictable inactivation kinetics, permitting the demonstration of the necessary sterility assurance level (SAL) [19].
The NO2 sterilization process is completed in a vacuum chamber and begins with evacuation of the air in the sterilization chamber. After evacuation, the sterilization chamber is filled with the sterilization process gases which consist of NO2 and humidified air. For most medical devices and drug delivery devices that can tolerate a deep vacuum, the chamber is evacuated to a relatively low vacuum pressure (e.g., 20 mbar; see Figure 1). Some drug delivery devices, such as prefilled syringes, are susceptible to piston movement during the sterilization cycle, due to the influence of the evacuated pressure, which causes the gases in the prefilled syringe to expand. The movement of the syringe piston may allow for the contamination of the syringe contents with both microorganisms and sterilant gas. To address this issue, a shallow-vacuum cycle has been developed. This type of shallow-vacuum cycle uses a much higher minimum process pressure (e.g., about 500 mbar minimum process pressure) compared to typical minimum process pressure levels used with deep vacuum cycles.
With the NO2 process, both deep vacuum and shallow-vacuum cycles have similar process stages. These stages are described below and illustrated in Figure 1. The process stages are:
-
Evacuation: The chamber is evacuated down to the designated minimum process pressure, which removes air from the sterilization chamber, permitting the addition of the process gases.
-
Humidity Addition: Relative Humidity (RH) is added to the chamber until the desired percentage is reached, typically between 65% RH and 80% RH.
-
NO2 Injection: the NO2 is metered into the sterilization chamber, with a target concentration typically between 5 mg/L and 20 mg/L.
-
Pressurization: Dry air is added to the sterilization chamber to reach the target process pressure (dwell pressure), which is typically between 700 mbar and 800 mbar.
-
NO2 Exposure Stage: The NO2 and RH remain in the chamber for the set dwell time of the NO2 exposure. The dwell time ranges from 5 min to 30 min.
-
Aerations: Aeration consists of repeated stages of evacuating and refilling the chamber with air until the sterilant is removed from the chamber.
Some sterilization cycles define the stages listed above as a half-cycle, whereby the biological indicators are found to be sterile after completing these stages. For a sterilization full-cycle, these stages are repeated to achieve the needed sterility assurance level, or SAL, as described in Annex D of ISO 14937:2009 [20]. Therefore, a sterilization cycle may consist of two sterilant exposure stages.
To test material compatibility, polymers were exposed to either three full-cycles (six exposure stages), referred to as the Mid-Range Exposure or 3× cycle, or six full-cycles (twelve exposure stages), referred to as the High-Range Exposure or 6× cycle.

1.2. NO2 Reactions

Chemical reactions between polymers and NO2 can be anticipated by examining the polymeric molecular structure. NO2 is an oxidizer and a free radical because of the lone electron on the nitrogen atom, written as NO2. At the temperature and concentration of the NO2 sterilization process, the possible oxidative reaction mechanisms with polymers include addition to the double bond pi structure of a phenyl group, the abstraction of hydrogen from the backbone of the polymeric chain, and electron transfer mechanisms [21].
For example, with nylon and polyurethane, the N–H group in the polymer chain is susceptible to oxidation by NO2 because of the functional amide group in nylon (–CO–NH–) and the carbamate group in polyurethane (–NH–CO=O–) [22]. In Delrin (polyoxymethylene, or POM), the oxidative degradation pathways lead to chain scission at the methylene group (–O–CH2–) and result in formaldehyde release [23].
As an example, NO2 can add to double-carbon bonds in a polymer backbone, leading to the formation of a c-centered radical (Figure 2).
The resulting reducing C-centered radical may undergo further oxidizing reaction to produce the cation and NO2- ions or produce di-nitro compounds and nitro nitriles (Figure 3). In this example, the reaction between the double-carbon bond and the NO2 resulted in radical formation.
As a further example of potential reactions, NO2 can react with the vinylene and vinylidene groups found in polyethylene and polypropylene, respectively. Published studies have also demonstrated that NO2 can interact with the phenyl, carbonyl, ester, and ketone groups in various polymers [16].
As a final example of potential reactions, at relatively high temperatures (above 310 K), the reactivity between NO2 and polymers increases. It has been reported that carbonyl groups and hydroxyl groups can be formed [24].
Other reactions that might be anticipated are low-level surface reactions, such as carbamate oxidation, acetal cleavage, or amine nitrosation, which may occur when exposed to gaseous NO2 under ambient conditions [24]. For polymers with more oxidizable linkages, the overall extent of the reaction is kinetically limited, which is consistent with the observed preservation of mechanical and optical properties following NO2 sterilization. For these other reactions, and despite the oxidizing properties of NO2, reactions between NO2 and these polymers require overcoming activation-energy barriers, which can be achieved only at elevated temperatures [16,24].
With this knowledge of possible reactions, we have chosen polymers for this study with carbon backbones, methylene groups, phenyl groups, double-bonded oxygens, and other features that are representative of potential reaction sites with NO2. These materials are not reflective of all medical device materials but represent the specific molecular structures we are investigating.
The selected polymers are Amorphous Polyester, Polysulfone (PSU), Polycarbonate (PC), PolyEtherEtherKetone (PEEK), PolyArylEtherKetone (PAEK), and Polypropylene (PP), as representatives of hydrocarbon aliphatic polymers, polycarbonate, and polyester groups. Table 1 summarizes the identifiable molecular structures and functional groups of each material, along with the common medical devices in which these materials are used.
This work addresses gaps in the literature regarding NO2 gas exposure at lower temperatures (10–30 °C), high concentrations, and long exposure times to extensively evaluate the materials and their compatibility with the NO2 sterilization process. This work is only the start of answering gaps in the literature. Additional polymers will be considered elsewhere and will include studies of exposed polymers aged in real time (not accelerated).

2. Results

2.1. NO2 Achieved Parameters

All material exposure cycles were completed according to the parameters and tolerances listed in table of Section 4.4. A total of seven Mid-Range exposure cycles and ten High-Range exposure cycles were completed. The average exposure conditions achieved are summarized in Table 2. The vacuum level reached 20 torr for all cycles. The exposure pressure was consistently 592 torr, which is 2 torr above the targeted pressure but is within the set tolerance. The relative humidity was also consistently at 79.6%, which is only 0.4% off the target. While most validated cycles have a temperature range 18–22 °C, the material exposure cycles in this study had an average exposure temperature of 24 °C, providing a “worst case” for polymer exposure.
The uncertainty for all cycle parameter values was calculated from the standard deviation of all the NO2 concentrations achieved for each group.

2.2. NO2-Induced Chemical Reactions

2.2.1. Surface Characterization Changes

The surface of each material was characterized using FTIR-ATR Spectroscopy, as shown in Figure 4. No significant changes in the FTIR-ATR spectra were observed in the comparison of High-Range-exposed samples with control samples. The variations were analyzed using a two-sample t-test using a 95% confidence interval, which showed no shift of absorbance wavenumbers or significant changes in absorbance (peak height). Insignificant variations in peak height (less than 5%) are attributed to instrumental factors such as spectral resolution, signal-to-noise ratio, and variability in ATR contact rather than actual differences in the samples themselves [37,38]. This indicates that there were no molecular changes in the polymers tested after NO2 exposure.

2.2.2. Free Radicals

Each material was analyzed for the presence of free radicals using EPR. Only the NO2 High-Range-exposed Amorphous Polyester samples exhibited a signal at 3515 G, as shown in Figure 5. For this signal, the g-value was calculated using Equation (3) and was found to be 2.0029. The number of radical electron spins per gram of material was calculated with Equation (4) and was found to be 1.71 × 1014 spins/g.

2.2.3. Change in Morphology

The morphology of each material was evaluated using DSC. For amorphous materials, the glass transition temperature (Tg) was measured, and for semi-crystalline materials, the melting temperature (Tm) was measured. All reported peaks agree with previously reported literature values. The variations were analyzed using a two sample t-test using a 95% confidence interval, which showed that no samples exhibited a significant difference in melting or glass transition temperatures (as is shown in Table 3 and Table 4), indicating no change in bulk morphology with NO2 exposure. The reported literature values for the glass transition temperature for Polyester [39], PSU [40], and PC [41] are found in Table 3, and the reported literature values for the melt temperature of PEEK [42], PAEK [43], and PP [44] are found in Table 4.
For the semi-crystalline polymers, the crystallinity of the test samples was calculated using literature reference values for 100% crystalline Polypropylene [45], PAEK [46], and PEEK [47]. There was no significant difference between the percent crystallinity of Mid-Range-exposed samples and High-Range-exposed samples when compared to the control percent crystallinity (Table 5), further confirming the lack of change in the polymer structure upon NO2 exposure.

2.2.4. Mechanical Changes

Changes in the mechanical properties of exposed materials were evaluated using measured mechanical properties from tensile testing. For the five samples tested from each group, the average true yield strength, true strength at fracture, and true percent elongation at fracture of each polymer are compared in Table 6. Elastic Modulus and strain at yield or fracture are not reported since an extensometer was not used during testing. A comparison of the Mid-Range-exposed samples and High-Range-exposed samples to the control samples shows no significant difference (significance level = 0.05) using an unpaired two-tailed t-test.

2.2.5. Shore Hardness

No significant change in hardness (Durometer) was observed between the control, Mid-Range-exposed, and High-Range-exposed samples. The Shore D hardness values for the materials Amorphous Polyester, PSU, PC, PEEK, PAEK, and PP are summarized in Table 7.

2.2.6. Color Change

The materials in this study were evaluated for visual change using a color reader that provided a ΔE value, comparing the color of the control samples with the exposed samples. The results from these measurements are shown in Table 8. Polypropylene and Polycarbonate exhibited the greatest degree of color change, with a ΔE value greater than 4. The other samples had a ΔE value less than 2 after exposure.

2.2.7. Surface Residuals

All materials were extracted in the specified manner (see Materials and Methods, Section 4.5) and the extraction medium was evaluated for the concentration of extracted nitrates and nitrites. The limit of quantification (LOQ) was 0.043 ppm for nitrates and 0.038 ppm for nitrites, translating to 0.034 µg/cm2 and 0.030 µg/cm2, respectively. The values of extracted sterilant residuals are shown in Table 9. All the controls were low, less than 0.660 ug/cm2 for nitrates and 0.0746 ug/cm2 for nitrites.
The nitrates are average values of the 6×exposed group and the nitrite values are the averages of 3× and 6× groups based on the highest average measured for each group.

2.2.8. Biocompatibility

The observed cytotoxicity grade, based on the resulting cell culture conditions corresponding to the numerical cytotoxicity grade for each material, can be found in Table 10. The interpretation of the grade is in table of Section 4.5. All other materials were found to be non-cytotoxic.

3. Discussion

3.1. FTIR

The FTIR-ATR measurements did not detect chemical changes in the polymers. However, an absorbance band in the range of 3200 cm−1 to 3750 cm−1 was observed. This absorbance band is shown in Figure 6 and suggests the presence of O-H or N-H stretching vibrations. However, following the washing of samples with water, this absorbance band was not detected. This observation correlates with the analysis of the sterilant residuals, extracted in aqueous solutions. Because we can extract varying levels of nitrate and nitrite ions, as shown in Table 9, we know there is residual nitrate and nitrite on the polymer surfaces. This suggests that the observation of absorbance in the range of 3200 cm−1 to 3750 cm−1 after exposure is not a result of a chemical reaction; rather, sterilant residue was present on the material surfaces.
Polyethylene (PE), Amorphous Polyester, PAEK, and PEEK each show a change in the spectrum from 3200 cm−1 to 3750 cm−1. PC did not show a change in spectra between unexposed and exposed samples.
The NO2 sterilization process uses an atmosphere of NO2 and humidified air. It is known that humidity interacts with materials, forming monolayers of water, within which the reactions between water and NO2 occur [17]. The thickness of the water monolayers on a material surface is dependent on the surface material and the partial pressure of water at the surface [48]. The monolayers of water form because the process adds humidity to the chamber, resulting in water molecules forming these monolayers on the material surfaces. When NO2 gas encounters water, one of the reactions that can occur is the formation of two acids: HNO3 (nitric acid) and HNO2 (nitrous acid), as shown in Equation (1), below.
2NO2(g) + H2O(aq) → HNO3(aq) + HNO2(aq)
HNO2(aq) → NO(g) + HNO3(aq) + H2O(l)
HNO2 is inherently unstable and decomposes rapidly into NO (nitric oxide), HNO3, and additional water (Equation (2)). The instability of HNO2 is illustrated in the sterilant residual testing results shown in Table 9. In this table, the amount of recovered nitrite is much lower than that of nitrate, regardless of the expectation of equal stoichiometry implied by Equation (1).
At the end of the sterilization process, the vacuum in the sterilization chamber vaporizes water, leaving an amount of HNO3. The HNO3 contains a highly polar –OH group and a strongly electronegative –NO2 moiety. On polymers with hydrogen bond acceptors (e.g., –C=O, –O–, –N–), HNO3 can form directional hydrogen bonds. When immersed in water, the water can break these hydrogen bonds through competitive solvation and molecular displacement. Therefore, this is not a chemical reaction between NO2 and the polymers.

3.2. EPR

The EPR spectrum shown in Figure 5 displays a signal with a g-value of 2.0029, which is consistent with carbon-centered radicals, particularly those adjacent to electronegative atoms such as oxygen [49,50]. Because there is no oxygen neighboring a carbon in the phenyl group, we conclude that the NO2 abstracts a hydrogen atom from the polyester backbone and not the phenyl group. The potential reactions between NO2 and the Polyester backbone are shown in Figure 7.
This observation of an EPR signal is important because it could elucidate a critical polymer degradation pathway. The published literature on this topic has numerous examples of degradation mechanisms leading to decreased tensile strength, embrittlement, and reduced molecular weight, the effects of which may not be immediately apparent but manifest as long-term aging and performance loss. This is caused by the free radicals generated in polymers during radiation sterilization. In the presence of oxygen, these primary C-centered free radicals react to form peroxyl radicals, which then undergo further reactions, leading to the degradation of the polymer chains, particularly at low dose rates common in some sterilization processes [51].
After the first sample of Amorphous Polyester was observed to exhibit radicals in the EPR spectrum, subsequent testing was performed. This included three (3) samples of Polyethylene Terephthalate Glycol (a modified Polyester), three (3) samples of Crystalline Polyester, and two (2) samples of Amorphous Polyester. All samples in this subsequent testing were placed on dry ice immediately following NO2 exposure until EPR testing. Only one of the two samples of Amorphous Polyester during this subsequent testing exhibited an EPR signal showing radical formation.
The two samples of Amorphous Polyester that had shown evidence of radicals were retested: the first retest after 25 days and the second retest after 6 days. No EPR spectrum was observed in either sample, suggesting that the C-centered radicals undergo crosslinking or some other relaxation reaction and decay rapidly.
No changes in functional properties were observed for the Amorphous Polyester, and the material only exhibited low surface residual levels and a non-cytotoxic response. Other Crystalline Polyester samples from different batches did not exhibit EPR signals. The lack of EPR signal may be because the Crystalline Polyester samples contain certain antioxidants, or the NO2 diffusion through the sample bulk was impeded because of the crystallinity. Table 11 lists the additional EPR testing performed and the material information of the samples tested.

3.3. Tensile Testing and Shore Hardness Durometer

There was no significant change in mechanical properties or in the measured hardness after NO2 exposure with the materials reported herein. The value of establishing a correlation would be for screening material impact with a simple hardness test. Other published reports have correlated changes in hardness to changes in mechanical properties. It has been shown that changes in polymer durometer can occur from cross-linking after radiation sterilization [5]. Additionally, research from Tomas Bata University examined polyamide 12 (PA12) irradiated with e-beam, where they found that nano-hardness increased by 61% at 132 kGy, directly correlating with increased crosslink density [5]. A white paper from Saint-Gobain Performance Plastics studied the effects of EO sterilization on silicone rubbers. While EtO is less aggressive than radiation, the paper notes that crosslinking reactions can still occur, especially in peroxide-cured silicones, resulting in increased durometer hardness and tensile modulus [52]. However, for the materials reported here, we did not observe changes in mechanical properties or in hardness. Therefore, the correlation between mechanical properties and hardness is inconclusive.
Across all polymers, changes are within ±1 Shore D unit, which is well within measurement variability. This suggests that NO2 sterilization does not cause substantial embrittlement or softening, preserving mechanical properties. Hardness correlates with surface durability and resistance to deformation. Since hardness remains stable, the polymers should maintain their load-bearing capability, wear resistance, and dimensional stability post-sterilization.

3.4. Color Change

The materials in this study were evaluated for visual change using a color reader that provided a ΔE value by comparing the difference between control and exposed samples. ΔE (Delta E) is a value that quantifies the perceptual difference between two colors. The threshold for what is considered a noticeable color change depends on the ΔE value and the context of viewing. This is described in Table 12 [53]. Only Polycarbonate and Polypropylene had moderate color change after the High-Range Exposure, the rest of the materials had none or only slight color change.
Material color change is observed with other modes of sterilization, most notably following radiation sterilization. To compensate for small color changes, some polymers are blended with the colorants formulated to account for the color change [6]. Color change can be an indication of a reaction of the sterilant with the polymer, or it can be a reaction with the colorants and other additives added to the polymer (e.g., antioxidants, stabilizers, lubricants, etc.) [7]. For example, a common type of organic antioxidant found in medical-grade polymers is phenolic antioxidants [55]. These antioxidants are widely used primary antioxidants in medical-grade polyolefins (like polyethylene and polypropylene) due to their high efficiency, low volatility, and good toxicological clearance [56]. The phenolic antioxidant acts as a free radical scavenger, interrupting the polymer degradation process by undergoing a hydrogen-transfer mechanism. Figure 8 shows the reaction between the phenolic antioxidant and NO2.
The resulting alkoxyl radical is less reactive than the original radical and can react further with another NO2 radical or undergo other reactions, often leading to nitrated phenols or quinone methide structures, which may cause discoloration (yellowing) of the polymer [57]. This discoloration is not necessarily a sign of polymer degradation but rather the sacrificial transformation of antioxidants. Any additional oxidation removes the unpaired electron from the oxygen, requiring the double-bonding of the O to the aromatic group, resulting in the quinone structure.

3.5. Residuals and Cytotoxicity

Cytotoxicity is a foundational test in demonstrating the biocompatibility of a material [58]. Following NO2 sterilization, some level of residuals will remain on the material surface. Residual retention is material-dependent. However, we can use the measured values matched with biocompatibility data to determine levels of residuals that would be harmful. The highest level of residuals shown in Table 9 is for Amorphous Polyester, which had extracted nitrate at 90 µg/cm2. This level of residuals was not cytotoxic.
Unlike with EO, there is no established limit for the acceptable level of residuals on a medical device. While none of the NO2-exposed materials exhibited cytotoxicity, H2O2 is found to leave cytotoxic residual chemicals on the surface of polymers. In the report by Ikarashi et al. [59], H2O2 residuals recovered above 1 µg/cm2 were cytotoxic. Similarly, EO residuals are found to be cytotoxic at low residual levels. From ISO 10993-7, the maximum permissible limit of EO residuals is 10 µg/cm2 [60]. This is supported by the report from Lucas et al. [61], where they found that materials with about 1 µg/cm2 cause a cytotoxic response in cell cultures. These results demonstrate that the sterilant residuals recovered from polymer surfaces after NO2 sterilization are less cytotoxic than the same mass density of residuals from either H2O2 or EO [59].

3.6. Material Resistance

Whereas reactions potentially might occur between NO2 and the identified molecular structures (e.g., double oxygen bonds, etc.), we did not always observe these reactions. Therefore, we conclude that some materials do not degrade in the presence of an oxidative gas like nitrogen dioxide due to factors like availability and steric hindrance. For example, where one might expect a reaction between NO2 and ether linkages (-O-Ar) in Polysulfone, it instead exhibits notable chemical resistance during NO2 sterilization, in part due to steric hindrance arising from its rigid, aromatic backbone [62]. The polymer’s repeating unit—composed of phenyl rings flanking sulfone and ether linkages—creates a spatially crowded environment that limits NO2 access to reactive sites. This steric shielding reduces the likelihood of hydrogen abstraction and electrophilic attack, which are common oxidative pathways for NO2 degradation in polymers such as nylon and polyurethane. Unlike aliphatic or polar polymers, Polysulfone lacks readily accessible hydrogen atoms adjacent to electron-rich centers, and its sulfone group is already in a fully oxidized state, making further oxidation energetically unfavorable. As a result, NO2-induced chain scission is minimal. Mechanistically, NO2 reactions typically proceed via radical initiation, where NO2 abstracts a hydrogen atom to form a polymer-centered radical and HNO2.
This paper evaluates compatibility between NO2 with selected polymers soon after the polymers were exposed to NO2. At the time that this manuscript is being prepared, samples exposed during this study and stored at ambient indoor temperature and humidity are being prepared for testing after one year of real-time aging. Also, additional polymers and elastomers are being evaluated to expand on the learning gained during this testing reported.

4. Materials and Methods

4.1. Materials

The materials reported herein were selected for their constituent chemical bonds and as representative examples from common polymer groups that are used in medical devices. These polymers are listed in Table 13. As discussed above, these polymers could potentially react with NO2 [21,24]. While in no way exhaustive, this group of polymers will provide insight into the general compatibility of polymers with the NO2 sterilization process and instruct material selection when designing drug delivery devices.

4.2. Material Characterization Methods

The testing of the materials post-exposure to NO2 consisted of seven characterization methods. These characterization methods evaluated the change in chemical and mechanical properties after exposure to the NO2 sterilization process. These characterization methods were as follows:
  • FTIR–ATR: Fourier Transform InfraRed Attenuated Total Reflectance (FTIR-ATR) Spectroscopy is a technique used to analyze the molecular structure of materials. The FTIR-ATR absorption spectrum was normalized to provide consistent evaluation across samples and groups. FTIR-ATR measures to a depth of 0.5 × 10−6 m to 5 × 10−6 m.
  • EPR: Electron Paramagnetic Resonance (EPR) was used for evaluating the formation of radicals on the material after NO2 exposure. The presence of an EPR signal indicates free radicals in the polymer. Comparing the spectra from control samples with the spectra of exposed samples can detect the creation of radicals in the polymer due to the exposure to the NO2 process.
  • DSC: Differential Scanning Calorimetry (DSC) was used to determine whether a change in morphology occurred—specifically, a change in the glass transition temperature or melt temperature from chain scissions or cross-linking. The critical temperatures were evaluated for each material and reported with the uncertainty values. Variations in these measurements indicate alterations in polymer structure, such as changes in crystallinity, crosslinking, or scissions.
  • Mechanical Properties: Tensile testing was used to determine the yield strength, strength at break, and elongation at break for unexposed control, Mid-Range-exposed, and High-Range-exposed samples. The results indicate whether the sterilization process affected the bulk mechanical properties of the materials.
  • Hardness (Durometer): Shore Hardness Durometer was used to evaluate the bulk surface characteristic of the unexposed control, Mid-Range-exposed samples, and High-Range-exposed samples. Shore D was used to measure all of the materials hardness values. Hardness changes can indicate a change in polymer morphology. The change in polymer hardness from cross-linking after radiation sterilization is well studied [5].
  • Color Change: Color change (ΔE) was measured between the control and Mid-Range-exposed samples and between and the control and High-Range-exposed samples. The exposure of polymers to NO2 can lead to discoloration (yellowing). This discoloration has multiple mechanisms, including reaction with additives, oxidative degradation, and surface adsorption [63]. NO2 reacts with additives like phenolic-type stabilizers in the polymer matrix, forming quinone-like structures or nitroso derivatives [64]. These compounds absorb light in the visible spectrum, causing a yellow or pink tint. Additionally, NO2 is an oxidizer and can initiate chain scission or crosslinking, which can alter the polymer’s chemical structure and create chromophoric groups that absorb visible light. Finally, NO2 can permeate into the polymer surface and cause color change from the NO2 molecules present in the material.
  • Surface Residuals: Surface residuals were measured using a colorimetric Griess reagent assay. Relating the level of surface residuals (nitrates and nitrites) to the cytotoxicity provides insight into whether biocompatibility will depend on reactive byproducts between NO2 and the polymer blend (including additives), or from the sterilant residuals. Published literature has shown that sterilant residuals are cytotoxic above threshold levels [59,61].
  • Cytotoxicity: Cytotoxicity was used as an indicator of biocompatibility, combined with measured surface residuals (nitrates and nitrites). Cytotoxicity was performed in an external laboratory following GLP measures.

4.3. Material Preparation

All polymeric materials were received from the material suppliers in the form of ASTM Type 1 tensile bars [65]. The samples were inspected, cleaned by wiping with distilled water, and then air-dried. The dry samples were sealed in Tyvek/Mylar pouches (A7320 Roll Stock from Beacon Converters) and labeled. The packaged samples were divided into three groups. The first group was set aside as unexposed control samples. A second group of samples was exposed to the Mid-Range NO2 exposure cycle, while the third group of samples were exposed to the High-Range NO2 exposure cycle, following the parameters outlined in Table 2.
Samples for DSC and EPR testing had to be cut from the tensile bars into smaller pieces, weighing less than 5 mg. These samples were wiped with distilled water and air-dried before being placed in pouches and labeled.

4.4. Material Exposure

The exposure cycles were completed in the 730 L volume R&D sterilizer (Noxilizer Model RTS-360, Noxilizer, Hanover, MD, USA). The exposure parameters are listed in Table 14.

4.5. Post-Exposure Handling and Testing

After the exposure cycles were completed, the samples for FTIR-ATR testing were washed to remove exposure process residues. This washing process consisted of placing the samples in extraction bags with water and incubating for 72 h +/− 2 h at 50 °C, following ISO 10993-12 instructions for material extraction [66]. This process was adopted for sample washing because this is the washing process used to extract residuals from the material surfaces for the residual sterilant testing and for the testing cytotoxicity of the materials. For EPR, used for the detection of radicals like NO2, washing the NO2 from the surface would remove any signal from NO2 persisting within the material, hindering the detection of NO2 free radicals that could contribute to ongoing degradation.
After drying the washed samples, the samples were placed in resealable bags and transported to the University of Maryland (UMD) Department of Materials Science and Engineering Laboratory for Radiation and Polymer Science. The FTIR-ATR testing was performed using a Thermo-Scientific Nicolet IS50 FTIR following ASTM E168-16 (2023) [67]. Five control samples, five Mid-Range-exposed samples, and five High-Range-exposed samples were evaluated with FTIR-ATR. The spectra from control samples were compared to the spectra of exposed samples. The changes in the FTIR spectra were reviewed to identify new or missing chemical bonds in the polymer.
The EPR samples were transported to the UMD Laboratory for Radiation and Polymer Science. One control sample and one High-Range-exposed sample were evaluated with EPR. The EPR spectra of all samples were measured using a Bruker EMX EPR spectrometer (Bruker, Billerica, MA, USA), equipped with an X-band (9–10 GHz) microwave source. EPR spectra were measured 24 h after exposure to NO2. All measurements were conducted at room temperature. 2,2-diphenyl-1-picrylhydrazyl (DPPH) was used as a standard for the quantification of spins within the polymer samples tested. The g-value for the EPR results was calculated using Equation (3):
g = h v μ B × B
Here, h is Planck’s constant, ν is the microwave frequency, μB is the Bohr magneton, and B is the magnetic field strength at which resonance occurs. The number of spins in the sample was calculated by integrating the EPR signal of a polymer sample and comparing this to the standard sample of DPPH with a known spin concentration (Equation (4)).
x = s t d A x R x s c a n x 2 G s t d M s t d g s t d 2 s s + 1 s t d A s t d R s t d s c a n s t d 2 G x M x g x 2 s s + 1 x
In Equation (4), [std] is the known concentration of the spins/g in the DPPH, A represents the area under the EPR absorption curve, and R is the receiver gain. The term g refers to the g-value of the radical (Equation (3)), which serves as a unique fingerprint for identifying radical species, and s is the electron spin quantum number. The instrumental parameters used in the calculation include G, the modulation frequency (measured in Hz); M, the modulation amplitude (in Gauss); and scan, the magnetic field sweep width (also in Gauss). The subscripts “std” and “x” refer to the standard DPPH sample and the experimental polymer sample, respectively.
Differential Scanning Calorimetry (DSC) using a Perkin Elmer PYRIS Diamond DSC was used to assess the changes in morphology induced by NO2 exposure by evaluating changes in melting temperature (Tm), crystallization temperature (Tc), or glass transition temperature (Tg). This characterization followed ASTM D3418-21 [68]. Three control samples, three Mid-Range-exposed samples and three High-Range-exposed samples were transported to the UMD Laboratory for Radiation and Polymer Science to be evaluated using DSC. The DSC testing program consisted of an initial heat ramp above Tm to erase thermal history, a cooling ramp back to the starting temperature (below Tg), and a final heat ramp above Tm. All heat ramps were run at 10 °C/min. Peaks were identified by analyzing the derivative of the second heat ramp heat flow curve.
Mechanical properties were characterized by measuring yield strength, strength at fracture, and elongation at fracture using the Tinius Olsen H25KT Universal Testing Machine, following guidance from the ASTM D638-22 [65]. Five control samples, five Mid-Range-exposed samples, and five High-Range-exposed samples were transported to The UMD Laboratory for Radiation and Polymer Science to be evaluated for mechanical property characterization. The samples were fixed in the jaws to align with the end tabs. The machine was set to a strain rate of 50 mm/min and recorded force (N) as a function of displaced position (mm). The data were analyzed by graphing the true stress and strain from the force and displacement data. The graph was used to determine the yield strength, strength at break, and elongation at break. Uncertainty using a 95% confidence interval test is reported, and an unpaired two-tailed t-test was used to evaluate the consistency of the data [69].
The hardness characterization of materials reported here used a PosiTector Shore Hardness Durometer (Model SHDD1) with a DeFelsko Shore Hardness D Durometer Probe (DeFelsko, Ogdensburg, NY, USA) following the Durometer Hardness Test described in ASTM D2240-15(R21) [70]. For each material, hardness was tested 15 times: five tests on each of the three samples from each of the control, Mid-Range-exposed and High-Range-exposed samples. The DeFelsko Shore Hardness D Durometer probe is a hardened steel rod with a 30° angle conical point indenter. Uncertainty, calculated using a 95% confidence interval test, is reported.
Color change analysis was performed using a Konica Minolta Color Reader CR-10 Plus (Konica Minolta, Tokyo, Japan) with a White Tile and Polaroid Photo Box. The procedure was guided by ASTM E1347-06 (2020) [71]. The measurement of ΔE is a comparison of two scans, which in this study is a comparison of a scan of a control sample to the scan of a Mid-Range- or High-Range-exposed sample. Three samples for Mid-Range-exposed and High-Range-exposed groups were evaluated against a control sample. For each sample, color was tested three times. Uncertainty calculated using a 95% confidence interval test is reported.
Surface Residuals Analysis employed a colorimetric method (Griess Reagent assay) to quantify the extracted residuals, such as nitrate (NO3-) and nitrite (NO2-) residuals on the surface of materials [72]. When paired with the results of biocompatibility testing, this analysis of surface residuals can support the establishment of acceptable limits for residuals on the surfaces of NO2-exposed materials. Five control samples, five Mid-Range-exposed samples and five High-Range-exposed samples were evaluated for surface residuals. Control samples, Mid-Range-exposed samples, and High-Range-exposed samples were placed in extraction bags with water and incubated at 50 °C for 72 h +/- 2 h, following ISO 10993-12 instructions for material extraction [66]. After extraction, the extraction media were placed in cold storage (2 °C to 8 °C) until testing. Uncertainty, calculated using a 95% confidence interval test, is reported.
Cytotoxicity testing was performed by Millstone Medical Testing (Bloomfield, CT, USA) using the elution method described in ANSI/AAMI/ISO 10993-5: 2009/(R) 2014 [73]. A material may be considered non-cytotoxic with a grade of 0, 1, or 2, as shown in Table 15. A control sample and a High-Range-exposed sample were tested for cytotoxicity for all six polymers, and a Mid-Range-exposed sample was tested for three of the polymers.

5. Conclusions

Medical devices and drug delivery devices require materials that are compatible with terminal sterilization. The current portfolio of medical device polymers represents a form of natural selection driven by sterilization requirements. Over decades, materials incompatible with EO or gamma radiation have been phased out or reformulated, leaving a universe of polymers that reliably withstand these processes. NO2 sterilization joined the sterilization landscape with a need to partner with materials suppliers to investigate polymers and polymer additives that are compatible with NO2 as well as other sterilization methods like H2O2.
These detailed results will aid device designers and regulators in the types of chemical reactions that may occur with polymers during NO2 sterilization. The data demonstrate the compatibility of Amorphous Polyester, PSU, Polycarbonate, PEEK, PAEK, and Polypropylene with the NO2 sterilization process at room temperature. Despite the oxidizing properties of NO2, as well as it being a free radical, reactions between NO2 and these polymers require overcoming activation-energy barriers, which can be achieved only at elevated temperatures [16,24]. No significant chemical changes were observed, and the mechanical testing showed no changes in their morphologies, mechanical properties, or surface hardness. Even though Amorphous Polyester had a low signal for radicals, the additional data showed no significant changes, indicating that the radicals were not a source of degradation for the material. Color changes were observed with the Polypropylene and Polycarbonate samples, which were graded as moderate. All other samples exhibited only a slight change that would be only noticeable to trained observers or under controlled conditions. The color change seen with materials may be due to reactions between NO2 and additives in the polymer blends. Based on our results, none of the samples tested had a residual sterilant density greater than 90 µg/cm2. Additionally, unlike EO-sterilized materials and H2O2-sterilized materials, none exhibited cytotoxicity, whereas more than 1 µg/cm2 of residual EO or H2O2 is unacceptable.
This evaluation of six materials demonstrates the rigorous testing and analytical methods needed to establish material compatibility. The materials tested for this report are compatible with use in drug delivery systems and medical devices. The methodology presented in this study can be applied to evaluate additional polymer systems for compatibility with NO2 sterilization. Ongoing work is evaluating additional materials to build the library of material compatibility and will include 1-year real-time aging results from the materials reported here.

Author Contributions

Conceptualization, D.O. and M.A.-S.; Methodology, N.R., J.D.S., A.A., D.L., T.S., D.O. and M.A.-S.; Validation, D.O. and M.A.-S.; Formal analysis, J.D.S., A.A., N.S., D.O. and M.A.-S.; Investigation, N.R., J.D.S., A.A., N.S., D.L., M.A.K., T.B., T.S., D.O. and M.A.-S.; Resources, D.L., M.A.K., T.S., D.O. and M.A.-S.; Data curation, J.D.S., A.A., N.S., M.A.K., T.B. and T.S.; Writing—original draft, N.R., J.D.S., A.A. and N.S.; Writing—review & editing, N.R., D.L., D.O. and M.A.-S.; Supervision, D.O. and M.A.-S.; Project administration, D.O.; Funding acquisition, D.O. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by Maryland Industrial Partnerships (MIPS), grant number 7302, and Noxilizer, Inc.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding authors.

Acknowledgments

The authors would like to thank Robert Bonenberger for access to testing equipment in the George E. Dieter Jr. Materials Instructional Laboratory at the University of Maryland, the staff at the Noxilizer Laboratory for their support with testing, and Joseph Naft and Ronnie Gist of the Maryland Industrial Partnerships (MIPS) Program for their financial support and assistance under grant number 7302.

Conflicts of Interest

Julia Diane Schexnayder, Aiysha Ashfaq, Nusrat Sarwahrdy, Trevor Batman, and Mohamad Al-Sheikhly received grant funding from Maryland Industrial Partnerships (MIPS) to support the research described in this manuscript. Noelle Ray, David Opie, Delaney Lisco, Tom Sadler, and Minufar Abdollahi Khabisi were involved in the administration of the grant funding from Noxilizer, Inc. and contributed to the study. All parties affirm that the funding source had no influence on the design, analysis, interpretation, or reporting of the results beyond the agreed terms of the grant. Author Noelle Ray, Delaney Lisco, Minufar Abdollahi Khabisi, Tom Sadler and David Opie were employed by the company Noxilizer, Inc. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Abbreviations

The following abbreviations are used in this manuscript:
DPPH2,2-diphenyl-1-picrylhydrazyl
DSCDifferential Scanning Calorimetry
EOEthylene Oxide
EPRElectron Paramagnetic Resonance
FTIR-ATRFourier Transform InfraRed Attenuated Total Reflectance Spectroscopy
HNO2Nitrous Acid
HNO3Nitric Acid
H2O2Hydrogen Peroxide
NO2Nitrogen Dioxide
PAEKPolyArylEtherKetone
PCPolycarbonate
PEPolyethylene
PEEKPolyEtherEtherKetone
PETGPolyethylene Terephthalate Glycol
PPPolypropylene
PSUPolysulfone
RHRelative Humidity
SALSterility Assurance Level

References

  1. Skalko-Basnet, N. Biologics: The Role of Delivery Systems in Improved Therapy. Biol. Targets Ther. 2014, 8, 107. [Google Scholar] [CrossRef] [PubMed]
  2. Altasciences. Terminal Sterilization of Pharmaceutical Products. The Altascientist 2021, 21. Available online: https://www.altasciences.com/sites/default/files/2021-12/The-Altascientist_issue-21_Terminal-Sterilization.pdf (accessed on 16 October 2025).
  3. Parenky, A.C.; Wadhwa, S.; Chen, H.H.; Bhalla, A.S.; Graham, K.S.; Shameem, M. Container Closure and Delivery Considerations for Intravitreal Drug Administration. AAPS PharmSciTech 2021, 22, 100. [Google Scholar] [CrossRef] [PubMed]
  4. Mollah, M.Z.I.; Rahaman, M.S.; Faruque, M.R.I.; Khandaker, M.U.; Osman, H.; Alamri, S.; Al-Assaf, S. Effects of Radiation Sterilization Dose on the Molecular Weight and Gelling Properties of Commercial Alginate Samples. Front. Mater. 2021, 8, 761995. [Google Scholar] [CrossRef]
  5. Naikwadi, A.T.; Sharma, B.K.; Bhatt, K.D.; Mahanwar, P.A. Gamma Radiation Processed Polymeric Materials for High Performance Applications: A Review. Front. Chem. 2022, 10, 837111. [Google Scholar] [CrossRef]
  6. Styron LLC. Advances in Polycarbonate Color Compensation Technology to Withstand Sterilization by Irradiation Methods. Med. Des. Briefs 2012, September Issue. Available online: https://www.medicaldesignbriefs.com/component/content/article/14641-advances-in-polycarbonate-color-compensation-technology-to-withstand-sterilization-by-irradiation-methods (accessed on 11 August 2025).
  7. Fujiwara, S.; Ueda, T.; Ishii, N.; Abe, Y.; Hayashihara-Kakuhou, K.; Takeda, K.; Kurinomaru, T.; Fukuhara, A.; Uchiyama, S.; Carpenter, J.F. Chemical-Gas Sterilization of External Surface of Polymer-Based Prefilled Syringes and Its Effect on Stability of Model Therapeutic Protein. J. Pharm. Sci. 2022, 111, 41–50. [Google Scholar] [CrossRef]
  8. Wang, W. Lyophilization and Development of Solid Protein Pharmaceuticals. Int. J. Pharm. 2000, 203, 1–60. [Google Scholar] [CrossRef]
  9. Ebrahim, A.; DeVore, K.; Fischer, T. Limitations of Accelerated Stability Model Based on the Arrhenius Equation for Shelf Life Estimation of In Vitro Diagnostic Products. Clin. Chem. 2021, 67, 684–688. [Google Scholar] [CrossRef]
  10. Phatak, A.; Burns, C.M.; Huang, R.Y.M. Transport of Ethylene Oxide through Polymer Films. J. Appl. Polym. Sci. 1987, 34, 1835–1859. [Google Scholar] [CrossRef]
  11. Chempedia. Gas Permeability of Polymers and Rubbers. Available online: https://chempedia.info/info/gas_permeability_of_polymers_and_rubbers/ (accessed on 20 June 2025).
  12. Gitsov, I.P.I.; Liu, Y.; Saylor, D.M.; Hendrick, A.L.; Gronda, A.M.; Sinha, M.N.; Long, E.; Liu, S.; Hu, R.; Weeks, J.W.; et al. Diffusion of Hydrogen Peroxide Through Medical Grade Poly(Ether)Urethane: Analyzing Mechanisms of Sorption and Transport to Support Sterilization With Vapor Hydrogen Peroxide. J. Biomed. Mater. Res. B Appl. Biomater. 2025, 113, e35567. [Google Scholar] [CrossRef]
  13. Mehta, S.B.; Cook, J.; Liu, W.; Brisbane, C. Risk Mitigation of Plunger-Stopper Displacement Under Low Atmospheric Pressure by Establishing Design Space for Filling-Stoppering Process of Prefilled Syringes: A Design of Experiment (DoE) Approach. J. Pharm. Sci. 2022, 111, 2038–2048. [Google Scholar] [CrossRef]
  14. Evans, C.; Oni, Y.; Paniagua, D.; Franck, J.; Dahlheim, C.; Kulshrestha, A. Stopper Movement and Headspace (Air Bubble Size) Limitations for 2.25 mL Prefilled Syringe. PDA J. Pharm. Sci. Technol. 2023, 77, 472–484. [Google Scholar] [CrossRef] [PubMed]
  15. Gillet, A.; DeMent, A.; Haghedooren, E.; Offermann, E.; Tsakas, R. Nitrogen Dioxide Sterilization is a New State-of-the-Art Sterilization Technology to Control Contamination Risks from Your Prefilled Devices and Combination Drug Products. PDA J. Pharm. Sci. Technol. 2025, 79, 244–245. [Google Scholar] [CrossRef]
  16. Zaikov, G.; Davydov, E.; Pariiskii, G.; Gapanova, I.; Pokholok, T. Interaction of Polymers with Polluted Atmosphere: Nitrogen Oxides; Smithers Rapra: Shropshire, UK, 2009; ISBN 9781847353795. Available online: http://ndl.ethernet.edu.et/bitstream/123456789/58268/1/Gennady%20Zaikov_2009.pdf (accessed on 16 October 2025).
  17. McDonnell, G.; Hansen, J. Nitrogen Dioxide. In Block’s Disinfection, Sterilization, and Preservation; McDonnell, G., Hansen, J., Eds.; Lippincott Williams & Wilkins: Philadelphia, PA, USA, 2020; pp. 726–755. [Google Scholar]
  18. U.S. Food and Drug Administration, Center for Drug Evaluation and Research. Approval Package for: Pavblu (Aflibercept-Ayyh) Injection; U.S. Food and Drug Administration: Silver Spring, MD, USA, 2024. Available online: https://www.accessdata.fda.gov/drugsatfda_docs/nda/2025/761298Orig1s000Approv.pdf (accessed on 16 October 2025).
  19. Richards, T.P.; Lisco, D.; Bianchi, T.; Shahine, G.; Nyugen, H.; Simmons, N.; Dufresne, S.; Opie, D. Nitrogen Dioxide Sterilization Follows Log-Linear Microbial Inactivation Kinetics Using Geobacillus Stearothermophilus Biological Indicators. PDA J. Pharm. Sci. Technol. 2025, 79, 379–390. [Google Scholar] [CrossRef] [PubMed]
  20. ISO 14937:2009; Sterilization of Health Care Products—General Requirements for Characterization of a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices. International Organization for Standardization: Geneva, Switzerland, 2009.
  21. Polyakov, A.A.; Trazhetsets’ka, T.A.; Arbuzov, K.N.; Akhumova, A.A.; Chepurov, K.P. Bactericidal Action of Nitrogen Dioxide on the Vegetative and Sporous Forms of Bacillus anthracis. Mikrobiol. Zh 1963, 24, 43–45. (In Ukrainian) [Google Scholar]
  22. de Souza, F.M.; Kahol, P.K.; Gupta, R.K. Introduction to Polyurethane Chemistry. In Polymer Chemistry for Advanced Applications; American Chemical Society: Washington, DC, USA, 2021; pp. 1–24. [Google Scholar] [CrossRef]
  23. Majka, T.M.; Berkowicz-Płatek, G.; Żukowski, W. Modeling of the Kinetics of Polyoxymethylene Decomposition under Oxidative and Non-Oxidative Conditions. Materials 2021, 14, 2281. [Google Scholar] [CrossRef]
  24. Davydov, E.; Gapanova, I.; Pariiskii, G.; Pokholok, T.; Zaikov, G. Reactivity of Polymers on Exposure to Nitrogen Dioxide. Chem. Chem. Technol. 2010, 4, 281–290. [Google Scholar] [CrossRef]
  25. Brentwood Industries. Top 8 Materials for Medical Device Packaging. 2024. Available online: https://www.brentwoodindustries.com/resources/learning-center/medical/top-8-materials-for-medical-device-packaging/ (accessed on 16 October 2025).
  26. Aprios. PETG: The Go-To Resin for Clarity and Durability in Diverse Applications. 2025. Available online: https://www.aprios.com/insights/petg-the-go-to-resin-for-clarity-and-durability-in-diverse-applications (accessed on 21 November 2025).
  27. Genesis Medical Plastics. Plastics for Implantable Medical Devices. 2024. Available online: https://www.genesismedicalplastics.com/materials/implantable-medical-devices/ (accessed on 16 October 2025).
  28. Foster Corporation. Polysulfones in Healthcare Applications. 2023. Available online: https://www.fostercomp.com/polysulfones-in-healthcare-applications/ (accessed on 21 November 2025).
  29. Covestro AG. High-Quality Plastics for Drug Delivery: Medical-Grade Polycarbonates, PC+ABS Blends and Thermoplastics for Drug Delivery Devices. 2021. Available online: https://solutions.covestro.com/en/highlights/articles/theme/applications/drug-delivery (accessed on 16 October 2025).
  30. Invibio Biomaterial Solutions. Active Implants & Drug Delivery—Application Areas. 2025. Available online: https://www.invibio.com/en/application-areas/medical-devices/ (accessed on 16 October 2025).
  31. Sprovieri, J. New Plastics for Medical Devices. 2018. Available online: https://www.assemblymag.com/articles/94276-new-plastics-for-medical-devices (accessed on 21 November 2025).
  32. Drake Plastics. Life Science & Medical—High-Performance Thermoplastics for Life Science, Medical and Analytical Equipment Applications. 2019. Available online: https://drakeplastics.com/life-science-medical/ (accessed on 16 October 2025).
  33. Solvay Specialty Polymers USA, LLC. ConMed Linvatec Launches First UL-Approved Autoclavable Battery for Powered Surgical Tools. 2013. Available online: https://www.solvay.com/sites/g/files/srpend616/files/tridion/documents/ConMed_Linvatec_Lithium_Battery_System_Uses_AvaSpire%20PAEK_Final_EN.pdf (accessed on 21 November 2025).
  34. Sacks, G.; Shah, V.; Yao, L.; Yan, C.; Shah, D.; Limeta, L.; DeStefano, V. Polyaryletherketones: Properties and Applications in Modern Medicine. Biomed. Technol. 2024, 6, 75–89. [Google Scholar] [CrossRef]
  35. Reno, F.E.; Edwards, C.N.; Jensen, M.B.; Török-Bathó, M.; Esdaile, D.J.; Piché, C.; Triest, M.; Carballo, D. Needle-free nasal delivery of glucagon for treatment of diabetes-related severe hypoglycemia: Toxicology of polypropylene resin used in delivery device. Cutan. Ocul. Toxicol. 2016, 35, 242–247. [Google Scholar] [CrossRef]
  36. ResMart. Applications of Medical-Grade Polypropylene: A Versatile Solution for Healthcare Needs. 2024. Available online: https://resmart.com/news/post/applications-of-medical-grade-polypropylene-a-versatile-solution-for-healthcare-needs1.html (accessed on 21 November 2025).
  37. Nicolau, S.T.; Matzger, A.J. An Evaluation of Resolution, Accuracy, and Precision in FT-IR Spectroscopy. Spectrochim. Acta A Mol. Biomol. Spectrosc. 2024, 319, 124545. [Google Scholar] [CrossRef]
  38. Ekgasit, S.; Padermshoke, A. Optical Contact in ATR/FT-IR Spectroscopy. Appl. Spectrosc. 2001, 55, 1352–1359. [Google Scholar] [CrossRef]
  39. Nisticò, R. Polyethylene Terephthalate (PET) in the Packaging Industry. Polym. Test. 2020, 86, 106707. [Google Scholar] [CrossRef]
  40. Black, J.; Hastings, G. (Eds.) Handbook of Biomaterial Properties; Springer: Boston, MA, USA, 1998; Available online: https://link.springer.com/book/10.1007/978-1-4615-5801-9 (accessed on 15 September 2025).
  41. Brunelle, D.J. Polycarbonates. In Encyclopedia of Polymer Science and Technology; Wiley: Hoboken, NJ, USA, 2006. [Google Scholar] [CrossRef]
  42. Day, M.; Supaphol, T.; Cooney, J.D.; Wiles, D.M. Thermal Degradation of Poly(Aryl-Ether-Ether-Ketone) (PEEK): A Differential Scanning Calorimetry Study. J. Appl. Polym. Sci. 1988, 36, 951–959. [Google Scholar] [CrossRef]
  43. Audouit, J.; Rivière, L.; Dandurand, J.; Louison, A.; Dantas, E.; Lacabanne, C. Thermal, Mechanical and Dielectric Behaviour of Poly(Aryl Ether Ketone) with Low Melting Temperature. J. Therm. Anal. Calorim. 2018, 135, 2147–2157. [Google Scholar] [CrossRef]
  44. Maddah, H.A. Polypropylene as a Promising Plastic: A Review. Am. J. Polym. Sci. 2016, 6, 1–11. [Google Scholar] [CrossRef]
  45. Wunderlich, B. (Ed.) Thermal Analysis of Polymeric Materials; SpringerLink Bücher; Springer: Berlin/Heidelberg, Germany, 2005; ISBN 978-3-540-23629-0. [Google Scholar]
  46. Yi, N.; Chaplin, A.; Grasmeder, J.; Ghita, O. Adaptable Polyaryletherketones (PAEKs) with Competing Crosslinking and Crystallisation Mechanisms. Sci. Rep. 2024, 14, 679. [Google Scholar] [CrossRef]
  47. Alvaredo, A.; Martín, M.; Castell, P.; Guzmán de Villoria, R.; Fernández-Blázquez, J. Non-IsothermalCrystallization Behavior of PEEK/Graphene Nanoplatelets Composites from Melt and Glass States. Polymers 2019, 11, 124. [Google Scholar] [CrossRef]
  48. Liebe, H.J.; Wolfe, V.L.; Howe, D.A. Test of Wall Coatings for Controlled Moist Air Experiments. Rev. Sci. Instrum. 1984, 55, 1702–1705. [Google Scholar] [CrossRef]
  49. Tian, L.; Koshland, C.P.; Yano, J.; Yachandra, V.K.; Yu, T.; Lee, S.C.; Lucas, D. Carbon-Centered Free Radicals in Particulate Matter Emissions from Wood and Coal Combustion. Energy Fuels 2009, 23, 2523–2526. [Google Scholar] [CrossRef]
  50. Green, U.; Shenberger, Y.; Aizenshtat, Z.; Cohen, H.; Ruthstein, S. Exploring the Radical Nature of a Carbon Surface by Electron Paramagnetic Resonance and a Calibrated Gas Flow. J. Vis. Exp. 2014, 51548. [Google Scholar] [CrossRef]
  51. Azevedo, A.M.; Mendonça da Silveira, P.H.P.; Lopes, T.J.; Costa, O.L.B.; Monteiro, S.N.; Veiga-Júnior, V.F.; Silveira, P.C.R.; Cardoso, D.O.; Ben-Hur da Silva Figueiredo, A. Ionizing Radiation and Its Effects on Thermoplastic Polymers: An Overview. Polymers 2025, 17, 1110. [Google Scholar] [CrossRef]
  52. Saint-Gobain Performance Plastics. Effect of Sterilization on the Mechanical Properties of Silicone Rubbers; Saint-Gobain Biopharm: Plymouth, MN, USA, 2025; Available online: https://www.biopharm.saint-gobain.com/sites/hps-mac3-lifesciences-bioprocess/files/effect-of-sterilization-on-the-mechanical-properties-of-silicone-rubbers.pdf (accessed on 14 August 2025).
  53. Wild, N. Everything You Need to Know about Color—Delta E. 2025. Available online: https://blog.hybridhelix.com/everything-you-need-to-know-about-color-delta-e/ (accessed on 21 November 2025).
  54. Sharma, G.; Wu, W.; Dalal, E.N. The CIEDE2000 Color-Difference Formula: Implementation Notes, Supplementary Test Data, and Mathematical Observations. Color Res. Appl. 2005, 30, 21–30. [Google Scholar] [CrossRef]
  55. BASF SE. Antioxidants—Plastic Additives for Plastics and Rubber. Available online: https://plastics-rubber.basf.com/global/en/plastic_additives/products/antioxidants (accessed on 21 November 2025).
  56. Wang, Y.; Ren, H.; Yan, Y.; He, S.; Wu, S.; Zhao, Q. Hindered Phenolic Antioxidants as Heat-Oxygen Stabilizers for HDPE. Polym. Polym. Compos. 2020, 28, 662–673. [Google Scholar] [CrossRef]
  57. Schmid, D.; Hupa, M.; Paavola, M.; Karlström, O. NO2 Absorption in Aqueous Solutions with Phenolic Additives: Hydroquinone and 4-Methylaminophenol Sulfate. Ind. Eng. Chem. Res. 2023, 62, 21007–21012. [Google Scholar] [CrossRef]
  58. Jhamb, S.K.; Goyal, A.; Pandey, A.; Bhowmik, A. A Review of Cytotoxicity Testing Methods and In Vitro Study of Biodegradable Mg-1%Sn-2%HA Composite by Elution Method. J. Mater. Sci. Mater. Med. 2025, 36, 92. [Google Scholar] [CrossRef]
  59. Ikarashi, Y.; Tsuchiya, T.; Nakamura, A. Cytotoxicity of medical materials sterilized with vapour-phase hydrogen peroxide. Biomaterials 1995, 16, 177–183. [Google Scholar] [CrossRef]
  60. ISO 10993-7; Biological Evaluation of Medical Devices—Part 7: Ethylene Oxide Sterilization Residuals. ISO: Geneva, Switzerland, 2008.
  61. Lucas, A.D.; Merritt, K.; Hitchins, V.M.; Woods, T.O.; McNamee, S.G.; Lyle, D.B.; Brown, S.A. Residual ethylene oxide in medical devices and device material. J. Biomed. Mater. Res. B Appl. Biomater. 2003, 66, 548–552. [Google Scholar] [CrossRef]
  62. Athipatla, V. Polysulfones and Their Applications. J. High Sch. Sci. 2023, 7, 2. [Google Scholar] [CrossRef]
  63. Nova Chemicals. Preventing Gas Fading in Polyethylene; Nova Chemicals: Calgary, AB, Canada, 2019; Available online: https://www.novachem.com/wp-content/uploads/NOVA-Chem-Tech-Bulletin-GasFading-FINAL-2019.pdf (accessed on 16 October 2025).
  64. Cai, B.; Wang, Y.; Yang, X.; Li, Y.; Zhai, J.; Zeng, Y.; Ye, J.; Zhu, L.; Fu, T.-M.; Zhang, Q. Rapid Aqueous-Phase Dark Reaction of Phenols with Nitrosonium Ions: Novel Mechanism for Atmospheric Nitrosation and Nitration at Low pH. PNAS Nexus 2024, 3, pgae385. [Google Scholar] [CrossRef]
  65. ASTM D638-22; Standard Test Method for Tensile Properties of Plastics. ASTM International: West Conshohocken, PA, USA, 2022.
  66. ISO 10993-12:2021; Biological Evaluation of Medical Devices—Part 12: Sample Preparation and Reference Materials. International Organization for Standardization: Geneva, Switzerland, 2021.
  67. ASTM E168-16 (2023); Standard Practices for General Techniques of Infrared Quantitative Analysis. ASTM International: West Conshohocken, PA, USA, 2023.
  68. ASTM D3418-21; Standard Test Method for Transition Temperatures of Polymers by Differential Scanning Calorimetry. ASTM International: West Conshohocken, PA, USA, 2021.
  69. ASTM E2655-14 (R20); Standard Guide for Reporting Uncertainty of Test Results and Use of the Term Measurement Uncertainty in ASTM Test Methods. ASTM International: West Conshohocken, PA, USA, 2020.
  70. ASTM D2240-15 (R21); Standard Test Method for Rubber Property—Durometer Hardness. ASTM International: West Conshohocken, PA, USA, 2021.
  71. ASTM E1347-06 (2020); Standard Test Method for Color and Color-Difference Measurement by Tristimulus (Filter) Colorimetry. ASTM International: West Conshohocken, PA, USA, 2020.
  72. García-Robledo, E.; Corzo, A.; Papaspyrou, S. A Fast and Direct Spectrophotometric Method for the Sequential Determination of Nitrate and Nitrite at Low Concentrations in Small Volumes. Mar. Chem. 2014, 162, 30–36. [Google Scholar] [CrossRef]
  73. ISO 10993-5:2009; Biological Evaluation of Medical Devices—Part 5: Tests for In Vitro Cytotoxicity. International Organization for Standardization: Geneva, Switzerland, 2009.
Figure 1. Graph of pressure (mbar) vs. time (minutes) for an example NO2 sterilization cycle.
Figure 1. Graph of pressure (mbar) vs. time (minutes) for an example NO2 sterilization cycle.
Pharmaceuticals 18 01869 g001
Figure 2. NO2 addition reaction.
Figure 2. NO2 addition reaction.
Pharmaceuticals 18 01869 g002
Figure 3. (1) NO2 addition reaction or (2) Electron Transfer Mechanism.
Figure 3. (1) NO2 addition reaction or (2) Electron Transfer Mechanism.
Pharmaceuticals 18 01869 g003
Figure 4. FTIR summary spectra for each polymer sample control, High-Range-exposed.
Figure 4. FTIR summary spectra for each polymer sample control, High-Range-exposed.
Pharmaceuticals 18 01869 g004
Figure 5. EPR spectra of Amorphous Polyester before NO2 exposure (red) and after NO2 exposure (black). The exposed sample exhibited an EPR spectrum with a g-value of 2.0029.
Figure 5. EPR spectra of Amorphous Polyester before NO2 exposure (red) and after NO2 exposure (black). The exposed sample exhibited an EPR spectrum with a g-value of 2.0029.
Pharmaceuticals 18 01869 g005
Figure 6. Example of signal around 3500 cm−1 for the High-Range-exposed sample washed and not washed for Amorphous Polyester.
Figure 6. Example of signal around 3500 cm−1 for the High-Range-exposed sample washed and not washed for Amorphous Polyester.
Pharmaceuticals 18 01869 g006
Figure 7. Possible reaction mechanisms between NO2 and polyester.
Figure 7. Possible reaction mechanisms between NO2 and polyester.
Pharmaceuticals 18 01869 g007
Figure 8. The reaction between the phenolic antioxidant and NO2 to make the phenolic hydroxyl group and HONO.
Figure 8. The reaction between the phenolic antioxidant and NO2 to make the phenolic hydroxyl group and HONO.
Pharmaceuticals 18 01869 g008
Table 1. Summary of Materials, Suppliers, Molecular Structures, and Their Uses in Medical Devices.
Table 1. Summary of Materials, Suppliers, Molecular Structures, and Their Uses in Medical Devices.
MaterialSupplierMolecular Structure/Functional GroupUsed in (Medical Device)
Amorphous PolyesterEnsingerContains ester
(–COO–) linkage; amorphous, transparent structure
Rigid trays for pre-filled syringes, packaging of devices and pharmaceutical products, implants [25,26]
Polysulfone (PSU)EnsingerSulfone (–SO2–) group with aromatic ringsCatheters, access port implants, dialysis cartridges, surgical instrument trays [27,28]
Polycarbonate (PC)CovestroCarbonate
(–O–(C=O)–O–) linkage between aromatic rings
Disposable auto-injectors, portable drug delivery devices, on-body drug delivery devices [29]
PolyEtherEtherKetone (PEEK)EnsingerKetone (C=O) and ether (–O–) groups in aromatic backboneImplantable drug-delivery pumps, on-body e-drug delivery devices, surgical instruments, orthopedic implants [30,31]
PolyArylEtherKetone (PAEK)SyensqoAromatic backbone with ether and ketone groupsStructural components of drug-delivery devices, orthopedic surgical tools, orthopedic implants [32,33,34]
Polypropylene (PP)Ensinger Aliphatic hydrocarbon chain with methyl (–CH3) side groupsSyringes, IV bags, catheters, Intranasal delivery actuator [35,36]
Table 2. Average Achieved Cycle Parameters with Uncertainty Values.
Table 2. Average Achieved Cycle Parameters with Uncertainty Values.
GroupVacuum Level (Torr)NO2
Injection
(mg/L)
Relative Humidity (%RH)NO2
Exposure Time (mm:ss)
Exposure Temperature
(°C)
Exposure
Pressure (Torr)
No. of Pulses
Mid-Range2015.4 ± 0.179.615:0024.43 ± 0.115926
High-Range2015.3 ± 0.179.615:0124.54 ± 0.0759212
Within
Tolerance (Y/N)
YYYYYYY
Table 3. Summary of average glass transition temperatures for each amorphous polymer sample comparing unexposed control, Mid-Range-exposed, and High-Range-exposed. Literature-reported glass transition temperature ranges are included for reference.
Table 3. Summary of average glass transition temperatures for each amorphous polymer sample comparing unexposed control, Mid-Range-exposed, and High-Range-exposed. Literature-reported glass transition temperature ranges are included for reference.
PolymerLiterature
Reported Tg (°C)
Measured Tg
Control (°C)
Measured Tg Mid-Range-Exposed (°C)Measured Tg High-Range-Exposed (°C)
Amorphous Polyester67–125105.1 ± 0.1103.7 ± 0.5103 ± 2
Polysulfone (PSU)180–230 191.3 ± 2.0189.2 ± 2.5187.4 ± 3.0
Polycarbonate (PC)140–155146.1 ± 0.9146.7 ± 0.4146.3 ± 0.7
Table 4. Summary of average melting transition temperatures for each semi-crystalline polymer sample comparing unexposed control, Mid-Range-exposed, and High-Range-exposed. Literature-reported melting transition temperature ranges are included for reference.
Table 4. Summary of average melting transition temperatures for each semi-crystalline polymer sample comparing unexposed control, Mid-Range-exposed, and High-Range-exposed. Literature-reported melting transition temperature ranges are included for reference.
PolymerLiterature
Reported Tm (°C)
Measured Tm
Control (°C)
Measured Tm Mid-Range-Exposed (°C)Measured Tm High-Range-Exposed (°C)
PolyEtherEtherKetone (PEEK)340–350343.0 ± 0.7343 ± 2343.5 ± 0.2
PolyArylEtherKetone (PAEK)300–400344 ± 4341 ± 3342 ± 6
Polypropylene (PP)160–166159.1 ± 0.6159.9 ± 0.3160 ± 3
Table 5. Average percent difference in crystallinity for Mid-Range-exposed and High-Range-exposed from control average percent crystallinity for semi-crystalline polymer samples.
Table 5. Average percent difference in crystallinity for Mid-Range-exposed and High-Range-exposed from control average percent crystallinity for semi-crystalline polymer samples.
PolymerMid-Range-Exposed Percent
Difference in Crystallinity (%)
High-Range-Exposed Percent
Difference in Crystallinity (%)
PolyEtherEtherKetone (PEEK)−0.6 ± 2−0.1 ± 0.6
PolyArylEtherKetone (PAEK)−1.5 ± 51.6 ± 3
Polypropylene (PP)0.4 ± 0.4−0.35 ± 0.09
Table 6. Summary of average true yield strength, true strength at fracture, and true percent elongation at fracture for each polymer comparing unexposed control samples, Mid-Range-exposed samples, and High-Range-exposed samples.
Table 6. Summary of average true yield strength, true strength at fracture, and true percent elongation at fracture for each polymer comparing unexposed control samples, Mid-Range-exposed samples, and High-Range-exposed samples.
PolymerGroupYield Strength (MPa) Strength at
Fracture (MPa)
Elongation at Fracture (%)
Amorphous PolyesterControl49.9 ± 0.9112 ± 492 ± 2
Mid-Range-Exposed50.7 ± 0.4110 ± 690 ± 3
High-Range-Exposed50.1 ± 0.6110 ± 490 ± 2
Polysulfone (PSU)Control89.8 ± 1.056.2 ± 3.614 ± 3
Mid-Range-Exposed90.2 ± 0.359.1 ± 2.014 ± 2
High-Range-Exposed88.0 ± 2.455.5 ± 1.813 ± 1
Polycarbonate (PC)Control65.6 ± 0.3117 ± 1268 ± 5
Mid-Range-Exposed64.9 ± 0.5112 ± 1965 ± 10
High-Range-Exposed64.7 ± 0.6120 ± 1369 ± 6
PolyEtherEtherKetone (PEEK)Control116 ± 181 ± 316 ± 2
Mid-Range-Exposed117 ± 182 ± 116.3 ± 0.9
High-Range-Exposed117.2 ± 0.380 ± 213.7 ± 0.8
PolyArylEtherKetone (PAEK)Control101 ± 1094 ± 739 ± 10
Mid-Range-Exposed95 ± 590 ± 330 ± 8
High-Range-Exposed112 ± 6112 ± 650 ± 5
Polypropylene (PP)Control39 ± 441 ± 237 ± 8
Mid-Range-Exposed37 ± 542 ± 240 ± 6
High-Range-Exposed35 ± 338 ± 3 36 ± 5
Table 7. Summary of average Shore D Hardness measurements for each polymer comparing control, Mid-Range-exposed, and High-Range-exposed.
Table 7. Summary of average Shore D Hardness measurements for each polymer comparing control, Mid-Range-exposed, and High-Range-exposed.
MaterialControlMid-Range-ExposedHigh-Range-Exposed
Amorphous Polyester78 ± 078 ± 179 ± 1
Polysulfone (PSU)85 ± 085 ± 086 ± 0
Polycarbonate (PC)85 ± 084 ± 185 ± 0
PolyEtherEtherKetone (PEEK)89 ± 189 ± 190 ± 0
PolyArylEtherKetone (PAEK)88 ± 089 ± 088 ± 0
Polypropylene (PP)75 ± 075 ± 076 ± 0
Table 8. Summary of average color change (ΔE) for each polymer comparing control with Mid-Range-Exposed and with High-Range-Exposed.
Table 8. Summary of average color change (ΔE) for each polymer comparing control with Mid-Range-Exposed and with High-Range-Exposed.
MaterialΔE Between
Control and
Mid-Range-Exposed
ΔE Between
Control and
High-Range-Exposed
Amorphous Polyester1.1 ± 0.32.0 ± 1.1
Polysulfone (PSU)1.0 ± 0.30.5 ± 0.1
Polycarbonate (PC)3.1 ± 0.14.1 ± 0.1
PolyEtherEtherKetone (PEEK)1.2 ± 0.11.9 ± 0.2
PolyArylEtherKetone (PAEK)0.8 ± 0.11.8 ± 0.1
Polypropylene (PP)3.7 ± 0.54.8 ± 0.1
Table 9. Values of nitrate or nitrite surface residuals on each polymer across the control, Mid-Range-Exposed, and High-Range-Exposed samples.
Table 9. Values of nitrate or nitrite surface residuals on each polymer across the control, Mid-Range-Exposed, and High-Range-Exposed samples.
MaterialNitrate Levels (µg/cm2)Nitrite Levels (µg/cm2)
Amorphous Polyester90.8 ± 3.10.065 ± 0.004
Polysulfone (PSU)70.7 ± 2.70.029 ± 0.004
Polycarbonate (PC)35.1 ± 0.4<LOQ
PolyEtherEtherKetone (PEEK)18.8 ± 0.90.073 ± 0.000
PolyArylEtherKetone (PAEK)16.5 ± 0.10.076 ± 0.005
Polypropylene (PP)0.412 ± 0.0700.265 ± 0.027
Table 10. Cytotoxicity grades for each polymer.
Table 10. Cytotoxicity grades for each polymer.
Material Control Grade Mid-Range-Exposed Grade High-Range-Exposed Grade
Amorphous Polyester 0 -- * 0
Polysulfone (PSU)0-- *0
Polycarbonate (PC)0 0 **0
PEEK 0 -- * 0
PAEK0-- * 0
Polypropylene 0 0 1
* Mid-Range-Exposed samples were not tested for these materials since the High-Range-Exposed samples were graded 0. ** The control and High-Range-Exposed samples for Polycarbonate had to be retested, in the meantime, the Mid-Range-Exposed sample was also tested.
Table 11. EPR testing material information and testing results.
Table 11. EPR testing material information and testing results.
TestMaterialMaterial
Supplier
Samples TestedSignal Found
InitialAmorphous
Polyester
(Tecadur MT TR)
Ensinger1Yes
Retest, Dry Ice Storage1No
Retest, Dry Ice Storage21 Yes
1 No
Retest, Dry Ice StorageSemi-Crystalline
Polyethylene Terephthalate Glycol (PETG)
E&T Plastics3No
Crystalline
Polyester (Tecapet PET Natural)
Ensinger3No
Table 12. ΔE Thresholds for Perceptible Color Change.
Table 12. ΔE Thresholds for Perceptible Color Change.
ΔE ValuePerceptibilityDescription
<1Not perceptibleColor difference is imperceptible to the human eye
1–2SlightOnly noticeable to trained observers or under controlled conditions
2–3SmallVisible when colors are compared side-by-side
3–5ModerateNoticeable under normal viewing conditions
5–10SignificantClear color difference; easily perceived
>10MajorColors appear distinctly different
Note: These thresholds are based on the CIEDE2000 color difference formula, which is considered the most accurate for human visual perception [54].
Table 13. List of Polymeric Material Suppliers.
Table 13. List of Polymeric Material Suppliers.
MaterialProduct NameSupplier
Amorphous PolyesterTecadur MT TREnsinger (Nufringen, Germany)
Polysulfone (PSU)Tecason S Natural Udel P1700Ensinger (Nufringen, Germany)
Polycarbonate (PC)Makrolon 2458Covestro (Leverkusen, Germany)
PolyEtherEtherKetone (PEEK)Tecapeek NaturalEnsinger (Nufringen, Germany)
PolyArylEtherKetone (PAEK)Avaspire AV651Syensqo (Brussels, Belgium)
Polypropylene (PP)Tecapro MTEnsinger (Nufringen, Germany)
Table 14. Material Exposure Cycle Conditions.
Table 14. Material Exposure Cycle Conditions.
GroupVacuum Level (Torr)NO2
Injection (mg/L)
Relative
Humidity (%RH)
NO2
Exposure Time (mm:ss)
Exposure Temperature (°C)Exposure
Pressure (Torr)
No. of Pulses
Mid-Range20158015:00245906
High-Range20158015:002459012
Tolerance±10±1.5±10%±00:10±1.5±30±0
Table 15. Cytotoxicity Test Scoring.
Table 15. Cytotoxicity Test Scoring.
GradeReactivityConditions of all Cultures
0NoneDiscrete intracytoplasmic granules, no cell lysis, no reduction in cell growth.
1SlightNot more than 20% of the cells are round, are loosely attached and without intracytoplasmic granules, or show changes in morphology; occasional lysed cells are present; only slight growth inhibition observable.
2MildNot more than 50% of the cells are round or devoid of
intracytoplasmic granules; no extensive cell lysis; not more than 50% growth inhibition observed.
3ModerateNot more than 70% of the cell layers contain rounded cells or are lysed; cell layers not completely destroyed, but more than 50% growth inhibition was observed.
4SevereNearly complete or complete destruction of the cell layers.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Ray, N.; Schexnayder, J.D.; Ashfaq, A.; Sarwahrdy, N.; Lisco, D.; Abdollahi Khabisi, M.; Bateman, T.; Sadler, T.; Opie, D.; Al-Sheikhly, M. Drug Delivery Device Design and Compatibility with Nitrogen Dioxide Gas Sterilization. Pharmaceuticals 2025, 18, 1869. https://doi.org/10.3390/ph18121869

AMA Style

Ray N, Schexnayder JD, Ashfaq A, Sarwahrdy N, Lisco D, Abdollahi Khabisi M, Bateman T, Sadler T, Opie D, Al-Sheikhly M. Drug Delivery Device Design and Compatibility with Nitrogen Dioxide Gas Sterilization. Pharmaceuticals. 2025; 18(12):1869. https://doi.org/10.3390/ph18121869

Chicago/Turabian Style

Ray, Noelle, Julia Diane Schexnayder, Aiysha Ashfaq, Nusrat Sarwahrdy, Delaney Lisco, Minufar Abdollahi Khabisi, Trevor Bateman, Tom Sadler, David Opie, and Mohamad Al-Sheikhly. 2025. "Drug Delivery Device Design and Compatibility with Nitrogen Dioxide Gas Sterilization" Pharmaceuticals 18, no. 12: 1869. https://doi.org/10.3390/ph18121869

APA Style

Ray, N., Schexnayder, J. D., Ashfaq, A., Sarwahrdy, N., Lisco, D., Abdollahi Khabisi, M., Bateman, T., Sadler, T., Opie, D., & Al-Sheikhly, M. (2025). Drug Delivery Device Design and Compatibility with Nitrogen Dioxide Gas Sterilization. Pharmaceuticals, 18(12), 1869. https://doi.org/10.3390/ph18121869

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop