1. Introduction
Cataract surgery is one of the most frequently performed surgical procedures worldwide and represents the standard treatment for lens opacification, the leading cause of reversible blindness globally [
1]. Modern cataract surgery is routinely performed using phacoemulsification followed by implantation of an intraocular lens (IOL), providing excellent visual rehabilitation with high safety and efficacy [
2,
3]. Currently, four principal biomaterials are available for IOL manufacturing: polymethylmethacrylate (PMMA), hydrophilic acrylic, hydrophobic acrylic, and silicone [
4,
5].
One of the most common long-term complications following cataract surgery is posterior capsule opacification (PCO), commonly referred to as secondary cataract, which may progressively reduce visual acuity, contrast sensitivity, and overall visual quality, significantly affecting patients’ daily activities [
6,
7]. PCO develops because the complete removal of lens epithelial cells (LECs) during cataract extraction is impossible. Residual equatorial LECs retain proliferative potential, migrating toward the posterior capsule where they undergo proliferation, transdifferentiation into myofibroblasts, extracellular matrix deposition, and fibrotic remodeling, ultimately leading to visually significant capsule opacification [
8,
9,
10,
11,
12,
13].
The incidence of PCO is strongly influenced by patient age. Epidemiological studies report rates approaching 70% in patients younger than 40 years, approximately 37% in adults older than 40 years, and nearly 100% in pediatric patients, reflecting the greater proliferative capacity of younger lens epithelial cells [
14,
15].
Nd laser posterior capsulotomy remains the gold-standard treatment for visually significant PCO because it is rapid, minimally invasive, and highly effective in restoring visual function [
16]. Despite continuous improvements in cataract surgical techniques and IOL design, approximately 15–33% of patients undergoing uncomplicated phacoemulsification eventually require laser capsulotomy during postoperative follow-up [
17,
18,
19,
20,
21].
Although Nd capsulotomy is generally considered safe, one of its most frequent procedure-related complications is inadvertent laser impact on the implanted IOL optic rather than the posterior capsule, resulting in localized surface defects commonly referred to as IOL pitting [
20]. Experimental investigations have demonstrated that the morphology and severity of laser-induced defects depend on the optical material. PMMA lenses typically develop characteristic “star-crater” lesions, whereas silicone IOLs exhibit a “splash-like” damage pattern following laser exposure [
19,
22].
The susceptibility of an IOL to laser-induced damage is influenced by several factors, including biomaterial composition, refractive index, polymer architecture, manufacturing process, and surface characteristics [
23]. These variables determine the laser damage threshold, defined as the minimum laser energy capable of producing permanent structural alterations within the optic. Optical microscopy analyses have demonstrated that silicone IOLs exhibit the lowest resistance, with a damage threshold of approximately 0.37 mJ, followed by acrylic lenses (0.52–0.66 mJ), whereas PMMA lenses appear to be the most resistant, with reported thresholds around 0.68 mJ [
24].
To date, no universally accepted laser energy threshold has been identified that reliably separates the energy required for effective capsulotomy from that capable of producing clinically relevant IOL damage [
25,
26,
27,
28,
29]. Consequently, current recommendations emphasize the use of the lowest effective laser energy, accurate focusing posterior to the capsule, and meticulous aiming to minimize the risk of inadvertent IOL injury.
Fortunately, most cases of IOL pitting produce little or no clinically significant visual impairment. Nevertheless, when defects involve the visual axis or become extensive, patients may experience glare, halos, reduced contrast sensitivity, decreased visual acuity, or the perception of fixed opaque spots. In rare but severe cases, extensive optic damage may necessitate surgical IOL explantation and exchange [
30,
31].
While previous studies have primarily characterized the morphological appearance and physical properties of laser-induced IOL damage, considerably less is known about its functional optical consequences. In particular, the effects of laser pitting on retinal light distribution, forward light scatter, image quality, and optical performance remain incompletely understood. Previous experimental investigations have demonstrated that Nd:YAG laser impacts can produce material-dependent surface craters, microfractures, and irregularities in acrylic, silicone, and polymethyl methacrylate IOLs [
29,
32,
33,
34]. However, most available studies have focused primarily on damage morphology, laser-damage thresholds, or surface characterization. Objective optical consequences, including changes in point spread function, modulation transfer function, forward light scatter, contrast transmission, and retinal image quality, have been evaluated less extensively. Therefore, the relationship between the structural appearance of laser-induced pits and the resulting degradation of optical performance remains incompletely defined. In particular, although surface defects may theoretically increase straylight and alter contrast transmission, direct comparisons between untreated and laser-damaged IOLs using complementary structural and optical measurements remain limited. The present study was therefore designed as an exploratory optical bench investigation integrating SEM, thermographic imaging, digital beam-displacement analysis, PSF assessment, and MTF evaluation.
The CT LUCIA and AcrySof IQ models were selected because both are commonly used hydrophobic acrylic monofocal IOLs with the same nominal dioptric power, while differing in polymer formulation, refractive index, manufacturing characteristics, and surface microarchitecture. Their inclusion allowed an exploratory comparison between two clinically established hydrophobic acrylic platforms while limiting variability related to IOL category and refractive power.
Therefore, the purpose of the present experimental study was to investigate how Nd laser-induced pitting alters light transmission through hydrophobic acrylic intraocular lenses using a standardized optical bench model. Specifically, we evaluated whether laser-induced surface defects modify retinal light distribution and quantified the resulting optical degradation through point spread function (PSF) and modulation transfer function (MTF) analyses, providing objective evidence of optical alterations that may contribute to, but cannot by themselves demonstrate, the visual disturbances occasionally reported after Nd:YAG laser capsulotomy.
2. Materials and Methods
This experimental laboratory study was conducted through a collaboration between the Department of Ophthalmology and the Physics Laboratory of the “Gaetano Martino” University Hospital, Messina, Italy. The aim of the study was to investigate the structural and optical consequences of Nd:YAG laser-induced pitting in hydrophobic acrylic intraocular lenses (IOLs) using a multimodal experimental approach integrating scanning electron microscopy (SEM), infrared thermal imaging, optical bench analysis, and digital image processing.
The experimental setup included sixteen monofocal hydrophobic acrylic intraocular lenses, comprising eight CT LUCIA IOLs (Carl Zeiss Meditec, Jena, Germany) and eight AcrySof IQ IOLs (Alcon Laboratories Inc., Fort Worth, TX, USA), a customized IOL holder, a VISULAS YAG III laser system (Carl Zeiss Meditec, Jena, Germany) for the induction of laser pitting, and a PhysioGo.Lite laser platform (ASTAR Industry, Bielsko-Biała, Poland) for standardized optical illumination during thermographic acquisition, an optical bench equipped with a collimated LED light source, a retinal simulation screen consisting of a white A4 sheet (emissivity = 0.70), a complementary metal–oxide–semiconductor (CMOS) image sensor, and a FLIR T440bx infrared thermal camera (FLIR Systems, Wilsonville, OR, USA). Morphological evaluation of laser-induced defects was performed using scanning electron microscopy, whereas digital image processing and quantitative analyses were carried out using MATLAB R2024a (MathWorks, Natick, MA, USA).
2.1. Nd:YAG Laser Procedure
Laser-induced pitting was produced using a VISULAS YAG III Nd:YAG laser system (Carl Zeiss Meditec, Jena, Germany). Each IOL was positioned on a custom-designed holder and exposed to standardized laser applications using a pulse energy of 1.2 mJ, reproducing the energy levels commonly employed during posterior capsulotomy. All laser procedures were performed on isolated intraocular lenses mounted in air. No aqueous-filled model eye, balanced salt solution, or capsular bag simulation was used during laser exposure. For each IOL model, two lenses received five laser impacts, two received ten impacts, and two received fifteen impacts. Two additional untreated lenses of each model served as controls. All laser applications were performed under identical experimental conditions by the same experienced operator to ensure reproducibility and minimize procedural variability. The laser impacts were applied within the central optic at adjacent non-overlapping locations to reproduce multiple inadvertent laser hits while avoiding repeated impacts within the same crater.
2.2. Thermographic Image Acquisition
Following laser treatment, infrared images were acquired using a FLIR T440bx thermal imaging camera. During image acquisition, the PhysioGo.Lite laser platform (ASTAR Industry) was employed as a standardized optical illumination source. The system was operated under constant conditions with a spot size of 0.3 cm2, an output power of 200 mW, an irradiation time of 18 s, and an energy density of 12 J/cm2, thereby ensuring reproducible illumination for all experimental measurements. Thermal images were acquired at a frequency of 60 Hz over a 30 s acquisition period and stored in TIFF format before subsequent processing using MATLAB. Infrared thermography was not used to quantify heat-related material damage or temperature-dependent optical degradation. Its purpose was to provide a reproducible visualization of the spatial distribution of the transmitted light spot under standardized illumination conditions. The thermographic images were subsequently used for digital image registration, subtraction, and centroid-displacement analysis. Therefore, this component of the experiment assessed changes in light propagation rather than the thermal mechanism of Nd:YAG-induced IOL damage.
2.3. IOLs Characteristics
A total of sixteen monofocal hydrophobic acrylic intraocular lenses (IOLs) were included in the study, comprising eight CT LUCIA lenses (Carl Zeiss Meditec, Jena, Germany) and eight AcrySof IQ lenses (Alcon Laboratories Inc., Fort Worth, TX, USA). All IOLs had an identical refractive power of +19.0 diopters and were selected from the models routinely implanted at our institution. These models were selected because they represent two widely used hydrophobic acrylic monofocal IOL platforms with the same nominal refractive power but different proprietary polymer formulations, refractive indices, manufacturing processes, and surface characteristics. The comparison was intended to explore whether two clinically established hydrophobic acrylic platforms might exhibit different descriptive patterns of laser-induced structural and optical alteration. The study was not designed or powered to establish definitive superiority or material-specific resistance.
Each IOL was mounted on a custom-designed lens holder to ensure stable positioning and reproducible alignment throughout the experimental procedures. Laser treatment was subsequently performed according to the protocol described in
Section 2.1. Untreated and laser-treated lenses were then subjected to morphological, thermographic, and optical bench analyses.
2.4. FLIR T440bx Thermal Imaging CAMERA
The FLIR T440 thermal imaging camera was utilized in this study. It is a long-wave thermal imager equipped with an uncooled microbolometer focal array. The spectral range of sensitivity spans from 7.5 to 13 μm, with a thermal sensitivity of 0.1 °C and an FPA (focal plane array) of 320 × 240 pixels. The thermal camera captured images at a rate of 60 Hz, with each acquisition lasting for 30 s. The images were initially acquired in .tiff format and subsequently converted to .jpeg to facilitate further post-processing.
2.5. Scanning Electron Microscope
The analysis to assess the morphological characteristics of the lenses under study was conducted using a scanning electron microscope, specifically the Jeol JMC-6000 (Jeol Co., Akishima, Tokyo, Japan). Each lens was air-dried prior to examination. The scanning was performed in high-vacuum mode, scanning each sample along the x-y directions. The electron beam was accelerated at 15 kV to achieve a magnification of 300×. After each acquisition, all scanned images were saved in Tagged Image File Format (.tiff) without any further processing.
2.6. Optical Bench, PSF, and MTF Measurements
Optical quality was evaluated using a customized optical bench comprising a collimated LED illumination source, an adjustable aperture, a custom-designed IOL holder, an image-acquisition plane, and a CMOS sensor. Each IOL was positioned with its optical axis aligned with the illumination and imaging axes. Untreated and laser-treated IOLs were evaluated under identical geometrical and illumination conditions. Optical measurements were likewise performed in air. Consequently, the refractive environment differed from that of an intraocular lens immersed in aqueous humor within the pseudophakic eye.
The optical bench configuration included a 3.0 mm pupil aperture and monochromatic illumination with a central wavelength of 546 nm. The distance between the IOL and the image-acquisition plane was set at 50 mm, while images were recorded using a CMOS sensor with a resolution of 1920 × 1200 pixels. All measurements were performed in a controlled dark environment to minimize background illumination, stray reflections, and variations in image contrast.
For each IOL, three repeated acquisitions were obtained. The lens was removed and realigned between consecutive acquisitions to account for potential positioning variability. Measurement repeatability was assessed by calculating the mean and standard deviation of the values obtained from the three acquisitions.
PSF images were acquired using a point-like illumination target positioned along the optical axis. The resulting images were analyzed in terms of central peak intensity, halo extension, radial light distribution, and the presence of peripheral light spikes. Where applicable, the full width at half maximum was calculated from the intensity profile passing through the PSF centroid.
MTF curves were derived from the acquired PSF images using Fourier-transform analysis and were evaluated over a spatial-frequency range of 0–100 cycles/mm. Numerical MTF values were extracted at 10, 20, 30, 50, and 100 cycles/mm. The untreated control lens for each model was used as the reference for calculating the relative percentage reduction in MTF after Nd:YAG laser exposure.
The optical bench was configured according to the general principles described in ISO 11979-2:2024 [
14] for laboratory evaluation of IOL optical properties. However, because the system was customized for this exploratory investigation and did not reproduce all elements of a certified model-eye testing platform, it should not be considered a fully ISO-compliant or certified optical bench. A schematic representation of the complete optical bench configuration is provided in
Figure 1.
The available optical acquisition system was primarily designed for qualitative assessment of retinal light distribution and optical image degradation. Consequently, PSF analysis focused on descriptive features, including central peak intensity, halo enlargement, radial light distribution, and peripheral light spikes. Quantitative PSF metrics such as full width at half maximum (FWHM), Strehl ratio, and encircled energy were not calculated because the experimental setup was not specifically calibrated for validated quantitative PSF characterization.
2.7. MATLAB Software
MATLAB software was used for image registration, digital subtraction, and quantitative centroid analysis. All thermographic images were acquired at a resolution of 320 × 240 pixels. Before subtraction, images from untreated and laser-treated IOLs were converted to grayscale and normalized to the same intensity range.
To minimize errors caused by differences in camera or sample positioning, each image pair was subjected to rigid image registration. Registration was performed using fixed reference points within the acquisition field and translational correction along the x- and y-axes. No rotational or non-linear transformation was applied. Registered images were visually inspected before subtraction to confirm adequate alignment.
The image obtained from the corresponding untreated IOL was subtracted from the image obtained from each laser-treated IOL. Pixels with identical or closely similar intensity values were suppressed, whereas pixels showing intensity differences remained visible in the subtraction image.
The center of the transmitted light spot was identified using an intensity-weighted centroid algorithm. The centroid coordinates were calculated as
and
where
represents the pixel intensity at coordinates
and
.
The displacement between the untreated and laser-treated conditions was calculated as:
This value represents the Euclidean displacement of the light-spot centroid on the image plane. Δ-pixel displacement was used as an experimental surrogate of altered light propagation and was not intended as a direct measure of retinal displacement, visual acuity, or clinical dysphotopsia.
To verify the stability of the centroid calculation, the analysis was repeated on 30 independent frames for each IOL condition. The reported Δ-pixel values represent the mean centroid displacement calculated across the 30 consecutive frames, while the corresponding standard deviation was used as an indicator of measurement repeatability. The same intensity threshold, image registration procedure, and centroid detection algorithm were applied to all images to ensure methodological consistency and minimize operator-dependent variability.
2.8. Statistical and Descriptive Analysis
This study was designed as an exploratory proof-of-concept laboratory investigation. Because only two IOLs were available for each model and experimental condition, no inferential statistical analysis was performed. Quantitative measurements are therefore reported descriptively as individual values or mean ± standard deviation, where replicate measurements were available. No p-values or claims of statistical significance were generated. Comparisons between IOL models and laser-shot conditions should consequently be interpreted as preliminary descriptive observations rather than statistically demonstrated differences.
4. Discussion
Cataract surgery with phacoemulsification and intraocular lens (IOL) implantation is currently one of the most frequently performed surgical procedures worldwide and remains the gold standard for the treatment of cataract [
1]. Posterior capsule opacification (PCO) is the most common long-term postoperative complication and continues to represent the leading indication for Nd laser posterior capsulotomy [
6,
7,
8]. Although Nd:YAG posterior capsulotomy is considered a safe and highly effective procedure for the treatment of posterior capsule opacification, postoperative visual complaints such as glare, halos, reduced contrast sensitivity, and persistent dysphotopsias continue to be reported in a subset of patients despite apparently uncomplicated procedures [
6,
7,
8,
20]. In many cases, the underlying mechanism responsible for these symptoms remains poorly understood, particularly when no clinically significant complications are evident on slit-lamp examination. Therefore, identifying the optical consequences of inadvertent IOL pitting may provide useful mechanistic information for understanding postoperative visual disturbances, although direct clinical relevance cannot be established without psychophysical testing and patient-reported outcomes [
20,
31]. A successful Nd capsulotomy depends on meticulous surgical technique, including accurate focusing on the posterior capsule, appropriate posterior offset (typically 100–250 μm), the use of the lowest effective laser energy, and minimization of the total number of laser pulses. Any anterior displacement of the focal point, excessive laser energy, or inaccurate aiming may result in unintended laser impacts on the IOL surface. The present findings indicate that even minimal laser misfocusing is sufficient to produce permanent structural damage capable of altering the optical performance of the implanted lens.
The primary objective of this study was to investigate whether laser-induced IOL pitting modifies the propagation of light through hydrophobic acrylic intraocular lenses and whether these alterations may influence retinal image formation. Using a multimodal experimental approach combining thermal imaging, scanning electron microscopy (SEM), optical bench analysis, and digital image processing, we demonstrated that laser-induced pits alter the normal optical pathway by deviating transmitted light from its expected trajectory. Thermal imaging combined with MATLAB-based digital subtraction analysis demonstrated measurable displacement of the transmitted light beam after Nd-laser exposure. Comparison between untreated and laser-treated IOLs consistently revealed displacement of the projected light spot, indicating that laser-induced surface irregularities modify the refractive behavior of the optic. Notably, this phenomenon was observed even after a limited number of laser impacts, suggesting that relatively small surface defects are sufficient to produce measurable optical alterations. From a clinical perspective, these findings suggest that even limited laser-induced surface defects may redirect incident light toward non-physiological retinal locations, thereby increasing forward light scatter without necessarily affecting conventional visual acuity. This mechanism may represent one possible contributor to persistent visual disturbances in patients with apparently minor IOL damage. However, the present ex vivo study did not evaluate visual function, contrast sensitivity, straylight perception, or patient-reported dysphotopsia and therefore cannot establish a causal relationship.
Descriptively, the two hydrophobic acrylic IOL models exhibited different optical response patterns after laser exposure. The AcrySof IQ lenses showed progressively greater displacement of the transmitted light beam (Δ-pixel analysis) with increasing numbers of laser impacts, whereas the CT LUCIA lenses demonstrated relatively stable spot displacement after the initial laser exposures. These preliminary observations raise the hypothesis that differences in polymer composition, refractive index, manufacturing process, or surface microarchitecture may influence the optical response to Nd:YAG-induced damage. However, the study was not designed to isolate the contribution of each material characteristic.
The optical bench analysis further supported these observations. PSF measurements demonstrated progressive enlargement of the central halo, increased forward light scatter, and the appearance of directional light spikes originating from laser-induced pits. Similarly, MTF analysis revealed a progressive reduction in optical quality following laser exposure, indicating deterioration of image contrast and spatial resolution. The greatest reduction in MTF was observed at intermediate spatial frequencies, which are particularly relevant for functional vision and contrast sensitivity in pseudophakic patients. In the limited samples evaluated, the AcrySof IQ lenses showed descriptively greater cumulative alterations in some optical outcomes than the CT LUCIA lenses. This observation should be considered preliminary and requires confirmation in larger experimental series.
Since both PSF enlargement and MTF reduction are closely associated with decreased retinal image quality and impaired contrast sensitivity, the present findings extend previous morphological observations by demonstrating their potential functional relevance. These optical changes provide a plausible experimental mechanism that could contribute to subjective complaints such as glare, halos, or reduced visual quality. Nevertheless, the study did not measure psychophysical visual performance or patient-reported symptoms, and no direct clinical correlation can therefore be inferred.
Previous studies of Nd:YAG-induced IOL damage have primarily focused on surface morphology and laser-damage thresholds. Newland et al. demonstrated material-dependent differences in damage formation among acrylic, PMMA, and silicone IOLs, while Borkenstein and Borkenstein characterized the morphology and optical-surface alterations produced by Nd:YAG impacts. Meduri et al. used SEM and X-ray spectrometry to document structural and compositional changes in PMMA IOLs after laser exposure. The present study extends these morphological observations by combining surface imaging with experimental assessment of light-spot displacement, PSF alteration, and MTF reduction. However, unlike standardized large-sample optical bench studies, the current investigation remains exploratory and does not establish statistically validated differences between materials.
Although most cases of IOL pitting are considered clinically insignificant, the present findings provide an optical explanation for the dysphotopsias occasionally reported after Nd capsulotomy, including glare, halos, reduced contrast sensitivity, and localized visual disturbances. Even when laser-induced pits do not directly involve the visual axis, the resulting increase in forward light scatter may contribute to subtle degradation of retinal image quality that is not readily detectable using conventional visual acuity testing alone.
Patient satisfaction after cataract surgery increasingly depends on visual quality rather than visual acuity alone. Modern pseudophakic patients frequently report subtle visual disturbances despite excellent Snellen acuity. Therefore, preservation of optical quality has become a major objective of contemporary cataract surgery and postoperative management. The present findings suggest that preventing even limited Nd:YAG-induced IOL damage may contribute to maintaining postoperative optical quality and reducing the risk of dysphotopsias.
Several limitations of this study should be acknowledged. First, the experimental model employed a simplified optical bench incorporating a retinal simulation screen and thermographic imaging, which cannot fully reproduce the complexity of retinal image formation or the optical and perceptual environment of the pseudophakic human eye. In particular, the experimental setup did not account for physiological variables such as tear-film instability, corneal and internal aberrations, aqueous and vitreous light propagation, pupil size and dynamics, IOL decentration or tilt, retinal sampling, neural processing, neuroadaptation, or the subjective perception of glare, halos, and other dysphotopsias. Consequently, the imaging plane used in this study should not be regarded as a direct surrogate for the human retina. Nevertheless, the highly controlled laboratory configuration enabled reproducible comparisons of the relative optical alterations induced by Nd:YAG laser damage. A major limitation of this study is that both the Nd:YAG laser treatment and the optical measurements were performed in air rather than in a fluid-filled model eye. Consequently, laser-induced pit morphology, cavitation dynamics, and optical scattering may differ from those occurring in vivo. Therefore, the present findings should not be directly extrapolated to the pseudophakic human eye.
Second, only two IOLs were available for each model and laser-exposure condition, precluding inferential statistical analysis. Accordingly, all findings should be interpreted as descriptive observations, and the apparent differences between CT LUCIA and AcrySof IQ lenses should be considered preliminary and hypothesis-generating rather than definitive evidence of material-dependent susceptibility.
Third, the study was conducted under ex vivo laboratory conditions, and Δ-pixel displacement represents an experimental image-plane parameter that has not been validated against clinically relevant functional outcomes such as retinal image displacement, contrast sensitivity, straylight, or patient-reported dysphotopsias. Therefore, direct extrapolation of these findings to postoperative visual performance should be avoided.
Another limitation is that the experimental setup was not calibrated to provide validated quantitative PSF metrics, such as Strehl ratio, full width at half maximum (FWHM), or encircled energy. Consequently, PSF analysis was restricted to qualitative assessment of changes in halo extension, central peak intensity, and overall light distribution. Similarly, complete numerical MTF datasets at individual spatial frequencies were not available. Optical performance was therefore evaluated descriptively using representative MTF curves rather than comprehensive quantitative optical characterization. Future investigations should employ standardized optical bench systems capable of generating validated quantitative PSF and MTF measurements.
Finally, only two hydrophobic acrylic monofocal IOL models were investigated, limiting the generalizability of the present findings to other IOL materials and designs. Despite these limitations, this study combines scanning electron microscopy, thermographic imaging, MATLAB-based image analysis, PSF assessment, and MTF evaluation within a single experimental platform, providing a comprehensive qualitative characterization of the structural and optical consequences of Nd:YAG laser-induced IOL pitting and generating hypotheses for future quantitative and clinical investigations.
To our knowledge, it is among the first investigations to combine scanning electron microscopy, infrared thermal imaging, digital image subtraction, PSF, and MTF analyses within a single experimental model to comprehensively evaluate the structural and functional consequences of Nd laser-induced IOL damage. This statement refers specifically to the combined use of these modalities within a single exploratory experimental protocol and does not imply that PSF, MTF, or optical effects of IOL damage have never been investigated previously.
Unlike previous studies that primarily described the morphology of laser-induced pits, the present work provides descriptive experimental evidence that laser-induced surface alterations are accompanied by measurable changes in light distribution and optical bench performance. Furthermore, the quantitative assessment of light beam displacement using Δ-pixel analysis provides an objective parameter linking physical surface damage with altered optical behavior. The comparative evaluation of two widely used hydrophobic acrylic IOLs also suggests material-dependent susceptibility to laser-induced optical degradation, which may have implications for future IOL design and material engineering.