Uncertainty of Postmortem Time Estimation Based on Potassium Ion Determination in Vitreous Humor Using Potentiometric Ion-Selective Electrode and Microwave-Induced Plasma with Optical Emission Spectrometry Methods

by
Sonia Zięba
1,†,
Marek Wiergowski
1,*,†,
Bartłomiej Michał Cieślik
2,
Jacek Sein Anand
3,4 and
Marta Krzyżanowska
1
1
Department of Forensic Medicine, Faculty of Medicine, Medical University of Gdańsk, M. Skłodowskiej-Curie 3a Str., 80-210 Gdańsk, Poland
2
Department of Analytical Chemistry, Faculty of Chemistry, Gdańsk University of Technology, Narutowicza 11/12 Str., 80-233 Gdańsk, Poland
3
Division of Clinical Toxicology, Faculty of Health Sciences with the Institute of Maritime and Tropical Medicine, Medical University of Gdańsk, M. Skłodowskiej-Curie 3a Str., 80-210 Gdańsk, Poland
4
Pomeranian Center of Toxicology, Kartuska 4/6 Str., 80-104 Gdańsk, Poland
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Separations 2023, 10(3), 201; https://doi.org/10.3390/separations10030201
Submission received: 12 February 2023 / Revised: 6 March 2023 / Accepted: 10 March 2023 / Published: 14 March 2023
(This article belongs to the Section Forensics/Toxins)

Abstract

:
There is a need for a reliable and independent evaluation and confirmation of the postmortem interval (PMI) based on objective factors other than only postmortem changes or temperature measurements. Estimating the PMI by examining the concentration of potassium ions in the vitreous humor (VH) has a tradition in forensic toxicology dating back to the mid-20th century. So far, the methods for determining the presence of potassium ions have not been characterized in terms of the measurement uncertainty of types A and B, which directly affect the estimation of time of death uncertainty. The study evaluated the uncertainty of the determination of potassium ion concentrations using potentiometric ion-selective electrode (ISE) and microwave-induced plasma with optical emission spectrometry (MIP-OES) methods. In addition, the influence of the estimated measurement uncertainty on the results of the analysis of potassium ions in the VH was investigated. The estimated values of the expanded uncertainty determined by the type A experimental method indicate lower uncertainty in the determination of potassium ion concentration in the case of MIP-OES than ISE; that is, for concentrations of 2, 10, and 25 mg/L of potassium ions, the expanded uncertainties by MIP-OES were 1.2%, 2.2%, and 2.5% and the uncertainties by ISE were 12.2%, 6.1%, and 3.1%, respectively. Similarly, in the case of estimating the uncertainty of type B, the MIP-OES method compared to ISE was characterized by almost two times lower expanded uncertainty; that is, for MIP-OES, it was 2.53%, and for ISE, it was 4.75%. Both methods of uncertainty estimation, types A and B, can be used interchangeably, as they provide comparable results.

Graphical Abstract

1. Introduction

Estimating the postmortem interval (PMI, also called “time of death” [TOD] or “time since death” [TSD]) is an important element in prosecution proceedings (e.g., in the case of searching for a person suspected of murder). Witness testimonies are often associated with a high risk of inaccuracy, manipulation, or even forgery. Witnesses may also have prejudices against the suspect. However, PMI determination can verify the course of events, including limiting the number of suspects or verifying their alibi [1,2].
There are many methods for estimating the PMI. Nowadays, more and more often attempts are made to use microRNAs as useful tool to estimate TSD [3]. The frequently used ones include temperature measurement (which is usually performed in the anus), evaluation of rigor mortis, and assessment of livor mortis. Among these, the temperature measurement method has the widest scope of application. Unfortunately, the time window for the PMI in which it can be used is 36 h [1]. Methods with longer time windows for the PMI include biochemical processes in which the determination of the TOD is based on the correlation of the concentration of substances (e.g., potassium ions) in biological material (most often in the vitreous humor [VH]). The method of assessing potassium concentration in the VH enables us to determine the TOD within 120 h [4]. The VH is a valuable postmortem material because it is composed primarily of water and the autolysis process takes place much slower in it than in peripheral blood. However, this type of material also has disadvantages, such as low volume (1–2 mL) and relatively high viscosity. The analysis of the concentration of potassium ions concerning the PMI dates back to the 1960s [5,6,7]. Since then, researchers have proposed more than a dozen formulas that can be used to estimate the PMI (the uncertainty of this time estimate was difficult to determine) (Table 1).
The most frequently used method to measure the concentration of potassium ions in the VH is direct potentiometry in the form of an ion-selective electrode (ISE) and, less frequently, the method of flame photometry, for example, in the form of microwave-induced plasma with optical emission spectrometry (MIP-OES) (Table 2).
To the best of the authors’ knowledge, both ISE and MIP-OES methods have not yet been characterized in terms of measurement uncertainty of types A and B, which directly affect the estimation of PMI uncertainty. Measurement uncertainty is the numerically expressed dispersion of the measured quantity, in which it is expected, with a certain probability, to find the true measurement value. Measurement uncertainty can be determined using two methodologies: type A and type B. The type A method most often uses a series of measurements and determines the variation of the results (e.g., standard deviation), considering the coverage factor equal to 2 and the corresponding confidence interval. The type B method uses data contained in the specifications of measuring devices, safety data sheets for reagents, laboratory equipment, and previous tests described in publications and test reports. The feature that distinguishes the type B from the type A method is the lack of necessity to conduct experimental tests to estimate the uncertainty.
This study aimed to estimate the uncertainty of potassium ion determination using ISE and MIP-OES methods and to examine the impact of the estimated measurement uncertainty on the analysis of potassium ions in the VH and, consequently, on the PMI estimation.

2. Materials and Methods

2.1. Sample Collection

During the medicolegal autopsy in the Department of Forensic Medicine of the Medical University of Gdańsk in 2021–2022, VH samples were initially collected from 42 deceased individuals. The deceased with head injuries (risk of VH contamination by peripheral blood) and those who died more than 120 h before dissection were excluded from the study. After sampling, the material was transferred to 2 mL microcentrifuge tubes (Eppendorf-type vials) and clearly labeled. The collected VH samples were centrifuged shortly after dissection for 10 min at 3500 rpm and transferred to 2 mL Eppendorf vials, which were then stored at −20 °C until analysis.
Table 3 summarizes the data of the deceased: gender, age of death, blood alcohol content and other biological materials, cause of death, TOD or disclosure of the body, and time of postmortem examination.
The results of the deceased who died due to hanging, who may also rupture the blood vessels of the eyes, should be interpreted with great care. Samples from 26 deceased individuals were taken for further analysis.
From the TOD to the autopsy examination, the time usually did not exceed a few days. During the time preceding the autopsy, the bodies were stored at +4 °C. In the case of analytical methods, the time from the moment of collecting the vitreous humor during autopsy to the moment of taking measurements is important. The storage time of the vitreous humor samples at −20 °C is also essential, as we assume no significant changes in the chemical composition during this time, which is not necessarily accurate.

2.2. VH Preparation

The potassium ion determination procedure was developed based on the research results of Mihailovic et al. [15] and Yang et al. [27] (Table 2). Below, we present the methodology optimized to our conditions. Samples with the centrifuged VH (supernatant) were left at room temperature for about 1 h after removal from the −20 °C freezer. Each sample was vortexed for 15 s. Then, 50 μL of the supernatant was collected twice into two 5 mL Eppendorf vials and refilled (completely) with deionized water. As a result, an 80-fold dilution was achieved. Each sample was vortexed for 15 s. Depending on the method, the remaining steps of the procedure differed from each other and are described in Table 4.

2.3. Instrumentation

Measurements were carried out using two methods: direct potentiometry (ISE) and plasma-excited optical emission spectrometry (MIP-OES). Measurements by direct potentiometry were conducted using a combination of ISE (HI 4114, Hanna Instruments) and a pH meter (S220, Mettler Toledo SevenCompact). When performing measurements using the MIP-OES method with plasma excitation, the Agilent 4210 MP-AES (the older and trade name derived from “microwave plasma–atomic emission spectroscopy”) device was applied, and mineralization was performed with the Multiwave GO Plus by Anton Paar.

3. Results

Calibration curves obtained for potassium ion concentrations ranging from 2 to 100 mg/L by the ISE method in three measurement series (performed on three different days) carried out with six replications for each concentration are presented in Figure 1.
Calibration curves obtained for potassium ion concentrations ranging from 1 to 25 mg/L by the MIP-OES method for two different wavelengths are shown in Figure 2.
The measurement uncertainty, and hence the reliability of PMI estimation, is influenced by several factors, including the properties of the biological material, the measurement method applied, the process of sample preparation, the sample concentration, and the cause of death (Figure 3 and Figure 4).
To obtain the A-type uncertainty, water samples with low (2 mg/L), medium (10 mg/L), and high (25 mg/L) potassium ion concentrations were tested using the ISE and MIP-OES methods, and the results with the estimated uncertainty are presented in Table 5.
The components of the uncertainty budget of ISE and MIP-OES methods to estimate type B uncertainty are presented in Table 6 and Table 7, respectively.
Using the components of the uncertainty budget listed in Table 6 and Table 7, the expanded uncertainty using the ISE and MIP-OES methods was estimated (Table 8).
Table 9 presents 26 cases for which the PMI was determined (based on the data from Table 3), and the concentration of potassium ions was determined using the ISE and MIP-OES methods. The concentration units were converted from mg/L to mmol/L.

4. Discussion

4.1. Properties of Biological Material

VH is characterized by a much lower rate of decomposition changes (or increased resistance to autolysis) than blood. However, the described type of material has limitations related to its relatively high viscosity and small volume (up to 2 mL). Because of the above aspects, the samples were diluted to reduce the matrix effect. The first factor influencing measurement uncertainty is interfering ions, which include NH4+, Li+, Na+, and Ca2+. To reduce the interference, the samples were diluted 80-fold.
It is important to keep the temperature constant during the entire test (in the range of 20–25 °C) because, depending on the temperature, the measurement results may fluctuate. According to the manufacturer, the temperature range of the ISE operation is equal to 0–40 °C. Maintaining the pH of the measuring environment also influences the measurement, and the environment is stable in the range of 1.5–12.00 pH.
VH samples were collected using a metal, pointed (35 × 2 mm) injection needle. The time the samples were collected varied according to the duration of the autopsy. The bodies of the deceased were stored in autopsy refrigerators (4 °C) at various times.

4.2. Measurement Uncertainty by ISE and MIP-OES

In this study, an attempt was made to estimate the uncertainty components (A and B types) related to sample preparation and instrumental analysis using ISE and MIP-OES (Table 10).
The estimated values of the expanded uncertainty determined by the experimental method A (UA) indicate a few to even ten-times lower uncertainty in the determination of potassium ion concentration in the case of the MIP-OES method. Similarly, in the case of estimating the expanded uncertainty with the B (UB) method, the MIP-OES method is characterized by twice the lower uncertainty. Based on our experience in determining the concentration of potassium ions using the ISE and MIP-OES methods, it is worth considering their advantages and disadvantages before implementing them into diagnostic practice (Table 11).

4.3. Causes of Death

The causes of death that affect the postmortem concentration of potassium ions in VH are hanging, diabetic coma, and electrolyte disturbances. The increased concentration of potassium ions in hanging cases is due to the tightness of the loops in the neck, which causes a rupture of the jugular system and a significant increase in venous pressure. The cause of death strongly influences the change in the concentration of potassium ions after death, which is mainly related to the possibility of blood contamination of the VH. Venous circulation can lead to capillary hyperemia and thus increase vascular leakage from the capillaries into the retina [23,30]. In addition, if metabolic acidosis occurs, hyperkalemia is expected, which may make it impossible to reliably estimate the PMI from the concentration of potassium ions.
Differences were observed in lethal diabetic coma. It is known that, in a diabetic coma, there is a potassium liberation from the cell to the outside due to insulin deficiency and acidosis [21,31].
Any sudden mechanical movement during or after death can introduce the breaking of blood vessels in the eyes and changes in the concentration of potassium ions. The uncertainty of determination related strictly to the ISE and MIP-OES methods is relatively small concerning the other factors influencing the concentration of potassium ions (cause of death, ambient temperature, and cadaver storage conditions). However, based on our studies, the uncertainty of type A (experimental) for low concentrations (2 mg/L) of potassium ions can be as high as 12.2% using the ISE method. Reducing the uncertainty of PMI estimation can be performed by introducing additional indicators that determine the TOD zystkie surowe dane mogą być udostępnione na prośbę skierowaną do autora korespondencyjnego. (muscle degradation products by means of quantitative MRI).

5. Conclusions

The postmortem concentration of potassium ions in the VH has a limited PMI time range, which usually does not exceed four days, and may be influenced by interfering factors, including analytical factors (e.g., sample viscosity, the necessity of dilution, failures, and nonselectivity of the ISE) and nonanalytical factors (e.g., age, cause of death, and ambient temperature).
To determine the concentration of potassium ions in the VH, the analytical uncertainty of types A and B should be taken into account, as it generally has a significant effect on the estimation of PMI. Both A and B estimates give similar results, which may prove the usefulness of both methods in estimating the measurement uncertainty of the concentration of potassium ions, and thus PMI estimation.
The cause of death also influences the postmortem concentration of potassium ions in the VH. Caution should be exercised if death occurs due to hanging or electrolyte disturbances, as they may significantly affect the measurement results and thus the estimated TOD.

Author Contributions

Conceptualization, S.Z., M.W. and B.M.C.; methodology, S.Z., M.W., B.M.C. and M.K.; validation, S.Z., B.M.C. and M.W.; investigation, S.Z., M.W. and B.M.C.; resources, S.Z. and M.W.; writing—original draft preparation, S.Z., M.W., B.M.C., J.S.A. and M.K.; writing—review and editing, S.Z., M.W., B.M.C., J.S.A. and M.K.; supervision, M.W. and J.S.A. All authors have read and agreed to the published version of the manuscript.

Funding

The publication of the article was supported by the project POWR.03.02.00-00-I014/17-00 co-financed by the European Union through the European Social Fund under the Operational Programme Knowledge Education Development 2014–2020.

Informed Consent Statement

Patient consent was waived due to the consent of the bioethics committee of the Medical University of Gdańsk, dated 28 February 2020 (NKBBN/18/2020).

Data Availability Statement

All research data can be made available upon request send to the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Henßge, C.; Madea, B. Estimation of the time since death in the early post-mortem period. Forensic Sci. Int. 2004, 144, 167–175. [Google Scholar] [CrossRef]
  2. Mathur, A.; Agrawal, Y.K. An overview of methods used for estimation of time since death. Aust. J. Forensic Sci. 2011, 43, 275–285. [Google Scholar] [CrossRef]
  3. Maiese, A.; Scatena, A.; Costantino, A.; Di Paolo, M.; La Russa, R.; Turillazzi, E.; Frati, P.; Fineschi, V. MicroRNAs as useful tools to estimate time since death. A systematic review of current literature. Diagnostics 2021, 11, 64. [Google Scholar] [CrossRef] [PubMed]
  4. Madea, B. Methods for determining time of death. Forensic Sci. Med. Pathol. 2016, 12, 451–485. [Google Scholar] [CrossRef] [PubMed]
  5. Markowska, J.; Szopa, M.; Zawadzki, M.; Piekoszewski, W. Ciało szkliste oka—Rutynowy czy alternatywny materiał do badań w medycynie sądowej. Arch. Med. Sadowej Kryminol. 2017, 67, 201–213. [Google Scholar] [CrossRef] [PubMed]
  6. Madea, B. Is there recent progress in the estimation of the postmortem interval by means of thanatochemistry? Forensic Sci. Int. 2005, 151, 139–149. [Google Scholar] [CrossRef]
  7. McCleskey, B.C.; Dye, D.W.; Davis, G.G. Review of postmortem interval estimation using vitreous humor: Past, present, and future. Acad. Forensic Pathol. 2016, 6, 12–18. [Google Scholar] [CrossRef]
  8. Adelson, L.; Sunshine, I.; Rushforth, N.B.; Mankoff, M. Vitreous potassium concentration as an indicator of the postmortem interval. J. Forensic Sci. 1963, 8, 503–514. [Google Scholar]
  9. Coe, J.I. Vitreous potassium as a measure of the postmortem interval: An historical review and critical evaluation. Forensic Sci. Int. 1989, 42, 201–213. [Google Scholar] [CrossRef]
  10. Ross, J.; Hoadley, P.; Sampson, B. Determination of postmortem interval by sampling vitreous humour. Am. J. Forensic Med. Pathol. 1997, 18, 158–162. [Google Scholar]
  11. Barús, J.I.M.; Suárez-Peñaranda, J.M.; Otero, X.L.; Rodríguez-Calvo, M.S.; Costas, E.; Miguéns, X.; Concheiro, L. Improved estimation of postmortem interval based on differential behaviour of vitreous potassium and hypoxantine in death by hanging. Forensic Sci. Int. 2002, 125, 67–74. [Google Scholar] [CrossRef] [PubMed]
  12. Zhou, B.; Zhang, L.; Zhang, G.; Zhang, X.; Jiang, X. The determination of potassium concentration in vitreous humor by low pressure ion chromatography and its application in the estimation of postmortem interval. J. Chromatogr. B Anal. Technol. Biomed. Life Sci. 2007, 852, 278–281. [Google Scholar] [CrossRef] [PubMed]
  13. Jashnani, K.D.; Kale, S.A.; Rupani, A.B. Vitreous humor: Biochemical constituents in estimation of postmortem interval. J. Forensic Sci. 2010, 55, 1523–1527. [Google Scholar] [CrossRef] [PubMed]
  14. Bortolotti, F.; Pascali, J.P.; Davis, G.G.; Smith, F.P.; Brissie, R.M.; Tagliaro, F. Study of vitreous potassium correlation with time since death in the postmortem range from 2 to 110 hours using capillary ion analysis. Med. Sci. Law 2011, 51 (Suppl. S1), 20–23. [Google Scholar] [CrossRef] [PubMed]
  15. Mihailovic, Z.; Atanasijevic, T.; Popovic, V.; Milosevic, M.B.; Sperhake, J.P. Estimation of the postmortem interval by analyzing potassium in the vitreous humor: Could repetitive sampling enhance accuracy? Am. J. Forensic Med. Pathol. 2012, 33, 400–403. [Google Scholar] [CrossRef]
  16. Siddhamsetty, A.K.; Verma, S.K.; Kohli, A.; Verma, A.; Puri, D.; Singh, A. Exploring time of death from potassium, sodium, chloride, glucose & calcium analysis of postmortem synovial fluid in semi arid climate. J. Forensic Leg. Med. 2014, 28, 11–14. [Google Scholar] [CrossRef] [PubMed]
  17. Bohra, B.; Verma, R.; Mathur, I.B.; Sharma, Y.; Khangwal, V.P. Estimation of postmortem interval by measuring potassium level in vitreous humor. J. Indian Acad. Forensic Med. 2014, 36, 374–378. [Google Scholar]
  18. Foster, S.N.; Smith, P.R.; Biggs, M.; Rutty, G.N.; Hollingbury, F.E.; Morley, S.R. Estimation of postmortem interval using vitreous potassium levels in cases of fatal road traffic collision. Arch. Med. Sadowej Kryminol. 2016, 66, 71–82. [Google Scholar] [CrossRef]
  19. Murthy, A.S.; Das, S.; Thazhath, H.K.; Chaudhari, V.A.; Adole, P.S. The effect of cold chamber temperature on the cadaver’s electrolyte changes in vitreous humor and plasma. J. Forensic Leg. Med. 2019, 62, 87–91. [Google Scholar] [CrossRef]
  20. Focardi, M.; Lanzilao, L.; Bonari, A.; Lazzeretti, M.; Lorubbio, M.; Ognibene, A.; Gualco, B.; Fanelli, A.; Pinchi, V. Improvement in estimation of time since death by albumin and potassium concentrations in vitreous humor. Forensic Sci. Int. 2020, 314, 110393. [Google Scholar] [CrossRef] [PubMed]
  21. Zilg, B.; Bernard, S.; Alkass, K.; Berg, S.; Druid, H. A new model for the estimation of time of death from vitreous potassium levels corrected for age and temperature. Forensic Sci. Int. 2015, 254, 158–166. [Google Scholar] [CrossRef] [Green Version]
  22. Rognum, T.O.; Hauge, S.; Øyasaeter, S.; Saugstad, O.D. A new biochemical method for estimation of postmortem time. Forensic Sci. Int. 1991, 51, 139–146. [Google Scholar] [CrossRef] [PubMed]
  23. Chandrakanth, H.V.; Kanchan, T.; Balaraj, B.M.; Virupaksha, H.S.; Chandrashekar, T.N. Postmortem vitreous chemistry—An evaluation of sodium, potassium and chloride levels in estimation of time since death (during the first 36 h after death). J. Forensic Leg. Med. 2013, 20, 211–216. [Google Scholar] [CrossRef]
  24. Cordeiro, C.; Ordóñez-Mayán, L.; Lendoiro, E.; Febrero-Bande, M.; Vieira, D.N.; Muñoz-Barús, J.I. A reliable method for estimating the postmortem interval from the biochemistry of the vitreous humor, temperature and body weight. Forensic Sci. Int. 2019, 295, 157–168. [Google Scholar] [CrossRef] [PubMed]
  25. Hansson, L.; Uotila, U.; Lindfors, R. Potassium content of the vitreous body as an aid in determining the time of death. J. Forensic Sci. 1966, 11, 390–394. [Google Scholar]
  26. Moriya, F.; Hashimoto, Y. Endogenous γ-hydroxybutyric acid levels in postmortem blood. Leg. Med. 2004, 6, 47–51. [Google Scholar] [CrossRef] [PubMed]
  27. Yang, M.; Li, H.; Yang, T.; Ding, Z.; Wu, S.; Qiu, X.; Liu, Q. A study on the estimation of postmortem interval based on environmental temperature and concentrations of substance in vitreous humor. J. Forensic Sci. 2018, 63, 745–751. [Google Scholar] [CrossRef]
  28. Stephen, R.J. Vitreous humor chemistry. The use of potassium concentration for the prediction of postmortem interval. J. Forensic Sci. 1987, 32, 503–509. [Google Scholar] [CrossRef]
  29. Eppendorf.com. Eppendorf Research Instruction Manual. 2016. Available online: http://www.sun-way.com.tw/Files/DownloadFile/Researchplus.pdf (accessed on 28 February 2016).
  30. Hanna Instruments. Instruction Manual Potassium Ion Selective Electrode HI 4114, MAN4114 07/06 REV3. 2007. Available online: https://www.hannainst.com/hi4114-potassium-combination-ion-selective-electrode.html (accessed on 23 November 2022).
  31. Kitabchi, A.E. Hyperglycemic crises in diabetes. Diabetes Care 2004, 27 (Suppl. S1), s94–s102. [Google Scholar] [CrossRef] [Green Version]
Figure 1. Calibration curves for the determination of potassium ion concentration by the ISE method in three different measurement series: 10 March 2022 (a), 11 March 2022 (b), and 14 March 2022 (c).
Figure 1. Calibration curves for the determination of potassium ion concentration by the ISE method in three different measurement series: 10 March 2022 (a), 11 March 2022 (b), and 14 March 2022 (c).
Separations 10 00201 g001aSeparations 10 00201 g001b
Figure 2. Calibration curves for the determination of potassium ion concentration by the MIP-OES method for two wavelengths at 766.49 nm (dotted line) and 769.89 nm (solid line).
Figure 2. Calibration curves for the determination of potassium ion concentration by the MIP-OES method for two wavelengths at 766.49 nm (dotted line) and 769.89 nm (solid line).
Separations 10 00201 g002
Figure 3. Cause-and-effect diagram for ion potassium concentration in VH using the ISE method, showing the key components that contribute to measurement uncertainty.
Figure 3. Cause-and-effect diagram for ion potassium concentration in VH using the ISE method, showing the key components that contribute to measurement uncertainty.
Separations 10 00201 g003
Figure 4. Cause-and-effect diagram for ion potassium concentration in VH using the MIP-OES method, showing the key components that contribute to measurement uncertainty.
Figure 4. Cause-and-effect diagram for ion potassium concentration in VH using the MIP-OES method, showing the key components that contribute to measurement uncertainty.
Separations 10 00201 g004
Table 1. Formulas for estimating the PMI based on the concentration of potassium ions ([K+] in mmol/L).
Table 1. Formulas for estimating the PMI based on the concentration of potassium ions ([K+] in mmol/L).
No.Author(s) and YearFormula Proposed
1Adelson et al. (1963) [8]PMI = 5.88 [K+] − 31.53
2Coe (1969) [9]PMI = 6.15 [K+] − 38.10
3Henßge and Madea (1989) [1]PMI = 5.26 [K+] − 30.90
4Ross et al. (1997) [10]PMI = 4.32 [K+] − 18.35
5Muñoz Barús et al. (2002) [11]PMI = 3.63 [K+] − 17.33
6Zhou et al. (2007) [12]PMI = 5.88 [K+] − 32.71
7Jashnani et al. (2010) [13]PMI = 1.08 [K+] − 2.82
8Bortolotti et al. (2011) [14]PMI = 5.77 [K+] − 13.28
9Mihailovic et al. (2012) [15]PMI = 2.75 [K+] − 11.98
10Siddhamsetty et al. (2014) [16]PMI = 4.75 [K+] − 27.9
11Bohra et al. (2014) [17]PMI = 3.75 [K+] − 16.22
12Foster et al. (2016) [18]PMI = 6.42 [K+] − 40.94
13Murthy et al. (2019) [19]PMI = 5.26 [K+] − 30.9
14Focardi et al. (2020) [20]PMI = 6.16 [K+] − 32.49
Table 2. Methods applied to determine the postmortem concentration of potassium ions in the VH.
Table 2. Methods applied to determine the postmortem concentration of potassium ions in the VH.
No.Volume [mL]Sample PreparationInstrumental Analysis (Commercial Name of Instrument)Ref.
10.10Samples were stored at −20 °C and centrifuged at 16,000 rpm for 10 min; only the supernatant was decantedPhotometric, potentiometric, turbidimetric (ARCHITECT 8000)[7]
20.10Samples were stored at −20 °C and diluted at 1:20 with a 40-mg/mL aqueous solution of barium chloride (internal standard)Capillary electropherogram with a UV absorbance detector (A PACE MDQ; capillary column: 75 mm ID × 50 cm, 60 cm)[17]
3n/aSamples were centrifugedIon-selective electrode (ADVIA 2400 Chemistry System)[9]
40.20Samples were not pre-treated (without dilution, centrifugation, or sonication)Ion-selective electrode (ABL 625 radiometer with UniCel DxC 800)[21]
51.50–2.00Blood was collected in test tubes with EDTA, the VH was placed in sterile plain vials, and samples were centrifuged at 3500 rpm for 10 minIon-selective electrode (Analyser AU680)[11]
60.10Samples were stored at −70 °CTurbidimetric (Humalyzer Junior)[22]
72.00Samples were centrifuged at 4500 rpm and the supernatant was transferred to another containerIon-selective electrode, flow-through, liquid membrane electrode (Roche 9180 Electrolyte Analyzer)[23]
80.15n/aIndirect potentiometry (Advia 2400).[24]
93.00–4.00Samples were stored at 4 °C and centrifuged at 2000 rpm for 5 minFlame photometry (FLM3, Biolyte 2000)[25]
101.50–2.00Samples were centrifuged at 3500 rpm for 10 minIon-selective electrode (indirect potentiometry method)[26]
11n/aSamples were stored at 4 °C and centrifuged at 3000 rpm for 10 minIndirect potentiometry (BM/747)[27]
12n/aSamples were centrifuged at 13,000× g for 10 min, supernatant solutions were stored at −80 °C and vortexed for 10 s, and viscous VH samples were diluted with deionized waterPhotometric, potentiometric, turbidimetric (ARCHITECT c16000)[15]
132.00Samples were stored at −18 °C and −70 °C and centrifuged at 3000 rpm for 10 minIon-selective electrode (Beckman auto-analyzer)[28]
142.00Samples were stored at −80 °C, vortexed for 30 s using the highest level, and centrifuged at 1650× g for 8 min; the supernatant was divided into four aliquotsPotentiometric method (VLYTE1 Integrated Multisensor K800A)[20]
UV, ultraviolet detection; EDTA, ethylenediaminetetraacetic acid.
Table 3. Selected data of the deceased individuals autopsied in 2021–2022.
Table 3. Selected data of the deceased individuals autopsied in 2021–2022.
NoSex 1Age (y)Circumstances, Mode, and Possible Cause of DeathConcentration of Ethanol (‰)Estimated Time of Death Time of AutopsyPMI (h)
1M40Acute circulatory and respiratory failure, cirrhosis, ethanol poisoning0.92 (blood),
1.13 (VH)
6 June 2020 19:478 June 2020 12:0040
2M39Acute circulatory and respiratory failure, asphyxiation (by hanging), ethanol poisoning3.18 (blood),
3.77 (urine)
19 July 2020 21:2021 July 2020 09:0036
3M50Acute cardiorespiratory failure, urosepsis, acute bacterial interstitial nephritisNegative 226 July 2020 05:3530 July 2020 08:0098
4M32Acute circulatory and respiratory failure, asphyxiation (by hanging), ethanol poisoning0.69 (blood)
0.22 (urine)
26 July 2020 22:3028 July 2020 09:3035
5F58Acute circulatory and respiratory failure, acid–base disorders, ethanol addiction, cachexiaNegative3 August 2020 13:597 August 2020 12:0094
6M63Acute circulatory and respiratory failure, asphyxiation (blood aspiration), oral cancerNegative6 August 2020 18:4510 August 2020 10:3088
7M57Acute circulatory and respiratory failure, bleeding from esophageal varices, cirrhosis, ethanol addiction, ethanol poisoning1.01 (blood)24 August 2020 11:0027 August 2020 10:0071
8M41Acute circulatory and respiratory failure, alcohol-induced liver damage, hygiene negligenceNegative6 September 2020 14:5010 September 2020 09:4591
9M54Acute circulatory failure, hypertrophic and dilated cardiomyopathy, condition after aortic valve implantation, obesityNegative14 September 2020 14:2016 September 2020 10:3044
10M42Acute heart failure, hypertrophic cardiomyopathy, arrhythmiaNegative14 September 2020 11:0016 September 2020 09:1546
11M55Acute circulatory and respiratory failure, disturbances in acid–base and water–electrolyte balance, suspected alcohol ketoacidosis, ethanol addictionNegative15 September 2020 01:1017 September 2020 09:0056
12M37Acute circulatory and respiratory failure, bleeding from the duodenum ulcer, alcohol-induced liver damageNegative16 September 2020 22:2021 September 2020 09:00107
13M46Acute circulatory and respiratory failure, myocardial infarction, pulmonary edema, chronic coronary artery disease, ethanol poisoning2.82 (blood)
4.15 (urine)
20 September 2020 05:5024 September 2020 08:3099
14M56Acute circulatory and respiratory failure, asphyxiation (by hanging)Negative19 September 2020 20:4023 September 20 08:3084
15M63Acute circulatory failure, hypertrophic and dilated cardiomyopathy, generalized atherosclerosisNegative4 October 2020 12:107 October 2020 09:3069
16M49Acute circulatory failure, chronic coronary artery disease, arrhythmiaNegative5 October 2020 15:008 October 2020 10:3068
17M33Acute circulatory and respiratory failure, asphyxiation (by hanging), ethanol poisoning1.73 (blood), 2.85 (urine)17 October 2020 14:4020 October 2020 10:3068
18M57Acute circulatory and respiratory failure, cirrhosis, ascites, chronic pancreatitis, COPD (chronic obstructive pulmonary disease), ethanol addictionNegative25 October 2020 18:0027 October 2020 10:0040
19F61Acute circulatory and respiratory failure, myocardial infarctionNegative13 December 2020 00:0017 December 2020 08:30104
20M59Acute circulatory and respiratory failure, ethanol poisoning4.18 (blood), 3.96 (urine)11 April 2021 14:3014 April 2021 09:0067
21F50Acute circulatory and respiratory failure, venous hemorrhage, ethanol poisoning2.62 (blood), 2.57 (urine)21 April 2021 15:0523 April 2021 09:0042
22F99Acute circulatory and respiratory failure, death from natural causesNegative10 October 2021 11:5514 October 2021 08:1592
23M67Acute circulatory and respiratory failure, arrhythmia, vascular cardiomyopathyNegative31 October 2021 14:153 November 2021 11:0069
24M34Acute circulatory and respiratory failureNegative1 November 2021 12:204 November 2021 10:0070
25M42Acute cardiorespiratory failure, ethanol poisoning, liver damage, pancreatic damage, ethanol addiction3.55 (blood)
3.48 (urine)
4 November 2021 11:309 November 2021 08:15117
26M45Acute circulatory and respiratory failure, asphyxiation (by hanging), ethanol poisoning0.33 (blood)
0.56 (urine)
7 November 2021 12:3010 November 2021 08:1568
1 M, male; F, female. 2 Regarding Polish legal limits, “negative” means below 0.2‰ of ethanol.
Table 4. Procedure for determining potassium ion concentration in VH by ISE and MIP-OES methods.
Table 4. Procedure for determining potassium ion concentration in VH by ISE and MIP-OES methods.
StepISEMIP-OES
1Take a sample of VH during medicolegal autopsy (1–2 mL) using a metal, pointed (35 × 2 mm) injection needle.
2Centrifuge VH at 3500 rpm for 10 min.
3Pipette 50 μL of the supernatant twice in two 5 mL Eppendorf tubes.
4Fill Eppendorf tubes to 4 mL (i.e., 3950 µL) with deionized water.
5Mix on a vortex shaker for 15 s.
6Pour the contents into an intermediate vial (5 mL).
7Add 80 µL of ionic strength adjuster (ISA) solution (NaCl solution) and mix.Quantitatively pour the entire beaker into the mineralizer tube.
8Rinse the ion-selective electrode membrane with deionized water and dry it.Rinse the test tube with nitric acid (3–4 mL) and pour it into the mineralizer.
9Place the electrode in the beaker and wait for the measurement to stabilize for 1 min.Perform microwave-assisted mineralization for 40 min at 180 °C.
10Write down and save the result.After mineralization and reaching the room temperature of the solution, pour the mineralization tube contents quantitatively into 10 mL measuring flasks and then wash the tube with deionized water.
11After the measurements, place the ISE in a solution with a concentration of 100 mg/L of K+.Pour the contents of the beaker into a 15 mL plastic test tube and provide MIP-OES measurements.
12After analysis, store VH samples at −20 °C.After analysis, store VH samples at −20 °C.
Table 5. Type A uncertainty estimation for the determination of potassium ion concentrations of 2 mg/L (low), 10 mg/L (medium), and 25 mg/L (high) using the ISE and MIP-OES methods.
Table 5. Type A uncertainty estimation for the determination of potassium ion concentrations of 2 mg/L (low), 10 mg/L (medium), and 25 mg/L (high) using the ISE and MIP-OES methods.
Concentration of K+Standard Deviation (SD) of ISE (mg/L)Standard Deviation (SD) of MIP-OES (mg/L)Expanded Uncertainty (UA) 1 by ISE (%)Expanded Uncertainty (UA) by MIP-OES (%)
Low
(2 mg/L)
±0.12±0.012±12.2±1.2
Medium
(10 mg/L)
±0.30±0.111±6.1±2.2
High
(25 mg/L)
±0.39±0.308±3.1±2.5
1 Expanded uncertainty is defined as U A = 2 · S D c o n c e n t r a t i o n · 100 % .
Table 6. Components of the uncertainty budget of the ISE method [28,29].
Table 6. Components of the uncertainty budget of the ISE method [28,29].
Procedure Stage and CommentsVolume (V)Precision (P)Accuracy (A)
Pipette 50 µL twice into Eppendorf tubes; variable-capacity pipette 10–100 µL, precision and accuracy for 50 µL0.05 mL±0.7%<0.3%
Fill an Eppendorf tube to 4 mL (i.e., 3950 µL) with deionized water; variable-capacity pipette 500–5000 µL, precision and accuracy for 2500 µL3.95 mL±0.6%<0.15%
Add 80 µL of ISA solution (NaCl solution) and mix; variable-capacity pipette 10–100 µL, precision and accuracy for 50 µL0.08 mL±0.8%<0.2%
Influence of temperature fluctuations on ISE measurementn/an/a<2%
Table 7. Components of the uncertainty budget of the MIP-OES method [28,29].
Table 7. Components of the uncertainty budget of the MIP-OES method [28,29].
Procedure Stage and CommentsVolume (V)Precision (P)Accuracy (A)
Pipette 50 µL twice into Eppendorf tubes; variable-capacity pipette 10–100 µL, precision and accuracy for 50 µL0.05 mL±0.7%<0.3%
Fill an Eppendorf tube to 4 mL (i.e., 3950 µL) with deionized water; variable-capacity pipette 500–5000 µL, precision and accuracy for 2500 µL3.95 mL±0.6%<0.15%
After mineralization and reaching the room temperature of the solution, pour the thimble contents quantitatively into 10 mL volumetric flasks and then wash the thimble with deionized water.10 mL±0.08 mLn/a
Table 8. Results of the B-type expanded uncertainty estimation for the determination of potassium ion concentration using ISE and MIP-OES methods.
Table 8. Results of the B-type expanded uncertainty estimation for the determination of potassium ion concentration using ISE and MIP-OES methods.
Concentration of K+UB for ISE (mg/L)UB for MIP-OES (mg/L)Expanded Uncertainty (UB) 1 by ISE (%)Expanded Uncertainty (UB) by MIP-OES (%)
Low
(2 mg/L)
±0.24±0.051±4.75±2.53
Medium
(10 mg/L)
±0.48±0.253±4.75±2.53
High
(25 mg/L)
±1.19±0.633±4.75±2.53
1 Expanded uncertainty of type B (UB) was defined as U B = k · C · P 2 + A 2 2 , where k = 2, C is the nominal concentration of potassium ions, P is precision, and A is accuracy.
Table 9. Results of the PMI assessment (based on the data from Table 3) and the determination of potassium ion concentration in the VH using ISE and MIP-OES methods in 26 deceased individuals.
Table 9. Results of the PMI assessment (based on the data from Table 3) and the determination of potassium ion concentration in the VH using ISE and MIP-OES methods in 26 deceased individuals.
No.Concentration of K+ by ISE (mmol/L)Concentration of K+ by MIP-OES (mmol/L)PMI (h)
117.015.140
222.444.236
319.231.498
419.919.723
518.621.982
618.724.488
710.516.971
830.827.791
921.335.644
107.5927.046
1110.416.872
1224.221.0107
1323.822.099
1416.623.084
1523.529.470
1612.212.757
1711.021.968
187.917.4240
1937.139.0104
2015.823.367
2116.625.542
2216.024.893
2323.321.769
2428.225.770
2518.718.3117
2625.629.481
Table 10. Summary of the uncertainty budget results of types A (experimental) and B (nonexperimental) for determining potassium ion concentration using ISE and MIP-OES methods.
Table 10. Summary of the uncertainty budget results of types A (experimental) and B (nonexperimental) for determining potassium ion concentration using ISE and MIP-OES methods.
Concentration of K+Expanded Uncertainty (%)
ISEMIP-OES
UAUBUAUB
Low
(2 mg/L)
±12.2±4.75±1.2±2.53
Medium
(10 mg/L)
±6.1±4.75±2.2±2.53
High
(25 mg/L)
±3.1±4.75±2.5±2.53
Table 11. Advantages and disadvantages of ion-selective electrode (ISE) and microwave-induced plasma with optical emission spectrometry (MIP-OES) methods.
Table 11. Advantages and disadvantages of ion-selective electrode (ISE) and microwave-induced plasma with optical emission spectrometry (MIP-OES) methods.
MethodAdvantagesDisadvantages
ISE
-
High selectivity to potassium ions
-
Wide pH range in aqueous solutions (from 2.5 to 11)
-
Wide potassium ion concentration range (from 10−6 mol/L to 1 mol/L or from 0.04 mg/L to 39,000 mg/L)
-
Simple and direct analysis
-
Low costs of equipment and analysis
-
Nondestructive measurements
-
Relatively low volume of samples required
-
Limit of detection (LOD) and limit of quantification (LOQ) are not good as other techniques (e.g., optical emission spectrometry)
-
Problems due to interferences. If the electrode is exposed to high concentrations of interfering ions, drift and long response times may occur. Popular cations such as Cs+, NH4+, Tl+, H+, Ag+, Tris+ (cation of tris (hydroxymethyl) aminomethane), Li+, and Na+ can cause an error of up to 10% for different levels of potassium ion concentration
-
Temperature changes affect the electrode potential. For this reason, both samples and standard solutions should not differ in temperature by more than ±1 °C. At a potassium concentration level of 10−3 mol/L, a temperature difference of 1 °C causes an error of more than 2.5%
-
It cannot be used for a long time due to the deterioration of the membrane (the real-life time of the ISE module is about six months). If the membrane is damaged, the potentiometric response becomes extremely sluggish or the slope of the calibration curve decreases significantly (the ISE module should be replaced)
MIP-OES
-
Despite the selectivity issues, the possibility of performing measurements in several spectral ranges
-
Possibility of performing analysis in a broader range for alternative wavelengths (from 20 mg/L up to 1000 mg/L for 404.414 nm, 344.738 nm, 404.721 nm, and 693.877 nm)
-
Low cost of routine analysis
-
Very simple and automatized analysis
-
Low volume of samples required (1–3 mL)
-
Moderately low influence of the temperature on the analysis result
-
Possibility of performing a wide range of elemental analyses during one measurement
-
Narrow range of analysis (from 0.1 mg/L up to 100 mg/L for regularly used wavelengths: 766.491 nm and 769.897 nm) in comparison to other plasma-based methods
-
Possible occurrence of spectral interferences from other elements (La—766.434 nm and Yb—769.949 for most often used wavelengths)
-
Moderately high costs of apparatus
-
Destructive measurement
-
Limitations connected to the determination of elements with high ionization potentials
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Zięba, S.; Wiergowski, M.; Cieślik, B.M.; Anand, J.S.; Krzyżanowska, M. Uncertainty of Postmortem Time Estimation Based on Potassium Ion Determination in Vitreous Humor Using Potentiometric Ion-Selective Electrode and Microwave-Induced Plasma with Optical Emission Spectrometry Methods. Separations 2023, 10, 201. https://doi.org/10.3390/separations10030201

AMA Style

Zięba S, Wiergowski M, Cieślik BM, Anand JS, Krzyżanowska M. Uncertainty of Postmortem Time Estimation Based on Potassium Ion Determination in Vitreous Humor Using Potentiometric Ion-Selective Electrode and Microwave-Induced Plasma with Optical Emission Spectrometry Methods. Separations. 2023; 10(3):201. https://doi.org/10.3390/separations10030201

Chicago/Turabian Style

Zięba, Sonia, Marek Wiergowski, Bartłomiej Michał Cieślik, Jacek Sein Anand, and Marta Krzyżanowska. 2023. "Uncertainty of Postmortem Time Estimation Based on Potassium Ion Determination in Vitreous Humor Using Potentiometric Ion-Selective Electrode and Microwave-Induced Plasma with Optical Emission Spectrometry Methods" Separations 10, no. 3: 201. https://doi.org/10.3390/separations10030201

APA Style

Zięba, S., Wiergowski, M., Cieślik, B. M., Anand, J. S., & Krzyżanowska, M. (2023). Uncertainty of Postmortem Time Estimation Based on Potassium Ion Determination in Vitreous Humor Using Potentiometric Ion-Selective Electrode and Microwave-Induced Plasma with Optical Emission Spectrometry Methods. Separations, 10(3), 201. https://doi.org/10.3390/separations10030201

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop