Next Article in Journal
Automatic Suppression Method for Water Surface Glints Using a Division of Focal Plane Visible Polarimeter
Previous Article in Journal
Surface Defect Detection of Bearing Rings Based on an Improved YOLOv5 Network
Previous Article in Special Issue
Comparison of Chemical and Electrochemical Approaches to Abacavir Oxidative Stability Testing
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor

1
Department of Chemical Engineering, University of Utah, Salt Lake City, UT 84112, USA
2
Department of Chemistry, Snow College, Ephraim, UT 84627, USA
*
Author to whom correspondence should be addressed.
Sensors 2023, 23(17), 7444; https://doi.org/10.3390/s23177444
Submission received: 9 August 2023 / Revised: 23 August 2023 / Accepted: 23 August 2023 / Published: 26 August 2023
(This article belongs to the Collection Electrochemical Sensors and Platforms: Design and Application)

Abstract

:
Formaldehyde is a known human carcinogen and an important indoor and outdoor air pollutant. However, current strategies for formaldehyde measurement, such as chromatographic and optical techniques, are expensive and labor intensive. Low-cost gas sensors have been emerging to provide effective measurement of air pollutants. In this study, we evaluated eight low-cost electrochemical formaldehyde sensors (SFA30, Sensirion®, Staefa, Switzerland) in the laboratory with a broadband cavity-enhanced absorption spectroscopy as the reference instrument. As a group, the sensors exhibited good linearity of response (R2 > 0.95), low limit of detection (11.3 ± 2.07 ppb), good accuracy (3.96 ± 0.33 ppb and 6.2 ± 0.3% N), acceptable repeatability (3.46% averaged coefficient of variation), reasonably fast response (131–439 s) and moderate inter-sensor variability (0.551 intraclass correlation coefficient) over the formaldehyde concentration range of 0–76 ppb. We also systematically investigated the effects of temperature and relative humidity on sensor response, and the results showed that formaldehyde concentration was the most important contributor to sensor response, followed by temperature, and relative humidity. The results suggest the feasibility of using this low-cost electrochemical sensor to measure formaldehyde concentrations at relevant concentration ranges in indoor and outdoor environments.

1. Introduction

Globally, formaldehyde is the most abundant carbonyl [1]. It is also a known human carcinogen [2,3]. More than 25 million people in the US are exposed to formaldehyde levels that exceed the cancer risk threshold, making it the biggest driver of cancer risk among hazardous air pollutants [4]. Hazardous air pollutants are defined by the US EPA as compounds that known to cause cancer and other serious health impacts [5]. Formaldehyde also plays a key role in ozone and secondary particulate matter formation—both of which are associated with significant adverse health effects [6,7,8]. Should formaldehyde sources be better characterized and controlled, it will reduce health risks to the local citizens and help address critical air pollution challenges: PM2.5 and ozone.
The existing measurement techniques for formaldehyde can be broadly categorized as satellite-based, spectrometric, and optical techniques. Satellite-based estimates of average annual formaldehyde concentration have been made at a grid resolution of approximately 5 × 5 km2 [9]. This resolution is excellent for monitoring regional and global trends, but is unsuitable for identifying local emission sources. Chromatographic technologies, such as ion mobility spectrometry (IMS) and gas chromatography-mass spectrometry (GC-MS) are gold standards for aldehyde identification [10]. In practice, samples are usually collected in canisters, which are analyzed by contract labs at a high cost per sample and results can be delayed for weeks to months. Although portable systems exist, they tend to be prohibitively expensive. Optical techniques include differential optical absorption spectroscopy (DOAS), infrared spectrophotometry, Raman spectroscopy, fluorescence, and colorimetric [1,11]. These methods provide sensitive measurements for a wide range of VOCs, but they are labor-intensive and require daily maintenance. Overall, these strategies for formaldehyde measurement are either expensive or labor-intensive, which hamper the ability to identify formaldehyde sources.
Low-cost gas sensors have been emerging in recent years, which can complement measurements from regulatory or research-grade measurements [12,13,14,15,16,17,18,19]. These sensors can be easy to deploy and provide real-time measurement with higher spatial coverage than conventional measurements [18]. Common low-cost formaldehyde sensors can be categorized as semiconductor sensors and electrochemical sensors [16,18,20]. Semiconductor sensors can detect formaldehyde by enabling a redox reaction and then measuring the resistance change [21]. These sensors have high sensitivity and stability. However, some of them, such as metal oxide sensors, require high operating temperature (>100 °C) and have low selectivity [20,21,22]. On the other hand, electrochemical sensors detect formaldehyde by enabling an electrochemical reaction and then measuring the current change [23]. These sensors can have high selectivity and accuracy. In addition, they consume less power compared to metal oxide sensors because the sensors generally work under room or ambient temperature. However, the signals of electrochemical sensors can be influenced by temperature and relative humidity (RH) [24,25].
Some laboratory studies have evaluated the performance of low-cost formaldehyde sensors. Chattopadhyay et al. [26] evaluated the performance of two electrochemical sensors and three metal oxide sensors in a laboratory chamber over the concentration range of 10–800 ppb, a temperature range of 22–50 °C, and a RH range of 8–85%. Deng et al. [27] developed a metal oxide sensor based on hierarchical flower-like CuO nanostructure and evaluated the sensor response to a formaldehyde concentration range of 50–1000 ppb. Gautam et al. [28] developed a Si-chip assisted MOS/SiNWs nanocomposite-based sensor and evaluated the sensor response to a formaldehyde concentration range of 0.01–1000 ppm. Li et al. [29] synthesized SnO2 microspheres and evaluated the sensor response to a formaldehyde concentration range of 1–500 ppm. Hu et al. [30] reported batch fabrication of formaldehyde sensors based on LaFeO3 film and evaluated sensor response over a concentration range of 0.05–1 ppm. However, these studies have some limitations. First, most of the studies focused on sensor fabrication and provided limited sensor performance metrics. Most of the studies reported the linearity of response [28,29,30], but only a few studies reported limit of detection (LOD) [27,30], sensor accuracy [26], or sensor repeatability [28,29]. Second, most of the studies focused on formaldehyde concentration ranges that were higher than typical indoor (~17 ppb) [31] or ambient (~3 ppb) [32] levels in the United States. In addition, previous studies have reported the influence of temperature and RH on the performance of electrochemical sensors [25,33,34,35,36]. However, some studies report contradictory results regarding the effect of RH [37], and many studies only report qualitative results [30,38,39,40]. The effects of temperature and RH on electrochemical formaldehyde sensors have not been systematically investigated, which is crucial to understanding sensor performance in the ambient environment.
This study aims to evaluate the laboratory performance of eight low-cost electrochemical formaldehyde sensors (Sensirion SFA30®, Staefa, Switzerland) over an environmentally relevant concentration range of 0–76 ppb, compared to a high-accuracy broadband cavity enhanced absorption spectrometer (BBCEAS) as the reference instrument [41]. It provides comprehensive performance metrics including linearity of response, LOD, accuracy, repeatability, response time, and inter-sensor variability. This study also aims to systematically investigate the effects of temperature and RH on sensor performance by implementing a Box-Behnken experimental design and developing a multiple linear regression (MLR) model. Understanding the sensor performance in a laboratory environment paves the way to future studies of applying the sensors to field measurement of formaldehyde.

2. Materials and Methods

2.1. Sensor Preparation

This study evaluated eight Sensirion SFA30® formaldehyde sensors (Sensirion AG, Staefa, Switzerland). These sensors had integrated temperature and RH modules. The laboratory evaluation entailed exposing the low-cost sensors to target concentrations of formaldehyde, temperature, and RH. Each sensor was placed in a 3D-printed polylactic acid (PLA) chamber, which was sealed with polytetrafluoroethylene (PTFE) tape to minimize leakage. These eight sensors were divided into two sets. Each set included four sensors that were connected to an Arduino board via a multiplexer (I2C interface), as shown in Figure S1. This setup allowed real-time viewing and recording of the sensor signal, temperature, and RH using the Arduino interface on a computer.

2.2. Broadband Cavity-Enhanced Absorption Spectrometer

A BBCEAS provided the reference formaldehyde measurements [42] for this study. The BBCEAS consisted of a light source module, an optical cavity module, and a detection module, as shown in Figure S2. A deep ultraviolet LED (DUV-325, Roithner LaserTechnik, Vienna, Austria) centered at 325 nm provided the light source. The LED was aligned within a PTFE optical cavity, which contained two highly reflective mirrors (Layertec GmbH, Mellingen, Germany). An optical fiber collected light at the rear of the cavity, and this optical fiber was connected to a high-sensitivity optic spectrometer (AvaSpec-Hero, Avantes, Inc., Apeldoorn, Netherlands ). Spectra were saved every minute.
The BBCEAS measurements were given by:
α λ = 1 R λ d + α λ R a y l e i g h , Z A I Z A λ I λ I λ
where α λ is the extinction coefficient of the transmitted light through the cavity, λ is the wavelength of light, d is the cavity length, R λ is the mirror reflectivity, α λ R a y l e i g h , Z A is the extinction due to Rayleigh scattering of zero air, I Z A λ is the reference spectrum of zero air, and I λ is the measured spectrum at each wavelength.
We obtained the mirror reflectivity by flowing 1 LPM of nitrogen for 10 min followed by flowing 1 LMP of helium for 10 min into the BBCEAS. Mirror reflectivity was obtained using the following relationship:
R λ = 1 d · I N 2 · n N 2 · σ R a y , N 2 λ I H e · n H e · σ R a y , H e λ I H e λ I N 2 λ
where d is the cavity length, I N 2 λ and I H e λ are the measured spectral intensities when the cavity is filled with nitrogen and helium, respectively, σ R a y , N 2 λ and σ R a y , H e λ are the Rayleigh scattering cross section of nitrogen and helium, respectively.
Subsequently, we obtained the reference spectra by flowing 1 LPM of zero air into the BBCEAS for 20 min. For each formaldehyde sample gas, the sampling time was 75 min. Extinctions from the cavity were fitted using standard literature absorption cross-sections for formaldehyde [43]. All spectra were post-processed using Equations (1) and (2) with a custom-made Python package to obtain the formaldehyde concentration. To obtain formaldehyde concentration from the BBCEAS, we averaged 30 min of stable spectra for the concentration-only tests and 10 min of stable spectra for the environmental simulation tests.

2.3. Laboratory Evaluation System

We designed a laboratory system to evaluate four formaldehyde sensors simultaneously in parallel (Figure 1). The system contained two gas lines, a blank line, and a sample line. Both the blank and sample lines were set to 1 LPM using a mass flow controller (MFC, Cole Parmer, Inc., Vernon Hills, IL, USA). The blank line consisted of zero air, which provided the baseline signal for both the sensors and the BBCEAS. This blank line was also used to flush the system. The sample line started from the permeation oven (Dynacalibrator® Model 340, VICI, Inc., Houston, TX, USA), which generated formaldehyde with a permeation tube of paraformaldehyde (325 ng/min, VICI, Inc., Houston, TX, USA) heated to 90 °C. By adjusting the dilution rate of the permeation oven, formaldehyde concentrations ranging from 0–76 ppb were obtained. A switching valve (Parker, Inc., Mayfield Heights, OH, USA) allowed the two lines to alternate.
The evaluation system was capable of generating different temperature and RH conditions with its environmental chamber, which consisted of a temperature module (water bath, Precision, Thermo Fisher Scientific, Waltham, MA, USA) and an RH module (Nafion tube, Perma Pure LLC, Lakewood, NJ, USA), which provided a temperature range of 0–40 °C and an RH range of 15–75%. The water bath provided the heat source or sink for the environmental chamber. A 6” Nafion (OD 0.110”) tube immersed in deionized water inside a closed glass container provided humidity. A humidity bypass was implemented to adjust the dry/humid gas ratio and provide the target humidity. The sample gas with target concentration, temperature, and RH flowed past the four sensors and then entered the BBCEAS.
For all tests in this study, we used 75 min as the sampling time to obtain relatively stable formaldehyde concentration, sensor signal, temperature, and RH. In between any two samples, we flushed the system with zero air for 20 min. The sensor signal, temperature, and RH from the Arduino interface were recorded and post-processed by taking the average of each stable signal (30-min average for the concentration-only tests and 10-min average for the environmental simulation tests).

2.4. Experimental Design

We evaluated the sensor response to formaldehyde concentration alone (concentration-only tests) and their response to different formaldehyde concentrations, temperature, and RH conditions (environmental simulation tests).
  • Concentration-only tests
The concentration-only tests included target formaldehyde concentrations of 10, 20, 30, 40, and 50 ppb at 24 °C and approximately 15% RH.
b.
Effects of temperature and RH
To further investigate the sensor performance under different temperature and RH conditions, we performed environmental simulation tests with a Box-Behnken experimental design (Table 1). This Box-Behnken design effectively reduced the total number of tests needed to evaluate the influence of temperature and RH on sensor response to formaldehyde.

2.5. Data Analysis

We applied ordinary least squares (OLS) to evaluate the effects of the different parameters on sensor response. All eight sensors were evaluated both individually and as a group.
  • Linear regression (LR) model
For the concentration-only tests, a linear regression (LR) model provided the relationship between the sensor response and the reference measurements (BBCEAS):
y ^ = k c + b
where c is the formaldehyde concentration from the BBCEAS, k is the slope of the linear regression, b is the intercept of the linear regression and y ^ is the predicted sensor response based on the LR model.
b.
Multiple linear regression model
An MLR model provided the relationship between the sensor response to formaldehyde concentration, temperature, and RH:
y ^ = k 1 c + k 2 T + k 3 R H + b
where y ^ is the predicted sensor signal, c is the formaldehyde concentration from the BBCEAS, T is the temperature, R H is the relative humidity, b is the intercept, and k 1 , k 2 , k 3 are the coefficients of the environmental factors.

2.6. Sensor Performance Metrics

Although the performance guidelines for low-cost formaldehyde sensors do not exist, we discussed typical performance metrics (linearity of response, LOD, accuracy, and repeatability), which have been used to evaluate other low-cost air quality sensors [16,18]. In addition, the U.S. environmental protection agency (EPA) also uses these metrics for the evaluation of low-cost gas sensors [44]. We also investigated into the response time and inter-sensor variability.
  • Linearity of response
In this study, we used the coefficient of determination (R2) to evaluate linearity of response:
R 2 = 1 y i y ^ i 2 y i y ¯ 2
where y i is the observed sensor signal, y ¯ is the mean of the observed sensor signal, y ^ i is the predicted sensor signal.
b.
Limit of detection
The LOD was calculated based on the linear relationship between each sensor and the BBCEAS [45]:
L O D = 3.3 σ k
where σ is the standard error of the linear regression and k is the slope of the linear regression from Equation (3).
c.
Sensor accuracy
We used root mean squared error (RMSE) and normalized root mean squared error (NRMSE) to evaluate sensor accuracy:
R M S E = 1 N i = 1 N y ^ i c i 2
N R M S E = R M S E c m a x c m i n
where y ^ i is the predicted sensor signal, c i is the observed formaldehyde concentration from the BBCEAS, c m a x is the maximum of the observed formaldehyde concentration and c m i n is the minimum of the observed formaldehyde concentration, N is the total number of measurements.
In addition to, some other studies of formaldehyde sensor performance have presented mean absolute error (MAE) [26]. MAE is given by:
M A E = y ^ i c i N
where y ^ i is the predicted sensor signal, c i is the observed formaldehyde concentration from the BBCEAS, N is the total number of measurements.
d.
Sensor repeatability
To evaluate the ability of the sensors to generate reproducible responses to the same formaldehyde concentrations, the concentration-only tests were repeated three times. Coefficient of variation (CV) is a common metric for low-cost sensor repeatability [16], and it is given by the following equation:
C V = σ μ × 100 %
where σ is the standard deviation of repeated measurements and μ is the mean of repeated measurements.
However, in this study, the target formaldehyde concentrations could not be reproducibly achieved because of two reasons. First, the dilution rate of the permeation oven could not be precisely controlled. Secondly, the BBCEAS provided formaldehyde concentration only after post-processing at the end of each test; hence the formaldehyde concentration could not be dynamically adjusted to the target concentration. Consequently, we conservatively estimated the CV by calculating the mean and standard deviation of the sensor responses for each target formaldehyde concentration.
e.
Response time
The response time of the sensors was estimated by t90, which is the time needed for the sensors to reach 90% of the final stable signal [46].
f.
Inter-sensor variability
We implemented intraclass correlation coefficient (ICC) to examine inter-sensor variability [47]. Widely used in reliability testing, ICC is a number ranging from 0 to 1, which refers to the correlations within a class of data [48]. Based on the 95% confidence interval of the ICC estimate, values less than 0.5, between 0.5 and 0.75, between 0.75 and 0.9, and greater than 0.90 are indicative of poor, moderate, good, and excellent reliability, respectively [49]. In this study, we used a linear mixed effect model implemented in R Studio® (version 4.2.2) to calculate the ICC, where formaldehyde concentration was the fixed effect and the sensor IDs were the random effect:
y = k c + b + b s + ε
where y is the sensor signal, c is the formaldehyde concentration from the BBCEAS, k is the slope of the linear regression, b is the intercept of the linear regression, b s is the random intercept that captures the variability of sensor IDs, ε is the random error.
Consequently, ICC was calculated as:
I C C = σ b s 2 σ b s 2 + σ ε 2
where σ b s 2 is the variance of the random effect and σ ε 2 is the variance of the random error.

2.7. Preliminary Cross-Sensitivity Tests

Although electrochemical sensors generally have high selectivity, previous studies have reported cross sensitivity to common atmospheric trace gases [12,25,28,50,51]. In this study, we considered common ambient trace gases, including carbon monoxide (CO), nitric oxide (NO), nitrogen dioxide (NO2) and isobutylene. The tests were performed using the same experimental setup as the concentration-only tests, except for the gas generation and reference instrument. Four sensors were exposed to each target gas at the following concentrations: CO (39.7 ppm), NO (101 ppb), NO2 (83 ppb), and isobutylene (100 ppb). The target CO, NO, and NO2 concentrations were based on US National Ambient Air Quality Standards [52]. Each target gas concentration was obtained by diluting a calibration gas cylinder with zero air. The concentrations of CO and isobutylene were measured by a Q-Trak 7575-X (TSI Inc., Shoreview, MN, USA). The concentrations of NO and NO2 were measured by Model 42i NO-NO2-NOX Analyzer (Thermo Fisher Scientific Inc., Waltham, MA, USA). The tests were performed at a temperature of 24 °C and a RH of approximately 15%. We compared sensors signals between zero air and each target gas individually using a Student’s t-test.

3. Results and Discussion

3.1. Concentration-Only Tests

Figure 2 shows the response of the eight sensors to formaldehyde concentrations ranging from 10.7 ppb to 70.9 ppb (at 24 °C and approximately 15% RH). Figure S3 shows the sensor responses for the other two repeated tests. In general, the sensor response increased linearly as formaldehyde concentration increased. Previous laboratory studies of formaldehyde sensors found a linear relationship between sensor response and formaldehyde concentration [29,30,53]. These previous studies included a LaFeO3 thin film sensor, a SiO2 microsphere sensor, and a Ni-Doped SnO2 nanoparticle sensor. In this study, a linear regression model was applied to the concentration-only tests. As shown in Table S2, the slopes of the three repeated concentration-only tests are 1.11 ± 0.097, 1.13 ± 0.086, and 1.07 ± 0.090. The intercepts of the three repeated tests are −7.39 ± 2.44, −8.98 ± 2.02 and −6.68 ± 2.93 ppb.
We evaluated the performance of the formaldehyde sensors both as a group and individually (Section 2.6 describes the performance metrics). Table 2 shows the performance metrics for the eight sensors evaluated as a group. Figure 3 and Table S2 show the performance metrics for individual sensors.
  • Linearity of response
As shown in Figure 2 and Figure 3A, the sensors demonstrated an R2 > 0.98 for individual sensors and an R2 > 0.95 as a group within the target formaldehyde concentration range of 0–50 ppb. Linearity of formaldehyde sensor response has been widely reported in previous laboratory studies, and linear regression models have often been applied to evaluate formaldehyde sensors compared to research-grade instrumentation. Alonso et al. [37] reported an R2 = 0.775 using OLS in their evaluation of low-cost micro fuel cell formaldehyde sensors over the concentration range of 17.3–477.1 ppb (22–606 µg/m3). Hu et al. [30] fabricated a LaFeO3 thin-film formaldehyde sensor and reported an R2 = 0.993 over the concentration range of 0.05–1 ppm in a laboratory test. An R2 value greater than 0.75 is generally considered as a strong agreement between a low-cost sensor and a reference instrument [16,18]. Therefore, our sensors demonstrate a good linearity of response.
b.
Limit of detection
In this study, the LOD (Figure 3B) was calculated based on the linear regression of the sensor response vs. the BBCEAS. Our results show that the individual LOD of the sensors ranged from 2.86–9.73 ppb. As a group, the sensors demonstrated an LOD of 11.3 ± 2.07 ppb. Overall, the sensors exhibit an LOD that is lower than those reported in previous laboratory studies. Alolaywi et al. [38] evaluated an electrochemical formaldehyde sensor and reported a LOD of 60 ppb. Evaluations of nanomaterial-based formaldehyde sensors have reported LODs ranging from 50–90 ppb [27,40,53]. It is important for low-cost sensors to be capable of measuring relevant levels of formaldehyde. Hun et al. [31] reported a mean indoor formaldehyde concentration of 17 ppb. Typical ambient formaldehyde concentrations in urban areas of the U.S. range from 11–20 ppb [54] with some industrial areas of the U.S. reporting higher levels of formaldehyde (24–66 ppb) [55]. Our low LOD suggests the potential of using the sensors to measure indoor and ambient levels formaldehyde.
c.
Sensor accuracy
The accuracy of the sensors was characterized by RMSE (Figure 3C) and NRMSE (Figure 3D). In the three repeated concentration-only tests, the individual RMSE ranged from 1.84 ppb to 10.1 ppb, and the individual NRMSE ranged from 2.9% to 16.8%. Sensor S6 and S8 exhibited higher RMSEs (8.76 ± 1.01 ppb and 7.22 ± 1.95 ppb, respectively) compared to the rest of the sensors. As a group, the eight sensors demonstrated an RMSE of 3.96 ± 0.33 and an NRMSE of 6.23 ± 0.33%, indicating low errors between the measurements of the low-cost sensors and the reference instrument. Although accuracy criteria for low-cost formaldehyde sensors do not exist, for discussion purposes, we compared the RMSE and NRMSE values of the formaldehyde sensors to EPA’s criteria for low-cost ozone sensors (RMSE ≤ 5 ppb) [56] and particulate matter sensors (NRMSE ≤ 30%) [57]. These comparisons suggest that the sensors evaluated in this study would meet EPA guidelines for low-cost sensor accuracy.
Limited laboratory studies have reported RMSE or NRMSE for formaldehyde sensors. Song et al. [58] developed an yttrium-doped ZnO sensor array with a back propagation neural network model, reporting an RMSE of 0.892 ppm over a formaldehyde concentration range of 5–45 ppm in a chamber test. In this study, the individual MAE of the sensors in three repeated tests ranged from 1.53 ppb to 10.1 ppb, as shown in Table S2. Sensor S6 and S8 demonstrated higher MAEs (8.66 ± 1.02 ppb and 7.04 ± 1.82 ppb, respectively) compared to the rest of the sensors in the three repeated concentration-only tests. Chattopadhyay et al. [26] evaluated both electrochemical (MAE ranging from 70.8 ppb to 78.8 ppb) and metal oxide (MAE ranging from 154 ppb to 335 ppb) formaldehyde sensors over a concentration range of 10–800 ppb in a laboratory environment. Our results demonstrated lower measurement error between the low-cost sensors and the reference instrument.
d.
Sensor repeatability
In addition to accuracy, repeatability is another important factor in the evaluation of low-cost sensors. As discussed in Section 2.6, we conservatively estimated the CV of the eight sensors, which ranged from 1.36 to 6.41% (Table S3). Some previous studies [28,29,59] reported that the formaldehyde sensors demonstrated good repeatability by repeatedly exposing the sensors to the same formaldehyde concentration. However, these studies lack quantitative evaluation of the repeatability. Alonso et al. [37] reported a CV of 20%, 14.86% and 14.82% for three different models of a micro fuel cell formaldehyde sensor. Although CV criteria for low-cost formaldehyde sensors do not exist, EPA recommends CV ≤ 30% for both ozone [56] and PM2.5 [57] sensors. Our conservative estimation indicates that the low-cost formaldehyde sensors have acceptable repeatability.
e.
Response time
Table S4 summarizes the response time for sensor S1-S4 in the concentration-only tests. In general, the response time increased as the formaldehyde concentration increased. We observed a minimum response time of 131 s and a maximum response time of 439 s over the formaldehyde concentration range of 17.2–60.1 ppb. Our sensors demonstrate comparable response times to previous studies. Descamps et al. [39] evaluated a hand-held fluorescence-based sensor in the laboratory and reported a response time of 180 s for formaldehyde concentrations ranging from 0–30 ppb. Evaluations of thin-film semiconductor sensors presented response times of less than 120 s for ppm-level formaldehyde concentrations [60].
f.
Inter-sensor variability
Although the eight sensors were purchased from the same distributor at the same time, some inter-sensor variability still existed. As a result, even though the sensors were exposed to the same formaldehyde concentration, they exhibited slightly different responses, which were reflected in their slopes, intercepts, R2 values, LODs, RMSEs, NRMSEs and MAEs (Table S2). This study also evaluated inter-sensor variability using the ICC. Sensors S1-S8 showed an ICC of 0.551, which corresponds to moderate variability (ICC: 0.50–0.75) [49].

3.2. Environmental Simulation Tests

We evaluated the effects of temperature and RH on sensor response and developed an MLR model for sensor response (Figure 4 and Figure S4). MLR is a commonly used model to estimate gas concentrations from the low-cost sensors, and common parameters include target gas concentration, temperature, RH and interferent gases [18]. This evaluation used a Box-Behnken experimental design (described in Section 2.4).
The metrics of R2 and RMSE helped determine the optimum input parameters for the MLR model. Specifically, we observed the R2 and RMSE as we added input variables to the model in the sequence of concentration, temperature, RH, and intercept (Figure 5 and Table S5). Group R2 was the best (R2 = 0.878) when including concentration, temperature, and RH in the MLR model. R2 dropped to 0.751 for the group after adding the intercept term. Individually, R2 for each sensor was also the best when ignoring the intercept term. In addition, the p-value and 95% confidence interval (95% CI) of the intercept indicated that this intercept term was not significant (Table S6). The group RMSE was similar with and without the intercept term (RMSE = 13.32 ppb, RMSE = 13.28, respectively). Consequently, the intercept term was ignored in the MLR model.
The resulting MLR relationship for the eight sensors was given by:
y ^ = 0.552 c + 0.399 T 0.142 R H
where y ^ is the predicted sensor signal, c is the formaldehyde concentration, T is the temperature, and R H is the relative humidity. The equation indicates that sensor response is largely dependent on formaldehyde concentration (p-value = 1.01 × 10−31), positively related to temperature (p-value = 3.06 × 10−12) and negatively related to RH (p-value = 1.02 × 10−7). If the low-cost sensors are used to predict formaldehyde concentration, the MLR model could be converted into:
c = 1.812 y ^ 0.723 T + 0.257 R H
Temperature and RH can affect the baseline and sensitivity of electrochemical sensors [50,61]. Some previous studies have shown that temperature can possibly affect the electrodes of the electrochemical sensors [25] and RH may affect the humidity equilibrium between sample gas and the electrodes [24]. Over the long term, RH also changes the electrolyte chemistry, resulting in a change in zero current and sensitivity [33,34,36]. In addition, sensor type, sensor material, experimental setup and model selection could also affect the sensor response to different temperature and RH conditions.
  • Sensor response to temperature and RH
Alonso et al. [37] evaluated the effects of temperature and RH on the performance of a micro fuel cell formaldehyde sensor using three different models (OLS, ML, and REML). Their laboratory evaluation included a formaldehyde concentration range of 17.3–477.1 ppb (22–606 µg/m3), a temperature range of 19.9–27.7 °C and an RH range of 26.8–72.6%. Their OLS model indicated that both temperature and RH were significant factors for the sensor performance (temperature negatively related and RH positively related). However, the other two models in their study showed that RH was not a significant parameter; temperature significant and was positively related to sensor performance.
Chattopadhyay et al. [26] compared the performance of three different formaldehyde sensors over the temperature range of 22–50 °C and the RH range of 8–85% by looking into the MAE of each sensor compared to a reference instrument (Gasera One Formaldehyde). Their study showed that both metal oxide sensors (SGP30, BME680) and electrochemical sensor (ZE08-CH2O) were sensitive to increasing temperature. Specifically, the metal oxide sensors exhibited larger MAEs when temperature was above 45 °C. The electrochemical sensor (ZE08-CH2O) appeared to be insensitive to RH and exhibited a consistently low measurement MAE (29.43 ppb) over the RH range (8–85%). The BME680 sensor showed low MAE (48.33 ppb) at higher humidity (RH% > 45%) but high MAE (> 500 ppb) when RH is below 40%. The SGP30 sensor exhibited high sensitivity to humidity with an increasing MAE as humidity increased. However, they did not quantitatively evaluate the relationship of temperature and RH on sensor response.
b.
Effects of RH
Previous studies reported different effects of RH on sensor response. Descamps et al. [39] found that an increased RH from 0% to 72.5% reduced their fluorescence-based sensor signal for indoor formaldehyde measurements. Chen et al. [60] reported that a Ga-doped ZnO sensor with Ag paste and Pt wires was not significantly affected by RH (0 to 70 ± 10%). Alolaywi et al. [38] studied the influence of RH on a MoOx/carbon nanocomposite-based electrochemical sensor at room temperature and found that increasing RH from 65% to 90% led to an increased signal response when exposed to 1 ppm formaldehyde gas.

3.3. Preliminary Cross-Sensitivity Tests

Table S7 summarizes the preliminary cross-sensitivity results. It shows a decrease in sensor signal when exposed to CO, NO, or NO2, compared to the average sensor signal when exposed to zero air. This decrease was statistically significant (t-test, p-value < 0.01). Isobutylene at a concentration of 100 ppb did not appear to affect the sensor signal (t-test, p-value > 0.06). In general, previous studies have reported cross sensitivity of electrochemical sensors to common air pollutants, including ozone (O3) cross sensitivity to NO2, and CO cross sensitivity to NO and NO2 [50]. Previous studies have also reported laboratory cross-sensitivity results for different types of formaldehyde sensors. Chang et al. [12] reported responses to ethanol, water vapor, oxygen and acetone that were 7 times lower than to formaldehyde for their functionalized phosphomolybdic acid (PMA)-based sensor. Li et al. [59] found that their hollow TiO2 microsphere sensor had a 10 times greater response to formaldehyde compared to methylbenzene, methanol, ethanol, acetone, and ammonia (3 ppm of formaldehyde and 3 ppm of potential interferent gas, at a temperature of 20 °C and an RH of 40%). Gautam et al. [28] reported that their Si-chip assisted MOS/SiNWs nanocomposite-based sensor had double the response to formaldehyde, compared to methanol, ethanol and acetone (1 ppm of interferent gas and 1 ppm of formaldehyde at room temperature). In summary, our study suggests that the SFA30® is also sensitive to CO, NO and NO2. However, additional study is needed to further understand this sensor’s selectivity to formaldehyde in the presence of other relevant atmospheric trace gases.

3.4. Limitations

This study has several limitations. First, the experimental setup did not permit the study of sensor performance when temperature was below 0 °C or RH exceeded 77%. Second, we performed preliminary cross-sensitivity tests that included only a limited number of species (CO, NO, NO2, and isobutylene) concentrations. This preliminary cross-sensitivity study did not evaluate e the magnitude of this cross sensitivity or other potentially relevant species or mixtures of species. In the ambient environment, formaldehyde is part of a complex mixture of traces gases. However, this is a first step at understanding sensor performance. More sophisticated models [15,26] might be developed to evaluate sensor performance in an ambient environment.

4. Conclusions

This study comprehensively evaluated eight low-cost electrochemical formaldehyde sensors over a formaldehyde concentration range of 0–76 ppb in the laboratory, with a high-accuracy reference instrument (BBCEAS). In general, the sensors exhibited a good linearity of response (R2 > 0.98 individually and > 0.95 as a group), low LOD (2.86–9.73 ppb individually and 11.3 ± 2.07 ppb as a group), good accuracy (1.84–10.1 ppb RMSE and 2.9–16.8% NRMSE individually, 3.96 ± 0.33 ppb RMSE and 6.23 ± 0.33% NRMSE as a group), acceptable repeatability (1.36–6.41% CV individually), moderate inter-sensor variability (ICC = 0.551) and reasonably rapid response (131–439 s). The effects of temperature (target range of 0–40 °C) and RH (target range of 0–75%) were also systematically investigated with a Box-Behnken experimental design and an MLR model. The MLR model indicates that sensor response is largely dependent on formaldehyde concentration, positively related to temperature and negatively related to RH. Overall, this study is important for understanding the performance of low-cost electrochemical formaldehyde sensors. It also reveals the feasibility of applying the sensors to field measurement of formaldehyde in the future.

Supplementary Materials

The supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/s23177444/s1.

Author Contributions

Conceptualization, Z.P., R.T. and K.E.K.; methodology, Z.P., R.T. and K.E.K.; software, Z.P., R.T. and K.E.K.; formal analysis, Z.P. and K.E.K.; investigation, Z.P., M.B. and K.E.K.; writing-original draft preparation, Z.P., R.T. and K.E.K.; writing-reviewing and editing, Z.P., M.B., R.T. and K.E.K.; supervision, K.E.K.; project administration, K.E.K.; funding acquisition, K.E.K. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Science Foundation, grant No. 10054273, CAREER: Community-engaged, sensor network for identifying air pollution sources.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study may be obtained from the authors upon reasonable request.

Conflicts of Interest

Kerry E. Kelly has a financial interest in the company Tellus Networked Solutions, LCC, which commercializes solutions for environmental monitoring. Their technology was not used as part of this work.

References

  1. Hak, C.; Pundt, I.; Trick, S.; Kern, C.; Platt, U.; Dommen, J.; Ordóñez, C.; Prévôt, A.S.H.; Junkermann, W.; Astorga-Lloréns, C.; et al. Intercomparison of Four Different In-Situ Techniques for Ambient Formaldehyde Measurements in Urban Air. Atmos. Chem. Phys. 2005, 5, 2881–2900. [Google Scholar] [CrossRef]
  2. McLaughlin, J.K. Formaldehyde and Cancer: A Critical Review. Int. Arch. Occup. Environ. Health 1994, 66, 295–301. [Google Scholar] [CrossRef]
  3. Nielsen, G.D.; Wolkoff, P. Cancer Effects of Formaldehyde: A Proposal for an Indoor Air Guideline Value. Arch. Toxicol. 2010, 84, 423–446. [Google Scholar] [CrossRef]
  4. EPA, U.S. Summary of Results for the 2005 National—Scale Assessment; U.S. Environmental Protection Agency: Washington, DC, USA, 2012; pp. 1–7.
  5. Initial List of Hazardous Air Pollutants with Modifications | US EPA. Available online: https://www.epa.gov/haps/initial-list-hazardous-air-pollutants-modifications (accessed on 17 August 2023).
  6. Ailshire, J.A.; Crimmins, E.M. Fine Particulate Matter Air Pollution and Cognitive Function among Older US Adults. Am. J. Epidemiol. 2014, 180, 359–366. [Google Scholar] [CrossRef]
  7. Beard, J.D.; Beck, C.; Graham, R.; Packham, S.C.; Traphagan, M.; Giles, R.T.; Morgan, J.G. Winter Temperature Inversions and Emergency Department Visits for Asthma in Salt Lake County, Utah, 2003–2008. Environ. Health Perspect. 2012, 120, 1385–1390. [Google Scholar] [CrossRef]
  8. Weinmayr, G.; Romeo, E.; de Sario, M.; Weiland, S.K.; Forastiere, F. Short-Term Effects of PM10 and NO2 on Respiratory Health among Children with Asthma or Asthma-like Symptoms: A Systematic Review and Meta-Analysis. Environ. Health Perspect. 2010, 118, 449–457. [Google Scholar] [CrossRef]
  9. Zhu, L.; Jacob, D.J.; Keutsch, F.N.; Mickley, L.J.; Scheffe, R.; Strum, M.; González Abad, G.; Chance, K.; Yang, K.; Rappenglück, B.; et al. Formaldehyde (HCHO) As a Hazardous Air Pollutant: Mapping Surface Air Concentrations from Satellite and Inferring Cancer Risks in the United States. Environ. Sci. Technol. 2017, 51, 5650–5657. [Google Scholar] [CrossRef] [PubMed]
  10. Król, S.; Zabiegała, B.; Namieśnik, J. Monitoring VOCs in Atmospheric Air I. On-Line Gas Analyzers. Trends Anal. Chem. 2010, 29, 1092–1100. [Google Scholar] [CrossRef]
  11. Suzuki, Y.; Nakano, N.; Suzuki, K. Portable Sick House Syndrome Gas Monitoring System Based on Novel Colorimetric Reagents for the Highly Selective and Sensitive Detection of Formaldehyde. Environ. Sci. Technol. 2003, 37, 5695–5700. [Google Scholar] [CrossRef] [PubMed]
  12. Chang, C.Y.; Cheng, Y.H.; Ho, C.Y. Surface Engineering of a Triboelectric Nanogenerator for Room Temperature High-Performance Self-Powered Formaldehyde Sensors. J. Mater. Chem. A Mater. 2022, 10, 22373–22389. [Google Scholar] [CrossRef]
  13. Lou, C.; Lei, G.; Liu, X.; Xie, J.; Li, Z.; Zheng, W.; Goel, N.; Kumar, M.; Zhang, J. Design and Optimization Strategies of Metal Oxide Semiconductor Nanostructures for Advanced Formaldehyde Sensors. Coord. Chem. Rev. 2022, 452, 214280. [Google Scholar] [CrossRef]
  14. Shooshtari, M.; Vollebregt, S.; Vaseghi, Y.; Rajati, M.; Pahlavan, S. The Sensitivity Enhancement of TiO2-Based VOCs Sensor Decorated by Gold at Room Temperature. Nanotechnology 2023, 34, 255501. [Google Scholar] [CrossRef] [PubMed]
  15. Sayahi, T.; Garff, A.; Quah, T.; Lê, K.; Becnel, T.; Powell, K.M.; Gaillardon, P.E.; Butterfield, A.E.; Kelly, K.E. Long-Term Calibration Models to Estimate Ozone Concentrations with a Metal Oxide Sensor. Environ. Pollut. 2020, 267, 115363. [Google Scholar] [CrossRef]
  16. Kang, Y.; Aye, L.; Ngo, T.D.; Zhou, J. Performance Evaluation of Low-Cost Air Quality Sensors: A Review. Sci. Total Environ. 2022, 818, 151769. [Google Scholar] [CrossRef] [PubMed]
  17. Collier-Oxandale, A.M.; Thorson, J.; Halliday, H.; Milford, J.; Hannigan, M. Understanding the Ability of Low-Cost MOx Sensors to Quantify Ambient VOCs. Atmos. Meas. Tech. 2019, 12, 1441–1460. [Google Scholar] [CrossRef]
  18. Karagulian, F.; Barbiere, M.; Kotsev, A.; Spinelle, L.; Gerboles, M.; Lagler, F.; Redon, N.; Crunaire, S.; Borowiak, A. Review of the Performance of Low-Cost Sensors for Air Quality Monitoring. Atmosphere 2019, 10, 506. [Google Scholar] [CrossRef]
  19. Castell, N.; Dauge, F.R.; Schneider, P.; Vogt, M.; Lerner, U.; Fishbain, B.; Broday, D.; Bartonova, A. Can Commercial Low-Cost Sensor Platforms Contribute to Air Quality Monitoring and Exposure Estimates? Environ. Int. 2017, 99, 293–302. [Google Scholar] [CrossRef]
  20. Chung, P.R.; Tzeng, C.T.; Ke, M.T.; Lee, C.Y. Formaldehyde Gas Sensors: A Review. Sensors 2013, 13, 4468–4484. [Google Scholar] [CrossRef]
  21. Nikolic, M.V.; Milovanovic, V.; Vasiljevic, Z.Z.; Stamenkovic, Z. Semiconductor Gas Sensors: Materials, Technology, Design, and Application. Sensors 2020, 20, 6694. [Google Scholar] [CrossRef]
  22. Wang, C.; Yin, L.; Zhang, L.; Xiang, D.; Gao, R. Metal Oxide Gas Sensors: Sensitivity and Influencing Factors. Sensors 2010, 10, 2088–2106. [Google Scholar] [CrossRef]
  23. Dhall, S.; Mehta, B.R.; Tyagi, A.K.; Sood, K. A Review on Environmental Gas Sensors: Materials and Technologies. Sens. Int. 2021, 2, 100116. [Google Scholar] [CrossRef]
  24. Sun, L.; Wong, K.C.; Wei, P.; Ye, S.; Huang, H.; Yang, F.; Westerdahl, D.; Louie, P.K.K.; Luk, C.W.Y.; Ning, Z. Development and Application of a next Generation Air Sensor Network for the Hong Kong Marathon 2015 Air Quality Monitoring. Sensors 2016, 16, 211. [Google Scholar] [CrossRef]
  25. Cross, E.S.; Williams, L.R.; Lewis, D.K.; Magoon, G.R.; Onasch, T.B.; Kaminsky, M.L.; Worsnop, D.R.; Jayne, J.T. Use of Electrochemical Sensors for Measurement of Air Pollution: Correcting Interference Response and Validating Measurements. Atmos. Meas. Tech. 2017, 10, 3575–3588. [Google Scholar] [CrossRef]
  26. Chattopadhyay, A.; Huertas, A.; Rebeiro-Hargrave, A.; Fung, P.L.; Varjonen, S.; Hieta, T.; Tarkoma, S.; Petaja, T. Low-Cost Formaldehyde Sensor Evaluation and Calibration in a Controlled Environment. IEEE Sens. J. 2022, 22, 11791–11802. [Google Scholar] [CrossRef]
  27. Deng, H.; Li, H.-R.; Wang, F.; Yuan, C.-X.; Liu, S.; Wang, P.; Xie, L.-Z.; Sun, Y.-Z.; Chang, F.Z. A High Sensitive and Low Detection Limit of Formaldehyde Gas Sensor Based on Hierarchical Flower-like CuO Nanostructure Fabricated by Sol–Gel Method. J. Mater. Sci. Mater. Electron. 2016, 27, 6766–6772. [Google Scholar] [CrossRef]
  28. Gautam, V.; Kumar, A.; Nagpal, S.; Jain, V.K. Ultrasensitive Detection of Formaldehyde at Room Temperature Using Si-Chip Assisted MOS/SiNWs Nanocomposite Based Sensor. J. Alloys Compd. 2022, 919, 165840. [Google Scholar] [CrossRef]
  29. Li, Y.; Chen, N.; Deng, D.; Xing, X.; Xiao, X.; Wang, Y. Formaldehyde Detection: SnO2 Microspheres for Formaldehyde Gas Sensor with High Sensitivity, Fast Response/Recovery and Good Selectivity. Sens. Actuators B Chem. 2017, 238, 264–273. [Google Scholar] [CrossRef]
  30. Hu, J.; Chen, X.; Zhang, Y. Batch Fabrication of Formaldehyde Sensors Based on LaFeO3 Thin Film with Ppb-Level Detection Limit. Sens. Actuators B Chem. 2021, 349, 130738. [Google Scholar] [CrossRef]
  31. Hun, D.E.; Corsi, R.L.; Morandi, M.T.; Siegel, J.A. Formaldehyde in Residences: Long-Term Indoor Concentrations and Influencing Factors. Indoor Air 2010, 20, 196–203. [Google Scholar] [CrossRef]
  32. Weisel, C.P.; Zhang, J.; Turpin, B.J.; Morandi, M.T.; Colome, S.; Stock, T.H.; Spektor, D.M.; Korn, L.; Winer, A.; Alimokhtari, S.; et al. Relationship of Indoor, Outdoor and Personal Air (RIOPA) Study: Study Design, Methods and Quality Assurance/Control Results. J. Expo. Anal. Environ. Epidemiol. 2005, 15, 123–137. [Google Scholar] [CrossRef]
  33. Ouyang, B. First-Principles Algorithm for Air Quality Electrochemical Gas Sensors. ACS Sens. 2020, 5, 2742–2746. [Google Scholar] [CrossRef] [PubMed]
  34. Farquhar, A.K.; Henshaw, G.S.; Williams, D.E. Understanding and Correcting Unwanted Influences on the Signal from Electrochemical Gas Sensors. ACS Sens. 2021, 6, 1295–1304. [Google Scholar] [CrossRef] [PubMed]
  35. Wei, P.; Ning, Z.; Ye, S.; Sun, L.; Yang, F.; Wong, K.C.; Westerdahl, D.; Louie, P.K.K. Impact Analysis of Temperature and Humidity Conditions on Electrochemical Sensor Response in Ambient Air Quality Monitoring. Sensors 2018, 18, 59. [Google Scholar] [CrossRef]
  36. Hitchman, M.L.; Saffell, J.R. Considerations of Thermodynamics and Kinetics for the Effects of Relative Humidity on the Electrolyte in Electrochemical Toxic Gas Sensors. ACS Sens. 2021, 6, 3985–3993. [Google Scholar] [CrossRef] [PubMed]
  37. Justo Alonso, M.; Madsen, H.; Liu, P.; Jørgensen, R.B.; Jørgensen, T.B.; Christiansen, E.J.; Myrvang, O.A.; Bastien, D.; Mathisen, H.M. Evaluation of Low-Cost Formaldehyde Sensors Calibration. Build. Environ. 2022, 222, 109380. [Google Scholar] [CrossRef]
  38. Alolaywi, H.Y.; Duanghathaipornsuk, S.; Kim, S.S.; Li, C.-H.; Jinschek, J.R.; Kim, D.-S.; Alba-Rubio, A.C. Electrochemical MoOx/Carbon Nanocomposite-Based Gas Sensor for Formaldehyde Detection at Room Temperature. J. Electrochem. Soc. 2021, 168, 067525. [Google Scholar] [CrossRef]
  39. Descamps, M.N.; Bordy, T.; Hue, J.; Mariano, S.; Nonglaton, G.; Schultz, E.; Tran-Thi, T.H.; Vignoud-Despond, S. Real-Time Detection of Formaldehyde by a Sensor. Sens. Actuators B Chem. 2012, 170, 104–108. [Google Scholar] [CrossRef]
  40. Song, Y.; Zhang, Y.; Ma, M.; Ren, J.; Liu, C.; Tan, J. Visible Light-Assisted Formaldehyde Sensor Based on HoFeO3 Nanoparticles with Sub-Ppm Detection Limit. Ceram. Int. 2020, 46, 16337–16344. [Google Scholar] [CrossRef]
  41. Washenfelder, R.A.; Attwood, A.R.; Flores, J.M.; Zarzana, K.J.; Rudich, Y.; Brown, S.S. Broadband Cavity-Enhanced Absorption Spectroscopy in the Ultraviolet Spectral Region for Measurements of Nitrogen Dioxide and Formaldehyde. Atmos. Meas. Tech. 2016, 9, 41–52. [Google Scholar] [CrossRef]
  42. Bhardwaj, N.; Kelsch, A.; Eatough, D.J.; Thalman, R.; Daher, N.; Kelly, K.; Jaramillo, I.C.; Hansen, J.C. Sources of Formaldehyde in Bountiful, Utah. Atmosphere 2021, 12, 375. [Google Scholar] [CrossRef]
  43. Meller, R.; Moortgat, G.K. Temperature Dependence of the Absorption Cross Sections of Formaldehyde between 223 and 323 K in the Wavelength Range 225-375 Nm. J. Geophys. Res. Atmos. 2000, 105, 7089–7101. [Google Scholar] [CrossRef]
  44. Clements, A.; Duvall, R.; Green, D.; Dye, T. Enhanced Air Sensor Guidebook; U.S. Environmental Protection Agency: Washington, DC, USA, 2022.
  45. Shrivastava, A.; Gupta, V. Methods for the Determination of Limit of Detection and Limit of Quantitation of the Analytical Methods. Chron. Young Sci. 2011, 2, 21. [Google Scholar] [CrossRef]
  46. Spinelle, L.; Aleixandre, M.; Gerboles, M. Protocol of Evaluation and Calibration of Low-Cost Gas Sensors for the Monitoring of Air Pollution; Publications Office of the European Union: Luxembourg, 2013.
  47. Cicchetti, D.V. Guidelines, Criteria, and Rules of Thumb for Evaluating Normed and Standardized Assessment Instruments in Psychology. Psychol. Assess. 1994, 6, 284–290. [Google Scholar] [CrossRef]
  48. Liljequist, D.; Elfving, B.; Roaldsen, K.S. Intraclass Correlation—A Discussion and Demonstration of Basic Features. PloS ONE 2019, 14, e0219854. [Google Scholar] [CrossRef] [PubMed]
  49. Koo, T.K.; Li, M.Y. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J. Chiropr. Med. 2016, 15, 155–163. [Google Scholar] [CrossRef]
  50. Mead, M.I.; Popoola, O.A.M.; Stewart, G.B.; Landshoff, P.; Calleja, M.; Hayes, M.; Baldovi, J.J.; McLeod, M.W.; Hodgson, T.F.; Dicks, J.; et al. The Use of Electrochemical Sensors for Monitoring Urban Air Quality in Low-Cost, High-Density Networks. Atmos. Environ. 2013, 70, 186–203. [Google Scholar] [CrossRef]
  51. Lewis, A.C.; Lee, J.D.; Edwards, P.M.; Shaw, M.D.; Evans, M.J.; Moller, S.J.; Smith, K.R.; Buckley, J.W.; Ellis, M.; Gillot, S.R.; et al. Evaluating the Performance of Low Cost Chemical Sensors for Air Pollution Research. Faraday Discuss. 2016, 189, 85–103. [Google Scholar] [CrossRef]
  52. NAAQS Table | US EPA. Available online: https://www.epa.gov/criteria-air-pollutants/naaqs-table (accessed on 19 August 2023).
  53. Yang, Z.; Cao, W.; Peng, C.; Wang, T.; Li, B.; Ma, H.; Su, Y.; Zhou, Z.; Yang, J.; Zeng, M. Construction, Application and Verification of a Novel Formaldehyde Gas Sensor System Based on Ni-Doped SnO2Nanoparticles. IEEE Sens. J. 2021, 21, 11023–11030. [Google Scholar] [CrossRef]
  54. Myers, G.E.; Nagaoka, M. Environmental Protection Agency EPA Hazard Summary. In Formaldehyde; U.S. Environmental Protection Agency: Washington, DC, USA, 2000. [Google Scholar]
  55. Parrish, D.D.; Ryerson, T.B.; Mellqvist, J.; Johansson, J.; Fried, A.; Richter, D.; Walega, J.G.; Washenfelder, R.A.; De Gouw, J.A.; Peischl, J.; et al. Primary and Secondary Sources of Formaldehyde in Urban Atmospheres: Houston Texas Region. Atmos. Chem. Phys. 2012, 12, 3273–3288. [Google Scholar] [CrossRef]
  56. Duvall, R.M.; Clements, A.L.; Hagler, G.; Kamal, A.; Kilaru, V.; Goodman, L.; Frederick, S.; Barkjohn, K.K.; VonWald, I.; Greene, D.; et al. Performance Testing Protocols, Metrics, and Target Values for Ozone Air Sensors; EPA/600/R-20/279, 2021; U.S. EPA Office of Research and Development: Washington, DC, USA, 2021.
  57. Duvall, R.M.; Clements, A.L.; Hagler, G.; Kamal, A.; Kilaru, V.; Goodman, L.; Frederick, S.; Barkjohn, K.K.; VonWald, I.; Greene, D.; et al. Performance Testing Protocols, Metrics, and Target Values for Fine Particulate Matter Air Sensors: Use in Ambient, Outdoor, Fixed Site, Non-Regulatory Supplemental and Informational Monitoring Applications; EPA/600/R-20/280, 2021; U.S. EPA Office of Research and Development: Washington, DC, USA, 2021.
  58. Song, H.; Ma, L.; Pei, S.; Dong, C.; Zhu, E.; Zhang, B. Quantitative Detection of Formaldehyde and Ammonia Using a Yttrium-Doped ZnO Sensor Array Combined with a Back-Propagation Neural Network Model. Sens. Actuators A Phys. 2021, 331, 112940. [Google Scholar] [CrossRef]
  59. Li, X.; Li, X.; Wang, J.; Lin, S. Highly Sensitive and Selective Roomerature Formaldehyde Sensors Using Hollow TiO2 Microspheres. Sens. Actuators B Chem. 2015, 219, 158–163. [Google Scholar] [CrossRef]
  60. Chen, D.; Yuan, Y.J. Thin-Film Sensors for Detection of Formaldehyde: A Review. IEEE Sens. J. 2015, 15, 6749–6760. [Google Scholar] [CrossRef]
  61. Hitchman, M.L.; Cade, N.J.; Kim Gibbs, T.; Hedley, N.J.M. Study of the Factors Affecting Mass Transport in Electrochemical Gas Sensors. Analyst 1997, 122, 1411–1417. [Google Scholar] [CrossRef]
Figure 1. Flow chart of the laboratory evaluation system. All the tubing inside the system was PTFE tubing (OD 1/4″).
Figure 1. Flow chart of the laboratory evaluation system. All the tubing inside the system was PTFE tubing (OD 1/4″).
Sensors 23 07444 g001
Figure 2. Response of eight sensors vs. the BBCEAS to formaldehyde concentrations ranging from 10.7 ppb to 70.9 ppb at 24 °C and approximately 15% RH. Error bars in X-axis denote the standard deviation of the BBCEAS measurement and error bars in Y-axis denote the standard deviation of the sensor response.
Figure 2. Response of eight sensors vs. the BBCEAS to formaldehyde concentrations ranging from 10.7 ppb to 70.9 ppb at 24 °C and approximately 15% RH. Error bars in X-axis denote the standard deviation of the BBCEAS measurement and error bars in Y-axis denote the standard deviation of the sensor response.
Sensors 23 07444 g002
Figure 3. Performance metrics for each sensor in concentration-only tests at 24 °C and approximately 15% RH: (A) R2; (B) LOD; (C) RMSE; (D) NRMSE. Error bars denote the standard deviation for three repeated tests. The different patterns in the bars denote different sensors.
Figure 3. Performance metrics for each sensor in concentration-only tests at 24 °C and approximately 15% RH: (A) R2; (B) LOD; (C) RMSE; (D) NRMSE. Error bars denote the standard deviation for three repeated tests. The different patterns in the bars denote different sensors.
Sensors 23 07444 g003
Figure 4. Sensor response vs. formaldehyde concentration, temperature, and RH: (A) Temperature (°C) and (B) RH (%). Experimental conditions selected by a Box-Behnken design. Note that this figure showed actual testing conditions. Target testing conditions can be found in Table 1.
Figure 4. Sensor response vs. formaldehyde concentration, temperature, and RH: (A) Temperature (°C) and (B) RH (%). Experimental conditions selected by a Box-Behnken design. Note that this figure showed actual testing conditions. Target testing conditions can be found in Table 1.
Sensors 23 07444 g004
Figure 5. R2 and RMSE for the MLR model variable selection: (A) R2 and (B) RMSE. The following parameters were added sequentially: concentration (C), temperature (T), relative humidity (RH) and the intercept. Results were included for the group of all eight sensors as well as for the individual sensors.
Figure 5. R2 and RMSE for the MLR model variable selection: (A) R2 and (B) RMSE. The following parameters were added sequentially: concentration (C), temperature (T), relative humidity (RH) and the intercept. Results were included for the group of all eight sensors as well as for the individual sensors.
Sensors 23 07444 g005
Table 1. Box-Behnken design for environmental simulation tests of the sensors.
Table 1. Box-Behnken design for environmental simulation tests of the sensors.
Test No.Target Concentration * (ppb)Target Temperature * (°C)Target RH * (%)
125015
2254015
325075
4254075
50045
604045
750045
8504045
902015
1002075
11502015
12502075
13252045
14252045
15252045
* Note that these were target concentrations, temperatures, and RHs. Actual conditions can be found in Table S1.
Table 2. Performance metrics of all sensors as a group in three repeated concentration-only tests.
Table 2. Performance metrics of all sensors as a group in three repeated concentration-only tests.
MetricsAverageStandard Deviation
R20.9640.012
LOD (ppb)11.32.07
RMSE (ppb)3.960.33
NRMSE%6.230.33
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Pei, Z.; Balitskiy, M.; Thalman, R.; Kelly, K.E. Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor. Sensors 2023, 23, 7444. https://doi.org/10.3390/s23177444

AMA Style

Pei Z, Balitskiy M, Thalman R, Kelly KE. Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor. Sensors. 2023; 23(17):7444. https://doi.org/10.3390/s23177444

Chicago/Turabian Style

Pei, Zheyuan, Maxim Balitskiy, Ryan Thalman, and Kerry E. Kelly. 2023. "Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor" Sensors 23, no. 17: 7444. https://doi.org/10.3390/s23177444

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop