GantryMate: A Modular MR-Compatible Assistance System for MR-Guided Needle Interventions

by
Andreas Reichert
1,*,
Michael Bock
1,
Michael Vogele
2 and
Axel Joachim Krafft
1
1
Department of Radiology, Medical Physics, Medical Center – University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany
2
Interventional Systems GmbH, Kitzbühel, Austria
*
Author to whom correspondence should be addressed.
Tomography 2019, 5(2), 266-273; https://doi.org/10.18383/j.tom.2019.00007
Submission received: 7 March 2019 / Revised: 3 April 2019 / Accepted: 3 May 2019 / Published: 1 June 2019

Abstract

Percutaneous minimally invasive interventions are difficult to perform in closed-bore high-field magnetic resonance systems owing to the limited space between magnet and patient. To enable magnetic resonance–guided needle interventions, we combine a small, patient-mounted assistance system with a real-time instrument tracking sequence based on a phase-only cross-correlation algorithm for marker detection. The assistance system uses 2 movable plates to align an external passive marker with the anatomical target structure. The targeting accuracy is measured in phantom experiments, yielding a precision of 1.7 ± 1.0 mm for target depths up to 38 ± 13 mm. In in vivo experiments, the possibility to track and target static and moving structures is demonstrated.
Keywords: interventional magnetic resonance imaging; MR-guided interventional procedures; percutaneous procedures; MRI-compatible assistance system interventional magnetic resonance imaging; MR-guided interventional procedures; percutaneous procedures; MRI-compatible assistance system

Share and Cite

MDPI and ACS Style

Reichert, A.; Bock, M.; Vogele, M.; Krafft, A.J. GantryMate: A Modular MR-Compatible Assistance System for MR-Guided Needle Interventions. Tomography 2019, 5, 266-273. https://doi.org/10.18383/j.tom.2019.00007

AMA Style

Reichert A, Bock M, Vogele M, Krafft AJ. GantryMate: A Modular MR-Compatible Assistance System for MR-Guided Needle Interventions. Tomography. 2019; 5(2):266-273. https://doi.org/10.18383/j.tom.2019.00007

Chicago/Turabian Style

Reichert, Andreas, Michael Bock, Michael Vogele, and Axel Joachim Krafft. 2019. "GantryMate: A Modular MR-Compatible Assistance System for MR-Guided Needle Interventions" Tomography 5, no. 2: 266-273. https://doi.org/10.18383/j.tom.2019.00007

Article Metrics

Back to TopTop