Next Article in Journal
Recovery of Optical Transport Coefficients Using Diffusion Approximation in Bilayered Tissues: A Theoretical Analysis
Previous Article in Journal
Delayed Choice for Entangled Photons
Previous Article in Special Issue
The Efficacy of Erbium-Ion, Diode, and CO2 Lasers in Debonding Attachments Used During Overlay Orthodontic Treatment and the Risk of Hard Tooth Tissue Damage Compared to Traditional Methods—An In Vitro Study
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Influence of Uncertainties in Optode Positions on Self-Calibrating or Dual-Slope Diffuse Optical Measurements

1
Department of Biomedical Engineering, Tufts University, 4 Colby Street, Medford, MA 02155, USA
2
Department of Power Electronics & Intrumentation Engineering, Jorhat Institute of Science and Technology, Chenijaan, Jorhat 785010, Assam, India
*
Author to whom correspondence should be addressed.
Photonics 2025, 12(7), 697; https://doi.org/10.3390/photonics12070697
Submission received: 1 May 2025 / Revised: 3 July 2025 / Accepted: 8 July 2025 / Published: 10 July 2025
(This article belongs to the Special Issue Photonics: 10th Anniversary)

Abstract

Self-calibrating and dual-slope measurements have been used in the field of diffuse optics for robust assessment of absolute values or temporal changes in the optical properties of highly scattering media and biological tissue. These measurements employ optical probes with a minimum of two source positions and a minimum of two detector positions. This work focuses on a quantitative analysis of the impact of errors in these source and detector positions on the assessment of optical properties. We considered linear, trapezoidal, and rectangular optode arrangements and theoretical computations based on diffusion theory for semi-infinite homogeneous media. We found that uncertainties in optodes’ positions may have a greater impact on measurements of absolute scattering versus absorption coefficients. For example, a 4.1% and 19% average error in absolute absorption and scattering, respectively, can be expected by displacing every optode in a linear arrangement by 1 mm in any direction. The impact of optode position errors is typically smaller for measurements of absorption changes. In each geometrical arrangement (linear, trapezoid, rectangular), we identify the direction of the position uncertainty for each optode that has minimal impact on the optical measurements. These results can guide the optimal design of optical probes for self-calibrating and dual-slope measurements.

1. Introduction

In Frequency-Domain (FD) Near-InfraRed Spectroscopy (NIRS) strategies for the retrieval of the bulk absolute optical properties of optically turbid media are based on multi-frequency [1] or multi-distance [2] measurements. In both cases a calibration of the experimental apparatus on a phantom of known optical properties is required. This is not a trivial step, and the success of the calibration procedure for the quantification of the absolute optical properties of a target medium is based on several assumptions: (1) that a calibration phantom of known (i.e., high accuracy) optical properties and robust procedures for its quantification exists [3]; (2) that the phantom is optically homogeneous; (3) that one is able to reproduce the optical coupling between the probe and the calibration phantom also during the experiment on a medium of unknown optical properties; (4) that the laser power and the detector gain are stable. These are not conditions that are always easy to satisfy. For example, experiments that last several hours might need recalibration of the system due to laser power and detector gain fluctuations. Ideally, pressure sensors to control the optical coupling might be needed. During in vivo experiments on human subjects, the requirement of stable coupling might be particularly difficult to achieve, due to uncontrolled subject motion. Therefore, methods that bypass the limiting factors of calibration and that provide accurate estimates of the optical properties are highly welcome.
Hueber et al. [4] first proposed the Self-Calibrating (SC) method for the retrieval of the absolute optical properties of tissues in FD NIRS (namely the μ a and the μ s ). The proposed method provided a robust way to retrieve the optical properties with the advantage of being highly insensitive to the limiting factors intrinsic in a calibration procedure. Since then, the SC method has found a number of applications for the estimation of StO2 [5,6,7], for the measurement of Δ μ a of target tissues during protocols that elicit hemodynamic changes [8], and spectral measurements of the absolute optical properties of tissues [9,10]. When implemented to recover Δ μ a from a single type of optical data (e.g., only the ϕ in FD), the method is referred to as Dual-Slope (DS), which can been proposed in Time-Domain (TD) [11] and implemented in FD [12]. The basic “unit” for SC measurement features two sources and two detectors in a special geometrical arrangement. The four optodes identify four source–detector separations, and the prerequisite of a SC unit is that the two short and the two long source–detector separations must be equal. If this requirement is met (with some spatial constraints due to the dynamic range of detectors), one is free to choose between several spatial arrangements of the optodes. Typical ones are the linear (i.e. with the detectors sandwiched between the sources along a line), rectangular and trapezoidal (i.e., the optodes occupy the vertices of a rectangle and a trapezoid, respectively) [13]. Examples of these various arrangements can be found in the literature: linear [4,5,6,10,11,12,13,14,15,16,17]; rectangular [4,7,8,13,18,19,20]; and trapezoidal [13,20,21].
Since its inception the SC method has been also used with Continuous-Wave (CW) instrumentation. For example, Jenny et al. [18] developed a custom-made CW system with a probe having two concentric rectangular SC units and three wavelengths at each source location. The instrument was used for assessing the precision of StO2 measurements in vivo on a neonatal population. The same instrument was modified by Kleiser et al. [22] in a custom-made CW instrument (OxyPrem v. 1.3) featuring two collinear SC units, i.e., having two detectors and four source locations (four wavelengths per source location) sandwiched between the detectors in a linear array. The instrument was used to determine the accuracy of StO2 measurements on a blood-lipid phantom. Kleiser et al. [6] used the same instrument for StO2 precision assessment in a neonatal population. Chincarini et al. [21] used OxyPrem v. 1.4 featuring two SC units in a trapezoidal arrangement, and measured changes of oxy- and deoxyhemoglobin on sheep. Raoult et al. [23] used OxyPrem for measuring changes in oxy- and deoxyhemoglobin in dogs. Another commercial instrument, the Nonin EQUANOX, has used a probe with the SC arrangement to measure cerebral oxygenation in humans [14], but it is unclear if the SC method was implemented in the data analysis. Xu et al. [8] developed a custom-made CW system with eight linear SC arrangements (two wavelength per source location) combined in a hand-held probe for dynamic characterization of biological tissues. Scholkmann et al. [15] used OxyPrem to test the resilience of measurements towards induced motion artifacts. Perekatova et al. [17] used a custom-built CW ultra-broadband (460 nm to 1030 nm) featuring a SC linear unit and retrieved the μ e f f at different wavelengths. The experimental setup was tested for resistance toward different instrumental perturbations. Wu et al. [7] developed a CW system with a probe comprising a SC unit in a rectangular arrangement (with two wavelengths per source location). The system was used for studying pulsatile cerebral blood flow together with a Diffuse Correlation Spectroscopy (DCS) system [19].
We note that when CW instrumentation is used for measuring StO2 or Δ μ a , one must make assumptions about the values and wavelength dependence of the reduced scattering coefficient. This source of error and how it propagates to the target parameters has not been investigated. Also, there has not been a thorough investigation into the errors on the source–detector separations and how it propagates to the target parameters. This error was briefly addressed by Hueber et al. [4] for the linear arrangement by considering (a) errors in only one short distance; (b) errors in all the short distances; (c) errors in all short and long distances (by keeping the difference of the distances unchanged); (d) errors in all short and long distances by keeping the average distance unchanged. In this work, we consider a wider range of cases for the errors in the source–detector separation and their propagation for the quantification of absolute optical properties (i.e., μ e f f or μ a and μ s ), or absorption changes (i.e., Δ μ a ). We also consider four typical SC arrangements: linear, asymmetric-linear, trapezoidal, diagonal-rectangular.

2. Materials and Methods

2.1. Dual-Slope Arrangements

In this work, four types of DS arrangement geometry will be considered (Figure 1). The first arrangement, LINeaR (LINR; Figure 1a), is the simplest and most common type of DS set. The next arrangement, Asymmetric-LINear (ALIN; Figure 1b), is a slight modification on LINR with the detectors shifted to the left. ALIN has not been implemented in as many experiments as the other sets, but has been shown to have a similar region of spatial sensitivity as LINR [13]. The last two sets, TRAPezoidal (TRAP; Figure 1c) and Diagonal-ReCTangular (DRCT; Figure 1c), are the two types of sets found in the modular hexagonal imaging arrangement proposed by our group [20]. TRAP is of particular interest since most DS imaging arrays are composed of mainly trapezoidal sets. All of these arrangements have the same mean ρ of 31 mm.

2.2. Analytical Forward Model

We generated forward FD NIRS data in the form of the R ˜ for each source–detector pair in the arrangement being assessed (i.e., 1A, 1B, 2A, and 2B). These R ˜ data were computed using an analytical diffusion model for a semi-infinite homogeneous medium [1]. The primary inputs to the forward model are μ a , μ s , and ρ , of which the former two were varied and the latter one was determined by the arrangement. Optical property ranges were chosen to encompass bulk tissue values recovered by FD NIRS on the human forehead [2]. The remaining model parameters were the f m o d and the n, which were fixed at 100 × 10 6 Hz and 1.4, respectively. Importantly, when considering an optode position error, data were generated with ρ s corresponding to the optode in the error position, not the nominal position.
We also considered changes in R ˜ resulting from small changes in μ a to test DS’s ability to recover Δ μ a s. These were computed by finding R ˜ for the baseline optical properties and for the same parameters except for μ a , which was increased by 0.0001 mm 1 (i.e., true Δ μ a of 0.0001 mm 1 ).

2.3. Inverse Models

2.3.1. Absolute Optical Properties

For a given simulation case, the forward model generated four values of R ˜ , one for each source–detector pair in the arrangement of interest. These four R ˜ s are input into the inverse models for recovering the μ a and the μ s to implement the SC method. The remaining parameters which must be assumed for the inverse models were the ρ s, the f m o d , and the n. The ρ s were either assigned the correct values (given the optode position considered in the forward model) or incorrect values. The cases where the ρ s were assumed incorrectly correspond to cases where optodes were not in their nominal position to generate data with the forward model, but the nominal ρ s were assumed in the inverse model. This represents a possible real-life scenario where the optode arrangement was affected by position errors, but the data were analyzed assuming the nominal ρ s. The remaining two parameters, the f m o d and the n, were assumed to be known in all cases.
In this work, we have tested two inverse models based on the SC R ˜ data to recover the μ a and the μ s . We will refer to the first method as the “slopes” method since it is based on the slopes, computed with SC/DS methods, of the ln ρ 2 I and the ϕ with respect to ρ [2]. The second method we will refer to as the “iterative” method since it is based on iteratively solving the forward model for R ˜ as a function of ρ [16], where R ˜ = I e i ϕ . The “slopes” method solves the following equation:
ln ρ 2 R ˜ = μ ˜ e f f ρ
where the μ ˜ e f f is defined as:
μ ˜ e f f = 3 ( μ s + μ a ) ( μ a 2 π f m o d i / v )
where v is the speed of light in the medium. The “iterative” method, on the other-hand, iteratively solves this expression:
ln f ˜ ( ρ , μ a , μ s ) R ˜ = μ ˜ e f f ρ 2 + 1 / μ s 2
by updating the μ a and the μ s every iteration. Details on the iterative method and the complex function f ˜ ( ρ , μ a , μ s ) can be found in Reference [16]’s Equation (4). The slope method is based on stricter assumptions than the iterative method but is less computationally intense and more stable against artifacts in noisy data.

2.3.2. Changes in Absorption

The forward model considered changes in the μ a by generating R ˜ s for both the baseline optical properties and perturbed optical properties for each of the four source–detector pairs in the DS arrangement of interest. From these baseline and perturbed R ˜ s, the changes in the DS of ln I 0 / I (i.e., ln ( | R ˜ | ) ) or ϕ (i.e., R ˜ ) were calculated and converted to Δ μ a using previously reported methods [12]. It is worth noting that these methods require an assumption of μ a and μ s to calculate the L ˜ s. The values of μ a and μ s found using the iterative method on the forward data were used in this case, meaning errors in the recovered μ a and μ s may propagate through L ˜ s to an error on the recovered Δ μ a . The inverse model for absorption changes also requires an assumption of the ρ s. These ρ s were assumed in the same way as described for the absolute optical property inverse model, with the nominal ρ s always used in the inverse problem even when an error is imposed on the optode positions to generate the data with the forward model.

2.4. Variable Definitions

The results in this work focus on errors in recovered parameters as a result of incorrect assumed values of ρ due to geometrical optode position errors. Therefore, we must be careful to use a notation which differentiates true parameters from parameters recovered either with correct or incorrect assumptions. Accordingly, variables with ( ρ ) in superscript indicate values recovered by the inverse model assuming the correct ρ s, while variables with ( × ρ ) in superscript indicate values recovered assuming the incorrect ρ s. Variables without either ( ρ ) or ( × ρ ) in superscript indicate the true values used in the forward model to generate simulated data. For example, μ a indicates the true absorption coefficient used in the forward model, while μ a ( ρ ) or μ a ( × ρ ) indicate the absorption coefficient recovered by the inverse model, assuming the correct or incorrect ρ s, respectively.
We also aim to assess to what extent an error in an optode’s position affects the values recovered in general. To quantify this we define the root-mean-square-error in either the μ a or the μ s by varying the optode positions in a 2 mm diameter circle which orbits their nominal position ( σ μ a 2 mm or σ μ s 2 mm , respectively) as follows:
σ μ a 2 mm = i = 1 n 2 mm μ a , i ( × ρ ) μ a ( ρ ) 2 n 2 mm
σ μ s 2 mm = i = 1 n 2 mm μ s , i , ( × ρ ) μ s , ( ρ ) 2 n 2 mm
where μ a ( ρ ) and μ s , ( ρ ) are the absorption coefficient and the reduced scattering coefficient recovered by the inverse model with the optode in its nominal position and assuming the nominal ρ s. However, μ a ( × ρ ) and μ s , ( × ρ ) are the absorption coefficient and the reduced scattering coefficient recovered by the inverse model with the optode not in its nominal position but assuming the nominal ρ s. To realize this, the optode’s (e.g., source 1’s) position is varied through n 2 mm points in a 2 mm diameter circle surrounding its nominal position (e.g., a circle orbiting source 1’s nominal position with a 1 mm radius) and computing the forward then inverse models for each point. To assess the influence of a position error in multiple optodes (e.g., source 1 and detector A), each optode considers n 2 mm m positions (e.g., n 2 mm ) in a circle around its nominal position, where m is the number of optodes considered (e.g., 2). These positions are co-varied for each optode resulting in n 2 mm values of μ a ( × ρ ) and μ s , ( × ρ ) , which may be used to compute σ μ a 2 mm or σ μ s 2 mm with Equations (4) and (5). It should be noted that this definition quantifies deviation from the recovered nominal value instead of deviation from the true value to remove any effect of systematic error in the inversion.

3. Results

3.1. Self-Calibrated Recovery of Absolute Optical Properties

Before diving into the errors introduced by recovery method or optode position, we present the expected errors in μ a and μ s from other common experimental considerations to form a point of comparison. First, the recovery methods typically assume point-like optical sources and detectors. However, in reality optode areas are finite. Therefore, to understand the error introduced by this assumption, we have simulated a situation with source radii of 300 μm and detector radii of 1.5 mm. In the case of a μ a of 0.01 mm 1 , a μ s of 1 mm 1 , and ρ s of [ 25 , 25 , 37 , 37 ] mm, assuming these finite optode sizes creates an additional error in μ a and μ s of 0.1% and 0.2%, respectively, when using the iterative recovery method. Since this consideration of optode area creates little additional error, we modeled optodes as point-like for the remainder of this work. However, we do note that if the optode sizes are known, then this consideration can be included in the inverse model, which may be relevant at shorter source–detector separations of the order of 10 mm or less. Second, in reality, random instrumental noise propagates to a random error in the recovered μ a and μ s . Considering typical instrumental noise values of 0.1% for the I and 0.1° for the ϕ results in random error in μ a and μ s of 0.6% and 0.5%, respectively, using the same parameters as the finite optode case above. Since instrumental noise is also expected to propagate to small errors in optical properties, this additional consideration is also neglected for the remainder of the manuscript.

3.1.1. Error from the Choice of Optical Property Recovery Method

We have compared the accuracy of two methods to recover the μ a and the μ s from SC FD NIRS data. These two methods are the slopes method [2] and the iterative method [16], where the slope method is simpler but contains stronger assumptions in its derivation. Table 1 shows the accuracy of μ a ( ρ ) and μ s , ( ρ ) recovered using each method for a range of true values of μ a and μ s . We conducted this exercise for two sets of ρ s. The first set, ρ = [ 25 , 25 , 37 , 37 ] mm, corresponds to the LINR, TRAP, and DRCT sets (Figure 1a,c,d). The second set, ρ = [ 20 , 30 , 32 , 42 ] mm, corresponds to the ALIN set (Figure 1b).
Overall, the results in Table 1 show that the slopes method is less accurate than the iterative method. To summarize, averaging the error in μ a ( ρ ) for all optical properties and ρ sets yields 6% and 0.01% for the slopes and iterative methods, respectively. Similarly, averaging the error in μ s , ( ρ ) for all optical properties and ρ sets yields 6 % and 1 % for the slopes and iterative methods, respectively. Both methods have a positive bias for μ a ( ρ ) and a negative bias for μ s , ( ρ ) . In either case, the accuracy of the method depends strongly on the optical properties and only depends weakly on the set of ρ s, so we will focus on interpreting the results for the ρ = [ 25 , 25 , 37 , 37 ] mm set of ρ s here. The accuracy of the μ a ( ρ ) recovered by the slopes method depends heavily on the true μ a , with worse accuracies occurring at low μ a s. Meanwhile, the accuracy of the μ a ( ρ ) recovered by the iterative method more strongly depends on the true μ s , and only weakly depends on the true μ a . For both the slopes and iterative methods, the accuracy of μ s , ( ρ ) depends on the true value of μ s , with lower true μ s s corresponding to worse accuracy. The purpose of these results is to indicate how accurate these methods are even in the case where the ρ s are known. From these results, we decided to focus mainly on the iterative method and only show select results for the slopes method due to the iterative method’s substantially better accuracy.

3.1.2. Error from Optode Position Displacement

Single-Optode Displacement
  • Linear (LINR) Arrangement
We begin investigating optode position errors by exploring errors in a single optode’s position in the LINR arrangement. Due to the symmetry in this arrangement, displacing either source 1 or source 2 will result in the same error. Similarity, displacement in either detector A or detector B will also result in the same error. Therefore, we only show results for displacements of source 1 or detector A. Figure 2 shows the error in the recovery optical properties ( μ a ( × ρ ) and μ s , ( × ρ ) ) relative to the true optical properties ( μ a and μ s ) due to displacing the optode from its nominal position in a 20 mm × 20 mm square. The recovered optical properties always assume the nominal optode position even if the optode is not in that position, thus the use of the ( × ρ ) superscript.
To interpret the results in Figure 2, we will focus on the direction of maximum change in the error and the range of the error from displacement in the 20 mm × 20 mm square. Concentrating on the subplots for source 1 (Figure 2a–e), we find four color maps for each combination of the two recovery methods “slopes” or “iterative” (Figure 2b,c or d,e, respectively) and two optical properties μ a or μ s (Figure 2b,d or c,e, respectively). All four of these subplots show that larger errors occur if the optode is displaced away or toward the detectors (along the linear line of the LINR arrangement). In each case, the range of the error is about 10%. Now focusing on errors from displacement of detector A (Figure 2f–j), we see again that movement along the linear line of the arrangement creates large changes in the error. However, the magnitude of the error from displacing detector A is much larger than source 1. Displacement of detector A in the 20 mm × 20 mm square results in a error range of about 30% for μ a and >100% for μ s (for both the slopes and iterative recovery methods). Overall, Figure 2 shows that displacement of optodes along the line of the LINR arrangement creates the largest error and the position of the detectors is much more critical than the position of the sources, but the two recovery methods do not have a significantly different dependence on optode position error.
Table 2 represents the magnitude of the error in the recovered optical properties from a 1 mm displacement in any direction. This is done using the metrics σ μ a 2 mm / μ a ( ρ ) or σ μ s 2 mm / μ s , ( ρ ) (Equation (4) or Equation (5)), which represent the average error resulting from displacing an optode 1 mm in any direction from its nominal position relative to the value the method would recover with the optode in the nominal position ( μ a ( ρ ) or μ s , ( ρ ) ). These values are important because they represent expected errors that could result from probe construction, since it is reasonable that optode positions in a real probe may be incorrect on the order of 1 mm. Therefore, these errors can be thought of as expected systematic errors from the practical limitations of building a NIRS probe, for a given arrangement, optical property, and recovery method. Specifically for the LINR arrangement, we see that an error in the detector position propagates to a worse error in optical properties in all cases. Additionally, the effect of a 1 mm displacement does not seem significantly different between the slopes and iterative methods. Aside from that, the remaining conclusions are somewhat expected from the limitations of diffusion theory. That is, overall larger errors in μ s than μ a (specifically for the case of detector displacement) and worse errors at low true values of μ a and μ s . For overall guidance, from a 1 mm optode position error in any direction, we can expect about a 0.7% or 3% error, for source or detector displacement, respectively, in μ a and about a 0.3% or 10% error, for source or detector displacement, respectively, in μ s (using the iterative recovery method) for the LINR arrangement.
  • Asymmetric-Linear (ALIN) Arrangement
Second, we explore errors in a single optode’s position in the ALIN arrangement. Due to the lack of symmetry in this arrangement, displacement of each optode must be looked at individually.
Interpreting the results in Figure 3, we again focus on the direction of maximum change in the error and the range of the error in the simulated square. Similar to the LINR arrangement in Figure 2, all of the subplots show that larger errors occur if the optode is displaced away or toward the detectors (along the linear line of the ALIN arrangement). Looking at the range of the error, we see that the position of the detectors is most critical with an error range of about 60% for μ a and >100% for μ s . However, the position of source 1 is slightly more critical than the position of source 2. Source 1 shows an error range of about 20% for μ a and μ s , while source 2 shows a range of about 10%. Overall, Figure 3 shows that displacement of optodes along the line of the ALIN arrangement creates the largest error and the position of the detectors is much more critical than the position of the sources, same as the LINR arrangement (Figure 2).
Table 3 presents the magnitude of the error in the recovered optical properties from a 1 mm displacement in any direction for the ALIN arrangement (Equation (4) or Equation (5)). Similar to the LINR arrangement, we see for the ALIN arrangement that errors in the detector position propagate to a worse error in optical properties in all cases. To give overall guidance, from a 1 mm optode position error in any direction, we can expect about a 0.7% or 3% error, for source or detector displacement, respectively, in μ a and about a 0.4% or 10% error, for source or detector displacement, respectively, in μ s (using the iterative recovery method) for the ALIN arrangement.
  • Trapezoidal (TRAP) Arrangement
Next, we explore errors in a single optode’s position in the TRAP arrangement. The symmetry of the TRAP arrangement lets us focus on only source 1 (since it is equivalent to source 2) and detector A (since it is equivalent to detector B).
Focusing on the results in Figure 4, we again emphasize the direction of maximum change in the error and the range of the error in the simulated square. In this arrangement the directions which cause a maximum change in the error are not as trivial as they were in the LINR and ALIN arrangements. Looking at the range of the error, we again see that the position of the detectors is most critical with an error range for detector A of about 60% for μ a and >100% for μ s . Instead, source 1 shows an error range of about 30% for μ a and about 70% for μ s . Overall, Figure 4 shows that the direction of the optode displacement which causes the maximum error is not as trivial as it is with the LINR or ALIN arrangements. We will discuss a way to use knowledge of these directions in Section Design of Optode Support Structure.
Table 4 shows the magnitude of the error in the recovered optical properties from a 1 mm displacement in any direction for the TRAP arrangement (Equation (4) or Equation (5)). Similar to the LINR and ALIN arrangement, we see for the TRAP arrangement that errors in the detector position propagate to a worse error in optical properties in all cases; however, the difference in the dependence on source or detector is not as strong with the TRAP arrangement. To give overall guidance, from a 1 mm optode position error in any direction, we can expect about a 0.9% or 2% error, for source or detector displacement, respectively, in μ a and about a 2% or 9% error, for source or detector displacement, respectively, in μ s (using the iterative recovery method) for the TRAP arrangement.
  • Diagonal-Rectangular (DRCT) Arrangement
Finally, we explore errors in a single optode’s position in the DRCT arrangement. The symmetry of the DRCT arrangement lets us focus on only source 1 (since it is equivalent to source 2, detector A, and detector B).
Figure 5 shows the results for the DRCT arrangement. Similar to the TRAP arrangement, the directions which cause a maximum change in the error are not as trivial as they were in the LINR and ALIN arrangements. However, roughly speaking, it seems that displacement of an optode towards or away from the center causes little error (i.e., movement orthogonal to that creates a large change in the error). Since, for this arrangement, displacement of any optode is equivalent, there is no error dependence preference between sources and detectors (i.e., their positions are equally important). Looking at the range of the error, we see an error range of about >100% for μ a and >100% for μ s from moving with the 20 mm × 20 mm square. However, the large (i.e., >100%) errors occur for μ a only in the lower right corner of the square; ignoring that region, the μ a error range is about 50%.
Table 5 shows the magnitude of the error in the recovered optical properties from a 1 mm displacement in any direction for the DRCT arrangement (Equation (4) or Equation (5)). In this case there is no dependence on optode, since displacement of any optode is equivalent. To give overall guidance, from a 1 mm optode position error in any direction, we can expect about a 2% in μ a and about a 9% in μ s (using the iterative recovery method) for the DRCT arrangement.
Multi-Optode Displacement
Table 6 shows the average errors which may be expected if multiple optodes are displaced by 1 mm in any direction assuming 1.0 mm 1 for μ s and 0.01 mm 1 for μ a . Certain combinations were omitted due to symmetry as follows (the “≡” symbol is used to indicate that an optode combination is equivalent): for LINR and TRAP 1A ≡ 2B, 1B ≡ 2A, 1AB ≡ 2AB, and 12A ≡ 12B; for ALIN no combination is omitted; and for DRCT 1A ≡ 2B, 1B ≡ 2A, 12 ≡ AB, and 1AB ≡ 2AB ≡ 12A ≡ 12B. The errors in this table are, for the most part, no more than twice the worst-case error from a single optode displacement. Furthermore, moving from two optodes to three to four does not significantly change the magnitude of the error, with more dependence on if a critical optode is displaced (e.g., the detectors in the LINR arrangement). For example, in the LINR case, the lowest error occurs by displacing 12, which is the only case that does not include a detector displacement. Furthermore, in this case 12A shows an error more compatible to the 1A and 1B cases then the 1AB case; therefore, the main influence in the magnitude of the error appears to be how many detectors are displaced (i.e., more generally, how many critical optodes are displaced). Overall, these results point to the TRAP arrangement being slightly more robust against optode displacement than the other arrangements, possibly because it de-emphasizes the criticality of the detector positions. This can be seen in the case where all four optodes are displaced (12AB), which results in TRAP average errors of about 3% and 14%, for μ a and μ s , respectively, while this average error is 4% and 19%, for μ a and μ s , respectively, for the LINR, ALIN, and DRCT arrangements.

3.2. Dual-Slope Recovery of Relative Changes in the Absorption Coefficient ( μ a )

In the following section the figures are in the same format as Section 3.1 but show results relevant to DS measurements of Δ μ a instead of SC measurements of the absolute μ a and μ s . These results show errors in the recovered Δ μ a (shown with a ( × ρ ) in superscript to indicate that the nominal position is always assumed) relative to the true Δ μ a simulated in the medium. Using DS, Δ μ a can be recovered with either only I data or only ϕ data; therefore, results are shown for both cases. Figure 6, Figure 7, Figure 8 and Figure 9 show these results for the LINR, the ALIN, the TRAP, and the DRCT arrangements, respectively. As before, we suggest that one focus on the direction of the change in the error and the magnitude of the range in the error from displacing an optode in a 20 mm × 20 mm square around its nominal position.
Summarizing these results, we note that, overall, the relative errors in Δ μ a due to optode position errors are less then those for absolute optical properties. For DS I the error is often on the order of 1% from moving the optode in the 20 mm × 20 mm square. However, for DS ϕ the order is 10%. Therefore, the ability for DS ϕ to accurately recover values of Δ μ a depends much more on the optode position than DS I by an order of magnitude.

4. Discussion

Based on the results in this work, we can provide guidance on what type of SC/DS arrangement is more robust against optode position errors. Additionally, we can provide guidance on what absolute optical property recovery method to use. Regarding the latter, we advocate for the iterative absolute optical property recovery method [16] on FD SC data. We justify this by comparing the accuracy errors of the iterative and slopes method in a case where the optode positions are assumed correctly (Table 1). Focusing on the error in the recovered μ a , the iterative method showed an average error of 6%, while the iterative method showed 0.01%, with a similar relationship for the error in μ s . Additionally, if we consider how robust these two methods are against optode position errors (Table 2), we see that the iterative method either performs the same as the slopes method or has a slight advantage. Therefore, we conclude that the iterative method is preferable over the slopes method.
For a comparison of the SC/DS arrangements, we can examine the results in both Section 3.1 and Section 3.2. Considering all of these together, we conclude that arrangements typically have a critical optode type, and movement of the critical optode causes large errors in the recovered values. The critical optode type is characterized by the type of optodes placed near the arrangement’s line of symmetry or centroid (e.g., detectors A and B in LINR). This is because the position of an optode with this characteristic more heavily influences the difference between the short and long ρ s in the SC/DS set, and this difference is known to be a critical parameter for these measurements [4]. The critical type may be either sources or detectors since sources and detectors may be interchanged (e.g., if sources and detectors are switched in LINR then sources 1 and 2 are critical). For LINR and ALIN the critical optodes are optodes A and B, while for TRAP the importance of optodes 1 and 2 and optodes A and B is more equally balanced but optodes A and B are still slightly more critical than optodes 1 and 2. The DRCT arrangement is not ideal since all the optodes’ positions are critical, since displacement of any optode is geometrically equivalent to the displacement of another. Additionally, the results in Table 6 further suggest that the TRAP arrangement may be more robust against multiple optode position errors than the other arrangements. A second conclusion is that the error in μ s or Δ μ a recovered by DS ϕ is more dependent on the position of the critical optodes than μ a or Δ μ a recovered by DS I, respectively.
Results in Section 3.1 unsurprisingly show that μ s is more susceptible to errors, but from the recovery method itself and from optode position errors. This agrees with our experience working with experimental SC data [16]. Additionally, we have confirmed that the spatial dependence of the error on optode displacement behaves the same for experimental data as it does for both these simulations and the error’s order of magnitude agree. Specifically, if we consider an in vivo dataset [20] and assume incorrect optode displacements, we recover errors of the same order of magnitude and direction as the simulation results herein (results not included here for brevity). A more unexpected result is that from Section 3.2, which shows that Δ μ a recovered by DS ϕ is more susceptible to errors than DS I by an order of magnitude. This may have implications when comparing Δ μ a s recovered by either DS I or DS ϕ since a small optode position error could propagate to almost no error in DS I’s results but a large error in DS ϕ ’s results, making comparison of the data-types tenuous.

Design of Optode Support Structure

The figures in Section 3.1 and Section 3.2 reveal information about which optode displacement direction causes the largest error and which direction causes no change in the error. If an optode moves along its iso-line (i.e., line of constant error from the nominal optode position) in these figures, the recovered value will not change. Therefore, we wish to design a probe which prevents optode movement in directions which introduce errors but allows movement in directions that will not change the recovered value. Figure 10 shows these lines of zero change for the LINR and TRAP arrangements. These lines show positions where the optode can be placed that will not change the recovered value of a particular optical property. The lines for different optical properties intersect at the nominal optode position since this is the only position which will cause no error change for any of the optical properties. We can use Figure 10 to determine which directions it is allowable for optodes to move in (i.e., along the lines of zero change) in a real-life probe design.
Figure 11 shows how these lines of zero change can be used to design a DS/SC probe. Since movement along the line of zero change is allowable but movement perpendicular to that causes an error, one can design rigid supports which are perpendicular to the tangent of the line of zero change. We then extend these supports until they intersect to show the proposed skeleton of rigid supports. This proposed support structure should reduce errors induced by incorrect assumptions of the optode position, since they do not allow the optodes to move in the direction of maximum error. For the LINR arrangement (Figure 11a) these supports are trivial since they run along the linear line of the probe. However, for the TRAP arrangement (Figure 11b), this method of designing supports results in a non-obvious structure. This approach to design SC/DS probes will result in measurements more robust against optode position errors and probe deformation.

5. Conclusions

Self-calibrating and dual-slope measurements present a number of significant practical advantages by not requiring calibration of source emission and detector sensitivity [4], and by being relatively insensitive to movement artifacts [15]. These measurements rely on the fulfillment of specific geometrical requirements for the location of optodes [4,13] and knowledge of the location of all optodes. This work investigated the impact of uncertainties in the location of the optodes, which is especially relevant in the case of in vivo applications to biological tissues where the optical probe is required to conform to the shape of the investigated tissue. The significance of this work is twofold. First, it allows one to estimate the impact of uncertainties in the optode positions on absolute or relative optical measurements from continuous-wave or frequency-domain data in dual-slope configurations. Second, it guides the design of optical probes to minimize probe deformation along directions that result in optode displacements with the greatest impact on the optical measurements. Future work may include the investigation of other less impactful geometric sources of error for these diffuse optical measurements such as sample curvature (for relatively broad curvatures such as those relevant to the adult human head) or finite optode size.

Author Contributions

Conceptualization, G.B., A.S. and S.F.; methodology, G.B. and A.S.; software, G.B. and A.S.; validation, G.B. and A.S.; formal analysis, G.B. and A.S.; investigation, G.B., A.S., and T.D.; resources, S.F.; data curation, G.B., A.S., and T.D.; writing—original draft preparation, G.B., A.S., T.D., and S.F.; writing—review and editing, G.B., A.S. and S.F.; visualization, G.B.; supervision, A.S. and S.F.; project administration, S.F.; funding acquisition, S.F. All authors have read and agreed to the published version of the manuscript.

Funding

This work is supported by National Institutes of Health (NIH) awards R01-NS095334 and R01-EB029414. G.B. would also like to acknowledge support from NIH award K12-GM133314. The content is solely the authors’ responsibility and does not necessarily represent the official views of the awarding institutions.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Supporting code for this manuscript can be found at the following link: https://github.com/DOIT-Lab/DOIT-Public/tree/master/SelfCalibratingPositionErrors accessed on 7 July 2025.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Haskell, R.C.; Svaasand, L.O.; Tsay, T.T.; Feng, T.C.; McAdams, M.S.; Tromberg, B.J. Boundary Conditions for the Diffusion Equation in Radiative Transfer. J. Opt. Soc. Am. A 1994, 11, 2727–2741. [Google Scholar] [CrossRef] [PubMed]
  2. Blaney, G.; Sassaroli, A.; Pham., T.; Krishnamurthy, N.; Fantini, S. Multi-Distance Frequency-Domain Optical Measurements of Coherent Cerebral Hemodynamics. Photonics 2019, 6, 83. [Google Scholar] [CrossRef]
  3. Pifferi, A.; Torricelli, A.; Bassi, A.; Taroni, P.; Cubeddu, R.; Wabnitz, H.; Grosenick, D.; Möller, M.; Macdonald, R.; Swartling, J.; et al. Performance Assessment of Photon Migration Instruments: The MEDPHOT Protocol. Appl. Opt. 2005, 44, 2104–2114. [Google Scholar] [CrossRef] [PubMed]
  4. Hueber, D.M.; Fantini, S.; Cerussi, A.E.; Barbieri, B.B. New Optical Probe Designs for Absolute (Self-Calibrating) NIR Tissue Hemoglobin Measurements. In Proceedings of the Optical Tomography and Spectroscopy of Tissue III, San Jose, CA, USA, 24 January 1999; SPIE: Bellingham, WA, USA, 1999; Volume 3597, pp. 618–631. [Google Scholar] [CrossRef]
  5. MacLeod, D.B.; Ikeda, K.; Vacchiano, C.; Lobbestael, A.; Wahr, J.A.; Shaw, A.D. Development and Validation of a Cerebral Oximeter Capable of Absolute Accuracy. J. Cardiothorac. Vasc. Anesth. 2012, 26, 1007–1014. [Google Scholar] [CrossRef]
  6. Kleiser, S.; Ostojic, D.; Nasseri, N.; Isler, H.; Bucher, H.U.; Bassler, D.; Wolf, M.; Scholkmann, F.; Tanja Karen, M.D. In Vivo Precision Assessment of a Near-Infrared Spectroscopy-Based Tissue Oximeter (OxyPrem v1.3) in Neonates Considering Systemic Hemodynamic Fluctuations. J. Biomed. Opt. 2018, 23, 067003. [Google Scholar] [CrossRef]
  7. Wu, K.C.; Tamborini, D.; Renna, M.; Peruch, A.; Huang, Y.; Martin, A.; Kaya, K.; Starkweather, Z.; Zavriyev, A.I.; Carp, S.A.; et al. Open-Source FlexNIRS: A Low-Cost, Wireless and Wearable Cerebral Health Tracker. NeuroImage 2022, 256, 119216. [Google Scholar] [CrossRef]
  8. Xu, R.X.; Qiang, B.; Mao, J.J.; Povoski, S.P. Development of a Handheld Near-Infrared Imager for Dynamic Characterization of in Vivo Biological Tissue Systems. Appl. Opt. 2007, 46, 7442. [Google Scholar] [CrossRef] [PubMed]
  9. Lanka, P.; Lanka, P.; Segala, A.; Segala, A.; Farina, A.; Farina, A.; Sekar, S.K.V.; Nisoli, E.; Valerio, A.; Taroni, P.; et al. Non-Invasive Investigation of Adipose Tissue by Time Domain Diffuse Optical Spectroscopy. Biomed. Opt. Express 2020, 11, 2779–2793. [Google Scholar] [CrossRef]
  10. Pham, T.; Wei, L.L.; Roblyer, D. Cross-Wavelength Calibrating Method for Real-Time Imaging of Tissue Optical Properties Using Frequency-Domain Diffuse Optical Spectroscopy. Biomed. Opt. Express 2024, 15, 4963–4979. [Google Scholar] [CrossRef]
  11. Sawosz, P.; Liebert, A. Method to Improve the Depth Sensitivity of Diffuse Reflectance Measurements to Absorption Changes in Optically Turbid Medium. Biomed. Opt. Express 2019, 10, 5031–5041. [Google Scholar] [CrossRef]
  12. Blaney, G.; Sassaroli, A.; Pham, T.; Fernandez, C.; Fantini, S. Phase Dual-Slopes in Frequency-Domain near-Infrared Spectroscopy for Enhanced Sensitivity to Brain Tissue: First Applications to Human Subjects. J. Biophotonics 2020, 13, e201960018. [Google Scholar] [CrossRef]
  13. Fantini, S.; Blaney, G.; Sassaroli, A. Transformational Change in the Field of Diffuse Optics: From Going Bananas to Going Nuts. J. Innov. Opt. Health Sci. 2019, 13, 1930013. [Google Scholar] [CrossRef] [PubMed]
  14. Davie, S.N.; Grocott, H.P. Impact of Extracranial Contamination on Regional Cerebral Oxygen Saturation: A Comparison of Three Cerebral Oximetry Technologies. Anesthesiology 2012, 116, 834. [Google Scholar] [CrossRef]
  15. Scholkmann, F.; Metz, A.J.; Wolf, M. Measuring Tissue Hemodynamics and Oxygenation by Continuous-Wave Functional near-Infrared Spectroscopy—How Robust Are the Different Calculation Methods against Movement Artifacts? Physiol. Meas. 2014, 35, 717–734. [Google Scholar] [CrossRef] [PubMed]
  16. Blaney, G.; Donaldson, R.; Mushtak, S.; Nguyen, H.; Vignale, L.; Fernandez, C.; Pham, T.; Sassaroli, A.; Fantini, S. Dual-Slope Diffuse Reflectance Instrument for Calibration-Free Broadband Spectroscopy. Appl. Sci. 2021, 11, 1757. [Google Scholar] [CrossRef]
  17. Perekatova, V.; Kostyuk, A.; Kirillin, M.; Sergeeva, E.; Kurakina, D.; Shemagina, O.; Orlova, A.; Khilov, A.; Turchin, I. VIS-NIR Diffuse Reflectance Spectroscopy System with Self-Calibrating Fiber-Optic Probe: Study of Perturbation Resistance. Diagnostics 2023, 13, 457. [Google Scholar] [CrossRef]
  18. Jenny, C.; Biallas, M.; Trajkovic, I.; Fauchère, J.C.; Bucher, H.U.; Wolf, M. Reproducibility of Cerebral Tissue Oxygen Saturation Measurements by Near-Infrared Spectroscopy in Newborn Infants. J. Biomed. Opt. 2011, 16, 097004. [Google Scholar] [CrossRef] [PubMed]
  19. Wu, K.C.; Martin, A.; Renna, M.; Robinson, M.; Ozana, N.; Carp, S.A.; Franceschini, M.A. Enhancing Diffuse Correlation Spectroscopy Pulsatile Cerebral Blood Flow Signal with Near-Infrared Spectroscopy Photoplethysmography. Neurophotonics 2023, 10, 035008. [Google Scholar] [CrossRef]
  20. Frias, J.; Blaney, G.; Tavakoli, F.; Sassaroli, A.; Fantini, S. Dual-Slope near-Infrared Spectroscopy for Selective Sensitivity to Cerebral Hemodynamics. In Proceedings of the Optical Tomography and Spectroscopy of Tissue XVI, San Francisco, CA, USA, 27 January 2025; SPIE: Bellingham, WA, USA, 2025; Volume 13314, pp. 157–162. [Google Scholar] [CrossRef]
  21. Chincarini, M.; Qiu, L.; Spinelli, L.; Torricelli, A.; Minero, M.; Dalla Costa, E.; Mariscoli, M.; Ferri, N.; Giammarco, M.; Vignola, G. Evaluation of Sheep Anticipatory Response to a Food Reward by Means of Functional Near-Infrared Spectroscopy. Animals 2019, 9, 11. [Google Scholar] [CrossRef]
  22. Kleiser, S.; Nasseri, N.; Andresen, B.; Greisen, G.; Wolf, M. Comparison of Tissue Oximeters on a Liquid Phantom with Adjustable Optical Properties. Biomed. Opt. Express 2016, 7, 2973–2992. [Google Scholar] [CrossRef]
  23. Raoult, C.; Gygax, L. Valence and Intensity of Video Stimuli of Dogs and Conspecifics in Sheep: Approach-Avoidance, Operant Response, and Attention. Animals 2018, 8, 121. [Google Scholar] [CrossRef] [PubMed]
Figure 1. The four types of DS arrangement considered in this work. Sources are shown as red squares labeled with numbers and detectors are blue circles labeled with letters. Distances are measured from optode center to optode center. Drawn as a top-view such that the surface of the sample is represented as the plane of the page and the optodes are in contact with the sample. (a) The LINeaR (LINR) arrangement with ρ s of [ 25 , 25 , 37 , 37 ] mm. (b) The Asymmetric-LINear (ALIN) arrangement with ρ s of [ 20 , 30 , 32 , 42 ] mm. (c) The TRAPezoidal (TRAP) arrangement with ρ s of [ 25 , 25 , 37 , 37 ] mm. (d) The Diagonal-ReCTangular (DRCT) arrangement with ρ s of [ 25 , 25 , 37 , 37 ] mm.
Figure 1. The four types of DS arrangement considered in this work. Sources are shown as red squares labeled with numbers and detectors are blue circles labeled with letters. Distances are measured from optode center to optode center. Drawn as a top-view such that the surface of the sample is represented as the plane of the page and the optodes are in contact with the sample. (a) The LINeaR (LINR) arrangement with ρ s of [ 25 , 25 , 37 , 37 ] mm. (b) The Asymmetric-LINear (ALIN) arrangement with ρ s of [ 20 , 30 , 32 , 42 ] mm. (c) The TRAPezoidal (TRAP) arrangement with ρ s of [ 25 , 25 , 37 , 37 ] mm. (d) The Diagonal-ReCTangular (DRCT) arrangement with ρ s of [ 25 , 25 , 37 , 37 ] mm.
Photonics 12 00697 g001
Figure 2. Error in the accuracy of the “slopes” or “iterative” optical recovery method for the LINeaR (LINR) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ae) Errors from displacement of source 1 (equivalent to source 2). (fj) Errors from displacement of detector A (equivalent to detector B). (b,c,g,h) Results using the “slopes” [2] recovery method. (d,e,i,j) Results using the “iterative” [16] recovery method. (b,d,g,i) Error in the μ a . (c,e,h,j) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 2. Error in the accuracy of the “slopes” or “iterative” optical recovery method for the LINeaR (LINR) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ae) Errors from displacement of source 1 (equivalent to source 2). (fj) Errors from displacement of detector A (equivalent to detector B). (b,c,g,h) Results using the “slopes” [2] recovery method. (d,e,i,j) Results using the “iterative” [16] recovery method. (b,d,g,i) Error in the μ a . (c,e,h,j) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g002aPhotonics 12 00697 g002b
Figure 3. Error in the accuracy of the iterative optical recovery method for the Asymmetric-LINear (ALIN) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1. (df) Errors from displacement of detector A. (gi) Errors from displacement of detector B. (jl) Errors from displacement of source 2. (b,e,h,k) Error in the μ a . (c,f,i,l) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 3. Error in the accuracy of the iterative optical recovery method for the Asymmetric-LINear (ALIN) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1. (df) Errors from displacement of detector A. (gi) Errors from displacement of detector B. (jl) Errors from displacement of source 2. (b,e,h,k) Error in the μ a . (c,f,i,l) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g003
Figure 4. Error in the accuracy of the iterative optical recovery method for the TRAPezoidal (TRAP) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2). (df) Errors from displacement of detector A (equivalent to detector B). (b,e) Error in the μ a . (c,f) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 4. Error in the accuracy of the iterative optical recovery method for the TRAPezoidal (TRAP) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2). (df) Errors from displacement of detector A (equivalent to detector B). (b,e) Error in the μ a . (c,f) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g004
Figure 5. Error in the accuracy of the iterative optical recovery method for the Diagonal-ReCTangular (DRCT) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2, detector A, and detector B). (b) Error in the μ a . (c) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 5. Error in the accuracy of the iterative optical recovery method for the Diagonal-ReCTangular (DRCT) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2, detector A, and detector B). (b) Error in the μ a . (c) Error in the μ s . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g005
Figure 6. Error in the accuracy of recovering Δ μ a for the LINeaR (LINR) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2). (df) Errors from displacement of detector A (equivalent to detector B). (b,e) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c,f) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 6. Error in the accuracy of recovering Δ μ a for the LINeaR (LINR) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2). (df) Errors from displacement of detector A (equivalent to detector B). (b,e) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c,f) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g006
Figure 7. Error in the accuracy of recovering Δ μ a for the Asymmetric-LINear (ALIN) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1. (df) Errors from displacement of detector A. (gi) Errors from displacement of detector B. (jl) Errors from displacement of source 2. (b,e,h,k) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c,f,i,l) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 7. Error in the accuracy of recovering Δ μ a for the Asymmetric-LINear (ALIN) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1. (df) Errors from displacement of detector A. (gi) Errors from displacement of detector B. (jl) Errors from displacement of source 2. (b,e,h,k) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c,f,i,l) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g007
Figure 8. Errorin the accuracy of recovering Δ μ a for the TRAPezoidal (TRAP) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2). (df) Errors from displacement of detector A (equivalent to detector B). (b,e) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c,f) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 8. Errorin the accuracy of recovering Δ μ a for the TRAPezoidal (TRAP) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2). (df) Errors from displacement of detector A (equivalent to detector B). (b,e) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c,f) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g008
Figure 9. Error in the accuracy of recovering Δ μ a for the Diagonal-ReCTangular (DRCT) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2, detector A, and detector B). (b) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 9. Error in the accuracy of recovering Δ μ a for the Diagonal-ReCTangular (DRCT) arrangement. In the color maps the black dashed iso-lines represent the color bar tick-mark values and the nominal optode position is shown as a black asterisk. (ac) Errors from displacement of source 1 (equivalent to source 2, detector A, and detector B). (b) Error in the Δ μ a recovered by Dual-Slope (DS) I. (c) Error in the Δ μ a recovered by DS ϕ . Symbols: Displacement of optode from the nominal position ( [ Δ x , Δ y ] ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g009
Figure 10. Lines for which, if the optode is moved along them, a recovered optical property will not change (i.e., lines of zero change), e.g., the solid red line which passes through source 1 shows positions where source 1 can be placed and the μ a recovered by assuming the nominal ρ s will be the same. (a) The LINeaR (LINR) arrangement. (b) The TRAPezoidal (TRAP) arrangement. Symbols: Reduced scattering coefficient ( μ s ), absorption coefficient change ( Δ μ a ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 10. Lines for which, if the optode is moved along them, a recovered optical property will not change (i.e., lines of zero change), e.g., the solid red line which passes through source 1 shows positions where source 1 can be placed and the μ a recovered by assuming the nominal ρ s will be the same. (a) The LINeaR (LINR) arrangement. (b) The TRAPezoidal (TRAP) arrangement. Symbols: Reduced scattering coefficient ( μ s ), absorption coefficient change ( Δ μ a ), optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g010
Figure 11. Proposed support structure, which was designed by allowing moving along the lines of zero change (Figure 10) but restricting movement perpendicular to the line. Lines in this figure considered zero change in μ a . (a) The LINeaR (LINR) arrangement. (b) The TRAPezoidal (TRAP) arrangement. Symbols: Optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Figure 11. Proposed support structure, which was designed by allowing moving along the lines of zero change (Figure 10) but restricting movement perpendicular to the line. Lines in this figure considered zero change in μ a . (a) The LINeaR (LINR) arrangement. (b) The TRAPezoidal (TRAP) arrangement. Symbols: Optical property recovered with the optode displaced from the nominal position ( ( × ρ ) in superscript).
Photonics 12 00697 g011
Table 1. Comparison of the accuracy of the slopes [2] versus the iterative [16] recovery method.
Table 1. Comparison of the accuracy of the slopes [2] versus the iterative [16] recovery method.
Recovery Method
ρ s * μ a μ s Slopes Iterative
(mm) ( mm 1 ) ( mm 1 ) μ a ( ρ ) μ a μ a (%) μ s , ( ρ ) μ s μ s (%) μ a ( ρ ) μ a μ a (%) μ s , ( ρ ) μ s μ s (%)
[25, 25, 37, 37]0.0050.511−120.024−0.99
0.0051.09.3−2.90.0058−0.50
0.0051.57.0−1.20.0013−0.33
0.0100.55.0−120.029−2.0
0.0101.04.7−3.50.0038−0.99
0.0101.53.6−1.70.0046−0.67
0.0150.52.9−130.037−3.0
0.0151.03.1−4.00.0040−1.5
0.0151.52.4−2.10.0025−1.00
[20, 30, 32, 42]0.0050.511−130.025−1.00
0.0051.09.7−3.20.0070−0.50
0.0051.57.4−1.30.0016−0.33
0.0100.54.7−130.032−2.0
0.0101.04.9−3.80.0046−0.99
0.0101.53.8−1.80.0011−0.66
0.0150.52.6−140.041−3.0
0.0151.03.3−4.30.0048−1.5
0.0151.52.6−2.20.0030−1.00
Symbols: Source-detector distance ( ρ ), absorption coefficient ( μ a ), reduced scattering coefficient ( μ s ), true optical property of the medium ( μ ), optical property recovered with the optode in the nominal position ( ( ρ ) in superscript). * [25, 25, 37, 37] corresponds to the LINR, TRAP, and DRCT arrangements; [20, 30, 32, 42] corresponds to the ALIN arrangement.
Table 2. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the LINeR (LINR) arrangement.
Table 2. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the LINeR (LINR) arrangement.
Recovery Method
μ a μ s Optode Slopes Iterative
( mm 1 ) ( mm 1 ) σ μ a 2 mm μ a ( ρ ) (%) σ μ s 2 mm μ s , ( ρ ) (%) σ μ a 2 mm μ a ( ρ ) (%) σ μ s 2 mm μ s , ( ρ ) (%)
0.0050.51 or 21.10.561.20.44
0.0050.5A or B5.1165.013
0.0051.01 or 20.820.520.920.50
0.0051.0A or B3.5154.614
0.0051.51 or 20.680.460.750.45
0.0051.5A or B2.9143.814
0.0100.51 or 20.680.370.710.29
0.0100.5A or B3.4153.013
0.0101.01 or 20.530.370.560.35
0.0101.0A or B2.3142.813
0.0101.51 or 20.440.330.460.32
0.0101.5A or B1.9142.313
0.0150.51 or 20.530.270.540.22
0.0150.5A or B2.8152.312
0.0151.01 or 20.420.290.440.28
0.0151.0A or B1.9142.213
0.0151.51 or 20.340.270.360.26
0.0151.5A or B1.5131.813
Symbols: Absorption coefficient ( μ a ), reduced scattering coefficient ( μ s ), root-mean-squared-error from orbiting the optode position around the nominal optode position in a 2 mm diameter circle ( σ 2 mm ), optical property recovered with the optode in the nominal position ( ( ρ ) in superscript).
Table 3. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the Asymmetric-LINear (ALIN) arrangement.
Table 3. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the Asymmetric-LINear (ALIN) arrangement.
μ a μ s
( mm 1 ) ( mm 1 ) Optode σ μ a 2 mm μ a ( ρ ) (%) σ μ s 2 mm μ s , ( ρ ) (%)
0.0050.511.50.56
0.0050.520.880.34
0.0050.5A5.313
0.0050.5B4.613
0.0051.011.30.69
0.0051.020.670.37
0.0051.0A5.014
0.0051.0B4.314
0.0051.511.10.64
0.0051.520.540.33
0.0051.5A4.214
0.0051.5B3.714
0.0100.510.950.37
0.0100.520.540.23
0.0100.5A3.213
0.0100.5B2.812
0.0101.010.810.48
0.0101.020.410.26
0.0101.0A3.113
0.0101.0B2.713
0.0101.510.680.45
0.0101.520.340.24
0.0101.5A2.614
0.0101.5B2.313
0.0150.510.730.28
0.0150.520.420.18
0.0150.5A2.412
0.0150.5B2.112
0.0151.010.630.38
0.0151.020.320.21
0.0151.0A2.413
0.0151.0B2.113
0.0151.510.520.37
0.0151.520.260.19
0.0151.5A2.013
0.0151.5B1.713
Symbols: Absorption coefficient ( μ a ), reduced scattering coefficient ( μ s ), root-mean-squared-error from orbiting the optode position around the nominal optode position in a 2 mm diameter circle ( σ 2 mm ), optical property recovered with the optode in the nominal position ( ( ρ ) in superscript).
Table 4. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the TRAPezoidal (TRAP) arrangement.
Table 4. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the TRAPezoidal (TRAP) arrangement.
μ a μ s
( mm 1 ) ( mm 1 ) Optode σ μ a 2 mm μ a ( ρ ) (%) σ μ s 2 mm μ s , ( ρ ) (%)
0.0050.51 or 21.52.5
0.0050.5A or B3.79.3
0.0051.01 or 21.22.7
0.0051.0A or B3.39.9
0.0051.51 or 21.02.7
0.0051.5A or B2.79.9
0.0100.51 or 20.912.4
0.0100.5A or B2.28.9
0.0101.01 or 20.772.6
0.0101.0A or B2.09.5
0.0101.51 or 20.632.6
0.0101.5A or B1.79.5
0.0150.51 or 20.692.4
0.0150.5A or B1.78.8
0.0151.01 or 20.592.5
0.0151.0A or B1.69.3
0.0151.51 or 20.492.5
0.0151.5A or B1.39.3
Symbols: Absorption coefficient ( μ a ), reduced scattering coefficient ( μ s ), root-mean-squared-error from orbiting the optode position around the nominal optode position in a 2 mm diameter circle ( σ 2 mm ), optical property recovered with the optode in the nominal position ( ( ρ ) in superscript).
Table 5. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the Diagonal-ReCTangular (DRCT) arrangement.
Table 5. Errors in the recovered optical properties from a 1 mm displacement in any direction relative to the value recovered in the nominal position, for the Diagonal-ReCTangular (DRCT) arrangement.
μ a μ s
( mm 1 ) ( mm 1 ) Optode σ μ a 2 mm μ a ( ρ ) (%) σ μ s 2 mm μ s , ( ρ ) (%)
0.0050.51, 2, A, or B3.79.3
0.0051.01, 2, A, or B3.39.9
0.0051.51, 2, A, or B2.79.9
0.0100.51, 2, A, or B2.28.9
0.0101.01, 2, A, or B2.09.5
0.0101.51, 2, A, or B1.79.5
0.0150.51, 2, A, or B1.78.8
0.0151.01, 2, A, or B1.69.3
0.0151.51, 2, A, or B1.39.3
Symbols: Absorption coefficient ( μ a ), reduced scattering coefficient ( μ s ), root-mean-squared-error from orbiting the optode position around the nominal optode position in a 2 mm diameter circle ( σ 2 mm ), optical property recovered with the optode in the nominal position ( ( ρ ) in superscript).
Table 6. Errors in the recovered optical properties from a simultaneous 1 mm displacement by multiple optodes in any direction relative to the value recovered in the nominal positions.
Table 6. Errors in the recovered optical properties from a simultaneous 1 mm displacement by multiple optodes in any direction relative to the value recovered in the nominal positions.
ArrangementOptodes σ μ a 2 mm μ a ( ρ ) (%) σ μ s 2 mm μ s , ( ρ ) (%)
LINR1 and A2.913
1 and B2.913
1 and 20.800.49
A and B4.019
1, A, and B4.019
1, 2, and A2.913
1, 2, A, and B4.119
ALIN1 and A3.213
1 and B2.813
2 and A3.113
2 and B2.713
1 and 20.910.55
A and B4.119
1, A, and B4.219
2, A, and B4.119
1, 2, and A3.213
1, 2, and B2.813
1, 2, A, and B4.219
TRAP1 and A2.29.8
1 and B2.29.8
1 and 21.13.6
A and B2.913
1, A, and B3.014
1, 2, and A2.310
1, 2, A, and B3.114
DRCT1 and A2.913
1 and B2.913
1 and 22.913
1, A, and B3.516
1, 2, A, and B4.119
Symbols: Absorption coefficient ( μ a ), reduced scattering coefficient ( μ s ), root-mean-squared-error from orbiting the optode position around the nominal optode position in a 2 mm diameter circle ( σ 2 mm ), optical property recovered with the optode in the nominal position ( ( ρ ) in superscript).
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Blaney, G.; Sassaroli, A.; Das, T.; Fantini, S. Influence of Uncertainties in Optode Positions on Self-Calibrating or Dual-Slope Diffuse Optical Measurements. Photonics 2025, 12, 697. https://doi.org/10.3390/photonics12070697

AMA Style

Blaney G, Sassaroli A, Das T, Fantini S. Influence of Uncertainties in Optode Positions on Self-Calibrating or Dual-Slope Diffuse Optical Measurements. Photonics. 2025; 12(7):697. https://doi.org/10.3390/photonics12070697

Chicago/Turabian Style

Blaney, Giles, Angelo Sassaroli, Tapan Das, and Sergio Fantini. 2025. "Influence of Uncertainties in Optode Positions on Self-Calibrating or Dual-Slope Diffuse Optical Measurements" Photonics 12, no. 7: 697. https://doi.org/10.3390/photonics12070697

APA Style

Blaney, G., Sassaroli, A., Das, T., & Fantini, S. (2025). Influence of Uncertainties in Optode Positions on Self-Calibrating or Dual-Slope Diffuse Optical Measurements. Photonics, 12(7), 697. https://doi.org/10.3390/photonics12070697

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop