1. Introduction
Gene therapy holds promise for treating over 10,000 monogenic disorders worldwide [
1] and is now being explored for use in more complex polygenic disorders, broadening its therapeutic reach [
2,
3]. However, reliable vector quality control (QC) remains a major barrier to clinical translation [
4]. Adeno-associated viruses (AAVs) have emerged as the leading delivery vehicle, with more than 250 ongoing human trials [
5] and 7 FDA-approved treatments already in use, including Luxturna and Zolgensma, with dozens more in the regulatory pipeline [
6]. A critical challenge in AAV manufacturing is the accurate determination of DNA loading status within capsids [
7]. Key metrics—such as capsid and genome titers, and the population fractions of empty (AAV
empty), partially filled with DNA (AAV
partial), or fully loaded capsids (AAV
full)—are difficult to assess with existing technologies [
8]. Accurately determining capsid fill efficiency is essential for precise dosing, as under-filled vectors can reduce therapeutic potency while over-dosed preparations increase the risk of toxicity, immune activation, and other adverse events. In addition to distinguishing genome load, the ability to discriminate between genome forms such as single-stranded DNA (ssDNA) versus self-complementary DNA (scDNA) is crucial because ssDNA vectors are more immunogenic [
9]. Furthermore, ssDNA vectors require second-strand synthesis before expression, whereas scDNA vectors bypass this rate-limiting step and enable faster, more efficient transgene expression [
10].
Several analytical approaches are currently employed to characterize AAVs, each with distinct strengths and limitations. Analytical ultracentrifugation (AUC) is considered the most reliable method for distinguishing empty, partial, and full capsids, as well as detecting aggregates and free nucleic acids [
11]. Quantitative polymerase chain reaction (qPCR) or digital droplet PCR (ddPCR) are used to quantify genome titer [
12], and enzyme-linked immunosorbent assay (ELISA) is used for capsid titer [
13]. Dynamic light scattering (DLS) can provide rapid and reproducible measurements of particle size distribution and aggregation [
14,
15], but it lacks the resolution required to distinguish subtle differences between particles of different DNA load but similar AAV size. Mass spectrometry (MS), including native and charge-detection approaches, enables sensitive analysis of capsid content and offers orthogonal validation of genome composition [
16]. More recently, new technologies have entered the AAV QC field, including the Stunner system (Unchained Labs) that operates based on the thermo-diffusion properties of different nanoparticles [
17], and mass photometry (Samux MP, Refeyn) that offers sensitive discrimination between AAV particles of different mass [
18,
19]. Despite these advances, attaining comprehensive, high-sensitivity, low sample volume QC of AAV particles in heterogeneous solutions remains a challenge. Variability in the ability to estimate the AAV
empty, AAV
partial, and AAV
full population fractions in biomanufactured AAV solutions, which are unavoidably heterogeneous, by current QC assays can result in overdosing, where empty capsids trigger immune responses, or underdosing, which limits therapeutic efficacy despite curative potential.
To improve discrimination capacity between AAV particles with different DNA load status (AAV
empty, AAV
partial, and AAV
full) and DNA load cargo type (AAV
scDNA versus AAV
ssDNA), we have developed a plasmonic nanopore sensor that integrates optical trapping with nanopore resistive-pulse sensing for single-nanoparticle characterization. The device consists of a solid-state nanopore embedded within a plasmonic nanostructure, enabling simultaneous optical and electrical readout of individual nanoparticles, proteins and protein complexes, as reported previously [
20,
21,
22]. In this work, we report further development of this nanosensor technology for discriminating between AAV9 particles with different DNA load status and type, label-free, at single particle resolution. Specifically, by analyzing the concurrently acquired optical-electrical data with machine-learning, this sensor enabled sensitive discrimination between AAV
scDNA and AAV
ssDNA, even though they are designed to have identical mass. This was accomplished by probing the effects of DNA conformation on capsid deformation. Differences in the conformation of the ssDNA versus scDNA genomes can influence the overall shape and rigidity of the AAV9 capsid. These conformational differences also shift the distribution of internal charges, which in turn modify the capsid’s effective dielectric properties and surface charge, producing distinct optical and electrical signatures. Furthermore, the nanosensor measurements could discriminate between AAV
empty, AAV
partial, and AAV
full in solutions containing only AAV
scDNA or AAV
ssDNA, and in an equimolar mixture of AAV
empty, AAV
scDNA and AAV
ssDNA. Additionally, these measurements were performed at 50 μL sample volumes and fM AAV concentrations. These results support the promise of this plasmonic nanosensor technology as a new path forward towards more sensitive QC for gene therapy formulations.
2. Materials and Methods
2.1. Sensor Fabrication
Plasmonic nanopore sensors were fabricated from intrinsic ⟨100⟩ orientation silicon (Si) wafers, 4-inch diameter, 525 ± 25 µm thick (Nova Electronic Materials, Flower Mound, TX, USA). A 500 nm thermal oxide layer (SiO2) was grown using a Tytan series thermal oxidation furnace (Tystar Corporation, Garden Grove, CA, USA). Subsequently, a 60 nm silicon nitride (SixNy) film was deposited by low-pressure chemical vapor deposition (LPCVD) in the Tytan LPCVD system (Tystar Corporation, Garden Grove, CA, USA). For the plasmonic metal layer, wafers were coated with a negative photoresist (NR7-1500PY, Futurrex, Franklin, NJ, USA) and patterned using a backside mask aligner (Model 806 Manual Front/Back Contact Mask Aligner Series 800, OAI, Milpitas, CA, USA). After development in RD-6 (Futurrex, Franklin, NJ, USA) and an oxygen plasma descum (Trion DRIE system, Trion Technology, Clearwater, FL, USA), a 5 nm Chromium (Cr) adhesion layer and 100 nm gold (Au) layer were deposited using an ATC Orion series electron-beam evaporation system (AJA International Inc., Hingham, MA, USA). Lift-off in sonication yielded patterned Au features. For backside patterning, wafers were coated with a positive photoresist (PR1-1000A; Futurrex, Franklin, NJ, USA) and again aligned with the OAI 806 mask aligner. After development in RD-6 and an oxygen descum (Trion DRIE system, Clearwater, FL, USA), the exposed SixNy and SiO2 layers were etched using a CF4 plasma (Trion DRIE system, Clearwater, FL, USA) to open access windows through to the silicon substrate.
The wafers were then loaded into an illuminated wafer holder (ART0042 IL-HAL, Advanced Micromachining Tools, Frankenthal, Germany) designed for 4-inch wafers and immersed in a 40% potassium hydroxide (KOH) bath at 80 °C for ~7 h to anisotropically etch through the intrinsic Si. Once optical transmission through the windows confirmed complete breakthrough, the etch was extended by an additional ~2 h to ensure removal of residual SiO2 and ensure the membrane was released. Wafers were rinsed and cleaned before being separated into individual chips. Cleaving was performed manually with a glass scribe, leveraging cleavage lines defined during the nitride etch. In some cases, these lines were further deepened by KOH etching to facilitate controlled fracture and clean chip separation.
Final nanostructuring was performed at Oak Ridge National Laboratory using a focused ion beam helium ion (He-Ne) microscope (ORION NanoFab Helium Ion Microscope; Carl Zeiss SMT, Oberkochen, Germany). Double nanohole (DNH) antennas were milled using the neon beam, producing apertures ~100 nm in diameter with cusp spacing of ~42 nm. A single nanopore was then milled at the center of each DNH using the helium ion beam, targeting a diameter of ~30 nm (typical diameter achieved ~32 nm). The resulting structure is shown in
Figure 1b.
The DNH geometry was selected to enable self-induced back action (SIBA) trapping for nanoparticles in the 5–50 nm size range, which includes the ∼25 nm AAV9 capsid [
23]. The SIBA effect occurs when a nanoparticle or molecule trapped near a plasmonic nanostructure modifies the local optical field in a way that dynamically enhances and stabilizes its own trapping potential through feedback between the particle’s position and the resonator’s optical response [
24]. The gap in the DNH concentrates the electromagnetic field into a highly confined hotspot, enabling stable SIBA trapping with low optical power. The aperture diameter (~100 nm) and cusp spacing (~42 nm) were chosen based on prior work by Chen et al. [
25], which demonstrated that this DNH geometry supports a plasmonic resonance near 820 nm. This matches the trapping laser wavelength used in our system and therefore maximizes field enhancement and trapping efficiency at the nanopore.
Photomasks used in both lithography steps were designed in-house and fabricated by the Nanotech Institute at the University of Texas at Dallas.
2.2. Optical-Electrical Plasmonic Nanopore Sensor Setup
The experimental platform consisted of a dual-mode optical and electrical detection system centered on a plasmonic nanopore (
Figure 1). Optical trapping was achieved using an MTS Mini Ti:Sapphire laser operated in continuous-wave (CW) mode tuned at 820 nm with ~300 mW output power (KMLabs, Boulder, CO, USA). A 5 W Verdi pump laser at 532 nm provided excitation for the Ti:Sapphire cavity (Coherent, Saxonburg, PA, USA). Residual pump light was removed with an ET665lp barrier long-pass filter (Chroma Technology Corp., Bellows Falls, VT, USA) shown as BF in
Figure 1a.
Along the optical path after the laser source, a 1″ diameter BSF10-B fused silica beam sampler (FBS in
Figure 1a) provided a ~1% diagnostic pick-off to a spectrometer (Ocean Optics, Orlando, FL, USA) shown as Spect in
Figure 1a. A second fused silica sampler directed light to a PDA36A2 amplified photodiode (Thorlabs, Newton, NJ, USA), shown as PD ref in
Figure 1a, for intensity tracking. Polarization was controlled using a WPQ05M-808 zero-order quarter-wave plate (optimized for 780–820 nm), a GTH10M Glan–Thompson polarizer (transmission 350 nm–2.3 µm), and a WPH05M-830 zero-order half-wave plate (optimized for 780–830 nm) (Thorlabs, Newton, NJ, USA) shown as QWP, GTP and HWP in
Figure 1a, respectively. A beam splitter (BS) in
Figure 1a transmitted the trapping beam to the sensing plane while routing back-reflected light to a CS135MUN CMOS camera (Thorlabs, Newton, NJ, USA) shown as CMOS in
Figure 1a, for optical alignment.
The beam path was then raised with a periscope and focused onto the nanosensor using a C-Apochromat 63×/1.2 W Corr water-immersion apochromat objective (Carl Zeiss Microscopy GmbH, Jena, Germany). Alignment to the DNH sensor was guided by fiducial markers patterned during chip fabrication and finalized by adjusting the half-wave plate to maximize transmitted intensity. On the transmission side, light was collected with an RMS10X 10× dry objective, NA 0.25, WD 10.6 mm (Thorlabs, Newton, NJ, USA) and focused onto a second PDA36A2 photodiode (Thorlabs, Newton, NJ, USA) shown as PD trans in
Figure 1a.
Ionic currents were measured using an Axopatch 200B patch-clamp amplifier and digitized with a Digidata 1500B interface (Molecular Devices, San Jose, CA, USA). During all trapping experiments, a +110 mV command voltage was applied across the nanopore.
2.3. Flow Cell Fabrication and Assembly
Flow cells were fabricated using polydimethylsiloxane (PDMS; Sylgard 184, Dow Corning, Midland, MI, USA). PDMS was cast in a custom-machined aluminum mold (Hubs) and allowed to cure overnight at 65 °C. After curing, the PDMS block was demolded, washed with isopropanol, and bonded to a glass slide via corona plasma treatment (BD-20; Electro-Technic Products, Inc., Chicago, IL, USA) for strong adhesion. The assembled flow cell was mounted into a custom 3D-printed caddy to interface with the microscope stage. Nanopore chips were affixed to the PDMS chamber using SecureSeal Adhesive Sheets (SAS-1L, 120 µm thickness; Grace Bio-Labs, Bend, OR, USA). Over the chip, a 170 µm glass coverslip was bonded via SecureSeal SAS-2L (240 µm thickness), ensuring the sensor region lies within the 270 µm working distance of the Zeiss 63× water-immersion objective. A schematic of the flow cell assembly is shown in
Figure 1c. For the experiments, the
trans chamber was filled with 300 mM KCl electrolyte. Before sealing the
cis side with the coverslip, 50 µL of sample was applied directly onto the sensor surface also diluted in 300 mM KCl.
2.4. AAV Constructs and Sample Preparation
A non-therapeutic model of AAV9 capsids was loaded with different forms of the green fluorescent protein (GFP) gene as provided by the University of Texas Southwestern (UTSW) AAV Core. The GFP load serves as an effective analog genome for characterizing AAV loading efficiency and when used in cell culture, the resulting green glow from successful transfections can be used as an indicator of transfection efficiency. Single-stranded (AAV
ssDNA) and self-complementary (AAV
scDNA) vectors were produced separately, each packaging a GFP transgene expression cassette. The cassette fit within the 2.2 kb size limit of AAV
scDNA particles. To generate AAV
ssDNA, an additional 2.2 kb of stuffer DNA was included. The two plasmid constructs used for AAV production differed only in one inverted terminal repeat (ITR, wild type for AAV
ssDNA, mutated for AAV
scDNA) and the presence of the stuffer sequence. Once packaged, both vectors contained a total ssDNA length of ~4.7 kb and so had a near-identical total mass, but different internal DNA load conformations. As a note, AAV9 particles have a diameter of ~25 nm and can be slightly asymmetrical in three dimensions due to their deformability, depending on capsid protein composition [
26]. The DNA load status for all AAV9 solutions was validated against reference standards available at the UTSW AAV Core [
8,
27], namely AUC to separate empty from filled capsids and gel electrophoresis to verify DNA load integrity. The same AAV9 preparations and their detailed characterization by AUC has been reported on in prior work [
28]. The variability of AAV sizes is attributed to the range of DNA load sizes from full, or even over-full, genomes to partial genomes, small fragments, or combinations thereof, as observed, for example, in gel electrophoresis smear patterns [
29].
The samples were provided in 20 µL aliquots of 490 nM, 183 nM and 61 nM of AAVempty, AAVscDNA and AAVssDNA, respectively. All samples were diluted to 10 fM in 300 mM KCl in 100:1 dilution steps prior to measurement and pipetted in the 50 μL PDMS flow cell holding the sensor. For the experiments with heterogeneous solutions, equimolar mixtures were prepared by combining 100 µL of each single-species solution (10 fM), resulting in a final equimolar concentration of 3.33 fM per particle type (AAVempty, AAVscDNA, AAVssDNA).
2.5. Event Detection and Preprocessing
Raw optical and electrical time series were first filtered with a digital 8-pole Bessel filter with a 50 Hz cutoff frequency to reduce high-frequency noise before event tagging. Event identification was guided by the annotated schematic in
Figure 2. Points a and b marked the pre- and post-event optical signal levels and were used to calculate the optical step change in the signal. The time interval between points a and c defined the trapping duration. Point d represented the baseline electrical current prior to optical trapping, while points e and f corresponded to the peaks of the electrical entry and exit spikes, respectively. Point g indicated the post-event electrical baseline. Together, these parameters captured both optical and electrical signatures of each trapping event. To minimize artifacts, events were screened using z-score trimming. The z-score for each event feature was calculated using Equation (1):
In this equation, x is the feature value, μ is the mean, and σ is the standard deviation of the full dataset. Events with ∣z∣ > 5 were discarded as outliers. This z-correction step effectively removed spurious events such as large aggregates, electrical noise spikes, and manual mis-tagging errors introduced during event annotation. We note that this method assumes a unimodal Gaussian distribution, whereas our event features were often multimodal. Consequently, trimming was applied relative to the overall mean and was more effective at excluding large deviations than at rejecting subtle outliers closer to zero amplitude.
2.6. Statistical Analysis and Clustering
Univariate event distributions were modeled with Gaussian mixture models (GMMs) to identify subpopulations [
30,
31]. For multidimensional analysis, event features were min–max normalized across five dimensions: Trapping duration, Transmission optical step change, Reflectance optical step change, Electrical entry spike amplitude, Electrical exit spike amplitude. Spectral clustering was then applied across the five normalized dimensions following established approaches [
32,
33]. To evaluate stability, clustering was repeated multiple times for each candidate cluster number
k, and stability was quantified using the mean pairwise Normalized Mutual Information (NMI), defined in Equation (2). The minimum useful (separable) number of clusters was identified as the
k that maximized the mean NMI (Equation (3)). This stability-based approach to model selection follows principles described by Volkovich and Avros [
34]. The theoretical consistency of spectral clustering under sampling assumptions has also been demonstrated in the variational framework of García Trillos and Slepčev [
35]. Stability curves
(Equation (2)) were plotted as a function of the number of clusters
k (Equation (3)) along with the computed stability uncertainty (±1 SD). Inflection points or plateauing behavior in
with increasing
k were interpreted as diminishing returns in increased stability at the cost of parsing data into a higher number of clusters.
To optimize the clustering, we incremented k stepwise, starting at the minimum useful cluster described above, and tracked how clusters evolved across iterations. Clusters that persisted with high membership overlap across successive k values were deemed stable and retained. Conversely, when increasing k produced new clusters that were roughly even splits of previously stable groups, daughter clusters with similar sizes and near-identical feature distributions, we treated this as evidence of over-clustering (arbitrary subdivision rather than discovery of new structure). In such cases, we stepped back to the last k at which all retained clusters remained stable and well separated.
For visualization, clusters were displayed both in the two most separable original features (transmission and reflectance step changes) and in the eigenspace projection defined by spectral clustering.
4. Discussion
Traditional plasmonic trapping approaches rely almost exclusively on a detected optical step change as a proxy for particle size [
38,
39]. Because the optical step change amplitude is dependent on multiple physical properties of a particle (size, shape, and refractive index—a proxy for density), much of the underlying AAV physical property heterogeneity is effectively lost. As a result, populations that differ meaningfully in structure may collapse onto overlapping measurement values, masking critical distinctions. This work highlights both the strengths and limitations of the plasmonic nanopore approach as a novel way of probing additional data dimensions for characterizing AAV DNA load status and type. As expected, the single-species analytes generally produced Gaussian-like distributions, but each also contained secondary distribution components (
Figure 3). Closer inspection of the time-series data suggested that one of these component Gaussians in each dataset represented complex trapping events, referred to as trap-on-trap events (
Figure 3). Such events were consistently observed across the empty, AAV
scDNA, and AAV
ssDNA datasets, as well as in the equimolar mixture and they are a result of the experimental conditions used in this work.
The relative positions of the dominant population fraction Gaussians in
Figure 3 support the expected size hierarchy of capsid populations. Empty capsids yielded the smallest mean optical step change (4.23%), AAV
scDNA particles were intermediate (4.79%), and AAV
ssDNA particles produced the largest values (5.55%). This ordering is consistent with the physical characteristics of the genomes: The AAV
ssDNA construct is single-stranded and therefore less compact. On the other hand, the self-complementary AAV
scDNA construct, could adopt denser packing. Empty capsids, lacking DNA load, appeared at the smallest size values since they had no negatively charged DNA load pushing the capsid radially outwards. In addition, the lack of DNA load reduced the charge density of the capsid [
40,
41]. Without the negatively charged DNA, empty AAV9 capsids have a higher isoelectric point (pI), around 6.3, whereas for full capsids it is 5.9. At physiological pH (7.4), where experiments were performed, the difference between the pI and the solution pH was therefore smaller for empty relative to full capsids, resulting in lower net negative charge, and a greater tendency to self-interact and aggregate [
42]. Being closer to the pI point was consistent with lower solubility and the more ‘sticky’ behavior observed at the nanosensor. In the equimolar mixture (
Figure 3d) four Gaussian components were observed, corresponding to the three reference populations plus the trap-on-trap events. However, the overlap between AAV
scDNA and AAV
ssDNA distributions underscored the limitation of optical-only sensing when particles have similar mass.
Where our sensor demonstrated its true advantage was in its ability to integrate multiple data dimensions to improve AAV species separability. By combining optical and electrical signals with a reference channel that captured backscattered light, the platform moved beyond one-dimensional descriptors and revealed asymmetries in particle scattering. This effect was particularly apparent in the AAV
scDNA dataset (
Figure 5b), where clustering revealed mirror-image populations that were tagged as +Full and −Full AAV
scDNA. The emergence of these paired populations suggested an orientation-dependent light scattering response: The self-complementary genome likely imposed greater internal rigidity, producing a slightly flattened, or oblate, capsid geometry. The broader and narrower faces of this geometry could interact differently with the plasmonic field, leading to measurable differences in backscattered light. Because optical paths to and from the trap are not symmetric, some scattered light is redirected off-axis and lost, resulting in unequal total intensity between orientations. The difference in population sizes between +Full and −Full clusters therefore reflected a preferred or more stable scattering configuration rather than a defined geometric orientation. The absence of this asymmetry in the AAV
partial or AAV
empty species suggests that it is a unique feature of the fully loaded self-complementary capsid and its associated scDNA rigidity.
Surface charge also influenced how deeply a capsid would settle inside the optical trap: Because the Au layer was not grounded, or held at a defined potential, mirror charges were induced by the applied electric fields, which could repel the more highly charged particles. Furthermore, electrostatic particle repulsion could have been influenced by the trapping laser itself, as the local E-field at the DNH Au cusps could establish a locally focused plasmon of higher electron density contributing to the mirror-charge response of the sensor. However, laser fluctuations and external vibrations, potentially introduced additional drift and noise into the measurements.
Another factor contributing to the observed asymmetries was the geometry of the AAV9 capsid itself, which is a function of its deformability [
43]. Depending on how an individual capsid orients within the trap—oblate versus prolate cross-sections—the optical signatures may shift [
44]. These shape differences alter the light-scattering profiles that are used to deduce the apparent particle volume. We hypothesize that the +Full and −Full populations seen in
Figure 5b were driven by the relative structural rigidity of the scDNA capsid load, whereas no such distinction was made in
Figure 5c for AAV
ssDNA because its capsid was less asymmetric.
The equimolar mixture was intended as a validation test, designed to demonstrate whether the nanosensor could resolve an expected distribution of roughly one-third AAVssDNA, one-third AAVscDNA, and one-third empty capsid population fractions. A slightly higher weighting toward the empty species was expected, because both the AAVssDNA and AAVscDNA solutions would inevitably also contain an empty capsid fraction. However, comparison with AUC data revealed that the equimolar solution data obtained from the nanosensor was far more complex than anticipated, producing a feature-rich dataset that extended beyond the simple tripartite expectation. In practice, a larger number of measured events would be required to improve on statistical uncertainty of population fractions and fully characterize this expanded feature set. The most important outcome from the equimolar experiments was that our sensor could indeed differentiate between populations in a complex mixture, but at the cost of increased ambiguity at interpreting the identity of higher numbers of clusters and a higher incidence of trap-on-trap events, possibly due to the competition between different AAV types to enter the trap. These observations highlight the dual nature of the platform: enhanced dimensionality enables richer resolution but also demands more rigorous controls.
Lastly, a few special cases in the time-series data were noted, which highlighted additional factors that must be considered when interpreting plasmonic nanopore data. Events with very small optical step changes likely represented free proteins, partial capsids, or free-floating DNA. It is possible that this debris transiently adhered to the nanosensor non-specifically [
45], or trapped and un-trapped quickly due to its smaller mass range. In addition, rapid on-off step-changes in optical signals, interpreted as rapid entry–reentry events, may have reflected unstable confinement near the nanopore sensor. Alternatively, as our nanosensor is known to create a traffic jam of analytes around the optical trap [
22], it is possible that multiple particles could line up like a ‘string of pearls’ and sequentially pass through the nanopore. The multi-dimensional feature space makes it easier to distinguish between the former and the latter cases making a ‘string of pearls’ scenario unlikely. Gradual particle entry or exit events from the optical trap likely corresponded to capsids interacting with the nanopore during translocation. It is also possible that particles moved laterally between the two optical hot spots of the DNH cusps [
38]. These positional effects could modulate transmission and backscatter amplitudes, adding variability to weak exit events. The gradual traps were usually not tagged as events due to poor definition of the leading edge of the trap and low electrical response. Weak exit traps were included as we could ensure there was a clearly defined optical step change and trapping current.
Lastly, drift in optical alignment across long experiments and noise introduced by electrical alignment contribute to baseline variability between runs. Although electrical channel data do improve cluster separability, they are not yet reliable as a standalone discriminator due to higher noise. Future methodological refinements, such as introducing controlled flow to fight trap-on-trap analyte settling, improving optical trap stability with automated alignment, coating the Au layer with a hydrophilic coating [
46], using carefully designed Faraday cages [
47], and implementing new protocols for more accurate baseline corrections in optical and electrical data, will help mitigate sources of uncertainty in mixed-population measurements.
5. Conclusions
This work demonstrated that the use of a plasmonic nanopore sensor shows significant promise as a higher-sensitivity QC platform for AAV9 gene delivery vectors. Existing methods provide limited discriminatory capacity when attempting to resolve closely related capsid species, such as AAV
full and AAV
partial, particularly when their masses overlap. With further refinements—such as removal of aggregates from the analysis pipeline and expansion of sample sizes—we anticipate even clearer differentiation between these species. In more realistic scenarios, where partial loads are likely to be appreciably smaller in mass than fully packaged capsids, the separation should be even more pronounced. Comparisons between single-analyte samples and the equimolar mixture revealed a distribution that closely mirrored the expected ratios, validating this approach for use in more complex mixtures. Several technical limitations remain, which were discussed. Looking forward, this approach can be expanded to incorporate alternating current interrogation [
48], which may differentially probe between the inside and outside of a nanoparticle [
49] (DNA load versus the capsid and its surroundings) to further improve separation between AAV species. Such developments would push this nanosensor system closer to an all-in-one impedance-like spectroscopy for nanoparticle quality control, enabling precise, low-volume characterization of viral vectors in both research and manufacturing settings.
While single-particle resolution is a core strength of this platform, its current throughput is modest, typically 100–300 events per hour under the trapping conditions used in this study. Nevertheless, several clear paths exist for scaling throughput to levels more suitable for QC workflows. These include adjusting laser power and command voltage to shorten trapping durations while retaining robust optical–electrical signatures and parallelizing the architecture through arrays of plasmonic nanopores operating simultaneously. Although these engineering developments lie beyond the scope of the present manuscript, they represent a practical roadmap toward higher-throughput implementations compatible with industrial needs.
The optical sensing component of this technology is being considered for potential future applications in protein drug discovery, protein dynamics, and interaction analysis [
50], and has even been tested for egg white protein composition [
51]. The potential of the combined optical electrical nanosensor technology presented in this manuscript has an even broader potential range of possible applications due to the increased discrimination capacity offered by combining information from concurrently acquired multi-modal data. Although this technology has not yet been explored more broadly, it could in principle find use cases to any therapeutic nanoparticle and more broadly to the detection of contaminants such as microplastics among many possible applications.