In Vitro Comparison of Initial and 7-Day System-Level Resultant Forces in Clear Aligners with Different Layer Structures
Abstract
1. Introduction
2. Materials and Methods
3. Results
3.1. Descriptive Time Course of Resultant Force
3.2. Endpoint Resultant Forces Under Attachment-Free and Attachment-Integrated Conditions
4. Discussion
5. Conclusions
Supplementary Materials
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Weir, T. Clear aligners in orthodontic treatment. Aust. Dent. J. 2017, 62, 58–62. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wiechens, B.; Meyer-Marcotty, P.; von Waldthausen, E.; Mai, L.-M.; Schmid, J.Q.; Brockmeyer, P.; Hahn, W.; Quast, A. Biomechanical force profiles and force decrease of direct printed versus thermoformed aligners with different trimlines: An in vitro analysis. Prog. Orthod. 2026, 27, 23. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Azaripour, A.; Weusmann, J.; Mahmoodi, B.; Peppas, D.; Gerhold-Ay, A.; Van Noorden, C.J.F.; Willershausen, B. Braces versus Invisalign®: Gingival parameters and patients’ satisfaction during treatment: A cross-sectional study. BMC Oral Health 2015, 15, 69. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jiang, Q.; Li, J.; Mei, L.; Du, J.; Levrini, L.; Abbate, G.M.; Li, H. Periodontal health during orthodontic treatment with clear aligners and fixed appliances: A meta-analysis. J. Am. Dent. Assoc. 2018, 149, 712–720.e12. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yassir, Y.A.; Nabbat, S.A.; McIntyre, G.T.; Bearn, D.R. Clinical effectiveness of clear aligner treatment compared to fixed appliance treatment: An overview of systematic reviews. Clin. Oral Investig. 2022, 26, 2353–2370. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lombardo, L.; Martines, E.; Mazzanti, V.; Arreghini, A.; Mollica, F.; Siciliani, G. Stress relaxation properties of four orthodontic aligner materials: A 24-hour in vitro study. Angle Orthod. 2017, 87, 11–18. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Albertini, P.; Mazzanti, V.; Mollica, F.; Pellitteri, F.; Palone, M.; Lombardo, L. Stress relaxation properties of five orthodontic aligner materials: A 14-day in vitro study. Bioengineering 2022, 9, 349. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sayahpour, B.; Zinelis, S.; Polychronis, G.; Eliades, T.; Goteni, M.; Kopp, S.; Eslami, S. Effects of intraoral aging on mechanical properties of directly printed aligners vs. thermoformed aligners: An in vivo prospective investigation. Eur. J. Orthod. 2024, 46, cjad063. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rossini, G.; Parrini, S.; Castroflorio, T.; Deregibus, A.; Debernardi, C.L. Efficacy of clear aligners in controlling orthodontic tooth movement: A systematic review. Angle Orthod. 2015, 85, 881–889. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Haouili, N.; Kravitz, N.D.; Vaid, N.R.; Ferguson, D.J.; Makki, L. Has Invisalign improved? A prospective follow-up study on the efficacy of tooth movement with Invisalign. Am. J. Orthod. Dentofac. Orthop. 2020, 158, 420–425. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Castroflorio, T.; Parrini, S.; Rossini, G. Aligner biomechanics: Where we are now and where we are heading for. J. World Fed. Orthod. 2024, 13, 57–64. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kohda, N.; Iijima, M.; Muguruma, T.; Brantley, W.A.; Ahluwalia, K.S.; Mizoguchi, I. Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances. Angle Orthod. 2013, 83, 476–483. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Alexandropoulos, A.; Al Jabbari, Y.S.; Zinelis, S.; Eliades, T. Chemical and mechanical characteristics of contemporary thermoplastic orthodontic materials. Australas. Orthod. J. 2015, 31, 165–170. [Google Scholar] [CrossRef] [Scilit]
- Elshazly, T.M.; Bourauel, C.; Ismail, A.; Ghoraba, O.; Aldesoki, M.; Salvatori, D.; Elattar, H.; Alhotan, A.; Alkabani, Y. Effect of material composition and thickness of orthodontic aligners on the transmission and distribution of forces: An in vitro study. Clin. Oral Investig. 2024, 28, 258. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- U.S. Food and Drug Administration. 510(k) Premarket Notification K233544: MESHEET; U.S. Food and Drug Administration: Silver Spring, MD, USA, 2024. Available online: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMN/pmn.cfm?ID=K233544 (accessed on 21 August 2026).
- Bay Materials, LLC. Zendura FLX® Thermoformable Sheet Materials: Instructions for Use with Pressure Thermoforming Equipment; IFU 1003, Rev. F; Bay Materials, LLC: Fremont, CA, USA, 2026. [Google Scholar]
- Takara, Y.; Haga, S.; Kimura, H.; Maki, K. Mechanical analysis of factors affecting clear aligner removability. Dent. Mater. J. 2022, 41, 534–544. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Seo, J.-H.; Kim, M.-S.; Lee, J.-H.; Eghan-Acquah, E.; Jeong, Y.-H.; Hong, M.-H.; Kim, B.; Lee, S.-J. Biomechanical efficacy and effectiveness of orthodontic treatment with transparent aligners in mild crowding dentition—A finite element analysis. Materials 2022, 15, 3118. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Seo, J.-H. Biomechanical Efficacy and Effectiveness of Orthodontic Treatment with Transparent Aligners in Severe Crowding Dentition. Ph.D. Thesis, Inje University, Gimhae, Republic of Korea, 2021. (In Korean) [Google Scholar]
- ATI Industrial Automation. Nano17 Force/Torque Sensor. Available online: https://www.ati-ia.com/products/ft/ft_models.aspx?id=Nano17 (accessed on 21 August 2026).
- Taheri, R.; García-Marín, C.; Aizpuru-Arotzena, M.; Iglesias-Linares, A. 3D volumetric changes and force delivery at the attachment region of different aligner materials. Clin. Oral Investig. 2026, 30, 106. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nakornnoi, T.; Srirodjanakul, W.; Chintavalakorn, R.; Santiwong, P.; Sipiyaruk, K. The biomechanical effects of clear aligner trimline designs and extensions on orthodontic tooth movement: A systematic review. BMC Oral Health 2024, 24, 1523. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ren, Y.; Maltha, J.C.; Kuijpers-Jagtman, A.M. Optimum force magnitude for orthodontic tooth movement: A systematic literature review. Angle Orthod. 2003, 73, 86–92. [Google Scholar] [CrossRef] [Scilit]
- Hahn, W.; Engelke, B.; Jung, K.; Dathe, H.; Fialka-Fricke, J.; Kubein-Meesenburg, D.; Sadat-Khonsari, R. Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor. Angle Orthod. 2010, 80, 239–246. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Elkholy, F.; Panchaphongsaphak, T.; Kilic, F.; Schmidt, F.; Lapatki, B.G. Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation. J. Orofac. Orthop. 2015, 76, 460–475. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Toms, S.R.; Eberhardt, A.W. A nonlinear finite element analysis of the periodontal ligament under orthodontic tooth loading. Am. J. Orthod. Dentofac. Orthop. 2003, 123, 657–665. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ryokawa, H.; Miyazaki, Y.; Fujishima, A.; Miyazaki, T.; Maki, K. The mechanical properties of dental thermoplastic materials in a simulated intraoral environment. Orthod. Waves 2006, 65, 64–72, Erratum in Orthod. Waves 2006, 65, 173. https://doi.org/10.1016/j.odw.2006.09.002. [Google Scholar] [CrossRef] [Scilit]





| Material | Manufacturer | Architecture/Composition | Thickness Before Thermoforming | Regulatory/Source Note |
|---|---|---|---|---|
| DURAN® | Scheu-Dental GmbH, Iserlohn, Germany | Single-layer PET-G | 0.75 mm total | Commercial sheet used in the study |
| Zendura FLX | Bay Materials LLC., Fremont, CA, USA | Copolyester/elastomeric polyurethane/copolyester | 0.76 mm nominal total thickness | Manufacturer IFU [16] |
| MESHEET | DENTIS Co., Ltd., Daegu, Republic of Korea | PET-G (outer, non-tooth-contacting)/mesh-structured TPU (middle)/PET-G (inner, tooth-contacting) | 0.25/0.20/0.30 mm (0.75 mm total) | Sheet specification used in this study; FDA 510(k) clearance [15] |
| Material | Attachment | Day 0 Mean ± SD (N) | Day 0 95% CI (N) | Day 7 Mean ± SD (N) | Day 7 95% CI (N) |
|---|---|---|---|---|---|
| DURAN | Without | 2.386 ± 0.328 | 2.092–2.680 | 1.298 ± 0.277 | 1.098–1.498 |
| ZENDURA FLX | Without | 2.511 ± 0.217 | 2.217–2.805 | 1.413 ± 0.199 | 1.213–1.613 |
| MESHEET | Without | 2.306 ± 0.411 | 2.012–2.600 | 1.291 ± 0.221 | 1.091–1.491 |
| DURAN | With | 2.914 ± 0.255 | 2.620–3.208 | 1.415 ± 0.186 | 1.215–1.615 |
| ZENDURA FLX | With | 3.211 ± 0.338 | 2.917–3.505 | 1.664 ± 0.197 | 1.464–1.864 |
| MESHEET | With | 2.806 ± 0.326 | 2.512–3.100 | 1.558 ± 0.206 | 1.358–1.758 |
| Endpoint | Effect | df | F | p | Partial η2 |
|---|---|---|---|---|---|
| Day 0 (0 h) | Material | 2, 24 | 2.403 | 0.112 | 0.167 |
| Day 0 (0 h) | Attachment | 1, 24 | 24.510 | <0.001 | 0.505 |
| Day 0 (0 h) | Material × Attachment | 2, 24 | 0.289 | 0.752 | 0.024 |
| Day 7 (168 h) | Material | 2, 24 | 1.806 | 0.186 | 0.131 |
| Day 7 (168 h) | Attachment | 1, 24 | 7.175 | 0.013 | 0.230 |
| Day 7 (168 h) | Material × Attachment | 2, 24 | 0.362 | 0.700 | 0.029 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Seo, J.-H.; Park, E.-Y.; Lee, J.-H.; Kim, W.-G.; Lee, S.-J.; Kim, B. In Vitro Comparison of Initial and 7-Day System-Level Resultant Forces in Clear Aligners with Different Layer Structures. Bioengineering 2026, 13, 1081. https://doi.org/10.3390/bioengineering13091081
Seo J-H, Park E-Y, Lee J-H, Kim W-G, Lee S-J, Kim B. In Vitro Comparison of Initial and 7-Day System-Level Resultant Forces in Clear Aligners with Different Layer Structures. Bioengineering. 2026; 13(9):1081. https://doi.org/10.3390/bioengineering13091081
Chicago/Turabian StyleSeo, Jeong-Hee, Eun-Yeong Park, Jeong-Hyeon Lee, Won-Gi Kim, Sung-Jae Lee, and Bongju Kim. 2026. "In Vitro Comparison of Initial and 7-Day System-Level Resultant Forces in Clear Aligners with Different Layer Structures" Bioengineering 13, no. 9: 1081. https://doi.org/10.3390/bioengineering13091081
APA StyleSeo, J.-H., Park, E.-Y., Lee, J.-H., Kim, W.-G., Lee, S.-J., & Kim, B. (2026). In Vitro Comparison of Initial and 7-Day System-Level Resultant Forces in Clear Aligners with Different Layer Structures. Bioengineering, 13(9), 1081. https://doi.org/10.3390/bioengineering13091081