Next Article in Journal
Novel Europium-Grafted 3D Covalent Organic Framework for Selective and Sensitive Fluorescence-Enhanced Detection of Levofloxacin
Next Article in Special Issue
Portable Measurement System for the Characterization of Capacitive Field-Effect Sensors
Previous Article in Journal
An Algorithm for Mining the Living Habits of Elderly People Living Alone Based on AIoT
Previous Article in Special Issue
PULSE: A Fast Portable Unit for Lab-on-Site Electrochemistry
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

A Highly Portable Smartphone-Based Capillary Electrophoresis with Capacitively Coupled Contactless Conductivity Detection

1
School of Sensing Science and Engineering, School of Electronic Information and Electrical Engineering, Shanghai Jiao Tong University, Shanghai 200240, China
2
Student Innovation Center, Shanghai Jiao Tong University, Shanghai 200240, China
3
State Key Laboratory of Microbial Metabolism, School of Life Sciences and Biotechnology, Shanghai Jiao Tong University, Shanghai 200240, China
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
Sensors 2025, 25(7), 2303; https://doi.org/10.3390/s25072303
Submission received: 6 March 2025 / Revised: 28 March 2025 / Accepted: 3 April 2025 / Published: 4 April 2025
(This article belongs to the Special Issue Sensors from Miniaturization of Analytical Instruments (2nd Edition))

Abstract

:
Work has rarely been reported on a highly portable smartphone-based capillary electrophoresis (CE) with capacitively coupled contactless conductivity detection (C4D). Herein, a highly portable phone-based CE (130 mm × 190 × 70 mm, 1.4 kg) with C4D and Bluetooth communication, as well as user-interface software, was developed for portable analysis. To demonstrate the device, six metal ions were selected as model analytes for verification and successfully applied to the detection of ions in tap water. The analytical performance highlighted that the runs and data analysis of the CE-C4D device could be controlled via the user interface based on smartphones. Furthermore, the experiments showed that (i) the linear ranges of six metal ions were between 6 and 1500 μmol/L with a correlation coefficient of more than 0.9934; (ii) the limit of detection (LOD) values were within 1.84–4.33 μmol/L; (iii) the intra-day deviations of migration time and peak area were 2.40–5.24% and 0.75–2.82% (n = 5), respectively. Although the LOD is not the most optimal among current portable devices, the results still indicated the satisfactory analytical performance and potential of the developed device, software, and method for portable separation and quantitation of analytes from various fields.

1. Introduction

Portable analytical devices contribute to ensuring the living conditions of everyone and enhancing their well-being [1]. They not only eliminate the need for complex sample storage and transportation but also enable on-site assays for prompt decision-making [2,3,4,5,6]. For these reasons, several portable analytical techniques have been developed in the last two decades, such as lateral chromatography [7,8], colloidal gold bioassay [9], electrochemical sensors [10], portable/handheld NIR [11], and Raman spectrophotometers [12,13], as well as electrophoresis titration sensors [14,15].
Capillary electrophoresis (CE) has the advantages of high separation efficiency, short analysis time, low chemical consumption, and easy installation, operation, and maintenance [16], while capacitively coupled contactless conductivity detection (C4D) also has the merits of small size, low power consumption, high versatility, and easy construction and operation [17,18,19]. Thus, the combination of CE and C4D is particularly suitable for developing portable analytical devices. In 2001, Kappes et al. [20] first proposed a portable CE device with amperometric, potentiometric, and conductometric detection, with dimensions of 340 × 175 × 175 mm and a weight of 7.5 kg. After six years, Kubáň et al. [21] proposed an improved CE-C4D device with a size of 310 × 220 × 260 mm and a reduced limit of detection (LOD). Since then, portable CE devices have received increasing attention from research teams. In 2013, Mai et al. [22] designed a portable CE-C4D device with a size of 450 × 350 × 150 mm and a weight of 8 kg. One year later, Kobrin et al. [23] and Nguyen et al. [24] designed a CE device and a C4D controlled via a computer, respectively. In 2016, Greguš et al. [25] constructed a small CE-C4D device with dimensions of 200 × 330 × 170 mm and a weight of less than 5 kg. Koenka et al. [26] presented a thermostatted dual-channel portable CE-C4D device with dimensions of 520 × 340 × 180 mm and a weight of less than 15 kg. Opekar et al. [27] designed an original dual-channel CE-C4D device on a supporting board. In 2018, Fuiko et al. [28] proposed a CE-C4D water quality analyzer integrated in a 19 inch rack. In 2024, Li et al. [29] constructed a compact and high-performance setup of CE-C4D with a size of 245 × 220 × 95 mm. Despite the use of LCD tablets or computers for signal acquisition and analysis, the developed CE-C4D devices still suffered from unsatisfactory portability for real on-site analysis.
To enhance the portability and practicality of the CE-C4D device for field applications, it is essential to focus on miniaturization, operational simplification, and intelligent data processing. Leveraging the ubiquity and user-friendliness of smartphones, these devices can serve as effective platforms for data recording and processing, eliminating the dependency on traditional computers. Notably, smartphones have already demonstrated their potential in advancing bioassay technologies. In 2017, Kanakasabapathy et al. [30] proposed an alternative to blood sample flow cytometry based on smartphones for testing HIV/AIDS. In 2020, Calabretta et al. [31] designed a portable biosensor coupled with paper and smartphone devices for detecting bacterial ATP, Nie et al. [32] demonstrated an electrochemi-luminescence sensor with zinc-doped MoS2 quantum dots (QDs), and Xu et al. [33] developed a smartphone-based point-of-care testing (POCT) system for detection of metal ions. In 2022, Zheng et al. [34] presented a smartphone-based multilayered paper-based microfluidic analysis device for simultaneously determining glucose and uric acid in the blood, and Fiore et al. [35] proposed a smartphone-assisted electrochemical device capable of quantitative analysis of tyrosine in serum samples. In 2023, Yu et al. [36] developed a colorimetric and electrochemical dual-mode smartphone-sensing platform based on MOF-818 nanozyme for in situ detection of hydrogen peroxide and hydrogen sulfide released by living cells. Despite the widespread adoption of smartphones in advancing bioassay technologies, the development of smartphone-based portable CE-C4D devices remains unexplored, primarily due to the technical challenges associated with processing complex electrophoretogram data on mobile platforms and achieving seamless integration between CE-C4D hardware and smartphone software.
Herein, a highly portable CE-C4D device with a smartphone and user interface was designed for real portable analysis. As a proof of concept, six metal ions were selected as model analytes to validate the performance of the developed device. As shown below, the experiments were conducted to systematically validate the device, software and method of CE-C4D. Finally, the developed device was successfully used for the actual analysis of drinking water. All of these demonstrate the potential of the developed device, software, and method for portable separation and quantitative analysis of a wide range of analytes in the field.

2. CE-C4D Device

2.1. Design of CE-C4D

Figure 1 shows the schematic diagram of the portable smartphone-based CE-C4D with three parts of CE separation and cleaning (A), the online C4D (B), and the smartphone (C). The first part consists of a capillary (a), a sample vial (b), an anode vial (c) and a cathode vial (d) for background electrolyte (BGE), a micro diaphragm pump (e), and a high-voltage power supply (f), etc. The second part consists of a C4D-sensing section with an excitation electrode and a detection electrode (g) and a control and signal processing module (h) with an alternating current (AC) source, an excitation amplifier, an I/V converter, an AC amplifier and rectifier, a baseline regulator, a direct-current (DC) amplifier and filter, a 24-bit analog to digital converter (ADC), a microcontroller unit (MCU), and a Bluetooth module, etc. The last part is the smartphone, which includes an Android phone with the user-interface software for communicating with the CE-C4D via Bluetooth.
The CE separation and cleaning part is shown in Figure 1A. Herein, the high voltage electrode and the anode end of the capillary (a) are inserted into the anode BGE vial (c), and when injecting the sample, the anode end of the capillary is inserted from the anode BGE vial into the sample vial (b). The grounding electrode of the power supply and the cathode end of the capillary are inserted into the cathode BGE vial (d) through a rubber plug. In order to achieve automatic injection, the inlet pipe of the diaphragm pump (e) is inserted into the cathode BGE vial through the rubber plug, and the outlet pipe is kept outside. Using a high-voltage power supply (f) with controllable and positive voltage output as the CE power supply.
Figure 1B shows the online C4D part. The sensing section (g) is the core component of the C4D. Its performance depends not only on its own parameters but also on its position on the capillary. The experiments reveal that the closer the sensing section is to the cathode, the better the ion separation is. The closer the sensing section is to the anode, the faster the ion peak appears. Thus, the sensing section was fixed at 10 cm of the capillary near the cathode based on the preliminary test results. The control and signal-processing module (h) integrates an excitation module, a control and detection module, and a Bluetooth module. The excitation module consists of an AC source and an excitation amplifier. The signal of the AC source is amplified by the excitation amplifier and loaded to the excitation electrode of the sensing section. The control and detection module consists of an I/V converter, an AC amplifier and rectifier, a baseline regulator, a DC amplifier and filter, a 24-bit ADC, and a MCU. The detection electrode converts the received current into a voltage by the I/V converter, amplifies the signal by the amplifier, and finally rectifies it into a DC signal by the rectifier. The DC signal is connected to the DC amplifier and filter after baseline adjustment. The amplified signal is digitized by the 24-bit high-speed ADC and sent to the MCU. The micro diaphragm pump is driven by PWM waves generated by the MCU, and at the same time, feedback pulse signals (6 pulses per cycle) are sent back to the MCU to accurately obtain the pumping volume of the diaphragm pump, thereby improving the accuracy of sample injection. The integrated Bluetooth module is responsible for wireless communication with a smartphone (C) to monitor electrophoresis signal in real-time.
The smartphone is equipped with the user-interface software, which can not only control the run of the CE-C4D device, such as controlling the high-voltage power supply and the diaphragm pump, but also receive real-time data from the online C4D detector, draw the electrophoretic graphs, and perform data processing, as shown in Figure 1C.

2.2. Fabrication of CE-C4D Device

Figure 2 shows the real photograph of the portable smartphone-based CE-C4D device, including top view (A), capillary winding diagram (B), sensor board diagram (C), T-tube diagram (D), and internal diagram (E). The developed CE-C4D device mainly includes a capillary (a), a sample vial (b), an anode BGE vial (c), a cathode BGE vial (d), a micro diaphragm pump (e), a high-voltage power supply (f), a sensor board (g), a main board (h), two potentiometers for baseline adjustment and amplification (i), a T-tube (j), a lithium battery (k), and a metal shell (m).
The capillary (a) (Ruifeng Chromatograohinc Co., Ltd., Handan, China) with a length of 60 cm (effective length of 50 cm), an outer diameter of 365 μm, and an inner diameter of 75 μm was used. It is threaded through the C4D electrode on the sensor board (g), and after spiral winding in the isolation space above the device by relying on the elasticity of the capillary, both ends protrude from the metal shell (m), as shown in Figure 2B,C. One end of the capillary is inserted into the anode BGE vial (c) or the sample vial (b), and the other end is inserted through the rubber plug into one end of the upper outlet of the T-tube (j) and finally extended from the lower outlet of the T-tube and inserted into the cathode BGE vial (d), as shown in Figure 2A,D. Due to the spiral winding of the capillary inside the device, the wound part of the capillary can be pulled out or retracted into the device, thus achieving the purpose of inserting into the anode BGE vial and the sample vial at different positions. Herein, the vials are all Duchenne tubules with a diameter of 6 mm and a length of 30 mm. They are made of glass, can hold approximately 350 μL of solution, and are secured to both sides of the device by adhesive-covered elastic tube clamps. The T-tube (j) is made of polyethylene, and the lower outlet of the T-tube is connected to the cathode BGE vial through a latex sleeve, as shown in Figure 2D. One end of the upper outlet of the T-tube is inserted with the capillary and sealed with the rubber plug, and the other end is connected to the inlet tube of the micro diaphragm pump (e) (H012, Nidec Corporation, Kyoto, Japan) through the latex sleeve to ensure the sealing of the connection part. At the same time, the outlet tube of the diaphragm pump, which operates at a vacuum power of 3 W, is kept outside the device to discharge air, as shown in Figure 2E. When the diaphragm pump is working, it will extract air from the cathode BGE vial, causing the internal pressure to drop below atmospheric pressure, thereby relying on the atmospheric pressure difference between the two ends of the capillary to inject the solution from the anode end into the capillary. Herein, the T-tube and the inlet tube of the diaphragm pump can also serve as air pressure buffers. The high-voltage power supply (f) (KDHM-Q-12S15000P-VI, Xi’an Koso Electronic Technology Co., Ltd., Xi’an, China) has controllable voltage output, with an output voltage range of 0 ~ +15 kV and a volume of only 128 × 40 × 24 mm.
The C4D sensor board (g) includes a sensing section, an AC source, and a voltage detection module. It is vertically fixed on the inner side of the device, with the sensing section located about 10 cm away from the cathode end of the capillary. The electrodes of the sensing section are made of copper pipes with an inner diameter of 0.4 mm and an outer diameter of 1 mm. In order to improve the sensitivity of C4D, the electrode parameters were selected as a length of 15 mm and a gap of 1 mm according to our previous work [37]. There is a grounded copper plate between the two electrodes as a Faraday shield to eliminate the influence of stray capacitance, as shown in Figure 2C. The AC source used a Wien bridge oscillator base on LF351 to generate a 40 kHz sine wave (Figures S1 and S2), and the excitation amplifier used an operational amplifier AD711 to increase the amplitude of the excitation signal to 10 V. The voltage detection module includes an I/V converter and an amplifier and rectifier. The I/V converter used an operational amplifier OPA637 with a 1 MΩ feedback resistance to convert the picked current into voltage. In addition, another operational amplifier OPA637 is used to further amplify the AC voltage signal. At last, the rectifier was relied on a diode peak detection circuit to convert the voltage signal into a DC signal.
The main board (h) includes a baseline regulator, a DC amplifier and filter, a 24-bit ADC, a MCU, a Bluetooth module, and a power management module. The baseline regulator is obtained via a subtraction circuit based on the OPA227 chip. The DC amplifier and filter are implemented using an active second-order low-pass filtering circuit based on two OPA228 chips, with a signal magnification of 1–500 times. The baseline and magnification can be adjusted by adjusting the potentiometers (i), as shown in Figure 2A. The use of this filter that can both amplify ion peaks and act as a filter has improved the sensitivity of the C4D. The ADC uses the ADS1251 to achieve high-precision conversion of analog signal to digital signal. The MCU uses a low-power STM32L053 microprocessor chip with a built-in DA converter for controlling the output of the high-voltage power supply. A low-power Bluetooth module based on TI-CC2541 is used for wireless communication with the smartphone. The device is powered by a 12 V, 3000 mAh lithium battery (k) and can also be connected to an external adapter for charging and power supply. When the adapter is connected, the lithium battery automatically charges. The power management module efficiently distributes power to all components, including the sensor board, main board, and high-voltage module. During high-voltage electrophoresis operation at 15 kV, the device consumes approximately 490 mA of current, enabling the lithium battery to theoretically sustain continuous operation for up to 6 h. Furthermore, the device is housed in an aluminum metal shell (m), with the circuit ground connected to the shell to effectively shield against external signal interference.

2.3. Phone-Based User Software

A popular smartphone (Xiaomi 6X, Shanghai, China) was used as the user interface. The mobile software is based on Android system, using “activity + fragment + service” components, developed in Java language, and will use the Bluetooth, WiFi, or data traffic function of the mobile phone in the running process. It contains two user interfaces, Bluetooth connectivity and main interface, which also included four sub-interfaces: Home, Settings, Monitor, and Data, as shown in Figure 3. The Settings interface can be used to input or select default electrophoresis voltage and electrophoresis time, control the on/off (“Set”/“HV stop”) of the high voltage, and control the diaphragm pump to complete sample injection and capillary flushing, as shown in Figure 3B. Data visualization in real-time is completed via the Monitor interface, as shown in Figure 3C. The Data interface is developed for data optimization, storage, and calculation, as shown in Figure 3D. Herein, the electropherogram saved in the interface of Monitor can be plotted on the upper display window by clicking the “Draw” button, and then the start and end time of the analysis and the cut-off frequency (Fc) of the filter can be set. By clicking the “Optimize” button, the built-in algorithm can obtain the fitted baseline and the electropherogram after removing baseline and noise within the start and end time, which are plotted on the upper electropherogram and the lower display window, respectively. The data of plotted the fitted baseline can also be saved by clicking the above “Save Data” button. Afterwards, the peak area of the ion peak to be calculated can be obtained by setting the approximate time when the peak appears and clicking the “Calculate” button. If necessary, the data and graph of the electropherogram after removing baseline and noise can be saved in file format on the smartphone by clicking the below “Save Data” button and “Save Image” button, respectively.
The built-in algorithms mainly include the baseline estimation and denoising with sparsity (BEADS) [38] and the Gaussian-peak-fitting algorithm. The BEADS algorithm does not need to carry the distribution characteristics of the baseline, making it more versatile for various electropherograms. It can simultaneously perform baseline estimation and denoising, obtaining baseline and ion peaks. However, the BEADS algorithm can only handle sparse and non-overlapping ion peaks, which cannot handle the overlapping peaks that occur during electrophoresis. The Gaussian-peak-fitting algorithm can accurately separate and reconstruct ion peaks that overlap to a certain extent, making up for the shortcomings of the BEADS algorithm in practical applications. Specifically, the built-in algorithm can automatically identify peaks and valleys based on the derivative of ion peaks and combine with the baseline obtained by the BEADS algorithm to determine whether there are overlapping peaks. If there are overlapping peaks, Gaussian-peak-fitting algorithm is used to separate the overlapping parts and reconstruct them into complete ion peaks. Thus, this algorithm can be used to calculate the area of these peaks and obtain the concentration of the corresponding substances. In addition, due to the insufficient support libraries of Java in fields such as data processing, signal analysis, and numerical computation, the built-in algorithm was implemented on a computer using Python 3.7 and deployed on a cloud server. When processing data, users can set the start and end time of the analysis based on the distribution of the ion peaks to avoid system peaks, set the cut-off frequency of the filter in the BEADS algorithm, and then click the “Optimize” button. The setting parameters and raw data will be uploaded to the cloud server, and the processing results received from the cloud server will be displayed on the smartphone. This process is usually completed within 5 s.

3. Materials and Methods

3.1. Instruments

An electronic balance (Mettler-Toledo, Greifensee, Switzerland) was used for weighing the sample. The electric blast dryer (Shanghai YIHENG Technical Co., Ltd., Shanghai, China) was used for drying the chemicals. The commercial contact conductivity meter was DDS-307 with DJS-1C conductivity electrode (REX, Shanghai, China). A YD1940A high-voltage digital voltmeter (Changzhou Yangzi Electronics Co., Ltd., Changzhou, China) was used for measuring high voltage.

3.2. Chemicals

The chemicals are of analytical grade and can be used without further purification. Standard solutions of the six cations (Mg2+, K+, Na+, Ca2+, Zn2+, Mn2+) were prepared from their sulfates or chlorides. Magnesium chloride was purchased from Bide Pharmatech Co., Ltd. (Shanghai, China). Potassium chloride, sodium chloride, calcium chloride, and zinc chloride were bought from Aladdin Biochemical Technology Co., Ltd. (Shanghai, China). Lactic acid (Lac) was obtained from Titan Technology Co., Ltd. (Shanghai, China). β-Ala was purchased from Sinopharm Chemical Reagent Co., Ltd. (Shanghai, China). All solutions were prepared with deionized water (resistivity of 18.2 MΩ·cm at 25 °C) from SZ-93 automatic distiller (Yarong, Shanghai, China). Tap water was obtained from the laboratory building on campus.

3.3. Procedure of CE-C4D

First, Lac of 2.815 g and β-Ala of 2.2273 g were weighted, respectively, and were put into a volumetric flask of 1 L to prepare the Lac-β-Ala BGE [39] with a concentration of 25 mmol/L and a pH of 3.6. Store this BGE in a reagent flask for later testing.
Then, stock solutions of 10 mL with concentrations of 4000 μmol/L were prepared for each of the six cations mentioned above. In the experiment, a certain amount of the stock solutions was diluted to 2~1000 folds to obtain sample or mixed-sample solutions with the desired concentrations of each cation for a later test.
Before a run, first, open the developed device and mobile software and connect the phone to the device via Bluetooth. Second, check the rubber plug and latex sleeve to ensure sealing, and click the “Flush” button to flush the capillary. Prior to use, the new capillaries were rinsed sequentially with 0.1 mol/L NaOH, deionized water, and BGE for 10 min each. In the separation process, the capillaries were treated with deionized water and BGE for 2 min each. After rinsing, loosen the rubber plug to balance the air pressure inside and outside the T-tube and then reinsert it. Third, transfer the anode end of the capillary from the anode BGE vial to the sample vial. Fourth, click the “Injection” button to activate the diaphragm pump, which will automatically stop after running for one second. Fifthly, wait for two seconds to achieve the optimal injection time of three seconds (as shown in Figure S3), then release the rubber plug of the T-tube to balance the air pressure and reinsert it to complete the sample injection. Sixthly, move the anode end of the capillary back from the sample vial to the anode BGE vial. Finally, input the required electrophoresis voltage and time in the mobile software settings interface, and click the “Set” button to initiate the CE-C4D analysis.
The smartphone can receive the real-time data via Bluetooth and draw electropherogram in the Monitor interface of the mobile software during electrophoresis. At the beginning, users can adjust the baseline to the appropriate position via the baseline potentiometer and wait for the appearance of the ion peaks. After electrophoresis is completed, save the experimental data in the Monitor interface, then perform baseline estimation and denoising and calculate the peak area in the Data interface. If a quantitative calibration curve for a certain ion has been pre-stored, the built-in algorithm can also directly calculate the content of that ion in the sample.

4. Results and Discussion

4.1. Feasibility of Smartphone-Based CE-C4D

Figure 4 shows the developed device (A) and the core steps, such as Bluetooth scanning (B), Bluetooth connecting (C), and controlling the power supply using a smartphone with a 15 kV output (D) of the CE run. The developed CE-C4D device has total dimensions of 130 × 190 × 70 mm (Figure S4), weighs 1.4 kg, and can be held in one hand. At the beginning of the CE run, the smartphone should first search the CE-C4D device which was marked with blue name and connected through Bluetooth to the CE-C4D device, as shown in panel B. The successfully connected signal on the phone and the indicator light of the device helped the user to make sure the connection was a success (Panel C). Then, after the sample injection, the CE-C4D run started, and the real-time data could be visualized on the smartphone, as shown in Figure 3C. After the testing, the user could also use the software developed to conduct the data processing (Figure 3D).
The control of the power supply was performed by the sub-interface of Setting as shown in Figure 4D. A high-voltage meter was used to measure the output of the power supply controlled by the smartphone. The real outputs of the power supply were consistent with the setting voltage, further indicating the feasibility of smartphone-based CE runs.

4.2. Analytical Performance

Figure 5 presented the electropherograms for BGE with six single cations and their mixtures. The electropherograms of the mixtures showed that K+, Mg2+, Mn2+, and Zn2+ could achieve baseline separation within 400 s, with retention times of 199 s, 284 s, 298 s, and 367 s, respectively. It is noteworthy that the reduced ion concentration after mixing enhances electrophoretic separation efficiency, achieving baseline separation of Mg2+ and Mn2+ in Figure 5g, despite the partial overlap observed in their individual runs (Figure 5d,e). The retention times of Ca2+ and Na+ were 257 s and 267 s, respectively, but they have not fully achieved baseline separation. Baseline separation of Ca2+ and Na+ can be achieved by optimizing the separation conditions, such as reducing the capillary inner diameter and simultaneously replacing the BGE (as demonstrated in Figure S5). However, a smaller capillary inner diameter may lead to extended cleaning times and increased susceptibility to clogging. Herein, the Gaussian-peak-fitting algorithm was used to separate and reconstruct ion peaks that overlap to a certain extent, achieving satisfactory results (as shown in Figure S6) and avoiding the aforementioned issues. Overall, the electrophoretogram had a good peak shape, and each peak detected for single metal ion solution could match the peak detected for their mixture. The experimental results fully verified that the CE-C4D device had good selectivity for six ions.
Table 1 shows the linear regression equation, coefficients of determination (R2), linear range, and LOD of six ions. The results revealed that R2 reached 0.9934, and the linear range depended on the species of ions. Herein, the mixture was diluted gradually to obtain the ratio of peak height and noise (S/N) at S/N = 3, and the correspondent concentration of each cation was defined as the LOD for the corresponding cation (Table 1). Accordingly, the LODs for K+, Ca2+, Na+, Mg2+, Mn2+, and Zn2+ were 1.9, 1.8, 3.3, 2.7, 3.6, and 4.3 μmol/L, respectively, implying a good LOD.
Table 2 showed the intra-day and inter-day deviations of the migration time and peak area of the six ions in the smartphone-based CE-C4D runs, revealing good stability. Accordingly, the intra-day and inter-day deviations of migration time (MT) of six ions were 2.40–5.24% and 4.17–5.53%, respectively. The intra-day and inter-day deviations of peak area (PA) of six ions were 0.75–2.82% and 1.02–4.15%, respectively.

4.3. Application

The safety of drinking water can be evaluated by detecting the contents of K+, Ca2+, Na+, and Mg2+ in tap water. Herein, a certain amount of tap water from the campus was diluted eight times with BGE to prepare a sample solution, and CE-C4D experiments were performed according to the previously described procedure. Figure 6 showed the electrophoretograms for BGE with diluted tap water via the portable smartphone-based CE-C4D device. The electrophoretogram of tap water showed that the retention times of K+, Ca2+, Na+, and Mg2+ were 203 s, 266 s, 276 s, and 290 s, respectively, which were basically the same as the retention times of the corresponding ions in Figure 5.
Table 3 showed the results of eightfold dilution of tap water via the developed CE-C4D device and ion chromatography (IC). The CE-C4D results showed some deviation from IC measurements, mainly attributed to significant baseline drift in the electrophoretograms and limitations in the data-processing algorithm. Notably, K+ exhibited a relatively high recovery rate of 128%, likely influenced by matrix interference and algorithmic inaccuracies. However, the recovery rates of Ca2+, Na+, and Mg2+ ranged between 95% and 106%, demonstrating close agreement with IC results and confirming the reliability of the portable smartphone-based CE-C4D device.

4.4. Merits and Disadvantages

The developed CE-C4D device has the following merits as compared with the existing CE-C4D devices (Table 4). First, the developed CE-C4D device was of high portability in contrast to the existing CE-C4D devices. Its volume was less than 1/3 of the volume of the existed CE-C4D portable devices, and its weight was less than 1/4 of the weight of the existing CE-C4D portable devices. Second, the developed CE-C4D device was controlled by a smartphone in contrast to the existing CE-C4D devices. A software developed for control of the smartphone-based CE-C4D device facilitates the operation of CE run. Third, the use of the BEADS and Gaussian peak fitting algorithm that can to some extent handle overlapping peaks has improved the applicability of the device. All these made the developed CE-C4D device have good portability, facility, and practicability.
While the developed device demonstrates promising performance, its LOD still requires improvement compared to high-performance CE-C4D systems [25,29]. Additionally, the migration time repeatability is currently limited by ambient temperature variations and the direct contact between the capillary and the metal shell. Future development efforts should focus on these areas to achieve trace-level detection capabilities and enhanced reproducibility.

5. Conclusions

The following conclusions were achieved: First, a highly portable smartphone-based CE-C4D device as well as user-interface software was developed for portable analysis. Second, the BEADS algorithm was combined with the Gaussian-peak-fitting algorithm to accurately separate and reconstruct the corresponding partially overlapping ion peaks, avoiding the problems of long capillary cleaning time and easy blockage caused by using smaller inner-diameter capillaries. Third, the experiments of the CE-C4D device were systemically performed, which well demonstrated the validity of the device, software, and method of CE-C4D. Finally, the developed CE-C4D device was successfully applied to the real analysis of tap water, and the results were basically consistent with the IC. The developed CE-C4D device had the following merits of good linearity, low LOD, good repetition, and high accuracy. Moreover, the developed CE-C4D device has better portability in contrast to the existed CE-C4D portable devices. All these indicate that the developed device, software, and method have the potential for portable separation and quantification of analytes in various fields.

Supplementary Materials

The supporting information can be downloaded at https://www.mdpi.com/article/10.3390/s25072303/s1, Figure S1: Schematic diagram of the C4D; Figure S2: Output and sensitivity characteristic curves of the C4D with different excitation frequencies; Figure S3: Electrophoretograms with different injection times; Figure S4: Left- and right-side photograph of the CE-C4D device; Figure S5: Electrophoretogram for BGE with mixed solution of six ions; Figure S6: The accuracy of fitting-ion peaks with different degrees of overlap; Table S1: Ion chromatography results of tap water.

Author Contributions

Conceptualization, Z.T. and Q.Z.; methodology, Z.T. and Q.Z.; software, Z.T.; validation, X.D. and Y.C.; formal analysis, Y.C.; investigation, Q.Z.; resources, L.F.; data curation, Z.T.; writing—original draft preparation, Z.T.; writing—review and editing, W.L.; visualization, Y.W.; supervision, C.C. and W.L.; project administration, W.L.; funding acquisition, Q.Z. and C.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by National Natural Science Foundation of China (No. 22104082 and No. 22074091).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The original contributions presented in this study are included in the article/Supplementary Material. Further inquiries can be directed to the corresponding author(s).

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Van Schepdael, A.V. Capillary Electrophoresis as a Simple and Low-Cost Analytical Tool for Use in Money-Constrained Situations. TrAC Trends Anal. Chem. 2023, 160, 116992. [Google Scholar] [CrossRef]
  2. Saar-Reismaa, P.; Erme, E.; Vaher, M.; Kulp, M.; Kaljurand, M.; Mazina-Šinkar, J. In Situ Determination of Illegal Drugs in Oral Fluid by Portable Capillary Electrophoresis with Deep UV Excited Fluorescence Detection. Anal. Chem. 2018, 90, 6253–6258. [Google Scholar] [CrossRef] [PubMed]
  3. Pan, J.-Z.; Fang, P.; Fang, X.-X.; Hu, T.-T.; Fang, J.; Fang, Q. A Low-Cost Palmtop High-Speed Capillary Electrophoresis Bioanalyzer with Laser Induced Fluorescence Detection. Sci. Rep. 2018, 8, 1791. [Google Scholar] [CrossRef]
  4. Saar-Reismaa, P.; Brilla, C.-A.; Leiman, K.; Kaljurand, M.; Vaher, M.; Kulp, M.; Mazina-Šinkar, J. Use of a Newly-Developed Portable Capillary Electrophoresis Analyser to Detect Drugs of Abuse in Oral Fluid: A Case Study. Talanta 2020, 211, 120662. [Google Scholar] [CrossRef] [PubMed]
  5. Zhang, M.; Smejkal, P.; Bester, N.; Robertson, J.C.; Atia, M.A.; Townsend, A.T.; Guijt, R.M.; Breadmore, M.C. Inexpensive Portable Capillary Electrophoresis Instrument for Monitoring Zinc(II) in Remote Areas. J. Chromatogr. A 2022, 1668, 462895. [Google Scholar] [CrossRef]
  6. Kaljurand, M.; Mazina-Šinkar, J. Portable Capillary Electrophoresis as a Green Analytical Technology. TrAC Trends Anal. Chem. 2022, 157, 116811. [Google Scholar] [CrossRef]
  7. Xiao, R.; Lu, L.; Rong, Z.; Wang, C.; Peng, Y.; Wang, F.; Wang, J.; Sun, M.; Dong, J.; Wang, D.; et al. Portable and Multiplexed Lateral Flow Immunoassay Reader Based on SERS for Highly Sensitive Point-of-Care Testing. Biosens. Bioelectron. 2020, 168, 112524. [Google Scholar] [CrossRef]
  8. Tran, V.; Walkenfort, B.; König, M.; Salehi, M.; Schlücker, S. Rapid, Quantitative, and Ultrasensitive Point-of-Care Testing: A Portable SERS Reader for Lateral Flow Assays in Clinical Chemistry. Angew. Chem. Int. Ed. 2018, 58, 442–446. [Google Scholar] [CrossRef]
  9. Hassantabar, F.; Zorriehzahra, M.J.; Firouzbakhsh, F.; Thompson, K.D. Development and Evaluation of Colloidal Gold Immunochromatography Test Strip for Rapid Diagnosis of Nervous Necrosis Virus in Golden Grey Mullet (Chelon aurata). J. Fish Dis. 2021, 44, 783–791. [Google Scholar] [CrossRef]
  10. Raziq, A.; Kidakova, A.; Boroznjak, R.; Reut, J.; Öpik, A.; Syritski, V. Development of a Portable MIP-Based Electrochemical Sensor for Detection of SARS-CoV-2 Antigen. Biosens. Bioelectron. 2021, 178, 113029. [Google Scholar] [CrossRef]
  11. Correia, R.M.; Andrade, R.; Tosato, F.; Nascimento, M.T.; Pereira, L.L.; Araújo, J.B.S.; Pinto, F.E.; Endringer, D.C.; Padovan, M.P.; Castro, E.V.R.; et al. Analysis of Robusta Coffee Cultivated in Agroforestry Systems (AFS) by ESI-FT-ICR MS and Portable NIR Associated with Sensory Analysis. J. Food Compos. Anal. 2020, 94, 103637. [Google Scholar] [CrossRef]
  12. Deidda, R.; Sacre, P.-Y.; Clavaud, M.; Coïc, L.; Avohou, H.; Hubert, P.; Ziemons, E. Vibrational Spectroscopy in Analysis of Pharmaceuticals: Critical Review of Innovative Portable and Handheld NIR and Raman Spectrophotometers. TrAC Trends Anal. Chem. 2019, 114, 251–259. [Google Scholar] [CrossRef]
  13. Hara, R.; Ishigaki, M.; Kitahama, Y.; Ozaki, Y.; Genkawa, T. Use of the Product of Mean Intensity Ratio (PMIR) Technique for Discriminant Analysis of Lycopene-Rich Vegetable Juice Using a Portable NIR-Excited Raman Spectrometer. Food Chem. 2018, 241, 353–357. [Google Scholar] [CrossRef]
  14. Wang, C.; Zhang, Q.; Liu, X.; Li, G.; Kong, H.; Khan, M.I.; Xiao, H.; Wang, Y.; Liu, W.; Cao, C. Double Inner Standard Plot Model of an Electrophoresis Titration Chip for a Portable and Green Assay of Protein Content in Milk. Lab A Chip 2019, 19, 484–492. [Google Scholar] [CrossRef]
  15. Chen, L.; Zhang, Q.; Liu, W.; Xiao, H.; Liu, X.; Fan, L.; Wang, Y.; Li, H.; Cao, C. A Facile Thermometer-like Electrophoresis Titration Biosensor for Alternative miRNA Assay via Moving Reaction Boundary Chip. Biosens. Bioelectron. 2021, 171, 112676. [Google Scholar] [CrossRef]
  16. Poboży, E.; Trojanowicz, M. Application of Capillary Electrophoresis for Determination of Inorganic Analytes in Waters. Molecules 2021, 26, 6972. [Google Scholar] [CrossRef] [PubMed]
  17. Kubáň, P.; Hauser, P.C. Fundamental Aspects of Contactless Conductivity Detection for Capillary Electrophoresis. Part I: Frequency Behavior and Cell Geometry. Electrophoresis 2004, 25, 3387–3397. [Google Scholar] [CrossRef]
  18. Kubáň, P.; Hauser, P.C. Fundamental Aspects of Contactless Conductivity Detection for Capillary Electrophoresis. Part II: Signal-to-noise Ratio and Stray Capacitance. Electrophoresis 2004, 25, 3398–3405. [Google Scholar] [CrossRef]
  19. Brito-Neto, J.G.A.; Fracassi da Silva, J.A.; Blanes, L.; do Lago, C.L. Understanding Capacitively Coupled Contactless Conductivity Detection in Capillary and Microchip Electrophoresis. Part 1. Fundamentals. Electroanalysis 2005, 17, 1198–1206. [Google Scholar] [CrossRef]
  20. Kappes, T.; Galliker, B.; Schwarz, M.A.; Hauser, P.C. Portable Capillary Electrophoresis Instrument with Amperometric, Potentiometric and Conductometric Detection. TrAC Trends Anal. Chem. 2001, 20, 133–139. [Google Scholar] [CrossRef]
  21. Kubáň, P.; Nguyen, H.T.A.; Macka, M.; Haddad, P.R.; Hauser, P.C. New Fully Portable Instrument for the Versatile Determination of Cations and Anions by Capillary Electrophoresis with Contactless Conductivity Detection. Electroanalysis 2007, 19, 2059–2065. [Google Scholar] [CrossRef]
  22. Mai, T.D.; Pham, T.T.T.; Pham, H.V.; Sáiz, J.; Ruiz, C.G.; Hauser, P.C. Portable Capillary Electrophoresis Instrument with Automated Injector and Contactless Conductivity Detection. Anal. Chem. 2013, 85, 2333–2339. [Google Scholar] [CrossRef] [PubMed]
  23. Kobrin, E.; Lees, H.; Fomitšenko, M.; Kubáň, P.; Kaljurand, M. Fingerprinting Postblast Explosive Residues by Portable Capillary Electrophoresis with Contactless Conductivity Detection. Electrophoresis 2014, 35, 1165–1172. [Google Scholar] [CrossRef] [PubMed]
  24. Nguyen, T.A.H.; Pham, T.N.M.; Doan, T.T.; Ta, T.T.; Sáiz, J.; Nguyen, T.Q.H.; Hauser, P.C.; Mai, T.D. Simple Semi-Automated Portable Capillary Electrophoresis Instrument with Contactless Conductivity Detection for the Determination of β-Agonists in Pharmaceutical and Pig-Feed Samples. J. Chromatogr. A 2014, 1360, 305–311. [Google Scholar] [CrossRef]
  25. Greguš, M.; Foret, F.; Kubáň, P. Portable Capillary Electrophoresis Instrument with Contactless Conductivity Detection for On-Site Analysis of Small Volumes of Biological Fluids. J. Chromatogr. A 2016, 1427, 177–185. [Google Scholar] [CrossRef]
  26. Koenka, I.J.; Küng, N.; Kubáň, P.; Chwalek, T.; Furrer, G.; Wehrli, B.; Müller, B.; Hauser, P.C. Thermostatted Dual-channel Portable Capillary Electrophoresis Instrument. Electrophoresis 2016, 37, 2368–2375. [Google Scholar] [CrossRef]
  27. Opekar, F.; Tůma, P. Dual-Channel Capillary Electrophoresis for Simultaneous Determination of Cations and Anions. J. Chromatogr. A 2016, 1446, 158–163. [Google Scholar] [CrossRef]
  28. Fuiko, R.; Saracevic, E.; Koenka, I.J.; Hauser, P.C.; Krampe, J. Capillary Electrophoresis for Continuous Nitrogen Quantification in Wastewater Treatment Processes. Talanta 2019, 195, 366–371. [Google Scholar] [CrossRef]
  29. Li, L.; Song, Y.-P.; Ren, D.-D.; Li, T.-X.; Gao, M.-H.; Zhou, L.; Zeng, Z.-C.; Pu, Q. A Compact and High-Performance Setup of Capillary Electrophoresis with Capacitively Coupled Contactless Conductivity Detection (CE-C4D). Analyst 2024, 149, 3034–3040. [Google Scholar] [CrossRef]
  30. Kanakasabapathy, M.K.; Pandya, H.J.; Draz, M.S.; Chug, M.K.; Sadasivam, M.; Kumar, S.; Etemad, B.; Yogesh, V.; Safavieh, M.; Asghar, W.; et al. Rapid, Label-Free CD4 Testing Using a Smartphone Compatible Device. Lab A Chip 2017, 17, 2910–2919. [Google Scholar] [CrossRef]
  31. Calabretta, M.M.; Álvarez-Diduk, R.; Michelini, E.; Roda, A.; Merkoçi, A. Nano-Lantern on Paper for Smartphone-Based ATP Detection. Biosens. Bioelectron. 2020, 150, 111902. [Google Scholar] [CrossRef] [PubMed]
  32. Nie, Y.; Zhang, X.; Zhang, Q.; Liang, Z.; Ma, Q.; Su, X. A Novel High Efficient Electrochemiluminescence Sensor Based on Reductive Cu(I) Particles Catalyzed Zn-Doped MoS2 QDs for HPV 16 DNA Determination. Biosens. Bioelectron. 2020, 160, 112217. [Google Scholar] [CrossRef] [PubMed]
  33. Xu, Z.; Liu, Z.; Xiao, M.; Jiang, L.; Yi, C. A Smartphone-Based Quantitative Point-of-Care Testing (POCT) System for Simultaneous Detection of Multiple Heavy Metal Ions. Chem. Eng. J. 2020, 394, 124966. [Google Scholar] [CrossRef]
  34. Zheng, J.; Zhu, M.; Kong, J.; Li, Z.; Jiang, J.; Xi, Y.; Li, F. Microfluidic Paper-Based Analytical Device by Using Pt Nanoparticles as Highly Active Peroxidase Mimic for Simultaneous Detection of Glucose and Uric Acid with Use of a Smartphone. Talanta 2022, 237, 122954. [Google Scholar] [CrossRef]
  35. Fiore, L.; De Lellis, B.; Mazzaracchio, V.; Suprun, E.; Massoud, R.; Goffredo, B.M.; Moscone, D.; Arduini, F. Smartphone-Assisted Electrochemical Sensor for Reliable Detection of Tyrosine in Serum. Talanta 2022, 237, 122869. [Google Scholar] [CrossRef]
  36. Yu, K.; Li, M.; Chai, H.; Liu, Q.; Hai, X.; Tian, M.; Qu, L.; Xu, T.; Zhang, G.; Zhang, X. MOF-818 Nanozyme-Based Colorimetric and Electrochemical Dual-Mode Smartphone Sensing Platform for in Situ Detection of H2O2 and H2S Released from Living Cells. Chem. Eng. J. 2023, 451, 138321. [Google Scholar] [CrossRef]
  37. Liu, W.; Liang, Z.; Wang, Y.; Cao, J.; Zhang, Q.; Liu, X.; Wang, Y.; Cao, C. A Facile Online Multi-Gear Capacitively Coupled Contactless Conductivity Detector for an Automatic and Wide Range Monitoring of High Salt in HPLC. Analyst 2022, 147, 496–504. [Google Scholar] [CrossRef]
  38. Ning, X.; Selesnick, I.W.; Duval, L. Chromatogram Baseline Estimation and Denoising Using Sparsity (BEADS). Chemom. Intell. Lab. Syst. 2014, 139, 156–167. [Google Scholar] [CrossRef]
  39. Tang, M.; Xu, J.; Xu, Z. Simultaneous Determination of Metal Ions by Capillary Electrophoresis with Contactless Conductivity Detection and Insights into the Effects of BGE Component. Microchem. J. 2019, 147, 857–862. [Google Scholar] [CrossRef]
Figure 1. Schematic diagram of the portable smartphone-based CE-C4D with three parts of CE separation and cleaning (A), the online C4D (B), and the smartphone (C). Panel A: (a) capillary, (b) sample vial, (c) anode BGE vial, (d) cathode BGE vial, (e) micro diaphragm pump, (f) high-voltage power supply, (g) C4D-sensing section, and (h) control and signal-processing module.
Figure 1. Schematic diagram of the portable smartphone-based CE-C4D with three parts of CE separation and cleaning (A), the online C4D (B), and the smartphone (C). Panel A: (a) capillary, (b) sample vial, (c) anode BGE vial, (d) cathode BGE vial, (e) micro diaphragm pump, (f) high-voltage power supply, (g) C4D-sensing section, and (h) control and signal-processing module.
Sensors 25 02303 g001
Figure 2. Real photograph of the portable smartphone-based CE-C4D device, including top view (A), capillary winding diagram (B), sensor board diagram (C), T-tube diagram (D), and internal diagram (E). Herein, (a) capillary, (b) sample vial, (c) anode BGE vial, (d) cathode BGE vial, (e) micro diaphragm pump, (f) high-voltage power supply, (g) sensor board, (h) main board, (i) potentiometers for baseline adjustment and amplification, (j) T-tube, (k) lithium battery, and (m) metal shell.
Figure 2. Real photograph of the portable smartphone-based CE-C4D device, including top view (A), capillary winding diagram (B), sensor board diagram (C), T-tube diagram (D), and internal diagram (E). Herein, (a) capillary, (b) sample vial, (c) anode BGE vial, (d) cathode BGE vial, (e) micro diaphragm pump, (f) high-voltage power supply, (g) sensor board, (h) main board, (i) potentiometers for baseline adjustment and amplification, (j) T-tube, (k) lithium battery, and (m) metal shell.
Sensors 25 02303 g002
Figure 3. User interface of conductivity detection mobile software for controlling of developed CE-C4D device of Bluetooth scanning (A), CE condition setting (B), CE run (C), and data processing (D).
Figure 3. User interface of conductivity detection mobile software for controlling of developed CE-C4D device of Bluetooth scanning (A), CE condition setting (B), CE run (C), and data processing (D).
Sensors 25 02303 g003
Figure 4. The developed device (A) and the core steps, such as Bluetooth scanning (B), Bluetooth connecting (C), controlling the power supply with the smartphone with a 15 kV output (D) of the CE run. Herein, the YD1940A high-voltage digital voltmeter was used to verify the high-voltage output set on the smartphone.
Figure 4. The developed device (A) and the core steps, such as Bluetooth scanning (B), Bluetooth connecting (C), controlling the power supply with the smartphone with a 15 kV output (D) of the CE run. Herein, the YD1940A high-voltage digital voltmeter was used to verify the high-voltage output set on the smartphone.
Sensors 25 02303 g004
Figure 5. Electrophoretograms for BGE with six single ions and their mixture. a—100 μmol/L K+, b—100 μmol/L Ca2+, c—100 μmol/L Na+, d—100 μmol/L Mg2+, e—100 μmol/L Mn2+, f—200 μmol/L Zn2+, g—mixture. BGE: 25 mmol/L Lac-β-Ala (pH 3.6). Capillary: length of 60 cm (effective length of 50 cm), o.d. of 365 μm and i.d. of 75 μm. Separation voltage: 15 kV. Injection time: 3 s. Excitation source: peak–peak voltage of 20 V and frequency of 40 kHz. DC signal magnification: 100.
Figure 5. Electrophoretograms for BGE with six single ions and their mixture. a—100 μmol/L K+, b—100 μmol/L Ca2+, c—100 μmol/L Na+, d—100 μmol/L Mg2+, e—100 μmol/L Mn2+, f—200 μmol/L Zn2+, g—mixture. BGE: 25 mmol/L Lac-β-Ala (pH 3.6). Capillary: length of 60 cm (effective length of 50 cm), o.d. of 365 μm and i.d. of 75 μm. Separation voltage: 15 kV. Injection time: 3 s. Excitation source: peak–peak voltage of 20 V and frequency of 40 kHz. DC signal magnification: 100.
Sensors 25 02303 g005
Figure 6. Electrophoretograms for BGE with diluted tap water via the portable smartphone-based CE-C4D device, including the baseline fitting (A), and peak area calculation (B). The dark gray line, red line, and gray line represent the electrophoretic curve, the fitted baseline, and the reconstructed Gaussian peak, respectively. BGE: 25 mmol/L Lac-β-Ala (pH 3.6). Capillary: length of 60 cm (effective length of 50 cm), o.d. of 365 μm and i.d. of 75 μm. Separation voltage: 15 kV. Injection time: 3 s. Excitation source: peak–peak voltage of 20 V and frequency of 40 kHz. DC signal magnification: 100.
Figure 6. Electrophoretograms for BGE with diluted tap water via the portable smartphone-based CE-C4D device, including the baseline fitting (A), and peak area calculation (B). The dark gray line, red line, and gray line represent the electrophoretic curve, the fitted baseline, and the reconstructed Gaussian peak, respectively. BGE: 25 mmol/L Lac-β-Ala (pH 3.6). Capillary: length of 60 cm (effective length of 50 cm), o.d. of 365 μm and i.d. of 75 μm. Separation voltage: 15 kV. Injection time: 3 s. Excitation source: peak–peak voltage of 20 V and frequency of 40 kHz. DC signal magnification: 100.
Sensors 25 02303 g006
Table 1. Linear regression equation, coefficients of determination, linear range, and LOD of six ions.
Table 1. Linear regression equation, coefficients of determination, linear range, and LOD of six ions.
IonsRegression EquationR2Linear Range (μmol/L)LOD
(μmol/L)
K+y = 0.1958 × x + 1.51120.99956–8001.9
Ca2+y = 0.3742 × x + 21.6020.99346–10001.8
Na+y = 0.2081 × x + 4.45880.995310–10003.3
Mg2+y = 0.3261 × x + 4.76850.998910–10002.7
Mn2+y = 0.1868 × x + 5.56350.999012–15003.6
Zn2+y = 0.1845 × x + 0.64900.998215–15004.3
Table 2. Intra-day and inter-day deviations of migration time and peak area of six ions in the phone-based CE-C4D runs.
Table 2. Intra-day and inter-day deviations of migration time and peak area of six ions in the phone-based CE-C4D runs.
IonsRSD of Intra-Day (%) (n = 5)RSD of Inter-Day (%) (n = 5)
Migration TimeAreaMigration TimeArea
K+2.402.534.683.34
Ca2+5.241.385.531.87
Na+4.960.755.341.02
Mg2+3.052.034.522.69
Mn2+3.832.705.224.15
Zn2+3.542.824.173.03
Table 3. Results of eightfold dilution of tap water via the developed CE-C4D device and IC.
Table 3. Results of eightfold dilution of tap water via the developed CE-C4D device and IC.
IonsDeveloped CE-C4DIC (Table S1)Recovery
Peak AreaDetected (μmol/L)Detected (μmol/L)
K+6.163423.7618.53128%
Ca2+70.9018131.75127.30103%
Na+57.9601257.09241.41106%
Mg2+19.885246.3648.7695%
Table 4. Comparison of developed CE-C4D device with existing pCE-C4D devices.
Table 4. Comparison of developed CE-C4D device with existing pCE-C4D devices.
DevicesSize
(mm3)
Weight (kg)Phone
Based
LOD
(μmol/L)
RSD
(MT)
RSD
(PA)
Recovery
(%)
AnalytesRef.
CE-C4D prototype340 × 175 × 1757.5No0.7–5.6---Inorganic ions, organic acids[20]
New portable CE-C4D310 × 220 × 260-No0.2–1.0-0.82–6.9-Inorganic cations and anions[21]
Suitcase CE-C4D450 × 350 × 1508No1.5–17.0 0.6–2.493–116Inorganic cations and anions[22]
Purpose-built CE-C4D270 × 246 × 124-No3.7–35.70.3–0.63.6–9.9-Inorganic cations and anions[23]
semi-auto CE-C4D400 × 280 × 2106No2.1–3.4 a0.8–0.94.1–6.286.6–113.9β-agonists[24]
Portable CE-C4D200 × 330 × 170<5No0.04–0.360.42–1.062.85–5.05-Inorganic and organic anions[25]
Dual channel CE520 × 340 × 180<15No2.8–18-6.3–11.6-Inorganic cations and anions[26]
On-board CE200 × 250 b-No6.9–10.61.5–3.52.0–5.295.2–114.5Inorganic cations and anions[27]
SIA-CE-C4D c19 inch rack-No0.5–3.40.43–0.960.52–1.3490–108NO2, NO3, NH4+[28]
Compact CE-C4D245 × 220 × 95-No0.029–0.41--97.1–103.7Lanthanide rare-earth ions[29]
Phone-based CE-C4D130 × 190 × 701.4Yes1.8–4.32.40–5.530.75–4.1595–106(K+:128)Inorganic cationsThis work
a The LOD of this was transformed from the determination of β-agonists with a unit of ppm. b the device was set on a 200 × 250 board and the depth was not mentioned. c SIA was the brief name of sequential injection analysis.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Tao, Z.; Zhang, Q.; Cao, Y.; Duan, X.; Wu, Y.; Fan, L.; Cao, C.; Liu, W. A Highly Portable Smartphone-Based Capillary Electrophoresis with Capacitively Coupled Contactless Conductivity Detection. Sensors 2025, 25, 2303. https://doi.org/10.3390/s25072303

AMA Style

Tao Z, Zhang Q, Cao Y, Duan X, Wu Y, Fan L, Cao C, Liu W. A Highly Portable Smartphone-Based Capillary Electrophoresis with Capacitively Coupled Contactless Conductivity Detection. Sensors. 2025; 25(7):2303. https://doi.org/10.3390/s25072303

Chicago/Turabian Style

Tao, Zhimin, Qiang Zhang, Yiren Cao, Xunjie Duan, Yuyang Wu, Liuyin Fan, Chengxi Cao, and Weiwen Liu. 2025. "A Highly Portable Smartphone-Based Capillary Electrophoresis with Capacitively Coupled Contactless Conductivity Detection" Sensors 25, no. 7: 2303. https://doi.org/10.3390/s25072303

APA Style

Tao, Z., Zhang, Q., Cao, Y., Duan, X., Wu, Y., Fan, L., Cao, C., & Liu, W. (2025). A Highly Portable Smartphone-Based Capillary Electrophoresis with Capacitively Coupled Contactless Conductivity Detection. Sensors, 25(7), 2303. https://doi.org/10.3390/s25072303

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop