Design, Calibration, and On-Site Validation of an LCVR-Driven Fast-Tunable Lyot Filter for the YOGIS Coronagraph
Abstract
1. Introduction
2. Design of the YOGIS-Lyot Filter
2.1. Design Principle
2.2. Filter Bandwidth Properties
2.3. Optical Configuration
2.4. Effects of Orientation Errors of Polarizers and Wave Plates
2.5. Transmission Band Wavelength Tuning
2.5.1. Tuning Mechanism
2.5.2. LCVR Configuration and Phase Control
- Independent phase adjustment: Each LCVR modulates the phase of its corresponding Lyot unit, enabling targeted shifts in the transmission band without affecting other stages;
- Proportional phase modulation: Maintaining a 2:1 ratio of phase shifts between the 1 Å and 2 Å units ensures a single transmission peak that can be tuned within the ±2 Å range, with minimal sideband leakage (Figure 5).
2.5.3. Transmission Band Modes and Application Scenarios
- 1.
- Single-peak mode: Both Lyot units are phase-aligned to center the transmission peak at 5303 Å, enabling high-SNR imaging of the coronal green line core;
- 2.
- Dual-peak mode: A 180° phase shift is applied to the 2 Å Lyot unit, generating two symmetric peaks separated by 2 Å. This mode is used to measure the continuous spectrum of atmospheric scattered light (sky background) for subsequent subtraction;
- 3.
- Line-wing mode: By tuning the phase shift ratio of the two LCVRs, the single transmission peak is offset by Å relative to 5303 Å, capturing images of the green line’s wings. These images are critical for calculating the Doppler shift of coronal plasma.
2.5.4. Tuning Performance
2.6. Calibration of LCVRs
2.6.1. Rationale for Calibration
- 1.
- Voltage nonlinearity: The phase retardation () does not vary linearly with driving voltage (Figure 6), due to non-uniform liquid crystal alignment under electric fields. Without calibration, this would lead to deviations from target wavelengths (e.g., ±0.45 Å line wings).
- 2.
- Fast-axis misalignment: The LCVR’s fast-axis azimuth () may shift relative to the Lyot unit’s polarization components, reducing interference contrast and lowering peak transmittance.
- 3.
- Temperature drift: The operating temperature of LCVRs differs from factory calibration conditions, altering liquid crystal birefringence and requiring temperature-compensated calibration.
2.6.2. Calibration Methodology
- 1.
- Sample Integration: The LCVR under test is mounted between the MMSP’s polarization generator and analyzer, ensuring alignment with the optical axis.
- 2.
- Polarization Modulation and Data Acquisition: The rotating stage adjusts the fast-axis angles of the generator and analyzer, generating a sequence of polarized light inputs. The spectrometer records light intensity data across 5290–5310 Å (centered on 5303 Å) for each polarization state.
- 3.
- Mueller Matrix Extraction: Using Stokes vector inversion, the LCVR’s full 4 × 4 Mueller matrix is derived from intensity data—quantifying and at each wavelength.
- 4.
2.6.3. Calibration Outcomes and Application
- Performance metrics: At 5303 Å and 40.6 °C, the calibrated LCVRs achieve a maximum of ±180° (covering the Å tuning range) with a fitting accuracy of . Fast-axis deviation is reduced to <0.5°, minimizing polarization loss. Figure 6 shows the finally calibrated response curves of actual voltage versus phase retardation angle for the two LCVRs.
- On-site voltage calculation: During observation, the required (e.g., 63° for −0.45 Å wing, Table 2) is converted to driving voltage via:
- 1.
- Base voltage calculation using polynomial coefficients (Equation (7)):
- 2.
- Temperature compensation using and real-time temperature (Equation (8)):
This ensures precise control for fast wavelength tuning.
2.7. Transmission Band Wavelength Calibration
2.7.1. Calibration Principle and Equipment
2.7.2. Calibration Workflow
- 1.
- Transmission profile alignment:
- Fix the driving voltage of the 2 Å Lyot unit (2–4 V) to maintain a stable reference transmission band.
- Adjust the LCVR voltage of the 1 Å unit to shift its transmission band until it aligns with the 2 Å unit—resulting in a symmetric main peak, maximum peak intensity, and minimal sideband leakage.
- 2.
- Wavelength zero-point correction:
- Calculate the wavelength offset () between the aligned peak and the reference 5303 Å zero-point.
- Use Equation (9) () to compute the required for LCVRs, then adjust voltages via calibrated coefficients to center the transmission band at 5303 Å.
2.7.3. Statistical Analysis of the Errors on Filter Performance
2.7.4. Calibration Results
- Central wavelength accuracy: Deviation from 5303 Å is <0.1 Å, meeting the 1 Å FWHM requirement.
- Mode-specific phase settings: Calibrated values for key observation modes are documented in Table 4 (e.g., Single mode: LCVR1 = −18°, LCVR2 = 40°; +0.45 Å wing: LCVR1 = 261°, LCVR2 = −0.5°), enabling direct recall during on-site observation.
2.8. Electronic Control System
2.8.1. Thermostatic Control Subsystem
- 1.
- Design RationaleTemperature variations alter the wave plate length and refractive index, resulting in band shifts [13]:Around , a temperature increase of 1°C induces an approximate band shift of in calcite. Therefore, for narrowband applications, to ensure the stability of the transmission band wavelength, it is generally necessary to equip a high-precision thermostatic device and house the entire filter assembly in a thermostat to enhance stability. As long as the temperature control accuracy of the thermostatic device is within 0.1 °C, the impact caused by temperature drift of the entire filter is negligible.
- 2.
- Hardware and Performance
- Sensing and actuation: Three platinum resistance temperature detectors (RTDs) are embedded between the filter core sleeve and optical assembly to monitor temperature in real time. Heating resistors (wrapped around the sleeve) and a thermal insulation layer ensure uniform heating.
- Control algorithm: A proportional–integral–derivative (PID) controller locks the chamber temperature at 40.6 °C (higher than Lijiang Observatory’s ambient ∼15 °C, enabling passive cooling) with a stability precision of 0.01 °C.
- Integration: The subsystem connects to the host computer via a reserved RS232 interface, enabling remote temperature monitoring and parameter adjustment.
2.8.2. LCVR Voltage Regulation Subsystem
- 1.
- Design Constraints and SolutionsLCVR liquid crystal molecules are damaged by long-term DC exposure; thus, the subsystem uses AC square-wave signals (2 kHz frequency, 50% duty cycle) to avoid molecular orientation degradation. The 2 kHz frequency exceeds the liquid crystal response speed (1 ms), ensuring stable control.
- 2.
- Hardware and Performance
- Signal generation: A custom FPGA-based multi-channel signal generator provides ±10 V adjustable amplitude signals, with 16-bit AD converters enabling 1 mV voltage resolution and <5 mV DC offset (via automatic DC balance).
- Temperature feedback: Two integrated temperature sensors (attached to LCVRs) monitor real-time LCVR temperature, enabling dynamic voltage compensation via Equation (8) to counter thermal drift.
- Expandability: The system supports six-channel waveform output (for future filter upgrades) and communicates with the host computer via RS232, enabling synchronized control with the YOGIS v1.0 observation software.
2.8.3. System Integration
3. On-Site Validation with the YOGIS
3.1. Validation System Configuration
3.2. Validation Methodology
- 1.
- System Preparations: The YOGIS was aligned with the Sun to ensure stable light input. The filter’s thermostatic chamber was preheated for 2 h to reach thermal equilibrium; calibration coefficients and target LCVR phase retardation values were loaded into the control software for one-click mode switching.
- 2.
- Multi-Mode Sequential Imaging: Leveraging the filter’s fast-tuning capability, four key observation bands were captured in sequence (total imaging time: <10 s per cycle to avoid solar activity drift).
- 3.
- Image Post-Processing:
- Dark current correction: Dark-field images (telescope cover closed, same exposure time) were subtracted to eliminate sensor noise.
- Background subtraction: S-D (Single-Double) image fusion was performed to suppress atmospheric scattered light and instrumental stray light.
- SNR quantification: SNR was calculated as , where = mean gray value of coronal signal regions, = mean gray value of off-limb background regions, and = standard deviation of the background.
3.3. Validation Results
- 1.
- Wavelength Tuning Accuracy and Speed
- Tuning accuracy: Spectral analysis showed that the filter’s actual central wavelength deviated by <0.2 Å (single-peak mode, <0.05 Å, dual-peak mode, <0.2 Å) from the target in all modes—consistent with the transmission band calibration results, confirming the effectiveness of the LCVR calibration and temperature compensation model.
- Tuning speed: The average switching time between any two modes was ms, meeting the design requirement (<60 ms) and outperforming traditional mechanical-tunable filters (switching time: 10–20 s). This fast tuning enabled sequential imaging of four bands within 10 s, avoiding artifacts from coronal loop evolution.
- 2.
- Image SNR and Quality
- Core mode (single mode): SNR of coronal loops in the range reached (low solar activity) and (moderate solar activity)—a 2.3 and 2.1 improvement, respectively, compared to the original YOGIS (SNR: ). This validates the filter’s 1 Å FWHM narrowband transmission and background suppression capability.
- Wing modes: The calculation of the Doppler velocity field is given by: , where v denotes the Doppler velocity, represents the wavelength shift of the spectral line, is the central wavelength of the spectral line in the rest frame, and c stands for the speed of light. Residual asymmetry between the Å wings may introduce Doppler velocity bias. The Doppler velocity v exhibits a linear proportional relationship with the spectral line wavelength shift . The line wing employs single-peak modulation with a modulation error of less than , and the deviation of the Doppler velocity field is also within . Through the superposition processing of multiple images, this deviation will not affect the calculation results. Despite weaker line wing emission, the SNR of Å wing images remained (low activity) and (moderate activity)—sufficient for Doppler shift calculation. Figure 16 presents the measured image of the coronal Doppler velocity field.
- Background suppression: The S-D fused images reduced background noise by 72% compared to single-mode images alone, enabling detection of weak coronal plumes in the range (previously undetectable with the original system).
4. Conclusions
- The optical design of the filter is fundamental to its exceptional performance. By employing a four-stage cascaded structure (FWHMs: 1 Å, 2 Å, 4 Å, 8 Å) in conjunction with two nematic liquid crystal variable retarders (LCVRs), the system achieves non-mechanical wavelength tuning. This design mitigates sensitivity to vibrations and the slow switching associated with mechanical systems, resulting in a tuning range of Å centered around 5303 Å and a tuning time of less than 60 ms. The compact and lightweight optical assembly, combined with a thermostatic chamber, facilitates seamless integration with the 10 cm aperture YOGIS coronagraph while ensuring stable operation in high-altitude (3200 m) environments and under significant diurnal temperature variations.
- A two-step calibration workflow guarantees precise regulation of wavelength. The calibrated phase retardation values for key modes (Single, Double, Å wings) facilitate one-click switching during on-site observations, thereby ensuring reliable performance.
- On-site validation at the Lijiang Observatory confirmed the filter’s full compatibility with the YOGIS optical system, thereby enabling rapid multi-band sequencing of the green-line core, wings, and sky-background channels. The resulting coronal images exhibit a notable enhancement in performance, with the core-band signal-to-noise ratio (SNR) improving by a factor of 2–2.3 compared to the original system. Background-suppressed images obtained in single–double mode further reveal faint coronal structures extending to , thereby validating the effectiveness of the fast-tuning and dual-channel acquisition strategy. The integration of the filter’s control system with the YOGIS observation software allows for coordinated multi-device control, thereby facilitating the efficient acquisition of coronal images and Doppler velocity field data.
Author Contributions
Funding
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Gong, Q.; Socker, D. Theoretical study of the occulted solar coronagraph. In Proceedings of the Optical Systems Degradation, Contamination, and Stray Light: Effects, Measurements, and Control, SPIE, Denver, CO, USA, 15 October 2004; Volume 5526, pp. 208–219. [Google Scholar]
- Zhang, H.X.; Lu, Z.W.; Xia, L.D.; Liu, H.; Li, P. Stray light suppressing of optical system in white light coronagraph. Opt. Precis. Eng. 2009, 17, 2371. [Google Scholar]
- Sun, M.; Zhang, H.; Bu, H.; Wang, X.; Ma, J.; Lu, Z. Computation of the diffracted field of a toothed occulter by the semi-infinite rectangle method. J. Opt. Soc. Am. A 2013, 30, 2140–2149. [Google Scholar] [CrossRef] [PubMed][Green Version]
- Liu, D.Y.; Zhang, H.X.; Sun, M.Z.; Huang, Z.H.; Xia, L.D.; Liu, W.X.; Fu, H. A New Method for Monitoring Scattered Stray Light of an Inner-occulted Coronagraph. Res. Astron. Astrophys. 2024, 24, 025020. [Google Scholar] [CrossRef]
- Zhang, X.F.; Liu, Y.; Zhao, M.Y.; Song, T.F.; Wang, J.X.; Li, X.B.; Li, Z.H. On the relation between coronal green line brightness and magnetic fields intensity. Res. Astron. Astrophys. 2022, 22, 075007. [Google Scholar] [CrossRef]
- Beckers, J.M.; Dickson, L.; Joyce, R.S. Observing the sun with a fully tunable Lyot-Ohman filter. Appl. Opt. 1975, 14, 2061–2066. [Google Scholar] [CrossRef] [PubMed]
- Kopp, G.A. Tunable birefringent filters using liquid crystal variable retarders. In Proceedings of the Polarization Analysis and Measurement II, SPIE, San Diego, CA, USA, 14 September 1994; Volume 2265, pp. 193–201. [Google Scholar]
- Tomczyk, S.; Card, G.; Darnell, T.; Elmore, D.; Lull, R.; Nelson, P.; Streander, K.; Burkepile, J.; Casini, R.; Judge, P. An instrument to measure coronal emission line polarization. Sol. Phys. 2008, 247, 411–428. [Google Scholar] [CrossRef]
- Wang, X. Study on the Measurement Method of Solar Photosphere Magnetic Field Based on Liquid Crystal Filter. Ph.D. Thesis, University of Chinese Academy of Sciences, Beijing, China, 2023. [Google Scholar]
- Bei, L.; Dennis, G.I.; Miller, H.M.; Spaine, T.W.; Carnahan, J.W. Acousto-optic tunable filters: Fundamentals and applications as applied to chemical analysis techniques. Prog. Quantum Electron. 2004, 28, 67–87. [Google Scholar] [CrossRef]
- Glebov, A.L.; Mokhun, O.; Rapaport, A.; Vergnole, S.; Smirnov, V.; Glebov, L.B. Volume Bragg gratings as ultra-narrow and multiband optical filters. In Proceedings of the Micro-Optics 2012, SPIE, Brussels, Belgium, 8 May 2012; Volume 8428, pp. 42–52. [Google Scholar]
- Cao, W.; Denker, C.J.; Wang, H.; Ma, J.; Qu, M.; Wang, J.; Goode, P.R. Characteristic evaluation of a near-infrared Fabry-Pérot filter for the InfraRed Imaging Magnetograph (IRIM). In Proceedings of the Telescopes and Instrumentation for Solar Astrophysics, SPIE, San Diego, CA, USA, 4 February 2004; Volume 5171, pp. 307–315. [Google Scholar]
- Ai, G. Birefringent filter and its application to astronomy. Prog. Astron. 1987, 5, 317–329. [Google Scholar]
- Shabtay, G.; Eidinger, E.; Zalevsky, Z.; Mendlovic, D.; Marom, E. Tunable birefringent filters—Optimal iterative design. Opt. Express 2002, 10, 1534–1541. [Google Scholar] [PubMed]
- Miller, P.J.; Hoyt, C.C. Multispectral imaging with a liquid crystal tunable filter. In Proceedings of the Optics in Agriculture, Forestry, and Biological Processing, SPIE, Boston, MA, USA, 6 January 1995; Volume 2345, pp. 354–365. [Google Scholar]
- Hou, J.F.; Deng, Y.Y.; Wang, D.G.; Lin, J.B.; Zhang, Y.; Sun, W.J.; Zhang, Z.Y.; Sun, Y.Z. Application and prospect of liquid crystal modulation technology in solar magnetic field detection. Spacecr. Environ. Eng. 2021, 38, 296–304. [Google Scholar]
- Hagino, M.; Ichimoto, K.; Kimura, G.; Nakatani, Y.; Kawate, T.; Shinoda, K.; Suematsu, Y.; Hara, H.; Shimizu, T. Development of a universal tunable filter for future solar observations. In Proceedings of the Advances in Optical and Mechanical Technologies for Telescopes and Instrumentation, SPIE, Montréal, QC, USA, 28 July 2014; Volume 9151, pp. 1977–1984. [Google Scholar]
- Liu, Y.; Zhang, X. The coronal green line monitoring: A traditional but powerful tool for coronal physics. Proc. Int. Astron. Union 2018, 13, 169–170. [Google Scholar] [CrossRef]
- Sha, F.; Liu, Y.; Zhang, X.; Song, T. Characterization and correction of the scattering background produced by dust on the objective lens of the Lijiang 10-cm coronagraph. Sol. Phys. 2023, 298, 139. [Google Scholar] [CrossRef]
- Song, T.; Liu, Y.; Zhang, X.; Zhao, M.; Li, X.; Luo, Q.; Sha, F.; Liu, Q.; Oloketuyi, J.; Wang, X. Toward Automated Coronal Observations: A New Integrated System Based on the Lijiang 10 cm Coronagraph. Universe 2025, 11, 154. [Google Scholar] [CrossRef]
- Xin, Y.X.; Bai, J.M.; Lun, B.L.; Fan, Y.F.; Wang, C.J.; Liu, X.W.; Yu, X.G.; Ye, K.; Song, T.F.; Chang, L.; et al. Astronomical site monitoring system at Lijiang observatory. Res. Astron. Astrophys. 2020, 20, 149. [Google Scholar] [CrossRef]
- Ichimoto, K.; Noguchi, M.; Tanaka, N.; Kumagai, K.; Shinoda, K.; Nishino, T.; Fukuda, T.; Sakurai, T.; Takeyama, N. A new imaging system of the corona at norikura. Publ. Astron. Soc. Jpn. 1999, 51, 383–391. [Google Scholar] [CrossRef]
- Wang, X.; Liang, Y.; Zhang, J.; Wu, J.; Yuan, S.; Fu, Y.; Lü, Z.; Mao, W.; Jin, Z. Development of Rapid Tunable Lyot Filter and Its Observation on 1 m New Vacuum Solar Telescope. Acta Opt. Sin. 2023, 43, 2423001. [Google Scholar]
- Optics, M. Meadowlark Optics. Available online: https://www.meadowlark.com/ (accessed on 7 March 2019).
- Gu, N.; Chen, H.; Tang, A.; Fan, X.; Noda, C.Q.; Xiao, Y.; Zhong, L.; Wu, X.; Zhang, Z.; Yang, Y.; et al. Embedded solar adaptive optics telescope: Achieving compact integration for high-efficiency solar observations. Opto-Electron. Adv. 2025, 8, 250025. [Google Scholar] [CrossRef]
- Yang, J.; Tang, A.; Lian, B.; Xiong, Z.; Ma, Z.; Gu, N. Optimal multi-spectrum polarization detection with full-Stokes in a single channel by using standard retarders. Measurement 2025, 242, 116052. [Google Scholar]
- Hou, J.; Sun, Y.; Lin, J.; Zhang, Y.; Wang, D.; Deng, Y.; Zhang, Z. In Situ Calibration of the Tunable Liquid-crystal Birefringent Filter. Acta Photonica Sin. 2023, 52, 0552208. [Google Scholar]
- Frenkel, D.; Smit, B. Understanding Molecular Simulation: From Algorithms to Applications; Elsevier: Amsterdam, The Netherlands, 2023. [Google Scholar]
















| Parameter | Value |
|---|---|
| Central Wavelength | 5303 Å |
| FWHM (Full-Width at Half-Maximum) | 1 Å |
| Effective Clear Aperture | 30 mm |
| Free Spectral Range (FSR) | 2 Å |
| Wavelength Tuning Time | <60 ms |
| Thermostatic Chamber Operating Temperature | 40.6 °C |
| Temperature Stability Precision | 0.01 °C |
| Coeff_A1 | ||||||||
| 0.307 | 1.29 × 10−3 | −3.82 × 10−6 | 2.17 × 10−8 | −6.39 × 10−12 | −5.74 × 10−13 | 2.30 × 10−15 | −2.51 × 10−18 | |
| 0.324 | 1.48 × 10−3 | −4.66 × 10−6 | −7.12 × 10−9 | 1.41 × 10−10 | 7.02 × 10−13 | −6.75 × 10−15 | 1.17 × 10−17 | |
| Coeff_A2 | ||||||||
| 0.320 | 1.07 × 10−3 | −4.67 × 10−6 | 2.55 × 10−9 | 2.24 × 10−10 | −1.45 × 10−13 | −6.10 × 10−15 | 1.36 × 10−17 | |
| 0.341 | 1.10 × 10−3 | −4.27 × 10−6 | 1.09 × 10−8 | 2.16 × 10−10 | −3.44 × 10−13 | −5.87 × 10−15 | 1.62 × 10−17 | |
| Retardance Error | Std of Center Wavelength | Peak Transmittance | FWHM | Sidelobe Proportion |
|---|---|---|---|---|
| 5° | Å | Å | 0.15% | |
| 10° | Å | Å | 0.42% | |
| 20° | Å | Å | 1.93% |
| Mode | Retardation | |
|---|---|---|
| LCVR1 | LCVR2 | |
| Single | −18° | 40° |
| Double | −18° | 210° |
| −0.45 Å | 63° | 80.5° |
| +0.45 Å | 261° | −0.5° |
| −1 Å | 162° | 130° |
| +1 Å | 162° | −50° |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Song, T.; Liu, Y.; Zhang, X.; Zhao, M.; Li, Z. Design, Calibration, and On-Site Validation of an LCVR-Driven Fast-Tunable Lyot Filter for the YOGIS Coronagraph. Photonics 2026, 13, 76. https://doi.org/10.3390/photonics13010076
Song T, Liu Y, Zhang X, Zhao M, Li Z. Design, Calibration, and On-Site Validation of an LCVR-Driven Fast-Tunable Lyot Filter for the YOGIS Coronagraph. Photonics. 2026; 13(1):76. https://doi.org/10.3390/photonics13010076
Chicago/Turabian StyleSong, Tengfei, Yu Liu, Xuefei Zhang, Mingyu Zhao, and Zhen Li. 2026. "Design, Calibration, and On-Site Validation of an LCVR-Driven Fast-Tunable Lyot Filter for the YOGIS Coronagraph" Photonics 13, no. 1: 76. https://doi.org/10.3390/photonics13010076
APA StyleSong, T., Liu, Y., Zhang, X., Zhao, M., & Li, Z. (2026). Design, Calibration, and On-Site Validation of an LCVR-Driven Fast-Tunable Lyot Filter for the YOGIS Coronagraph. Photonics, 13(1), 76. https://doi.org/10.3390/photonics13010076

