2.2. Sample Preparation, ESR Measurements and Evaluation of Results
All spectra were taken in standard 5 mm outer diameter quartz ESR sample tubes on a Bruker EMX X-band spectrometer with an E4105DR double resonator at room temperature, 2 mW microwave power, and 0.05–0.15 mT modulation at 100 kHz. Four working samples were prepared: 1.00 mg dry NN-polyphenylene (MW* = 628–703), 0.25 mg dry NN-GNR (MW* = 616–692), 0.27 mg dry cupric acetylacetonate (Cu(acac)2) (MW = 262), 3.0 µg 2,2-diphenyl-1-picrylhydrazyl (DPPH) (MW = 394) in 20 µL toluene. The sample mass was determined by weighing an empty tube on laboratory balances (Ohaus Analytical Plus) to an accuracy of 0.01 mg, putting the sample powder or liquid aliquot inside and weighing the tube again.
As preliminary experiments demonstrated (see below), ESR spectra of both NN-polyphenylene and NN-GNR consisted of a narrow signal in the region g = 2 and a very broad and weak background that on the one hand is practically unavoidable, while on the other hand can hide a broad signal from bulk paramagnetic powder [
11]. Therefore, two standard samples for the absolute spin count were prepared. The copper salt, Cu(acac)
2, was used as a standard due to its wide ESR spectrum. It was taken in an amount corresponding to the order of magnitude to the degree of polymer spin labeling (number of spins per molecular unit) of about 100%, to check whether any substantial ESR intensity was hidden in the background. The other reference compound, DPPH, is a standard with a narrow ESR spectrum. It was taken in an amount corresponding to the order of magnitude to the degree of polymer spin labeling of about 1% to check the absolute numbers of spins which generated the intensity contained in the narrow spectrum. Since the required amount of this low-molecular-weight compound was too small for reliable weighing, a solution was prepared and used for sample preparation. While DPPH is stable as a dry powder, DPPH solution in toluene gradually degrades with a characteristic lifetime of about one week, changing color from purple-blue to pale yellow. Fresh solution and sample were prepared as needed. The sample could be considered as a stable standard over the course of measurement taken over the course of several hours.
Additionally, two witnesses having a wide and a narrow ESR spectra of intensities comparable to the four samples were prepared using small crystals of dry Cu(acac)2 and DPPH, respectively, without noting their exact masses. To avoid additional ambiguities related to different sensitivities of the two sections of the double resonator, the spectra were taken using the witnesses with similar spectral types, Cu(acac)2 for wide spectra or DPPH for narrow spectra, as follows. One of the witnesses was placed in the reference chamber (front section) of the double resonator, then the four samples were placed into the other chamber. The spectrometer was tuned and pairs of the spectra for the witness and for the sample were recorded. The series was then repeated with the other witness. The measurements were independently repeated over three days, each time preparing a fresh sample of DPPH. All spectra were recorded in identical conditions within the series with wide and narrow witnesses, covering the range of 240 mT for the wide witness and 10 mT for the narrow witness centered at g = 2. The spectra from the samples were normalized by the maximum of the spectrum of the witness for the narrow witness series and by the second integral of the spectrum of the witness for the wide witness series. In the spectra and their integrals, given below, the absolute scales of vertical axes can be directly compared within each group of spectra of the same type.
The described procedure of measuring with standards and against witnesses may seem unnecessarily complicated, but this was done deliberately to handle the problem of the broad background as effectively as possible. The very weak and very broad signals arise due to a combination of factors, including the resonator, sample tubes and eddy currents, and are present as the slowly varying baseline in any real instrument. The resonator is clean, but still has traces of paramagnetic ions from the ambient atmosphere on its walls, as does the waveguide. Sample tubes, although made from suprasil and having calibrated dimensions, still establish slightly different distributions of microwave field in the cavity. Even different repeats (with sample tube removal and installation in the cavity) of the same sample yielded slightly varying backgrounds at the scales reported in this work. To demonstrate what provides information and what is unavoidable natural variation, two characteristic series of (double-integrated) spectra are shown in Figure 3. Furthermore, the background was not identical in the two chambers of the dual resonator, which is the reason for performing the experiments with additional witnesses, in such a manner so as to never compare spectra from different resonators. Instead, a sample with a spectrum similar to that which one would expect was placed in the reference cavity, and then all the analyzed samples—including standards—were placed, in turn, in the working cavity, measured, and normalized to this witness. This completely bypasses the problem of different and unaccountable variations in the backgrounds in two halves of the resonator. Finally, the background was not subtracted before performing integrations of the spectra, because in this case it was more detrimental than helpful. Instruments normally subtract backgrounds automatically, which is very convenient when working with commonly recorded narrow spectra. However, when dealing with very weak and broad spectra—for which a second additional integral will be needed—subtracting a background can be disastrous, as very slight variations in what is subtracted can produce wildly varying results after two integrations. It is much safer to leave everything as recorded and judge the results by eye
The purpose of performing these experiments that have issues with a noisy background was to verify that there is not much useful signal, if any, contained in the broad spectrum, so that the quantitative estimates can then be confined to much less complicated narrow spectra. A broad spectrum could, in principle, have been expected for these systems, as they might have some form of ferromagnetic resonance signals. As described further below, none were found (at least at room temperature), and so the observed narrow spectra could indeed be interpreted as spectra from localized spin labels that bear the useful signal. Therefore, the relative degrees of spin labeling (R) were estimated from the second integrals of the narrow spectra (I), the masses of the samples (m) and molecular weights of the compounds (M) as R1/R2 = (I1M1m2)/(I2M2m1). Then, taking the degree R for DPPH as 100%, the degrees R for NN-polyphenylene and NN-GNR were estimated as 0.8% and 1.3%, respectively.
The evaluation described in detail above was performed in our Novosibirsk Lab. The same samples were also independently evaluated in a different laboratory in the UK on a different Bruker EMX system to give a radical substitution of 1.2% for NN-polyphenylene and a radical substitution of 1.3% for NN-GNR (preliminary data from Prof. Lapo Bogani, University of Oxford). This data is based on the double integral and a 3-D EPR image determination by Bruker. This match between two independent evaluations in different laboratories is most encouraging, and lends us a certain degree of confidence in these numbers and the procedure itself. Therefore, it was deemed desirable to be prepared for this publication. This is certainly just an estimate and only provides a ballpark value, so no error evaluation for the given values was attempted. A reader interested in the possibility of the statistically consistent evaluation of quantitative ESR measurements is referred to works by Nicola Yordanov [
12,
13,
14] and to [
15]. The next section provides a concise step-by-step exposition of our study.